Midmark Corporation
60 Vista Drive
                              Infection Control in the Dental Practice through Proper Sterilization
P.O. Box 286
Versailles, Ohio 45380-0286
937-526-3662                  In today's world, the need for infection control has never been greater. Dentists, their staff and patients
midmark.com
                              are more concerned about the transmission of infection than ever before. Controlling bacterial
                              contamination through sterilization has long been considered the most essential component in the
                              infection control process and vital to patient safety. The result of proper instrument sterilization is the
                              protection of the patient, physician and staff from various infectious diseases.
                              The sterilization process, however, can be time-consuming, the equipment can be expensive and
                              mistakes can result in the unwanted spread of infection. Proper design of the instrument sterilization
                              process and the processing area can help make sterilization more efficient, minimize environmental
                              contamination, reduce errors, maintain the sterility of instruments and ensure staff and patient safety.
                              This article outlines some basic guidelines to follow for setting up an instrument processing area in a
                              dental practice and an instrument sterilization workflow process. These processes are modeled after
                              the Center for Disease Control and Prevention (CDC) guidelines, and are designed to be used in
                              accordance with general state board guidelines. If discrepancies exist, local state board requirements
                              should take precedence.
                              THE INSTRUMENT PROCESSING AREA
                              The dental instrument processing area in a practice should be centrally located within the office,
                              separated from the operatories, but still easily accessed by staff. To reduce potential introduction of
                              contamination into sterile environments, the area should have a traffic pattern that restricts it to
                              authorized personnel only.
                              The work environments for processing equipment should be separated into four key areas:
                              decontamination, packaging, sterilization and storage. The areas should be designed to separate the
                              sections where contaminated items are received, processed and packaged from the areas where clean
                              items are sterilized and stored. This separation can be achieved through partitions or spatially if the
                              area is well designed and marked.
                              When laying out an instrument processing workspace, it is important to allow for a sink for hand
                              washing, storage cabinets or drawers, adequate lighting and electrical service, floors that are easily
                              cleanable and countertops that are water- and heat-resistant. It is also important to allow for proper
                              ventilation when designing the space not only to maintain a comfortable room temperature, but also to
                              remove any odors or vapors from chemicals or waste. Portable fans should not be used as they can
                              recirculate dust and microbes from the floor and work surfaces, increasing the chance of
                              recontamination.
                              The workspace layout should facilitate a natural workflow that encompasses all of the important steps
                              in the sterilization process. Establishing a smooth workflow from dirty to clean in an instrument
                              processing area can help contain contamination and also enhance the efficiency of the entire
                              sterilization process.
                              The following steps will identify the proper procedure for cleaning and sterilizing instruments and
                              handpieces:
Instrument Sterilization Workflow Chart
                                                                                                                                 Midmark M11 UltraClave             Midmark M9 UltraClave
                                                                                                                                    Automatic Sterilizer               Automatic Sterilizer
                                                                                                       Follow handpiece
                  Midmark Steri-Center                                          Wipe external           manufacturers
                                                                                surfaces and            recommended
                                                                                remove burs.             maintenance
                                                                                                            protocol
                                                            Handpieces
                                  Transport                Sort instruments
   Examination                instruments and              and handpieces.                 Flow of                        Pouch or wrap       Sterilize*** and dry          Store in a dry,
  or procedure.                handpieces to                 Dispose of                  Instruments                       instruments.          instruments.              protected area.
                                Steri-Center.              waste properly*.
                                                            Instruments
                                                                                                           Rinse with
                                                                                    Soak and
                                                                                                         clean** water
                                                                              ultrasonically clean
                                                                                                            and dry
                                                                                  instruments.
                                                                                                         instruments.
    This workflow chart is designed to be used in accordance
    with general state board guidelines. If discrepancies are
    found, you must, in all cases, follow your local state
    dental board requirements.
* 	 Biohazardous waste must be disposed according to your
    states EPA requirements.
** 	 Certain situations may require rinsing with distilled water.
*** Chemical and biological sterilization monitoring must
    follow your state dental board and/or CDC requirements.                             Midmark Soniclean
                                                                                        Ultrasonic Cleaners
Instrument Sterilization Workflow
Step 1  Transport Instruments and Handpieces to Steri-Center
Contaminated instruments and handpieces should be transported to the processing area in a way that minimizes the risk of
exposure to staff, patients and the environment. Once the contaminated instruments are finished being used in the operatory,
they should be transported from the room in a closed, rigid, leak-proof container, and the staff who is transporting the
instruments should be wearing appropriate personal protective equipment.
Step 2  Sort Instruments and Handpieces - Dispose of Waste Properly
Separate the reusable instruments from the disposables and remove the handpieces for sterilization. Disposable instruments
should be properly disposed of and not sterilized for reuse. Sort the reusable instruments based on the type of sterilization
that is going to be required and the materials the instruments are constructed of (e.g. carbon steel instruments should be
sterilized separately from stainless steel to avoid corrosion). Some instruments may require a cold soak prior to sterilization.
If instruments cannot be cleaned immediately, pre-soaking or maintaining them in a moist environment may improve the
cleaning process. Handpieces have a separate cleaning procedure prior to sterilization. Dispose of any excess waste in a
biohazard waste receptacle according to state EPA requirements.
Step 3  Instruments: Soak and Ultrasonically Clean Instruments
To minimize the risk of personnel injury, instruments should be cleaned with a hands-free mechanical process such as an
ultrasonic cleaner or instrument washer prior to sterilization. Ultrasonic cleaning is the preferred process since it is safer and
more effective than manual cleaning and more efficient in penetrating inaccessible areas such as crevices and joints. It is
important to ensure that instruments are rinsed thoroughly and the majority of bioburden is washed off prior to placing them
in an ultrasonic cleaner. Removal of debris may be easier if the instrument has had time to soften while in the holding solution.
Visually inspect the instruments for residual debris, and damage and reclean
or replace any instruments as appropriate. Hand scrubbing of instruments is
not recommended because of the risk of sticks from sharp instruments.
                                                                                     Selecting an Ultrasonic Cleaner 
Personal protective equipment should also be used in the ultrasonic                  Different Options for Different Needs
process.Instruments are placed in an ultrasonic cleaning basket or cassette         Every practice has different cleaning
rack that assures positioning of the instruments at the proper distance from         needs, depending on the number of
the bottom of the tank while keeping them completely immersed in ultrasonic          patients and the types of procedures
solution. Follow the manufacturers recommended instructions to ensure               being performed. There are multiple size
optimum results.                                                                    options available in ultrasonic cleaners,
                                                                                     and dentists can choose the cleaner that
While some microorganisms may be destroyed in the ultrasonic cleaner, this           best meets their offices individual needs.
process should not be considered an appropriate substitute for disinfection          It is also important to evaluate the ease
or sterilization.Additionally, only those chemicals that are specifically labeled   of use and the quality of the cleaner as
as ultrasonic solutions should be used. Detergents, disinfectants and various        well as the noise level. This is important to
liquids not chemically prepared for such use should not be substituted.              note because while ultrasonic cleaning is
Solutions should be changed frequently to assure the continued effectiveness         probably the fastest, most consistent and
of the cleaner.                                                                      safest method of cleaning instruments and
                                                                                     components, most cleaners can be noisy,
Handpieces: Wipe External Surfaces and Remove Burs                                   emitting a high-pitched sound during the
The burs should be removed from each handpiece. They should be properly              cleaning cycle. When selecting an ultrasonic
cleaned and sterilized per the bur manufacturers requirements. The external         cleaner, review the power of the transducers
surface of the handpiece should be wiped to remove any debris.                       and the uniformity of the coverage of
                                                                                     cavitation. More powerful transducers
                                                                                     and more uniform cavitation coverage
                                                                                     will result in a more thorough cleaning of
                                                                                     the instruments.
Step 4  Instruments: Rinse with Clean Water and Dry Instruments
After being ultrasonically cleaned, the instruments should be rinsed in clean water, and in certain situations, in distilled water.
Then, dry the instruments before packaging by either allowing them to air dry, by patting them down or through the use of an
instrument dryer. If required, follow manufacturers recommendations to lubricate or use rust inhibitors that are appropriate for
sterilization.
Handpieces: Follow Handpiece Manufacturers Recommended Maintenance Protocol
Each handpiece manufacture has specific protocols for cleaning their handpieces. Follow their recommended guideline for
cleaning and lubricating the handpiece prior to sterilization.
Step 5  Pouch or Wrap Instruments
In order to prevent instrument recontamination, it is important to package items since placement of unwrapped sterilized
instruments in a contaminated drawer, tray or other receptacle undermines the purpose of infection control.
All instruments to be sterilized should be packaged in pouches or wraps unless they will be used immediately after
sterilization.There are a variety of packaging options available depending on a practices needs. Packaging must be sealed to
maintain sterility and compatible with the type of sterilization used.
Step 6  Sterilize and Dry Instruments
There are a variety of different sterilization methods including steam sterilization (known as autoclaving or moist heat under
pressure), dry-heat sterilization (electric oven) and chemical vapor sterilization. Critical patient care instruments (instruments
that have direct contact with the bloodstream or tissues under the skin) must be sterilized prior to reuse to avoid possible
transmission of infections. If the use of critical patient care instruments is a routine occurrence in the practice, it is important to
have more than one method of sterilization or multiple sterilizers available to use as a backup in the event equipment breaks
down, supplies run low, electricity is unavailable or to avoid disruption in the practice.
Steam sterilization is the most commonly used and recommended method of sterilization. To achieve optimal sterilization
conditions inside the chamber of steam sterilizers, it is necessary to remove the air trapped inside the chamber once the
sterilizer door is closed. Traditionally, steam sterilizers have been classified by the method used to remove this air. Gravity
displacement sterilizers rely on gravity to force the air out of the bottom of the chamber through a mechanical valve at the
beginning of the heating phase. Once specific conditions (temperature, pressure, etc.) inside the chamber are reached, this
valve closes and remains closed throughout the remainder of the cycle. Thus there is the potential for small amounts of air
to remain trapped in the chamber after the valve closes. This trapped air can result in incomplete sterilization under certain
conditions.
Another traditional steam sterilizer type is the pre- and post-vacuum sterilizer. These sterilizers use a vacuum pump to remove
air from the sterilizer chamber prior to and during the heating phase. Because this method uses a pump to draw the air out
of the chamber and its contents instead of relying on gravity to cause the air to be moved to the bottom of the chamber and
escape through an open valve, it can remove more air from the sterilizer chamber than gravity displace sterilizers. In this type
of steam sterilizer, the vacuum pump is also usually used at the end of the sterilization cycle to remove moisture from the
sterilizer contents to shorten the drying time. Pre- and post-vacuum sterilizers tend to be significantly more expensive and
also more costly to maintain because of the added components and complexity.
The third and newest type of steam sterilizer is the Steam-Flush Pressure-Pulse model. This sterilizer uses an electronically
operated valve in place of the mechanical valve used on gravity displacement sterilizers to remove the air from the sterilizer
chamber and its contents. Since the valve is electronically controlled it can be opened numerous times as required to vent
the air from the chamber, thus maximizing air removal. This type of steam sterilizer offers the low maintenance and cost of the
gravity displacement sterilizers with the more effective air removal of the prevacuum sterilizers. This allows some sterilizers
of this type to pass air removal tests that are only required by prevacuum sterilizers (an example would be the Midmark M11
UltraClave Automatic Sterilizer), which removes more air from the chamber vs. gravity displacement sterilizers.
Whatever type of sterilizer is being used, follow the sterilizer manufacturers instructions for specific instrument packaging,
sterilizer operation and maintenance to ensure reliability of the equipment and correct operating procedures.
Chemical Indicators should be used in every cycle to validate that the sterilizer reached the appropriate cycle conditions
necessary to achieve sterilization. Biological Indicators should be used in the sterilizer at least once a week to validate that the
sterilizer is killing all microorganisms.
When loading the sterilizer, care must be taken not to pierce the instrument packaging with instruments, marking devices
or other sharp objects. Correct loading of the sterilizer is required to achieve consistent, effective sterilization. Consult the
sterilizer operators manual for directions on maximum load limits and proper placement of items to be sterilized.
A normal sterilization cycle includes four phases: a heating phase (the period of time when the sterilizer is heating up to
achieve appropriate sterilization temperature); an exposure phase (the actual time required for sterilization of the load); a
venting phase (the period of time when the chamber vents the steam and depressurizes); and a drying phase (the period
of time when the sterilizer is drying the instruments). The instruments should not be removed from the sterilizer until the full
operating cycle is complete and the instruments and/or packaging is dry to prevent compromising packaging and
re-contaminating the instruments.
  What to Look For In a Sterilizer:
  Reliability. Efficiency. Ease of Use.
  No matter what the needs of a practice are, it is important to use sterilization that is reliable and efficient. So what should
  dentists look for when selecting the best sterilizer for them? Its important to do research and find a sterilizer that offers
  fast and efficient instrument processing for reliable and effective quality infection control. Operation should be intuitive
  with easy-to-use controls that help the user easily select the cycle needed. Controls should be programmable to allow
  for the creation of different cycle parameters based on the practices sterilization needs. Fully automatic sterilizers are
  recommended for efficiency in the practice. Once the sterilizer is set, the operator can return to other duties in the office
  without having to monitor the sterilizer. Sterilizers should also offer easy draining and filling without moving the equipment
  to access a reservoir fill port or drain tube. A sterilizer should properly dry the instruments and/or packaging at the end of
  the cycle. Dentists should work with their local dealer to determine which sterilizer is right for their practice.
Step 7  Store in a Dry, Protected Area
Sterilized instruments should be stored in a clean, dry and protected place that has minimal airflow. Instruments should not
be stored unpackaged as this will cause them to become contaminated from hands and airborne microorganisms when
doors or drawers are opened. Packaging should be placed on clean shelves or in clean drawers. Instrument packages should
be rotated on a first in, first out basis. To minimize the possibility of contamination, instruments should remain packaged
until they are required for a procedure and the packaging should be inspected prior to use of the instruments to assure the
packaging is in tact. If the sterile barrier (packaging) has been punctured or gotten wet, the instruments should be re-sterilized
before using them on a patient.