Caliper Calibration
Purpose
The purpose of this procedure is to provide general instructions for the calibration
of calipers including dial, vernier, digital, or a combination thereof.
Scope
All calipers used to measure, gage, test, inspect and control part compliance to
customer specifications and drawings.
Procedure
Preliminary Instructions and Notes
   1. Read this entire procedure before beginning the calibration.
   2. Calibration shall be performed in an environment that conforms to
      Manufacturer Specifications.
   3. The caliper will hereafter be referred to as the Instrument Under Test (IUT).
   4. Verify that the IUT is clean.
   5. Visually examine the IUT for any condition that could cause errors in the
      calibration.
   6. If any of the requirements cannot be met, refer to the applicable
      manufacturer manual.
   7. If a malfunction occurs or a defect is observed while calibration is in
      progress, the calibration shall be discontinued and necessary corrective
      action taken; if corrective action affects a measurement function
      previously calibrated, the function shall be recalibrated before the
      remainder of the procedure implemented.
   8. Use extreme care when removing dirt, fingerprints and other foreign
      substances so that cleaning solutions do not enter internal portions of the
      IUT.
Reference Material
Applicable Manufacturers Manual or Brochures.
Specifications
The specifications of the IUT are determined by the applicable manufacturer's
documentation. If the manufacturer's documentation is not available, then the
specifications identified in this procedure are used.
Equipment Required
The Standards listed below should be selected on the basis of their higher
accuracy level when compared to the unit under test. Equivalent Standards must
be equal to or better than the Minimum-Use-Specification.
Minimum-Use-Specifications for Standards listed are 1/4 the accuracy required by
the IUT.
      Cleaning solution
      Hard arkansas stone
      Lint free cloth
      Gage block set
      Fine wire or nylon brush
      Gage oil
Detailed Procedure
   1. Carefully examine all gaging surfaces for nicks, burrs, scratches, wear or
      other signs of mishandling that may affect accuracy of the instrument.
      Remove any positive surface imperfections utilizing a hard arkansas stone.
   2. Apply sufficient amounts of cleaning solution to dampen a lint free cloth.
      Carefully clean all exposed surfaces removing all foreign substances and
      particulate matter. (Use brush on pinion track if necessary.) Lightly apply
      gage oil to cleaned surfaces.
   3. Check O.D. jaws when in the closed position. Hold calipers to light source
      and visually examine for light between the jaws. Light will not be seen if
      jaws are parallel.
   4. Check I.D. knife-edge jaws for wear when in the closed position. Caliper
      must be positioned squarely in line of sight to detect air gaps or surface
      imperfections between jaws.
   5. Verify zero setting of caliper by observing the indicated reading with jaws
      in the closed position. If caliper does not read 0.000, dial calipers can be
      easily adjusted to zero by loosening the bezel lock screw, then rotating
      bezel to align the zero graduation to the pointer. Some vernier calipers have
      sliding scales that may be adjusted.
   6. Check the step-measuring feature (if model is of this type) using gage
      blocks. Note dial readings when checked at zero and at the measured value
      of the gage block(s).
   7. Check depth measuring feature (if model is of this type) using gage blocks.
      Note readings when checked at zero and at the measured value of the gage
      block(s).
   8. If any defects have been found at this point that would affect accuracy or
      function of the gage, discontinue calibration and refer to section 3.6 for
      remedial action.
   9. Check four quadrants equally spaced throughout the entire measurement
      range of the caliper. e.g., 1.000", 3.000", 4.000", and 6.000".
          a. Align the beam of the caliper to be as nearly parallel to the line of
             measurement as possible.
          b. Tighten clamp screw on the movable jaw without disturbing the
             "feel" between the caliper and gage block(s) to ensure precision
             measurements. The clamp screw must be tightened due to the "play"
             within the sliding member on the movable jaw.
          c. Read caliper (if possible) without disturbing placement. If not,
             carefully remove caliper to obtain reading.
          d. Repeat the measurement steps a sufficient number of times to rule
             out any obviously incorrect readings.
   10. Record all above readings on the Equipment Calibration Record.
   11. Label each instrument with calibration date, due date and person
       performing calibration.
Gage Correction
Any gage exceeding specified tolerances at any time during calibration shall be
repaired and recalibrated, returned as is for restricted use, or scrapped.
All repair work shall be performed by a qualified individual.
Reason for Reissue
Revision A - First Release