This document provides a procedure for calibrating angle and bevel protectors in a calibration laboratory. It outlines the purpose, scope, references, definitions, responsibilities, calibration conditions, general guidelines, calibration procedure, required records, and attachments for the calibration process. The procedure specifies placing the unit under test and reference standards in a controlled environment, inspecting the unit, taking measurements at various angles using gauges, comparing results to specifications, and recording measurements on checklists and certificates.

Uploaded by

Manivasagan Vasu
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as DOC, PDF, TXT or read online on Scribd
100% found this document useful (2 votes)
832 views5 pages

Angle - Bevel Protector

This document provides a procedure for calibrating angle and bevel protectors in a calibration laboratory. It outlines the purpose, scope, references, definitions, responsibilities, calibration conditions, general guidelines, calibration procedure, required records, and attachments for the calibration process. The procedure specifies placing the unit under test and reference standards in a controlled environment, inspecting the unit, taking measurements at various angles using gauges, comparing results to specifications, and recording measurements on checklists and certificates.

Uploaded by

Manivasagan Vasu
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as DOC, PDF, TXT or read online on Scribd
You are on page 1/ 5

CALIBRATION

LABORATORY

ANGLE / BEVEL PROTECTOR

REVISION RECORD SHEET


NOTES

i) Revisions after Rev. 0 are denoted by a vertical line in the left hand margin against
the revised text, with the revision number displayed next to the revision line as shown
on the left.

ii) Revision numbers are A, B, C, etc. up to the issue for management approval and
implementation, which is Rev. 0. Subsequent revisions shall be 1, 2, 3, etc.

No. Date Description of Revision Prepared Review Approved

Issued for approval &


0 01-Jan-2019 MS JR JC
implementation
TITLE : PROTECTOR CALIBRATION PROCEDURE

Table of Contents

S/No. Description Page No.

1.0 Purpose 4

2.0 Scope 4

3.0 Reference 4

4.0 Definitions 4

5.0 Responsibilities 4

6.0 Calibration Conditions 4

7.0 General Guidelines 5

8.0 Calibration Procedure 6

9.0 Records 6

10.0 Attachments 6

Document No. Rev. No. Issue No. Date of Issue Page No.
0 1 01-Jan-2019 2 of 6
TITLE : PROTECTOR CALIBRATION PROCEDURE

1.0 PURPOSE
1.1 The purpose of this work instruction is to ensure that standard practices / methods of
Calibration are followed to obtain reproducible and comparable results and this
document shall be revised as and when required to meet specific requirements of the
project to become a project specific work instruction.

2.0 SCOPE
2.1 This work instruction is applicable for Calibration Laboratory to carryout calibration of
Angle Protector at Lab. temperature as per applicable international
standards/methods.

3.0 REFERENCE
3.1 Documents used:
 BS 1685 - Bevel Protractors (Mechanical & optical) - Requirements and
Test Methods
3.2 Angle Gauge
3.3 Protractor Digital
3.4 Surface Plate

4.0 DEFINITIONS
4.1 UUT: Unit Under Test
4.2 BS: British Standard

5.0 RESPONSIBILITIES
5.1 All the Operators/Technicians perfoming the Calibration shall be trained by.

6.0 CALIBRATION CONDITIONS


6.1 Lab. Temperature : (21±3) °C
6.2 Relative Humidity : (<70) % RH

7.0 GENERAL GUIDELINES

Document No. Rev. No. Issue No. Date of Issue Page No.
0 1 01-Jan-2019 3 of 6
TITLE : PROTECTOR CALIBRATION PROCEDURE

7.1 The calibration should be performed in the controlled environment only. Both the
UUT and Reference Standard equipment should be placed in controlled
environmental condition as mentioned above for at least 30 min warm up.
7.2 UUT should be inspected visually before starting the calibration. Visual inspection is
done for
 Any physical damage.
 Proper Display (in Digital Bevel Protractor)
 Proper functioning of push buttons (in Digital Bevel Protractor)
 Clarity in Magnifying Glass (For Vernier type bevel protractor)
 Rusting in the blades.
 If there is any rusting, it should be removed properly in such a fashion that
flatness of the surface should not be affected.
7.3 If it is not satisfactory, it is to be recorded in the Data Sheet.

Bevel Protractor – Vernier type


Angle Protractor

Digital Bevel Protractor Digital Angle Protractor

8.0 CALIBRATION PROCEDURE


8.1 Wipe clean the blade properly.
8.2 Wipe clean the surface plate properly.
Document No. Rev. No. Issue No. Date of Issue Page No.
0 1 01-Jan-2019 4 of 6
TITLE : PROTECTOR CALIBRATION PROCEDURE

8.3 For Bevel protractors, place the angle gauge between the measuring face of the
blade and the base of the protractor and note down the reading of the protractor in
datasheet.
8.4 For Digital Angle Protractor place the angle gauge on the surface plate and put the
measuring face of the protractor on the angle gauge to get the reading.
8.5 For Mechanical type Angle Protractor place the angle gauge between the measuring
faces of the protractor and note the reading.
8.6 Similarly take the reading for other sizes of the angle gauge.
8.7 Take the reading one by one in such a fashion that it covers the whole range of the
UUT.
8.8 Compare the maximum error found with the instrument accuracy as per BS 1685 (for
Bevel Protractor) and as per manufacturer’s specification (for angle protractor) and
the accuracy statement is made accordingly in the calibration certificate.

9.0 RECORDS
9.1 Measurement/reading shall be recorded in the appropriate Check List/Formats given
the Attachment.
9.2 The test report should contain the following information as minimum.
 Client/Customer Name
 Instrument Manufacturer, Make, Model, Range, Serial No & Accuracy
 Test results

10.0 ATTACHMENT

S.No Format Format No. Description

1 Certificate Angle / Bevel Protector

2 Check List Angle / Bevel Protector