NX7500 Multi-parameter                                                                                          Applications
• Wastewater
scanning UV probe
One step ahead of rising nitrite concentrations
Hach®’s NX7500 scanning UV probe provides you with a high level of confidence in your process by
alerting you to high nitrite concentrations before your process is affected.
Improved control of your process                                     Low maintenance
Continuous monitoring of nitrite gives you an accurate and           The measuring method does not require any reagents or
complete picture of your process, allowing you to make rapid         sample preparation. With the probe submerged in the water,
and data-driven decisions. With short measurement intervals          the nano-coated measuring windows are automatically
along a full UV spectrum, you’ll get more results faster - keeping   cleaned to prevent fouling of the optical path that can diminish
you one step ahead of rising nitrite concentrations negatively       accuracy.
impacting your process.
Easy operation and expert support
Thanks to factory pre-calibration, you will gain a full UV
spectrum view of your water quality from day one with easy
setup. Plus, Hach experts are standing by to support you any
time.
2	                                                                                   NX7500 Multi-parameter scanning UV probe
Technical Data*
Parameter                                                      Nitrate NO₃-N, Nitrite NO2-N, CODeq, BODeq, DOCeq, TOCeq
Application                                                                  Wastewater inlet/aeration and outlet
Path length                                                                          0.3/1/2/5/10/50 mm
Parameter                                                   NO₃-N                NO2-N              CODeq, BODeq               DOCeq, TOCeq
                           Path length 0.3 mm           1 - 330 mg/L        1.7 - 500 mg/L          100 - 7300 mg/L            17 - 3300 mg/L
                           Path length 1 mm            0.3 - 100 mg/L       0.5 - 150 mg/L          30 - 2200 mg/L              5 - 1000 mg/L
                           Path length 2 mm            0.15 - 50 mg/L       0.25 - 75 mg/L          15 - 1100 mg/L             2.5 - 500 mg/L
Measuring range
                           Path length 5 mm            0.06 - 20 mg/L        0.1 - 30 mg/L           6 - 440 mg/L               1 - 200 mg/L
                           Path length 10 mm           0.03 - 10 mg/L       0.05 - 15 mg/L           3 - 220 mg/L              0.5 - 100 mg/L
                           Path length 50 mm           0.006 - 2 mg/L        0.01 - 3 mg/L           0.6 - 44 mg/L              0.1 - 20 mg/L
                           Path length 0.3 mm              1 mg/L               1.7 mg/L                100 mg/L                   17 mg/L
                           Path length 1 mm               0.3 mg/L              0.5 mg/L                 30 mg/L                    5 mg/L
                           Path length 2 mm               0.15 mg/L            0.25 mg/L                 15 mg/L                   2.5 mg/L
Detection limit            Path length 5 mm               0.06 mg/L             0.1 mg/L                  6 mg/L                    1 mg/L
                           Path length 10 mm              0.03 mg/L            0.05 mg/L                  3 mg/L                   0.5 mg/L
                           Path length 50 mm             0.006 mg/L            0.01 mg/L                0.6 mg/L                   0.1 mg/L
                                     Detection limit specifications refer to standard reference solutions under lab conditions.
Controller compatibility                                                                 CD300, CD500
Measuring interval                                                                           ≥ 1 min
Measuring principle                                               UV absorption measurement: 200 - 360 nm (0.8 nm/pixel)
Operating temperature range                                                                 2 - 40 °C
Sample temperature                                                                          2 - 40 °C
Storage conditions                                                                         -20 - 80 °C
Pressure range                                                              Max. 3 bar (in the flow cell max. 1 bar)
Flow rate                                                                                  0.1 - 10 m/s
                                                                                      Stainless steel: 3 kg
Weight
                                                                                         Titanium: 2 kg
Diameter                                                                                    48.3 mm
Length                                                     470 mm (0.3/1/2/5/10 mm path length) or 510 mm (50 mm path length)
Material                                                         Sensor enclosure: Stainless steel 1.4571/1.4404 or Titanium
Light source                                                                           Xenon flash lamp
Cable length                                                                    10 m fixed cable with M12 plug
Interface                                                         Ethernet (TCP/IP), RS232 or RS485 (Modbus RTU, ASCII)
Power requirements (Voltage)                                                         12 - 24 V DC (±10%)
Power consumption                                                                             <8W
Certifications                                                                                 CE
                                                                                                                   *Subject to change without notice.
www.hach.com
NX7500 Multi-parameter scanning UV probe	                                                                                                 3
Principle of Operation
The probe uses a xenon flash lamp as a broadband light source. The light goes through the optical path in the medium, and the medium
absorbs some light. The spectrometer records the remaining, spectrally resolved light and determines the intensity at different wavelengths
in a wavelength range.
Dimensions
                                                                                                                          www.hach.com
4	                                                                                      NX7500 Multi-parameter scanning UV probe
Order Information
Probes
LXV529.99.11xxx           NX7500 Multi-parameter scanning UV probe, 0.3 mm path length, stainless steel or titanium
LXV529.99.21xxx           NX7500 Multi-parameter scanning UV probe, 1 mm path length, stainless steel or titanium
LXV529.99.31xxx           NX7500 Multi-parameter scanning UV probe, 2 mm path length, stainless steel or titanium
LXV529.99.41xxx           NX7500 Multi-parameter scanning UV probe, 5 mm path length, stainless steel or titanium
LXV529.99.51xxx           NX7500 Multi-parameter scanning UV probe, 10 mm path length, stainless steel or titanium
LXV529.99.61xxx           NX7500 Multi-parameter scanning UV probe, 50 mm path length, stainless steel or titanium
Software Packages
LXZ529.99.M0001           NX7500 Calibration Software Wastewater Inlet/Aeration, CODeq
LXZ529.99.M0002           NX7500 Calibration Software Wastewater Inlet/Aeration, CODeq, BODeq
LXZ529.99.M0003           NX7500 Calibration Software Wastewater Inlet/Aeration, CODeq, TOCeq
LXZ529.99.M0004           NX7500 Calibration Software Wastewater Inlet/Aeration, NO₃-N
LXZ529.99.M0005           NX7500 Calibration Software Wastewater Inlet/Aeration, NO₃-N, CODeq
LXZ529.99.M0006           NX7500 Calibration Software Wastewater Inlet/Aeration, NO₃-N, CODeq, BODeq
LXZ529.99.M0007           NX7500 Calibration Software Wastewater Inlet/Aeration, NO₃-N, NO2-N
LXZ529.99.M0008           NX7500 Calibration Software Wastewater Inlet/Aeration, NO₃-N, NO2-N, CODeq
LXZ529.99.M0009           NX7500 Calibration Software Wastewater Inlet/Aeration, NO₃-N, NO2-N, CODeq, BODeq
LXZ529.99.M0010           NX7500 Calibration Software Wastewater Inlet/Aeration, NO₃-N, NO2-N, CODeq, TOCeq
LXZ529.99.M0011           NX7500 Calibration Software Wastewater Outlet, CODeq
LXZ529.99.M0012           NX7500 Calibration Software Wastewater Outlet, CODeq, BODeq
LXZ529.99.M0013           NX7500 Calibration Software Wastewater Outlet, CODeq, TOCeq
LXZ529.99.M0014           NX7500 Calibration Software Wastewater Outlet, CODeq, TOCeq, BODeq
LXZ529.99.M0015           NX7500 Calibration Software Wastewater Outlet, CODeq, DOCeq, TOCeq, BODeq
LXZ529.99.M0016           NX7500 Calibration Software Wastewater Outlet, NO₃-N
LXZ529.99.M0017           NX7500 Calibration Software Wastewater Outlet, NO₃-N, CODeq
LXZ529.99.M0018           NX7500 Calibration Software Wastewater Outlet, NO₃-N, CODeq, BODeq
LXZ529.99.M0019           NX7500 Calibration Software Wastewater Outlet, NO₃-N, NO2-N
LXZ529.99.M0020           NX7500 Calibration Software Wastewater Outlet, NO₃-N, NO2-N, CODeq
LXZ529.99.M0021           NX7500 Calibration Software Wastewater Outlet, NO₃-N, NO2-N, CODeq, BODeq
Several options for probe and calibration software combinations are available.
Please ask your local Hach contact to identify the most suitable option for your application.
Accessories
LZX914.99.11110           Stainless steel chain mounting set
LZY232                    Stainless steel chain with lock, 5 m
6860100.99.0002           High Output Airblast, 230 V
LZY499                    Cable Kit for High Output Airblast
LXZ529.99.B0001           Flow cell, panel, NX7500, 0.3 to 10 mm path length
LXZ529.99.0003A           Flow cell ultrasonic, panel, NX7500, 0.3 to 10 mm path length
                                                                                                                                   DOC053.52.35139.Jan18
LXZ529.99.C0001           Extension cable, 8-pole connector, 10 m
LXZ529.99.C0002           Extension cable, 8-pole connector, 25 m
LXZ529.99.0001A           CD300 controller, 2 sensors
LXZ529.99.A002A           CD500 Controller, 4 sensors
www.hach.com