Oscilloscope
Digital
Features:
• Go-No Go and auto setup sequence.
• FFT function.
• Built-in help menu, multi-language and PC software.
• Standard interface : USB, RS-232C, printer port.
• Option : GPIB interface.
Specifications
Display device Colour (320 x 240) 5.7 inches LCD
Display contrast Adjustable
Display system
Waveform display graticule 8 x 10 divisions (8 x 12 div, when menu off)
Display mode Dot, vector, accumulate
Bandwidth 150MHz (-3dB)
Channels 2
Vertical resolution 8-bit
Vertical sensitivity 2mV/div to 5V/div
Vertical accuracy ±3%
Rise time <2.3ns
Input coupling AC, DC, ground
Vertical system
Input impedance 1MΩ ±2% to 22pF
Polarity Positive and negative
Maximum voltage between signal and common at input BNC 300V (DC + AC peak), CATII
Waveform signal process CH1 + CH2, CH1 - CH2, FFT
2mV/div to 50mV/div : ±0.5V
Offset range 100mV/div to 500m/V/div : ±5V
1V/div to 5V/div : ±50V
BW limit 20MHz (-3dB)
1ns/div to 10s/div
Time base range
Roll : 250ms/div to 10s/div
Time base mode Main, window, window zoom, roll, X to Y
Horizontal system
Time base accuracy ±0.01%
Pre-trigger : 20 div maximum
Delay range
Post-trigger : 1000 div
http://www.farnell.com
http://www.newark.com
http://www.cpc.co.uk
Page <1> 23/10/09 V1.1
Oscilloscope
Digital
Specifications
Real-time sample rate 100MS/s maximum on each channel
Equivalent sample rate 25GS/s ET maximum on each channel
Signal Acquisition Record length 125k/CH
System Peak detection 10ns (500ns/div to 10s/div)
Acquisition mode Sample, peak detect, average
Average 2, 4, 8, 16, 32, 64, 128 and 256
Trigger source CH1, CH2, line, ext
Auto level, auto, normal, single, TV, time delay, event delay, edge,
Mode
pulse width
Trigger
Coupling AC, DC, HF, LF, noise reject
DC to 25MHz : approximately 0.35 div or 3.5mV
Sensitivity
25MHz to 150MHz : approximately 1.5 div or 15mV
X-axis input / Y-axis input Channel 1/Channel 2
X-Y mode
Phase shift ±3° at 100kHz
Auto voltage measurement Vpp, Vamp, Vavg, Vrms, Vhi, VIo, Vmax, Vmin
Frequency, period, rise time, fall time, positivie width, negative
Auto time measurement
Cursor and width, duty cycle
measurement
Voltage difference between cursors (∆V)
Cursor measurement Time difference between cursors (∆T)
Frequency different between cursors (1/∆T)
Readout resolution 6 digits
Frequency range AC coupled, 10Hz minimum to rated bandwidth
Frequency counter
All available trigger source except the pulse width and video
Signal source
trigger mode
Range ±15V
DC to 30MHz : 50mV
Sensitivity
30 to 150MHz : 100mV
External trigger
Input impedance 1MΩ ±2% to 22pF
Maximum input 300V (DC + AC peak), CATII
http://www.farnell.com
http://www.newark.com
http://www.cpc.co.uk
Page <2> 02/05/09 V1.1
Oscilloscope
Digital
Specifications
Autoset can adjust vertical (volt/div)
Autoset Horizontal (sec/div)
Control panel Trigger level automatically
function
Save/recall Up to 15 sets of measurement conditions can be save and recalled
Waveform trace save/recall 2 sets of waveform can be saved and recalled
Power source 100V to 240V ac, 48Hz to 63Hz, auto selection
Power cord x 1, operation manual x 1
Accessories
GTP-150A probe (10:1/1:1) x 2
Dimensions (D) 254 x (H) 142 x (W) 310mm
Part Number Table
Description Part Number
Oscilloscope, Digital, Colour, 150MHz 72-7235 EURO
isclaimer This data sheet and its contents (the "Information") belong to the Premier Farnell Group (the "Group") or are licensed to it. No licence is granted for the use of it other than for information purposes
in connection with the products to which it relates. No licence of any intellectual property rights is granted. The Information is subject to change without notice and replaces all data sheets previously supplied.
The Information supplied is believed to be accurate but the Group assumes no responsibility for its accuracy or completeness, any error in or omission from it or for any use made of it. Users of this data
sheet should check for themselves the Information and the suitability of the products for their purpose and not make any assumptions based on information included or omitted. Liability for loss or damage
resulting from any reliance on the Information or use of it (including liability resulting from negligence or where the Group was aware of the possibility of such loss or damage arising) is excluded.
This will not operate to limit or restrict the Group's liability for death or personal injury resulting from its negligence. TENMA is the registered trademark of the Group. © Premier Farnell plc 2009.
http://www.farnell.com
http://www.newark.com
http://www.cpc.co.uk
Page <3> 02/05/09 V1.1