Unicorn lyte
(Na⁺, K⁺, Cl⁻, Ca²⁺, Ph)
Unicorn lyte
• Unicorn lyte 3 (Na⁺, K⁺, Cl⁻)
• Unicorn lyte 5 (Na⁺, K⁺, Cl⁻,
Ca²⁺, Ph)
Unicorn lyte
SPECIFICATIONS :
Method : Multi-Biosensor Technology
Sample Volume : 120-150 µl
Througput : 80 Test/Hour
Sample type : Serum, Urine, Whole blood
Display : 4 line LCD with backlight
Printer : Inbuilt thermal printer
3 RS 232 port : For upgrade software, for barcode, for data
output
Working temperature : 15-30°C
Unicorn lyte
Power requirement : 1 KVA Online UPS with isolated
transformer, Earth voltage ≤ 5 Volt
Reagent pack : 1270 ml
Standard A : 530 ml
Standard B : 210 ml
Flush Solution : 220 ml
Reference solution : 310 ml
Unicorn lyte
The Unicorn lyte analyzer is required to be turn on all the
times.
If you do not use analyzer for a long time, you need to
remove reagent pack and then perform priming with DI
water and drain the all liquid from tubes.
Unicorn lyte
How to use MBM correctly ? Please operate
according to following requirements.
Store the MBM card at 2 – 30⁰C.
If you take the MBM from refrigerator then
restore it to room temperature before install
in instrument. During this period do not tear
the aluminium foil bag.(about 25 minutes).
Unicorn lyte
Installation
Install reagent pack
Make sure that reagent pack
connected properly
Unicorn lyte
Install pump tube
Install pump tube on plunger and ensure it should not be
overstretched or overlapped
Unicorn lyte
Install MBM
MBM
Push the tongue of MBM box Install new MBM in MBM box
and open it
Unicorn lyte
Install MBM
Insert it properly in box Push it gently to close and
lock
Unicorn lyte
Switch on analyzer
After switch on the instrument, analyzer shows READY screen
READY
Not Cal’ V23.1.40
27-Sep-2019
13:07:53
Press NO key to enter
in Menu
Unicorn lyte
1 PRINT 1 CALIBRATION After Reset The
2 SAMPLE/QC 2 PRIMING Electrode Please
close the door
3 OPERATION 3 CHANGE MBM And Press YES
4 MAINTENANCE > 4 CHANGE PACK >
PRESS 3, THE PRESS 2, FOR PRIMING PRESS YES, So
ANALYZER WILL BE AND REPEAT SAME FOR ANALYZER WILL
ENTERED INTO 3 TIMES AND CHECK ALL TAKE ACTIVATION
‘OPERATION’ MENU 4 TUBES SHOLD BE AND CLAIBRATION
FILLED WITH LIQUID. PROCESS
THEN PRESS 3, CHENGE AUTOMATICALLY
MBM DISPLAY WILL
SHOW :
Unicorn lyte
After calibration analyzer will show calibration slope on display
K: 62.2
Na: 61.5
Cl: -67.4
OK
The slope range is 35 -80. After calibration now it is ready for
sampling
Unicorn lyte
RFID Details
After completion Reagent pack and MBM installation check
and take picture of RFID details
Unicorn lyte
Sampling :
After the calibration of the analyzer it is in the status of
“READY”, lift the sample door, the prompt, “Probe in Serum?”
will be displayed, press YES and pump will begin to aspirate.
Unicorn lyte
After beep sound take out the sample and clean probe with
lint-free tissue and close the sample door
Unicorn lyte
After closing the sample door a brief countdown will begin.
“Analyzing…” will be displayed. Upon completion of analysis,
the test results will displayed and printed
Analyzing… ID: 23 Serum
35 K: 2.17 mmol/L
Na: 118.7 mmol/L
ID: 23 Cl: - 79.6 mmol/L
Note : Probe should not touch the bottom surface of sample tube
Unicorn lyte
Cleaning Procedure : (after 300 samples or 30 days
whichever is earlier
Press 4 Press > key
Unicorn lyte
Put EZ cleaner and It will take 10 minutes
Press 5 press YES to complete
Unicorn lyte
CALIBRATION PROCEDURE :
In READY mode press
NO key for menu
Press 4 Press 1 and enter Press 1
service code 666
Unicorn lyte
Input target Press 4, auto
value from Chart Press YES
factors
and press YES,
Then go in
Maintenance →
Auto Factors
Unicorn lyte
Press YES Enter 1569 in
Press YES
service code
Unicorn lyte
Lift probe and aspirate
Press YES Press YES QCB-1. Every level QCB
should be tested three
times. New factors will
be saved automatically
Unicorn lyte
To check new factors
in Manual mode
Unicorn lyte