Is 9000
Is 9000
/nciian Standard
METHOD FOR THE DETERMINATION OF THE
PROOF AND THE COMPARATIVE TRACKING
INDICES OF SOLID INSULATING MATERIALS
(Second Revision )
@ BIS 2007
..~. —. .—_—.._. ~
Solid Electrical Insulating Materials and Insulating Systems Sectional Committee, ETD 02
NATIONAL FOREWORD
This Indian Standard (Second Revision) which is identical with IEC 60112 : 2003 ‘Method for the 4.
determination of the proof and the comparative tracking indices of solid insulating materials’ issued by
the International Electrotechnical Commission (lEC) was adopted by the Bureau of Indian Standards
on the recommendation of the Solid Electrical Insulating Materials and Insulating Systems Sectional
Committee and approval of the Electrotechnical Division Council. .
This standard was originally published in 1964 and subsequently revised in 1975 with considerable
assistance derived from IEC Publication 112 (1 971). This second revision has been undertaken to
harmonize this standard with latest edition of IEC 60112:2003.
The text of IEC Standard has been approved as suitable for publication as an Indian Standard without
deviations. Certain conventions are, however, not identical to those used in Indian Standards.
Attention is particularly drawn to the following:
a) Wherever the words ‘International Standard’ appear referring to this standard, they should
be read as ‘Indian Standard’.
b) Comma (,) has been used as a decimal marker in the International Standards, while in
Indian Standards, the current practice is to use a point (.) as the decimal marker.
In this adopted standard, reference appears to certain International Standards for which Indian
Standards also exist. The corresponding Indian Standards, which are to be substituted in their
respective places, are listed below along with their degree of equivalence for the editions indicated:
Indian Standard
METHOD FOR THE DETERMINATION OF THE
PROOF AND THE COMPARATIVE TRACKING
INDICES OF SOLID INSULATING MATERIALS
1 scope
( Second Revision)
This International standard specifies the method of test for the determination of the proof and
comparative tracking indices of solid insulating materials on pieces taken from parts of
equipment and on plaques of material usirlg alternating vr31tages.
NOTE 1 The proof tracking index is used as an acceptance criterion as well as a means for the quality control of
materials and fabricated parts The comparative tracking index is mainly used for the basic characterization and
comparison of the properties of materials
Test results cannot be used directly for the evaluation of safe creepage distances when
designing electrical apparatus.
NOTE 2 This test discriminates between materials with relatively poor resistance to tracking, and those with
moderate or good resistance, for use in equipment which can be used under moist conditions. More severe tests,
of longer duration are required for the assessment of performance of materials for outdoor use, utilizing higher
voltages and larger test specimens (see the inclined plane test of IEC 60587). Other test methods such as the
Incllned method may rank materials in a different order from the drop test given in this standard
2 Normative references
The following referenced documents are indispensable for the application of this document.
For dated references, only the edition cited applies. For undated references, the latest edition
of the referenced document (including any amendments) applies.
IEC 60589:1977, Methods of test for the determination of ionic impurities in electrical
insulating materials by extraction with liquids
IEC Guide 104:1997, The preparation of safety publications and the use of basic safety
publications and group safety publicafioms
For the purposes of this International Standard, the following definitions apply:
3.1
tracking
progressive formation of conducting paths, which are produced on the surface andlor within
a solid insulating material, due to the combined effects of electric stress and electrolytic
contamination
3.2
tracking failure
failure of insulation due to tracking between conducting parts
NOTE In the present test, tracking is Indicated by operation of an over-current device due to the passage of
a current of at least 0,5 A for at least 2 s across the test surface and/or wkhin the specimen.
3.3
electrical erosion
wearing away of insulating material by the action of electrical discharges
3.4
air arc
arc between the electrodes above the surface of the specimen
3.5
comparative tracking index
CTI
numerical value of the maximum voltage at which five test specimens withstand the test
period for 50 drops without tracking failure and without a persistent flame occurring and
including also a statement relating to the behaviour of the material when tested using
100 drops (see 11 .4)
NOTE 1 The crlfena for CTI may also require a statement concerning the degree of erosion.
NOTE 2 Although a non-persistent flame IS allowed In the test without constituting failure, matertals which
generate no flame at all are preferred unless other factors are considered to be more important, See also Annex A.
3.6
persistent flame
in case of dispute – one which burns for more than 2 s
3.7
praof tracking index
PTI
numerical value of the proof voltage in volts at which five test specimens withstand the test
period for 50 drops without tracking failure and without a persistent flame occurring
NOTE Although a non-persistent flame is allowed In the test wkhout constituting failure, materials which generate
no flame at all are preferred unless other factors are considered to be more Important. See also Annex A
2
IS 2824:2007
IEC 60112:2003
4 Principle
The upper surface of the test specimen is supported in an approximately horizontal plane and
subjected to an electrical stress via two electrodes, The surface between the electrodes is
subjected to a succession of drops of electrolyte either until the over-current device operates,
or until a persistent flame occurs, or until the test period has elapsed.
The individual tests are of short duration (less than 1 h) with up to 50 or 100 drops of about
20 mg of electrolyte falling at 30 s intervals between platinum electrodes, 4 mm apart on the
test specimen surface.
An a,c. voltage between 100 V and 600 V is applied to the electrodes during the test.
During the test, specimens may also erode or soften, thereby allowing the electrodes to
penetrate them. The formation of a hole through the test specimen during a test is to be
reported together with the hole depth (test specimen thickness). Retests may be made using
thicker test specimens, up to a maximum of 10 mm.
NOTE The number of drops needed to cause failure by tracking usually Increases with decreasing applled voltage
and, below a critical value, tracking ceases to occur
5 Test specimen
Any approximately flat surface may be used, provided that the area is sufficient to ensure that
during the test no liquid flows over the edges of the test specimen.
NOTE 1 Flat surfaces of not less than 20 mm x 20 mm are recommended to reduce the probabillfy of electrolyte
loss over the specimen edge although smaller sizes may be used, subject to no electrolyte loss, e g ISO 3167,
15 mm x 15 mm multi-purpose test specimens,
NOTE 2 It IS preferable to use separate test specimens for each test If several tests are to be made on the same
test piece, care should be taken to ensure that the testing points are sufficiently far from each other so that
splashes or fumes from the testing point WIII not contaminate the other areas to be tested
The thickness of the test specimen shall be 3 mm or more. Individual pieces of material may
be stacked to obtain the required thickness of at least 3 mm,
NOTE 3 The values of the CTI obtained on specimens with a thickness below 3 mm may not be comparable with
those obta[ned on thicker specimens because of greater heat transmission to the glass support through thinner test
specjmens For this reason, stacked specimens are allowed
Test specimens shall have nominally smooth and untextured surfaces which are free from
surface imperfections such as scratches, blemishes, impurities, etc, unless otherwise stated
in the product standard. If this is impossible, the results shall be reported together with a
statement describing the surface of the specimen because certain characteristics on the
surface of the specimen could add to the dispersion of the results.
3
lll-_.— —.. _. L.. .— -—-------
IS 2824:2007
IEC 60112:2003
lFor tests on parts of products, where it is impossible to cut a suitable test specimen from a
part of a product, specimens cut from moulded plaques of the same insulating material may
be used. In these cases care should be taken to ensure that both the part and the plaque are
produced by the same fabrication process wherever possible. Where the details of the final
fabrication process are unknown, methods given in ISO 293, ISO 294-1 and ISO 294-3 and
ISO 295 may be appropriate.
NOTE 4 The use of different fabrication conditions/processes may lead to different levels of performance In
the PTI and CTI test.
NOTE 5 Parts moulded using different flow directions may also exhibit different levels of performance in the PTI
and CT I test
In special cases, the test specimen may be ground to obtain a flat surface.
Where the direction of the electrodes relative to any feature of the material is significant,
measurements shall be made in the direction of the feature and orthogonal to it. The direction
giving the lower CTI shall be reported, unless otherwise specified.
Unless otherwise specified, the test specimens shall be conditioned for a minimum of 24 h at
23 “C*5K, with (50* 10) %’. RH.
7 Test apparatus
7.1 Electrodes
Two electrodes of platinum with a minimum purity of 99 % shall be used (see Annex B). The
two electrodes shall have a rectangular cross-section of (5 f 0,1) mm x (2 * 0,1) mm, with
one end chisel-edged with an angle of 30” f 2° (see Figure 1). The sharp edge shall be
removed to produce an approximately flat surface, 0,01 mm to 0,1 mm wide.
NOTE 1 A microscope with a calibrated eyepiece has been found suitable for checking the size of the en”d
surface
NOTE 2 It IS recommended that mechamcal means are used to re-furbish the electrode shape after a test to
ensure that the electrodes maintain the required tolerances, especially with respect to the edges and corners.
At the start of the test, the electrodes shall be symmetrically arranged in a vertical plane, the
total angle between them being 60° * 5° and with opposing electrode faces approximately
vertical on a flat horizontal surface of the test specimen (see Figure 2 ), Their separation
along the surface of the test specimen at the start of the test shall be 4,0 mm t 0,1 mm.
4
IS 2824:2007
IEC 60112:2003
A thin metal rectangular slip gauge shall be used to check the electrode separation. The
electrodes shall move freely and the force exerted by each electrode on the surface of the
test specimen at the start of the test shall be 1,00 N + 0,05 N. The design shall be such that
the force can be expected to remain at the initial level during the test.
One typical type of arrangement for applying the electrodes to the test specimen is shown in
Figure 3. The force shall be verified at appropriate intervals.
Where tests are made solely on those materials where the degree of electrode penetration is
small, the electrode force may be generated by the use of springs. However, gravity should be
used to generate the force on general purpose equipment (see Figure 3).
NOTE 3 W!th most, but not all designs of apparatus, If the electrodes move during a test due to softening or
erosion of the specimen, their tips WIII prescribe an arc and the electrode gap will change. The magnitude and
dlrectlon of the gap change WIII depend on the relatlve positions of the electrode pivots and the electrode/specimen
contact points. The slgnlflcance of these changes will probably be material dependent and has not been
determined. DlfferenGes in design could give rise to differences in inter-apparatus results.
The electrodes shall be supplied with a substantially sinusoidal voltage, variable between
100 V and 600 V at a frequency of 48 Hz to 62 Hz. The voltage measuring device shall
indicate a true r.m. s. value and shall have a maximum error of 1,5 ‘A. The power of the source
shall be not less than 0,6 kVA. An example of a suitable test circuit is shown in Figure 4.
A variable resistor shall be capable of adjusting the current between the short-circuited
electrodes to (1,0 * 0,1) A and the voltage indicated by the voltmeter shall not decrease by
more than 10 YO when this current flows (see Figure 4). The instrument used to measure the
value of the short-circuit current shall have a maximum error of *3 O/O.
The over-current device shall operate when a current with an r.m.s. value of 0,50 A with
a relative tolerance of +10 0!4, has persisted for 2,00 s with a relative tolerance of ~10 Yo.
Solution A:
Dissolve approximately 0,1 ‘7. by mass of analytical reagent grade anhydrous ammonium
chioride (N H4CI), of a purity of not less than 99,8 Y., in de-ionized water, having a conductivity
of not greater than 1 mS/m to give a resistivity of (3,95 * 0,05) Qm at (23 ~ 1) “C,
NOTE 1 Select the quantity of ammonium chlorlde to give a solution in the required range of resistivity
IS 2824:2007
IEC 60112:2003
Solution B:
NOTE 2 Select the quantity of ammonium chloride to give a solution in the required range of resistivity
Solution A is normally used, but where a more aggressive contaminant is required, Solution B
is recommended. To indicate that Solution B was used, the CT I or PTI value shall be followed
by the letter “M”.
The conductivity of the solutions shall be measured with alternating voltage at a frequency in
the range 1 kHz to 2 kHz. The procedure is described in IEC 60589.
Drops of the test solution shall fall on to the specimen surface at intervals of 30 s * 5 s. The
drops shall fall more or less centrally between the electrodes from a height of 35 mm t 5 mm.
The time for 50 drops to fall on to the “specimen shall be (24,5 * 2) min.
The mass of a sequence of 50 drops shall lie between 0,997 g and 1,147 g. The mass of a
sequence of 20 drops shall lie between 0,380 g and 0,480 g.
NOTE 1 The mass of the drops may be determined by weighing with the appropriate laboratory balance
NOTE 2 For Solution A, a length of thin walled stainless steel tubing (e g hypodermic needle tubing), having an
outer diameter of between 0,9 mm and 1,2 mm, dependent upon the dropping system, has been found to be
suitable for the tip of the dropping device. For Soluhon B, tubes having outer diameters over the range 0,9 mm to
3,45 mm have been found to be necessary with the different dropping systems in use.
NOTE 3 The use of a drop detector/counter IS recommended to ascertain whether there are any double drops or
whether drops are missing
A glass plate or plates, having a total thickness of not less than 4 mm and of a suitable size
shall be used to support the test specimen during the test.
NOTE 1 In order to avoid the problem of cleaning the specimen support table, it is recommended that a
disposable glass microscope slide be placed on the specimen support table immediately under the test specimen
NOTE 2 The use of thin metal foil Conductors around the edge of the glass plate to detect electrolyte loss has
been found useful.
The specimen and its immediate electrodes shall be mounted in an essentially draught-free
space in an enclosure.
NOTE To keep the chamber reasonably free of fumes, it may be necessary, for certatn classes of materials, to
have a small air flow across the surface of the test specimen and between the electrodes. An air velocity of the
order ‘of 0,2 m/s before the start of the test and as far as possible during the test has been found suitable The alr
veloclty In other areas of the enclosure may be substantially higher to assist in fume removal. The alr velocity may
be measured with an appropriately scaled hot wire anemometer.
A suitable fume extraction system shall be provided to allow safe venting of the enclosure
after the test.
6
IS 2824:2007
IEC 60112:2003
8.1 General
Where the material is substantially anisotropic, tests shall be made in the direction of the
features and orthogonal to them. Results from the direction giving the lower values shall be
used, unless otherwise specified.
The result of a test where a hole is formed is considered to be valid, irrespective of the test
specimen thickness, but the formation of the hole shall be reported together with the depth of
the hole (the thickness of the test specimen or stack).
8.2 Preparation
After each test, clean the electrodes with an appropriate solvent and then rinse them with de-
ionized water. If necessary, restore their shape and give a final rinse before the next test.
Immediately before the test ensure, if necessary by cooling the electrodes, that their
temperature is sufficiently low so that they have no adverse effect on the specimen
properties.
Ensure freedom from visual contamination and ensure that the solution to be used conforms
to the conductivity requirements either by regular testing, or by measurement immediately
before the test.
NOTE 1 Residues on the dropping device from an earner test will probably contaminate the solution and
evaporation of the solution will increase its concentration – both of which may result in lower than true values In
such cases It may be advisable to clean the outside of the dropping device mechanically and/or with a solvent and
the inside by flushing through with conforming solution before each test. Flushing through some 10 to 20 drops
depending upon the delay between tests will normally remove any non-conforming liquid.
In case of dispute, the cleaning procedures used for the electrodes and dropper tube shall be
agreed between purchaser and supplier.
Place the test specimen, with the test surface uppermost and horizontal on the specimen
support table. Adjust the relative height of the test specimen and electrode mounting
assembly, such that on lowering the electrodes on to the specimen, the correct orientation is
achieved with a separation of 4,0 mm * 0,1 mm. Ensure that the chisel edges make contact
with the surface of the specimen with the required force and over their full width.
NOTE 2 It may be helpful to place a light behind the electrodes when making this check visually
Set the test voltage to the required value which shall be an integer multiple of 25 V, and
adjust the circuit parameters so that the short-circuit current is within the permitted tolerance.
7
IS 2824:2007
IEC 60112:2003
8.3 Test procedure
Start the dropping system so that drops fall on to the test surface and continue the test until
one of the following occurs:
After completion of the test, vent the chamber of noxious fumes and remove the test
specimen.
9 Determination of erosion
When required, specimens which have not failed at the 50 drop point shall be cleaned of any
debris or loosely attached degradation products and placed on the platform of a depth gauge.
The maximum depth of erosion of each specimen shall be measured in millimetres to an
accuracy of 0,1 mm, using a 1,0 mm nominal diameter probe having a hemispherical end. The
result is the maximum of the five measured values.
In the case of tests according to Clause 10, when required the erosion shall be measured on
the specimens which withstood 50 drops at the specified voltage.
In the case of tests according to Clause 11, when required the erosion shall be measured on
the five specimens tested at the maximum 50 drop voltage.
10.1 Procedure
Where, in IEC standards for material or for electrical equipment specifications, or in other
standards, a proof test only is required, 50 drop tests shall be made in accordance with
Clause 8 but at the single voltage specified. The required number of specimens shall
withstand the test period up to at least 25 s after the fiftieth drop has fallen without tracking
failure, and without a persistent flame occurring.
Operation of the over-current device by air arcs does not constitute a tracking failure.
11.1 General
Determination of the comparative tracking index requires the determination of the maximum
voltage at which five specimens withstand the test period for 50 drops without failure and
whether, at a voltage of 25 V lower than the maximum 50 drop figure, the specimen
withstands 100 drops. If this is not the case, the maximum 100 drop withstand voltage has to
be determined.
NOTE 1 The wording of the previous edition of this standard implied that determinations of the maximum 50 drop
withstand voltage had to be made before any 100 drop determinations.
NOTE 2 It is recognized that the cost of testing may be reduced by firstly determining the maximum 100 drop
withstand voltage and therefore this procedure is recommended in this standard,
Using the basic procedure described in Clause 8, set the voltage at a selected level and make
the test until at least 25 s have elapsed after the one hundredth drop has fallen or until
previous failure occurs.
9
IS 2824:2007
IEC 60112:2003
If the behaviour of the material is unknown, it is recommended that the starting voltage
be 350 V.
If the over-current device operated due to the occurrence of an air arc above the test
specimen, the test was invalid. Repeat the test procedure at the same voltage using a new
test specimen or site after cleaning the apparatus and following the procedure in Clause 8.
If the same event occurs, repeat the test at progressively lower and lower voltages until a
valid failure or pass occurs. Report the details of the tests (see 11 .4).
NOTE 1 It may be impossible to determine the CTI of some materials because a valid failure cannot be achieved,
the characteristic behaviour moving directly from withstanding the test period at one voltage to exhibiting air arcs
at the next highest test voltage.
If the over-current device operated due to the passage of an excessive current across the
surface of the test specimen, or if a persistent flame occurred, the specimen failed the test at
that voltage. Repeat the test on a new sitelspecimen using a lower test voltage after cleaning
the apparatus, etc. as described in Clause 8.
If none of the above occurred and at least 25 s elapsed after the one hundredth drop had
fallen without the over-current device operating, the test is valid and the test specimen is
considered to have passed. Repeat the test on new siteslspecimens at progressively higher
and higher voltages until the maximum voltage is established at which no failure occurred
during the test period of up to at least 25 s after the one hundredth drop has fallen in the first
five tests at that voltage. Five separate specimens or five sites on one plaque may be used
for the tests after cleaning the apparatus and following the procedure described in Clause 8.
If a hole appeared through the test specimen, record the result noting both that a hole was
formed and the depth of the hole (the thickness of the test specimen or stack), and then
continue the tests as described above.
NOTE 2 Where a hole is generated during a test, the further tests may be made on thicker specimens (up to a
maximum thickness of 10 mm) to gain additional information after cleaning the apparatus, etc., as described in
Clause 8.
Where the properties of the test specimen are unknown, increases in test voltage at voltages
above 400 V shall be limited to 50 V per test.
Record, as the 100 drop result, the maximum voltage at which five specimens withstood the
100 drop period without failure.
By inference from the 100 drop data, repeat the test procedure at an appropriate test voltage,
using a new site/specimen and determine whether the specimen withstands the test for the
period up to at least 25 s after the fiftieth drop has fallen.
If the over-current device operated due to the occurrence of an air arc above the test
specimen, the test was invalid. Repeat the test procedure al the same voltage using a new
site/test specimen after cleaning the apparatus and following the procedure as described in
Clause 8. If the same event occurs, repeat the test at progressively lower and lower voltages
until a valid failure or pass occurs. Report the details of the tests (see 11 .4).
NOTE 1 It may be impossible to determine the CTI of some materials because a valid failure cannot be achieved,
the characteristic behaviour moving directly from withstanding the test period at one voiiage to exhibiting air arcs
at the next highest test voltage.
10
IS 2824:2007
IEC 60112:2003
If the over-current device operated due to the passage of an excessive current across the
surface of the test specimen, or if a persistent flame occurred, the specimen failed the test at
that voltage. Repeat the test on a new site/specimen using a lower test voltage after cleaning
the apparatus, etc. as described in Clause 8.
If none of the above occurred and at least 25 s elapsed after the fiftieth drop had fallen
without the over-current device operating, the test was valid and the test specimen is
considered to have passed.
If a hole has not formed through the test specimen during the test, repeat the test on new
sites/specimens, at progressively higher and higher voltages until the maximum voltage is
established at which no failure occurred during the test period of up to at least 25 s after the
fiftieth drop has fallen in the first five tests at that voltage. Five specimens or five sites on one
plaque may be used for the tests after cleaning the apparatus and following the procedure
described in Clause 8.
If a hole appeared through the test specimen, record the result noting both that a hole was
formed, and the depth of the hole (the thickness of the test specimen or stack), and then
continue the tests as described above.
NOTE 2 Where a hole IS generated during a test, the further tests may be made on thicker specimens (up to a
maximum thickness of 10 mm) to gain additional information after cleaning the apparatus, etc., as described in
Clause 8.
The result of tests where a hole formed, irrespective of the test specimen thickness, are
considered to be valid, but the formation of the hole shall be reported together with the depth
of the hole (the thickness of the test specimen stack).
Record, as the 50 drop result, the maximum voltage at which five specimens withstood the
50 drop period without failure.
11.4 Report
. CTI the numerical value of the maximum 50 drop voltage, obtained in five consecutive
tests (the numerical value of the highest 100 drop voltage determined in five
consecutive tests, if more than 25 V below the maximum 50 drop figure), when
appropriate followed by the letter “M” to indicate that Solution B was used.
11
IS 2824:2007
IEC 60112:2003
EXAMPLE ‘CTI 175’, ‘CTI 175 M’, or ‘CTI 400(350) M’
● CTI the numerical value of the maximum 50 drop voltage, obtained in five consecutive
tests, (the numerical value of the highest 100 drop voltage determined in five
consecutive tests, if more than 25 V below the maximum 50 drop figure), when
appropriate followed by the letter “M” to indicate that Solution B was used – the
maximum depth of erosion being in millimetres.
If, for some reason (such as extensive flaming) the erosion cannot be measured, this shall be
reported,
Where a hole developed through the specimen, its formation shall be reported together with
an indication of its depth (specimen or stack thickness).
Where the tests were invalid due to air arcs, this shall be reported.
12
IS 2824:2007
IEC 60112:2003
@-----l
ab
60° t 5“
3~ * 2“
N
1
Al
,1 Flattened edge
—
-lk- 0,01 to 0,1
3 I I
Key
1 Platinum electrode 2 Brass extension (optional)
3 Table 4 Tip of dropping device
5 Specimen 6 Glass specimen support
13
IS 2824:2007
IEC 60112:2003
Key
1 Platinum electrode 2 Brass extension (optional)
3 Table 4 TIP of dropping device
5 Specimen 6 Glass specimen support
m
1 Switch
LJ
2 AC source 100 V to 600 V
A 3 Delay over-current device
4 Variable resistor
5 Electrodes
6 Specimen
5
14
IS 2824:2007
IEC 60112:2003
Annex A
(informative)
The method may be used as published but there are several areas where product committees
may wish to exercise their options:
1. Whether the surface of specimens with rough surfaces may be smoothed by machining,
e,g, grinding (Clause 5).
5. Whether any special instructions need to be given concerning the method of cleaning the
apparatus between tests (Clause 8).
6. Where the material is anisotropic, results from the direction giving the lower values are
usually reported unless otherwise specified (8.1).
7. Number of specimens to be used in proof tests: usually five but a different number may
be preferred (10.2).
9. Whether the proof test should include a requirement for a minimum 100 drop test voltage.
10 Whether determination of erosion depth is required, and if so, the limits to be specified
(Clause 9).
11 Whether because of specific needs, the criteria for allowable flaming are” not suitable
for the application in mind. In those cases, alternative test methods should be
developed/used.
15
IS 2824:2007
IEC 60112:2003
Annex B
(informative)
B.1 Platinum electrodes have been selected for determining the comparative and proof
tracking indices because platinum is the most inert material commonly available. It interacts
least with the electrolyte and insulating materials used, allowing the characteristics of the
insulating material under test to become the main determining factor in arriving at the tracking
index
B.2 In order to simulate the hardware and insulating systems used in electrical devices and
to reduce the electrode cost, materials such as copper, brass, stainless steel, gold and silver
are sometimes used instead of platinum for appraising the tracking characteristics of the
particular electrode metal and insulating material combinations. These electrode materials
interact to varying degrees both with the electrolytes used and the insulating materials, and
thereby influence the test results. The results of tests made with alternatives to platinum
electrodes do not qualify as either comparative or proof tracking indices.
16
IS 2824:2007
IEC 60112:2003
Bibliography
IEC 60587:1984, Test methods for evaluating resistance to tracking and erosion of electrica/
insulating materials used under severe ambient conditions
IECITR 62062:2002, Results of the Round Robin series of tests to evaluate proposed
amendments to IEC 60112
17
(Continued trorn second cover)
The technical committee responsible for the preparation of this standard has reviewed the provisions
of the following International Standards referred in this adopted standard and has decided that they
are acceptable for use in conjunction with this standard:
IEC Guide 104:1997 The preparation of safety publications and the use of basic safety
publications and group safety publications
Only the English text of the International Standard has been retained while adopting it as an Indian
Standard, and as such the page numbers given here are not the same as in the IEC Publication.
For the purpose of deciding whether a particular requirement of this standard is complied with, the
final value, observed or calculated, expressing the result of a test or analysis, shall be rounded off in
accordance with IS 2 : 1960 ‘Rules for rounding off numerical values (revised)’. The number of
significant places retained in the rounded off value should be the same as that of the specified value
in this standard.
.. .,.....-—. ,,._ ——....—
—-...._+_
BIS is a statutory institution established under the Bureau of /rrdian Standards Act, 1986 to promote
harmonious development of the activities of standardization, marking and quality certification of goods
and attending to connected matters in the country.
Copyright
BIS has the copyright of all its publications. No part of these publications may be reproduced in any
form without the prior permission in writing of BIS. This does not preclude the free use, in course of
implementing the standard, of necessary details, such as symbols and sizes, type or grade
designations. Enquiries relating to copyright be addressed to the Director (Publications), BIS.
Amendments are issued to standards as the need arises on the basis of comments. Standards are
also reviewed periodically; a standard along with amendments is reaffirmed when such review
indicates that no changes are needed; if the review indicates that changes are needed, it is taken up
for revision. Users of Indian Standards should ascertain that they are in possession of the latest
amendments or edition by referring to the latest issue of ’61S Catalogue’ and ‘Standards: Monthly
Additions’.
This Indian Standard has been developed from Dot: No. ETD 02 (5734).
.—