Manual Tecnico Micros 45
Manual Tecnico Micros 45
P/n: RAA006BEN
HAN 677A
                          Please, take note of the modifications on next pages. Please, cross out the
                          appropriate sections in the technical manual prior to inserting this adden-
                          dum at the beginning of the manual.
                                                                                        Date: 22/4/08
FORM 0860 - rev 1
Micros 45 / abc Vet technical manual                                                                       RAM225AEN
           1. General information
              Technical informations related to maintenance and adjustments of the abc Vet instrument are not specified into
              the Micros 45 technical manual. The following addendum summarizes the necessary informations.
abc Vet is only available in opent tube mode with reagent pack.
In 16 parameters, from software version V2.41, use only the LMGE reagent pack.
2.3. Parameters
              From software version V2.6.1, abc Vet is available in 8 paramètres or 16 paramètres (with LMGE reagent pack) on
              all species.
2/4
Micros 45 / abc Vet technical manual                                                                               RAM225AEN
2.5. Reagents
  There is no quality control option. The smart card «QC» is similar to a species smart card, it contains only the
  correct thresholds for running control blood.
2.7. Calibration
Calibration is carried out in the same way than for Micros 45 wih exclusive use of Minocal.
Thresholds of the LMG curve are fixed. They do not vary according to temperature like on Micros 45.
2.9. Flags
  L1, M2, G1, G2 and G3 flags are not managed by abc Vet.
  AG1 and AG2 for «cat» species allow the user to determine the level of cold agglutinins or platelet aggregates on
  the WBC count (level 1 : AG1, level 2 : AG2). It is recommended to perform a manual count on a blood smear when
  these alarms occur.
There is no suspicion nor reject on WBC count when AG1 and AG2 occur.
  abc Vet is setup in factory with the «dog» species. If the instrument is turned ON when there is no smart card into
  the reader, abc Vet will be setup with the latest species specified. When the latest smart card used is a QC smart
  card, the instrument will be setup with the «dog» species.
  ◆ On software version lower than V2.41, once instrument has been set up with the animal species, it is not nec-
  essary to leave the card in place (to install a memory card for example).
  ◆ From software version V2.41, before each cycle, abc Vet enquires the smart card into the reader and affects the
  corresponding parameters.
                                                                                                                         3/4
Micros 45 / abc Vet technical manual                                                                         RAM225AEN
                     These 2 manuals includes the addendum RAM178A concerning the latest software version
                     V2.61.
The WBC volume adjustment on latex is given with the specific latex kit (vial + technical note).
           5. Technical manual
              The Micros 45 technical manual RAA006AF becomes Micros 45/abc Vet technical manual RAA006BEN.
4/4
MICROS 45 / abc Vet
Technical Manual
      P/n: RAA006BEN
MICROS 45 TECHNICAL MANUAL                                                                                          RAA006BEN
Introduction
1. Revisions
                        The information in this manual is distributed on an «As Is» basis, without warranty. While every precau-
                        tion has been taken in the preparation of this manual, Horiba ABX will not assume any liability to any per-
                        sons or entities with respect to loss or damage caused or alleged to be caused directly or indirectly by
                        not following the instructions contained in this manual or by using the computer software and hardware
                        products described herein in a manner inconsistent with our product labeling.
1.2. Trademarks
                        Other product names mentioned within this publication may be trademarks or registered trademarks of
                        other companies.
                        All rights reserved. No part of this book may be reproduced or transmitted in any form or by any means,
                        electronic, mechanical, photocopying, recording, or otherwise, without the prior written permission of
                        Horiba ABX.
                        Horiba ABX
                        Parc Euromédecine
                        Rue du caducée
                        B.P. 7290
                        34184 MONTPELLIER Cedex 4 - FRANCE
                        Tel: (33) (0)4 67 14 15 16
                        Fax: (33) (0)4 67 14 15 17
2/2
                                                                 SECTION 1
1. INTRODUCTION ............................................................................................................................2
       1.1. Preliminaries ...................................................................................................................2
       1.2. Description of the manual ...............................................................................................2
                                                                 SECTION 2
1. PRELIMINARIES .................................................................................................................................... 2
2. DESCRIPTION OF THE MANUAL.......................................................................................................... 2
1. INTRODUCTION................................................................................................................................... 2
2. HYDRAULIC PARTS .............................................................................................................................. 2
3. PNEUMATIC DIAGRAMS..................................................................................................................... 2
1. INTRODUCTION................................................................................................................................... 2
2. ELECTRIC PRINCIPLES ......................................................................................................................... 2
3. POWER SUPPLY ................................................................................................................................... 2
4. ELECTRONIC PRINCIPLES.................................................................................................................... 2
5. INTERCONNECTION DIAGRAMS........................................................................................................ 2
1. CUSTOMER MAINTENANCE
1.1 Introduction
WARNING ! : Never use solvent or an abrasive cleaning material to clean the system.
                                               Page 4/2
RAA 006 A Ind.F                                                                              10/03/98
                                                    4 - MAINTENANCE PROCEDURES
A - Hydraulic
B - Mechanical
                                         Page 4/3
RAA 006 A Ind.F                                                                    10/03/98
4 - MAINTENANCE PROCEDURES
3. PROCEDURES
                       CONCERNED PARTS :
                       - Hydraulic maintenance and adjustments
                       - Pneumatic maintenance and adjustments
                       - Electrical maintenance and adjustments
                       - Power supply maintenance and adjustments
                       - Electronic maintenance and adjustments
                       - Printer maintenance and setup
CAUTION : Each procedure has to be read entirely before the intervention is begun.
                                              Page 4/4
RAA 006 A Ind.F                                                                              10/03/98
                                                         4 - MAINTENANCE PROCEDURES
Fig.4.1
                                           Page 4/5
RAA 006 A Ind.F                                                                           10/03/98
4 - MAINTENANCE PROCEDURES
Fig.4.2
                                        Page 4/6
RAA 006 A Ind.F                                                                     10/03/98
                                                   4 - MAINTENANCE PROCEDURES
                                       Page 4/7
RAA 006 A Ind.F                                                                     10/03/98
MICROS                              DILUTION BLOCK MAINTENANCE
* CONCERNS
                       - Lyse dispenser O ring replacement
                       - Diluent dispenser O ring replacement
                       - Sampling needle dispenser O ring replacement
* REQUIRED TOOLS
                       - Allen key : 2,5
                                     2
                       - Dynamometric screw driver      A302 : MAG 019 A
                                                 A301 : MAG 020 A
* REQUIRED PRODUCTS
                       - Silicone grease : LAM 004 A
                       - Soft paper
* INTERVENTION TIME
                       - 15min
* FREQUENCY
                       - On request or or on a leak case.
                       - O rings replacement : 1/year
                                              Page 1/5
RAS 084 A Ind.C                                                            06.03.98
MICROS                             DILUTION BLOCK MAINTENANCE
* PROCEDURE
                       - Pull the piston assy in the upper position and clamp the diluent and lyse
                       tubing as described on the diagram 1.
                       - Disconnect the diluent/lyse/sampling tubings from the 3 syringes and the
                       tube on the sampling syringe side.
Diag.1
                                            Page 2/5
RAS 084 A Ind.C                                                                           06.03.98
MICROS                         DILUTION BLOCK MAINTENANCE
Diag.2
                  - Pull out the pistons    ,    and     from the body with their respective o
                  ring still around (Diagram 3).
                  - Replace the lyse and diluent O rings by new ones. Check the cleanliness of
                  the piston and of the syringe bodies. If necessary clean with a soft paper.
                                        Page 3/5
RAS 084 A Ind.C                                                                       06.03.98
MICROS                          DILUTION BLOCK MAINTENANCE
Diag.3
                  - Spread a little amount of silicone grease between two fingers and apply a
                  very thin film of grease on the 2 new sampling syringe O rings. Replace the
                  old ones      .
                                         Page 4/5
RAS 084 A Ind.C                                                                         06.03.98
MICROS            DILUTION BLOCK MAINTENANCE
* REMARKS
                      Page 5/5
RAS 084 A Ind.C                                06.03.98
  MICROS                     VACUUM/WASTE SYRINGE MAINTENANCE
* CONCERNS
                       - O ring replacement.
* REQUIRED TOOLS
                       - Allen key : 2,5
                       - Dynamometric screw driver A302 : MAG 019 A
* REQUIRED PRODUCTS
                       - Silicone grease : LAM 004 A
* INTERVENTION TIME
                       - 10min
* FREQUENCY
                       - O ring replacement :     - Type 1 : 1/year
                                          - Type 2 : 2/year
                                          - Type 3 : 3/year
                                               Page 1/4
RAS 085 A Ind.C                                                       10/03/98
 MICROS                     VACUUM/WASTE SYRINGE MAINTENANCE
* PROCEDURE
                       - Manually pull the syringe piston up in order to freed it from the syringe
                       body.
              NOTE :    The syringe has to be linked to the atmosphere, pressing the valve         , to
                        pull the piston out from the syringe body.
                        Turn the cylindrical gearing     by hand to help the raising of the piston
                        (See Diagram 1).
Diag.1
                                             Page 2/4
RAS 085 A Ind.C                                                                           10/03/98
 MICROS                VACUUM/WASTE SYRINGE MAINTENANCE
                  - Spread a little amount of silicone grease between two fingers and apply a
                  very thin film of grease on a new O ring.
                  - Reinstall in the reverse order. Apply the following torque to the screws   :
                  700 mN/m.
      CAUTION :    Push the piston back inside the syringe body before tightening the o ring
                   fixation screws      . Use the dynamometric screw driver to tighten the
                   screws      to 400 mN/m
                                        Page 3/4
RAS 085 A Ind.C                                                                         10/03/98
 MICROS           VACUUM/WASTE SYRINGE MAINTENANCE
* REMARKS
                          Page 4/4
RAS 085 A Ind.C                                 10/03/98
 MICROS CT                         SAMPLING NEEDLE MAINTENANCE
* CONCERNS
                      - Needle O ring replacement
                      - Sampling needle replacement
                      - Piercing needle replacement
* REQUIRED TOOLS
                      - Allen key : 2,5
                      - Dynamometric screw driver A302 : MAG 019 A
* REQUIRED PRODUCTS
                      - Silicone grease : LAM 004 A
* INTERVENTION TIME
                      - 15min
* FREQUENCY
                      - See frequency chart table for cleaning.
                      - O ring replacement : 1/year
                      - Needle replacement : On request only
                                            Page 1/5
RAS 086 A Ind.C                                                      09/03/98
 MICROS CT                    SAMPLING NEEDLE MAINTENANCE
* PROCEDURE
                  A - Needle or O ring replacement
                  - Disconnect the tube     from the top of the needle       . Manually lift up
                  the sampling needle (Diagram 1).
                  - Unscrew the 2 screws        in order to freed the needle rinsing block
                  from the carriage frame   .
Diag.1
                  - Remove the clip        and then the rinsing block/needle assy from the
                  carriage taking care not to bend the needle.
                                      Page 2/5
RAS 086 A Ind.C                                                                       09/03/98
 MICROS CT                          SAMPLING NEEDLE MAINTENANCE
                       - Lift up the O ring holder       and replace the O rings         by new ones
                       previously greased. Wipe all excess of grease away.
                       - If necessary clean the inner surface of the rinsing block with a little piece of
                       paper.
Diag.2
                                               Page 3/5
RAS 086 A Ind.C                                                                                09/03/98
 MICROS CT                           SAMPLING NEEDLE MAINTENANCE
  Diag.3
                       - Lift the needle   in the upper position.
                       - Loosen the 2 screws      just enough to enable the rinsing block to be lifted
                       up of about 5 mm.
                       - Pull the piercing needle  foreward and replace it by a new one.
                       - Reassemble in the reverse order. Apply the same torque : 700mN/m to
                       tighten the two screws      .
                                               Page 4/5
RAS 086 A Ind.C                                                                              09/03/98
 MICROS CT                     SAMPLING NEEDLE MAINTENANCE
* REMARKS
                  Blockage problems may occur on some instruments during the needle or
                  carriage motions giving some motor error messages.
                  Remove the axle screw of the puley and clean its 2 parts and the
                  washer.Reinstall the puley assy, the rounded edge facing the puley. Tighten
                  the screw with a torque of 400mN/m.
                  Check that the puley turns freely after the tightening. Add a drop of oil (LAM
                  007 A) if necessary.
A service kit (XEA 343 AS) will be available from May 1995 and will include :
                                        Page 5/5
RAS 086 A Ind.C                                                                         09/03/98
 MICROS OT                             SAMPLING NEEDLE MAINTENANCE
* CONCERNS
                       - Needle replacement
                       - O ring replacement
* REQUIRED TOOLS
                       - Allen key : 2,5
                       - Dynamometric screw driver A300 : MAG 013 A
* REQUIRED PRODUCTS
                       - Silicone grease : LAM 004 A
* INTERVENTION TIME
                       - 15 min
* FREQUENCY
                       - Needle replacement : On request only
                       - O ring replacements : 1/year
                                              Page 1/3
RAS 087 A Ind.C                                                       09/03/98
 MICROS OT                            SAMPLING NEEDLE MAINTENANCE
* PROCEDURE
                       - Disconnect the tube     from the top of the needle          . Manually lift up
                       the sampling needle (Diagram 1).
                       - Unscrew the 2 screws        in order to freed the needle rinsing block
                       from the carriage frame. Remove the clip        .
Diag.1
                       - Remove the rinsing block/needle assy from the carriage taking care not to
                       bend the needle.
                       - If necessary clean the inner surface of the rinsing block by means of a little
                       piece of soft paper. Spread a little amount of grease in between the rinsing
                       block     and its support.
                       - Lift up the O ring holder     and replace the O ring          by a new one
                       previously greased. Wipe all excess of grease away.
                                             Page 2/3
RAS 087 A Ind.C                                                                              09/03/98
 MICROS OT                       SAMPLING NEEDLE MAINTENANCE
* REMARKS
                  Blockage problems may occur on some instruments during the needle or
                  carriage motions giving some motor error messages.
                  Remove the axle screw of the puley and clean its 2 parts and the
                  washer.Reinstall the puley assy, the rounded edge facing the puley. Tighten
                  the screw with a torque of 400mN/m.
                  Check that the puley turns freely after the tightening. Add a drop of oil (LAM
                  007 A) if necessary.
A service kit (XEA 343 AS) will be available from May 1995 and will include :
                                        Page 3/3
RAS 087 A Ind.C                                                                         09/03/98
  MICROS                                  CHAMBER MAINTENANCE
* CONCERNS
                       - RBC, WBC/HGB, Mixing chamber cleaning
                       - Aperture O ring replacement
                       - Coaxial O ring replacement
* REQUIRED TOOLS
                       - Allen key : 2,5
                       - Soft paper
                       - Dynamometric screw driver : MAG 013 A
                       - Cutting pliers
* REQUIRED PRODUCTS
                       - Liquid soap
                       - Distilled water
                       - A scalpel
                       - A Micropipette tip
* INTERVENTION TIME
                       - 30 min
* FREQUENCY
                       - RBC, WBC, Mixing chamber cleaning : 2 (type 1 & 2) or 3/year (type 3).
                       - Aperture O ring replacement : 1/year
                       - Electrode O ring replacement : 1/year
                                              Page 1/7
RAS 088 A Ind.C                                                                           09/03/98
 MICROS                            CHAMBER MAINTENANCE
* PROCEDURE
                  1 - RBC chamber cleaning :
                  - Run a drain chamber cycle ("SERVICE" menu, "DRAIN CHAMBERS " Sub
                  menu).
                  - Record the tube positions before dismantling the chambers.
                  - Disconnect the chamber tubes.
                  - Unclip the RBC chamber. Dismantle the electrode loosening the 2 fixation
                  screws    (Diagram 1).
Diag.1
                                       Page 2/7
RAS 088 A Ind.C                                                                    09/03/98
 MICROS                             CHAMBER MAINTENANCE
                  - Clean the groove and replace the O ring , using the clamp as shown on
                  Diag.2.
Diag.2
                                          Page 3/7
RAS 088 A Ind.C                                                                     09/03/98
 MICROS                                 CHAMBER MAINTENANCE
- Remove the fixation screws from the counting head (See diag.3).
Diag.3
      WARNING ! :    Do not manipulate the aperture using hard instruments. Clean the aperture
                     with a piece of soft paper or preferably, in between 2 fingers.
                    - Clean the chamber and the counting head with liquid soap, do not
                    introduce any sharp instruments inside so as to avoid damaging the inside
                    of the chamber and the aperture.
                    - Rinse thoroughly with distilled water
                    - Dry the exterior of the chamber with a soft paper.
                    - Re-assemble the chamber. Replace the O ring          (Diag.3).
      WARNING ! :    Do not apply too much pressure on the counting head or on the electrode
                     fixation screws, as it can break the aperture (tightening torque =
                     100mN/m).
                                            Page 4/7
RAS 088 A Ind.C                                                                              09/03/98
 MICROS                            CHAMBER MAINTENANCE
                  2 - WBC chamber
                  - Run a drain chamber cycle ("SERVICE" menu, "DRAIN CHAMBERS" Sub
                  menu).
                  - Loosen the cover screws of the WBC chamber and remove the cover.
                  - Record the tube positions before dismantling the chamber.
                  - Disconnect the chamber tubes and top.
                  - Unclip the WBC chamber. Unscrew the photometer   fixation screws
                  (Diagram 4).
                  Remove, clean the WBC chamber and replace the O rings as described in
                  "1 - RBC chamber".
Diag.4
                                       Page 5/7
RAS 088 A Ind.C                                                                    09/03/98
 MICROS                            CHAMBER MAINTENANCE
                  3 - Mixing chamber
                  - Run a drain chamber cycle ("Service" menu, "Drain chambers" Sub menu).
                  - Record the tube positions before dismantling the chamber.
                  - Unclip the chamber and clean it with liquid soap.
                  - Rinse thoroughly with distilled water
                  - Dry the exterior of the chamber with a soft paper.
                  - Re install in the reverse order.
                                      Page 6/7
RAS 088 A Ind.C                                                                   09/03/98
 MICROS                            CHAMBER MAINTENANCE
* REMARKS
                  - Once the cleaning done on the chambers, start the instrument and check
                  for leaks.
                  - Check the calibration too.
                                      Page 7/7
RAS 088 A Ind.C                                                                  09/03/98
  MICROS                               LIQUID VALVE MAINTENANCE
* CONCERNS
                       - Liquid valve assy replacement
                       - Valve body replacement
* REQUIRED TOOLS
                       - Allen key
                       - Flat screw driver
* REQUIRED PRODUCTS
                       - Soft paper
* INTERVENTION TIME
                       - 15 min
* FREQUENCY
                       - On request only
                                             Page 1/4
RAS 089 A Ind.C                                                               09/03/98
 MICROS                           LIQUID VALVE MAINTENANCE
* PROCEDURE
                  - Several reasons can unabled the correct operations of the valves :
Diag.1
                                         Page 2/4
RAS 089 A Ind.C                                                                      09/03/98
 MICROS                              LIQUID VALVE MAINTENANCE
Diag.2 & 3
              NOTE :    When leaks occur on valves it is recommended to replace the entire valve
                        assembly by a new one.
                       - If only one valve has been damaged it is possible to dismantle the valve
                       body on its own as shown on the diagram 4 : use a screw driver to
                       disconnect the valve holder clip and remove the body.
Diag.4
                                            Page 3/4
RAS 089 A Ind.C                                                                         09/03/98
 MICROS                           LIQUID VALVE MAINTENANCE
* REMARKS
                    When piercing several times the same tube cap, some pieces of cork may
                    be dragged along towards the mixing chamber, the WBC chamber and
                    then the liquid valve .
                    This may damaged the operation of the valve.
                                        Page 4/4
RAS 089 A Ind.C                                                                   09/03/98
 MICROS                    POWER SUPPLY CHECK AND REPLACEMENT
* CONCERNS
                      - Voltage supply check
                      - Power supply module replacement
                      - Fan operation check
* REQUIRED TOOLS
                      - Allen key : 2,5
                                    2
                      - Flat screw driver
                      - Volmeter
* REQUIRED PRODUCTS
                      - None
* INTERVENTION TIME
                      - 15 min
* FREQUENCY
                      - 1/year
                                            Page 1/5
RAS 090 A Ind.C                                            09/03/98
 MICROS                 POWER SUPPLY CHECK AND REPLACEMENT
* PROCEDURE
                    1 - Supply voltage check
      CAUTION :      The supply voltage check has to be done with the power supply module
                     connected only.
                    - The supply voltage check is done on the mother board (XAA 355 A)
                    according to the following chart table :
* Ground on TP31, TP 30 or TP 29
      WARNING ! :    The two screws on the top of the coaxial cover holding the board on the
                     MICROS frame are the grounding connection of the mother board.
                     It is mandatory to chek the correct tightening of these screws to obtain
                     correct voltage values.
                                          Page 2/5
RAS 090 A Ind.C                                                                     09/03/98
 MICROS                      POWER SUPPLY CHECK AND REPLACEMENT
                                                                                              SUPPLIES
                                                                                             -12V +- 0.5
                                                                                                 +0.6 TP20
                                                                                              24V -0 TP22
                                                                                              5V +0.3
                                                                                                  -0 TP23
                                                                                             12V -+0.5TP21
SUPPLY
+5V
GND
                                                                                                    J26
                                                                                      +24V
                                                                                      +12V
                                                                                      -12V
Diag.1
                NOTE :    If the voltages values are not correct or among the ranges no adjustment
                          can be carried out either on the board or on the power supply module.
                          Replace the power supply module as described below :
                                              Page 3/5
RAS 090 A Ind.C                                                                                              09/03/98
 MICROS                 POWER SUPPLY CHECK AND REPLACEMENT
Diag.2
      WARNING ! :    The power supply module internal fuse is not to be replaced even when
                     this one has blown down.
                    - Replace the power supply module by a new one and reinstall in the
                    reverse order.
                    - Switch on the instrument and check the voltages on the mother board as
                    described in the previous paragraph.
                                        Page 4/5
RAS 090 A Ind.C                                                                    09/03/98
 MICROS                POWER SUPPLY CHECK AND REPLACEMENT
* REMARKS
                  The four leds in front of the test points are lit to indicate a voltage presence
                  but whatever its value!!!
                  Check the operation of the fan as following : When the fan has stopped,
                  move the cursor to the 4 SERVICE menu and press ENTER : the fan should
                  start.
                                        Page 5/5
RAS 090 A Ind.C                                                                         09/03/98
        MICROS OT                            TECHNICIAN FUNCTIONS
* CONCERNS
     * REQUIRED TOOLS
          -   Flat screw driver
          -   thermometer
          -   Barflex
          -   Allen key : 2,5
          -   Voltmeter
     * REQUIRED PRODUCTS
          - WBC latex : LAD 001 AS
          - RBC and PLT latex : LAD 002 AS
          - Soft paper
          - Flat piece of stiff plastic
     * INTERVENTION TIME
          - 60 min
     * FREQUENCY
          - See maintenance chart table.
     * PROCEDURE
           The arborescence of the «technician functions» is the following :
                                                              1 - CAL. PHOTOMETER      1 - WBC ADJUST
                                  1 - VERSION
                                  2 - ADJUSTMENTS
                                                                                       2 - MCV <     >
                                                              7 - VACUUM
                                                                                       1 - VACUUM <          >
                                                                                       2 - CHECK
       8 - TECHNICIAN
                                                              1 - UNCAL. TEMPERATURE
                                  3 - TEMPERATURE             2 - CALIBRATE
                                                                                       1 - VISIBLE
                                                              3 - TEMPERATURE
                                                                                       2 - INVISIBLE
         MICROS OT
                                                              1 - CONTROL
                                  4 - RUN MODE
                                                              2 - USER
                                                              1 - YES
                                  5 - REAGENT PACK
                                                              2 - NO
6 - SERIAL NUMBER
1 - BURNING #
8 - BURNING 2 - BURNING
3 - CONTROL
          Once entering the «SERVICE» sub menu, move to «TECHNICIAN FUNCTIONS» and press                       .
          A specific password (421) is required to enter the sub menus.
          Move the cursor by means of                and           and choose the required menus pressing the
                    key.
II - Adjustments :
             NOTE          - If the WBC chamber has been dismantled previously make sure no liquid has flown
                           in between the spectrophotomer and the chamber.
                           Clean the inner surfaces of the spectrophotometer as well as the chamber.
                           Reassemble the assy and tighten the two screws to the following torque : 400mN/m
                           (see RAS 088 A : Chamber maintenance)
An HGB channel is displayed on the LCD screen representing a voltage value as shown below :
                                                           VALUE
                                                           237
                                 The HGB photometer calibration must be done 20min at least after the
         IMPORTANT               instrument has been switched on.
          - By means of R97 (See diagram 1) adjust the HGB channel according to the room temperature using
          the chart table given on next page.
IMPORTANT This adjustment must be done with the WBC chamber cover installed!!!
                                                                                      HGB
                                                                                 R97        TP27
TP19
                                                                                APERTURE R91
                                                                                 VOLTAGE
                                                                                 60V +0.1
                                                                                     -
Diag.1
2 - Aperture voltage
           - Once entering the «APERTURE VOLTAGE» menu, connect a voltmeter between the ground (TP30)
           and TP19.
3 - Pressure check
                                                       - If the results are not correct check the O ring and the
  Diag.2                                               tubing watertightness.
4 - WBC adjust
- Put the WBC latex to mix on a Vortex during 5min or shake thoroughly
- Wait for several results to be displayed and check the stability of both values.
           After one minute the last volume values displayed on the screen are printed out as well as the WBC,
           PLT and RBC histograms and the CBC “s results. Check that these printed values correspond to the
           following target values :
                   - LYM                 = 57 + 1
                   - GRA                 = 180 + 2
           - If not rerun a «WBC ADJUST» cycle after having previously mixed the latex vial again. By means
           of R74 (see diagram 4) adjust the volumes to the target values during the measuring phasis.
                                                                                                                     THRESHOLD
                                                                            THRESHOLD R75                              350mV+5   APERTURE R91
                                                                             400mV+5         TP13                            -   VOLTAGE
                          TP4                                                     -                                      R74 TP7
                                      GAIN
                                             R87   TP15
                                                                                       R81                          GAIN          60V +0.1
                                                                                                                                      -
               J17                                                                GAIN
                                                                                             TP11
                                                                                                                                                GND    TP29
               NU
                                                                 RBC PULSE PROCESS                  WBC PULSE PROCESS
Diag.4
- Put the RBC and PLT latex to mix on a Vortex during 5min or shake thoroughly
          - Present the vial of Latex to the open probe as shown on diagram 3 and press the sampling bar
          located behind the sampling needle : the needle directly delivers the latex sample in the RBC chamber
          dilution (for a usual analysis cycle the sample is first delivers to the mixing chamber) and a measuring
          phasis begins.
          - During the cycle measuring phasis (around 1 minute) the Platelet and the Red Blood cell volumes
          are displayed on the screen every 3 seconds as shown below :
- Wait for several results to be displayed and check the stability of both values.
          After one minute the last volume values displayed on the screen are printed out as well as the PLT
          and RBC histograms and the CBC “s results.
- Check that these printed values correspond to the following target values :
                  - RBC = 74 + 1
                  - PLT = 59 + 1
          - If not rerun a «RBC PLT ADJUST» cycle after having previously mixed the latex vial again.
          - Adjust the volumes to the target values during the measuring phasis by means of the following
          potentiometers (see Diagram 4) :
6 - Adjust sensors
* Needle sensor
Diag.5
Diag.6
           - For a current number of steps out of ranges, unloosen the 2 screws 2 (diagram 5) and gently move
           the sensor         - upward if the current value is too low
                              - downward if the current value is too high.
           - Tighten the screws 2 and rerun a «NEEDLE SENSOR» cycle. Check that the current value is
           correct.
            This function allows the adjustment of the needle heigth in the RBC chamber. The current number of
            steps is 630 and may be adjusted if reproducibilty on RBC/PLT is incorrect.
            In this case the value has to be raised 10 steps by 10 steps. Reproducibilty on PLT has to be controled
            between each value modification.
            - From the SENSOR ADJUST menu, move the cursor to the function <1> "NEEDLE SENSOR" and
            press ENTER.
            - Push the sampling carriage above the RBC/PLT chamber. Lower the sampling needle until it comes
            into contact with the chamber edge as shown on the Diag.7.
Diag.7
 - Press ENTER and record the number of step displayed on the LCD. Add 284 to the number displayed and
 record the calculated value.
- Move the cursor to the function <2> "NEEDLE" and press ENTER. Enter the calculated value and validate.
* Transfer sensor
Diag.8
           - For a current number of steps out of ranges, unloosen the 2 screws 2 (see diagram 8) and gently
           move the sensor
NOTE 10 steps correspond to around 1 mm. The target number of steps is 533 + 5
           - Tighten the screws 2 (see diagram 8) and rerun a «CARRIAGE» cycle. Check that the current value
           is correct.
7 - Vacuum
            Move to «1 - VACUUM < >» and press ENTER. Raise the value if the drain has to be faster. Lower
            the value if the drain has to be slower. Press ENTER to confirm. Rerun a «CHECK» cycle to control
            the drain of the chamber.
                          Compare to this «CHECK» cycle , the chamber is totally drained during an analysis
               NOTE
                          cycle.
When entering the «temperature adjustment» menu the following sub menus are displayed :
                  1 - Uncal. temperature
            When pressing the           key the sensor temperature value uncalibrated is displayed. This value
            should be equivalent to the diluent temperature + 2°.
                  2 - Calibrate :
            The temperature must be calibrated according to the diluent temperature :
- Plunge a thermometer directly into the diluent container and leave it for a while until stabilization.
- Run 2 diluent primes («SERVICE» menu, «PRIME» sub menu, «DILUENT» selection).
            - Enter the «CALIBRATE» menu. Note the temperature of the diluent and type in the value (if it is
            different from the previous on the instrument).
          3 - Temperature :
                          1 - Visible : Press          to validate the temperature display.
                                        The value displayed is the calibrated temperature.
                          2 - Invisible : Press         to cancel the temperature display.
                                        (Temperature invisible by default)
IV - Run mode
          The instrument must be configurated in the «USER» mode (configurated by default in the «USER «
          mode), the «CONTROL» mode intends for a factory use.
V - Reagent pack
This function is used to update the instrument from a bottle mode to a pack mode.
VI - Serial #
VII - Cycle #
          Displays the :        Startup number since the first use of the instrument.
                                Stand by number since the first use of the instrument.
                                CBC number since the first use of the instrument.
                                 The startup, stand by and CBC numbers are adjustable in this menu but
         IMPORTANT               it is mandatory to keep the initial values (useful for maintenance
                                 schedules).
VIII - Burning
        CAUTION :          This function which allows the burning of the instrument is intended to a factory
                           use.
*REMARKS
* CONCERNS
                      - Sensor replacements
                      - Needle motion check
                      - Carriage motion check
                      - dilution block motion check
                      - Vacuum/waste syringe motion check
                      - Valve operation check
                      - LCD contrast adjustment
                      - Piercing mechanism check (MICROS CT)
* REQUIRED TOOLS
                      - Allen key : 2,5
                      - Felt-pen
* REQUIRED PRODUCTS
                      - None
* INTERVENTION TIME
                      - 30 min
* FREQUENCY
                      - See maintenance chart table.
                                           Page 1/7
RAS 092 A Ind.C                                                    10/03/98
 MICROS                             MECHANIC FUNCTIONS
* PROCEDURE
                  Enter the "4 - SERVICE" menu and move to "6 - MECHANIC" sub menu.
                  The following operations are available :
Diag.1
                                      Page 2/7
RAS 092 A Ind.C                                                                  10/03/98
 MICROS                                 MECHANIC FUNCTIONS
                    - Manually raise the needle support       in the upper position. The "0" should
                    switch to "1". This indicates the correct operation of the needle sensor
                    (diagram 1).
                    If nothing happened try to move the needle up and down again to get the
                    commutation from "0" to "1".
                    If the test is still wrong, check the correct connection of the sensor on the
                    connector J7 of the mother board.
WARNING ! : The new sensor must be installed exactly on the same position.
                    Proceed as described in RAS 091 A (MICROS OT) or RAS 121 A (MICROS CT)
                    "Sensor adjustment" to check the correct position of the sensor.
Diag.2
                                          Page 3/7
RAS 092 A Ind.C                                                                            10/03/98
 MICROS                                 MECHANIC FUNCTIONS
                    If needed to replace, note its position with a felt-pen, unscrew the 2 screws
                        and remove the cell holder    .
                    Disconnect the wire from the connector J8 (See diagram 3)
                    Replace the previous sensor by a new one and reassemble in the reverse
                    order.
WARNING ! : The new sensor must be installed exactly on the same position.
                    Proceed as described in RAS 091 A (MICROS OT) or RAS 121 A (MICROS CT)
                    "Sensor adjustment" to check the correct position of the sensor.
Diag.3
                          - The motor ------> Check the connection on the board (connector J3)
                          - The syringes ------> Check the motion up/down
                          - The sensor ------> Check the connection on the board or replace it.
                                          Page 4/7
RAS 092 A Ind.C                                                                         10/03/98
 MICROS                                   MECHANIC FUNCTIONS
Diag.4
                                            Page 5/7
RAS 092 A Ind.C                                                                       10/03/98
 MICROS                               MECHANIC FUNCTIONS
                        - The motor ------> Check the connection on the board (connector J4)
                        - The syringe ------> Check the motion up/down
                        - The sensor ------> Check the connection on the board or replace it.
Diag.5
                  Run a "6 - VALVES" cycle. The number of the valves from 1 to 13 is displayed
                  and each valve is activated once.
                  Check their correct operations. If a valve is suspected proceed as
                  described in the procedure : RAS 089 A.
                                        Page 6/7
RAS 092 A Ind.C                                                                      10/03/98
 MICROS                                     MECHANIC FUNCTIONS
Enter the sub menu "7 - CHG. CONTRAST". Use the arrows to modify the
                       Enter the sub menu "8 - PIERCING". Follow the instructions : "PLEASE CLOSE
                       TUBE HOLDER DOOR" : a piercing operation is simulated.
                       This function gives the sample tube holder position and the heigth of the
                       needle in its lower position. See RAS 091 A (MICROS OT) or RAS 121 A
                       (MICROS CT)
              NOTE :    If the sample tube holder has been removed the following message is
                        displayed : "ERROR : NO SAMPLE TUBE HOLDER".
                        If the sample tube holder has been turned in between two piercing
                        positions (it means that the tube is not in front of the piercing needle) the
                        following message is displayed : "ERROR : TUBE HOLDER POSITION'".
* REMARKS
                                             Page 7/7
RAS 092 A Ind.C                                                                             10/03/98
 MICROS                               DRAIN DETECTION ADJUSTMENT
* CONCERNS
                        Drain detection sensor adjustment/replacement
* REQUIRED TOOLS
                        - Voltmeter
                        - Flat screw driver
                        - 5ml syringe
* REQUIRED PRODUCTS
                  - None
* INTERVENTION TIME
                        - 15min
* FREQUENCY
                        - On request or 1/year
                                                 Page 1/5
RAS 093 A Ind.D                                                         30/03/00
 MICROS                        DRAIN DETECTION ADJUSTMENT
* PROCEDURE
                  The Vacuum/Waste syringe and the chambers drains are controled by an infrared sensor
                  located below the syringe.
                  When a drain problem occurs during a cycle two error messages are possible :
Diag.1
                                             Page 2/5
RAS 093 A Ind.D                                                                                       30/03/00
 MICROS                              DRAIN DETECTION ADJUSTMENT
                       - Make sure the cell is perfectly drained (no bubble).
                       - Connect a voltmeter between the ground TP30 and TP6 (See Diag.2) and adjust the
                       voltage to 4,5V + 0,3V by means of R6.
Diag.2
                                                   Page 3/5
RAS 093 A Ind.D                                                                                30/03/00
 MICROS                                 DRAIN DETECTION ADJUSTMENT
                           - Fill the syringe up with diluent and connect it on the syringe waste tubing instead of the
                           syringe nipple (see Diag.3).
Diag.3
                           - Push diluent through the sensor and check that the voltage falls down below 1Volt on the
                           voltmeter.
                  NOTE :    Check the commutation from 4,5Volts to ~ 0Volt once again pushing and drawning
                            alternately liquid and air through the sensor by means of the syringe.
                           - If nothing happens switch off the instrument and disconnect the connector from J13.
                           - Replace the drain detection sensor by a new one.
                           - Switch on the instrument
                           - Carry out the new sensor adjustment as described above.
                                                     Page 4/5
RAS 093 A Ind.D                                                                                              30/03/00
 MICROS                       DRAIN DETECTION ADJUSTMENT
* REMARKS
                  However it is posible to control the correct operation of the sensor by means of the LED
                  located next to R6 (Upper one).
                  Indeed the LED should be ON when liquid circulates through the sensor and OFF with air.
                                           Page 5/5
RAS 093 A Ind.D                                                                                 30/03/00
     MICROS                     PCB VOLTAGE CHECKS/ADJUSTMENTS
* CONCERNS
* REQUIRED TOOLS
                      - Voltmeter
                      - Flat screw driver
* REQUIRED PRODUCTS
- None
* INTERVENTION TIME
- 20min
* FREQUENCY
- None
                                              Page 1/4
RAS 094 A Ind.E                                             04/11/1999
     MICROS                     PCB VOLTAGE CHECKS/ADJUSTMENTS
* PROCEDURE
                         1 - RBC, PLT, WBC threshold checks/ adjustments
                         - Ground on TP30.
                         - Adjust the thresholds according to the below chart table :
(See Diag.1)
                               NOTE :
                                         ABC VET : Different threshold for RBC
                                               RBC : 350 mV + 7
                         - Ground on TP31.
                         - Adjust the motor voltages according to the below chart table :
(See Diag.1)
                                                  Page 2/4
RAS 094 A Ind.E                                                                                   04/11/1999
     MICROS       PCB VOLTAGE CHECKS/ADJUSTMENTS
                         Page 3/4
RAS 094 A Ind.E                               04/11/1999
     MICROS              PCB VOLTAGE CHECKS/ADJUSTMENTS
* REMARKS
                  The mother board XAA 355 A has been updated to index C according to the following
                  chart table :
                  The mother board XAA 355 has been updated to the major index B with the stepper
                  motor voltage adjustments.
                                         Page 4/4
RAS 094 A Ind.E                                                                            04/11/1999
        MICROS CT                            TECHNICIAN FUNCTIONS
* CONCERNS
     * REQUIRED TOOLS
          -   Flat screw driver
          -   thermometer
          -   Barflex
          -   Allen key : 2,5
          -   Voltmeter
     * REQUIRED PRODUCTS
          - WBC latex : LAD 001 AS
          - RBC and PLT latex : LAD 002 AS
          - Soft paper
          - Flat piece of stiff plastic
     * INTERVENTION TIME
          - 60 min
     * FREQUENCY
          - See maintenance chart table.
         * PROCEDURE
                     The arborescence of the «technician functions» is the following :
                                                  1 - CAL. PHOTOMETER
                                                                                  1 - WBC ADJUST
                          1 - VERSION
                                                  2 - APERTURE VOLTAGE
                                                                                  2 - # LYM <      >
                                                  3 - CHECK PRESSURE
                                                                                  3 - # GRA <      >
                                                  4 - WBC ADJUST
                                                                                                              1 - RBC PLT ADJUST
                          2 - ADJUSTMENTS
                                                                           1 - NEEDLE SENSOR                  3 - MPV <    >
                                                  6 - SENSOR ADJUST
                                                                           2 - NEEDLE <   >
                                                                                                              2 - CHECK
                                                                           4 - TRANSFER
    8 - TECHNICIAN
                                                  8 - SAMPLING POSITION
                                                                                              1 - NEEDLE ADJUSTMENT
2 - NO 8 - PIERCING
                          6 - SERIAL #
                                                                                              1 - VISIBLE
                                                  1 - STARTUP # <   >
                                                                                              2 - INVISIBLE
                                                  2 - STANDBY # <   >
                          7 - CYCLE #
1 - BURNING #
                          8 - BURNING             2 - BURNING
                                                                                               MICROS CT
                                                  3 - CONTROL
          Once entering the «SERVICE» sub menu, move to «TECHNICIAN FUNCTIONS» and press                       .
          A specific password (421) is required to enter the sub menus.
          Move the cursor by means of                and           and choose the required menus pressing the
                    key.
II - Adjustments :
             NOTE          - If the WBC chamber has been dismantled previously make sure no liquid has flown
                           in between the spectrophotomer and the chamber.
                           Clean the inner surfaces of the spectrophotometer as well as the chamber.
                           Reassemble the assy and tighten the two screws to the following torque : 400mN/m
                           (see RAS 088 A : Chamber maintenance)
An HGB channel is displayed on the LCD screen representing a voltage value as shown below :
                                                           VALUE
                                                           237
                                 The HGB photometer calibration must be done 20min at least after the
         IMPORTANT               instrument has been switched on.
          - By means of R97 (See diagram 1) adjust the HGB channel according to the room temperature using
          the chart table given on next page.
IMPORTANT This adjustment must be done with the WBC chamber cover installed!!!
                                                                                      HGB
                                                                                 R97        TP27
TP19
                                                                                APERTURE R91
                                                                                 VOLTAGE
                                                                                 60V +0.1
                                                                                     -
Diag.1
2 - Aperture voltage
           - Once entering the «APERTURE VOLTAGE» menu, connect a voltmeter between the ground (TP30)
           and TP19.
3 - Pressure check
                                                         - If the results are not correct check the O ring and the
  Diag.2                                                 tubing watertightness.
4 - WBC adjust
           - Put the WBC latex to mix on a Vortex during 5min or shake thoroughly
                                As the WBC gain is a factory adjustment it is mandatory not to readjust it
           IMPORTANT            without taking the following precautions :
                                Carry out previously an autoconcentrated cleaning to make sure of the
                                cleanliness of the WBC counting circuit.
                                If necessary clean the WBC chamber aperture as described in RAS 088 A.
                                Make sure the Latex has been thoroughly mixed before.
           - Run a blank cycle to check the cleanliness of the instrument. Install the vial in the sampling position
           required for the blood control and turn the tube holder in the compartment inside position.
CAUTION : Make sure the cap of the vial has been removed.
Diag.3
- Wait for several results to be displayed and check the stability of both values.
           After one minute the last volume values displayed on the screen are printed out as well as the WBC,
           PLT and RBC histograms and the CBC “s results. Check that these printed values correspond to the
           following target values :
                  - LYM                 = 57 + 1
                  - GRA                 = 180 + 2
           - If not rerun a «WBC ADJUST» cycle after having previously mixed the latex vial again. By means
           of R74 (see diagram 4) adjust the volumes to the target values during the measuring phasis.
                                                                                                                    THRESHOLD
                                                                           THRESHOLD R75                              350mV+5           APERTURE R91
                                                                            400mV+5         TP13                            -            VOLTAGE
                         TP4                                                      -                                     R74 TP7
                                     GAIN
                                            R87   TP15
                                                                                      R81                          GAIN                  60V +0.1
                                                                                                                                             -
               J17                                                               GAIN
                                                                                            TP11
                                                                                                                                               GND    TP29
               NU
                                                                RBC PULSE PROCESS                  WBC PULSE PROCESS
Diag.4
- Put the RBC and PLT latex to mix on a Vortex during 5min or shake thoroughly
CAUTION : Make sure the cap of the vial has been removed.
          Enter the «RBC PLT ADJUST» sub menu and close the door of the piercing mechanism (Diag.3).
          The needle directly delivers the latex sample in the RBC chamber dilution (for a usual analysis cycle
          the sample is first delivers to the mixing chamber) and a measuring phasis begins.
          - During the cycle measuring phasis (around 1 minute) the Platelet and the Red Blood cell volumes
          are displayed on the screen every 3 seconds as shown below :
- Wait for several results to be displayed and check the stability of both values.
          After one minute the last volume values displayed on the screen are printed out as well as the PLT
          and RBC histograms and the CBC “s results.
- Check that these printed values correspond to the following target values :
                  - RBC = 74 + 1
                  - PLT = 59 + 1
- If not rerun a «RBC PLT ADJUST» cycle after having previously mixed the latex vial again.
          - Adjust the volumes to the target values during the measuring phasis by means of the following
          potentiometers (see Diagram 4) :
6 - Adjust sensors
* Needle sensor
           - For a current number of steps out of ranges, unloosen the 2 screws 5 (diagram 5) and gently move
           the sensor         - upward if the current value is too low
                              - downward if the current value is too high.
           - Tighten the screws 5 and rerun a «NEEDLE SENSOR» cycle. Check that the current value is
           correct.
            This function allows the adjustment of the needle heigth in the RBC chamber. The current number of
            steps is 630 and may be adjusted if reproducibilty on RBC/PLT is incorrect.
            In this case the value has to be raised 10 steps by 10 steps. Reproducibilty on PLT has to be controled
            between each value modification.
            - From the SENSOR ADJUST menu, move the cursor to the function <1> "NEEDLE SENSOR" and
            press ENTER.
            - Push the sampling carriage above the RBC/PLT chamber. Lower the sampling needle until it comes
            into contact with the chamber edge as shown on the Diag.6.
Diag.6
 - Press ENTER and record the number of step displayed on the LCD. Add 204 to the number displayed and
 record the calculated value.
- Move the cursor to the function <2> "NEEDLE" and press ENTER. Enter the calculated value and validate.
* Transfer sensor
Diag.7
NOTE 10 steps correspond to around 1 mm. The target number of steps is 350 + 5
           - Tighten the screws 2 (see diagram 7) and rerun a «CARRIAGE» cycle. Check that the current value
           is correct.
* Transfer
           Once entering the menu «ADJUST SENSORS» (Selection 6), enter the sub menu «TRANSFER».
           Then move the cursor to «2 - CHECK» and press ENTER. The needle moves above the RBC chamber.
           Check the centering of the needle over the chamber.
           If the position of the needle has to be adjust, move to «1 - TRANSFER < >». The value shown on
           the screen is the number of steps from its initial position to the RBC chamber position. The current
           value is 585.
           Press ENTER and raise the value to move the needle towards the left or lower the value to move the
           needle towards the right. 10 steps correspond to around 1mm. Rerun a «2 - CHECK» to control the
           centering of the needle.
7 - Vacuum
           Move to «1 - VACUUM < >» and press ENTER. Raise the value if the drain has to be faster. Lower
           the value if the drain has to be slower. Press ENTER to confirm. Rerun a «CHECK» cycle to control
           the drain of the chamber.
                        Compare to this «CHECK» cycle , the chamber is totally drained during an analysis
              NOTE
                        cycle.
8 - Sampling position
                                                          Two states 0/1 are possible for the 3 switches (see diag
                                                          11) :
                                                                  - Switched OFF : 1
                                                                  - Switched ON : 0
                          The code «0 0 0» means that the tube holder has been turned in between two sampling
              NOTE        positions. The code «1 1 1» means that the tube holder has been removed.
Four of the positions are used on the actual tube holder : (See Diagram 11)
Diag.12
                        The number of step value which is recorded is the maximum value found to the bottom
              NOTE
                        of the position less 3 mm (Diag.12).
            It is possible to enter directly the required number of steps for each sampling position. Proceed as
            following :
            Enter the sub menu that corresponds to the number of the needle : the current number of steps is
            displayed. Enter the new value.
            Increase the number of steps to move the needle deeper or decrease the value to raise the needle.
            Confirm the new value with          . The minimum and maximum step values are as follow :
                                                               NUMBER OF STEPS
                     NEEDLE
                                               MINI.                 DEFAULT                    MAXI.
                         1                       1                      762                         782
                         2                       1                      630                         650
                         3                       1                      860                         880
                         4                       1                      925                         945
                         5                       1                      1005                    1025
  Table 3                6                       1                      580                         600
* Piercing
          The piercing function allows to check the piercing operation on each sampling position. From the
          SAMPLING POSITION menu, move the cursor to the function <8> Piercing and press ENTER.
          Select the required position on the tube holder and close the sample door. The piercing cycle is
          carried out and the number of steps for this position is displayed :
                         NEEDLE 2
                         CURRENT : 635                   STANDARD : 630
When entering the «temperature adjustment» menu the following sub menus are displayed :
                  1 - Uncal. temperature
          When pressing the            key the sensor temperature value uncalibrated is displayed. This value
          should be equivalent to the diluent temperature + 2°.
                2 - Calibrate :
          The temperature must be calibrated according to the diluent temperature :
- Plunge a thermometer directly into the diluent container and leave it for a while until stabilization.
- Run 2 diluent primes («SERVICE» menu, «PRIME» sub menu, «DILUENT» selection).
          - Enter the «CALIBRATE» menu. Note the temperature of the diluent and type in the value (if it is
          different from the previous on the instrument).
          3 - Temperature :
                       1 - Visible : Press          to validate the temperature display.
                                     The value displayed is the calibrated temperature.
                       2 - Invisible : Press         to cancel the temperature display.
                                     (Temperature invisible by default)
IV - Run mode
          The instrument must be configurated in the «USER» mode (configurated by default in the «USER «
          mode), the «CONTROL» mode intends for a factory use.
V - Reagent pack
This function is used to update the instrument from a bottle mode to a pack mode.
VI - Serial #
VII - Cycle #
          Displays the :        Startup number since the first use of the instrument.
                                Stand by number since the first use of the instrument.
                                CBC number since the first use of the instrument.
                                 The startup, stand by and CBC numbers are adjustable in this menu but
         IMPORTANT               it is mandatory to keep the initial values (useful for maintenance
                                 schedules).
VIII - Burning
        CAUTION :          This function which allows the burning of the instrument is intended to a factory
                           use.
*REMARKS
* CONCERNS
                      - Changing the instrument language from the V1.2 version.
* REQUIRED TOOLS
                      - Pair of pliers.
* REQUIRED PRODUCTS
                  - None
* INTERVENTION TIME
                      - 5 min
* FREQUENCY
                      - On request only.
                                              Page 1/4
RAS 122 A Ind.C                                                                   27/10/1999
 MICROS              CHANGING THE INSTRUMENT LANGUAGE
* PROCEDURE
                  - Switch off the instrument.
- Remove both jumpers from E1 (on mother board, see diagram 1).
Diag.1
                                           Page 2/4
RAS 122 A Ind.C                                                                     27/10/1999
 MICROS              CHANGING THE INSTRUMENT LANGUAGE
JUMPER CONFIGURATION ON E1
ENGLISH
FRENCH
SPANISH
GERMAN
                                          Page 3/4
RAS 122 A Ind.C                                                                   27/10/1999
 MICROS           CHANGING THE INSTRUMENT LANGUAGE
* REMARKS
                          Page 4/4
RAS 122 A Ind.C                               27/10/1999
  MICROS                                         REAGENT PACK
* CONCERNS
                       - Replacement of the waste connector O ring
                       - Replacement of the diluent/clean/lyse O rings
* REQUIRED TOOLS
                       - Pair of pliers
* REQUIRED PRODUCTS
                       - None
* INTERVENTION TIME
                       - 15 min
* FREQUENCY
                       - O ring replacements : 1/year
                                              Page 1/3
RAS 123 A Ind.B                                                          09/03/98
 MICROS                                        REAGENT PACK
* PROCEDURE
                  - Disconnect the pack if this one is still connected.
Diag.1
                                          Page 2/3
RAS 123 A Ind.B                                                                 09/03/98
 MICROS              REAGENT PACK
* REMARKS
                  Page 3/3
RAS 123 A Ind.B                     09/03/98
                             TRANSFORMATION OF A MICROS
MICROS                       "BOTTLE" INTO MICROS "PACK"
     * CONCERNS
           This procedure allows the technician to transform a MICROS OT or CT "bottle" version into a MICROS
           OT or CT "pack" version.
                         1 - This procedure is available only for MICROS having a serial number above 600
    WARNING !
                         as some frame parts are not compatible below this number.
                         2 - If the MICROS is operating with a program version below V1.3, a specific
                         password is required to carry out the instrument setup at the end of the procedure.
                         If this password is unknown, the procedure has to be carried out by authorized
                         technician only.
       * REQUIRED TOOLS
            - Hexagonal key 2.5.
            - Large and flat screwdriver.
            - Cutters.
            - Pliers.
            - Ruler.
     * REQUIRED PRODUCTS
           - MICROS reagent packs.
     * INTERVENTION TIME
           1hour 30minutes
     * FREQUENCY
           On request
* PROCEDURE
Diag.1
Diag.2
Diag.3
Diag.4
           - Cut the remaining part of the lyse tube to a length of 480mm. This tube will be used for the diluent
           output of the reagent pack.
- Disconnect the blue tube from the detergent straw to the valve <4> input #2 and remove the tube.
           - Connect a tube 2.29x1100 to the valve <5> input #1. Route the tube from the valve to the top of the
           reagent compartment using the previous route of the reagent tubes.
Diag.6
Diag.8
IMPORTANT Cut the ends of the used tubes to have a better tigthening on the connectors.
  Diag.10
            - Install a MICROS reagent pack in its location according to the instructions given on the pack pac-
            kage. Check for leaks.
            - From the main menu, move the cursor to the function <4> SERVICE and press ENTER. Move the
            cursor to the function <8> TECHNICIAN and press ENTER. Enter the technician password (421) and
            press ENTER.
            - For MICROS instrument operating with a program version below V1.3, a specific password is required.
            Move the cursor in front of the hidden function and press ENTER. Enter this password and press
            ENTER. Move the cursor next to the function <2> PACK and press ENTER. Move the cursor next to the
            function <2> YES and press ENTER. Return to the SERVICE menu with the <ESC> key.
            - For MICROS instrument operating with a program version V1.3 or above, move the cursor next to the
            function <5> PACK and press ENTER. Move the cursor next to the function <2> YES and press
            ENTER. Return to the SERVICE menu with the <ESC> key.
           - From the SERVICE menu, move the cursor next to the function <3> PRIME and press ENTER. Move
           the cursor next to the function <1> CHANGE PACK and press ENTER.
                        The CHANGE PACK function allows a complete priming of the reagent lines and allows
              NOTE      the initialization of the cycle counter. The PRIME function is recommended only to
                        prime the reagent lines.
           - Follow the displayed instructions and check for leaks. Run an autocleaning cycle from the SERVICE
           menu, then several blank cycles.
* REMARKS
* ATTACHED DOCUMENTS
* CONCERNS
          - This procedure allows the technician to check and to lubricate the air syringe and the liquid syringe
          of the ABX MICROS.
     * REQUIRED TOOLS
          - Hexagonal keys.
* REQUIRED PRODUCTS
     * INTERVENTION TIME
          - 30 minutes.
     * FREQUENCY
          - Once a year.
- Absorbant paper.
* PROCEDURE
Diag.2
Diag.3
                         Place some absorbant paper at the instrument rear connections (diluent and
                  NOTE
                         waste) as some liquids may come out when the syringe is pushed.
Diag.4
                                Place some absorbant paper at the instrument rear connections (diluent and
                     NOTE
                                waste) as some liquids may come out when the syringe is pushed.
- Re-install the instrument cover, reconnect the waste and diluent tubes, reconnect the power cable.
*REMARKS
     * CONCERNS
          This procedure allows the technician to set up the LX300 printer according to the user's needs.
     * REQUIRED TOOLS
          - none.
     * REQUIRED PRODUCTS
          - none
     * INTERVENTION TIME
          - 15 minutes
     * FREQUENCY
          - on request
- none
* PROCEDURE
             NOTE   On the MICROS, it is necessary to select the printer format RESERVED 1 of the
                    "PRINTER" menu (function 4) accessible through the "OPTIONS" menu (function 5 of
                    the main menu) then "RESULTS" (function 1).
Diag.1
1 - Printer configuration :
           The printer configuration is printed out when pressing the key <FONT> when the printer is switched
           ON. The configuration used for the MICROS is the factory configuration :
Diag.2
           Each parameter can be modify by the corresponding parameter chart. Each chart is accessible using
           the keys <PAUSE>, <FONT> and <LF/FF> according to the control LED combinaisons :
Diag.3
Diag.4
           The control pannel keys allow the user to set up the main functions of the printer : paper advance,
           paper ejection, and font selection. Control LEDS indicate the printer status.
           LED PAUSE : The orange LED PAUSE lights when the printer stops printing. During each power ON,
           this LED blinks for few seconds and 4 audible beeps occur. When the printer runs out of paper, the
           LED blinks and 3 audible beeps occur. This LED lights also when the paper is in its tear off position.
           When a problem occurs, this LED lights ON and 5 audible beeps occur.
           LEDS FONT 1 and FONT 2 : These 2 green LEDS indicate the selected font. Refer to the printer
           user's manual to select the font.
           Key FONT : During normal operation, the FONT key allows the font selection. For each pressure on
           this key, the selection is modified. Refer to the printer user's manual to select the font. When this key
           is pressed during the printer power ON, the printer setup menu is entered.
           Key LF/FF : During normal operation, a quick pressure on this key allows a ligne feed of the paper.
           Keep the pressure on this key to feed a whole page. This key can be used to load or eject the paper.
           When this key is pressed during the printer power ON, the printing test starts.
           Key PAUSE : When this key is pressed during the printing, the printout stops. Press again on this key
           to restart the printout.
           PAPER PARK : If Z folded paper is used, the paper can be driven to its parking position when
           pressing simultaneously on the keys LF/FF and FONT.
           MICRO ADJUST : This function allows to adjust the loading paper position. See the user's manual
           for details.
4 - Printer description :
Diag.5
Diag.6
*REMARKS
     * CONCERNS
           This procedure allows the technician to transform a MICROS OT or CT "pack" version into a MICROS
           OT or CT "bottle" version.
     * REQUIRED TOOLS
           - Hexagonal key 2.5 or torx keys when applicable.
           - Cutters.
           - Pliers.
           - Ruler.
     * REQUIRED PRODUCTS
           - MICROS reagent bottles.
     * INTERVENTION TIME
           1hour 30minutes
     * FREQUENCY
           On request
* ATTACHED DOCUMENTS
* PROCEDURE
Diag.1
Diag.2
Diag.5
           - Remove the tube 2.29x510 from the detergent output of the reagent pack to the input #2 of the valve
           <4>. Replace it by a blue tube 2.29x1100 from the output #2 of the valve <4> to the detergent straw.
           - Route the tubes from the valves to the top of the reagent compartment using the previous route of
           thereagent tubes.
IMPORTANT Cut the ends of the used tubes to have a better tigthening on the connectors.
Diag.7
- Install the reagent straws at the end of each tube, waste tube included.
- Install the label on the reagent tube next to the straws : MINICLEAN on the blue tube.
           - Place the diluent container at the same level than the instrument, install the MINILYSE and MINICLEAN
           into the reagent compartment. Install the bottle stoppers and place the reagent straws.
           - Place the waste container below the level of the instrument. Place the rubber stopper and the waste
           straw into the container.
Diag.8
Diag.9
           - From the main menu, move the cursor to the function <4> SERVICE and press ENTER. Move the
           cursor to the function <8> TECHNICIAN and press ENTER. Enter the technician password (421) and
           press ENTER.
           - For MICROS instrument operating with a program version below V1.3, a specific password is required.
           Move the cursor in front of the hidden function and press ENTER. Enter this password and press
           ENTER. Move the cursor next to the function <2> PACK and press ENTER. Move the cursor next to
           the function <1> NO and press ENTER. Return to the SERVICE menu with the <ESC> key.
           - For MICROS instrument operating with a program version V1.3 or above, move the cursor next to the
           function <5> PACK and press ENTER. Move the cursor next to the function <1> NO and press EN-
           TER. Return to the SERVICE menu with the <ESC> key.
           - From the SERVICE menu, move the cursor next to the function <3> PRIME REAGENTS and press
           ENTER. Move the cursor next to the function <1> ALL REAGENTS, press ENTER and check for
           leaks. Run an autocleaning cycle from the SERVICE menu, then several blank cycles..
NOTE The ALL REAGENTS function allows a complete priming of the reagent lines.
* REMARKS
• REQUIRED TOOLS
               None.
M I C R O S
• REQUIRED PRODUCTS
None.
• INTERVENTION TIME
10 minutes.
• FREQUENCY
On request.
               - Installation kits :
                      PANASONIC : XEA 448 AS
                      DATALOGIC : XEA 594 AS
                1234567                                           TEST
M I C R O S
                EAN 8
                                                                    2/5
9076543210
1234567
                EAN 13
                                                             CODABAR
                123456789012
                                                                543210
EAN 13
C 128
                  789:;<=>567=
                                                                  98765
                123456789012
* PROCEDURE
                                                                                                       MICROS
  - Make sure that the BARCODE reader is properly fitted in the DIN plug of the RS adaptor
  wiring.
  - Make sure that the MICROS mother board XAA 355 A has an index H or above (MICROS
  standard).
  - Make sure that the MICROS mother board XAA 355 B has an index E or above (MICROS
  EMC).
  - Check the barcode reader works properly by means of the reading test on the previous
  page. If the reading test is correct, the procedure is completed. If not, continue below.
  - When the last label is read, the signal beeps 5 times in order to indicate the end of the setup.
  Check once more on the test labels.
WARNING
For code I2of 5, to avoid a bad interpretation, it is mandatory to hold the reader in order to read
the entire code.
                                                                                                       45
                - If the reading test has failed for the PANASONIC reader, proceed to the barcode setup as
4 5             follows :
No STF
                                   (5 beeps)
M I C R O S
1 bit stop Ø
Code identifier : no Ø
                                                                                                 Exit
                                  No start/stop code
5 Beeps
 - If the reading test has failed for the DATALOGIC reader, proceed to the barcode setup as
 follows :
                                                                                              MICROS
DATALOGIC SETUP
                                                                        4 low beeps
                                                                        +1 high
                                                                        + 4 low
                                                   $+CQ0$-
                                                                        4 low beeps
                                                                        +1 high
                                                                        + 4 low
                                                   $+CP0$-
                                                                      4 low beeps
                                                                      +1 high
                                                                      + 4 low
                                                   $+$*
 In case of a reading error, the
 barcode reader triggers three
 low beeps                                                          4 low beeps
$+
1 beep
CA0CC1AZ0
1 beep
AA1AB11
1 beep
AC110432
1 beep
                                                                   1 beep
                                                                                              45
AI11
1 beep
EA110DEC0
4 low beeps
$-
4 5
              • CONCERNS
• REQUIRED TOOLS
               None.
M I C R O S
              • REQUIRED PRODUCTS
               None.
• INTERVENTION TIME
10 minutes.
• FREQUENCY
On request.
4 5           A barcode reader DATALOGIC DLC6165-M0 is available for the MICROS 45 equipped with a
              mother board XAA 355 B Ind.E and a software version V2.31 or above.
The MICROS 45 barcode reader is set up in order to read the following barcodes :
              ATTENTION: MICROS having a mother board XAA 355 A modified to index H are also
              able to receive the barcode reader
INSTALLATION
- Install the jumper (included in the kit) on E10 as shown on the diagram 1.
               WARNING !
               Once the jumper installed on E10, the instrument data ouput receives 5 volts to supply
               the barcode reader. This voltage should cause damages on computer connection if
               this one is directly connected on the MICROS 45 data ouput. It is then mandatory to
               connect the computer connection only on the cable (DAC 023 AS shown on the diag.2)
               intended for it.
              - Connect the barcode reader as shown on the Diag.2 and the attached wiring diagram DAC
              023 AS (diag 3).
- Install the new EPROM V2.31 software version on U30 (Diag.1) when necessary.
              A barcode function has been added into the SETUP menu which allows to valid the checksum
              or not according to the type of barcode label in use.
              - Control the reading of the barcode reader as described in the following pages.
              The barcode setup can be printed out by the function [5] PRINT CONFIG of the menu [3]
              SPECIAL from menu [5] SETUP.
                                                    M I C R O S
                                      U30
4 5
Diag.1
4 5
Diag 3
                                                                                                           M I C R O S
    * PROCEDURE
- Make sure that the BARCODE reader is properly fitted in the DIN plug of the RS adaptor wiring.
    - Make sure that the MICROS mother board XAA 355 A has an index H or above (MICROS
    standard).
- Make sure that the MICROS mother board XAA 355 B has an index E or above (MICROS EMC).
    - Check the barcode reader works properly by means of the reading test on the previous
    page. If the reading test is correct, the procedure is completed. If not, continue below.
    - Switch on the instrument and within the first seconds, read from the top to the bottom
    the barcode labels located on page 7 of this procedure. The audible beep occurs after
    each reading. Read all the labels from top to bottom and from left to right.
    - When the last label is read, the signal beeps several times in order to indicate the end of
    the setup. Check once more on the test labels.
    WARNING
    For code I2of 5, to avoid a bad interpretation, it is mandatory to hold the reader in order
    to read the entire code.
    This barcode reader allows the direct entry of barcode identification, sample tubes, calibration
    blood lots, QC blood lots on the condition that the instrument is set up with the US identifica-
    tion mode (Identification of the sample is displayed and printed out).
    - Set the identification mode by means of the SPECIAL FUNCTIONS [7] ID MODE (see User
    manual 8.3.7. Identification mode).
    Analysis :
    - Press the         key then carry out the barcode reading. A "beep" is then triggered off and
    the patient identification is displayed. Press          to valid.
    - Run the analysis.
                                                                                                           4 5
    Quality control :
    When the QC run is performed without chip card, and if the sample is identified with a barcode
    label, It is possible from the function [2] QC [2] ANALYSIS to directly identify the lot with the
    barcode reader.
    Calibration :
    If the sample is identified with a barcode label, and if the calibration is carried out without chip
    card, press the           key and read the barcode label.
    The sample identification is displayed on the screen. Press            to valid : the menu turns to
    the expiration date.
1224488
121314151617
                             CODE 39
                   (aucun contrôle de check digit)
                    (no control on Check Digit)
                         12345ABCDEW
                                                                                    123456/$
                                                                          M I C R O S
                      $+CQ0$- Barcode module output mode
4 5
M I C R O S
                                               E
                                              G
                                           A
                                          P
                                      K
                                   N
                                 A
                               L
                             B
                                                       Table of Content
1. - Message Structure .......................................................................................................................... 4
2. - Details about the structure ............................................................................................................ 4
3. - Identifier list and their formats........................................................................................................ 4
3.1.          Hematologic numeric parameters ................................................................................. 4
3.1.1.                    Format description ............................................................................................ 4
3.1.1.1.                                  Numerical field ..............................................................................................4
3.1.1.2.                                  Parameter status .............................................................................................4
3.1.1.3.                                  Example .........................................................................................................4
3.1.2.                    Identifier list ........................................................................................................ 5
3.2.          Pathology........................................................................................................................... 6
3.2.1.                    Flags associated with parameters .................................................................. 6
3.2.1.1.                                  Format description .........................................................................................6
3.2.1.2.                                  Identifier list...................................................................................................6
3.2.2.                    Pathological messages .................................................................................... 6
3.2.2.1.                                  Format description .........................................................................................6
3.2.2.2.                                  Identifier list...................................................................................................7
3.3.          Histograms and matrix ...................................................................................................... 7
3.3.1.                    Format description ............................................................................................ 7
3.3.1.1.                                  Histograms .....................................................................................................7
3.3.1.2.                                  Matrix.............................................................................................................7
3.3.1.3.                                  Separation thresholds .....................................................................................7
3.3.2.                    Identifier list ........................................................................................................ 7
3.3.3.                    Format description of the threshold transmission .......................................... 8
3.3.3.1.                                  WBC identifier...............................................................................................8
3.3.3.2.                                  RBC identifier ................................................................................................8
3.3.3.3.                                  PLT identifier.................................................................................................8
3.3.3.4.                                  Basophil identifier..........................................................................................8
3.3.3.5.                                  Lmne matrix identifier....................................................................................9
3.3.3.6.                                  Reticulocyte matrix identifier.........................................................................9
3.4.          Patient result identification .............................................................................................. 10
3.4.1.                    Format description ............................................................................................ 10
3.4.2.                    Identifier list ........................................................................................................ 10
4. - Packet type ..................................................................................................................................... 11
4.1.          Identifiers List ...................................................................................................................... 11
4.2.          Data exported by the analyzer....................................................................................... 11
4.3.          Errors List.............................................................................................................................. 11
4.4.          Data interpreted by the analyzer ................................................................................... 12
1. - Message Structure
    STX
    Size + carriage return.
    Identifier followed by a space then the Load Type + carriage return.
    Identifier followed by a space then the Information + carriage return.
    Remainder of the other Identifiers and Informations + carriage returns.
    Identifier followed by a space then the CheckSum + carriage return.
    ETX
    Size :        5 bytes representing the total amount of the data except STX and ETX.
    Load :        A character chain indicating that this load is a result type.
    Identifier :  1 byte (moving about $21 to $FF, it describes the information type which follows this
    indicator).
    CheckSum : Sum modulo 65535 of all the characters except ETX and STX in the hexadecimal
    format on 4 bytes.
    3.1.1.3.        Example
    5.5 millions RBC with a counting error in the standard units :
    $32$20$30$35$2E$35$30$52$68$0D ou "0 05.5Rh" + carriage return.
    The length is fixed and is worth 2+7+1, that is to say 10 bytes for one parameter.
                                             Page 5/3
    RAA 006 A Ind.F                                                       09/03/98
    5 - DATA OUTPUT FORMAT
    3.2.   Pathology
3.2.1.     Flags associated with parameters
                                               Page 5/4
    RAA 006 A Ind.F                                                                       09/03/98
                                                                                         5 - DATA OUTPUT FORMAT
    3.3.1.1.            Histograms
    Histograms are transmitted on 128 or 256 channels. They are automaticaly rescaled to a 223
    maximum amplitude value. The zero amplitude value is $20, the maximum amplitude value is $FF.
    3.3.1.2.            Matrix
    2048 bytes matrix are transmitted on 4096 bytes giving by group of 2 bytes the value of 1 screen
    bitmap byte represented from the left to the right and from the top to the bottom of the screen
    (principles of the recovery of the 128x16 matrix bitmap on the VEGA LCD board).
                                                            Page 5/5
    RAA 006 A Ind.F                                                                                           09/03/98
    5 - DATA OUTPUT FORMAT
    Example 1 : output format of the WBC curve thresholds for an LMG sampling. In this example, the
    analyzer doesn't send the calculation thresholds of the L1 flag.
    "] 000 000 000 040 060" + carriage return.
    Example 2 : output format of the WBC curve thresholds for CBC and DIFF sampling. In this 5 parts
    DIFF analyzer, the LMG is not measured, the 4-5 thresholds are not significants.
    "] 005 008 020 000 000" + carriage return.
                      1     2 3 4        5
                          L1
1 2
%MIC %MAC
                                         PLAQ
                                                     1
1 2 3
BASO
                                                     Page 5/6
    RAA 006 A Ind.F                                                                            09/03/98
                                                                                          5 - DATA OUTPUT FORMAT
NOISE EOS
FNE
                                       NoN
                                                                                          NE
                                                            NEUTRO                   RN
                                                    LN
RMN
                                                                          MONO
                            NL
LYMPHO
                                                                                                DC
                                 NoL         LL               AL     LM              RM
                   FLN
FMN LMN MN
F4
F3
F2
F1
                                                                                                DC
                                  R1                                            R2
                                                           Page 5/7
RAA 006 A Ind.F                                                                                        09/03/98
    5 - DATA OUTPUT FORMAT
                                                  Page 5/8
    RAA 006 A Ind.F                                                                                  09/03/98
                                                                                 5 - DATA OUTPUT FORMAT
4. - Packet type
   The information described in the packet type allows the specification of the global message
   content; Analyzers being able to communicate in the bidirectionnal mode and supporting the
   remote control mode, this string of characters can be presented under several forms. Its length is 8
   characters.
   The version number is linked to the development of the hematological message identifiers.
   An error number is also available. It codifies some events which happened on the analyzer (see
   error number list here below).
   4.1.   Identifiers List
                                                      Page 5/9
   RAA 006 A Ind.F                                                                                  09/03/98
5 - DATA OUTPUT FORMAT
                                               Page 5/10
RAA 006 A Ind.F                                                                          09/03/98
                                              5 - DATA OUTPUT FORMAT
2. PIN ASSIGNMENTS
                                             COMPUTER
       MICROS                                  SIDE
    RS 232 OUTPUT                        DB9 CONFIGURATION
      5               GND               GND    5
                  6                            4        6
      4
                  7   TXD               TXD    3        7
      3
                  8   RXD               RXD    2        8
      2
                  9                                     9
      1                                        1
                                             COMPUTER
                                               SIDE
                                        DB25 CONFIGURATION
                                              13
       MICROS
    RS 232 OUTPUT
      5               GND               GND    7
                  6
      4
                  7   TXD               RXD    3
      3
                  8   RXD               TXD    2
      2
                  9
      1                                       1
                            Page 5/11
RAA 006 A Ind.F                                             09/03/98