0% found this document useful (0 votes)
270 views4 pages

Biolyzer-300 GB

The document describes the Biolyzer 300, a clinical chemistry analyzer with a throughput of 270 or 404 tests per hour. It has separate sample and reagent pipettors, clot detection, and tests for quality control. The system is operated via an intuitive interface and offers connectivity to laboratory information systems.

Uploaded by

YT
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
270 views4 pages

Biolyzer-300 GB

The document describes the Biolyzer 300, a clinical chemistry analyzer with a throughput of 270 or 404 tests per hour. It has separate sample and reagent pipettors, clot detection, and tests for quality control. The system is operated via an intuitive interface and offers connectivity to laboratory information systems.

Uploaded by

YT
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 4

Biolyzer 300

The advanced solution for the


routine clinical laboratory
new

• Space efficient fully auto­


mated clinical chemistry
analyzer
• 270 tests/hour
• Optional ISE unit available,
increases throughput to
404 tests/hour
• Clot detection
• Highly reliable and efficient

agile - affordable - accurate


new
Reliable and accurate
Biolyzer ® 300
• Separate sample and reagent pipettors
takes clinical chemistry automation a avoid carryovers and are equipped with
step further. With a throughput of up to liquid level sensors
404 tests per hour and 50 refrigerated
• Clot detection for the sample pipettor
reagent positions the newest genera-
tion of the Biolyzer® systems offers an • Crash detection for both pipettors
increased flexibility. • Direct heating of the 72 cuvettes
Meeting the demands for precise and • Levey­Jennings charts, QC statistics
reliable results the Biolyzer® 300 is and calibration curves
equipped with a number of sophisti-
cated features like high precision sam- • Serum indices to detect lipeamic,
ple and reagent pipettors, clot detec- icteric or hemolytic samples
tion and serum indices testing.
Optional extras include a built-in direct
ISE module for the detection of Na, K
and Cl ions as well as a built-in degas-
ser module. The system is operated via
a highly intuitive graphic user interface
designed with a clear focus on being
user-friendly and on facilitating the
daily routine laboratory work.

Easy-to-use and agile


• The Biolyzer® 300 features a smart
software, which is intuitive, easy-to-use
and straightforward
• Enables sample dilution and automatic
test reruns
• Easy connectivity to LIS
• Sample and Reagent carousel
are easily removable
• Auto wake­up function and sleep
modus
Efficient and
affordable
• Reduced daily main-
tenance (less than 5
minutes) improves
efficiency
• Reduced operating
costs due to semi-di-
posable resin cuvettes
• Intelligent and solid
design provides maxi-
mum output at low
running costs
• Simple installation and
implementation

Analyticon provides a complete solu-


tion of dedicated reagents, calibrators
and controls for the Biolyzer® 300:
• Routine Clinical Chemistry
• Specific Proteins
• Substrates and Enzymes
• Lipid Profile
• Diabetes
• Electrolytes

agile - affordable - accurate


Specifications
Type Desktop fully automated random access analyzer
Usage General chemistry as photometric assay
Immunology as photometric assay (Latex reagents available)
Assay type End point, 2 point end, Rate, 2 point rate and electrolytes
Throughput 270 tests per hour (Up to 404 tests per hour with ISE)
Incubation time Ten minutes after applying sample for one reagent assay
Five minutes after applying second reagent for two reagents assay
Sample type Serum, Plasma and Urine
Sample input Type: Removable tray with sample tube holder on a turntable
system Capacity: Up to 40 samples (handling of STAT available)
Controls and Standards: 10 positions on designated inner tray
Sampling pipette Type: Micro-pipette with liquid level detector
Rinsing: Inside and outside with purified water
Sampling pump Type: Micro syringe
1.5 to 35 µl (increment by 0.1 µl)
Reagent system Type: Removable tray with reagent bottle holder on a turntable
Capacity: 50 reagent positions on a tray
(Twenty five 70 ml and twenty five 20 ml bottle positions)
Inventory: Calculation of remaining reagent volume available
Cooling: Cooling with Peltier element (8 to 15 °C)
Reagent pipette Numbers of pipettors: One
Type: Micro-pipette with liquid level sensor
Rinsing: Inside and outside with purified water
Number of reagents: Up to two reagents
Inventory: Calculation of remaining reagent volume available
Reagent pump Type: Micro syringe
R1 Volume: 20 to 350 µl (increment by 1µl)
R2 Volume: 20 to 180 µl (increment by 1µl)
Cuvette system Number of cuvettes: 72 cuvettes on reaction line
Type: Semi-disposable cuvette with onboard washing system
Material: Resin
Volume for measurement: Minimum 100 µl
Maximum volume: 350 µl in total
Reaction system Type: Direct heating system
Temperature: 37 °C
Detector Method: Direct measurement of absorbance in cuvette
Bi-chromatic or Mono-chromatic
Wavelengths: 340, 380, 415, 450, 510, 546, 570, 600, 660, 700, 750, 800 nm
Light source: Halogen tungsten lamp
Stirring system Stick type rotating stirrer
Dimensions 870 mm (W) x 670 mm (D) x 582 mm (H) (Analyzer only)
Optional On board ISE module, on board degasser module, liquid level sensors for external tanks

Analyticon Biotechnologies AG
Am Muehlenberg 10 · 35104 Lichtenfels - Germany · Phone: +49 64 54 79 91-0 · Fax: +49 64 54 79 91-71
www.analyticon-diagnostics.com · info@analyticon-diagnostics.com