Technical Specifications                                                                                                                                 Bench Top Random Access
System type
  Throughput
                                Automated bench top random access clinical chemistry analyzer
                                200 tests / hour with a cycle time of 18 seconds for mono and two component tests
                                                                                                                                                           Clinical Chemistry Analyzer
                                360 tests / hour with ISE
  Sample types                  Serum, plasma, urine, CSF, whole blood
  Sample volume                 2 – 70 µL
  Reagent pipetting             Reagent 1: 50 – 300 µL      Reagent 2: 10 – 200 µL
  STAT analytics                Several sample positions available
  Ion measurement               Direct potentiometry: Na, K, Cl, Li (optional)
  Bar code identification       Automatic bar code scan for reagents and samples
  Measuring principle           Colorimetry (Rate / Endpoint); Immunoturbidimetric assay
  Calibration                   Linear, non-linear, multi-point
  Sample tray                   30 positions for bar coded patient samples including STAT positions, 9 positions
                                for blanks, calibrators, controls or samples without bar code and ISE solutions
  Sample tubes / cups           Most commonly used primary blood collection tubes and sample cups
  Sample dilution               Dilution ratio: 2- to 150-fold
  Reagent onboard capacity      30 different methods in bar coded mono or twin containers
                                for adapter free one grip loading, refrigerated
  Reaction temperature          37 ± 0.2°C
  Reaction unit                 Temperature controlled heated rotor with 45 reusable quartz glass cuvettes
                                (37 ± 0.2°C)
  Photometry                    8 wavelengths: 340, 405, 450, 505, 546, 578, 660 and 700 nm (mono and bichromatic)
  Photometric linearity         Linearity: 0 – 2.5 OD; Resolution: 0.0001 OD
  and resolution
  Water consumption             Up to 7.5 Liters per hour
  System interface              Analyzer PC: USB connectivity bi-directional; CPU: Pentium IV or higher
  LIS connectivity              Yes
  Power source /                AC 220 V ± 10%, 50 ± 1 Hz or AC 110 V ± 10%, 60 ± 1 Hz; 600 VA
  Power consumption             (excluding PC / printer / monitor)
  Dimensions                    81 cm (W) x 70 cm (D) x 60 cm (H)
  Weight                        Approximately 110 kg
Handed over by:
                                                                     klimaneutral                       Climate-neutral print
                                                                                                        (Carbon neutral)
                                                                     natureOffice.com | DE-626-724545
                                                                     gedruckt                           ID: DE-626-724545
                                                                                                        Printed on FSC-certified paper
                            DiaSys                                Phone: +49 6432 9146-0
                                                                                                                                   820102 I October 2019
                            Diagnostic Systems GmbH               Fax:    +49 6432 9146-32
                            Alte Strasse 9                        E-Mail: systems@diasys.de
                            65558 Holzheim                        www.diasys-diagnostics.com
                            Germany
                                                                                                                                                           Economical. Efficient. Precise.
                                                                                                                                                           Trusted Performance.
Reliable Results, Optimized Workflow
and High Efficiency
respons®920 is the evidence of our ambition to develop and produce outstanding system solutions
for the diagnostic laboratory. Bearing in mind the specific demands on throughput and flexibility,
this automated random access clinical chemistry system has been designed as a real all-rounder.
Due to its concept, respons®920 may easily be integratedin all types of laboratories for routine,
emergency as well as speciality analysis. respons®920 stands for reliable results, optimized
workflow and high efficiency. Achieved by the perfect match of analyzer, system reagents,
applications and our service.
  Guaranteed Throughput of 200 Tests / Hour
  · 200 tests / hour with a cycle time of 18 seconds for mono and two component tests
  · 360 tests / hour with optional ISE unit
  The answers to Your Needs
                                                                                                                  Pleasing to Your Budget
  · High onboard reagent capacity of 30 different methods in bar coded mono and twin containers
  · Long term reagent and calibration stabilities
                                                                                                                  ·	Minimized running costs
  · Large panel of high quality clinical chemistry and immunoturbidimetric tests
                                                                                                                  ·	Low maintenance system
     manufactured by DiaSys
                                                                                                                  ·	High onboard and shelf life stabilities of DiaSys reagents
  · Unique and convenient respons® system container concept
  · 30 positions for bar coded patient and STAT samples
  · Flexible sample matrix
                                                                                                                  Easy to Use
  · Low sample volume
  · Customer oriented menu extension for specific system adaption
                                                                                                                  ·	Automatic bar code scan for samples and reagents
  · Wide measuring ranges
                                                                                                                  ·	Ready to use liquid stable reagents
                                                                                                                  ·	Adapter free one grip loading of reagent containers
  Inter-assay Precision and Recovery                                                                              ·	Versatile software: Set up in eight languages
                                                                                                                  ·	Easy to learn
  Parameter            Target      Mean        Reco       Target   Mean    Reco   CV [%]   CV [%]   Patient
                       TLN*        TLN*        very        TLP**    TLP**   very    TLN*     TLP**    Conc./
                       value       value       [%]         value    value   [%]                       [CV%]
  AP [U/L]             74.2        72.5        97.7        244      230     94.3    0.64     0.64     54.8/0.5    Method Comparison Urea                                                               Method Comparison Creatinine
  AST-P5P [U/L]        37.2        38.2        103         200      186     93.2    1.00     0.38     29.7/1.1
  AMY [U/L]            72.0        72.2        100         273      265     97.2    0.55     0.36     39.4/0.7                                600                                                                                         7
                                                                                                                                                                                                       Creatinine [mg / dL] respons®920
  Ca-P [mg/dL]         9.54        9.24        96.9        12.2     11.9    97.5    0.51     1.46     9.40/1.5                                                                                                                            6
                                                                                                                 Urea [mg / dL] respons®920
                                                                                                                                              500
  CK [U/L]             133         134         100         543      520     95.8    0.73     0.47     144/0.6                                                                                                                             5
                                                                                                                                              400
  CREA-J [mg/dL]       1.13        1.11        98.2        7.73     7.20    93.1    0.57     0.97     0.91/1.8                                                                                                                            4
                                                                                                                                              300
  DBIL [mg/dL]         0.53        0.55        104         2.24     2.43    109     0.81     0.61     0.08/8.3                                                                                                                            3
                                                                                                                                              200
  TBIL [mg/dL]         1.00        0.96        96.0        5.45     5.56    102     1.57     1.35     0.08/5.7                                                                                                                            2
                                                                                                                                              100
  LDH [U/L]            144         138         95.6        394      374     94.9    0.64     0.49     141/0.8                                                                                                                             1
                                                                                                                                               0                                                                                          0
  Lipase [U/L]         42.1        45.2        107         80.9     80.5    99.5    0.88     0.61     43.2/1.6                                      0     100     200    300     400    500      600                                          0     1     2      3     4      5     6      7
  PO4 [mg/dL]          3.39        3.28        96.8        7.09     6.83    96.3    3.45     0.51     4.53/2.1                                                     Urea [mg / dL]                                                                         Creatinine [mg / dL]
                                                                                                                                                        Competitor clinical chemistry analyzer                                                    Competitor clinical chemistry analyzer
  TP [g/dL]            5.32        5.03        94.5        6.39     5.95    93.1    1.02     0.63     6.84/0.7
                                                                                                                         n = 89; Passing / Bablok Regression:                                          n = 100; Passing / Bablok Regression:
  UA [mg/dL]           6.33        6.23        98.4        9.44     9.18    97.2    0.37     0.41     4.30/0.5           Y = 0.999 X - 0.622 [mg / dL]; r = 0.9981                                     Y = 0.961 X + 0.018 [mg / dL]; r = 0.9949
* TruLab N »Normal« control   ** TruLab P »Pathological« control