Portable USB
vector network
   analyzer
miniVNA PRO&PRO Lite
Product Overview
  The professional VNA-PC software is used for data analysis and processing. The signal generator function is
   integrated. The measurement range is 100kHz~200MHz, and the S parameter measurement of S11 and S21
   is supported.
  The miniVNA PRO & PRO Lite has the software open-circuit, short circuit, load, and straight-through
   calibration functions (the corresponding calibration components need to be prepared or purchased by the
   customer).
  The software supports data export in multiple formats, including JPEG, Excel, ZPlots, S1P, S2P, and PDF.
  Windows, Linux, and Mac
  The miniVNA PRO Lite : CNC aluminum shell (red ); The miniVNA PRO : Plastic shell (blue).
  miniVNA PRO Lite does not support Bluetooth connection, because there is no built-in Bluetooth module
   and lithium battery.
 ** This product does not contain other accessories. If you need to use it, purchase it separately.
Product Features
 Frequency range: 100KHz~200MHz
 Frequency step: 1Hz
 Port configuration: Dual ports
 S parameter: S11 and S21
 Dynamic rang: 90dB
 Z: 1~1000 ohms
 Output power: 0dB
 Full phase measurement and integration with Smith chart
 With software open, short, load, and pass-through calibration
   functions, the test result is more accurate.
 Supports data export in multiple formats (JPEG, Excel, ZPlot, S2P, and
   PDF).
 Friendly user interface, supporting Windows, Linux, and Mac
 The RF port is an SMA connector (SMA-K).
 USB power supply
 Super mini size, easy to carry (ultra-light and ultra-thin)
Precautions
 Note 1: If the precision is high, you are advised to calibrate the device after the device is powered on and
  calibrate it 15 minutes after the device is powered on. Otherwise, the measurement precision may be
  inaccurate due to temperature changes.
 Note 2: Because the minVNA PRO&PRO Lite needs to dissipate heat effectively, do not cover sundries on
  the surface to ensure good heat dissipation on the surface. Because the surface temperature is high after
  the device is used for a long time, take protective measures against scalding.
 Note 3: Because the USB current of the device is large, you are advised to connect the USB cable to the
  USB port on the mainboard to increase the reliability of the USB power supply.
 Note 4: The USB port of the device must be stable. You are advised to use a fixed or uncommon USB port
  of the PC as the minVNA PRO&PRO Lite port.
 Note 5: All legends, scanning results, and measurement quantities in this document are for reference only
  because of the differences between devices, environments, proofs, and products. The test tools and
  accessories involved in this document must be purchased by the customer.
 Note 6: A similar attenuator must be added to the test amplifier. Otherwise, the device may be damaged.
Contents
       01                                                       Preparing for the Use
               JAVA installation      Installing the Driver      Querying          Software running
                                                                    Device Ports       and serial port test
       02                                                       Use the Equipment
       03                                                       FAQ
                                                                                           JAVA        Installing     Querying       Software running
                                                                                        installation   the Driver    Device Ports   and serial port test
JAVA installation
    The miniVNA PRO&PRO Lite vector network analyzer must be used in the Java environment. Therefore, you need to install
the Java software of version 7.0 or later on the computer. Select the WINDOWS/Linux/MAC system according to the actual
situation and install the Java software of the 7u80 version(http://www.eastchip.com/minivnasoft.asp). If the JAVA7.0 or a
later version has been installed on the PC, you do not need to install it again. If you need to install the Java of 7u80 or later,
download the corresponding version from the official website of Oracle (https://www.java.com/en/download/).
    The following uses the Windows 7 (64-bit) system as an example to describe how to install the Java. The file name is jre-
7u80-windows-x64.exe. Run the file as shown in the following figure to complete the installation.
    Click Install                          Waiting                     Continue waiting                             Click Close
                                                                                          JAVA        Installing    Querying       Software running
                                                                                       installation   the Driver   Device Ports   and serial port test
Installing the Driver
    Use a USB cable to connect the miniVNA PRO&PRO Lite analyzer to the computer. Select the WINDOWS/Linux/MAC
system according to the actual situation and download the corresponding driver according to the 32/64 bit. In this example,
click http://www.eastchip.com/minivnasoft.asp or https://www.ftdichip.com/Drivers/D2XX.htm to download the CDM
v2.12.12 WHQL Certified.zip. Run the file as shown in the following figure to complete the installation.
    Install the driver software for FT232 USB UART and USB Serial Port according to the pattern on the continue two pages.
                               JAVA        Installing    Querying       Software running
                            installation   the Driver   Device Ports   and serial port test
Installing FT232 USB UART
                                JAVA        Installing    Querying       Software running
                             installation   the Driver   Device Ports   and serial port test
Installing USB Serial Port
                                                                              JAVA        Installing    Querying       Software running
                                                                           installation   the Driver   Device Ports   and serial port test
Querying Device Ports
    Use a USB cable to connect the miniVNA PRO&PRO Lite analyzer to the computer. Right-click My Computer
 and choose Device from the shortcut menu. In the Device Manager window shown in the following figure,
 check whether USB Serial Port (COM**) exists in Port (COM and LPT). If yes, the driver installation is complete.
 In this example, the COM port is COM137.
                                                                                                           JAVA        Installing         Querying       Software running
                                                                                                        installation   the Driver        Device Ports   and serial port test
 Software running and serial port test
           Go to http://www.eastchip.com/minivnasoft.asp , download vna.3.1.9.cn.jar to the directory selected by
     the user, and double-click vna.3.1.9.cn.jar. (When vna.3.1.9.cn.jar is executed, the file is decompressed. For
     details, see FAQ 1 in this document.)
           Note: The folders, parameter settings, calibration files, and exported files automatically generated by the
     software are saved in the vna.3.1 folder of the C:\user\Admininstrator.
Click OK. (Note: In reflection or   Select Analyzer on the       In the dialog box that is       After the serial port              In the displayed dialog
transmission measurement            toolbar and click Setup...   displayed, select               test is passed (the                box, click OK to use the
mode, if there is no                                             miniVNA-pro in the left         status in the dialog box           software. (The preceding
corresponding calibration data,                                  column and COM** in             is green), click Update            figure is not displayed for
the software prohibits the                                       the right column (port          at the lower right                 the devices that have been
corresponding measurement.                                       number queried on the           corner.                            used and calibrated.)
For details, see the device                                      previous page, for
calibration and test examples                                    e x a m p l e , C O M 137 ) ,
in the specification.                                            and click Test.
Contents
       01                                                      Preparing for the Use
       02                                                      Use the Equipment
               Introduction to the      Device Calibration         Transmission              Secondary
                Software Interface                                    Calibration and Test       calibration
                                                                      Cases
       03                                                      FAQ
                                                                                          Software               Device              Transmission            Secondary
                                                                                          interface            Calibration            Calibration            calibration
Software Interface and Functions
   The software version vna.3.1.9.cn is used as an example. The menu
bar and three important icons are displayed. You can move the pointer
over the corresponding icon to display the corresponding functions.
              File               Tool           Calibration                 Export            Analyzer                                Preset
            Settings            Analyze            Frequency                   CSV                 Setup                                Load
             Scales             Beacon               Create                    JPEG                 Info                                Save
             Colors                                   Load                      Pdf             Reconnect
                             Cable length
            Language                                                        S-parameter        Single scan
                               Cable loss            Import
              Exit                                                          S-parameter    Free-running scan
                                  FFT                Export                   collector
                                                                                          Port extension parameter setting, usually used for reflection (S11)
    Smith icon. The test       Firmware                                       EXCEL
                               download
                                              Open the dialog box for                     measurement between device and target measurement device,
    result of the positive                    loading the XML                             extension cable between device and device (such as PCB board) (this
                                                                               XML        cable should be RF 50 ohm relatively low loss cable), and cable One end
    reflection S11 is          Generator      reference data. For                         can be directly connected to the DUT port of miniVNA PRO&PRO Lite.
    displayed in the                          example, after the last         ZPlots      For example, the other end of the cable needs to be soldered to the PCB
    Smith chart.               Scheduler      measurement is                              (including the ground wire of the cable to be soldered). In this dialog
                                                                             Settings     box, the speed factor and length of the cable can be entered, that is, in
                                              complete, export the                        the standard After the reflection is calibrated (see the reflex calibration
                              Multi-tune
                                              XML data. After the test      Auto-Export   instructions in the manual), use this operation to remove the
                                              is complete, load the           settings    measurement error caused by the connecting cable (the cable should be
                             Pad-Calculator                                               controlled in a few centimeters to ten centimeters, of course, the shorter
                                              XML data. The old data                      and better, the higher the frequency or the longer the cable, the more
                                              is displayed in the list in                 the error Large), cable length measurement can be measured in the
                                              the dialog box, you can                     [Tools] - [Cable Length] in the menu bar (including speed factor
                                                                                          measurement), the recommended cable type and speed factor are
                                              also change the file                        available to the cable supplier. Note: This operation is currently only for
                                              stop record to compare                      users to try, and does not guarantee the complete correctness of the
                                              the data before and                         test results!
                                              after the upgrade.
                                                                        Software      Device      Transmission   Secondary
                                                                        interface   Calibration    Calibration   calibration
Common parameters
 Common parameters on the main interface of the software (option selector)
         Name                        Content
         -none-                 No data is displayed.
           RL       Return loss (only for reflection measurement)
           RP       The signal is a reflection measurement phase.
           TL         Transmission loss (only for transmission
                                   measurement)
           TP        Phase of signal transmission measurement
          SWR                   Standing wave ratio
          RSS        Absolute value of the transmit sensor (not
                                supported currently)
           Rs              Rs series equivalent impedance
         Theta                      Phase angle
          Tgr                          Delay
           Xs              XS series equivalent reactance
           |Z|                 |Z| complex impedance
                                                                                                          Software         Device      Transmission     Secondary
                                                                                                          interface      Calibration    Calibration     calibration
Setting the main interface of the software
                                                  Preset settings on the main                     Selecting the main interface mode
                                                  interface of the software (in the
                                                  middle of the right part of the
                                                                                                  (lower right part of the software)
   Setting the start and end
frequencies of the software                       software)                                         Two measurement modes
main interface (upper right                                                                           Transmission (S21)
corner of the software)                                                                               Reflection (S11)
     You need to enter the frequency              Th e th re e i co n s i n di c ate add i n g,
to be tested in the start and end                 deleting, and using the preset test
frequencies based on the features and             frequency band respectively.
experience values of the target test                Add a preset test frequency segment.
components. The minimum star t                    Enter the value to be set in Start and
                                                  End Frequency. Click Add to add a
                                                                                                  Single scanning and repeated scanning (at the
frequency of miniVNA PRO&PRO Lite                                                                 lower right corner of the software) on the
is 100KHz and the maximum end                     preset test frequency segment.                  main interface of the software
frequency is 200MHz. The number and                 If the cursor is in the preset test           Single scan: Connect the measurement components and
K/M/G input are supported, generally,             frequency segment, click the icon to            set all the test parameters. Then, perform a single
i t i s n o t re c o m m e n de d t h a t a l l   delete it.                                      measurement
frequency bands be scanned.                         The mouse pointer is in the preset            Repeated scanning: You need to perform a one-time
                                                  test frequency segment. Click the icon          scan before selecting this option for onsite
                                                  to start a single scan.                         commissioning.
                                                                                                                Software         Device        Transmission     Secondary
                                                                                                                interface      Calibration      Calibration     calibration
Setting Interface Marking Points
   M: Move the cursor to the measurement result curve and the corresponding measurement data (automatic) of the current mouse position.
   Mark 1: Move the mouse pointer to the measure result curve, click the left mouse button in the desired position, and then move the mouse wheel.
   Triangle: Calculate the absolute difference between the data marked 1 and 2 (automatic).
   Mark 2: Move the cursor to the measure result curve, press and hold the Shift key on the keyboard, and press and hold the Shift key to move the
     mouse wheel.
   Mark 3: Move the cursor to the measure result curve, hold down the Ctrl key on the keyboard, and click the left mouse button. After the setting,
     press and hold the Ctrl key and move the mouse wheel.
  Flag 4: Move the cursor to the measure result curve, press and hold the Shift+Ctrl key on the keyboard, and press the Shift+Ctrl key on the
  keyboard to move the mouse wheel.
   To cancel the mark point, remove the tick () on the right of the software tag point table area.
                                                                                          On the right of the tag point parameter table, there are two icons.
The following figure takes the parameters in the following figure as an example. The RL
                                                                                          Click        , as shown in the following figure. You can view more
and RP parameters in the figure have the maximum and minimum automatic search             parameters, including the bandwidth, bandwidth start and end
functions. You only need to use the mouse to point to the RL or RP in the corresponding   frequencies, bandwidth limit (dB), and Q value, as shown in the following
flag parameter.                                                                           figure, for parameters such as C and L, you can view the parameters
                                                                                          related to different bandwidths by changing the value of Limit (dB), as
                                                                                          shown in the following figure (right).
 Click     , as shown in the following figure. You can view the corresponding
 frequency point and SWR value.
                                                                                                                Software         Device        Transmission     Secondary
                                                                                                                interface      Calibration      Calibration     calibration
Device Calibration
Definition:         Calibration is a prerequisite for accurate measurement. It helps users accurately measure parameters. It is
                    recommended that the miniVNA PRO&PRO Lite vector analyzer calibrate the equipment before each test. If the data
                    requirement is not high, the previous calibration data can be used.
       miniVNA PRO &PRO Lite can be used to measure the reflection and
transmission modes. Before calibration, you need to set and test the serial port. In
                                                                                           SMA Calibration
the following figure, select the reflection and transmission modes and calibrate them.
                                                                                           Part
Select the corresponding mode in the following figure.
                                                                                          Note: The three calibration parts in the preceding figure need to be purchased
                                                                                          separately (SMA connectors of the calibration parts are internal screws). The
                                                                                          three calibration parts are used for reflection (S11) calibration.
                                                                                           SMA
    Reflection measurement (for example, antenna measurement) and transmission             straight-
measurement [for example, a filter, a bandpass, an amplifier, and a duplexer               through
                                                                                           cable
                                                                                         Note: Generally, one or two SMA straight through cables are required in the
                                                                                         preceding figure. The two ends of the cable are inner screw pins, which are
                                                                                         shorter and better when the test conditions are met. A straight-through
                                                                                         conversion connector (both ends are external screw holes) is required, the
                                                                                         attenuator and other SMA connectors may also be required. other accessories
      In the preceding figure, two modes can be selected during calibration. The         (fewer accessories) on the non-target test samples must be added to the
default value is Mode2 (Mode1 is used when precise calibration is required for           calibration for calibration and removal. this is usually used for transmission
certain frequency bands, and the frequency band and number of Mode1 are set              (S21). The accessories in the preceding figure need to be purchased by the
accordingly. For details, see FAQ 2 in this document.) The number of scanning times      customer.
is greater than or equal to 1 and less than or equal to 10. The number of calibration
points must be greater than or equal to 100 and less than or equal to 20000. The
more the set points, the more the times, and the longer the calibration time.
                                                                                                                                            Software            Device                 Transmission               Secondary
                                                                                                                                            interface         Calibration               Calibration               calibration
  Reflection calibration 1/2
  DUT port. The 13.56M antenna is used as an example
                                                   (The three SMA proofs on the left are used                                                                                          Note: Each device
          SMA Calibration Part               for reflection calibration. The SMA connectors are                                                                                        has a difference. The
                                             internal screws. You only need to perform the                                                                                             calibration curve is
                                             corresponding operations.)                                                                                                                for reference only.
                                             In the right pane, select the reflection mode in
                                             the lower right corner, and then choose Toolbar >
                                             Calibration > Create.
Window displayed after the                 The DUT is connected to the OPEN calibration component. The     Connect the DUT to the SHORT calibration kit,       Connect the DUT to the LOAD calibration component and click
                                                                                                                                                               the read load in the preceding figure until the following figure is
creation                                   default number of calibration points is 10000. Click the read   and click read in the preceding figure to short-    displayed.
                                           button in the preceding figure until the following figure is    circuit the DUT until the following figure is       (Do not remove the LOAD calibration part after the
                                           displayed (2000-10000 point calibration is recommended).        displayed.                                          calibration is complete.)
Click Save. A dialog box is displayed, as shown in      Click Update at the lower               You can start to use reflection measurement to check whether the
the following figure. Change the file name at the       right corner of the following           reflection measurement calibration of the device is correct. The LOAD       The following figure shows
lower left corner and click Save at the lower right     figure.                                 calibration component is properly connected to the DUT interface.           the scanning result.
corner.                                                                                         Click at the lower right corner to perform a single scan.
                                                                                                              Software             Device           Transmission       Secondary
                                                                                                              interface          Calibration         Calibration       calibration
Reflection calibration 2/2                                                                            Note: Each device has a difference. The calibration
DUT port. Take the 13.56M antenna as an example                                                       curve is for reference only.
      Connect the target antenna to the DUT port on the miniVNA PRO&PRO              In the upper right corner of the main interface, set the frequency, and click
      Lite. If a feeder is connected, the correction point should be at the end of   Scan at the lower right corner. After the scanning is complete, click the Smith
      the feeder. If there are other extension components, the calibration point     icon     . The preceding figure is displayed. The calibration is complete.
      is also at the end of the miniVNA PRO&PRO Lite.
                                                                                                                              Software            Device         Transmission       Secondary
                                                                                                                              interface         Calibration       Calibration       calibration
 Transmission calibration 1/2                                                                                                                                        Note: Each device
 DUT to DET forward transmission. The filter is used as an example                                                                                                   has a difference. The
                                                                                                                                                                     calibration curve is for
                                                                  Two SMA straight through cables (both ends are inner screws,                                       reference only.
                                                                  shorter and better if the test conditions are met) and a straight-
                                                                  through conversion connector (both ends are external screw holes)
 Connect the DUT and DET to an SMA
 straight-through cable. Connect one                   Select the transmission mode in                 Select Create … Page after                      By default, the number of calibration
 straight-through cable to the end of the              the lower right corner, and then                                                                points in the preceding figure is 1000.
 straight-through cable connected to the               choose ToolsCalibration > Create.                                                               Click read in the preceding figure
 DUT port (for reference only).                                                                                                                        until the legend is displayed.
                                                                                                                                                  You can start to use the transmission
Connect the straight-through      In the point calibration dialog box, read     The following figure is             Click Update at the lower     measurement. Check whether the
cable between the DUT and the     the loop until the preceding figure is        displayed. Change the file          right corner.                 transmission calibration of the device is
DET port through the straight     displayed. (Do not remove or disconnect       name at the lower left corner                                     correct. The straight-through cable connects
connector at the end.             the connection after the straight line        and click Save at the lower                                       the DUT to the DET. Click the lower right
                                  calibration is complete.) Then, click Save.   right corner.                                                     corner to start a single scan.
                                                                                 Software            Device          Transmission   Secondary
                                                                                 interface         Calibration        Calibration   calibration
Transmission calibration 2/2
DUT to DET forward transmission. Take the filter as an example.
                                                                                             Note: Each device has a difference.
                                                                                             The calibration curve is for reference only.
After the scanning is complete, the   Cancel the automatic scaling in the main               Connect the low-pass filter to the
following figure is displayed.        interface of the software. The following               DUT and DET respectively.
                                      figure is displayed. (The transmission
                                      calibration of the device is normal.)
                                                                                   Set the frequency in the upper right
                                                                                   corner of the main interface, and
                                                                                   click Scan at the lower right corner.
                                                                                   After the scanning is complete,
                                                                                   disable the TP (°) parameter in the
                                                                                   right column, as shown in the
                                                                                   following figure.
Contents
       01                                                Preparing for the Use
       02                                                Use the Equipment
       03                                                FAQ
               Decompress the driver      Frequency bands and      Software parameter
                package based on the        points of Mode1           repair
                Java driver
Common problem 1:
After the JAVA is installed, the vna.3.1.9.cn.jar file is decompressed.
 Perform the solution in sequence. If the solution is solved, you do not need to perform the
 following steps:
     1. Right-click vna.3.1.9.cn.jar and choose --Java (TM) Platform SE binary from the shortcut
 menu.
     2. Save the file on the right to the folder where the vna.3.1.9.cn.jar file is stored and run the
                                                                                                           Save as… and
 file. (Note: The folder path cannot contain Chinese characters. In this case, vna.3.1.9.cn.jar cannot        rename .txt
                                                                                                         to .cmd before use
 be placed on the desktop because the file path on the desktop also contains Chinese characters.                this file.
 Note: In the preceding information, vna.3.1.9.cn.jar is the software version. If the version is
 vna.3.1.8.jar, right-click the saved file and choose Edit from the shortcut menu. Change
 vna.3.1.9.cn.jar to vna.3.1.8.jar in English.
Common problem 2:
How Do I Set the Frequency Segment and Number of Mode1 Points?
    Open the C:\ user \Admininstrator\vna.3.1\ presets file and open the CalRanges_miniVNA-pro.txt file. (You are
advised to back up the file.) The following table describes the parameters in red. You are advised to modify only the
number of scan points and the number of scanning times. Save the modification and restart the vna.3.1.9.cn.jar. (The
number of scanning times of each frequency band is greater than or equal to 1.) Less than or equal to 10 The number
of calibration points must be greater than or equal to 100 and less than or equal to 20000.
                             Start frequency (Hz)   End frequency (Hz)   Scan points Scan times
                                      100000              999999            4000          1
                                     1000000              9999999           4000          1
                                     10000000            29999999           10000         1
                                     30000000            200000000          10000         1
 Common problem 3:
 How to solve software parameter modification can not be restored?
   Close the software, open the C:\ user \Admininstrator directory, find the vna.3.1 folder,
delete the vna.3.1 folder (the exported files and calibration data in the folder can be saved as
required), and then run the software to generate the vna.3.1 folder again, all internal
configuration data is regenerated (the folder of vna3.2.X or a later version is vna.3.2).