Application note #25
Multi Pitting Corrosion
I- Introduction
                                                                       - 4 working electrodes made of steel,
The effect of the corrosion in our daily life can
                                                                       - Ag/AgCl electrode as a reference
be underlined by its cost for the society. For an
                                                                         electrode,
industrial country, the cost is estimated at 3-
                                                                       - platinum wire as a counter electrode.
4% of gross domestic product.
Most metals are sensitive to corrosion in the
presence of oxygen and humidity. The general
equation of this oxidative process is described
as follows:
       M → Mz+ + z e-
Corrosion is a general term which generalizes
various kinds of processes, such as:
 - uniform corrosion: corrosive attack on the
   entire surface area,
 - crevice corrosion: localized corrosive
   attack caused by oxygen,
 - galvanic corrosion: corrosion caused by
   two different materials in contact,
 - pitting corrosion: localized corrosive attack
   caused by aggressive anion.
In this note, only the pitting corrosion in
presence of chloride is investigated. For this
purpose, Linear Polarization (LP) and                                                   Fig. 1: N'Stat Box.
Multielectrode Potentiodynamic Pitting (MPP)
are performed.
II- Experimental conditions
Investigations are carried out with a VMP3 and
EC-Lab® software with 3% NaCl at room
temperature in aerated water. Before
measurements, the electrodes were polished
for 5 minutes using the PK-3 Electrode
Polishing kit and rinsed by distilled water and                     Fig. 2: “CE to ground connection”                    in
alcohol.                                                            “Advanced Settings” window.
A multi-working electrode set-up is required. In                    Note: All settings and raw data files presented
order to overcome the connection complexity                         hereafter are available in the Data Sample
due to the multielectrode investigation, N'Stat                     folder of EC-Lab® Software as LP_03.mpr,
Box is used (Fig. 1). Note that the utilization of                  LP_04.mpr,         LP_05.mpr,       LP_12.mpr,
such a device requires the “CE to ground”                           MPP_03.mpr, MPP_04.mpr, MPP_05.mpr,
connection (Fig. 2) which is described in                           and MPP_12.mpr,
technical note #23 [1]. Electrodes are as
follows:
       Bio-Logic Science Instruments, 1 rue de l'Europe, F-38640 Claix - tel: +33 476 98 68 31 – Fax: +33 476 98 69 09    1
                                                   Web: www.bio-logic.info
III- Results                                                                                                     LP_12.mpr
                                                                                                                              <I>, log spacing vs. Ew e
                                                                                                                              LP_01.mpr          LP_03.mpr           LP_04.mpr #
 III-1 LP investigations                                                                   1e0
LP investigations are performed in the anodic
                                                                                           1e-1
area using slow scan rate (Fig. 3) in order to
be under steady-state condition [2]. Moreover,                                             1e-2
                                                                     <I>/mA, log spacing
it is important to note that the E range is
                                                                                           1e-3
reduced from -2 V to 2 V in order to optimize
the potential resolution to 100 µV. Measured                                               1e-4
current is averaged in order to reduce noise.
                                                                                           1e-5
                                                                                           1e-6
                                                                                           1e-7
                                                                                                         -1.5                      -1                                -0.5
                                                                                                                                            Ewe/V
                                                                    Fig. 4: Four-electrode LP curves.
                                                                    An example of data determination is shown in
                                                                    Fig. 5. All results are summarized in Table 1.
                                                                                                                                   log ( <I> ) vs. Ew e
                                                                                                                             LP_03.mpr #           LP_03_tafel.mpp
                                                                                                  -1e0
                                                                                                  -2e0
                                                                                 log ( <I>/mA )
                                                                                                  -3e0
Fig. 3: “Parameters Settings” window for LP                                                       -4e0                                                        Ep = -742 mV
measurement.
                                                                                                  -5e0
From LP curves (Fig. 4), it is possible to
                                                                                                  -6e0
determine the corrosion potential (Ecorr) and
other related parameters, such as corrosion                                                       -7e0
current (Icorr), cathodic (βc) and anodic (βc)
                                                                                                  -8e0                                       Ecorr = -1,041 mV
Tafel constant and corrosion rate with the                                                                -1.5                       -1                                -0.5
“Tafel Fit” analysis tool. Moreover, the pitting                                                                                             Ewe/V
potential (Ep) can be directly determined by
tracking the cursor on the graph and by right
clicking on the pitting potential (Fig. 5).
Analysis of Table 1 shows similar Ecorr for the
four electrodes included between -1,152 and
-1,000 mV vs. Ag/AgCl with -1,078 mV as
mean. Concerning Ep, values are included
between -760 and -740 mV vs. Ag/AgCl with
an average of -747 mV vs. Ag/AgCl.
                                                                    Fig. 5: LP curves of the electrode number 3 and
                                                                    its corresponding “Tafel Fit” window.
       Bio-Logic Science Instruments, 1 rue de l'Europe, F-38640 Claix - tel: +33 476 98 68 31 – Fax: +33 476 98 69 09                                                             2
                                                   Web: www.bio-logic.info
  Table 1: Data resulting from LP investigations                                                          MPP_12.mpr
                                                                                                                       <I>, log spacing vs. Ew e
                                                                                                                       MPP_01.mpr        MPP_03.mpr   MPP_04.mpr #
  (Potentials are given in mV vs. Ag/AgCl).
                                                                                                   1e-1
Electrode            1           3             4            12
                                                                                                   1e-2
 Ecorr/mV         -1,152       -1,035       -1,124         -1,000
                                                                                                   1e-3
                                                                             <I>/mA, log spacing
  Icorr/nA         0.12         0.12         0.15           0.13
                                                                                                   1e-4
  βc/mV             265         306          290            323
                                                                                                   1e-5
  βa/mV             362         311          384            328
                                                                                                   1e-6
 Corrosion        -3      -3                         -3             -3
           0.13x10 0.12x10                0.15x10         0.13x10
rate /mmpy                                                                                         1e-7
  Ep/mV            -760         -742         -746           -740
                                                                                                   1e-8
                                                                                                             -1.5                                     -1
                                                                                                                                    Ewe/V
    III-2 MPP investigations and Statistics
                                                                          Fig. 7: MPP curves of the four electrodes.
  Parameters of MPP investigations are similar
  to those of LP investigations (Fig. 6).                                 Ep and Ecorr are determined simultaneously via
  According to the results of LP investigations,                          the “MultiPitting Statistics” function (Fig. 8).
  Ip, which is the limiting current, is set to 50 µA.                     Note that Eoc of the “Multi Pitting Statistics”
                                                                          corresponds to Ecorr. Ep and Ecorr computed
                                                                          thank to this statistics tool, are in agreement
                                                                          with those obtained with Tafel Fit from LP
                                                                          results. Indeed, Ecorr averages are -1,078 mV
                                                                          and -1,202 mV vs. Ag/AgCl for LP and MPP
                                                                          investigations respectively, and Ep averages
                                                                          are -747 mV and -716 mV vs. Ag/AgCl for LP
                                                                          and MPP investigations respectively.
  Fig. 6: “Parameters Settings” window for MPP
  measurements.
  MPP curves (Fig. 7) display the current shift
  due to the pitting process.
                                                                          Fig. 8: “Multi Pitting Statistics” window in
                                                                                                   ®
                                                                          “Analysis” menu in EC-Lab software.
             Bio-Logic Science Instruments, 1 rue de l'Europe, F-38640 Claix - tel: +33 476 98 68 31 – Fax: +33 476 98 69 09                                         3
                                                         Web: www.bio-logic.info
IV- Conclusion                                                      References
This note exhibits some corrosion techniques                        [1]     Technical    Notes     #23,           http://www.bio-
(LP and MPP) and corresponding analysis                             logic.info/potentiostat/notes.html
processes (“Tafel Fit” and “MultiPitting                            [2] Traité des Matériaux, Vol. 12 Corrosion et
Statistics”). These tools are complementary.                        chimie de surfaces des métaux, D. Landolt, ed.
                                                                    Presses     Polytechniques    et  Universitaire
Indeed, “Tafel Fit” allows the user to obtain the
                                                                    Romandes, 2003.
corrosion rate and other typical corrosion
parameters, whereas “MultiPitting Statistics”                       [3] Cinétique électrochimique, J.-P. Diard, B. Le
                                                                    Gorrec, C. Montella, ed., Hermann (Paris) 1996.
allows     the   user     to    compute      data
simultaneously from several data files.
Other investigations can be carried out with
EC-Lab® software such as Critical Pitting
Temperature with a thermostat controlled via
EC-Lab® software.