Biomixer BM 330: Operator's Manual
Biomixer BM 330: Operator's Manual
BM 330
Operator’s Manual
Important notes
Warranty
Ljungberg&Kögel AB hereby warrants to the original purchaser that Biomixer BM330 is manufactured in a professional and
quality manner and, when properly used, will be free from all defects in material and workmanship, for a period of 12 months
after delivery from Ljungberg&Kögel AB. The warranty includes equipment or components that proves to have faults during
the warranty period.
Ljungberg&Kögel AB will without cost for the customer, repair or replace the faulty equipment. The warranty is not valid if
the equipment has been repaired by anyone else than qualified personnel, approved by Ljungberg&Kögel AB. The warranty
is not valid for the battery since it is considered to be an expendable supply. The warranty is not valid if the equipment has
been changed in any way that according to Ljungberg&Kögel AB:s opinion, affects the reliability or stability of the
instrument. The warranty is not valid when the serial number has been changed, crossed-out or been removed, or if the fault
has been caused by misuse or abnormal use. In these cases Ljungberg&Kögel AB or their representative will inform the
customer about the decision, and if wished by the client, will repair the equipment for normal rate. An estimated price can be
given on request. One important condition for this warranty to be effective is that the BM330 is used in accordance with the
instructions in this operators manual.
Abelko Innovation is committed to develop high-quality equipment and technical services to all our customers.
We welcome any inputs on technical issues that are encountered so that they can be resolved quickly and in the
most appropriate manner. Please submit your comments/feedbacks through your local distributors or
alternatively email us directly at info@abelko.se
BM330 is developed and manufactured by Abelko Innovation, Industrivägen 17, SE- 972 54 Luleå Sweden.
Tel +46-(0)920-220360. Fax +46-(0)920-220068. Email: info@abelko.se, www.abelko.se
Abelko Innovation also manage logistics, service and technical support.
BM330 is owned and marketed by Ljungberg & Kögel AB, Box1032, SE-251 10 Helsingborg,
Tel+46-(0)42-139860, fax +46 (0) 42-132181, info@ljungberg-kogel.com , www.ljungberg-kogel.com
Spare parts are to be ordered by Ljungberg & Kögel.
2
BM 330 Operator’s manual
Revision history:
Rev 2005-12-23: Page 13: Collection may continue if something falls on tray. Collection time is measured until pressing Pause. Page 19: Do not use
bleach. Page 21: Collected volume 50-999ml
Re 2007-02-02: More languages included, A-version English, Spanish, Swedish, B-version English, Swedish, Danish. Page 2 : Battery excluded from
warranty. Page 9 : Min flow before alarm now settable. Page 11 : Blocking function to press START before scanning barcode introduced. Page 12-13
Different Notes. Page 22 Cable mixer-junction box now included.
Rev 2008-03-14: Signal converter and LAN material has been removed from Spare Parts page 22. See separate spare parts for USB Start Kit for network.
Rev 2008-09-08: Page 10 &19, position of reference weight
Rev 2009-08-21: Page 18, Up to 31 mixers left out, Ritex and Westermos signal converters expired and are left out. Page 7: Battery charger BM12-12
is left out. Spelling control with some changes has been made in the whole document. Page 23, change year of directive. Page 21
technical data battery charger changed
Rev 2010-02-24 Page 12, Warning for use of mobiles removed, not needed any longer.
Rev 2010-08-09 Page 2, elucidation of manufacturer and marketing companies, Page 23, change back year of directive
Rev 2012-11-13 Page 2, Added Abelko Innovation commitment concerning technical feedback
Rev 2013-06-0 Page 22: 10 hours operation time changed to 12 hours
Rev 2014-04-17 Page 24: Updated address of DoC.
Page 8: Added note about transport and placement
3
BM 330 Operator’s manual
Table of contents.
Important notes .......................................................................................................................................... 2
Warranty ................................................................................................................................................. 2
The operator’s manual ............................................................................................................................ 2
Revision history: ..................................................................................................................................... 3
System description ...................................................................................................................................... 5
Programmability ..................................................................................................................................... 5
Network connection ................................................................................................................................ 5
Flexibility................................................................................................................................................ 5
Features ....................................................................................................................................................... 5
Connections ................................................................................................................................................ 7
Installation .................................................................................................................................................. 7
Battery charger (mains adapter) .............................................................................................................. 8
Barcode reader ........................................................................................................................................ 8
Local Area Network (LAN) .................................................................................................................... 8
Battery charging...................................................................................................................................... 8
Start-up ................................................................................................................................................... 8
Keyboard functions ..................................................................................................................................... 9
Setup ......................................................................................................................................................... 10
Collection without barcode reader registration......................................................................................... 13
Power on ............................................................................................................................................... 13
Setting of desired volume ..................................................................................................................... 13
Start of collection. ................................................................................................................................ 13
Pause during collection. ........................................................................................................................ 14
Termination of blood collection ........................................................................................................... 14
Collection with barcode reader ................................................................................................................. 15
Registration before collection. .............................................................................................................. 15
Registration after collection.................................................................................................................. 16
Send data to PC ........................................................................................................................................ 17
Alarm and error messages ........................................................................................................................ 17
Alarm and error messages before and during collection ...................................................................... 17
Remaining error messages .................................................................................................................... 18
Network connection .................................................................................................................................. 19
Service and maintenance .......................................................................................................................... 20
Cleaning ................................................................................................................................................ 20
Calibration (check of weighing accuracy) ............................................................................................ 20
LCD contrast setting ............................................................................................................................. 20
Change of battery................................................................................................................................. 20
Tray parking position ............................................................................................................................ 21
Trouble shooting and actions ................................................................................................................ 21
Technical specifications ........................................................................................................................... 21
Cable to programming PC .................................................................................................................... 21
Cable to registration PC........................................................................................................................ 21
Technical data ....................................................................................................................................... 22
Spare parts list .......................................................................................................................................... 23
MDD Declaration ..................................................................................................................................... 24
4
BM 330 Operator’s manual
System description
Programmability
BM330 is today the most complete and flexible blood mixer offering a complete system for blood
collection and registration to meet the demands of today and in the future for total quality control and
complete traceability.
It’s unique programming capability makes the blood enters free to choose their own strategy for
collection monitoring and later change it according to new global or local regulations or after their own
wishes.
Network connection
BM330 is also prepared to be connected to a local area network (LAN) for transferring registration and
collection data to a central host computer, so called ”On-line registration”.
This will minimize the risk of exchanging blood bags and offers new and improved possibilities to
quickly adapt the succeeding blood analyze in an optimal way.
Flexibility
Since BM330 is powered by a mains adaptor or an internal rechargeable battery with 12 hours capacity
it can also be used in a conventional way, with or without barcode-reader, as stationary as well as
mobile collection units.
Features
Barcode-reader connection
For quickly registration of donor, blood bags, operator etc.
Network connection
Immediate transfer of scanned barcode information together with collected data i.e. volume, flow, time,
alarms and events to a central PC or host computer.
5
BM 330 Operator’s manual
Large display
Large LCD display with 2x20 characters for continues display of set volume, collected volume, flow
and time during collection. The display also shows alarms and guides the operator during the
registration. The text on display can be selected in different languages.
Automatic calibration
Calibration is done automatically by using a calibration weight.
Automatic taring
BM330 automatically deducts the weight of the blood bag and anticoagulant solution. All types of
bluebags or set of blood bags can be used.
Homogeneous mixing
The tray is moving in a well-tested seesaw motion to agitate and mix blood and anticoagulant in an
efficient way.
Automatic stop
When the predetermined volume of blood has been collected the tube clamp closes automatically and a
ready-signal sounds to inform of the completion of donation sequence.
6
BM 330 Operator’s manual
Connections
Installation
• All connections to BM330 must be done with the power switch in ”OFF” position.
• If BM330 has been cooled down to a temperature below +2 ºC, please wait at least 2 hours before
start up to make the damp evaporate.
7
BM 330 Operator’s manual
Barcode reader
Connect the barcode reader to the RS232 connector.
Always use barcode readers recommended by Ljungberg&Kögel AB. Connection of other barcode
readers may damage the reader as well as the mixer!
Battery charging
If the mixer is to be used mobile (without access to mains supply) the battery must be charged. Connect
the charger and charge the batteries for at least 6 hours if battery is emptied to <20%.
The capacity is shown on display and when it drops to less than 30% it should be recharged.
When capacity drops below 20% start of collection is prohibited and an alarm is activated (see also the
chapter Alarms and error messages . Started collection can however be completed.
If BM330 is to be used stationary with access to mains supply the charger should always be connected.
Placement
In stationary conditions, it is recommended to use the mixer with BIOLIFT 410 (separate manual).
BIOLIFT 410 is a mobile table, with a moving shelf (up and down) designed to fit the mixer.
Transport
Before transporting the mixer, it should be set in parking mode. In the main menu, press the
RESET key. The mixer should be transported in a transport case suitable for BM330 (available by
Ljungberg & Kögel).
Start-up
Put the switch in” ON”- position.
Now the BM330 functions are automatically checked and possible errors are indicated on display and
with an intermittent sound-signal.
The following checks are made:
• The clock and memory are checked.
• The stepping-motor operating the tube clamp are driven to its starting position (fully open).
• The tray makes one seesaw motion and stops in the weighing position.
• The display shows the start menu:
V2.18B (ENG,SWE,DAN)
Lab123
The upper row shows the current program version and the languages included and lower row shows
the mixer ID if it is programmed (see BM330 configuration program user’s manual).
8
BM 330 Operator’s manual
On the upper row date and time are shown together with the remaining battery capacity in %. The lower
row shows the set collection volume and the tube clamp position.
Keyboard functions
Start of collection. The tray makes a seesaw motion to secure that the tray is in weighing
START
position, taring occurs, clamp is opened and the mixing and time measuring starts. The
START-key is also used during calibration.
PAUSE Stops the mixing temporary during collection. If there is an alarm the sound-signal is
AL.OFF silenced with the first pressing and after that the mixing can be stopped with another
pressing. The display shows that mixing is stopped with the text PAUSE.
This button is also used to check calibration by putting the tray in constant weighing
position when activated from the main menu.
CLAMP Opens and closes the tube clamp. Display will show CLAMP OPEN respective CLAMP
CLOSED.
NEW Starts the collection registration when barcode registration is activated. Pre-programmed
scanning sequence is shown on display and guides the operator to correct barcode
scanning.
Used for setting of desired collection volume. Every press on SET switches between two
SET preset volumes. The pre-setting’s can be done before collection with the keys ”+” or ”-”
ending with SET. ”SET” is also used for some setup adjustments.
Stops already started registration with barcode reader. Can also be used to end a started
RESET
collection when being in PAUSE-mode. Also stops settings in setup-menus for return to
main menu. A press on RESET in main menu makes the tray take its parking position.
Used for setting of alarm limits, clock adjustment, calibration, barcode reader,
SETUP
communication parameters etc.
Setup
Settings of alarm limits and functions are done in SETUP-menus which are in English no matter of
choice of language for the collection menus. Scrolling through the menus is done with SETUP key.
Settable values or functions are shown blinking and increasing/decreasing is done with the keys + and -.
Scrolling between different input areas in the same menu is done with the SET key. The lower row gives
a guide for what keys to press. The SETUP can be stopped with RESET at any moment.
This are the existing SETUP menus:
• Clock settings.
• Min collected volume before flow alarm
• Alarm limit low flow.
• Alarm limit high flow.
• Alarm limit collection time.
• Sound level for alarm signal, (buzzer).
• Calibration.
• Communication speed.
• Barcode reader connected.
• Obligation scan before start
• Programming mode.
• Choice of language.
Note ! Changed values and functions are saved in memory when next menu is selected with ”SETUP”
or when SETUP-function is interrupted with the RESET key.
The mixer has a real time clock for precise time registration. Set the right minute, hour, year, month and
day. Confirm the settings with a press on ”SET” .
10
BM 330 Operator’s manual
The sound level from the buzzer is set with a value between 0-31. The buzzer starts sounding when
”+”or ”-” are pressed and sounds until the menu is left. Setting a value less than ca 12 can cause the
buzzer to cease.
Make sure the tray is empty and moves freely. Start the calibration by pressing ”START”, after which
zero balancing and taring is done. The display shows:
Put the reference weight for 450 ml on the tray, (make sure that the reference weight is placed in the
centre of the tray), and press ”SET”.
Note! The reference weight must be 450 ml i.e. 477g. (450x1.06). Reference weights can be ordered
from Ljungberg & Kögel AB.
TARE=102 VOL=441>450
OK PRESS SETUP
The calibration is now performed automatically and the result is shown on display.
The value TARE is the measured tare (dead weight) and should be between 50 and 150. (This value
corresponds to the weight of the scale after automatic zero positioning and is only of technical interest
if there should be any fault during calibration process).
VOL is the read volume and should be close to the referens weight 450 ml. The ratio for 450/(read
volume) gives a conversion factor.
The BM330 microprocessor automatically adjusts the read volume to be 450 ml by multiplying with this
conversion factor and the calibration is done when ”SETUP” is pressed.
If the divergence is to big (more then ±20%) an error message is shown :
TARE=102 VOL=349>450
ERR ADJUST GAIN
11
BM 330 Operator’s manual
If this happens, manual adjustment with the potentiometer GAIN must be performed. This GAIN-
potentiometer is accessible behind a plastic cover on the side of the mixer. Remove the plastic cover and
adjust with a small screwdriver until the read value becomes 450 ml. Remove the reference weight and
make a new taring by pressing once on RESET and after that START again. Put the reference weight
back again and repeat adjustments until the read value is within ±5ml.
Note! Always end with automatic calibration, without adjusting GAIN, to get the correct conversion
factor.
This settings are for both the barcode reader (RS232) and the local network (RS485).
BARCODE-READER YES
Barcode reader connected. CHANGE WITH +/-
PROGRAM MODE
Program mode. ENTER WITH START
Activate the program mode with a press on ”START”. The display will show:
PROGRAM MODE
SEND SETUP FROM PC
Now the mixer is ready for programming of the scanner sequence from the programming PC through the
RS232 input (see also: BM330 configuration program user’s manual).
Note! By pressing ”SEND” at this moment the actual scanning sequence settings in the mixer can be
sent to the PC.
Language. LANGUAGE:ENGLISH
CHANGE WITH +/-
Note! The language setting is only for collection menus, not the setup menus.
12
BM 330 Operator’s manual
A press on ”SET” makes the set volume change between the two preset volumes.
With a press on ”+” or ”-” this values can be changed. The set volume starts to blink on display and
further presses on ”+” or ”-” makes the value increase or decrease. If the key is pressed continuously
more then 1 second a rapid change with 10-ml each step will occur. To replace the earlier preset value
the changed settings must be confirmed with a press on ”SET”.
Start of collection.
• Place the blood bag or sets of blood bags on the tray so that the tube inlet of blood bags is upwards
when the tray is in home position.
• Put the tube in the tube clamp and leave at least 25 cm of tube free between the clamp and the tray
allowing the tube to move with no affection to the weighing accuracy.
• Press ”CLAMP” to close the tube clamp. CLAMP CLOSED is shown on display.
13
BM 330 Operator’s manual
• The mixing will continue again with another press on ”PAUSE AL.OFF” and the clamp will open
automatically if closed.
• In the PAUSE-state the current collection can be interrupted with a press on ”RESET”. To avoid
interruption by mistake it has to be confirmed with an additional press on ”RESET” according to the
menu:
STOP COLLECTION?
YES:RESET NO:SET
Note! If “CLAMP” is not pressed within 2 minutes from collection stopped the clamp will open
automatically and the mixer will make one cycle to ensure the blood in the tube is not coagulated. This
will increase the collected blood volume to be more than set volume. This will be repeated every two
minutes until “CLAMP” is pressed.
• Data from the collection are transferred to the registration-PC with a press on ”SEND”. If there is no
registration-PC connected, collection is concluded with a press on ”RESET” instead. The collection
data is stored in the memory and can later be transferred to a registration-PC.
• During the collection all possible errors are also registered. Those can be taring error, motor error,
low battery, flow fail or time fail. More about error handling in the chapter: Alarm- and error
messages,
14
BM 330 Operator’s manual
Note! Collection data are always registered and stored in BM330 irrespective of the fact that using
barcode reader or not. Data can be transferred after each collection or after any number of collections.
When the memory is filled the oldest collection data are overwritten. See chapter ”Send data to PC”.
Note! If something falls on the scale making the set volume reached, the collection can be continued
after removal of the object and pressing “Start” button.
SIGNATURE 1?
MRO OK The operators ID from barcode.
REGISTRATION READY!
VOL=450,CLAMP OPEN
15
BM 330 Operator’s manual
The registration can also be stopped by turning the power off. In both cases already scanned, but not
completed, information is erased.
When registration before collection is done the blood collection can proceed in the same way as in
chapter: Collection without barcode reader registration.
EXTRA TESTS? Repeated scanning has been selected to specify some additional
42 BLOOD GROUPING OK tests to be done with the blood. These are read from the
barcode table.
EXTRA TESTS?
43 ANTI-HBC OK
EXTRA TESTS?
45 ANTI-HTLV OK
Note! A press on ”RESET” makes the mixer return to main menu and registered data is stored in
memory. The stored data can anytime be sent to a PC through the local area network or directly.
16
BM 330 Operator’s manual
Send data to PC
• All stored information can be transferred to the LAN (Local Area Network) after completed
collection by pressing ”SEND”. ”SEND” can also be activated from main menu. Display shows:
DATA TRANSMIT ACTIVE
• Each separate collection is sent in a data batch and the transfer is acknowledged by registration PC.
• When all collection data is sent the display shows:
• Up to 31 mixers can be connected to the LAN (RS485). When SEND is activated from several
mixers at the same time the transfer will be queued to avoid collision. If transmit is in progress from
one mixer the other waits until the line is free. The display shows:
• If the line is free and a data batch is sent to PC the mixer expects an acknowledgement from the PC.
If no acknowledgement arrives in spite of 3 repetitions an error signal sounds and an error message is
shown:
DATA TRANSMIT ERROR
Note! If transmission by any reason fails the stored data are preserved in memory and will be sent the
next time a transfer is activated.
Battery alarm
If the battery capacity drops below 20% a warning is given with a sound signal and a blinking text :
”BATTERY<20%” . If a collection has started it can be completed but no further collections can be
started without battery recharging. Charge of battery can be done at the same time as collection.
If a battery alarm occur the battery should at once be recharged. If recharging can’t be done the
mixer has to be turned off because deep discharge is shortening the batteries length of life and
storage collection data in memory may get lost.
17
BM 330 Operator’s manual
Taring error
Automatic tare deduction is always done before collection. If this fails an error message is given with
blinking text and sound. Collection can not proceed. Check that the tray is in its weighing position and
moves freely.
Motor error
If the mixing motor do not start or not find the home-position (weighing position) there is an error
message and collection is stopped. Check that the tray can move freely.
Flow alarm
If the blood flow is below or exceeds the set limits an alarm is given with FLOW blinking in the
collection main menu and an intermittent signal sounds. Flow alarm will be stored in collection data
memory. Flow alarm are prohibited during the first 50 ml of collection because blood flow disturbance
is liable to occur.
Time alarm
If collection time exceeds set time limit, an alarm is activated with TIME blinking on the collection
main menu and an intermittent signal sounds. Time alarm is stored in collection data memory.
Volume=EEE
Indicates overweight, measuring range is exceeded. Check if something is weighing down the tray.
Volume= - - -
Indicates underweight. Measuring range is below limit. The tray has become lighter since tare deduction
(by any reason) or is touching something when it is in the weighing position.
Setup error.
When the mixer is powered on all settings and functions are checked. If the setup, stored in non-volatile
memory (EEPROM), has been changed i.e. checksum is not the expected, ”SETUP ERROR” is
displayed and a new setup has to be done. See chapter Setup.
Data error.
Stored data from collections are checked at power on. This data are stored in a memory supplied from
the built in battery and from a backup-capacitor (BUC). If the battery voltage is to low or if it is
disconnected for more than 24 hours there is a risk for lost data. ”DATA ERROR” can be restored by
turning the power off and on again.
Input error.
If the clock is set to an unreasonable value, for instance 31 Feb., this can not be accepted and ”INPUT
ERROR” is indicated.
Clock error.
The clock is repeatedly checked and if there is any fault detected ”CLOCK ERROR” occurs. The fail is
probably restored after a new setting.
18
BM 330 Operator’s manual
Network connection
BM330 mixers can be connected in a LAN to a common registration PC through an USB/RS485 or
RS232/RS485 signal converter. Every individual collection can be transferred without delay to the PC
with the windows program BM330 Information program running for further processing.
The LAN consists of suitable located junction boxes for the mixers wired together with a twisted pair
cable. The mixers connects to this boxes with a 4-parts cable with modular contacts.
See chapter: Technical specification for the cable detail description.
USB/RS485 or
RS232/RS485 Registration-
JBn JB2 JB1 Signal PC
converter
R S485 R S485
lin e o u t lin e in
GND B A GND
p1 p2 p3 p4
M o d u la rja c k 6 /4
to B M 3 3 0
Note! The signal converter shall have galvanic isolation between RS485 line and USB/RS232 line. It is
the person who installs the LAN that is responsible for the installation and that connected external
equipment (converter and PC) comply with relevant safety standard, EN60950, and that the whole
system comply with EN 60601-1-1, ”Safety requirements for medical electrical systems”. If in doubt,
contact qualified technician or Ljungberg&Kögel.
19
BM 330 Operator’s manual
Do not use liquids with bleach since this may affect the plastic cover quality and make it fragile.
Notes!
• In this ”continuous weighing mode” the mixer operates like a scale, showing the weight placed on
tray adjusted for the specific gravity of blood ( /1.06).
• If the mixer have been exposed to a big temperature difference, e.g. been moved from a cold
environment to a warm, it must have some time to obtain a stable temperature.
Change of battery
When the battery no longer is sufficient for one day collections it has to be changed.
• Turn the mixer off with the switch ON/OFF and pull out the contact from the charger.
• Place the mixer so it supports against the pillars on the backside.
• Unscrew the 2 screws holding the battery hatch.
• Take out the battery and change it to the new one that must be of the same type : Sonnenschein A512
3,5S or of a similar type. Connect the red wire to + and the black to -
• Put the battery back and screw up the hatch.
The battery contains lead and worn out batteries must be left to a recycling station.
20
BM 330 Operator’s manual
Technical specifications
Cable to programming PC
A null modem-cable shall be used for PC-connection via RS232.
To BM330 Signal- To PC To PC
9-pole D-sub female direction 9-pole D-sub female 25-pole D-sub female
(1)
2 ( RX ) ⇐ 3 ( TX ) 2 ( TX )
3 ( TX ) ⇒ 2 ( RX ) 3 ( RX )
(4)
5 (GND) ⇔ 5 (GND) 7 (GND)
(6)
7 (RTS) ⇒ 8 (CTS) 5 (CTS)
8 (CTS) ⇐ 7 (RTS) 4 (RTS)
9 (+5V) ⇒
Note! To the same contact is the barcode-reader connected. Pin 9 in the 9-pole contact is the power
supply (+5V) for the barcode-reader.
Cable to registration PC
This PC is connected to the LAN via a USB/RS485 signal converter, type ADAM-4561 or other isolated
converter complying with safety standards for medical equipment. The converter must be strapped for
TX controlling the data flow from PC.
Junction boxes are mounted on the wall closed to each mixer, and the mixer connects to this box with a
4-pole straight cable with modular 6/4 in each end. See chapter Network connection.
This is the connection for the 4-pole modular cable:
We recommend twisted pair cable with shield and return wire, e.g. type Alpha Wire 5471C, ELFA art.
no. 55-752-04. The return wire and the shield shall be connected to p1, p4 (GND) in every junction box.
The cable for a direct connection of a BM330 to a 485 converter is shown below. The modular contact
is seen from contact side.
21
BM 330 Operator’s manual
RS232/RS485-Converter
Line-connection
P1
P2
B P3
A
P4
Signal ground
Technical data
Power supply......................................................... Mains adaptor, 100-240VAC, BM12-12, (14V/1A
(also operates as a battery charger) and/or internal
battery 12V/3.5Ah for approximately 12 hours
operation between charges.
Power consumption............................................... Max 10VA
Fuses (built in)....................................................... PTC (self recovery for overload protection)
Collected volume................................................... (Collected blood) 50-999 ml. (PCB Version -02)
Accuracy................................................................ <1% of max weight ±1 ml.
Taring range .......................................................... 0-600g
Mixing cycle..... .................................................... 16 cycles/min.
Data storage........................................................... 32.000 characters
Tube clamp operation............................................ Linear step motor
Mixing plate operation........................................... AC-driven synchronous motor
PC connection........................................................ RS232 for programming, 9-pole Dsub male
Network connection............................................... RS485 (multidrop), 6/4 modular
Barcode reader connection..................................... RS232, 9-pole Dsub male
Extra output............................................................ External display,6/6 modular, (optional)
Relative humidity incl. storage and transportation. 10-95 %, not condensing
Operating temperature............................................. +10 - + 40 °C.
Storage and transportation temperature................... - 40 - + 70 °C.
Dimensions ............................................................. 290 (L) x 253 (W) x 150 (H) mm.
Weight… ................................................................ 4.8 kg incl. battery.
Protective classification........................................... Class II, type B.
Symbol: .............................................................. Protection type B equipment
0413
22
BM 330 Operator’s manual
18 15 11
9 7 5
8
10
2
28
40
3
12 25 26 27 29 30 31 41 32 39 6 24 22
23
BM 330 Operator’s manual
MDD Declaration
DECLARATION OF CONFORMITY
according to the
LFVS 2003:11 corresponding to the Medical Devices Directive, 93/42/EEC
STANDARDS TO WHICH CONFORMITY IS DECLARED:
EN 60 601-1 (SEMKO certificate 9725103/01)
EN 60 601-1-2 (Enator test report TR 976119)
Including: EN 55011 Kl B, EN 61000-4-2, EN 61000-4-3, EN 61000-4-4, EN 61000-4-5
Marketing: Ljungberg&Kögel AB
Adress: Box 1032, 251 10 Helsingborg
Telephone: +46 042-139860
Telefax: +46 042-132181
I, the undersigned, hereby declare that the equipment specified above conforms to the above Directive
and Standards.
Signature:
/ Thommy Lundström
Position/title: President
24
BM 330 Operator’s manual
25