R300 Series (K34* & K35* Firmware)
ZERO/SPAN Calibration Procedure
Start with indicator powered on and in normal weighing mode
Press (Power & “f” key) together for 2
1
+ seconds to access full setup mode
2 Press ZERO key (2X)
2x
3 Press TARE key (1X)
1x
Remove all weight from scale
4
1x Press SELECT Key (1X)
5
Press PRINT key (1X)
Z. in P
1x Captured Zero Value:
6
Press TARE key twice (2X)
1x
7
Press F key twice (2X)
Clear Linearity; Y=Yes/N=No
8
Add Weight Add weights to scale (live load showing)
9
Press PRINT key (1X)
1x
10 Enter Weight Value:
Use the Print key to change digit Use the
1x Select Key to move digit
11
Press the F key to Calibrate (1X)
S. in P
Captured Span Value Showing:
12
Press (Power & “f” key) together for 2
+ seconds to save and exit.
R300-681-110