R300 Series Calibration Guide

The document provides a 12 step procedure for performing a zero/span calibration on an R300 series scale with K34* or K35* firmware. The procedure involves pressing specific keys in sequence to access calibration mode, capture the zero value with no weight, add a known weight value, capture the span value, and save and exit calibration mode.

Uploaded by

Ionel Racaru
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
85 views1 page

R300 Series Calibration Guide

The document provides a 12 step procedure for performing a zero/span calibration on an R300 series scale with K34* or K35* firmware. The procedure involves pressing specific keys in sequence to access calibration mode, capture the zero value with no weight, add a known weight value, capture the span value, and save and exit calibration mode.

Uploaded by

Ionel Racaru
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 1

R300 Series (K34* & K35* Firmware)

ZERO/SPAN Calibration Procedure

Start with indicator powered on and in normal weighing mode

Press (Power & “f” key) together for 2


1
+ seconds to access full setup mode

2 Press ZERO key (2X)


2x

3 Press TARE key (1X)


1x

Remove all weight from scale


4
1x Press SELECT Key (1X)

5
Press PRINT key (1X)
Z. in P
1x Captured Zero Value:

6
Press TARE key twice (2X)
1x
7
Press F key twice (2X)
Clear Linearity; Y=Yes/N=No

8
Add Weight Add weights to scale (live load showing)

9
Press PRINT key (1X)
1x

10 Enter Weight Value:


Use the Print key to change digit Use the
1x Select Key to move digit
11
Press the F key to Calibrate (1X)
S. in P
Captured Span Value Showing:

12
Press (Power & “f” key) together for 2
+ seconds to save and exit.

R300-681-110