7Programming and Calibration
In this section you will learn how to program BX 1 according to your application,
Calibrations are explained. Symbols at the bottom right of the device
Shows the functions during programming. The basics of the Russians
The meanings are summarized in the table below
7.1 Introduction to the Programming and Calibration Menu
To enter the programming and calibration menu, press the <F> key followed by Info
We press. The display will show [FunC] and [PASSWr] respectively. Service
Enter the operator password (T) and press <Enter> key
Program. To access the parameters that can be entered with the service password,
The calibration jumper must be installed (see section 2.2). the password
If not entered, it is returned to the weighing mode. With operator clearance
Only the parameters related to the use can be changed. With service mode
The parameters that can be changed are displayed in the flow field on the next page;
The calibration jumper requirements are indicated by black squares in the upper right
corners.
The Programming and Calibration menu displays the main blocks in the form [X--] and the
bottom
Build from blocks. And  typing in the next yada to the previous main blog
accessible. When the main block you want to enter is selected and then press Enter,
The first sub-block within the block is accessed and the display [X0-] appears. And  tuna
To reach the desired sub-blog and press Enter to select the sub-block
You can enter the first parameter. Indicating the first parameter of the selected sub-block
The message [XY0] appears. In the same quadrant and in the next yada
You can reach the previous parameter. To change the parameter values, press 0?
Use. To enter a numeric value, use the <T> button to select the digit you want to change,
Change the value of this digit with the key.
7.2 Exit from the Programming and Calibration Menu
While at any stage within the Programming and Calibration menu, press <F>
When pressed, a top block is reached from the end point (parameter bottom block).
When the <F> key is pressed while in the main blocks, the message [SAVE] is displayed.
Press <Enter> to save changes made in the programming menu to memory,
To save the changes made until the device is switched off, or
The <F> key is pressed to ignore the changes made. In these three cases
The display will briefly display the message [WAIT] and the BX 1 returns to the weighing
menu.
Especially in legal metrological use, after the setting,
Do not forget to remove the calibration jumper.
With the INDFACE program you can change the settings of your device from your computer
and
You can perform operations
HRKT DDK.
21 _
[31-] Calibration Correction Parameters
In this section only zero gain adjustments can be made. Thus, the entire calibration process
There is no need to repeat.
[310] Zero Calibration
This parameter is only to reset the zero setting. Press <Enter> for
Is entered into the parameter. When the display [ZE.CO] message appears, the scale
The kefesi is emptied and the Enter key is pressed. Indication during zero setting [WAIT]
Message appears. In this process, there should not be a load on the weighing instrument,
It must be ensured that it is stable. The indication [SAVE] appears after the measurement.
Set zero is loaded into memory by pressing <Enter> key.
<F> key to exit without saving the transaction.
[311] Gain Calibration Setting
With this parameter, the gain setting of the device is renewed. By pressing <Enter>
Displayed [XXXXXX] posts appear. This number is set for gain calibration
Is the required load value. Display weight is a different calibration weight than the displayed
value
, The weight of the calibrated calibration is adjusted with the help of the <Tare> and <Zero>
The desired weight value is entered and the weight is placed on the scale and press <Enter>
Printed. As in zero calibration, you can gain as much as 10 seconds here
Calibration is done and the display shows [SAVE] at the end.
The gain setting made is loaded into the memory by pressing the <Enter> key. Calibration
<F> key to exit without saving the transaction.
[312] Load Gain Calibration Setting
In this step, without lifting the current load of a weighing instrument under load,
Is the gain adjustment made by loading the weight required for calibration. It
Parameter facilitates gain calibration especially for non-empty tanks.
When entering this parameter, the device will first determine the current state,
The instrument should not be touched and the calibration weight should not be added.
Press <Enter> key in this parameter, the display will show [P.ZEro]. It,
BX 1 is the stage of determining the current state of the weighing instrument. Temporary
Zero setting
Press <Enter> when ready for this phase. [WAIT] message
There shall be no additional load on the weighing instrument and the stability of the
weighing instrument
It must be provided. The device that detects the current state switches to gain calibration.
The display will briefly read [LoAd] and then add the calibration
[XXXXXX], which is the load value that should be maintained. Set if necessary
Calibration weight is entered with <Dara> and <Zero> keys, additional load on scale
And then press <Enter>. Earning up to 10 seconds
The [SAVE] message appears on the BX 1 indicator, which completes the calibration setting.
The gain setting made is loaded into the memory by pressing the <Enter> key. Calibration
<F> key to exit without saving the transaction.
[313] eCal Calibration
Caution: Before starting to eCal, the capacity and division from parameter [212]
Enter the value.
This parameter allows us to calibrate your scale metric without using any test weight
It provides. Weighing instrument capacity, total load cell capacity, load cell output and
weighing
The tool dead load values are the values we need for this calibration.
Automatic zero instead of entering dead load value if conditions are appropriate
We recommend you to calibrate. Press the <Dara> key for input.
[LC.CAP]
[XXXXXX]
Total Load Cell capacity is entered in kg in <Dara> and <Zero>
Press the <Enter> key to move to the next step.
Example: If the weight system contains 4 1000 kg Load Cells, LC.CAP = 4000.
[LC.oUt]
[XXXXXX]
The output value (mV / V) specified in the load cell certificate can be set with the <Tare> and
<Zero>
Enter. If there are more than one load cell in the system, the average load cell output
Value must be entered. Press <Enter> to move to the next step.
Example: Output values of load cells in 4 load cell system:
Mean value for LC1: 2.0010, LC2: 1.9998, LC3: 1.9986 and LC4: 2.0002:
LC output = (2.0010 + 1.9998 + 1.9986 + 2.0002) / 4 = 1.9999 mV / V.
[Zer.adj]
[XXXXXX]
Press <Enter> key to perform zero calibration. If this step is empty and zero
Calibration should be performed if possible. Displaying about 10 seconds
[WAIT] the message came to mind - now that the weighing instrument is empty and stable
- then [SAVE] step is passed.
If zero calibration is not possible, press  (Info) to enter the estimated dead load value.
Yada  (X10) to move to step [PrE-Ld].
[Pre-Ld]
[XXXXXX]
Enter the dead load value with the <Dara> and <Zero> keys and press <Enter>
Go to the next step.
[SAvE]
The calibration is done by pressing <Enter> key. Calibration
Press <F> to exit without saving the transaction.
Note: If you need to make a setting after the calibration is completed, either [310]
Set the parameter to zero; Or dead load value [read on empty balance
Value + dead load value you entered earlier].