Summit 325e Service Manual
Summit 325e Service Manual
                                            01496-000 Revision G 
                                               AUGUST 2005
http://slidepdf.com/reader/full/summit-325e-service-manual                                                  2/40
7/18/2019                                                                                      Summit 325E - Service Manual
TABLE OF CONTENTS
                   I.       INTRODUCTION ..................................................................................................2
                   II.      USER INFORMATION ......................................................................................... 3
                            C.    MAINTENANCE ........................................................................................... 7
                            D.    TECHNICAL INFORMATION ....................................................................... 9
G. CALIBRATION............................................................................................ 20
J. SCHEMATICS
                    
                   I. INTRODUCTION
                   The equipment, which is described in this manual, will perform reliably when installed,
                   operated, and maintained in accordance with the instructions of this manual.
                   This equipment is sold with the understanding that the user assumes sole responsibility
                   for radiation safety and that the manufacturer does not accept any responsibility for the
                   following:
                            1.      Equipment improperly installed.
                            2.      Equipment improperly used.
                            3.      Equipment improperly maintained or repaired.
                            4.      Equipment, which has been modified or altered in any way.
                            5.      Injury or damage to patient or other personnel for any of the above
                                    causes.
                   We are proud of our products and are confident they will provide many years of useful
                   and enjoyable service.
http://slidepdf.com/reader/full/summit-325e-service-manual                                                     4/40
7/18/2019                                                                    Summit 325E
                                                                                     325E- Service
                                                                                            X-Ray  Manual
                                                                                                     Generator
                 Summi t Industri es, Inc.                                    Model Numbers F280 / A700
                 This x-ray generator is both easy to use and very versatile. It consists of an x-ray
                 control and a high voltage generator (H.V. generator).
Refer to photo on page 6 for location of the following controls and indicators:
                         C) Bucky selector switch – Allows selection of either wall bucky, table bucky, or
                             no bucky.
                         D) Line check rocker switch – When depressed, this switch causes the kVp
                            meter (O) to indicate if the line voltage compensator (E) is adjusted properly.
                         F) Tube current (mA) selector switch – Rotary switch allows operator to select
                            the optimum tube current and focal spot for the desired technique.
                            H) X-ray exposure switch – Two level switch: Partial depression (PREP) causes
                               anode rotation and filament current boost. Full depression (EXPOSE)
                               initiates the production of x-rays.
                            I) Exposure time selector switch – Provides the operator capability to select any
                               one of 23 different exposure times.
                            J) Exposure interlock “ready” light – Indicator light lights when all exposure
                               interlocks are closed. (Collimator interlock, tube thermal protection, misc.
                               exposure interlock devices, etc.) When this light is “off”, exposures will not be
                               allowed.
                            K) Tube ratings exceeded light – Indicator light lights when internal electronic
                               circuitry detects that selected technique factors exceed the single exposure
                               ratings of the x-ray tube. Under these conditions, x-ray exposures are not
                               allowed.
http://slidepdf.com/reader/full/summit-325e-service-manual                                                          6/40
7/18/2019                                                                325E
                                                                 Summit 325E    X-Ray
                                                                             - Service   Generator
                                                                                       Manual
                   Summi t Industri es, Inc.                      Model Numbers F280 / A700
                                                      M
                                                             L
                                                                            K
                                                                                                     J
              A
                                                                                                         I
                                     D
http://slidepdf.com/reader/full/summit-325e-service-manual
                                                                                              G              7/40
                                                        E            F
7/18/2019                                                                            325E
                                                                             Summit 325E    X-Ray
                                                                                         - Service   Generator
                                                                                                   Manual
                   Summi t Industri es, Inc.                                   Model Numbers F280 / A700
                            1. Turn the system “on” by slowly snapping the power switch (A) to the right.
                               Power “on” light (B) should light.
                                Note: If you are unable to position the needle within this area, it is an
                                indication that the line voltage is not correct and the equipment should
                                not be used. Either wait for l ine voltage to change or call for service.
                            3. Determine the desired technique factors Note: Do not exceed the x-ray
                               tube ratings. Refer to x-ray tube rating charts provi ded with the x-ray
                               tube. If the technique factors are set beyond the x-ray tube limits “ E01”
                               shoul d appear in the digital display window , preventing exposure. 
                                Note: Proper warm-up of the anode must be done on a cold uni t prior t o
                                making exposur es. The specific details may be found in “ General
                                Information fo r Installing and Operating X-Ray Tubes” , that is includ ed
                                with each tube shipped. The warm-up procedure tags are attached to
                                each unit. Basically this procedure will reduce the chance of tube
                                damage due to t hermal shock.
                                Note: Select the desired tube current (mA) prior to selecting the desired
                                peak tube potential (kVp).
http://slidepdf.com/reader/full/summit-325e-service-manual                                                       8/40
                                If at any time a different tu be current is selected it wil l be necessary t o
7/18/2019
                                                                                       325E X-Ray Generator
                                                                                Summit 325E - Service Manual
                 Summi t Industri es, Inc.                                        Model Numbers F280 / A700
                        6. Observe the digital display (G) to confirm that the correct mAs has been
                           selected.
                        7. Select the desired peak tube potential (kVp) by observing the kVp meter (O)
                           while rotating the kVp selectors (M&N)
                        8. This generator is equipped with tube protect circuitry that compares the
                           selected technique factors to the “single exposure ratings” of the x-ray tube.
                           Exposures will only be allowed if the selected values are less than or equal to
                           the x-ray tube limits. Refer to the chart below for description of “E01” error.
                        9. If the x-ray system includes a bucky (wall or table or both) and it is desired to
                           use a bucky technique, select the appropriate bucky by depressing the bucky
                           selector switch. If a non-bucky technique is desired select the center “off”
                           position of the bucky selector switch.
                       10. This generator is also equipped with an indicator light identified as “exposure
                           interlocks ready” light, which monitors the status of all auxiliary exposure
                            interlock circuitry.
                            is not “on”,         When all will
                                         x-ray exposures  interlocks
                                                               not be are ready Such
                                                                      allowed.  the light will light.
                                                                                      interlocks      If the
                                                                                                   might  be light
                                                                                                              SID
                            interlock switches, tube thermal protectors, collimator interlock, etc.
                       11. Depress the exposure switch approximately halfway. This action will cause
                           the rotating anode of the x-ray tube to accelerate to full speed and the
                           selected filament of the x-ray tube to heat up to its proper emission
                           temperature. Depress the exposure switch fully and hold. A beep tone will
                            indicate the termination of the exposure, and the mA meter deflection will
                            provide a visual indication of x-ray production.
                             After the exposure terminates, but while the exposure switch is still
                             depressed, the digital display will flash “on” and “off” indicating the number of
                             x-ray pulses produced. This is primarily a service feature provided for the x-
                             ray serviceman to verify proper operation.
                   C. MAINTENANCE
                    
                   Periodic inspection and preventative maintenance of this equipment is necessary to
                   assure continued safe performance and compliance with applicable state and federal
                   regulations. The following maintenance program must be closely followed.
                            1. Verify that the knob on exposure time selector is mounted solidly and
                               indicates the proper value.
                            2. Verify that all push-button and rocker switches operate freely with no sticking
                               or hanging up.
                            3. Verify proper operation of all pilot lights.
                            4. Verify proper operation of audible and visual indicators of x-ray termination
                               and production respectively.
                            5. Verify the accuracy of exposure timer, tube current, and peak tube potential,
                              and adjust as necessary.
                           6. Perform reproducibility and linearity tests and verify compliance.
                           7. Inspect electro-mechanical contactor for pitting and/or burned contacts.
                              Replace if necessary.
                           8. Examine terminal connections of high voltage generator. All terminals should
                              be tight.
                           9. Examine oil level of H.V. generator and refill as necessary. There should be
                              approximately ½ to ¾” air space between the oil and the inside edge of the
                              generator cover.
                          10. Examine the plugs of the high voltage cables. They should be clean with no
                              signs of carbon tracking.
                                Note: This x-ray generator sho uld not b e altered or modified in any way
http://slidepdf.com/reader/full/summit-325e-service-manual                                                        10/40
                                as such action may invalidate its regulatory compliance and void the
                                                                                     325E X-Ray Generator
7/18/2019                                                                   Summit 325E - Service Manual
                   Summi t Industri es, Inc.                                    Model Numbers F280 / A700
                   D. TECHNICAL INFORMATION
                    
                   Rated line voltage – 240 VAC, 60 Hz., single phase.
Technique factors that constitute the maximum line current – 300 mA @ 125 kVp.
Generator rating-
                            100mA
                            200mA @
                                  @ 125
                                        kVp –– 2%
                                               4%
                                    125kVp
                            300mA @ 125kVp – 1%
Maximum Deviation
                   The following maximum deviation statements are broad and considered “worst case”
                   conditions.
                   The Maximum Deviation of the Peak Tube Potential during an exposure from its pre-
                   indicated value is + 20% of the kVp meter’s full-scale value. The measurement basis
                   upon which the peak tube potential is defined is a calibrated high voltage divider unit
                   such as the Machlett Dynalyzer.
Tube Current
                   The Maximum Deviation of the tube current during an exposure from its pre-indicated
http://slidepdf.com/reader/full/summit-325e-service-manual                                                   11/40
                   value is +10% of selected value within a kVp range of 50-125 kVp (inclusive) and +20%
                                                                                     325E X-Ray Generator
7/18/2019                                                                      Summit 325E - Service Manual
                   Summi t Industri es, Inc.                                    Model Numbers F280 / A700
Exposure Time
The maximum deviation of the measured exposure time from its pre-indicated values is
                                    + 1 pulse for selected times within the range of 1/120 – 1/10 sec.
                                    + 15% of selected times within the range of 3/20-6 sec.
                   The measurement basis upon which the exposure time is defined as the number of half
                   cycle impulses of the 60 Hz power line measured either at the primary of the high
                   voltage generator with an oscilloscope or in the secondary of the high voltage generator
                   by means of a high voltage divider unit and an oscilloscope.
                   The digital display of mAs provided is not a primary indication of technique factors. It is
                   intended solely as an auxiliary indication to verify the product of the primary control
                   settings: tube current and exposure time. As such, a statement of maximum deviation
                   is not provided.
http://slidepdf.com/reader/full/summit-325e-service-manual                                                       12/40
                                                                                        325E X-Ray Generator
7/18/2019          Summi t Industri es, Inc.                                   Summit
                                                                                 Model325E - Service Manual
                                                                                        Numbers      F280
                                                                                                      / A700
                   Our
                   will certified
                        not affectcontrols and generators,
                                    the compliance         whencomponents
                                                     of these   combined with
                                                                          whenother certified
                                                                                 these        components,
                                                                                          components are
                   installed, connected and adjusted in accordance with the applicable manufacturer’s
                   instructions and specifications.
                   B. RADIATION PROTECTION
                           
                   Most states and local governments have regulations, which specify minimum safety
                   requirements for radiation protection. Consult your local regulations for specific
                   requirements.
                   Note:      Alternate line voltages, which wil l provide normal operation, are 208-277
                              VAC. 
                  where VN represents the line voltage under “no load” conditions, and V L represents line
                  voltage under “full load” conditions.
                  Minimum Over Current Protection Rating – 50% of maximum line current rating or
                  greater. Assuming maximum line current to be 140 amps, this value would be 70
                  amperes or greater.
                  Wire Size from Power Transformer to Disconnect Switch – for 50 feet use #2AWG; for
                  100 feet use #00AWG; for 200 feet use 250mcm
                                            (mcm – thousand circular mils.)
                  shall be operable
                  Laboratories also from a location
                                    requires         readily
                                              that this      accessible
                                                        disconnecting    from the
                                                                       means    be x-ray control.
                                                                                   mounted        Underwriters
                                                                                            on either the wall
                  behind the x-ray control or the wall directly adjacent to it.
                      2. Verify that the oil level in the H.V. generator is approximately ½” to ¾” below the
                         underside of the cover. Refill with “Shell Diala oil AX” or equivalent as
                         necessary.
                      3. Once the H.V. generator is in its final location, loosen the screw on the vent plug
                         to allow for oil expansion. Cover the vent plug and flange with one of the loose
                         fitting plastic covers provided with the H.V. generator to prevent contaminants
                         from entering the oil.
                       5. From the inside of the control, remove the two thumbscrews and washers from
                          front panel being careful not to drop them into the cabinet.
http://slidepdf.com/reader/full/summit-325e-service-manual                                              15/40
                                                                                     325E X-Ray Generator
7/18/2019          Summi t Industri es, Inc.                                   Model
                                                                             Summit   Numbers
                                                                                    325E           F280 / A700
                                                                                         - Service Manual
                       6. Grasp the “lift tab” at the center of the front panel (near the top) and lift upwards
                          approximately 1” to unhook the bottom of the front panel from the base of the
                          cabinet. While holding the panel up, push out the bottom and lower it to the floor.
                          Then tilt the lid back to its normal position so that the front panel can be
                          removed.
E. PRELIMINARY ASSEMBLY
http://slidepdf.com/reader/full/summit-325e-service-manual                                                        16/40
                                                                                     325E X-Ray Generator
7/18/2019          Summi t Industri es, Inc.                                   Model  Numbers
                                                                             Summit 325E           F280 / A700
                                                                                         - Service Manual
                            a) High voltage cables (HV cables) – The high voltage generator requires the
                               use of H.V. cables constructed with federal terminals. Note: Use extreme
                               care when handling the H.V. cables to avoid damaging the plug or
                               terminal pins.
                                Be sure that “Anode” of H.V. generator connects to “Anode” of x-ray tube, and
                                “Cathode” to “”Cathode”. H.V. cable terminals must be thoroughly cleaned
                                and then coated with a thick coating of Vapor Proofing Compount (normally
                                provided with the x-ray tube) prior to insertion into the x-ray tube or H.V.
                                generator receptacles.
                                Insert the H.V. cable terminals into the appropriate receptacle and screw the
                                cable nut as tightly as possible by hand – do not use tools for t ightening.
b) Stator cable – This x-ray control will accept either a 3 or 5 wire tube stator
                                cable.
                                09, T5, The terminals of the x-ray control are on TB3 and are marked 07, 08,
                                        and T6.
                                    *If tube has thermal cutout switch, remove jumper across T5 and T6 and
                                    connect leads as indicated. If tube does not have thermal cutout switch,
                                    terminals T5 and T6 must be jumpered together. (Control is shipped with
                                    this jumper installed.)
http://slidepdf.com/reader/full/summit-325e-service-manual                                                       17/40
                                                                                       325E X-Ray Generator
7/18/2019
                   Summi t Industri es, Inc.                                     Model  Numbers F280 / A700
                                                                               Summit 325E - Service Manual
                            a) The 25’ primary cable for the H.V. generator is connected to the control at the
                               factory. The loose end, marked P1, P2, and GND, should be connected to
                               the corresponding terminals of the H.V. generator.
                            b) The 25’ filament supply cable for the H.V. generator is connected to the
                               control at the factory. The loose ends, marked XC, XS, XL, and MI are to be
                               connected to the corresponding terminals of the H.V. generator.
                            This x-ray control is provided with fused power supplies rated 24 VACm 10Amp
                            max. (terminals “C1 and “C2” of TB2) and 24 VDC 8Amp max. (terminal “DC+”
                            and “DC-“ of TB2).
                            This x-ray
                            auxiliary    control circuitry
                                      interlock   provideswhich
                                                           pairsinhibit
                                                                  of terminals which can These
                                                                        x-ray exposures. be used  to connect
                                                                                               terminal pairs
                            are marked T6/T7, and T7/F2 of TB3. Removable jumper clips are connected
                            across these terminals at the factory. If it is necessary to connect an interlock
                            switch into the exposure circuit, remove one of these jumpers and connect the
                            interlock leads across the opened terminals. Note: Any interlock switch to be
                            used must be rated 125 VAC, 2 Amps mi nimum . 
                       6. Bucky connections –
                            This control provides terminals which can be used to operate one or two buckys.
                            The terminals are designated B1W, B2W, B3W, and B4W for “wall” buckys; and
                            B1T, B2T, B3T, and B4T for “table” buckys. Since several types of buckys may
                            be used, be sure to connect bucky to appropriate terminals as indicated below.
                            All bucky’s must be configured for 115 VAC operation.
                            Terminals have been provided in the control for connecting an auxilary exposure
                            switch. They are on TB2 and are marked S1, s2, and S3. Refer to control circuit
                            schematic for connection details.
8. Power supply –
                            A 13 ft. line cable marked L1, L2, and GND is provided. One end is connected to
                            the control at the factory. The other end is to be connected to the disconnect
                            switch to the appropriate terminals. The disconnect switch should be “off” before
                            making connections. Be sure to connect the green “GND” lead to a “GOOD”
                            ground.
                            The 02249-000 timer board has a DIP switch (S1) to select the tube protect limits
                            for the x-tube type used at this installation. The new U1 chip will be labeled with
                            generator type (325 or 360), line frequency (50 or 60 Hz), and tube manufacturer
                            (Summit/Toshiba or Varian/Eureka). Verify the chip type installed is appropriate
                            for the generator, line frequency and tube type at this site. DIP SW1 positions
                            will select the specific type of x-ray tube insert at this site (see table below). For
                            the DIP switch “OFF” is to the left, “ON” is to the right. Any change to the DIP S1
                            settings must be done while the generator is turned OFF.
http://slidepdf.com/reader/full/summit-325e-service-manual                                                           19/40
                                                                                        325E X Ray Generator
                   Summi t Industri es, Inc.                                       Model Numbers F280 / A700 
7/18/2019                                                                        Summit 325E - Service Manual
                       3. Disconnect leads marked P1 and P2 from H.V. generator and insulate with
                          electrical tape. This will prevent accidental production of x-rays during initial
                          check-out. Switch the disconnect switch “on” and then switch the control’s “on-
                          off” switch “on”.
                   Note that the 02249-000 timer board will display error codes per the table below. If
                   more than one selection is invalid the highest number will take precedence. For
                   example, an “invalid mA” and “kVp too low” would display as E04. Once the mA error is
                   cleared the kVp low error E03 would then be displayed.
                         4. Depress and hold “line check rocker switch” and observe kVp meter. Rotate “line
                            voltage compensator” until kVp meter needle is in the black area of the scale
                            marked “V”. Release line check rocker switch.
                         5. While observing the tube filaments through the port of the x-ray tube, verify that
                            the correct filament is lit for each mA station.
                         6. Depress the exposure switch to the 1st  stage (prep.) and verify anode rotation
                            and filament boost.
                         7. The kVp meter circuit is factory adjusted, but should be checked to verify
                            approximate calibration.
                            a) With power “off” connect an “AC” voltmeter capable of reading 0-250 VAC
                               across the wipers (center terminals of tap switches) of the KV selector
                               switches (major and minor). This will allow measurement of the “no-load”
                               primary voltage.
                            b) Turn “on” power and adjust the kVp selector switches to result in 220 VAC
                               indicated on the voltmeter.
                            c) Compare the kVp meter indications for all mA stations with the values given
                               below:
http://slidepdf.com/reader/full/summit-325e-service-manual                                                       21/40
                                                                                      325E X-Ray Generator
                   Summi t Industri es, Inc.                                    Model Numbers F280 / A700
7/18/2019                                                                     Summit 325E - Service Manual
                                                     50                                  81
                                                    100                                  80
                                                    150                                  76
                                                    200                                  69
                                                    300                                  53
                                    If the kVp meter indication varies by more than 5 kVp from the values
                                    given above, proceed to the kVp calibration procedure (page 23).
                       8. Check proper operation of all auxiliary devices and interlocks (electric locks,
                          collimator, auxiliary exposure switch, bucky, etc.)
                       9. Switch the disconnect switch “off” and reconnect the leads P1 and P2 to the H.V.
                          generator.
                   G. CALIBRATION
                                  
                   This x-ray generator has been fully tested and calibrated using a “production test” x-ray
                   tube.    Because the tube characteristics of this tube may vary from the tube
                   characteristics of the specific tube being installed, the assembler must check the
                   accuracy of the tube current, and make the appropriate adjustments.
                   This
                   powerequipment
                         supply as has been below:
                                   described designed and calibrated based on proper connection to a
                   Calibration will normally involve only slight adjustments of the tube current and
                   verification of KV accuracy.
                   It is recommended that the assembler read and understand the information provided
                   with the x-ray tube prior to making any x-ray exposures. Particular attention should be
                   given to:
                            The leads connected to the resistor bands of RXS and RXL are marked with
                            numbers, which correspond to the mA stations. Moving one of these bands to
                            the right increases the tube current for the corresponding mA station.
                            The leads connected to the bands of RSCC are also marked with numbers,
                            which correspond the mA stations. Moving one of these bands to the right
                            causes a reduction of mA at high kVp levels as compared to low kVp levels for
                            the corresponding mA station. This is commonly referred to as the balance
                            adjustment.
                            Note:    If a band
                            corresponding   tubeofcurrent
                                                    RXS oris RXL is all
                                                             not up     the proper
                                                                    to the  way to level,
                                                                                     the right & th
                                                                                           it may   e
                                                                                                   be
                            necessary to reposition the right-most band of Boost Resistor, “ RB”.
                            Moving this band to the right causes tube current to increase for all mA
                            stations.
                            b) Adjust
                               (movingthe
                                       the band  of the
                                           band to   RXL   marked
                                                        right        150mA).
                                                              increases  to produce approximately 156 mA
                            e) If the mA at 120 kVp is higher than the mA at 50kVp, move the band of RSCC
                                marked 150 to the right. If lower, move the band to the left.
                            f) Repeat steps (a) through (f) until no further adjustments are necessary.
                            g) Check mA tracking from 50 to 120 kVp and fine tune as required. The unit is
                               capable by fine tuning to deviate no more than approximately 5% of selected
http://slidepdf.com/reader/full/summit-325e-service-manual                                                     23/40
                                                                                        325E X-Ray Generator
                   Summi t Industri es, Inc.                                   Model Numbers F280 / A700
7/18/2019                                                                     Summit 325E - Service Manual
mA
                                                                                  + 5%      SELECTED mA
                                                                                  - 5%
0 50 80 125 140 kV p
Note: In step (b), use the follow ing values in place of 156 mA.
                                                             MA STATION         MA
                                                                 50             52
                                                                100            105
                                                                200            210
                                                                300            315
                            b) Verify kVp METER COMP. CIRCUIT input voltage and write down factory kVp
                               METER offset calibration data.
                                1) Connect an AC voltmeter between TP-1 and Tp-2 of the kVp COMP. PCB.
                                   Verify that voltage is 100 VAC.
                                3) Repeat step 2 for all remaining mA stations using test points indicated in
                                   table below. Use results for future reference.
http://slidepdf.com/reader/full/summit-325e-service-manual                                                       25/40
                                                                                        325E X-Ray Generator
                   Summit Indus tries, Inc.                                   Model Numbers F280 / A700
7/18/2019                                                                    Summit 325E - Service Manual
                            c) Verify and adjust kVp meter accuracy
                                2) Adjust the 150 mA offset pot until the kVp meter’s preread indication
                                   matches the measured kVp output.
                                3) Select 120 kVp. Make an x-ray exposure and adjust the common slope
                                   pot until the kVp meter’s preread indication matches the measured kVp
                                   output. Repeat steps 1, 2 and 3 above until no further adjustments are
                                   necessary.
                                4) Select 50 mA and 90 kVp; Adjust the 50 mA offset pot until the kVp
                                   meter’s preread indication matches the measured kVp output.
6) Check final kVp meter tracking over the entire kVp range. Under normal
                                    conditions
                                    kVp output the kVp meter’s
                                                by more         indication will not deviate from the measured
                                                        than 5 kVp.
http://slidepdf.com/reader/full/summit-325e-service-manual                                                                             26/40
                                                                                                                325E X-Ray Generator
              Summit Indus tries, Inc.                                                       Model Num bers F280 / A700 
7/18/2019                                                    Summit 325E - Service Manual
25
http://slidepdf.com/reader/full/summit-325e-service-manual                                                                 27/40
                                                                      325E X-Ray Generator
                   Summi t Industri es, Inc.                                   Model Numbers F280 / A700
7/18/2019                                                                     Summit 325E - Service Manual
                   Perform these tests in accordance with NEMA publication #XR8-1979, Tests Methods
                   for Diagnostic X-Ray Machines for Use During Initial Installation. It is necessary that the
                   assembler document these tests to demonstrate at a later date that all tests were
                   performed and that the x-ray system was left in full compliance with the standard. The
                   test results obtained must verify that the following system specifications are met or the
                   system must be removed from service until corrective action is completed.
SYSTEM SPECIFICATIONS
                   BEAM QUALITY
                                 
                   The half-value layer of the useful beam shall not be less than 3.5mm of aluminum.
                   When the x-ray unit is operated on an adequate power supply (see section IIA): (1) the
                   estimated coefficient of variation of radiation exposure shall not be greater than .048 for
                   any specific combination of technique factors, and (2) the average ratios of exposure to
                   the indicated tube current exposure time product (mAs) obtained at any two consecutive
                   tube current settings shall not differ by more than .095 timed their sum, or
                                                                    X1 – X2
                                                                                    < 0.095
                                                                    X1 + X2
                   where X1 and X2 are the average mR/mAs values obtained at each of two consecutive
                   tube current settings.
http://slidepdf.com/reader/full/summit-325e-service-manual
                                                                                       325E X-Ray Generator      28/40
                   Summi t Industri es, Inc.                                 Model Numbers F280 / A700
7/18/2019                                                                  Summit 325E - Service Manual
                   TUBE CURRENT
                    
                   The maximum deviation of the tube current from its preindicated value, during an
                   exposure, when the equipment is connected to an adequate power supply (see section
                   IIA) shall not exceed +9% of selected tube current within the range of 50-125 kVp
                   (inclusive), and +19% of selected tube current within the range of 40-50 kVp (exclusive).
                   EXPOSURE TIME
                    
                   The maximum deviation of the exposure time from its preindicated value, during an
                   exposure, when the equipment is connected to an adequate power supply (see section
                   IIA) shall not exceed +0 pulses within the range of 1/120-1/10 (inclusive) +14% of
                   selected time within the range of 3/20-6 seconds (inclusive).
http://slidepdf.com/reader/full/summit-325e-service-manual
                                                                                  325E X-Ray Generator         29/40
                   Summi t Industri es, Inc.                                 Model Numbers F280 / A700
7/18/2019                                                                                 Summit 325E - Service Manual
ANCHORING SCREW
PLYWOOD PLATE
http://slidepdf.com/reader/full/summit-325e-service-manual                                                               30/40
                                              PART I.D. for 325                       360 GENERATOR LID ASSEMBLY
7/18/2019                                                                             Summit 325E - Service Manual
            01376-000:  Kit,
            Kilovolt Meter
            01381-000:
            Replacement                                                                                                                         HG11: GREEN,
            Instructions                                                                                                                            Ready Light
                                                                                                                                                A310: SWITCH,
             A365:  ON/OFF                                                                                                                      Time Select 325 
            Circuit Breaker
                                                                                                                                               C083:  SWITCH,
                                                                                                                                                Time Select 360 
            HG11: GREEN
            Power Indicator                                                                                                                        A311: KNOB,
            (same as Ready)                                                                                                                         With Pointer
1:02 PM 12/26/2006
http://slidepdf.com/reader/full/summit-325e-service-manual                                                                                                          31/40
7/18/2019                                                               Summit 325E - Service Manual
01204
G
http://slidepdf.com/reader/full/summit-325e-service-manual                                  33/40
7/18/2019                                                    Summit 325E - Service Manual
http://slidepdf.com/reader/full/summit-325e-service-manual                                  34/40
7/18/2019                                                                                Summit 325E - Service Manual
     1/4 WT
    RESISTOR
       SIZE
     JUMPER
                                                                                                                                        mA Station
                                                                                                                                  J4
                                                                                                                                  Pin   For    For
                                                                                                                                        325    360
                                                                                                                                        Gens   Gens
                                                                                                                                  1     200S   24S
                                                                                                                                  2       X    X
                                                                                                                                  3     5VDC 5VDC
                                                                                                                                  4     300L   300L
                                                                                                                                  5     200L   240L
                                                                                                                                  6     150L   120S
                                                                                                                                  7     100S   360L
                                                                                                                                  88    50S    48S
http://slidepdf.com/reader/full/summit-325e-service-manual                                                                                            35/40
7/18/2019                                                                Summit 325E - Service Manual
Component Layout for Timer/Driver PCBA (Summit Part No. 02249-002 Rev. C)
http://slidepdf.com/reader/full/summit-325e-service-manual                                                                   36/40
7/18/2019                                                    Summit 325E - Service Manual
http://slidepdf.com/reader/full/summit-325e-service-manual                                  37/40
7/18/2019                                                    Summit 325E - Service Manual
http://slidepdf.com/reader/full/summit-325e-service-manual                                  38/40
7/18/2019                                                    Summit 325E - Service Manual
http://slidepdf.com/reader/full/summit-325e-service-manual                                  39/40
7/18/2019                                                    Summit 325E - Service Manual
http://slidepdf.com/reader/full/summit-325e-service-manual 40/40