SP

Configuration
System

Configuration Guide Version VC10A

Valid for:
ARCADIS Varic
ARCADIS Orbic
ARCADIS Avantic
© Siemens 2008
The reproduction, transmission or use
of this document or its contents is not
permitted without express written
authority. Offenders will be liable for
damages. All rights, including rights
created by patent grant or registration
of a utility model or design, are
reserved.

Print No.: SP00-000.843.30.02.02 English


Replaces: SP00-000.843.30.01.02 Doc. Gen. Date: 05.08
2 Revision / Disclaimer
1Revision / Disclaimer

Document revision level


The document corresponds to the version/revision level effective at the time of system
delivery. Revisions to hardcopy documentation are not automatically distributed.
Please contact your local Siemens office to order current revision levels.

Disclaimer
The installation and service of equipment described herein is to be performed by qualified
personnel who are employed by Siemens or one of its affiliates or who are otherwise
authorized by Siemens or one of its affiliates to provide such services.
Assemblers and other persons who are not employed by or otherwise directly affiliated
with or authorized by Siemens or one of its affiliates are directed to contact one of the
local offices of Siemens or one of its affiliates before attempting installation or service pro-
cedures.

ARCADIS SP00-000.843.30.02.02 Page 2 of 112 Siemens


05.08 CS SD SP/CO Medical Solutions
Table of Contents 3
0Table of Contents

1 _______ Definitions and abbreviations _____________________________________ 6

Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6

2 _______ General ________________________________________________________ 8

General. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Reboot, restart and shutdown . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9

3 _______ Country-specific settings / Image Text Editor________________________ 10

Country-specific keyboard layout and user interface . . . . . . . . . . . . . . . . . . . . . . . . . . . 10


Image Text Editor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12

4 _______ Syngo ________________________________________________________ 13

Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Service software . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Open the service software. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Configuration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
System Options. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Local Host. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Site Info . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
TCP/IP LAN . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
Routing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
Users. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
Monitor Type . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
Security . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
Service . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Mail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
FTP . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
Auto Transfers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
Event log . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
Backup/Restore . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
Licensing. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
DICOM . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
Character Set . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
Structured Reports . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
Offline Devices . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
Network Nodes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
Print Devices. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
HC Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
LUT Files. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
HIS/RIS Nodes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
Import/Export . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
Directories. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
EPRI . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60

Siemens SP00-000.843.30.02.02 Page 3 of 112 ARCADIS


Medical Solutions 05.08 CS SD SP/CO
4 Table of Contents

Server . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60
External Devices . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61
Paper Printer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61
Printer Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62
PS LUT Files . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63
SystemMgmt . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 64
Master . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 64
Agent Controls. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 64
Applications. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 66
Pat. Registration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 66
Viewer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 67
CorRea . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 67
MPPS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68
Worklist Results . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 69

5 _______ System Devices ________________________________________________ 71

USB Settings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 71
CAN Bus . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 72
Footswitch . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 73
Trigger Injector . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 74

6 _______ Imaging System ________________________________________________ 75

Valid Mainsystem . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 76
Application Fields . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 77
Examination Set Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 78
Edge Enhancement Filter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 79
DICOM LUT Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 80
Windowing LUT. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 81
Logarithm LUT . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 82
Metering Sections File. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 83
Display Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 84
Subtraction Parameters. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 85
Motion Detection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 86
Power Off Parameters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 88
Brightness Control. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 89
Auto Dose Adjustment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 91
Auto Lut. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 95
Camera . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 96
Circle Mask . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 97
Hardcopy Configuration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 98
Homogenisation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 99

7 _______ Main System Varic / Orbic _______________________________________ 100

ARCADIS SP00-000.843.30.02.02 Page 4 of 112 Siemens


05.08 CS SD SP/CO Medical Solutions
Table of Contents 5

Configuration Varic / Orbic . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 101


Main System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 101
Load Counter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 101
Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 101
Current Limitation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 101
Buzzer. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 102
Image Intensifier . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 102

8 _______ Main System Avantic ___________________________________________ 103

Configuration Avantic . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 104


Main System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 104
Image Intensifier . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 104
Load Counter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 104
Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 104
Current Limitation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 105
Buzzer. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 105

9 _______ Backup of parameter changes ___________________________________ 106

10 ______ Network Checklist _____________________________________________ 107

Checklist . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 107
Person responsible for the network configuration at the customer site. . . . . . . . . . 107
System network data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 107
Network data for HIS/RIS network partner (only with HIS/RIS option) . . . . . . . . . . 108
Network data for reporting console/archiving system . . . . . . . . . . . . . . . . . . . . . . . 108
Network data for reporting console/archiving system . . . . . . . . . . . . . . . . . . . . . . . 109

11 ______ Changes to previous version ____________________________________ 111

Siemens SP00-000.843.30.02.02 Page 5 of 112 ARCADIS


Medical Solutions 05.08 CS SD SP/CO
6 Definitions and abbreviations
1-
Overview
1Definitions and abbreviations

Abbreviation Definition
BIOS Basic Input Output System, lowest level of PC configura-
tion
AET Application Entity Title - name of the DICOM service that
communicates either as the user (SCU) or the provider
(SCP).
CD Compact disc
CDR Compact Disc Recordable
DVD Digital Video Disc
DICOM Digital Imaging and Communications in Medicine
(ACR-NEMA standard)
DHCP Dynamic Host Configuration Protocol - procedure for
assigning numeric IP addresses to nodes, in which
blocks of addresses are leased to clients on a first-come,
first-served basis. Typically, DHCP is combined with
WINS so an IP address is both dynamically assigned
and automatically associated with a domain name. This
can be combined with DNS, resulting in a dynamically
assigned numeric IP address that is automatically asso-
ciated with both NetBIOS and Internet names.
DNS Domain Name Services - used for address parsing with
numeric IP addresses being linked to names in the Inter-
net format.
FTP File Transfer Protocol - a client-server application that
transfers files from one computer to another.
Gateway Controls data traffic between two networks above the
network layer.
HIS Hospital Information System
LAN Local Area Network (in this case: always Ethernet)
LMHOSTS Address parsing method in which numeric IP addresses
are statically linked to NetBIOS names. The LMHOSTS
file is checked to determine whether the name is
addressed.
NIC Network Interface Card - an expansion card that provides
a physical connection to the network.
PPP Point-to-Point Protocol - data connection and network
protocol for serial connections (such as dial-up
modems).

ARCADIS SP00-000.843.30.02.02 Page 6 of 112 Siemens


05.08 CS SD SP/CO Medical Solutions
Definitions and abbreviations 7

RAS Remote Access Services - a Windows NT service that


supports dial-in and dial-out connections using modems
and ISDN.
RIS Radiology Information System
Router A multi-port computer or a dedicated hardware compo-
nent that examines network layer addresses for data
packets and sends them to a different network as appro-
priate.
SCU Service Class User
SCP Service Class Provider
TCP/IP Transport Control Protocol/Internet Protocol - the stan-
dard protocol for networks.
UI User Interface
URL Uniform Resource Locator
VPN Virtual Private Network
WINS Windows Internet Name Service - used for address pars-
ing with IP addresses being linked dynamically to Net-
BIOS names. DNS can be combined with WINS.
WIP Work In Progress

Siemens SP00-000.843.30.02.02 Page 7 of 112 ARCADIS


Medical Solutions 05.08 CS SD SP/CO
8 General
2-
General
2General

NOTE Depending on the progress of the configuration, the "Next”


button can also be displayed as the "Finish” button.

NOTE Once the <Save> button has been selected in the individual
screens, the message "..... successfully saved” appears fol-
lowing a successful save.

• Select "OK".

NOTE If any problems arise when making changes in the Service


UI, attempt to resolve them first by powering the system
completely off and then powering it on again.

NOTE A shutdown is required after servicing and prior to handing


over to the customer. Hibernation alone is not sufficient.

ARCADIS SP00-000.843.30.02.02 Page 8 of 112 Siemens


05.08 CS SD SP/CO Medical Solutions
General 9

Reboot, restart and shutdown 2.1

NOTE To activate changes in the system, the system must always


be shut down; i.e., the complete system must be switched
off and switched back on again.
A shutdown must always be performed at the end of the con-
figuration process. Hibernation alone is not sufficient.

Siemens SP00-000.843.30.02.02 Page 9 of 112 ARCADIS


Medical Solutions 05.08 CS SD SP/CO
10 Country-specific settings / Image Text Editor
3-
Country-specific keyboard layout and user interface
3Country-specific settings / Image Text Editor

1. In the "Configuration" panel double-click “Regional and Language...”.


2. Select “Regional Options”.
3. Under “Standards and formats”, select the appropriate language from the drop-down
menu:
- English (United States)
- German (Germany)
- French (France)
- Spanish (Spain)
4. Under “Location”, select your corresponding location.
5. Click <Apply>

NOTE “English (United States)” is the default setting. As a result


“Apply” is not available.

6. Select the “Languages” tab card and click <Details...>.


7. Under “Installed services” click “Add...”
¹ The window “Add input language” appears.
8. Under “Installed services”, the following “input language” can be selected:
- English (United States)
- German (Germany)
9. Under “Keyboard layout / IME”, select the appropriate keyboard language from the
drop-down menu:
- US
- United Kingdom
- German
- Swedish
- Portuguese
- Italian
- French
- Spanish

NOTE If a keyboard language (e. g. “US”) is displayed under


“Installed services”, it disappears from the “Keyboard lay-
out/IME” drop-down menu.

NOTE The correct keyboard language can also be found in the


delivery note from the factory.

10. Click <OK> to close the “Add input language” window.


11. Click <Apply> and <OK> to close the window “Text Services and Input Languages”.

ARCADIS SP00-000.843.30.02.02 Page 10 of 112 Siemens


05.08 CS SD SP/CO Medical Solutions
Country-specific settings / Image Text Editor 11

12. Click <Apply> and <OK> to close the window “Regional and Language Options”.
¹ If the language has been changed the following message appears:
The following is displayed in the case that values have been changed:
“The regional settings have been changed.
Therefore a restart of the application is necessary.
Do you want to restart the application now?”
¹ In case of language change click <OK>. This will apply the settings and the sys-
tem restarts.
¹ Otherwise shut down the system via <Options>-<End Software Session>-<Shut-
down>.

Siemens SP00-000.843.30.02.02 Page 11 of 112 ARCADIS


Medical Solutions 05.08 CS SD SP/CO
12 Country-specific settings / Image Text Editor

Image Text Editor 3.1

1. In the syngo menu bar, select <Options>-<Configuration>.


2. Double-click “Image text editor”.
3. Under “View Name” (Basisformat), select “XA Image”.
- Deselect software version
¹ Software version = Q
4. Click <Apply>.
5. Under “View name”, select “ARCADISAcqView”.
6. Click “Customized text” (“Ausgewählte Texte”).
¹ “Image Date” = Q
¹ “Image Time” = Q
¹ “Image Number” = Q
7. Click <Apply>.
8. Click <OK> and close the Configuration panel.

ARCADIS SP00-000.843.30.02.02 Page 12 of 112 Siemens


05.08 CS SD SP/CO Medical Solutions
Syngo 13
4-
Configuration
4Syngo

System Type Depending on the system, select ARCADIS Varic, ARCA-


DIS Orbic, ARCADIS Orbic 3D or ARCADIS Avantic.

The following system options can be selected.

Attached to network TCP/IP LAN (local network) for local system


DICOM Print Devices DICOM Basic Print Service for laser cameras
DICOM Offline Devices DICOM Media Storage on offline media such as MOD and
CD-R
DICOM Networking DICOM Storage and Query/Retrieve via remote network
nodes
DICOM HIS/RIS DICOM Basic Worklist for radiology (hospital) info. sys-
tems
Image import/export Import/export of DICOM images from/to the file system
Paper Printer Format for local printers (LPT1 or USB)
EPRI Electronic Patient Record Interface
System Management System Management Agent
Litho report Functionality activated by Litho report
Enable help pages Help pages are displayed between the configuration
screens.

Siemens SP00-000.843.30.02.02 Page 13 of 112 ARCADIS


Medical Solutions 05.08 CS SD SP/CO
14 Syngo

Service software 4.1

Open the service software. 0

• Under Options, select "Service".


• Select "Local Service".
• Enter the service key under "Please enter password".
• Select "Set as Default" and confirm with "OK".
¹ The Service Home menu is displayed.

Configuration 0

• On the "Home Page", select the "Configuration" menu button.


¹ If a "Restore" or a backup is not available, an "Initial Setup" has to be performed.

NOTE In the case of an initial setup, import the correct license file
for available options first.

• Select the corresponding system options.


¹ “Attached to Network” and “DICOM Offline Devices” are permanently selected.

ARCADIS SP00-000.843.30.02.02 Page 14 of 112 Siemens


05.08 CS SD SP/CO Medical Solutions
Syngo 15

System Options 0

Fig. 1: System Options

System Type Depending on the system, select ARCADIS Varic, ARCA-


DIS Orbic, ARCADIS Orbic 3D or ARCADIS Avantic.

Attached to network TCP/IP LAN (local network) for local system


DICOM Print Devices DICOM Basic Print Service for laser cameras
DICOM Offline Devices DICOM Media Storage on offline media such as MOD and
CD-R
DICOM Networking DICOM Storage and Query/Retrieve via remote network
nodes
DICOM HIS/RIS DICOM Basic Worklist for radiology (hospital) info. sys-
tems
Image import/export Import/export of DICOM images from/to the file system
Paper Printer Format for local printers (LPT1 or USB)
EPRI Electronic Patient Record Interface
System Management System Management Agent

Siemens SP00-000.843.30.02.02 Page 15 of 112 ARCADIS


Medical Solutions 05.08 CS SD SP/CO
16 Syngo

Litho report Functionality activated by Litho report


Enable help pages Help pages are displayed between the configuration
screens.

ARCADIS SP00-000.843.30.02.02 Page 16 of 112 Siemens


05.08 CS SD SP/CO Medical Solutions
Syngo 17

Local Host 4.2

Site Info 0

Fig. 2: Site Info

Software installation
This date is set automatically during software installation and cannot be changed.

handed over
• Enter the date and time when the system was handed over to the customer.

System
Material No
• This material no. is entered automatically once the "System typel" has been selected
under “System Options”
- For ARCADIS Varic, “101 43 406” is displayed under “Material No”
- For ARCADIS Orbic, “101 43 407” is displayed under “Material No”
- For ARCADIS Avantic, “101 43 408” is displayed under “Material No”

Siemens SP00-000.843.30.02.02 Page 17 of 112 ARCADIS


Medical Solutions 05.08 CS SD SP/CO
18 Syngo

Rdiag prefix
• This name is entered automatically once the "system model" was entered in the previ-
ous screen. It is needed to generate a unique name for the remote transfer.
- For ARCADIS Varic, "cv" is displayed under "Rdiag Prefix".
- For ARCADIS Orbic, "co" is displayed under "Rdiag Prefix".
- For ARCADIS Avantic, "av" is displayed under "Rdiag Prefix".

Serial No
• Enter the system serial number.
- ARCADIS Varic 12000 to 19999
- ARCADIS Orbic 22000 to 29999
- ARCADIS Avantic 32000 to 39999

NOTE The serial number is mandatory. It may not include any char-
acters other than numbers from 0 to 9.

Department
• Enter the department where the system is located.

Station Name
• Enter the “Station Name” (xxx = last 3 digits of the system serial number).
- For ARCADIS Varic: ARCADISVA12xxx
- For ARCADIS Orbic: ARCADISOR22xxx
- For ARCADIS Avantic: ARCADISAV32xxx

NOTE The“Station Name” should be set similar to the “Computer


Name” under “TCP/IP LAN”. Possibly a system reboot is nec-
essary in between the “Station Name” and “Computer Name”
adaptation.

NOTE Changing “Site Info - Station Name” or “TCP/IP LAN - Com-


puter Name” with MNP (Managed Note Package) activated:
¹ Uninstall MNP previously to the rearrangement.
¹ Install MNP newly after the rearrangement.

Service
Remote Service Center
• The correct remote service center, Europe, USA or Asia, must be set.

ARCADIS SP00-000.843.30.02.02 Page 18 of 112 Siemens


05.08 CS SD SP/CO Medical Solutions
Syngo 19

Service Center
• In this field, enter the location of the Service Center responsible for this system. This
information is required for the generation of automatic reports.

Site identification number


• The syntax of the ID is different in each country, and can be found in your “Service
Management System“. The ID is used to define the local system in the Service Center.

Service center phone


• Enter the phone number of the appropriate Service Center.

Customer and Address


Name
• Enter the name of the customer. In certain cases, it may be the same as the hospital
name (see the adjacent field).
Id
• Enter the Identification number of the customer. The syntax of the ID is different in each
country and can be found in your “Service Management System“
Hospital
• Hospital, Street, Street No, ZIP Code, Phone No, City, District
Enter the appropriate data into these fields. This information may be helpful during a
remote service session and is shown in the report files (not included in this version).
Country
• Enter the country code (see document AXA4.010.814.02. .... country codes)
Administrator
• In this field, enter the name of the customer's administrator responsible for this system.
This information is helpful during remote and local service sessions.

Siemens SP00-000.843.30.02.02 Page 19 of 112 ARCADIS


Medical Solutions 05.08 CS SD SP/CO
20 Syngo

Site Info (2)

Fig. 3: Site Info2

Time zone
• Depending on the location of the system
NTP Time Server
• NTP Time Server: Q
¹ If the customer has requested that "NTP Time Server” be selected, you must also
enter the IP address of the NTP server.
DaylightTime
• DaylightTime automatic: H
¹ If the NTP time server was activated, these values are automatically accepted
and no changes can be made.
• Select the values for "Date and time".
• Click "Save".
¹ "You are going to change the time settings .......” appears.
• Click <OK>.

ARCADIS SP00-000.843.30.02.02 Page 20 of 112 Siemens


05.08 CS SD SP/CO Medical Solutions
Syngo 21

TCP/IP LAN 0

Fig. 4: TCP_IP LAN

Identification / Computer name


• “Computer name” (xxx = the last 3 digits system serial number):
- For ARCADIS Varic: ARCADISVA12xxx
- For ARCADIS Orbic: ARCADISOR22xxx
- For ARCADIS Avantic: ARCADISAV32xxx

NOTE Changing “Site Info - Station Name” or “TCP/IP LAN - Com-


puter Name” with MNP (Managed Note Package) activated:
¹ Uninstall MNP previously to the rearrangement.
¹ Install MNP newly after the rearrangement.

NOTE After changing entries under “Site Info” the “Computer


name” is may not changeable. Consequently perform a sys-
tem reboot (not Hibernation!).
¹ Now the field “Computer name” can be adapted
again.

Adapter / Select name


• Select “ (00000001) Broadcom NetXtreme Gigabit Ethernet".

Siemens SP00-000.843.30.02.02 Page 21 of 112 ARCADIS


Medical Solutions 05.08 CS SD SP/CO
22 Syngo

Obtain an IP address from the DHCP server


• "Obtain an IP address from DHCP server” is not selectable.

NOTE ARCADIS systems have not been released for use with a
DHCP server.
DHCP server is not selectable.

Specify an IP address
• After you select "Specify an IP address," the following input fields appear:
- IP address
- Subnet mask
- Gateways
IP address
• Under "IP address", enter a valid IP address to identify the system in the network. Con-
tact the network administrator to obtain a valid IP address.

NOTE Only one IP address may be entered.


If “Factory Default” IP address 192.xxx.x.x is entered, it has
to be deleted before the new IP address can be entered.

Subnet mask
• The subnet mask indicates the sections into which the address is divided. The data are
distributed via the router based on this division. Contact the network administrator to
obtain the subnet mask ID.
Gateway
• In order the leave the hospital network, it is necessary to inform the system (ARCADIS)
about this gateway (router). Contact the network administrator to obtain the gateway
address.
WINS / Primary WINS server
• Do not enter anything under "Primary WINS server".
Secondary WINS server
• Do not enter anything under "Secondary WINS server".
Enable DNS for Windows resolution
• Enable DNS for Windows resolution: Q
Enable LMHOSTS Lookup
• Enable LMHOSTS Lookup: H
DNS / Domain
• Do not enter anything under "Domain".
DNS service search order
• Do not select or enter anything under "DNS service search order".

ARCADIS SP00-000.843.30.02.02 Page 22 of 112 Siemens


05.08 CS SD SP/CO Medical Solutions
Syngo 23

Routing 0

Fig. 5: Routing
Hospital / Static Route to SRS
• Set the static route of the hospital.
Server Network Address Translation (NAT)
• The “NAT” functionality is may used in case of system management (see instruction
SRS / system management SPR2-310.843.30...)
Remote service center
• Depending of your settings under “Site Info” ( Europe, USA or Asia) the settings for the
corresponding service center will be automatically set.

Siemens SP00-000.843.30.02.02 Page 23 of 112 ARCADIS


Medical Solutions 05.08 CS SD SP/CO
24 Syngo

Users 0

Fig. 6: Users
Account for UI
• “meduser”
Default Password
• Default Password: H
Enable Autologin
• Enable Autologin: H
Account for PPP / Default Password
• Default Password Q
Account for Administrator
• Disable Account: Q (For the default password see the KB password list, a new list
can be issued at any time)

ARCADIS SP00-000.843.30.02.02 Page 24 of 112 Siemens


05.08 CS SD SP/CO Medical Solutions
Syngo 25

Monitor Type 0

Fig. 7: Monitor type

Monitor Type
• Select "color” and Siemens (Pseudo Color)"
¹ Regardless of whether the monitor is black-and-white or color.

Siemens SP00-000.843.30.02.02 Page 25 of 112 ARCADIS


Medical Solutions 05.08 CS SD SP/CO
26 Syngo

Security 4.3

Settings 0

NOTE If changes have been made and saved here, they may be
activ only by completely reinstalling the system software!

Fig. 8: Settings

Security properties
• User management: Q
• "Data access check: Not selectable
• Functional check: Q
• Auditing: Q

DICOM access
• Enable trusted host functionality: Q

Remote service access


• SSL encryption: Q

ARCADIS SP00-000.843.30.02.02 Page 26 of 112 Siemens


05.08 CS SD SP/CO Medical Solutions
Syngo 27

Service 4.4

NOTE If the system is connected for remote access, adhere to the


"Installation of Siemens Remote Service" document after
performing the service software configuration.

NOTE It is not possible to rename a DICOM node.

Work steps for creating a new DICOM node:


• Delete old DICOM node.
• Perform a system restart via <Options>-<End Session>.
• Enter the new DICOM node and restart the system again.

Mail 0

Fig. 9: Mail

Siemens SP00-000.843.30.02.02 Page 27 of 112 ARCADIS


Medical Solutions 05.08 CS SD SP/CO
28 Syngo

NOTE This menu item allows you to configure the mail target for
the Auto Report function and for the Service Center pro-
vided in the Customer User Interface at Option - Remote Ser-
vice.
For detail information how to configure this functionality
click on “Help” in the navigation bar.

• Click ">".

Mail (page 2)

Fig. 10: Mail 2

NOTE This menu item allows you to configure the mail target for
the Auto Report function and for the Service Center pro-
vided in the Customer User Interface at Option - Remote Ser-
vice.
For detail information how to configure this functionality
click on “Help” in the navigation bar.

ARCADIS SP00-000.843.30.02.02 Page 28 of 112 Siemens


05.08 CS SD SP/CO Medical Solutions
Syngo 29

FTP 0

Fig. 11: FTP

NOTE This menu item allows you to configure the file transfer tar-
get (FTP) for the Auto Report and for the File&Image Tool
function.
For detail information how to configure this functionality
click on “Help” in the navigation bar.

• Click ">".

Siemens SP00-000.843.30.02.02 Page 29 of 112 ARCADIS


Medical Solutions 05.08 CS SD SP/CO
30 Syngo

FTP (page 2)

Fig. 12: FTP2

NOTE This menu item allows you to configure the file transfer tar-
get (FTP) for the Auto Report and for the File&Image Tool
function.
For detail information how to configure this functionality
click on “Help” in the navigation bar.

ARCADIS SP00-000.843.30.02.02 Page 30 of 112 Siemens


05.08 CS SD SP/CO Medical Solutions
Syngo 31

Auto Transfers 0

Fig. 13: Auto Transfers

NOTE If the customer agrees, automatic transfer (i. e., without user
interaction) of Auto-reporting messages or System Manage-
ment Event messages may be enabled here.
As long “Mail” or “FTP” is not configured / enabled this set-
ting has no effect.

Siemens SP00-000.843.30.02.02 Page 31 of 112 ARCADIS


Medical Solutions 05.08 CS SD SP/CO
32 Syngo

Event log 0

Fig. 14: Eventlog


Log only service messages
• Log only Service messages: Q

ARCADIS SP00-000.843.30.02.02 Page 32 of 112 Siemens


05.08 CS SD SP/CO Medical Solutions
Syngo 33

Backup/Restore 0

Fig. 15: Backup Restore


• Do not make any entries.

Siemens SP00-000.843.30.02.02 Page 33 of 112 ARCADIS


Medical Solutions 05.08 CS SD SP/CO
34 Syngo

Licensing 0

Licensing Details

Fig. 16: Licensing


Several applications are protected with one license key. This "Licensing" function allows
you to display and modify existing application licenses or to add new application licenses
to the system.

NOTE A license can be obtained from your UPTIME Service Center.

• License details are shown on this page

ARCADIS SP00-000.843.30.02.02 Page 34 of 112 Siemens


05.08 CS SD SP/CO Medical Solutions
Syngo 35

Select New License File

Fig. 17: Select New License File

• Select new license file


- Browse to select the license file: H
- Place the license CD into the DVD drive.
- Select license via "Browse".
- Click <Save>.

Siemens SP00-000.843.30.02.02 Page 35 of 112 ARCADIS


Medical Solutions 05.08 CS SD SP/CO
36 Syngo

General Settings

Fig. 18: General Settings

• Use Local License File: H

ARCADIS SP00-000.843.30.02.02 Page 36 of 112 Siemens


05.08 CS SD SP/CO Medical Solutions
Syngo 37

DICOM 4.5

DICOM allows you to specify the parameters for the DICOM communication interfaces of
your system and the remote systems.
AET = Application Entity Title

General 0

Fig. 19: General

Local application entity titles


HIS/RIS
• Enter AET for HIS/RIS.
Study Transfer
• Enter AET for Study Transfer.
Print
• Enter AET for Print.
• Click ">".

Siemens SP00-000.843.30.02.02 Page 37 of 112 ARCADIS


Medical Solutions 05.08 CS SD SP/CO
38 Syngo

General DICOM settings

Fig. 20: General 2


• Do not make any changes on this page.

ARCADIS SP00-000.843.30.02.02 Page 38 of 112 Siemens


05.08 CS SD SP/CO Medical Solutions
Syngo 39

Character Set 0

Fig. 21: Character Set

DICOM default character set


• Select "Allow only one character set".
• Selcet the correct key board character set
The settings need to be changed only if the keyboard requires it.
- Latin Alphabet No. 1: European (ISO_IR100)
- Default Repertoire: USA (ISO_IR6)
- Japanese, Katakana: Japan (ISO_IR13)

Siemens SP00-000.843.30.02.02 Page 39 of 112 ARCADIS


Medical Solutions 05.08 CS SD SP/CO
40 Syngo

Structured Reports 0

Fig. 22: Structured Reports


• Structured Report configuration is not used for ARCADIS systems. A default viewer for
a viewing application is not defined.

Offline Devices 0

Offline devices allow you to make settings for DICOM offline devices (media storage) on
the local host.

ARCADIS SP00-000.843.30.02.02 Page 40 of 112 Siemens


05.08 CS SD SP/CO Medical Solutions
Syngo 41

Fig. 23: Offline Devices

WARNING Risk of data loss.


CD-R or DVD media are not permitted for long-term
archiving.
¹ "Archive Device" must not be activated.

• F: is set to “CD/DVD Combo” for the DVD drive.


• The available Media Storage devices of the system are automatically detected by
syngo and shown in the device table with the assigned drive letter.
- Drive letter DVD: “R”
- Logical Name: “DVD_RW”
- Used by syngo with logical name: H
- Export device (available in transfer menu): H
- Archive device (available in Transfer menu): Q
- Default device: H
- Recording Speed: “maximum”

Siemens SP00-000.843.30.02.02 Page 41 of 112 ARCADIS


Medical Solutions 05.08 CS SD SP/CO
42 Syngo

Network Nodes 0

Fig. 24: Network Nodes


The DICOM nodes for storage (send and receive) and query/retrieve are defined here.
Please note that DICOM printers and the Radiology Information System (RIS) are not
configured here.
The image transfer or query/retrieve targets entered can be selected in the viewer's or
patient browser's menu "Transfer - Send to...".
The entered query/retrieve targets can be selected in the viewer's or patient browser's
menu "Patient - Search".

NOTE It is not possible to rename a DICOM node.

Work steps for creating a new DICOM node:


• Delete old DICOM node.
• Perform a system restart via <Options>-<End Session>.
• Enter the new DICOM node and restart the system again.

NOTE If this system has 3D Navigation, a DICOM node must be


configured as described in the 3D installation instruction.

ARCADIS SP00-000.843.30.02.02 Page 42 of 112 Siemens


05.08 CS SD SP/CO Medical Solutions
Syngo 43

NOTE DICOM requires a license.

• Under "Select Host", select "define new".


Host name
• Enter Host name.
TCP/IP address
• Enter the IP address.
Connected by
• LAN: H
• Click <Save>.
• Click ">".

Network Nodes - Page 2

Fig. 25: Network Nodes 2

General Node Properties


Logical Name
• Select "define new".

Edit Name
• Enter the name under "edit name".

Siemens SP00-000.843.30.02.02 Page 43 of 112 ARCADIS


Medical Solutions 05.08 CS SD SP/CO
44 Syngo

Host
• Under "Host", select the name entered on the previous page.

Application Entity Properties

AE title
• Select "define new".
¹ The AETs that have already been configured are shown in the pull-down menu.

Edit AE Title
• Enter a name in the "edit AE Title" field.

Port number
• Enter the port number for the DICOM device.
• Click "Add".
• Do not select "Verification".

Supported DICOM services


Select the DICOM services purchased by the customer and to be run on the correspond-
ing system.

NOTE At a minimum, "Storage" must be selected.

Storage
• Storage H
• Retrieve: H (if the particular DICOM device is to support query)
• Do not make any changes under "Storage".
• Click <Save>.

Storage Commitment
• Select "Storage Commitment" if the particular DICOM device is to be configured as the
Storage Commitment provider.

Query
• Query: H (if the particular DICOM unit is to support query)
• Patient root: H (under "Provides DICOM query model")

Retrieve
• Retrieve: H (if the particular DICOM device is to support retrieve)
• Click <Save>.

ARCADIS SP00-000.843.30.02.02 Page 44 of 112 Siemens


05.08 CS SD SP/CO Medical Solutions
Syngo 45

Print Devices 0

Here you define the DICOM laser cameras that are connected to your system via DICOM
Basic Print. A local printer is NOT a DICOM printer!
This system does not support the SPCI/SPDI camera interface

NOTICE If a printer needs to be deleted, e.g., following an incorrect


entry, the correct sequence of print devices 3 to print
devices 1 must be observed.

¹ Otherwise, problems can occur, e.g., in the "Filming”


application card, in the user interface or when making
registration settings.

NOTE You will find the installation instructions for each released
camera on CB-DOC under Installation, in the "Camera” cata-
log.

NOTE Configuring "Print Devices" without the necessary DICOM


license will result in an error.

Siemens SP00-000.843.30.02.02 Page 45 of 112 ARCADIS


Medical Solutions 05.08 CS SD SP/CO
46 Syngo

Print Devices 1
• Verify that the required DICOM basic license is available.
- This can be checked as follows:
In the service software, go to --> Configuration --> Service --> Licensing.
ASPIA_DICOM_BASIC....”should be listed under Licenses.

Fig. 26: Print Devices

Host Properties
• Under "Select Host", select "define new".

Host name
• Enter host name of the DICOM camera.

TCP/IP address
• Enter the IP address.

Connected by
• LAN: H
• Under "Select Host", select the host name entered for the DICOM camera.
• Click <Save>.
• Click ">".

ARCADIS SP00-000.843.30.02.02 Page 46 of 112 Siemens


05.08 CS SD SP/CO Medical Solutions
Syngo 47

Print Devices 2

Fig. 27: Print Devices 2

General Node properties

Logical Name
• Select "define new".

Edit Name
• Enter the name under "edit Name".

Host
• Under "Host", select the name entered on the previous page.

Application Entity Properties

AE Title
• Select "Define new".
¹ The AETs that have already been configured are shown in the pull-down menu.

Edit AE Title
• Enter a name in the "edit AE title" field.

Siemens SP00-000.843.30.02.02 Page 47 of 112 ARCADIS


Medical Solutions 05.08 CS SD SP/CO
48 Syngo

Port number
• Enter the port number for the DICOM device.
• Do not select "Verification".
• Click "Add".

Supported DICOM services

Print
• Print: H
• Click <Save>.
• Click ">".

Print Devices 3

Fig. 28: Print Devices 3

Select HC device
• Select "define new" under "Select HC device".

HC Device
• Enter the name of the camera.
¹ This name will be displayed in the customer interface in the "Print" tab card.
Default Camera

ARCADIS SP00-000.843.30.02.02 Page 48 of 112 Siemens


05.08 CS SD SP/CO Medical Solutions
Syngo 49

• Select this option to specify the camera as the default camera.


The other cameras need to be selected via a pull-down menu in the "Print" tab card in the
customer user interface.

General settings

Type
• Select "DICOM printer".

Class
• Select the DICOM camera in the pull-down menu, and make no other changes on this
page.
If your camera is not yet released or not in the list, contact the responsible Headquar-
ters Support Center.
• Click <Save>.

HC Overview 0

NOTE This menu has no meaning for the service engineer since
the settings and tests can also be performed in the "Print
device" menu.

The "Hard Copy Overview" menu is intended to provide the customer with a menu to
manipulate and test certain camera settings.

Siemens SP00-000.843.30.02.02 Page 49 of 112 ARCADIS


Medical Solutions 05.08 CS SD SP/CO
50 Syngo

Fig. 29: HC Overview

• Do not make any changes under "HC Overview".

LUT Files 0

LUT files (lookup table) for DICOM Basic Print can be imported, deleted, or selected in
this menu.

ARCADIS SP00-000.843.30.02.02 Page 50 of 112 Siemens


05.08 CS SD SP/CO Medical Solutions
Syngo 51

Fig. 30: LUT Files


• Do not make any changes.

NOTE The LUTs are automatically installed along with the software.

• Click ">".

Siemens SP00-000.843.30.02.02 Page 51 of 112 ARCADIS


Medical Solutions 05.08 CS SD SP/CO
52 Syngo

LUT Files 2

Fig. 31: LUT Files 2

NOTE Do not make any changes since all values will be entered
automatically during the software installation.

• Presentation LUT: Q
• Correction LUT: H
• Show only configurable image types: H
• Overall maximum density: “3.00”
• Interpolation for printing: "Cubic”
• Mode: 0

ARCADIS SP00-000.843.30.02.02 Page 52 of 112 Siemens


05.08 CS SD SP/CO Medical Solutions
Syngo 53

HIS/RIS Nodes 0

NOTE DICOM "Worklist" license required.

Configuration of a Hospital Information System (HIS) or Radiology Information System


(RIS) that supports DICOM Basic Worklist or the Detached Study Transfer service. In
many cases, the RIS is connected via an interface box such as the Mitra Broker (Magi-
cLink HL7).

HIS/RIS Nodes

Fig. 32: HIS RIS Nodes

Host Properties
• Under "Select Host", select "Define new".

Host name
• Enter the HIS / RIS node host name.

TCP/IP address
• Enter the IP address.

Connected by
• LAN: H

Siemens SP00-000.843.30.02.02 Page 53 of 112 ARCADIS


Medical Solutions 05.08 CS SD SP/CO
54 Syngo

• Click <Save>.
• Under "Select Host", select the host name entered for HIS/RIS.
• Click ">".

HIS/RIS Nodes 2

Fig. 33: HIS RIS Nodes 2

General Node properties

Host
• Under "Host", select the name entered on the previous page.

Logical Name
• Select "define new".

Edit Name
• Enter the name under "edit Name".

Application Entity Properties

AE Title
• Select "define new".
¹ The AETs that have already been configured are shown in the pull-down menu.

ARCADIS SP00-000.843.30.02.02 Page 54 of 112 Siemens


05.08 CS SD SP/CO Medical Solutions
Syngo 55

Edit AE Title
• Enter a name in the "Edit AE title" field.

Port number
• Enter the port number for the DICOM device.
• Do not select Verification.
• Click "Add".

Supported DICOM services


Basic Worklist

NOTE At a minimum, Basic Worklist must be selected.

• Select "Basic Worklist".


Query Waiting Time
• Check whether 20 sec. is listed; if not, enter it.
Max Query Match Number
• Check whether 200 matches is listed; if not, enter it.
Query Interval
• Check whether 60 minutes is listed; if not, enter it.

NOTE To activate the automatic DICOM Worklist polling function, it


is necessary to select "HIS/RIS update worklist in back-
ground" in the customer user interface under "Option - Con-
figuration - Patient Registration".

Default Basic Worklist node


• Select "Default Basic Worklist node".
Performed Procedure Step

NOTE This item is to be edited only if an MPPS license is available


and MPPS is used.

At the beginning and following patient acquisition, the DICOM Service Modality Performed
Procedure Step (MPPS) sends data to the information system (e.g., RIS). Select this
option if the selected information system is to be used as the destination for this DICOM
service.
Default Catalog Name
• If codes are used, enter the name of the coding catalog used in the information system.
Default PPS node
• Select "Default PPS node".

Siemens SP00-000.843.30.02.02 Page 55 of 112 ARCADIS


Medical Solutions 05.08 CS SD SP/CO
56 Syngo

• Click <Save>.

ARCADIS SP00-000.843.30.02.02 Page 56 of 112 Siemens


05.08 CS SD SP/CO Medical Solutions
Syngo 57

Import/Export 4.6

Directories 0

The system provides the function to import or export DICOM images from/to the file sys-
tem. In this section, specify the location of the import/export directories. They will be
located on the local system or in a shared directory.
For shared directories, you first need to configure the host on which the shared directory
is located.
To import/export DICOM images you need to select "Transfer - import from offline" or
"Import/export to offline" on the customer user interface.

Fig. 34: Directories 1

Select Host
• Click “Select Host” drop down menu to see whether the host is already listed. If so, and
if the listed host is correct, your are done and may continue with the next menu.
¹ The “Host” must be already configured under “Site Info/Station Name” and also
under “TCP/IP LAN/Computer Name”.
• Click “>” to go to the next page.

Siemens SP00-000.843.30.02.02 Page 57 of 112 ARCADIS


Medical Solutions 05.08 CS SD SP/CO
58 Syngo

DVD Drive

Fig. 35: Directories 2

DVD Drive - Definition of Import / Export Directories


• Select under HD device: “CDR_OFFLINE”
¹ If this entry is not present, you can create it by typing it in under “Logical Name”.
• Logical Name: “CDR_OFFLINE”
• Local System: H
• Enter path: “C:\ASPIA\temp\CDR_OFFLINE”
• Drive: “C:\”
• Remote System: Q
• Click <Save>

ARCADIS SP00-000.843.30.02.02 Page 58 of 112 Siemens


05.08 CS SD SP/CO Medical Solutions
Syngo 59

USB Device

Fig. 36: USB Device

USB Device - Definition of Import / Export Directories


• Select under HD device: “USB”
¹ If this entry is not present, you can create it by typing it in under “Logical Name”.
• Logical name: “USB”
• Local system: H
• Enter path: “F:\”
• Drive: “C:\”
• Remote system: Q
• Click <Save>

Siemens SP00-000.843.30.02.02 Page 59 of 112 ARCADIS


Medical Solutions 05.08 CS SD SP/CO
60 Syngo

EPRI 4.7

The EPRI (Electronic Patient Record Interface) is a web interface (Internet Explorer
browser) to a server, which provides additional data about the selected patient.

Server 0

The customer user interface provides a button to open a web browser with a link to an
EPR server. The URL and the type of EPR server need to be specified in this menu.

Fig. 37: EPRI

EPR Interface

Location of EPRServer
• Do not make changes under "Location of EPRServer".

Type of EPRServer
• Do not make changes under "Type of EPRServer".

ARCADIS SP00-000.843.30.02.02 Page 60 of 112 Siemens


05.08 CS SD SP/CO Medical Solutions
Syngo 61

External Devices 4.8

Paper Printer 0

NOTE The local printer has to be installed according to the soft-


ware installation manual.

NOTE Even locally networked screen printers are labeled as paper


printers in syngo.

Fig. 38: Paper Printer


Available paper printer media formats
• Select “A4” and “8x10”.

Siemens SP00-000.843.30.02.02 Page 61 of 112 ARCADIS


Medical Solutions 05.08 CS SD SP/CO
62 Syngo

Printer Settings 0

Fig. 39: Printer Settings

NOTE Under Printer Settings, the background can be selected out-


side the image circle.

The color of the circular shutter


• Select “black”

ARCADIS SP00-000.843.30.02.02 Page 62 of 112 Siemens


05.08 CS SD SP/CO Medical Solutions
Syngo 63

PS LUT Files 0

Fig. 40: PS LUT Files

Postscript Printer LUT Configuration


• “Use LUT for Postscript printers” = H
• “Use floating Zoom factors” = H
• Sony UPD 970 and UPD 990 : Do not select an LUT under “PS LUT directory”.
¹ The “PS LUT directory” field should remain blank.

Siemens SP00-000.843.30.02.02 Page 63 of 112 ARCADIS


Medical Solutions 05.08 CS SD SP/CO
64 Syngo

SystemMgmt 4.9

The IP address of the system management to which the event management agent is to
send its messages must be configured under this item.

NOTE System management settings are not included in this con-


figuration.
A separate document contains the necessary instructions.

Master 0

Fig. 41: Master

Agent Controls 0

The configured “Remote Service Center” (Europe, USA or Asia) under “Site Info” is shown
at this page and tested at this page.

ARCADIS SP00-000.843.30.02.02 Page 64 of 112 Siemens


05.08 CS SD SP/CO Medical Solutions
Syngo 65

Fig. 42: Agent Controls

Installation State
The status is automatically displayed if system management was configured.

Siemens SP00-000.843.30.02.02 Page 65 of 112 ARCADIS


Medical Solutions 05.08 CS SD SP/CO
66 Syngo

Applications 4.10

Menus for setting the application parameters are available under "Applications".

Pat. Registration 0

Fig. 43: Pat Registration

Displayed Patient Registration Dialog Elements


• Make changes only upon customer request.

ARCADIS SP00-000.843.30.02.02 Page 66 of 112 Siemens


05.08 CS SD SP/CO Medical Solutions
Syngo 67

Viewer 0

Fig. 44: Viewer

Viewer configuration settings


• Close all studies when a new one is loaded: Q
• Close all series when a new one is loaded: Q

CorRea 0

In this menu, you can set the attributes that can be modified in the CorRea(=Correct &
Rearrange Platform).

Siemens SP00-000.843.30.02.02 Page 67 of 112 ARCADIS


Medical Solutions 05.08 CS SD SP/CO
68 Syngo

Fig. 45: CorRea

Displayed CorRea dialog elements


• Make changes only upon customer request.

MPPS 0

(Modality Performed Procedure Steps)

NOTE Required settings that go beyond this syngo configuration


can be performed only in cooperation with the responsible
hospital network administrator.

The attributes defined here are displayed on the customer user interface when an MPPS
message is sent to the RIS.

ARCADIS SP00-000.843.30.02.02 Page 68 of 112 Siemens


05.08 CS SD SP/CO Medical Solutions
Syngo 69

Fig. 46: MPPS

Displayed MPPS dialog elements


• Make changes only upon customer request.

Worklist Results 0

The attributes defined here are displayed in the patient registration form after a DICOM
Worklist request.

Siemens SP00-000.843.30.02.02 Page 69 of 112 ARCADIS


Medical Solutions 05.08 CS SD SP/CO
70 Syngo

Fig. 47: Worklist Results

Displayed HISRIS worklist result attributes


• Make changes only upon customer request.

ARCADIS SP00-000.843.30.02.02 Page 70 of 112 Siemens


05.08 CS SD SP/CO Medical Solutions
System Devices 71
5-
USB Settings
5System Devices

Fig. 48: USB Settings

• Allow all devices: Q


• Replace Whitelist: H )
• Append to Whitelist: Q
• Learn mode: “Off”
• Learn on Boot: Q

Siemens SP00-000.843.30.02.02 Page 71 of 112 ARCADIS


Medical Solutions 05.08 CS SD SP/CO
72 System Devices

CAN Bus 5.1

Fig. 49: CAN Bus

CAN Bus
• Bautrate: 125 kBit/s (Default)= Must be selected for Orbic 3D
• Bautrate: 500 kBit/s = Select if connecting a lithotripter (e.g., Variostar) with a baud rate
of 500 kBit/s.

ARCADIS SP00-000.843.30.02.02 Page 72 of 112 Siemens


05.08 CS SD SP/CO Medical Solutions
System Devices 73

Footswitch 5.2

Fig. 50: Footswitch

Here it can be selected if the fluoro pedal of the foot switch releases continious (CFC) or
pulsed fluoroscopy (PFC).
Footswitch Configuration
• Operating Mode of the fixed swtich: CFC

Siemens SP00-000.843.30.02.02 Page 73 of 112 ARCADIS


Medical Solutions 05.08 CS SD SP/CO
74 System Devices

Trigger Injector 5.3

Fig. 51: Trigger Injector

If the system is eqipped with a trigger output for a contrast injector, here this output can
be enabled and the pulse lenght can be adjusted.
• Trigger: Q
• Trigger durat on: 10 (x10ms)

ARCADIS SP00-000.843.30.02.02 Page 74 of 112 Siemens


05.08 CS SD SP/CO Medical Solutions
Imaging System 75
6Imaging System 6-

NOTE The Examination Set Editor may not be opened during con-
figuration of the imaging system.

Siemens SP00-000.843.30.02.02 Page 75 of 112 ARCADIS


Medical Solutions 05.08 CS SD SP/CO
76 Imaging System

Valid Mainsystem 6.1

NOTE During the initial configuration: Enter the serial number of


the main system and click <Save>.

Valid Mainsystem
• Serial Number:
¹ Varic = 12xxx
¹ Orbic = 22xxx
¹ Avantic = 32xxx

ARCADIS SP00-000.843.30.02.02 Page 76 of 112 Siemens


05.08 CS SD SP/CO Medical Solutions
Imaging System 77

Application Fields 6.2

NOTE For application fields that require a license, you must first
load the license. Then you will be able to activate this appli-
cation field as described below.

Application Fields
• To activate an application field, it must be selected under "Application Fields" and then
activated by pressing the "Activate" button.
¹ There are application fields that require a license, e.g., Vascular (DSA).
• Under “Activated Application Fields” select the corresponding application field and click
“Set Default.
¹ The "Default Application Field" is “Ortho/Trauma”.

NOTE A application field must always be entered under "Default


Application Field".

Siemens SP00-000.843.30.02.02 Page 77 of 112 ARCADIS


Medical Solutions 05.08 CS SD SP/CO
78 Imaging System

Examination Set Configuration 6.3

The minimum and maximum filters and LUT‘s for each examination set can be defined
here per the customer’s request.

Examination Set Configuration


• Maximal number of Edge Enhancement Filters per Examination Set: 4
• Maximal number of LUTs per Examination Set: 4

ARCADIS SP00-000.843.30.02.02 Page 78 of 112 Siemens


05.08 CS SD SP/CO Medical Solutions
Imaging System 79

Edge Enhancement Filter 6.4

NOTE Change only with instructions from the factory!

Importing edge enhancement filters


• Select the proper drive under "Drive".
• Select the corresponding file in "Directory and file".
• Click "Import".
¹ The file will be displayed under "Available files".

Siemens SP00-000.843.30.02.02 Page 79 of 112 ARCADIS


Medical Solutions 05.08 CS SD SP/CO
80 Imaging System

DICOM LUT Settings 6.5

DICOM - LUT Data Storage


• Value range of LUT sequence data
- 12 Bit (default)
- 16 Bit
• LUT Descriptor
- 4096/0/16 (default)
- 4096/0/12

NOTE You might solve problems with viewing workstations or


PACS systems, if you either scale up the LUT data to 16 bit
values according the LUT Descriptor or alternatively set the
third value of the LUT Descriptor to 12.
Please only change the default configurations in case of
customer demand or after consulting the service depart-
ment as changing the default configuration may results in
images not compliant on the DICOM standard.

ARCADIS SP00-000.843.30.02.02 Page 80 of 112 Siemens


05.08 CS SD SP/CO Medical Solutions
Imaging System 81

Windowing LUT 6.6

NOTE Change only with instructions from the factory!

Importing Windowing LUT


• Select the proper drive under "Drive".
• Select the corresponding file in "Directory and file".
• Click "Import".
¹ The file will be displayed under "Available files".

Siemens SP00-000.843.30.02.02 Page 81 of 112 ARCADIS


Medical Solutions 05.08 CS SD SP/CO
82 Imaging System

Logarithm LUT 6.7

NOTE Change only with instructions from the factory!

Importing windowing LUT

1. Select the drive under "Drive"


2. Select the file in "Directory and file".
3. Click "Import".
¹ The file will be displayed under "Available files".

Remove windowing LUT


1. Select the file under "Available files".
2. Click "Remove".
¹ The file will be removed under "Available files".

Activate windowing LUT


1. Select the file under "Available files".
2. Click "Activate".
¹ The file will be displayed under "Activated logarithm LUT file".
¹ The default logarithm LUT is “Log100Delog04...”

NOTE A file must always be entered under "Activated logarithm


LUT file".

ARCADIS SP00-000.843.30.02.02 Page 82 of 112 Siemens


05.08 CS SD SP/CO Medical Solutions
Imaging System 83

Metering Sections File 6.8

NOTE Change only with instructions from the factory!

Import the metering section file


1. Select the drive under "Drive".
2. Select the file under "Directory and file".
3. Click "Import".
¹ The file will be displayed under "Available files".

Remove the metering section file


1. Select the required file under "Available files".
2. Click "Remove".
¹ The file will be removed under "Available files".

Activate the metering section file


1. Select the required file under "Available files".
2. Click "Activate".
¹ The file will be displayed under "Activated metering section file".
¹ The default metering section file is “MeteringSection_KPK...”

NOTE A file must always be entered under "Activated Metering Sec-


tion file".

Siemens SP00-000.843.30.02.02 Page 83 of 112 ARCADIS


Medical Solutions 05.08 CS SD SP/CO
84 Imaging System

Display Settings 6.9

Collimator (graphic display for collimator)


• Collimator Fade out Delay: 3 (s)

DAP / Air Kerma

NOTE A DAP chamber with a resolution of 0.01 cGy cm*cm/Pulse is


the prerequisite for the "Air Kerma" monitor display.

NOTE If “Air Kerma” is configured, an adjustment must be made


under <Service>-<Main System>-<Adjustment>-<DAP Air
Kerma>.

• “Dose Area Product” or “Air Kerma” can be selected (as required by the customer).
- Display Type = “Dose Area Product” (default)

Metering Section (display for dominant))


• Color = 30
• Enable Fadeout = H
• Fade out Delay = 1 (sec)

ARCADIS SP00-000.843.30.02.02 Page 84 of 112 Siemens


05.08 CS SD SP/CO Medical Solutions
Imaging System 85

Subtraction Parameters 6.10

The minimum and maximum values for noise suppression for the individual phases are
set here.

Operating Mode (SUB)


Noise Reduction Factor
• Phase A: GGM (k = 32)
• Phase A Low: GGM (k = 32)
• Phase B: GGM (k = 8)
• Phase B Low: GGM (k = 4)
Duration
• Phase A: 75 (frames)
Peak OP
• Frame rate: 5 (fps)

• Click “>” for the next service page.

Operating Mode Roadmap


Noise Reduction Factor
• Phase A: GGM (k = 32)
• Phase A Low: GGM (k = 32)
• Phase B: GGM (k = 8)
• Phase B Low: GGM (k = 4)
• Phase C: GGM (k = 4)
• Phase C Low : GGM (k = 2)

k-Factor for LIH


• Phase C: OFF
• Phase C Low : K = 4

Duration
• Phase A: 75 (frames)

Peak OP
• Frame rate: 5 (fps)

Siemens SP00-000.843.30.02.02 Page 85 of 112 ARCADIS


Medical Solutions 05.08 CS SD SP/CO
86 Imaging System

Motion Detection 6.11

NOTE Change only with instructions from the factory!

MD1 Parameter
• K1: GGM (k = 8)
• K1 motion: GGM (k = 2)
• K2: GGM (k= 8)
• Threshold: 125
• Low Pass Kernel: 11

• Click “>” for the next service page.

MD2 Parameter
• K1: GGM (k = 8)
• K1 motion: GGM (k = 2)
• K2: GGM (k= 8)
• Threshold: 160
• Low Pass Kernel: 11

• Click “>” for the next service page.

MD3 Parameter
• K1: GGM (k = 8)
• K1 motion: GGM (k = 2)
• K2: GGM (k= 8)
• Threshold: 200
• Low Pass Kernel: 11

• Click “>” for the next service page.

MD4 Parameter
• K1: GGM (k = 8)
• K1 motion: GGM (k = 2)
• K2: GGM (k= 8)
• Threshold: 250

ARCADIS SP00-000.843.30.02.02 Page 86 of 112 Siemens


05.08 CS SD SP/CO Medical Solutions
Imaging System 87

• Low Pass Kernel: 11

• Click “>” for the next service page.

MD5 Parameter
• K1: GGM (k = 8)
• K1 motion: GGM (k = 2)
• K2: GGM (k= 8)
• Threshold: 300
• Low Pass Kernel: 11

Siemens SP00-000.843.30.02.02 Page 87 of 112 ARCADIS


Medical Solutions 05.08 CS SD SP/CO
88 Imaging System

Power Off Parameters 6.12

NOTE Change only with instructions from the factory!

NOTE When the "Grace Period" parameter is set to a value less


than "800" (= 8 seconds), a proper system startup is not
guaranteed.
Leave the "Grace period" parameter set at "800".

Power Off Parameters


• Power Off Enable: H
• Power Off Grace Period: 800 (x 10ms = 8 seconds)

ARCADIS SP00-000.843.30.02.02 Page 88 of 112 Siemens


05.08 CS SD SP/CO Medical Solutions
Imaging System 89

Brightness Control 6.13

NOTE Change only with instructions from the factory!

Brightness Control
• AGA unit disabled: Q
- "AGA unit disabled" should be selected only for service purposes.
¹ "AGA unit disabled" is reset by closing the service software.

Brightness Control
• Nominal Brightness Value: 700
• Nominal Brightness Value Limitation Min: 650
• Nominal Brightness Value Limitation Max: 750
• Nominal Brightness Value (Median, Autogamma) 2100
Brightness Control (Metal Button)
• Nominal Brightness Value: 500
• Nominal Brightness Value Limitation Min: 460
• Nominal Brightness Value Limitation Max: 540
• Nominal Brightness Value (Median, Autogamma) 1900

• Click “>” for the next service page.

AGA Unit Parameters (page 2)


AGA Unit Manual Setting
• Lut Source: “Brightness Control Unit”
• Lut Index: 50
Auto Gamma Configuration
• AGA Lamda: 800
• AGA Gamma Min: 20

• Click “>” for the next service page.

AGA Unit Parameter (page 3)


Auto User Curves are applied in
• Native Mode : Q
• Subtraction Mode: Q
• 3D Mode: Q

Siemens SP00-000.843.30.02.02 Page 89 of 112 ARCADIS


Medical Solutions 05.08 CS SD SP/CO
90 Imaging System

Import user curve files


1. Select the drive under "Drive".
2. Select the file under "Directory and file".
3. Click "Import".
¹ The file will be displayed under "Available files".
Remove user curve files
1. Select the required file under "Available files".
2. Click "Remove".
¹ The file will be removed under "Available file".
Activate user curve
1. Select the file under "Available files".
2. Click "Activate".
¹ The file will be displayed under "Activated user curves".
¹ The default user curve is “UserCurve.ucv”

3. Click “>” for the next service page.

Histogram Noise Level (page 4)


ADR
• Dose Level Low: 100
• Dose level Medium: 100
• Dose level High: 100
AGA
• Dose Level Low: 100
• Dose level Medium: 100
• Dose level High: 100

ARCADIS SP00-000.843.30.02.02 Page 90 of 112 Siemens


05.08 CS SD SP/CO Medical Solutions
Imaging System 91

Auto Dose Adjustment 6.14

NOTE Change only with instructions from the factory!

Auto Dose 1 of 6
• Auto dose: H
• Increase: “Absolute”
• Increase Min: 2
• Percent Decrease Max: 50
• Brightness Value: “Mean”
• Minimum Grey Level: 400
• Minimum Pixel Count: 10000
Histogram based brightness value
• Histogram based brightness value: H
• Increase: “Percent”
• Increase Min: 10
• Percent Decrease Max: 40
• Brightness Value: “Mean”
• Minimum Grey Level: 400
• Minimum Pixel Count: 10000

• Click “>” for the next service page.

Auto Dose 2 of 6
• Auto dose: H
• Increase: “Absolute”
• Increase Min: 2
• Percent Decrease Max: 0
• Brightness Value: “Mean”
• Minimum Grey Level: 400
• Minimum Pixel Count: 10000
Histogram based brightness value
• Histogram based brightness value: H
• Increase: “Percent”
• Increase Min: 10
• Percent Decrease Max: 0
• Brightness Value: “Mean”

Siemens SP00-000.843.30.02.02 Page 91 of 112 ARCADIS


Medical Solutions 05.08 CS SD SP/CO
92 Imaging System

• Minimum Grey Level: 400


• Minimum Pixel Count: 10000

• Click “>” for the next service page.

Auto Dose Adjustment 3 of 6


• Auto dose: H
• Increase: “Absolute”
• Increase Min: 2
• Percent Decrease Max: 30
• Brightness Value: “Mean”
• Minimum Grey Level: 400
• Minimum Pixel Count: 10000
Histogram based brightness value
• Histogram based brightness value: H
• Increase: “Absolute”
• Increase Min: 2
• Percent Decrease Max: 30
• Brightness Value: “Mean”
• Minimum Grey Level: 400
• Minimum Pixel Count: 10000

• Click “>” for the next service page.

Auto Dose Adjustment 4 of 6


• Auto dose: H
• Increase: “Absolute”
• Increase Min: 2
• Percent Decrease Max: 30
• Brightness Value: “Mean”
• Minimum Grey Level: 400
• Minimum Pixel Count: 10000
Histogram based brightness value
• Histogram based brightness value: H
• Increase: “Percent”
• Increase Min: 10
• Percent Decrease Max: 30
• Brightness Value: “Mean”

ARCADIS SP00-000.843.30.02.02 Page 92 of 112 Siemens


05.08 CS SD SP/CO Medical Solutions
Imaging System 93

• Minimum Grey Level: 400


• Minimum Pixel Count: 10000

• Click “>” for the next service page.

Auto Dose Adjustment 5 of 6


• Auto dose: H
• Increase: “Absolute”
• Increase Min: 2
• Percent Decrease Max: 35
• Brightness Value: “Mean”
• Minimum Grey Level: 400
• Minimum Pixel Count: 10000
Histogram based brightness value
• Histogram based brightness value: H
• Increase: “Absolute”
• Increase Min: 2
• Percent Decrease Max: 50
• Brightness Value: “Mean”
• Minimum Grey Level: 400
• Minimum Pixel Count: 10000

• Click “>” for the next service page.

Auto Dose Adjustment 6 of 6


• Auto dose: H
• Increase: “Absolute”
• Increase Min: 2
• Percent Decrease Max: 0
• Brightness Value: “Mean”
• Minimum Grey Level: 400
• Minimum Pixel Count: 10000
Histogram based brightness value
• Histogram based brightness value: H
• Increase: “Absolute”
• Increase Min: 10
• Percent Decrease Max: 0
• Brightness Value: “Mean”

Siemens SP00-000.843.30.02.02 Page 93 of 112 ARCADIS


Medical Solutions 05.08 CS SD SP/CO
94 Imaging System

• Minimum Grey Level: 400


• Minimum Pixel Count: 10000

ARCADIS SP00-000.843.30.02.02 Page 94 of 112 Siemens


05.08 CS SD SP/CO Medical Solutions
Imaging System 95

Auto Lut 6.15

NOTE Change only with instructions from the factory!

Auto LUT Adjustment


• Histogram equalization: H
• Freeze histogram equalization: Q

Import limiter curve file


1. Select the drive under "Drive".
2. Select the file under "Directory and file".
3. Click "Import".
¹ The file will be displayed under "Available files".

Remove limiter curve file


• Select the required file under "Available files".
• Click "Remove".
¹ The file will be removed under "Available".

Activate limiter curve file


1. Select the required file under "Available files".
2. Click "Activate".
¹ The file will be displayed under "Activated limiter file".
¹ The default limiter curve file is “LimiterCurve_0G35_F50.lcv”

NOTE A file must always be entered under "Activated limiter file".

Siemens SP00-000.843.30.02.02 Page 95 of 112 ARCADIS


Medical Solutions 05.08 CS SD SP/CO
96 Imaging System

Camera 6.16

NOTE Change only with instructions from the factory!

Camera Gain
• DR single: Avantic = 1 / Varic, Orbic = 2
• DR single sum: 2
• CFC: 2
• PFC IDL: 2
• PFC PDL: 2
• DCM: 1
• 3D: 1
• SUB/ROAD: 2

ARCADIS SP00-000.843.30.02.02 Page 96 of 112 Siemens


05.08 CS SD SP/CO Medical Solutions
Imaging System 97

Circle Mask 6.17

Circle Mask
• Radius: 508 (pixel)
• Circle Mask Display: “on”
¹ A switch to "transparent" can be made for service purposes.

Siemens SP00-000.843.30.02.02 Page 97 of 112 ARCADIS


Medical Solutions 05.08 CS SD SP/CO
98 Imaging System

Hardcopy Configuration 6.18

NOTE If “Hardcopy button sends image to film task card” is dese-


lected, printing starts immediately after the "Hardcopy" but-
ton is pressed. Fluoroscopy will be blocked during this
period.

Hardcopy Configuration
• Hardcopy button sends image to Filming Taskcard: H

ARCADIS SP00-000.843.30.02.02 Page 98 of 112 Siemens


05.08 CS SD SP/CO Medical Solutions
Imaging System 99

Homogenisation 6.19

NOTE Change only with instructions from the factory!

Homogenisation
• Homogenisation mode: “Brightness and Images”

Import matrix definition file


1. Select the drive under "Drive".
2. Select the file under "Directory and file".
3. Click "Import".
¹ The file will be displayed under "Available files".

Remove matrix definition file


1. Select the required file under "Available files".
2. Click "Remove".
¹ The file will be removed under "Available files".

Activate matrix definition file


1. Select the required file under "Available files".
2. Click "Activate".
¹ The file will be displayed under "Activated matrix file".
¹ The default matrix file is “HomoMatrixKPK.hom”

NOTE A file must always be entered under "Activated matrix file".

Siemens SP00-000.843.30.02.02 Page 99 of 112 ARCADIS


Medical Solutions 05.08 CS SD SP/CO
100 Main System Varic / Orbic
7Main System Varic / Orbic 7-

NOTE In the main system service pages, you must press the
"Enter” key after entering a parameter.
When this is done, the "Save” button is displayed.

NOTE Once the "Save" button has been selected in the individual
screens, the message "..... successfully saved” appears fol-
lowing a successful save.
--> Click "OK”.

NOTE Depending on the progress of the configuration, the "Next”


button can also be displayed as the "Finish” button.

ARCADIS SP00-000.843.30.02.02 Page 100 of 112 Siemens


05.08 CS SD SP/CO Medical Solutions
Main System Varic / Orbic 101

Configuration Varic / Orbic 7.1

In the service menu click <MainSystem> and <Next>.

Main System 0

• System Type: Check to make sure that the correct system type is displayed here.
- Depending on the system, "ARCADIS Varic”, "ARCADIS Orbic” or ARCADIS Orbic
3D” can be displayed here.
• Serial No.: Enter the system serial number
- Varic: 12xxx (xxx = last 3 digits of serial number)
- Orbic: 22xxx (xxx = last 3 digits of serial number)
• Disable X-Ray: "no” (default)
• Log old messages: "yes” (default)
• Foot switch fix Mode: "Right”
- defines the foot switch pedal for fluoro mode.
¹ change only on customer demand to “Left”

Load Counter 0

• New SERIAL No Single Tank: Enter the serial number of the "Single Tank”.
¹ The serial number can be found on the single tank housing.

Options 0

Laser
• Laser Existing: If a laser light is installed select "yes”.

Dose Area Product


• DAP chamber existing: If a DAP chamber is installed select "yes”.
• Constant: 5000 (Tw) = default
¹ Enter the value from chamber test
• Resolution: “ 0.01 cGy cm*cm/Puls”

Current Limitation 0

NOTE The "Tube Current Limitation” is required only for Denmark.


The required setting is performed during startup.

Siemens SP00-000.843.30.02.02 Page 101 of 112 ARCADIS


Medical Solutions 05.08 CS SD SP/CO
102 Main System Varic / Orbic

Tube Current Limitation


• Limit Max mA: 40 (x10)
• Lock Buttons: “no”

Buzzer 0

NOTE The required setting is country-specific and is therefore


made during start-up.

Buzzer settings
• Buzzer Mode (direct, cassette): “After Exposure”
• Buzzer Mode (indirect): “Buzzer Off”

Blocktime settings
• Warning Time Limit: 50 (min/10)
• Blocking Time Limit:
- For DHHS counties: 0 (min/10)
- For non DHHS countries: 50 (min/10)

NOTE With setting the “Blocking Time Limit” to “0”, the x-ray block-
ing is disabled.

Image Intensifier 0

• New serial number I.I.: Value see I.I. protocol


• Full format diameter: Depends on adjustment, default = 215
• Zoom 1 diameter: Depends on adjustment, default = 155
• SID_JJA (FFA): 1005
• GX at installation: Value see I.I. protocol
• Vignetting: 2 (check-up during IQ test)

ARCADIS SP00-000.843.30.02.02 Page 102 of 112 Siemens


05.08 CS SD SP/CO Medical Solutions
Main System Avantic 103
8Main System Avantic 8-

NOTE In the main system service pages, you must press the
"Enter” key after entering a parameter.
When this is done, the "Save” button is displayed.

NOTE Once the "Save" button has been selected in the individual
screens, the message "..... successfully saved” appears fol-
lowing a successful save.
--> Click "OK”.

NOTE Depending on the progress of the configuration, the "Next”


button can also be displayed as the "Finish” button.

Siemens SP00-000.843.30.02.02 Page 103 of 112 ARCADIS


Medical Solutions 05.08 CS SD SP/CO
104 Main System Avantic

Configuration Avantic 8.1

In the service menu click <MainSystem> and <Next>.

Main System 0

• System Type: The system type “ARCADIS Avantic-33” is displayed here.


• Serial No Main Unit: Enter the system serial number
- Avantic: 32xxx (xxx = last 3 digits of serial number)
• Enable internal tests: “No”
- "Yes” is required only by the Engineering Department.
• Disable X-Ray: "No”
• Foot switch fix Mode: "Left”
- defines the foot switch pedal for fluoro mode.
¹ change only on customer demand to “Right”

Image Intensifier 0

• New serial number I.I.: Value see I.I. protocol


• Full Format diameter: Depends on adjustment, default = 300
• Zoom 1 diameter: Depends on adjustment, default = 215
• Zoom 2 diameter: Depends on adjustment, default = 160
• Zoom 1 diameter: Depends on adjustment, default = 120
• SID_JJA (FFA): 1000
• GX at installation: Value see I.I. protocol
• Vignetting: 2 (check-up during IQ test)

Load Counter 0

• New SERIAL No Single Tank: Enter the serial number of the "Single Tank”.
¹ The serial number can be found on the single tank housing.

Options 0

Laser
• Laser existing: If a laser light is installed select "Yes”.

Dose Area Product


• DAP chamber existing: If a DAP Chamber is installed select "Yes”.

ARCADIS SP00-000.843.30.02.02 Page 104 of 112 Siemens


05.08 CS SD SP/CO Medical Solutions
Main System Avantic 105

• Constant: 5000 (Tw) = default


¹ Enter the value from chamber test
• Resolution: “ 0.01 cGy cm*cm/PULS”

Diaphragm
• Power on Pos: “+90 DEGREE”

Current Limitation 0

NOTE The "Tube Current Limitation” is required only for Denmark.


The required setting is performed during startup.

Tube Current Limitation


• Limit Max mA: 40 (x10)
• Lock Buttons: “No”

Buzzer 0

NOTE The required setting is country-specific and is therefore


made during start-up.

Buzzer Settings
• Buzzer mode during x-ray: “BUZZER OFF”

Blocktime Settings
• Warning Time Limit: 50 (min/10)
• Blocking Time Limit:
- For DHHS counties: 0 (min/10)
- For non DHHS countries: 50 (min/10)

NOTE With setting the “Blocking Time Limit” to “0”, the x-ray block-
ing is disabled.

Siemens SP00-000.843.30.02.02 Page 105 of 112 ARCADIS


Medical Solutions 05.08 CS SD SP/CO
106 Backup of parameter changes
9Backup of parameter changes 9-

NOTE After the parameters have been changed, a backup must be


performed.

NOTE The monitor trolley and the basic unit must be connected.
(Plug X10 is connected to socket X10.)

1. Switch on the system and wait until it has booted.


2. Open local service.
3. Insert an empty CD-R into the DVD drive.
4. Select “Command” and “Backup”.
5. As backup drive, select “(R) CD-R” in the “Drives” list field.
6. Back up the following packages:
¹ SW settings 02
¹ ASPIA Settings
¹ 3D Reconstruction settings (only for ARCADIS Orbic with 3D reconstruction)
¹ ExamSet
¹ Security settings (optional)
¹ Main system
7. After the backup, remove the CD-ROM from the DVD drive, write the system name,
system serial number, date, and time on the CD-ROM, and insert it into the system log-
book.

ARCADIS SP00-000.843.30.02.02 Page 106 of 112 Siemens


05.08 CS SD SP/CO Medical Solutions
Network Checklist 107
10-
Checklist
10Network Checklist

NOTE If the system is connected to a network, the following check-


list must be filled out and filed in the system binder.

Person responsible for the network configuration at the customer site 0

Name:
Phone/fax:
E-mail:

System network data 0

Identification: Enter under:


* Configuration/ Local Host/
Computer name
TCP IP address
IP address:
TCP/IP address * Configuration/ Local Host/
TCP IP address
* Configuration/ Local Host/
Subnet mask
TCP IP address
* Configuration/ Local Host/
Gateways
TCP IP address
General: Local AE title for:
* Configuration/DICOM/Gen-
HIS/RIS
eral
* Configuration/DICOM/Gen-
Study transfer
eral
* Configuration/DICOM/Gen-
Print
eral

Comment:

Siemens SP00-000.843.30.02.02 Page 107 of 112 ARCADIS


Medical Solutions 05.08 CS SD SP/CO
108 Network Checklist

Network data for HIS/RIS network partner (only with HIS/RIS option) 0

Configured: Yes No
Product: Version: Manufacturer:

Host properties:
* Configuration/DICOM/Net-
Host name
work nodes
* Configuration/DICOM/Net-
TCP/IP address
work nodes
General node properties
* Configuration/DICOM/Net-
Logical name
work nodes
Application entity
* Configuration/DICOM/Net-
AE title
work nodes
* Configuration/DICOM/Net-
Port number
work nodes
Comment:

Network data for reporting console/archiving system 0

Configured: Yes No
Product: Version: Manufacturer:

Host properties:
* Configuration/DICOM/Net-
Host name
work nodes
* Configuration/DICOM/Net-
TCP/IP address
work nodes
General node properties:
* Configuration/DICOM/Net-
Logical name
work nodes
Application entity:

ARCADIS SP00-000.843.30.02.02 Page 108 of 112 Siemens


05.08 CS SD SP/CO Medical Solutions
Network Checklist 109

* Configuration/DICOM/Net-
AE title
work nodes
* Configuration/DICOM/Net-
Port number
work nodes
Supported DICOM ser-
vices:
Yes No * Configuration/DICOM/Net-
Storage
work nodes
Yes No * Configuration/DICOM/Net-
Storage Commitment
work nodes
Node name: * Configuration/DICOM/Net-
work nodes
Yes No * Configuration/DICOM/Net-
Query
work nodes
Yes No * Configuration/DICOM/Net-
Retrieve
work nodes

Comment:

Network data for reporting console/archiving system 0

Configured: Yes No
Product: Version: Manufacturer:

Host properties:
* Configuration/DICOM/Net-
Host name
work nodes
* Configuration/DICOM/Net-
TCP/IP address
work nodes
General node properties:
* Configuration/DICOM/Net-
Logical name
work nodes
Application entity:

Siemens SP00-000.843.30.02.02 Page 109 of 112 ARCADIS


Medical Solutions 05.08 CS SD SP/CO
110 Network Checklist

* Configuration/DICOM/Net-
AE title
work nodes
* Configuration/DICOM/Net-
Port number
work nodes
Supported Dicom ser-
vices:
Yes No * Configuration/DICOM/Net-
Storage
work nodes
Yes No * Configuration/DICOM/Net-
Storage Commitment
work nodes
Node name: * Configuration/DICOM/Net-
work nodes
Yes No * Configuration/DICOM/Net-
Query
work nodes
Yes No * Configuration/DICOM/Net-
Retrieve
work nodes

Comment:

ARCADIS SP00-000.843.30.02.02 Page 110 of 112 Siemens


05.08 CS SD SP/CO Medical Solutions
Changes to previous version 111
11Changes to previous version 11-

Chapter 4 Changes
Import/Export - Directories with USB device configuration completed

Siemens SP00-000.843.30.02.02 Page 111 of 112 ARCADIS


Medical Solutions 05.08 CS SD SP/CO