R5 Troubleshooting Guide

The document describes an issue with a medical device. It states the model is an R5 with an horizontal band appearing over the scan area that disappears with the quick scan option. The reason given is a faulty TGC part of the control panel. The solution is to replace the control panel board.

Uploaded by

Osmel Hernandez
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 9

1.

Basic Info
- Model Name : R5
- S/W version : N/A
- F/W version : N/A

2. Symptom:

Horizontal band over scan area.The band


-

disappears on using Q SCAN option.

3. Reason:

- TGC part of CP faulty

4. Solution:

- Replace CP board.
1. Basic information

A. Model name :R5

B. S/W Version : NA

C. F/W version : NA

2. Symptom

- System hang up while booting


3. Reason

- Defective DDM

4. Solution

- Replace DDM module


1. Basic information

A. Model name : R5

B. S/W Version : NA

C. F/W version : NA

2. Symptom

- HDD not recognized

3. Reason

- PC failure

4. Solution

- Replace the PC
1. Basic information
A. Model name : R5
B. S/W Version : NA
C. F/W version : NA
2. Symptom

- Monitor Flickering after replaced the main board

3. Reason
- Dip switch mismatch
4. Solution
- Select LCD/LED switch on the main board, please see attached file.
1. Basic information

A. Model name : R5

B. S/W Version : NA

C. F/W version : NA

2. Symptom

- Booting failure

3. Reason

- HDD itself, main board (PC parts) malfunction

4. Solution

- Reload the SW

- Change the HDD

- Replace the main board (PC board)


1. Basic information

A. Model name : R5

B. S/W Version : NA

C. F/W version : NA

2. Symptom

- Booting failure

3. Reason

- HDD bad sector or main board malfunction

4. Solution

- Reinstall SW

- Change HDD

- Change main board


1. Basic Info
- Model Name : SonoAce R5
- S/W version : N/A
- F/W version : N/A

2. Symptom Machine does not boot up


- The problem is that whenever we tried to start the ultrasound, the machine starts but after 5
second it’s turned off automatically.

- Replaced the DDM. but nothing changed.

- Replaced the ADM board (Assy trans) but, the same symptom persists.

3. Reason
- Defected on Ass'y PC part

4. Solution

- Replace the PC part. (Not whole main board)


1. Basic information

A. Model name : R3

B. S/W Version : NA

C. F/W version : NA

2. Symptom

- Booting failure, error message


3. Reason

- SW failure

4. Solution

- Mostly, booting failure is from PC section (ex, SW, HDD itself or PC module)

- Please see attached manual to check the SW.


1. Basic information

A. Model name : R5

B. S/W Version : NA

C. F/W version : NA

2. Symptom

- 90˚ rotated image printed out


3. Reason

- “abnormal” printed image case in UP-D897

4. Solution

- Please try this setting when you faced any “abnormal” printed image case in UP-
D897