RG 2018170038
RG 2018170038
org
                                                                                                                        833
                                                                                                                              Imaging Physics
            Pictorial Review of Digital Radiog-
            raphy Artifacts1
Alisa I.Walz-Flannigan, PhD
Kimberly J. Brossoit, RT                            Visual familiarity with the variety of digital radiographic artifacts
Dayne J. Magnuson, RT                               is needed to identify, resolve, or prevent image artifacts from creat-
Beth A. Schueler, PhD                               ing issues with patient imaging. Because the mechanism for image
                                                    creation is different between flat-panel detectors and computed
Abbreviation: AAPM = American Association           radiography, the causes and appearances of some artifacts can be
of Physicists in Medicine                           unique to these different modalities. Examples are provided of ar-
RadioGraphics 2018; 38:833–846                      tifacts that were found on clinical images or during quality control
https://doi.org/10.1148/rg.2018170038
                                                    testing with flat-panel detectors. The examples are meant to serve
                                                    as learning tools for future identification and troubleshooting of
Content Codes:
                                                    artifacts and as a reminder for steps that can be taken for preven-
1
 From the Department of Radiology, Mayo             tion. The examples of artifacts provided are classified according to
Clinic, 200 First St SW, Rochester, MN 55901.
Presented as an education exhibit at the 2016       their causal connection in the imaging chain, including an equip-
RSNA Annual Meeting. Received March 6,              ment defect as a result of an accident or mishandling, debris or
2017; revision requested October 13 and re-
ceived November 18; accepted December 6. For
                                                    gain calibration flaws, a problematic acquisition technique, signal
this journal-based SA-CME activity, the authors,    transmission failures, and image processing issues. Specific artifacts
editor, and reviewers have disclosed no relevant    include those that are due to flat-panel detector drops, backscatter,
relationships. Address correspondence to
A.I.W.F. (e-mail: walzflannigan.alisa@mayo.edu).    debris in the x-ray field during calibration, detector saturation or
©
 RSNA, 2018
                                                    underexposure, or collimation detection errors, as well as a variety
                                                    of artifacts that are processing induced.
 SA-CME Learning Objectives                         ©
                                                     RSNA, 2018 • radiographics.rsna.org
After completing this journal-based SA-CME
     activity, participants will be able to:
■■Identify digital radiographic artifacts
and the circumstances that resulted in                                                Introduction
their appearance.                                  An artifact on an image is a feature that does not correlate with
■■Discuss   the cause of the artifact.             the physical properties of the subject being imaged and may
■■Describe methods to eliminate the arti-          confound or obscure interpretation of that image. In this ar-
fact and improve image quality.
                                                   ticle, examples of artifacts from flat-panel detector–based digital
    See www.rsna.org/education/search/RG.          radiographic systems are presented. The examples of artifacts are
                                                   paired with their cause and resolution as a means to aid in the
                                                   future identification, resolution, or prevention of digital radio-
                                                   graphic artifacts that may affect the quality of patient care. The
                                                   artifact examples provided were taken from clinical images as well
                                                   as routine quality control testing.
                                                      Flat-panel detector–based digital radiographic systems differ
                                                   in their image creation mechanism from computed radiography,
                                                   having an intrinsic pixel matrix and differences in signal detection
                                                   and processing. This difference can contribute to artifact types that
834  May-June 2018	                                                                                      radiographics.rsna.org
                                                                  Detector Drops
                                                                  Wireless flat-panel detectors get dropped.
                                                                  Drops can damage the detector system in
                                                                  several ways through fracturing of the flat-
                                                                  panel detector, through disruption of readout
                                                                  electronics, or through shifting gain and offsets.
                                                                  Some vendor systems may have integrated
                                                                  diagnostics software to alert the user when
                                                                  detector drops are of sufficient magnitude for
                                                                  a risk to image quality and may provide follow-
                                                                  up instructions for how to respond. Following
Figure 1.  Detector drop. Chest radiograph obtained with          instructions that direct a user to seek service
a portable radiographic unit shows a white band (arrow) at
                                                                  or discontinue use of a damaged detector can
the top of the image. The white band appeared on images
obtained after a detector drop.                                   prevent unnecessary repeated images. However,
                                                                  because drops occur that can be withstood by
                                                                  detectors without apparent damage and be-
                                                                  cause some systems do not provide user feed-
                                                                  back about the effect of a detector drop, the
                                                                  truth of the drop’s effect is sometimes found in
                                                                  an image itself. We provide examples of images
                                                                  affected subsequent to detector drops.
scintillator layer and the thin-film transistor      provide for greater ease of handling to avoid
layer. The image shown in Figure 3 is presumed       drops, such as incorporation of a handgrip. In
to be the result of minute shifts between the        addition, protective cases are available, which
scintillator layer and the thin-film transistor      some practices may find useful. Providing
layer of a gadolinium oxysulfide flat-panel detec-   technologists with a safe strategy and a dem-
tor; such shifts affect the effectiveness of the     onstration of proper detector bagging, detector
flat-field calibration.                              cleaning, or holding wireless flat-panel detectors
                                                     during battery changes may provide additional
Avoidance or Remedy.—Some detector drops             reduction in detector drop–related failures. For
might result in shifts in signal gain or offset      any practice, good communication detailing
that can be compensated for with recalibration       when detector drops occur and the situations in
of the detector. The appearance of a speckle         which they occur can help to identify targets for
pattern in Figure 3 subsequent to a detector         reducing the number of detector drop–related
drop was resolved by a detector recalibration.       artifacts.
Likewise, we have seen detector artifacts that
initially look like dead lines but were able to be   Liquid Contamination
addressed through recalibration, indicating that     Although it is not a novel concept that liquids do
signal readout was affected but not broken.          not mix well with sensitive electronic equipment,
Other detector drops may result in mechani-          unless such an incident is noticed, reported, and
cal failures in signal readout, creating artifacts   addressed, it may result in imaging failures dur-
like those seen in Figures 1 and 2. In these         ing examination of a patient.
circumstances, it is not possible to address the
failures through detector calibration. Instead,      Description of Artifact Appearance.—After a
the detector needs to be replaced.                   leak in a patient’s water mattress during imag-
   Some vendors advertise detectors that have        ing, an image was acquired that shows banding
greater robustness with drops or detectors that      throughout (Fig 4).
RG  •  Volume 38  Number 3	                                                                      Walz-Flannigan et al  837
Figure 5.  Backscatter. (a) Radiograph obtained with a portable radiographic unit shows a shadow of detector electronics, created
by backscattered x-rays. (b) Image obtained when the detector was exposed upside down shows the detector electronics. (c) High-
lighted radiograph shows the matching features (yellow) that appear on the image of the detector electronics (b) and on the patient
radiograph with the backscatter artifacts (a), to highlight the similarities.
Cause.—Although the detector had been placed                        Cause.—Backscatter artifacts are more likely to
into a plastic bag, the bag was not one that could                  appear in situations with more scatter (ie, with
be effectively sealed. A substantial leak from a                    large patients or wide-open collimation) or if the
patient’s water mattress resulted in water getting                  x-ray beam is not fully intercepted by the de-
inside the bag. It is presumed that water infiltra-                 tector. This backscatter-related artifact is more
tion into the detector caused damage or disrup-                     possible with certain imaging geometries that
tion in the readout electronics that resulted in a                  use a wireless detector outside a table or upright
lack of signal, or false signal, that appears along                 holder, a situation in which it may be more chal-
the readout lines.                                                  lenging to properly angle the detector or use tight
                                                                    collimation (as in some cross-table imaging).
Avoidance or Remedy.—For both infection
control and prevention of liquid infiltration, wire-                Avoidance or Remedy.—If projections of detector
less flat-panel detectors are often bagged. Using                   electronics are visible on a patient image, op-
watertight protection, such as a sealable plastic                   tions for a repeated image include using tighter
bag, can better protect a detector from liquid ex-                  collimation (decreasing the field of view), verify-
posure such as that resulting in the failure shown                  ing good imaging geometry, or placing additional
in Figure 4. Vendors of newer models of flat-panel                  shielding behind the detector, such as a lead
detectors also advertise better water sealing to                    apron or plate to block the backscatter from strik-
prevent damage. The detector cannot recover                         ing the detector.
from the damage demonstrated in Figure 4, and
the detector needed to be replaced.                                 Image Saturation
                                                                    Information may be lost (“clipped”) from an
      Artifacts Related to Acquisition                              image when an exposure exceeds the dynamic
                 Technique                                          range imposed by an image processing algorithm
                                                                    or in some circumstances in which the exposure
Backscatter                                                         exceeds the dynamic range of the image receptor.
Image artifacts can be created when backscat-                       It can be difficult to tell which is the source of the
tered radiation reaches detector elements through                   problem unless one is able to access an unpro-
the back of the detector (3).                                       cessed (“for-processing”) image. Although either
                                                                    computed radiography or digital radiography
Description of Artifact Appearance.—The                             provides greater exposure latitude than film, thus
shadow image of detector electronics appears                        reducing the likelihood of image receptor satura-
superimposed on a patient image in Figure 5a.                       tion within the anatomy of a patient image, such
Corresponding features between an image of the                      saturation may be more likely with digital radi-
detector electronics and the backscatter projec-                    ography than with computed radiography. The
tion of the electronics on the patient image are                    saturation point for digital radiography is typically
shown in Figure 5b and 5c.                                          much lower than that for computed radiography
838  May-June 2018	                                                                                  radiographics.rsna.org
                Figure 6.  Image saturation. (a) Abdominal radiograph obtained with a portable radiographic unit shows the
                effect of excessive image exposure that resulted in a patient’s femurs and surrounding soft tissue being lost from
                the image. (b) Upper abdominal radiograph shows image saturation (arrow) at the diaphragm, a location in
                which image saturation is more commonly seen. For both images, a simulated-grid acquisition was used. Both
                acquisitions used manual techniques for one size above what would have ideally been recommended on the
                basis of patient thickness, as listed on the technique charts. The exposure was approximately double what was
                recommended for the patient in a and was 40% greater than what was recommended for the patient in b.
(eg, a maximum image receptor air kerma for                      simulated-grid imaging, which does not use a
one flat-panel detector is approximately 90 μGy,                 grid but employs a scatter-suppression image
compared with approximately 440 μGy for a com-                   processing algorithm.
puted radiographic plate). (Dynamic range limita-
tions are as reported by a vendor representative.)               Cause.—The manual technique used to ob-
   Data from the report of the AAPM Imaging                      tain the image in Figure 6a was approximately
Physics Committee Task Group 232 provide                         twice what was recommended for the measured
some perspective on the likelihood of observing                  abdominal thickness but was only one size step
detector saturation (6). For imaging an abdo-                    larger on the manual technique chart that was
men, if the target image receptor air kerma is                   provided to the technologists. The image in
3.4 μGy and the image has a pixel value dy-                      Figure 6b also used a technique one size step
namic range of 50 (including tissues up to the                   larger than might have been recommended on
skin line), then with a detector saturation point                the basis of the patient thickness, which resulted
of 90 μGy, one would anticipate seeing detec-                    in using a setting of 40% more milliampere-
tor saturation for a median exposure of 12 μGy.                  seconds than would have been desired.
This value for a median exposure would not be                        We saw an increase in image saturation arti-
anticipated in a typical acquisition. However, in                facts with the initial use of simulated-grid acqui-
situations with higher image receptor air kerma                  sition, most frequently along the diaphragm at
targets or for patient images that may have a                    abdominal imaging, as shown in Figure 6b. The
greater pixel value dynamic range (greater dif-                  increased occurrence of image saturation arti-
ference between the thickest and thinnest ana-                   facts with simulated-grid images, compared with
tomic features), there could be a higher prob-                   gridded images, can be logically related to a
ability of detector saturation.                                  higher recommended detector air kerma relative
                                                                 to gridded techniques. (A higher recommended
Description of Artifact Appearance.—The                          target detector air kerma is inferred from
abdominal radiograph in Figure 6a represents a                   the default technique guides provided by the
dramatic case of image saturation (“clipping”)                   manufacturer, which show a 0%–35% reduc-
in which the lower part of the patient’s anatomy                 tion in the milliampere-second setting between
is completely missing from the image. Figure                     gridded and simulated-grid chest and abdomen
6b is an example of saturation-related clip-                     techniques for the same detector for a medium
ping at the diaphragm in an upper abdominal                      patient.) Technique recommendations, as pro-
radiograph. Attempts to reasonably recover the                   vided by the vendor, suggested between a 0%
image content and quality with image process-                    and 35% reduction in the milliampere-second
ing were unsuccessful for each example. Both                     setting between grid use and simulated-grid use
images in Figure 6 were acquired by using                        in acquisitions. However, maintaining the same
RG  •  Volume 38  Number 3	                                                       Walz-Flannigan et al  839
detector exposure without a grid would require           Avoidance or Remedy.—Some degree of limita-
a reduction of approximately 75% in technique.           tion in flat-field correction may be anticipated
The net effect is that the recommended image             when the detector receives exposures outside
receptor air kerma for simulated-grid imaging is         its dynamic range. Because this type of artifact
higher than that for imaging with a grid. Aiming         has caused some confusion in the past, it has
for a higher detector target in effect increases         been important to educate technologists about
the likelihood of image saturation, creating a           what this artifact looks like, so they understand
greater need for technique optimization and              that it is not related to a failure in grid move-
careful technique selection.                             ment or suppression and so they learn to not
                                                         reject images unless the artifact extends into the
Avoidance or Remedy.—The most basic tool for             anatomy. This type of artifact is also less likely
avoidance of image saturation is to ensure that          when well-optimized target exposures are used
the designated size-based techniques have been           for the images.
followed. For manual techniques, the source-
to-detector distances and the thickness of the           Inverse Focal Spots and Calibration
anatomy need to be measured. In addition, we             Artifacts
found that we needed to provide more size-spe-           Debris in the x-ray beam creates an image on the
cific tailoring and further optimization to avoid        detector, as might be expected of any x-ray–at-
image clipping with the simulated-grid acquisi-          tenuating object found between the x-ray source
tion technique. Technique optimization involved          and any image receptor. In particular, a small
extensive review of patient images, acquisition          speck of x-ray–attenuating material near the head
techniques, and technologist-recorded patient            of the x-ray tube can create an inverse image
sizes. With guidance from radiologists, techniques       of the focal spot, acting as a pinhole camera as
that provided adequate noise properties with a           described by Walker (8).
lower risk of artifacts were determined.
                                                         Description of Artifact Appearance.—A single
 Artifacts from Limitations or Flaws in                  trapezoidal opacity can be seen in the flat-field im-
          Detector Calibration                           age in Figure 8a, surrounded by the magnified and
                                                         offset appearance of its inverse (the area of lucency
Detector Calibration Limitations                         around the opacity). Figure 8b shows similarly ap-
                                                         pearing artifacts but of different sizes, locations, and
Description of Artifact Appearance.—The image            orientations. Both images were acquired as part of
in Figure 7 shows vertical striping in the area of raw   quality control testing of portable x-ray units, with
radiation. The lines are seen only in the background     no added material in the x-ray beam, at 80 kVp,
on the image and do not extend into the anatomy.         and with an image receptor air kerma of 17.5 μGy.
Cause.—The artifact lines demonstrate the                Cause.—The cause of these artifacts was found
limits of gain and offset corrections of the detec-      when the collimator was removed from the x-ray
tor calibration when the exposure is far from            tube (Fig 9). Lead shavings seen near the head
the calibration conditions (as is the case when          of the x-ray tube were created by the collimator
the detector is exposed to the unattenuated              scraping against the beam-shaping port cylinder.
x‑ray beam). This evidence of detector structure         The lead shavings created opaque-appearing
disappears at lower exposures. The threshold for         pinhole images of the focal spot, with the size and
the appearance of detector structure will differ         location depending on the location of the speck
for detectors that have different offset tolerances      relative to the focal spot and the detector. The
between detector elements.                               darker trapezoids (Fig 8) represent the former
840  May-June 2018	                                                                      radiographics.rsna.org
                            Figure 8.  Inverse focal spots and calibration artifacts. (a) Radiograph obtained with
                            flat-field acquisition shows a trapezoidal opacity. The trapezoidal opacity is the pinhole
                            image of the focal spot caused by lead debris near the head of the x-ray tube. Because
                            the image of the individual inverse focal spot was incorporated into the gain calibration
                            of the detector, its opposite is also seen (as the lucent area around the opacity), with
                            the differences in magnification and offset resulting from differences in the detector
                            position between gain calibration and subsequent image acquisition. (b) Radiograph
                            shows that as lead shavings build up near the head of the x-ray tube, a larger number
                            of inverse focal spots and their opposites are depicted.
                      Figure 9.  Lead debris was created near the head of the x-ray tube in a portable x-ray unit
                      when fully opened collimator blades rubbed against the lead beam-shaping port cylinder (ar-
                      row). This issue was discovered during the investigation of the cause of the image artifacts
                      shown in Figure 8.
RG  •  Volume 38  Number 3	                                                    Walz-Flannigan et al  841
need to change batteries during an examination,         tion. It is important that the image be repro-
when a technologist may not feel that there is          cessed after the correct shutter is applied because
time to wait for the offset calibration, it is advis-   the values of interest for processing have an effect
able to have a procedure for either scheduled           on the final image appearance.
detector battery changes or monitoring battery
life. In addition, it was helpful for technologists     Poor Identification of Values of Interest
to understand the effect of choosing the urgent
mode and to avoid doing so, if possible.                Description of Artifact Appearance.—The im-
                                                        age in Figure 14a appears washed out, giving the
Artifacts Related to Image Processing                   impression that the anatomy was poorly penetrated
A predominant variation between different               despite the use of a higher (large-patient) technique
digital radiographic (or computed radiographic)         for a small- to medium-sized patient. Another ver-
systems relates to their image processing algo-         sion of the same image, reprocessed after shutter-
rithms and the image processing settings that           ing to the area within the dashed lines, is shown in
are used. Failures in image processing or a sub-        Figure 14b. Figure 14b is more consistent with the
stantial lack of robustness in image processing         anticipated appearance of patella radiographs.
outcomes can create image artifacts that limit
or complicate viewers’ ability to see the imaged        Cause.—Processing outcomes are dependent on
anatomy in a consistent way.                            what values of interest are used by the image
                                                        processing algorithm. In this case, there ap-
Electronic Shutter Failure                              pears to have been a failure of proper exposure
                                                        recognition or perhaps an improper use of pixels
Description of Artifact Appearance.—A cross-            outside the anatomy in setting contrast enhance-
table lateral hip radiograph that was acquired          ment (9). The processing outcome was seen to
according to the standard procedure is shown in         vary depending on the amount of raw radiation
Figure 13a. Interpreting the image problem as an        area in the image, indicating that raw radiation
acquisition failure, the technologist repeated the      was being included in the values of interest used
acquisition, with the result obtained with proper       for image processing.
shuttering shown in Figure 13b.
                                                        Avoidance or Remedy.—Generally, image pro-
Cause.—The image processing algorithm incor-            cessing algorithms should not include large areas
rectly identified the collimator blades and incor-      of raw radiation in the values of interest used for
rectly applied electronic shutters.                     processing. If a correctly acquired image appears
                                                        to have inappropriate contrast and/or opacity, it
Avoidance or Remedy.—An imaging failure of              is wise to check that only appropriate pixels are
this type caused by electronic shutter failure can      contributing to the values of interest. Some ven-
be fixed through reshuttering and reprocessing          dor systems may display which pixels have been
the image, rather than requiring a new acquisi-         included as relevant, or the user may be able to
844  May-June 2018	                                                                                       radiographics.rsna.org
                 Figure 15.  Midgray clipping artifacts. (a) Knee radiograph shows the homogeneous gray appearance of the
                 cement (arrow), which indicates a loss of detail. (b) Knee radiograph obtained with a larger field of view shows
                 the area of the inset (rectangle). Inset: Magnified view shows the area of the tibia in which there is a loss of detail
                 (arrow). (c) Knee radiograph obtained with different processing from that used in a shows recovery of image
                 detail in the area of the cement and the cement-implant boundary. (d) Knee radiograph obtained with different
                 processing from that used in b shows recovery of image detail in the tibia.
constrain what is considered relevant by selecting                   appreciated and was found after more patient im-
a particular image region for processing.                            ages demonstrated the problem. The outcome of
                                                                     signal equalization also depends on the selection of
Midgray Clipping                                                     values of interest. For Figure 15a, it was seen that
                                                                     the poor processing outcome related to a change
Description of Artifact Appearance.—The im-                          in collimation between the image shown and the
age examples in Figure 15a and 15b show a loss                       images for which the processing was optimized
of information in parts of the knee radiographs                      (knee images with a larger field of view).
where multiple pixel values and structures are
anticipated but not visible. This loss is seen in the                Avoidance or Remedy.—Image processing set-
loss of detail in the cement and cement-implant                      tings were adjusted to avoid the loss of informa-
boundary in Figure 15a and a loss of detail along                    tion. This adjustment required adding another
the anterior edge of the tibia in Figure 15b.                        processing option for the images with a small
                                                                     field of view, which resulted in the image in Fig-
Cause.—The artifacts in Figure 15a and 15b                           ure 15c, and modifying the default processing
appear to be related to poor optimization of the                     setting for lateral knees, as shown in Figure 15d.
signal equalization processing, also known as
contrast enhancement (9–11). The image process-                      Metal Interface Artifacts
ing that had previously been set for the lateral                     For certain image processing algorithms, an at-
knee, shown in Figure 15b, was chosen for overall                    tempt to emphasize bone detail in the presence
contrast and detail preference with a clinical image                 of metal may lead to an artifact in the image of
sample. The loss of image information with this                      the metal or the metal-bone interface (10,12).
particular processing setting was not immediately                    Given the importance of being able to visualize
RG  •  Volume 38  Number 3	                                                                          Walz-Flannigan et al  845
Figure 16.  Metal interface artifacts. (a) Knee radiograph shows artifacts at the metal-bone interface (rectangle). Inset:
Magnified view shows artifacts at the metal-bone interface, creating a false lucency (arrow) as well as a stairstepped
edge that does not correspond to the structure of the object being imaged. (b) Knee radiograph was obtained with the
same source image data but with the use of an alternative processing setting to address the lucency artifact (rectangle).
Inset: Magnified view shows elimination of the lucency artifact depicted in a.
Figure 17.  Grid-line suppression failure. Chest radiograph obtained with a portable radio-
graphic unit shows a failure of the grid-line suppression software (square). Inset: Magnified
image shows detail of grid lines.
this boundary in arthroplasty imaging, it is an                        lucency; however, the adjustments did not satis-
important artifact to address.                                         factorily resolve the false structure on the edge of
                                                                       the hardware.
Description of Artifact Appearance.—In the
image in Figure 16a, the processing creates a                          Grid-Line Suppression Failure
false appearance of lucency between the bone
and metal, a finding that might be suggestive of a                     Description of Artifact Appearance.—Grid lines
loosening (Fig 16a, inset). In addition, the edge                      appeared on the chest radiograph obtained with a
of the implant appears falsely serrated.                               portable radiographic unit shown in Figure 17.
Cause.—Lucency artifacts (Fig 16a, arrow) are                          Cause.—The example in Figure 17 shows a failure
presumed to be related to excessive edge en-                           of a grid-line suppression algorithm related to using
hancement. The processing setting had worked                           a grid with a line frequency that was incompatible
well for knee images with a larger field of view                       with the grid-line suppression software algorithm.
but did not work well with the small field of view.
                                                                       Avoidance or Remedy.—Proper functioning of
Avoidance or Remedy.—In Figure 16b, the image                          the grid-line suppression software may depend
processing was adjusted to eliminate the false                         on automatic recognition that a grid is present.
846  May-June 2018	                                                                                     radiographics.rsna.org
                                                                                     TM
This journal-based SA-CME activity has been approved for AMA PRA Category 1 Credit . See www.rsna.org/education/search/RG.