Data Sheet M&EPCM500EN

New: PCM500 Series


Fluid Cleanliness Monitor
The PCM500 Fluid Cleanliness Monitor is a
portable diagnostic monitoring device that provides
a measurement of system fluid cleanliness.

As the successor to the popular PCM400, the new


improved PCM500 uses proven mesh blockage
technology to report accurate, reliable, 3 part ISO 4406
cleanliness codes for most types of fluids, in many
types of environment.
With the PCM500 you can:
t Monitor contamination levels in mineral, synthetic,
or water based fluids. Results are unaffected by the
presence of water, air, or dark fluids
t Get accurate, 3 part* ISO 4406 cleanliness code
results in under 6 minutes and to quickly take
preventative action
t Upload real-time results directly to mobile devices for
PCM500
analysis and action Fluid Cleanliness
Monitor
t ’Pass off’ cleanliness of new builds quickly and
confidently
Operation
t Protect your systems from catastrophic failure by
The color LCD touch screen allows simple menu driven
detecting abnormal fluid cleanliness conditions quickly
input of sample identification, monitor configuration and
The PCM500 can be permanently installed to monitor data output.
critical applications (including component test facilities) The HD screen displays real time data and test results
or used as a portable device for routine condition which are automatically stored for subsequent trending
monitoring of various fluid systems and evaluation. An optional bluetooth connected printer
allows the operator to produce a hard copy of the test
PCM500 Monitor Features results if required.
t Proven mesh blockage technology provides accurate All ancillary components for high and low pressure
3-part* ISO 4406, AS 4059 Table 1 (NAS 1638) or on-line sample monitoring are contained
AS 4059 Table 2 cleanliness codes within the unit, with sufficient
t Self cleaning procedure between each test ensures internal power to complete
optimum accuracy of results up to 35 tests between
charges. (AC power can
t Compact, robust, fully self contained portable design be used if preferred).
(fluid sampling pump included)
For further protection and
t Simple to use, color touch screen interface ease of transport, the
t Long battery life for extended operation in remote PCM500 is supplied in a
locations robust flight case.
t Measurement of fluid cleanliness, temperature,
viscosity, and optional water content.
*3 part code measured at 4 µm, 6 µm and 14 µm (c) per
ISO 16889.
Specifications Ordering information
Power supply: 90-260 VAC or integral Please select from the following part number options only.
12 VDC Lithium Ion battery
Battery life: Typically 35 samples Without Water Sensor With Water Sensor

Temperature 10 °C to 80 °C (50 °F to 176 °F) PCM500 M A PCM500W M A


Range: (dependent on fluid type) PCM500 M B PCM500W M B
Compatibility: Water glycols, aqueous solutions. PCM500 M D PCM500W M D
Petroleum and synthetic oils (hydraulic
lubricating, dielectric, etc.), fuels, industrial PCM500 U B PCM500W U C
phosphate esters. PCM500 U C PCM500W U D
1 2 1 2
Seals: Fluorocarbon
Operating 1.5 to 450 cSt (30 to 2,200 SUS)
Viscosity: References 1 2 refer to tables below.
Pressure: 0 to 315 bar (4570 psi) max
Table 1: Fitting Type
Monitoring range: ISO 4406: <11/9/7 to 23/21/17
SAE AS 4059 Table 1 Code Description
Class 1 to 12 (derived from NAS 1638) M ¼” BSPP Female Swivel fitting to metric test point
SAE AS 4059 Table 2 U ¼” NPT fitting and end cap
>4 µm 1A to 12A, >6 μm 1B to
12B >14 μm 1C to 12C Table 2: Power Lead
Water in Oil % RH: ± 2% at 5 to 95% RH (PCM500W) Code Description
Accuracy: ± 1/2 ISO 4406 Code A UK Power Lead
Communication 3 x USB’s (Data Acquisition, PC B European Power Lead
Ports: Setup, Printer), Ethernet and C USA Power Lead
RS232C (PLC Control)
D Australian Power Lead
Enclosure: IP 65 (NEMA 4)
Weight: 11 kg (24 lb) Printer Kit and Accessories
Dimensions: 400 x 260 x 250 mm PCM500-PRT
(15.8 x 10.2 x 10 inches)

Real time data is displayed Multiple test data can be stored


during test to show progress and displayed for subsequent
analysis and download

Pall Machinery and Equipment Visit us on the Web at www.pall.com


Pall Corporation has offices and plants throughout the world. For Pall representatives
25 Harbor Park Drive in your area, please go to www.pall.com/contact
Port Washington, NY 11050
+1 516 484 3600 telephone Because of technological developments related to the products, systems, and/or
+1 800 289 7255 toll free US services described herein, the data and procedures are subject to change without
notice. Please consult your Pall representative or visit www.pall.com to verify that
this information remains valid.
Portsmouth - UK
+44 (0)23 9233 8000 telephone
+44 (0)23 9233 8811 fax
industrialeu@pall.com © Copyright 2015, Pall Corporation. Pall and are trademarks of Pall Corporation.
® Indicates a trademark registered in the USA. Better Lives. Better Planet and
Filtration. Separation. Solution.SM are service marks of Pall Corporation.

M&EPCM500EN Printed in the UK September 2015