icountLCM20
Features & Benefits
Test time:	               2 minutes                   	 Data entry allows individual equipment test log
                                                        details to be recorded.
Particle counts:	         MTD 4+, 6+, 14+, 21+,
                          38+ and 70+ microns(c)      	 Data retrieval of test results from memory via hand
                          ACFTD 2+, 5+, 15+, 25+,       set display.
                          50+ and 100+ microns        	 Automatic test cycle logging of up to 300 tests can
International codes: 	    ISO 7-22, NAS 0-12            be selected via hand set display.
                                                      	 Totally portable, can be used as easily in the field
Data retrieval:	          Memory access gives
                                                        as in the laboratory.
                          test search facility
                                                      	 Automatic calibration reminder.
Max. working pressure:	   6000 psi (420 bar)
                                                      	 Instant, accurate results achieved with a 2 minute
Max. flow rate:	          106 GPM when used             test cycle.
                          with system 20 Sensors.     	 Data entry allows individual equipment footprint
                          Higher with single point      record.
                          sampler
                                                      	 Data graphing selectable via the integral printer.
Working conditions: 	     LCM will operate with the   	 Auto 300-test cycle logging via LCD handset input.
                          system working normally
                                                      	 RS232 to USB computer interface.
Computer compatibility: 	 Interface via RS232         	 Limit level output to control peripheral equipment
                          connection @ 9600 baud        such as off-line filtration via internal relay limit
                          rate.                         switches.
 	 Special ‘diagnostics’ are incorporated into the    	 Auto-testing allows for the conducting of
   icountLCM microprocessor control to ensure           automatic sequencing tests on flushing systems
   effective testing.                                   for example.
 	 Routine contamination monitoring of oil systems    	 Optional bar code swipe wand to allow handset
   with icountLCM saves time and saves money.           data loading.
 	 Contamination monitoring is now possible during    	 Worldwide service and technical support.
   application operation - icountLCM saves on         	 Re-calibration - Annual certification by an
   production downtime.                                 approved Parker Service Center.
Typical Applications
 	 Construction machinery
 	 Industrial plant
 	 Hydraulic equipment & system
   manufacturers
 	 Research & testing institutes
 	 Offshore & power generation
 	Marine
 	 Military equipment applications
                                                      2
icountLCM20
Specifications
Automatic Particle Counters (APC), have been
widely used for many years in condition monitoring
of hydraulic fluids. However, it is only recently that
APCs have become flexible enough to enable the
instruments to be taken out of the laboratory and
used on-line in order to obtain the most credible form
of results.
Unusually, the move from fixed laboratory use, to
portable field use has not been at the expense of
accuracy or user flexibility, but has actually enabled
the instruments to be used over a wider range of
applications and situations.
The most common monitoring technique used in
APCs is that of light obscuration or light blockage.
Here, a focused light source is projected through
a moving column of oil, (in which the contaminants
being measured are contained), causing an image of               Laser Optical Sensing
the contaminant to be projected on to a photo diode
cell, (changing light intensity to an electrical output).        icountLCM Proven Core technology
The electrical output of the photo diode cell will vary
in accordance with the size of the particles contained           The icountLCM portable particle counter features
in the column of oil; the larger the particle, the bigger        microprocessor controlled optical scanning for
the change in the photo diode electrical output.                 accurate contaminant measurement with a calibration
                                                                 range from ISO 7 to ISO 22 with no counter saturation.
On-line APCs must be able to test the oil sample at
whatever cleanliness it is delivered to the machine.
                                                                 How does icountLCM work?
Parker therefore had to develop technology to ensure
the on-line APC was able to test a sample without                – 	 The particles are measured by a photo diode that
the conventional laboratory technique which requires                 converts light intensity to a voltage output which is
dilution - a practice that would have been simply                    recorded against time.
impossible with a portable unit.                                 – 	 As the particle moves across the window the
By careful design and window sizing, 40,000                          amount of light lost is proportional to the size of
particles per ml can be achieved without making the                  the particle. This reduction in voltage is measured
instrument susceptible to counter saturation.                        and recorded.
                                                                 – 	 This “voltage” lost relates directly to the area
                                                                     of the particle measured, is changed into a
                                                                     “positive” voltage and then in turn changed into a
                                                                     capacitance value.
                                                                 – 	 This value is counted and stored in the icountLCM
                                                                     computer in one of 6 channels according to
                                                                     particle size.
                                                                 – 	 Readouts are displayed on the hand-held LCD in
                                                                     the accepted ISO and NAS standards ready for
                                                                     hard copy printing or RS232 computer download.
                                        A focused light source   – 	 The on-board computer allows storage of up to
                                        is projected through a
                                        moving column of oil.
                                                                     300 test results.
                                                                 3
icountLCM20
Portable Particle Counter
Specification
                                                                   LaserCM       LaserCM
                         Description                             (LCM20 20 22) (LCM20 20 62)
ABS structural foam and injection moulded case
                                                                                                                     4.96 (126mm)
ABS handheld display
Mechanical composition – Brass,
plated steel, stainless steel and aluminium
Fluorocarbon seals
Perfluoroelastomer seals
                                                                                                                 1.10 (28mm)
Nylon hoses (kevlar braided microbore)
Stainless steel armoured hose ends
1.2m fluid connection hose
Rechargeable battery pack
                                                                                                   1.26 (32mm)
                                                                                                                 9.80 (249mm)
12Vdc power supply
Fast blow fuse
Unique optical scanning system
Bonded glass optical window enclosed in SS plate
Micron channels analysis (Six)
Analysis range ISO 7 to 22 incl. (NAS 0 to 12)
32 character dot matrix LCD. Alpha numeric keypad
Data retrieval
Calibration to ISO standards*
Viscosity range 2 to 100 cSt. 500 cSt.with SPS
Operating temp. 41°F to 176°F (+5°C to +80°C)
Ambient temp. 41°F to 104°F (+5°C to +40°C)
2 minute test completion time
Memory store – 300 test memory
Battery operated 6 x 1.5 D cells                                                                                                                                                    dimensions in inch (mm)
Phosphate Ester group compatibility
Mineral oil & petroleum based fluid compatibility                                              Accessory Kit for icountLCM Classic
Up to 420 bar (6000 psi)
                                                                                                                                    Limit socket, fuse                      ParSmart &
Integral 16 column printer                                                                      Power supply                           & jack plug       Screwdriver        cable assy
RS232 to USB computer interface
Astra board case weight – (lb.)                                       11            11
Unit weight – (lb.)                                                  17.6          17.6
ParSmart software and cable link pack
Weather protector cover
CE certified
Auto logging
*Note: In compliance with international standards, all Parker portable particle counters can
meet the ISO Medium test dust standards. The icountLCM’s, in addition to the complete
range of Condition Monitoring products, are capable of achieving certification to ISO
4406:1999 and with traceability to ISO 11171 for SRM 2806, via ISO 11943.
                                                                                                                                       Instruction         Power lead,         Batteries &
                                                                                               Re-chargeable                         manual & bar         printer ribbon,      printer reel
                                                                                                battery pack                         code software       bar code pen &
                                                                                                                                                         weather cover
                                                                                               4
icountLCM20
Portable Particle Counter
   Why On-Site Fluid Contamination Monitoring?
    	 Certification of fluid cleanliness levels.
    	 Early warning instrument to help prevent
      catastrophic failure in critical systems.
    	 Immediate results with laboratory accuracy.
    	 To comply with customer cleanliness requirements
      and specifications.
    	 New equipment warranty compliance.
    	 New oil cleanliness testing.
   Data Download Management
   Dedicated software, provides the link between an
   icountLCM20 and your computer management system.
                                                          16-column printer for hard copy data. A feature of
                                                          the icountLCM is the on-board printout data graphing
                                                          option developed to support predictive maintenance
                                                          procedures.
                                                                                               icountLCM Test
                                                                                                ON LINE TEST
                                                                     icountLCM Test
                                                                                              TEST NUMBER 022
                                                                     ON LINE TEST
                                                                                                         D M Y
                                                                   TEST NUMBER 022         Date	       04-03-10
                                                                                           Time	15-52
                                                                       D M Y               NAS CLASS:	        7
                                                                Date	04-03-10
                                                                Time	15-52
                                                                ISO:	20/15/09                   Count / 100ml
                                                                      Count / 100ml        4/6µ (c)	       789157
                                                                                           6/14µ (c)	       31250
                                                                >4µ (c)	         820721    NAS CLASS	           7
                                                                >6µ (c)	          31564    14/21µ (c)	        250
                                                                >14µ (c)	           314    NAS CLASS	           3
                                                                >21µ (c)		           64    21/38µ (c)	         50
                                                                >38µ (c)		           14    NAS CLASS	           3
                                                                >70µ (c)		            0    38/70µ (c)	         14
                                                                                           NAS CLASS	           4
                                                                NOTES                      >70µ (c)		           0
                                                                                           NAS CLASS	           0
                                                                                           NOTES
                                                                      ISO 4406 - 1999     Correlation to NAS 1638
                                                      5
icountLCM20
Portable Particle Counter
 Introducing the new icountLCM ‘Classic’
 There is a new addition to the proven range – the                                                                              Classic icountLCM20
 icountLCM ‘Classic’. Only available from Parker, the
 ‘Classic’ retains all the technology that made the
 icountLCM one of the most accurate, reliable and
 popular portable particle counters available.
 Our design engineers have re-configured the
 icountLCM specification in a way that has reduced
 our manufacturing costs. These savings have been
 passed onto icountLCM ‘Classic’ customers.
                                                                                                                                                                                  Calibration
                                                                                                                                                                                  certificate
 How have we done this?
 Parker listened to our existing customers and then to the
 engineers and maintenance operatives to find out the
                                                                                                                  Empty battery pack
 features that make the icountLCM a unique predictive
 maintenance instrument.
                                                                                                                                       CD user guide
 Then, we removed peripheral items such as the                                                                                                                        Carton box supplied
                                                                                                                                                                          to house kit
 aluminium case and all the accessories, so a
 customer receives the icountLCM, with a CD user
 guide, professionally and securely boxed. One thing
 that has not altered is the icountLCM accuracy
 and icountLCM reliability. Our in-house software
 engineers have re-configured the EPROM, removing
 Data programming, User ID, Automatic Testing, Data
 retrieval, Alarm level settings, the barcode pen and
 Graph printing functions to reduce costs still further
 without in any way reducing the efficiency of the
 icountLCM. The icountLCM ‘Classic’ remains an
 instrument to be proud of.
 Ordering Information (icountLCM and ‘Classic’ icountLCM)
                       Model                                             Fluid Type                                                    Options
                    LCM2020                               2                 Hydraulic mineral                    1             icountLCM20 (ACFTD calibrated)
                                                           6                      Skydrol                        2               icountLCM20 (MTD calibrated)
                                                                                                                 3         icountLCM20 ‘classic’ (ACFTD calibrated)
                                                                                                                 4         icount LCM20 ‘ classic’ (MTD calibrated)
 Note 1: Part numbers featured with bold highlighted codes will ensure a ‘standard’ product selection.
 Note 2: Alternate displayed part number selection will require you to contact Parker Filtration for availability.
                  Part Number                                         Supersedes                                               Description
                  ACC6NE015                                               B84702                                               Printer roll x 5
                  ACC6NE014                                              P.843702                                               Printer ribbon
                    ACC6NE013                                             B84609                                        Re-chargeable battery pack
                    ACC6ND002                                            P849603                                          Weather protector cover
                    ACC6ND000                                             B84703                                       USB to RS232 Download Cable
                  LCM20.ACCKIT                                                                               Accessory Kit for LCM Classic (see p. 4 for content)
 Note 1: Part numbers featured with bold highlighted codes will ensure a ‘standard’ product selection.
 Note 2: Alternate displayed part number selection will require you to contact Parker Filtration for availability.