Icountlasercm20: Portable Particle Monitor
Icountlasercm20: Portable Particle Monitor
Principle of operation
Specification
A portable particle monitor designed to be used in the field Test time 2 minutes; repeat test time every 2 minutes
Particle counts MTD >4, >6, >14, >21, >38 and >70ei µm
ACFTD >2, >5, >10ii, >15, >25, >50 and >100ei µm
icountLCM20 is a proven answer to fluid system contamination monitoring Analysis rangeiii ISO 4406: 7 to 22 inclusive; NAS 1638: 0 to 12 inclusive; AS4059E(cum): 000 to 12 inclusive; AS4059E(diff): 00 to 12
inclusive; GOST 17216: 00–17 inclusive;
offering a 2-minute test procedure. Multi-standard ISO, NAS, AS4059E and
RepeatabilityIV ≤7% of measured counts for MTD Particle Sizes: 4, 6, 14µm
GOST 17216 cleanliness reporting, data entry, data graphing and integral Coincidence Error LimitV 23,000 particles / mL
printing are all standard on this world proven contamination monitor. Data entry 32-character two-line dot-matrix LCD. Full alphanumeric entry facility on keypad
Calibration The LCM and calibration master sample the same particle distribution suspension.
The LCM is calibrated to the master to meet specification at the measured points.
MTD – instrument calibrated using MTD reference material.
ACFTD – instrument calibrated using ACFTD reference material.
Consult Parker for recalibration.
Certification This product complies with all relevant EU declarations of conformity
Working Viscosity 2–100 centistokes (500 cSt with Single Point Sampler)
Product Features: CMP Max. Viscosity Range 200 Centistokes (100 cSt when reporting in GOST standard and using Heated Bath)
Oil temp. range +5°C to +80°C
• icountLCM20 is a proven answer to Operating temp. range +5°C to +40°C
Max. working pressure 2.5 bar when using Case Mounted Pump; up to 420 bar when CMP not used
fluid system contamination monitoring. Max. flow rate 30ml/min when using Case Mounted Pump.
System 20 Sensors flow up to 400 l/min.
• 2-minute test procedure. Single Point Sampler can be used for high flow. (Consult Parker)
System connection Via System 20 In-line Sensors or the Single Point Sampler
Memory store 300 test (scrolling memory) capacity
• Multi-standard ISO, NAS, AS4059E and
Integral printer 16-column printer for hard copy data
GOST 17216 cleanliness reporting. Computer compatibility Interface via RS232 connection at 9600 baud rate
Power Replaceable battery pack: Requires 6 x 1.5V D cells.
• Data entry, data graphing and Rechargeable battery pack: Supply voltage of 12V DC 1A max, 9mm jack plug with positive centre contact.
Unit: Supply voltage 12V DC 3A max, 9mm jack plug with positive centre contact.
integral printer. Battery performance Replaceable battery pack: maximum 25 tests before replacement
Rechargeable battery pack: maximum 40 tests before recharging
• 420 bar rated maximum pressure. Fuse 1.25A fast blow fuse included for overload protection (spare supplied)
icountLCM20 cover Weatherproof cover
Construction Case: ABS
• Supported by the offline UBS Hand-held display: ABS
and online SPS accessories. Keypad: Fluorosilicone rubber
Carrying case: Astraboard
Mechanical components: Brass, plated steel, stainless steel and aluminium
Seals: FKM (Viton). Other materials available - consult Parker
Hoses: Nylon (Kevlar braided microbore). Stainless steel armoured ends
Hose length Fluid connection hose: 1.2 metres (1 metre extensions can be used)
Hand-held display cable length: 1 metre. CMP standard hose length: 0.26 metre
Portability icountLCM20, 8kg; LCM with CMP 8.15kg. Carrying case 5kg approx.
Commissioning kit 6-off D cell batteries; 1-off print rolls (shrink wrapped); 2-off printer ribbons;
‘ParSmart Downloader’ software plus cable; weather protector cover; 12V DC power supply; Rechargeable battery pack
i The number of particles >70 μm (MTD), >100 μm (ACFTD) and in the ranges 50–100 μm and 100–200 μm (GOST) are not measured by this device. However, estimated values (indicated by the letter e on the
display) have been calculated using the following formulae: MTD: >70e μm = >38 μm × 0.06816; ACFTD: >100e μm = >50 μm × 0.06816; GOST: 50–100e μm = >50 μm(ACFTD) × 0.93184; GOST: 100-200e μm =
>50 μm(ACFTD) × 0.06816
iii The instrument only uses the shorthand in these standards for reporting contamination levels.
Typical Applications
The electrical output of the photo diode cell will vary in
Parker LaserCM Portable accordance with the size of the particles contained in
Particle Monitor the column of oil; the larger the particle, the bigger the
change in the photo diode electrical output.
• Construction machinery With 35 years’ experience in manufacturing the
world’s best-selling portable contamination monitor – On-line PCM’s must be able to test the oil sample at
• Industrial plant the progression to the icountLaserCM with its whatever cleanliness it is delivered to the machine. Parker
opto-mechanical, continuous wave single point therefore had to develop technology to ensure the on-line
• Hydraulic equipment & system source laser (SPSL) is both a natural and a customer PCM was able to test a sample without the conventional
manufacturers driven development. laboratory technique which requires dilution - a practice
that would have been simply impossible with a portable unit.
• Research & testing institutes
By careful design and window sizing, an ISO code 22 at
• Offshore & power generation >4 micron per 100 ml, (equivalent to up to 2.3 million
particles >4 micron per 100 ml), can be achieved without
• Marine making the instrument susceptible to counter saturation
or coincidence error.
• Military equipment applications
These high saturation point on-line PCM’s, whilst losing
none of the accuracy of their laboratory counterparts,
enable particle counting to be carried out quickly and
accurately.
z
Specifications
How does icountLaserCM work?
32 LaserCM LaserCM
Description (LCM20 20 22) (LCM20 20 62)
The icountLaserCM portable particle counter features microprocessor 126 ABS structural foam and injection
controlled optical scanning for accurate contaminant measurement with a • •
moulded case
calibration range from ISO 7 to ISO 22 with no counter saturation. ABS handheld display • •
258 Mechanical composition – Brass,
• The particles are measured by a photo diode that converts light intensity to a
voltage output which is recorded against time. plated steel, stainless steel and
• •
28 aluminium
• As the particle moves across the window the amount of light lost is Fluorocarbon seals •
proportional to the size of the particle. This reduction in voltage is measured Perfluoroelastomer seals •
and recorded. Nylon hoses
249 • •
(kevlar braided microbore)
• The electrical output of the photo diode cell will vary in accordance with the Stainless steel armoured hose ends • •
size of the particles contained in the column of oil; the larger the particle,
the bigger the change in the photo diode electrical output. 1.2m fluid connection hose • •
Rechargeable battery pack • •
240
• This value is counted and stored in the icountLaserCM computer in one of 12Vdc power supply • •
5 measured channels according to particle size.
Fast blow fuse • •
• Readouts are displayed on the hand-held LCD in the accepted ISO and NAS Unique optical scanning system • •
standards ready for hard copy printing or RS232 computer download. Bonded glass optical window
• •
enclosed in SS plate
• The on-board computer allows storage of up to 300 test results. Micron channels analysis to 5
• •
measured channels and the sixth
channel is calculated.
Analysis range ISO 7 to 22 incl. (NAS
• •
0 to 12)
32 character dot matrix LCD. Alpha
• •
numeric keypad
Data retrieval • •
Calibration - see note below • •
Product configurator
Note 1: Part numbers featured with bold highlighted codes will ensure a ‘standard’ product selection.
Note 2: Alternate displayed part number selection will require you to contact Parker Filtration for availability.
ISO 4406 reporting Correlation to NAS 1638
Note 3: Option 7 and 8 with CMP (Case mounted pump).
Parker Worldwide
Europe, Middle East, Africa IE – Ireland, Dublin North America
Tel: +353 (0)1 466 6370
AE – United Arab Emirates, parker.ireland@parker.com CA – Canada, Milton, Ontario
Dubai Tel: +1 905 693 3000
Tel: +971 4 8127100 IL – Israel
Tel: +39 02 45 19 21 US – USA, Cleveland
parker.me@parker.com
parker.israel@parker.com Tel: +1 216 896 3000
AT – Austria, St. Florian
Tel: +43 (0)7224 66201 IT – Italy, Corsico (MI)
Tel: +39 02 45 19 21
Asia Pacific
parker.austria@parker.com
parker.italy@parker.com AU – Australia, Castle Hill
AZ – Azerbaijan, Baku Tel: +61 (0)2-9634 7777
Tel: +994 50 2233 458 KZ – Kazakhstan, Almaty
parker.azerbaijan@parker.com Tel: +7 7273 561 000 CN – China, Shanghai
parker.easteurope@parker.com Tel: +86 21 2899 5000
BE/NL/LU – Benelux,
Hendrik Ido Ambacht NO – Norway, Asker HK – Hong Kong
Tel: +31 (0)541 585 000 Tel: +47 66 75 34 00 Tel: +852 2428 8008
parker.nl@parker.com parker.norway@parker.com
IN – India, Mumbai
BG – Bulgaria, Sofia PL – Poland, Warsaw Tel: +91 22 6513 7081-85
Tel: +359 2 980 1344 Tel: +48 (0)22 573 24 00
JP – Japan, Tokyo
parker.bulgaria@parker.com parker.poland@parker.com
Tel: +81 (0)3 6408 3901
BY – Belarus, Minsk PT – Portugal
KR – South Korea, Seoul
Tel: +48 (0)22 573 24 00 Tel: +351 22 999 7360
Tel: +82 2 559 0400
parker.poland@parker.com parker.portugal@parker.com
MY – Malaysia, Shah Alam
CH – Switzerland, Etoy RO – Romania, Bucharest
Tel: +60 3 7849 0800
Tel: +41 (0)21 821 87 00 Tel: +40 21 252 1382
parker.switzerland@parker.com parker.romania@parker.com NZ – New Zealand, Mt Wellington
Tel: +64 9 574 1744
CZ – Czech Republic, Klecany RU – Russia, Moscow
Tel: +420 284 083 111 Tel: +7 495 645-2156 SG – Singapore
parker.czechrepublic@parker.com parker.russia@parker.com Tel: +65 6887 6300
DE – Germany, Kaarst SE – Sweden, Borås TH – Thailand, Bangkok
Tel: +49 (0)2131 4016 0 Tel: +46 (0)8 59 79 50 00 Tel: +662 186 7000
parker.germany@parker.com parker.sweden@parker.com TW – Taiwan, Taipei
DK – Denmark, Ballerup SK – Slovakia, Banská Bystrica Tel: +886 2 2298 8987
Tel: +45 43 56 04 00 Tel: +421 484 162 252
parker.denmark@parker.com parker.slovakia@parker.com South America
ES – Spain, Madrid SL – Slovenia, Novo Mesto AR – Argentina, Buenos Aires
Tel: +34 902 330 001 Tel: +386 7 337 6650 Tel: +54 3327 44 4129
parker.spain@parker.com parker.slovenia@parker.com
BR – Brazil, Sao Jose dos Campos
FI – Finland, Vantaa TR – Turkey, Istanbul Tel: +55 800 727 5374
Tel: +358 (0)20 753 2500 Tel: +90 216 4997081
parker.finland@parker.com parker.turkey@parker.com CL – Chile, Santiago
Tel: +56 2 623 1216
FR – France, Contamine s/Arve UA – Ukraine, Kiev
Tel: +33 (0)4 50 25 80 25 Tel: +48 (0)22 573 24 00 MX – Mexico, Toluca
parker.france@parker.com parker.poland@parker.com Tel: +52 72 2275 4200