CS-05
Calibration switch shall be put on ON place when calibrating;
calibration switch is located at lower left corner of indicator front panel, spin the lead
sealing screw at anticlockwise direction to open.
ON: facing to display panel, turn the DIP switch to right, means calibration.
1.Turn the calibration switch to the ON position, press the Fn and CAL buttons at the same time,
the meter displays CAL x, see the figure below
Press the Enter button to confirm
Display the digital input as shown below,Press the Enter button.
Display the digital input according to the figure below, the division value is 10kg, and the meter
increases or decreases according to the weight of each 10kg,Press the Enter button.
Input range (the figure below assumes 70t), press OK, Input the range according to the actual
situation on site,Press the Enter button.
Calibration zero (tare),Press the Enter button.
Wait for the countdown 10S to display the following image
Load weights to the weighing platform, press the Enter button to confirm after the weighing
platform is stable,(Alternatives can be used in place of weights)
After waiting for 10s, the following picture will be displayed. Assuming a weight of 700kg, enter
a weight of 700kg. Press the Enter button.
After the first calibration is completed, the following figure is displayed. Press the zero button to
exit. Calibration ends.
Error indicator symbol
Item Display Possible reason Solution
1 E--1 Calibration code error
Wrong operation as:
Forbidden clear zero or
2 E--2 dynamic tare function or tare
function forbidden status to
press TARE button.
Under matching 3 status,
3 E--3 forbidden running tare
function.
Profibus interface Change Profibus interface
4 E--4
initialization error module
- 28 -
Check load cell cable, this
Signal wire cable wrong
5 E--5 symbol display only under
connect
calibration process.
Check load cell selection
Load cell sensitivity is not whether reasonable and
6 E--6
enough the
shield wire connect or not.
This symbol is only
appeared only under
7 E--7 Data input error. calibration, input value
mare than 0 and less than
rating capacity.
Exceed zero clearance
8 - n o-
range.
9 over Over load
Modulus conversion parts
10 ADC-Err Change Mainboard.
malfunction