DS2 Operators Manual
DS2 Operators Manual
Operator’s Manual
                IMPORTANT
  Please read carefully before using the DS2
Revision History
Manual Version                                                   Revision Date:
Revision D                                                         June 2008
This publication is for the operators of the DYNEX DS2 Automated Microplate
Processing System using DS Matrix software.
Due to continuing software development, dialog boxes displayed in this manual
may differ from those actually seen in the software screens. Every effort has
been made to ensure the information in this manual is accurate, updated and
consistent with the product it describes. DYNEX reserves the right to make
technical improvements to the DS2 and documentation without prior notice as
part of a continuous program of product development.
The information in this manual is of a proprietary nature. No part of this
publication may be reproduced, stored in a retrieval system, translated,
transcribed, or transmitted in any form or by any means without prior written
permission of DYNEX.
DYNEX does not assume liability for the use of this manual. If the DS2 is not
used according to the description in this manual DYNEX does not assume
responsibility for any consequent effects.
Table of Contents
    LIMITED WARRANTY........................................................................................................168
    DYNEX TECHNOLOGIES CONTACT INFORMATION ..........................................................169
Appendix A – Arm Calibration Tool........................................................................ 170
Appendix B – Labware Specifications.................................................................... 180
Appendix C – Curve Fit Equations......................................................................... 182
    LINEAR REGRESSION ........................................................................................................182
    POLYGON FIT ...................................................................................................................184
    QUADRATIC, CUBIC AND QUARTIC REGRESSION ..............................................................184
    CUBIC SPLINE CURVE FITTING .........................................................................................186
    AKIMA FIT ........................................................................................................................186
    EXTRAPOLATION OF NON-LINEAR CURVES ......................................................................187
    SIGMOID FIT .....................................................................................................................188
    LOG – LOGIT CURVE FIT ..................................................................................................189
    USING LOGARITHMIC AXES FITTING ................................................................................189
Appendix D – Equation Functions.......................................................................... 191
Chapter 1 – Overview
Introduction
The DS2 Automated ELISA System is a computer-controlled microplate
processing system that fully automates the following steps of microplate
ELISA assays:
   •   Sample distribution
   •   Reagent addition
   •   Incubation
   •   Plate washing
   •   Signal Detection
The DS2 is intended for use in clinical, research, and industrial laboratories.
DS2 Specifications
 DS2 Instrument Dimensions
 Width                          <540 mm (21.3 inches)
 Depth                          <680 mm (26.8)
 Height                         <660 mm (26.0)
 Weight                         <48 kg (106.0)
 Power Supply
 Voltage                        100-240 V auto-switching
 Frequency                      50/60 Hz
 Power Consumption              <300 VA
Reader Specifications
Washer Specifications
Wash Buffers 2 x 2L
 Input connector for external   Clean fluid bottle must be provided (any size)
 clean fluid bottle provided
Incubator Specifications
Instrument Capacity
Number of Plates 2
Number of Assays/Run 12
 Number of                  24 x 2 mL tubes
 Standards/Controls
The system cover encloses the workspace, the arm and the pipette module.
The cover should be closed during operation to prevent accidents.
To open the cover, lift the cover handle until the cover is in the upright
position. To close the cover, pull down on the cover handle until the cover is
fully closed. The system cover will rest on the side panels when it is
completely closed.
                                                        Cover
                                                        Handle
System Cover
       Cover Resting
          Point
Indicator Light
Power Switch
Workspace Components
                                                              Incubation
                                                              Chamber
                                                                    Reading
                                                                    Optics
                                                                           Pipetting/
                                                                           Washing Area
Wash Head
Barcode                                                              Incubator
Scanner                                                              Door Lever
Sliding Cover
The incubation chamber can hold up to two microplates during a run. During
an incubation step, the reader sliding cover and incubator door are closed to
ensure proper temperature equilibration.
Barcode Scanner
The barcode scanner reads barcode labels on the sample tubes as well as
the positioning barcode labels located on the sample tube racks.
The barcode scanner is an IEC60825-1+A2:2001 Class 1 Laser Product that
complies with 21 CFR1040.10 and 1040.11 except for deviations pursuant
to Laser Notice 50 dated 7-26-01.
                 CAUTION: The barcode scanner has a maximum radiated
                 power output of 1.0 milliwatt. Do not stare into the beam of
                 the barcode scanner without appropriate protective
                 equipment (e.g. protective glasses). Obey the warning label
                 (shown below) that is on the front of the barcode scanner.
Pipette Module
The Pipette Module travels in the x-, y-, and z- directions to pipette samples,
controls and standards, dispense reagents, and to perform dilutions. The
Pipette Module is also used to slide the cover to the Incubation Chamber
open and close, as well as to pick up and eject the plate Wash Head
Assembly.
The pipette module has the following sub-components and functions:
Component Function(s)
 Wash Head              Verifies that wash head assembly has been picked up
 Detection              or ejected.
The pipetting system of the DS2 Automated ELISA System includes ESP™
(Electronic Signature Pipetting) software. ESP™ software automatically
detects gross pipetting inaccuracies caused by blockages or bubbles.
Wash Head Assembly
The DS2 uses a modular Wash Head Assembly stored in the front left
corner of the workspace. The wash head assembly is picked up by the
pipetting arm to perform wash operations. The Wash Head Assembly is
designed to wash 8 well strips of a full or partial 8 x 12 microplate.
Pipetting Arm
                                             Wash Head
                                             Assembly
Wash Head
The Wash Head portion of the Wash Head Assembly contains two sets of
pins. The shorter pins (the dispense pins) dispense fluid and the longer pins
(the aspirate pins) aspirate fluid. The aspirate pins and the dispense pins
are closely spaced so that fluid can be aspirated from and dispensed into
wells of a microplate at the same time.
During operation, the wash head assembly is automatically lowered to insert
the pins into the microplate wells or raised to remove the pins from the
wells. Lowering the wash head assembly allows the aspiration of the well
contents, or performing a wash cycle at the well bottom. Raising the wash
head assembly allows the movement of the wash head to another column
for filling or washing.
Aspirate Pin
Dispense Pin
Aspirate Line
Dispense
Line
                Figure 5: Wash Head with Wash Pins
Different wash buffers may be placed in the two wash buffer containers (A
and B) located at the front left of the instrument.
The maximum capacity of the wash buffer containers is two liters. The wash
buffer container must have a minimum of 500mL of fluid for a wash
operation to take place.
    Quick
 Disconnect
   Fitting
Float Switch
  Connector
Wash Buffer
Container A
Distilled or deionized water may be used to clean the wash head after plate
washing is complete. The user must program this step into the assay using
the Assay Editor. A user supplied bottle (Container C) is attached to the rear
of the DS2 using the supplied tubing. The supplied tubing is attached to the
Quick Disconnect valve located at the right rear of the DS2.
                                                                      Wash Head
                                                                      Cleaning Fluid
                                                                      Container C
          Quick
     Disconnect
                  Container C Connection
                                 Tubing
Waste Containers
   Fluid that is used during wash operations is collected in the Liquid Waste
   Container. Used sample and reagent pipette tips are ejected into the Tip
   Waste Container. Both waste containers are located at the front of the
   instrument.
                                                                             Float Switch
                                                                             Connection
                                                                             (level sensor)
Tip Waste
Container
                                                                  Quick Disconnect
                                                                  Fittings
                                                   Liquid Waste
            Waste Cap                              Container
   The Liquid Waste Container holds up to two liters of waste. A waste fluid
   level sensor alerts the operator when the Liquid Waste Container is full and
   should be emptied.
Absorbance Module
The DS2 Reader is able to take readings in two different wavelength modes:
  • Single - using one analytical test wavelength
  • Dual - using one reference wavelength and one analytical test
     wavelength
The single wavelength mode is sufficient for most applications.
Blanking
The Reader allows subtraction of a reference value from well ODs. Air is
automatically used as a reference, but the absorbance of a reagent solution
can also be subtracted from the test result. Blank wells may be single wells
or an average of multiple wells.
Software Description
DS Matrix software is used to control the DS2 Automated ELISA System.
The software automates the sample distribution, incubation, reagent
addition, and washing and detection steps of assays as defined by the user.
The DS Matrix software also provides the user interface for configuration of
the DS2 and management of consumables.
The Matrix software includes assay definition options that allow the user to
customize assay parameters, assay steps, OD reading settings, result
calculations, quality control checks, and report formats.
Additional information about the software can be found in the DS2 Online
Help, which is accessible from the Help menu.
Software Functions
DS2 Configuration
The following DS2 System parameters can be configured from the Tools
menu: database maintenance, pipetting options, OD read limits, system
configuration, Levey Jennings assay parameters, absorbance filters, and
user notifications.
Consumables Definitions
The DS Matrix software stores parameters for all consumables and fluids
used on the DS2 in a database. A consumable or fluid must be defined in
the database before it can be selected for use in an assay.
The parameters for fluids and consumables that are stored in the database
are summarized below:
Consumable Parameters
 Neat Fluids           Fluid name and type, pipetting profile, load settings
                       (location, bottle type).
 (Sample and
 Reagent Fluids)
 Wash Fluids           Fluid name and load position
Assay Programming
The Assay Editor screen allows the user to program the sequence of
pipetting, incubation, dispensing, washing, and reading operations
performed on samples. Data reduction and reporting are also programmed
using the Assay Editor screen.
Worklist Creation
Worklist Execution
The Optical Density results for each sample ID in the worklist are analyzed
according to the criteria specified in the assay.
Results Reporting
Chapter 2 – Safety
Warning Labels
The DS2 Automated ELISA System or its components may contain labels
that warn the user of a hazard or an electrical connection. The description
of the labels is described below. Potential personal injury to the operator or
damage the DS2 can result if the labels are not followed.
Label Description
Caution, biohazard
Safety Precautions
The following information aids in the safe and efficient use of the DS2.
In addition to the warning labels and other cautions previously described in
this manual, consider the following:
         1. Appropriate precautions must be taken when working with
            biohazards. Technicians must be trained in the safe handling
            and clean up of potential blood borne pathogens. Universal
            precautions, appropriate hygiene, and decontamination of
            surfaces are recommended. Consult the reagent kit
            manufacturer for precautions on handling potentially hazardous
            substances.
         2. Appropriate personal safety precautions must be made when
            opening and closing the DS2 cover. A gas spring holds the
            tension to keep the cover open. The DS2 cover should be able
            to be opened 8 inches (approximately 203mm) without falling. If
            the cover drops instantly above this height then the gas spring
            should be replaced. The cover may creep down slowly from this
            point due to the nature of gas springs and this is acceptable.
            Gas springs will inevitably lose pressure depending on
            frequency of use, so it is important for the user to take note if
            the cover begins to give way over time and notify the service
            provider of this circumstance.
         3. When the cover is up and the run has started, do not encroach
            upon the work area unless prompted by the software for user
            input of materials or manual intervention.
         4. Attend to error messages when the system prompts and stops.
            These messages indicate a need for user action.
         5. Sample tubes must be pushed down in the sample racks to
            prevent the pipette module from being obstructed.
         6. Place the sample racks securely onto the DS2. Push the rack
            firmly towards the back of the DS2 the rack clicks in place.
         7. Periodically inspect the sample rack springs to ensure proper
            tube alignment. Replace the springs as necessary.
         8. Barcode quality is critical to successful sample tube barcode
            scanning. Scanning is in accordance with ASTM E1466-92
            defining barcode quality, position, and orientation of barcode
            labels. The use of non-standard barcodes or barcodes with
            poor print quality may be problematic. Barcode labels should
            be applied using a vertical orientation. The barcode label
            should be oriented so that it faces out of the opening in the
            sample rack.
Chapter 3 – Installation
Unpacking
Packing List for DS2 Installation
     2. Place the DS2 near the bench or table where it will be located.
     3. Examine the packaging to verify that all of the materials listed in the
        packing list have been removed. Store the packaging material for
        future use.
     4. Inspect the DS2 and its components for damage. If damage is
        observed, contact your shipper or service representative.
The computer system that is used for operation of the DS2 Automated
ELISA System must meet the following minimum specifications:
     1. Intel Core/Core2/Pentium 4/Celeron family or compatible
        processor recommended
     2. 2 GHz or higher processor clock speed recommended; 1.8 GHz
        minimum required
     3. 10 GB hard drive with at least 100 MB of free space
     4. Microsoft® Windows® XP Professional operating system with
        Service Pack 2 or later
     5. Microsoft® Windows® compatible display adapter (card or built
        into motherboard) with 32-bit color (Highest setting) at 1024x768
        or more resolution
     6. 1 gigabytes (GB) of random-access memory (RAM)
        recommended, 512 megabytes (MB) minimum required
     7. One unused USB port is required for connecting the computer to
        the DS2 Automated ELISA System
     8. Mouse or other pointing device supported by Windows®
     9. Microsoft® Windows® compatible CD-ROM or DVD Drive
       3. Plug the other end of the USB communication cable into the USB
          port on the right rear side of the instrument.
       4. Connect the keyboard, monitor, and printer cables to the
          computer.
       5. Connect the power cords to the computer, monitor, and printer.
                                               Cooling
                                               Fan
                                                     Power Cord
                                                     Connector
Initialization
The DS2 will perform a series of initializations before it is ready to start a
run, including picking up and dropping the wash head assembly, opening
and closing the incubator door, and moving the microplate carrier to and
from the incubation and reader chambers.
            Timeline
             Toolbar                                       Incubator
                                                          Temperature
The Timeline Toolbar contains icons for starting and stopping the run,
changing the display, and setting run options. Certain functions may not be
available depending upon the run status.
• Database Maintenance
• System Configuration
• User Notification
• Absorbance Filter
• Import/Export
• ESP Calibration
Database Maintenance
The DS Matrix software database includes a description of all consumables
(plates, racks, and tips) and fluids (reagent controls, standards, wash
buffers, stop solutions, substrates, etc.) used on the DS2. A consumable or
fluid must be defined in the database in order for it to be available for
selection from a drop-down list when creating or editing an assay.
               IMPORTANT:
                     It is the burden of the user to perform appropriate
                validation of any consumable parameters to ensure proper
                assay performance. Factory defined consumable settings
                       will not necessarily work with all consumables.
                                                        Edit
                                                        Consumables
        DELETE
                                         COPY          EXIT          HELP
          Delete        Add       Edit          Copy          Exit          Help
Adding a Consumable
  1. Click on the ADD button, enter the consumable name and click OK.
        a. Consumables from the same assay should have a similar
           beginning to the name so that they are grouped together in
           the drop down list. This will make fluids easier to locate when
           writing assays. For example:
                 1. Lyme Substrate
                 2. Lyme Positive Control
                 3. Lyme Cutoff Calibrator
        b. If fluids are used universally across many assays they may
           be given a generic name, such as the name of the kit
           manufacturer. For example:
                 1. Dynex Wash Buffer
                 2. Dynex Stop Solution
                 3. Dynex Substrate Solution
        c. Once the consumable name is entered, the data entry
           window opens (see Figure 14 for an example). Enter the
           required parameters..
  2. Click SAVE. The new consumable will be added to the database for
     selection from the drop down list.
  System Configuration
  System configuration parameters are universal for every assay that is run on
  the DS2. Therefore, changing any of the parameters on the System
  Configuration screen will update the parameters used for every assay.
      5. By enabling the clot detection check box, the system will alert the
         user if a clot (which prevents aspiration) is detected. Clot
         detection will not alert the user if subtle fluid aspiration changes
         occur.
      6. Enable the Save All ESP (Electronic Signature Pipetting)
         parameters option to create an ESP calibration profile for each
         fluid in the system. This option allows users with engineering
         capabilities to monitor fluid aspirations and dispenses for
         uniformity. Statistics generated by this feature can only be
         viewed in engineering mode.
      7. The ESP score is also used in engineering mode. The ESP
         score defines the passing calibration score for all fluids when
         monitoring the ESP calibration data. Ramp times, pipetting
         speeds and pressures are all taken into account when calculating
         the ESP score.
OD Limits
                                           Numerical
                                           Conversion
                                                 Enter Laboratory
                                                 Information
The DS2 Barcode Scanner is able to read 6 symbologies (See Figure 20).
     1. Enable the check box Use Check Digit if the last numerical digit is
        used to verify that the rest of the barcode is correct. If enabled, the
        Check Digit will not appear as part of the sample ID. If disabled the
        last number in the bar code will be part of the sample ID.
     2. Enable the check box Use Leading Zero if the first numeric digit in
        the bar code is a 0.
      2. Click the Enable LIS Support checkbox to activate the LIS option
         within Matrix.
          c. The user may specify the date and time when the specified
             backup frequency should begin.
          d. The Matrix software will display when the last back-up took
             place and when the next back-up will take place
       3. The user should empty the database backup folder, with the
          exception of one or two early clean databases and some of the
          more recent databases, on a monthly basis to save space.
                                                                               Sender
                                                                                E-mail
                                                                               Address
2. Click on the Setup tab. In the To textbox, type in the E-mail address for
   the recipient of the user notification E-mail sent from the DS2 Matrix
   software.
3. In the From textbox, type in the E-mail address of the sender (for
   example the e-mail address of the lab manager).
4. In the SMTP Server textbox, type in the SMTP server for the sender’s E-
   mail address.
5. Click on the Save button to save these parameters, or click the Send
   Test E-Mail button to test the settings.
6. Click on the Manual tab to send E-mails manually. The screen below will
   be displayed:
                                                                  E-mail
                                                                  Addresses
                                                                  E-mail
                                                                  Message
                                                                  Body
7. Fill in the recipient’s email address and the sender’s E-mail address,
   along with the E-mail text.
       8. Click the Send E-Mail button to send the e-mail.
Absorbance Filter
Four standard absorbance filters are included with the DS2. The
wavelengths of these filters are 405, 450, 490, and 620. The user may
order a custom filter set specific to their assay(s) by purchasing them from
the supplier. Up to 6 filters may be installed in the DS2 filter wheel.
Import/Export
Assay and OD data files may be imported or exported for use in other DS2
instruments or in other software applications.
                                                             Click to
                                                             Browse
Import Selection
Import
              Note: Plate Data files must have the *.XDB extension. Assay files
                             must have the *.XML extension
                Note: Plate Data files can only be imported that were created
                using DS Matrix and saved as *XDB files. Data cannot be
              imported into DS Matrix from other software applications such as
                                 spreadsheet (*.CSV) files.
and the user is prompted to select the most appropriate action in order to
correct the problem (see Figure 29).
                                                             Error
                                                             Recovery
                                                             Options
On the same screen (underneath the drop down list options), the user can
check the Auto Recovery checkbox to instruct Matrix to carry out the same
action the next time the exact same error is encountered. Matrix can then
carry out some corrective actions without any user interaction.
The Reset Auto Recovery (see Figure 30) function erases the settings for
the user-selected corrective action for a particular error. The next time the
error occurs Matrix will prompt the user to select a corrective action instead
of carrying out corrective actions without user intervention. To remove a
specified auto recovery action from Matrix, check the checkbox associated
with the undesired recovery action and then click Delete. Auto recovery will
no longer be performed the next time the error is encountered.
              Note: Use the Reset Auto Recovery function with great caution.
     
Event Log
The Event Log lists every function the DS2 performs during an assay. To
view the event log, click Tools > View Event Log, and the Event Log
displaying events which occurred since the current software session began
will be displayed. A new Event Log is created each time the software is
started.
   1. To view logs from previous software sessions performed on a prior
      date, click the View History button.
   2. Choose a date on the calendar, highlight the desired Event Log and
      click open. Each operation performed is recorded in the event log.
                     Levey-Jennings Criteria
Levy-Jennings charts monitor the performance of an assay over time. Levy-
Jennings charts include plots of the individual results versus assay data to
analyze the dispersion of results about the cumulative mean value.
Typically, a specific well or sample type is used, such as a control or other
sample type, for which an expected value is available.
Levey-Jennings charts on the DS2 can be displayed by Matrix software for
specified wells on an assay plate using raw OD values or calculated results
(concentrations or ratios). Westgard rules may also be applied.
 Time           Assay
 Range
5. Click Apply.
                                              Well               Calculation
                                              Selection          Method
                               Plates
ESP Calibration
Electronic Signature Pipetting (ESP™) is a sophisticated technique used by
the DS2 for detection of gross pipetting errors often caused by the presence
of particulate matter, clots, foam, or bubbles in the sample.
Correct use of ESP requires calibration of acceptable sample transfers for a
given fluid with a given pipetting profile and volume. This data is collected
and stored in a database and is used by Matrix as a reference for good data
to compare against subsequent dispenses. Insufficient sample transfer can
therefore be detected and reported to the operator during a run.
                                                Number of
           Fluid Name     Volume                Transfers
Run Button
                Message
                Center
                (operator
                prompts)
                                    Select Bottle
                                    Type
Click Accept
                       Bottle
                     Description
Operator Instructions
4. Click Accept
Number
Calibration Point
Enter Next
Decreasing
Weight
                                                 Programming
                                                    Toolbar
                              Report
                              Format
                              SetUp
Creating An Assay
The Assay Title Page
      1. Select File > Assay Editor from the menu bar
             •   If the user checks the check box for Request kit lot data at runtime, it is
                 not necessary to enter kit information on the Assay Title Page, however
                 the user will be required to enter lot and expiration information for all
                 consumables and fluids each time the assay is run. The user can choose
                 not to enter lot information when prompted, however the user will be
                 prompted to enter lot information for every fluid.
             •   If kit lot data is required to match at runtime, check the check box for
                 Require runtime kit lot to match and enter data for the kit lot in the Lot
             Data table. Upon running the assay the user will be to confirm matching
             kit lot and consumable data at runtime. If lot numbers and expiration
             dates have been entered previously for consumables, the user will be
             asked to confirm that these lot numbers have not changed and that the lot
             number stored in the system is a match for the lot number of consumable
             being used. If a lot number has not been entered for a particular
             consumable, the user may opt to enter the lot information or to click
             Cancel and no lot information will be entered.
            •   Note: If a curve fit is used in the data reduction, the default Standard well
                type must be used to define the assay standards on the template.
       4. Select the number of Replicates desired by clicking the up and down arrows.
       5. Determine if the replicates are to be oriented By Row or By Column by selecting
          the appropriate radio button in the Replicates area. When replicates are added
          by row, one replicate will be placed (for example) in column 1 (A1) and one
          replicate will be placed in column 2 (A2). Replicates which are added by column
          will be added to the template serially (A1, B1).
       6. The user can click on each well individually, or the user can click and drag the
          cursor over the desired wells to define them as the selected well type.
       7. Repeat steps 3 through 6 until the entire plate has the desired well definitions for
          standards, controls, empty, samples, etc.
       8. To clear an individual well, click the Empty radio button, and then click on the
          desired well location in the template. Click on the Clear All button to reset all well
          types.
       9. Choose the microplate type which will be used for the assay from the
          Microplates drop-down list. Only microplate types which are defined in the
          plates database will appear in the drop-down list.
      10. Click Validate when finished.
   Microplate                               Replicates
   Pipetting                                  Number
   Template                                                                       Well Type
                                                                                  Options
                                               Empty
                                               Single Well
                                                                 Toggle
                                                                 Blank
                          Changing a Well
                          Type Definition
                                                                          Clear All
                                                                          Well Types
                                                                          Button
                                                             Microplate
                                                             Choices
     Label             Definition
     S                 Standard
     C                 Control
     T                 Test Sample
     NC                Negative Control
     PC                Positive Control
     CO                Cut-off Control
     PR                Positive Reference
     SC                Serum Control
     AC                Antigen Control
     N                 Negative
                                               Function
                                               Buttons
Report Format
     1. Click on the Report Formats tab.
      2. Select Format Type by selecting the appropriate radio button (raw data, ratio,
         threshold, curve fit, or spreadsheet). Data will be reported according to one or
         more of these options as programmed in the assay. The user must define a
         format type for each of the different operation types that will appear in the assay
         report.
      3. Check the Create Matrix check box for results to be displayed in an 8 x 12 array
         format.
      4. Check the Create Table check box for data to be displayed in a tabular format.
         Use the check boxes in the Table Options area to indicate which data to include
         in the table. In the Results Order area of Table Options, select the order of
         results. Items can be added to the list via the following method:
          a. Highlight the desired well type, then click on the        button to add the
             parameter to the list.
          b. Highlight the desired well type from the list, and then click on the         button
             to remove it from the list.
          c. Click on the Add All         button to move all well types into the list.
          d. Click on the Remove All           button to remove all well types from the list.
          e. Click the up or down arrow next to the right column to move highlighted well
             types up or down in the list.
          f.   Click Validate when finished.
      5. In the Numeric Options area, check the check box to Average the replicates if
         needed. Enter the desired number of decimal places the numeric data should be
         rounded to. In the Mean area:
          a. Select the radio button for Arithmetic mean to perform calculations using
             addition (+), subtraction (-), multiplication (*), and division (/). The arithmetic
             mean of replicate determinations will be reported.
          b. Select the radio button for geometric mean to perform calculations using the
             antilog of the mean of the logarithms of replicates. Reporting the geometric
             mean is useful when there is imprecision built into the result data.
                                              Source of
                                              data being
                                              reported
Matrix or
Table
Option
Header Footer
        1. Click on the Header Footer tab.
        2. Select the desired Header and Footer options to be put at the top and /or
           bottom of each report page by checking the check boxes associated with each
           piece of information.
        3. Click Validate when finished.
                                                               Footer
                                                               Options
                                  Header
                                  Options
     Output to File
        1. Click on the Output to File tab.
        2. If the data report is to be used with software such as external spreadsheets,
           check the Enabled check box.
        3. Information can be divided for export using a tab, comma, space or semi-colon
           separator by choosing the radio button associated with the desired option in the
           Separator area.
        4. Select the data to be exported by checking the check box associated with the
           desired data in the Select Data to Export area.
5. Assign a file name and path where the output files will be exported by clicking
            on the       button in the Output File Name area. The selected data will
            automatically export as a CSV (text type) file every time the assay is run.
        6. Click Validate when finished.
         icon which is currently highlighted will have blue font. Text describing all other
         operations will have black font.
Programming Toolbar
                                                                  Text
                                                                  Detailing
                                                                  Assay
                                                                  Creation
Dispense Operation
      Defining Dispense Parameters
Dispense Icon
                                                                         Select
                                                                         Reagent
     Select
     Wells
                                                                                              Enter
                                                                                              Volume
• If a new fluid must to be added to the Fluids Database, click on the Fluids
          button. Click on the      button to add a new fluid to the database. (See
          Figure 49)
                                                                                  Fluids
                                                                                  Button
Pipetting Operation
About Diluting Samples
The DS2 Automated ELISA System can dilute samples, controls and standards. Dilutions
can be performed in a single stage in standard microplates or in two stages using deep-well
strips. The DS2 adds diluent to a plate or deep well before the sample is added. Clicking
once on a well or well type in the microplate pipetting template allows dilution in the deep-
well strips. The selected wells are outlined red. Clicking a second time on the template
allows dilution in the plate. The outlines around the selected wells will change from red to
green.
The dilution ratios allowed for various microplate types are described below:
 2. Click the well locations on the template which will contain a diluted sample, control or
    standard.
           a. If clicked once, a red circle will appear around the well, indicating that
              dilution will take place in a deep well.
           b. If clicked twice, then a green circle will appear around the well, indicating
              that dilution will take place directly in the well on the plate.
               •   Note: When a Test Well type is selected, the neat fluid will automatically
                   be designated as “sample” and cannot be changed by the user.
5. Choose the Diluent to be used from the drop-down menu.
6.. Select the dilution ratio or absolute volume.
Definition Tab
       Volume specified     The exact amount of volume of diluent will be pipetted into
                            deep-well strip or plate. Enter a neat fluid volume and a
                            diluent volume. If using a deep well for dilution, specify
                            the target volume. The target volume is the volume of
                            diluted specimen to be transferred to the microplate
                            following the dilution step.
8.    To perform a two stage dilution in the deep well strip, check the “Two Stage Dilution”
      checkbox. Enter in the dilution formula information on the initial dilution tab, then
      select the secondary dilution tab and enter in the dilution formula information for
      the secondary dilution tab.
9.    When performing the assay, if the user should be able to change the dilution
      volumes at runtime, enable the option to Allow sample dilutions to be changed at
      runtime
10.   When diluting specimens in the deep wells, the user may elect to have the
      specimens stay in the deep wells a specified length of time prior to being transferred
      to the plate by entering the desired time into the Let deep wells stand before
      pipetting field. Deep wells may incubate for up to 60 minutes prior to transfer.
      Alternatively, the user may elect to pipette diluted specimens immediately after
      dilution is completed by checking the pipette immediately after dilution check box.
      If neither option is selected, the software will perform all required dilutions prior to
      transferring diluted material to the plate.
11.   Click on the Optimizations tab.
   12. Select an option by checking the checkbox to the left of each option. When
   diluting in the microplate, the bottom three options are unavailable.
Options
          Option                                           Description
          Use new tip for dispense of sample               A new sample tip will be used to pipette
                                                           each well.
          Multishot for dispense sample                    The same sample tip will be used for
                                                           replicates. If fluid for more than one
                                                           dispense can be aspirated at one time,
                                                           the required number of dispenses will be
                                                           aspirated at once and pipetted into the
                                                           required wells.
          Share dilution wells for microplate replicates   Only one dilution will be made for an
          (volume will be scaled)                          individual sample. The dilution will be
                                                           used for all replicates on the plate
          Share dilution wells across multiple assays      Only one dilution will be made for an
          (volume will be scaled)                          individual sample. The dilution will be
                                                           used for replicates in all assays on the
                                                           current worklist that share a common
                                                           sample diluent.
          Force level detect before transfer (uses new     If selected, a liquid level sense is
          tip)                                             performed on each deep well prior to
                                                           pipetting the sample to the plate.
                                                           Choosing this option will increase the
                                                           pipetting time from the deep well strip to
                                                           the plate.
        7. By clicking on the View / Edit Pipette Order button, the user can program
           which wells will be pipetted to the plate first. For example, if the user
           intends to pipette a negative control in the beginning of the plate (NC1),
           then samples, then a negative control at the end of the plate (NC2), then
           the pipetting order must indicate this sequence. In addition, if the pipetting
           order is changed from the default order indicated by the software, the user
           must select Use new tip for dispense of sample when pipetting diluted
           controls or Use new tip when pipetting directly from the vial to the plate to
           ensure the pipetting order is followed.
        8. Max post pipette time is the acceptable amount of time that a plate can sit
           prior to the next scheduled event being performed once pipetting is
           completed. This feature is useful when pipetting plates in parallel, so that
           one plate is not inactive for extended periods of time while operations are
           being performed on the other plate.
The Pipette Operation pipettes a specified fluid into selected wells. To pipette a control,
standard, or specimen directly from the vial to the plate:
       1. Click and drag the Pipette icon onto the assay programming bar.
       2. Click on the desired (target) well.
       3. Select the desired Neat fluid from the drop down list.
  4. Select Multishot Allowed and/or Use New Tip. If Multishot is selected, fluid will
     be aspirated in one tip to dispense all replicates of an individual test sample or
     reagent (up to the volume capacity of the tip). If neither option is selected, if
     replicates are present on the microplate pipetting template, the same tip will be
     used to dispense all replicates, but multishot will not take place.
  5. Select ESP Plus Parameter (Electronic Signature tracking to ensure pipetting
     was done properly) if desired. In order for an ESP Plus Parameter to be
     available in the drop-down list, an ESP calibration must be performed for the
     sample type and the calibration values must be saved (Tools>ESP Calibration).
  6. See the “Setting Dilution Parameters” section of this manual for information on
     selections in the Plate Options area.
  7. Click Validate to save the information.
                                                      Fluid to
                                                      Pipette
              Target
               Well
                                                                       Use new         Multishot
                                                                       tip             allowed
Well Fill Verification is used to track accurate pipetting of fluids into specified wells based on
their OD readings.
                                                            Filter
                                                           Selection
                   Selected
                    Wells
Incubate/Shake Operation
The incubate/shake operation specifies a temperature and duration for plate incubation
during an assay, and also allows for shaking. For ambient (up to 25o C) incubations, the
plate will stay outside the incubation chamber. For higher temperatures the plate will move
into the Incubation chamber.
Incubation Icon
Incubate/Shake Settings
       1. Drag and drop the Incubation Operation Icon into the Programming Toolbar.
       2. Define the Duration by clicking on hours, minutes, or seconds, then clicking on
          the up and down arrows. Incubation durations can last up to 23 hours, 59
          minutes and 59 seconds.
       3. Set the Shortest and Longest duration allowed for incubation. The incubation
          duration may be modified from the specified duration to fall within the given
          shortest and longest limits by Matrix in order to best-fit assay operations.
   4. Set the Temperature. Matrix will monitor the temperature of the incubation
      chamber, showing the current temperature in the upper right corner of the
      Timeline.
   5. If the temperature selected is Ambient, the user may choose to time the
      incubation starting with the first transfer of specimen to the plate by checking the
      Start from first transfer to plate checkbox. This check box can only be selected
      when ambient temperature incubation is chosen and when there are no
      operations inserted in the program between the pipette operation and the
      incubation operation (for example, no dispense may be present between the
      pipette operation and the incubation operation).
   6. Select the radio button for the desired shaking option. The default for shaking
      while incubating is off or Do not shake. The user may choose to:
          a. Perform a short initial shake by selecting the Initial Shake radio button
             and specifying a shaking time.
          b. Shake the entire incubation time by selecting the Shake for entire
             duration radio button.
   7. Select the Speed at which the shaking will take place, (Low, Medium or High).
   8. Click Validate to save the settings.
Wash Operation
A wash operation can consist of Purge, Dispense, Fill, Aspirate, Clean and Soak settings.
Operation Description
 Clean          Dispenses fluid from the clean line from the dispense
                wash pins while the wash head is positioned over the
                purge tray. A Clean cycle is carried out at the end of
                the wash protocol to prevent buffers from crystallizing
                inside the tubing lines and causing blockages.
 Dispense       Aspirates the contents of the well and then dispenses
                a specified amount of fluid into the wells. If a bottom
                wash is specified, the wash head is then lowered to
                the bottom wash position. The fluid will be aspirated
                from the bottom of the wells as wash fluid is being
                dispensed.
 Fill           Allows the user to fill the wells with a specific volume.
 Aspirate       Removes the contents of a well by positioning the
                wash pins at the aspiration height in the well and
                aspirating the liquid from the wells.
 Sweep          The aspiration pins move from side to side inside the
                well during aspiration.
 Super          The aspiration pins move in a four-step X pattern
 Sweep          across the well during aspiration.
 Soak           The contents of the wells sit for the specified number
                of seconds after each dispense cycle. Once the soak
                cycle is complete, the fluid is aspirated.
                                                   Volume to Dispense
                                                   per cycle
                   Number of Cycles
Specify Fluid
Purge
Volume
Clean
Volume
           •   If washing Stripwise, all cycles are completed on one strip before moving to
               the next strip. For example: 3 cycles are programmed with a 5 second soak
               after each aspirate-dispense cycle. The DS2 will aspirate and dispense the
               first strip, wait 5 seconds, and then repeat this cycle 2 more times on the first
               strip prior to performing any cycles on the second strip.
      •   Sweep – Normal aspiration takes place, then the aspirate pins move once
          across the well in a right to left motion to remove wash buffer which may be
          pooled at the well edges.
      •   Sweep on last cycle only – Sweep takes place only on the last programmed
          cycle.
      •   Super sweep – Normal aspiration takes place, then the aspirate pins move
          across the well in an x-patterned aspiration.
      •   Super sweep on last cycle only – Super sweep takes place only on the last
          programmed cycle.
7. Set the Purge volume and select the desired wash buffer.
      •   Purging fills the tubing lines with the specified wash buffer prior to beginning
          the wash cycle.
      •   Purging prevents the first few strips from getting air or bubbles dispensed into
          them.
    8. Set the Clean volume and select the desired clean fluid.
      •   Note: A fluid must be added to the Washerfluid database with the load
          position specified as the “clean line” in order for options to appear in the drop
          down list.
      •   Cleaning uses fluid from the external bottle attached to the rear of the DS2
          (bottle C) to clean the tubing lines following the last wash cycle.
       10. The Synchronize Soak Delay from Transfer Timings check box can be
           enabled to minimize OD drift across the plate due to differing incubation times
           between the strips.
         •   The DS2 will synchronize column soak time based on the time taken to
             transfer samples to the microplate. For example, if pipetting takes 15
             minutes, and plate washing takes 3 minutes, the first samples will be
             incubated for the correct amount of time, while the later samples would
             incubate a shorter amount of time. This timing problem can be resolved by
             synchronization.
         •   There must be a Pipette step in the assay preceding the Wash operation in
             order for the synchronization checkbox to be enabled.
         •   There cannot be any operations in between the wash and the pipette
             operation in order for the synchronization option to be available. The only
             exception to this rule is the incubation operation. The incubation operation
             may be inserted into the timeline in between the pipette and wash operation;
             however the Start from first transfer to plate option must be enabled.
• Choose the wash buffer from the washerfluid database drop-down list.
         •   To add a wash buffer to the washerfluid database, click on the Fluids button.
             Follow the prompts to name and define the fluid. The user may select either
             the A or B bottle location, but it is recommended not to specify a location if
             multiple wash solutions will be used on the instrument.
       12. Use Bottom Wash if fast washing is required.
         •   Bottom wash will squirt bursts of buffer into the well, while aspirating at the
             same time. Bottom wash can save time and buffer if used instead of an extra
             wash cycle. Bottom wash may be used with either plate or strip washing
             mode.
         •   Specify the number of Cycles and the Volume of wash buffer to use by using
             the up and down arrow buttons next to the associated parameter. 0 - 5
             bottom wash cycles may be programmed using 0-1000 µL.
       13. Check the Fill check box to dispense fluid. The dispense and bottom wash
           options must be set to zero for the Fill option to become available. Fill may be
           used in either strip or plate-wise mode.
         •   The Fill operation may be used if a long soak time is needed. Program the
             Fill in the Wash operation, then program an Incubation operation for the
             length of time needed, then program another Wash operation into the assay.
         •   Fill may also be used for well fill verification to test the washer dispense
             accuracy.
         14. Check Aspirate if wells only need to be emptied of fluid and do not need
             washing. The Aspirate checkbox does need not be checked if wells are being
             washed normally.
Read Operation
   The Read operation allows the user to choose the appropriate wavelength filters for use
   with the assay and instructs the DS2 to read the absorbance of each well containing
   fluid.
                                                       Wavelength
                                                                         Shake Time
Wavelengths: Wavelengths for OD readings are determined by the specific filters installed
in the reader. The correct wavelength values must be entered for the installed filters in the
Absorbance Filters menu for valid results.
If using single-mode, a primary test filter wavelength must be entered. If using dual mode,
both filter wavelengths must be entered.
Dual Wavelength Mode:
The dual wavelength mode can be used if necessary to reduce errors caused by dirt and
scratches on the bottom of the wells as these artifacts tend to scatter radiation equally,
regardless of wavelength.
The choice of test and reference wavelengths for the dual wavelength mode depends on the
particular enzyme/substrate reaction being tested. However, the following rules should
usually be followed:
       1. The test wavelength (λt) should be at or near the maximum absorbance of the
          reaction product.
       2. The reference wavelength (λr) should lie outside the absorbance band of the
          reaction, but not far removed.
The Reader subtracts the absorbance at the reference wavelength (λr) from the absorbance
at the test wavelength (λt) to minimize the effect of background noise.
             405 nm                  λt              λr       690 nm
                                      FILTER WAVELENGTH
Quality Control Operation equations are used to ensure that an assay has worked properly
and that the data generated can be accepted. The QC equations place criteria on the raw
OD data, or data generated by a curve fit, threshold, ratio or spreadsheet operation.
       1. Well ODs are accepted or rejected using criteria such as comparison against a
          control well or against a numerical value.
       2. QC equations may be applied to any Well Type that has been defined in the
          template. If the QC equation fails, the user may elect to suppress all data from
          being output to subsequent operations.
       3. Quality Control operations are context sensitive, as some QC equations may only
          be applicable to the results of a Curve Fit operation or to the results of a
          Threshold operation.
     Quality
     Control                                    Functions        Report
      Icon                                        List           Options
                                   Equation
                                     Bar
       1. Drag the Quality Control Operation icon             into the programming toolbar.
       2. Double click on the desired Function to add it to the equation. Functions are
          used to generate the equations. Continue to add functions until the equation is
          defined.
            a. Mathematical functions (such as +, -, =, >, <) may be typed into the equation
               box.
       3. Type in a Fail Message if desired. This message will be displayed in the report
          when the equation criteria have not passed.
       4. Click on the Add button to add the equation to the list of quality control
          acceptance criteria.
       5. Click on the Clear button to reset the Equation and Fail boxes so additional
          equations can be created.
       6. To change an equation already added to the list, highlight the equation to be
          edited, make the desired changes in the equation text box, and click Change.
       7.   To remove an equation from the list, highlight the desired equation and click
            Remove.
       8. The Quality Control equation checks will be performed in the order they appear in
          the list. Therefore if one equation depends on another equation taking place first,
          the equation which must be performed first must appear in the list before the
          dependent equation (for example: the “I” function equation must appear closer to
          the top of the list than the “valid” function equation). To change the order of the
          equations, highlight the desired equation and click Move Up or Move Dn.
       9. Options for QC Report handling include the default reporting of all QC results
          (Full QC Report). If the QC fails, all other results can be suppressed by checking
          the Suppress results if QC fails checkbox. This feature may be important if
          data is automatically being exported from Matrix into other files. QC results can
          also be related to an individual reagent lot by checking the Lot specific
          checking of control values checkbox (the Request or Require kit lot data
          checkbox must be enabled in the Assay Title Page screen for the lot information
          to be available for reporting).
      10. Select options for Pos/Neg, Curve Fit, and/or Auto Outlier Removal, if the options
          are available. A Curve Fit or Threshold operation must to be included in the
          programming toolbar before Quality Control Operation icon in order for these
          choices to be enabled.
See Appendix C for function descriptions and examples which can be entered into Quality
Control Equations.
  1. If any NC OD value is greater than 25% higher or lower than the mean OD value of all
     the NC’s, it must be rejected, enter:
                 0.75*NC<=NCi<=1.25*NC
  2. If the test must be rejected if one of the NC OD value fails the first Quality Control
     equation, enter:
                      valid(NC)>=3
  3. If the average OD value of the Negative Controls must be more than 0.5 OD, enter:
                       NC>0.5
  4. If the average OD value of the Positive Controls must be greater than 20% of the
     average OD value of the Negative Controls, then write:
                       PC>0.2*NC
  5. If the average Negative Control OD value must be less than half of the average PC
     OD values, enter:
                      NC<PC/2
  6. If the CV% of the Negative Control must be less than 15%, enter:
                       CV(NC)*100<15
Ratio Operation
The Ratio equation is used to convert raw data using mathematical equations, which often
expresses sample results as a ratio of a control value. Ratio calculations are frequently used
in competitive inhibition assays or when detecting an increase in some factor over a normal
baseline value. Every value on the plate is fed into the equation and the resulting values are
reported and passed on to subsequent operations. Ratio operations can also be used to
multiply wells on the plate by a correction factor, or to perform any calculation which is
required for further data processing.
EX: In the equation “Sample/PC”, all sample OD values will be converted to a ratio of the
average OD value of the Positive Controls.
Results from the ratio equation may be flagged with a user defined “result” or “flag” by
clicking on the Results Flagging tab.
EX: In the equation “Sample>NC then SICK” data calculated from the “Sample/PC” equation
will be flagged based on whether the results are higher or lower than the calculated negative
control value.
Ratio Icon
   2. To create a Ratio equation, double-click on the Functions which will make up the
      working components of the equation.
   3. Common arithmetic functions, such as addition (+), subtraction (-), multiplication (*),
      or division (/), etc. may be typed into the equation box.
   Condition Expression
                             Results Flagging Tab
Function List
5. Click on the Results Flagging tab in order to specify a phrase that either replaces a
   numerical result or is displayed next to a result. The phrase will be initiated whenever
   the numerical result meets the criteria programmed. Results Flagging is optional and
   should not be used if an LIS is being used.
See Appendix D for possible mathematic and function entries which can be used to write
Ratio equations.
   1. To convert all sample OD values to a ratio of the average OD of the Positive Control
      replicates, enter the equation: Sample/PC.
   2. To convert all values into larger numbers than the OD units, enter an equation such
      as: sample/0.25.
4. There are two methods to multiply all values on the plate by a correction factor:
          •   Sample*CustomVariable
          •   Sample*VariableX (where X is replaced by an integer)
       When these equations are entered, the user will be prompted at the beginning of the
       load wizard to enter the value for the variable.
        Note: Check the Ratio radio button when setting up the Report
        Format in an assay to have ratio information included on the report
                        in either a Matrix or Table format.
        Note: When creating ratio equations, use the Well Symbols from
                     the Template to create the equations.
Threshold Operation
The Threshold Operation converts raw OD data into symbols denoting positive and/or
negative results which fall above or below a user defined threshold. In order to use this
operation, Threshold limits must be defined.
A negative OD limit is set using the “–” data label and a positive limit set using the “+” data
label. These positive and negative limits are required to use the limits function. Three
additional positive results levels (++, +++, and ++++) are also available, but not required.
Threshold limits may be entered directly into the equation textbox as absolute numbers. A
result that falls above or below the value entered will be recorded as positive (+) or negative
(-) symbol instead of the actual value. The value entered for the threshold limit may be
based off of raw data, ratio, curve fit or spreadsheet output.
       If the – value is lower than or equal to the + value then results below the –value are
       negative and results above the + value are positive.
       If the – value is higher than the + value, then results below the + value are positive
       and results above the – value are negative.
Alternatively, conditions may be introduced into the limits equations. For example, test
sample OD readings may be compared to control OD readings or the average of a number
of samples to determine if they are a positive or negative result.
DS Matrix leaves a grey zone when there is a gap present between the values programmed
for the – and + equations.
The user can change threshold labels to any descriptive words or numbers. For example,
the default “-“ label may be changed to “Negative” by clicking on the text box and typing in
the desired text.
Value Limit
        1. Drag and drop the Threshold Operation icon                 into the programming
           toolbar.
        2. Enter absolute values (OD, concentration, etc.) for the positive and negative limit
           levels (if values are being used from another operation (such as a curve fit), that
           operation must appear in the programming toolbar before the Threshold
           Operation).
        3. Alternatively, the user may double-click on a function in the Functions list box to
           enter this function. Arithmetic operators may be entered.
                        EX: PC*2 for level +++
        4. In the Label column, change the default labels to words or other labels as
           desired.
        5. When all equations and labels have been entered, click Validate.
The DS Matrix program ensures each equation is valid and contains no errors. If any of the
equations are not valid, the DS Matrix program will display an error message. If this occurs,
edit the equation as required for correctness.
In the Curve Fit operation, DS Matrix software constructs a standard curve graph using OD
values from standards of known concentrations.
From the graph, the concentrations of test samples are determined by finding the
concentration that corresponds to a given OD reading. Concentration data is always plotted
on the X-axis, and OD data is always plotted on the Y-axis.
To use the Curve Fit operation, standards of increasing concentrations must be included in
the assay template.
Only one curve fit operation can be programmed into any given assay. To test data using
more than one curve fit, another assay must be created. When the Report is generated from
the first assay, data can be recalculated and another curve fit generated by clicking on the
New Assay button in the report viewer toolbar and generating a report using different a
different assay.
                                                       Specify
                                                       Standard
                                                       Concentration
                           Standard
                           Concentrations
7. Units can be entered as mg/ml, µg/ml, etc. by entering the desired units in the Units
   box.
                        Axes
                        Selection
Fit Selection
1. Click on the Fit Type tab in the Curve Fit operation.
2. Highlight the type of regression DS Matrix should use to generate the graph. See
   Appendix C for equations defining curve fits.
      Calculation Options
        1. Check the Average the standard replicates checkbox to have the mean of the
           standard replicates used to generate the curve, rather than individual replicate
           points. Averaging the replicates will make the graph easier to interpret.
        2. Data can be extrapolated beyond the lower and higher ends of the graph if the
           Extrapolate the data checkbox is checked. Extrapolation is usually valid
           scientifically if a linear regression curve fit is used, but not if a sigmoidal fit is
           used.
        3. There are 4 axes options for data plotting. To select an Axis Scaling option, click
           on its corresponding radio button.
Axes Selection
Axes Scaling                X-Axis                       Y-Axis
        4. Any scaling options that are not compatible with the chosen curve fit type will be
           disabled.
EX: If a Sigmoid fit type is highlighted, the axes selections will be disabled, but the Sigmoid
Parameters options will be active.
       5. All four of the Sigmoid, S-shaped curve Sigmoid Parameters are user-definable.
          To define a parameter, check the box for the desired parameter and type a
          number in the textbox. If no options are selected, Matrix will set the A parameter
          to the lowest standard concentration and the D parameter to the highest standard
          concentration. Matrix will then define the other standard points to obtain an S-
          shaped curve.
       6. The number entered for maximum Iterations is the maximum number of times
          Matrix software will attempt to construct a satisfactory plot before aborting the
          operation. The default value of 20 is suitable for most applications , however the
          minimum allowable number of iterations is 1 and the maximum number of
          iterations is 29999.
Graphing Options
Graph Title
Axis Labels
   1. Click on the Graph tab in the Curve Fit operation. By default, a graph will always be
      produced based on the regression chosen.
   2. Enter a title for the graph (if desired) by entering the title in the Title textbox.
   3. Titles can also be entered for the vertical and horizontal axes by entering the desired
      titles in the “Label:” text boxes.
   4. By default Matrix will use Auto Scaling to determine the best fit for the vertical and
      horizontal axes based on the range of data.
   5. If Manual scaling is chosen, the user can then define the minimum and maximum
      numbers to be used to define the axis when the curve fit is displayed graphically. In
      the Minimum box the lowest value for either the standard concentration or OD may
Results Flagging
   1. Click on the Results Flagging tab in the Curve Fit operation.
   2. Results Flagging equations are used to specify criteria for generating a text flag
      when the specified results are encountered.
         The Condition is the Results Flagging equation.
         The Result is the text to be reported instead of the numerical value.
         Equation
          Entry
                                                     Add, Move,
                                                    Delete Buttons
   whole spreadsheet. A custom functionality is also built into the spreadsheet operation (Boolean
   Function Assay). This functionality should only be used by those performing certain assays.
                                                Spreadsheet
                                                Operation
                                      Spreadsheet Icon
Equation Entry
          1. Click on the Spreadsheet Operation Icon and drag it into the programming
             toolbar.
          2. Highlight a cell in the spreadsheet and enter a calculation or formula for that
             individual well, for example A1 – H12. (This will take the incoming data
             reduction result from the well at H12 and subtract it from the incoming data
             reduction result from the well at A1.) The resulting value will be the output for
             that cell. This result will be passed to the next data reduction step.
          3. Use the buttons to facilitate spreadsheet equation entry:
             • Select the Cut button to remove a calculation from a cell, and then use
                 the Paste button to paste the calculation into the desired cell.
             • Select Copy to copy a formula, and Paste to move the formula from one
                 cell to another (the calculation cells will offset to reflect the selected cell).
             • Select Delete to remove a calculation from a cell and permanently
                 remove the calculation from the spreadsheet.
             • Clear All will remove calculations from all cells in the spreadsheet
                 operation.
          4. Alternatively, the user may select the Auto Difference Mode button and
             select a formula to be applied across the entire spreadsheet. Auto Difference
             Mode offers the user pre-programmed automatic subtraction equations (see
             table: Auto Difference Equations)
Figure 73. Auto Difference Mode Equation applied to spreadsheet (Row A-B)
Results Flagging
  4. Click on the Results Flagging tab in the Spreadsheet operation.
  5. Results Flagging equations are used to specify criteria for generating a text flag
     when the specified results are encountered.
     The Condition is the Results Flagging equation.
      The Result is the text to be reported instead of the numerical
     value.
     The Flag is the text to be reported along with a numerical value.
  6. An unlimited number of Results Flagging expressions can be defined for an assay.
      Assay Name
       (required)
                                                                                        Assay
                                                                                      Category
                                                                                      (optional)
         Modifying an Assay
             1. Select File > Assay Editor from the main Matrix menu bar. Select File >
                Open from the Assay Editor menu bar. The Open dialog box is displayed
                (See Figure 75).
             2. Select the assay to be modified from the drop-down list, and then click OK.
             3. Modify the assay operations and/or data reduction steps as desired.
             4. After modifying assay settings, save the changes to the assay using File >
                Save.
Creating a Worklist
A Worklist is created to define which assays will be run on which specific specimens. The
worklist specifies the samples that are to be run, their location in the sample rack, and the
assay(s) to be run on each sample.
         Note: A worklist can be created only after at least one assay has
          been created. The worklist can include assays run on one or two
                                different microplates.
          Up to 12 assays can be run on one plate if the assays have the
          same incubation, reading, washing, and shaking specifications.
            Only one worklist can be open at a time, although multiple
                    worklists may be entered on the Timeline.
                                                                      Sample Batch
                                                                      Name/Identifier
                                      Enter number of
                                      samples
           Use Pattern to
           specify root of
           sample IDs
                                         Assign
                                         Sample IDs
     Rack
    Location
                                            Assays
                                            Added
                                            Assay
                                           Selection
                                          Sample
                                         Checkbox
      Choose Assay
        1. After the sample ID assignment is complete, click the NEXT button.
        2. Choose the assay to be run, then either click on the Add Assay button or double
           click on the assay name. An orange strip of check boxes will appear next to the
           sample IDs.
        3. Add additional assays as needed.
        4. By default Matrix assumes that all samples are going to be tested using the
           assays chosen. By unchecking the checkbox next to an individual sample ID, the
           sample will not be tested using that assay.
        5. Once the Sample Batch has been assigned to an Assay, it is called the Worklist.
        6. To save a worklist (optional), go to File > Save Worklist after selecting the assay
           but before clicking the Done button.
                                                                             Edit Dilution
                                                                               Button
                                                                                Enter
                                                                               Dilution
                                                                               Change
       3. Click OK. The sample dilution button (Dil) for the sample ID will turn green.
       4. The Revert button erases any edits made and the test sample dilution will be
          performed according to the original dilution values.
       5. Once the assay is selected, dilutions are modified (if required) and the Done
          button is clicked, DS Matrix sets up a schedule of assay events.
        6. At the top of the Timeline are buttons which allow changes to the Timeline
           Schedule. Active buttons will flash, while inactive buttons will remain greyed out..
                           Pause       Play
                                                      Edit
               Zoom            Abort          Skip    Plate      Add             Start
                                                                Assay
       7. Click the Accept button if no editing or assay additions are needed. The Timeline
          will go from inactive to active status.
Start Line
                                                                             Assay
                                                                           Operations
                                   Load
                                Consumables
             1. For a report on the status of consumables during a run, click on the      button
                at the bottom right of the Timeline window.
                a. Consumables include all fluids (excluding samples), tips (sample and
                   reagent), plates, washer fluids, and deep-well strips.
             2. The Consumables Status Screen (See Figure 82) will display the name of the
                reagent, the reagent volumes, the number of tips left, and which assay is
                running.
Wells Tested
Remaining Tips
       3. To show the status of a different neat fluid (reagent), click the       buttons
          beside the fluid tube. To show the status of a wash buffer, click the Washer
          Fluids Tab directly under the fluid tube.
       4. The plate layout shows the wells being tested. At the top of the plate layout is the
          file name of the assay being run.
The Report
After a plate is finished testing, reports become available to view, save and print. The list of
recent reports can be seen on the main DS Matrix screen.
                            Report
                            Button
Plate ID
              •   Highlight the well to be removed by clicking on it, and then check the
                  checkbox associated with the selected well. Click OK and the data will be
                  recalculated. More than one well can be removed at the same time
General Considerations
Sample Types
      1. Biological samples such as serum, plasma, urine, or spinal fluid are typically
         analyzed. Adequately prepared, stool and tissue homogenates can be run on the
         DS2.
      2. Sample IDs may be manually entered by the operator or read from barcode labels
         on the sample tubes.
Code 39 Standard 2 of 5
Code 93 Interleaved 2 of 5
    Quick Disconnect
Wash Buffer Container
                                                                                              Quick Disconnect
                                                                                              Waste Container
     Liquid Level
           Sense
      Connection
     Wash Buffer
        container
   Wash Buffer
   Container B
          3. Replace the wash buffer container in its storage location. Reconnect the liquid
             level sense connection and the quick disconnect fitting.
          4. Repeat Steps 1 through 3 for the second wash buffer container if necessary.
                    Note: Do not remove the Liquid Waste Container cap when the
                 container is in the DS2. Always remove the container first, stand it
                  upright on the bench with the cap pointed up, then remove the cap.
       4. Rinse the waste container with DEIONIZED or distilled water. Discard the rinse
          water.
       6. Place the waste container on the system and reconnect the quick disconnect
          fittings and the liquid level sensor.
2. Select a previously used sample batch or select to create a new sample batch (<new
   sample batch>). Click the OK button. The Worklist Editor screen will be displayed.
3. When creating a new sample batch, select the samples that will be run by clicking on
   individual tubes or by specifying the number of samples in the text box. By clicking the
   last tube in a sample rack, the entire rack of tubes can be selected.
                                                          Next
                                                         Button
                                                                           Assay
           Rack                                     Active                Category
         Indicator                                  Assay
                                                                             Saved
                                                                             Assays
                                     Omit
                                    Sample
                                   Checkbox
                                                                        Done
                                                                        Button
              Click the
            Accept Button
             to Start the
                 Run
                                                            Assay Steps
                                                             Grayed Out
   3. After the timeline has been accepted, the software will prompt the user to load
      sample tube racks and consumables. (Sample tube racks may be loaded prior to
      clicking the Accept button.)
   4. The Load Tube Rack Screen will display the number of samples in the Worklist,
      along with a diagram of the DS2 with tube racks highlighted. This prompts the
      operator to load the appropriate sample racks.
5. Load the sample racks and click the Green Check button.
Loading Consumables
   1. After loading the samples, the load consumables wizard begins. The first screen
      of the load wizard is a table which lists the reagents, standards and controls
      required for the run. The software will prompt the operator to load them in the
      appropriate position with the appropriate volume. Where possible, reagents
      should be in the appropriate bottles prior to clicking the Accept button.
   2. After an item is loaded in the designated position with the designated volume and
      any required information is entered, click the   button to display the next
      prompt.
The Timeline View Screen displays the scheduled timing of individual assay functions.
These time estimates are continuously updated as more runtime information is available to
the DS2.
3. To view the entire Timeline (as displayed above), click on the icon.
       1. For a report on the status of consumables during a run, click on the       Status
          button.
       2. The Consumables Status screen will be displayed with the tube types, fluids, and
          plate types that are being used in the run. During the run, the consumables status
          is updated as fluids and tips are used.
      Tube Types                                                  Plate Types
                                            Neat Fluids
                                            Tab
      Daily Maintenance
        1. Empty and clean the tip waste container.
   Weekly Maintenance
        1. Empty the wash buffer containers and rinse them several times
           with deionized water.
   Six-month Maintenance
       1. Replace the dispense tubing.
       2. Replace the aspiration tubing.
   To Replace a Bulb
      1. Remove the filter / bulb access panel from the rear of the DS2 by
         unscrewing the two screws (one at the top and one at the bottom
         of the panel).
                                                                Filter
                                                                Access
                                                                Panel
Bulb
Filter Wheel
Tweezers
Bulb
   To Remove a Filter
      1. Remove the filter access panel from the rear of the DS2 by
         unscrewing the two screws (one at the top and one at the bottom
         of the filter access panel).
      2. Remove the filter access panel. The filters are mounted on the
         filter wheel as shown in Figure 103.
      3. Locate the filter that is to be removed. Note: absorbance filter
         405 must stay in slot one on the filter wheel or the DS2 will fail
         the self-test on start-up.
      4. Firmly grasp the exterior filter housing with a pair of needle nose
         pliers.
      5. Pull the filter out of the spring-loaded slot.
      6. Replace the filter access panel.
      7. Go to the Matrix main menu toolbar and select the Tools menu.
         Select Absorbance Filter and change the wavelength in the
         position the filter was removed from to 0.
      8. Click OK to save the changes.
                                                           Needle Nose
                                                               Pliers
 Screws
                                                          Grasp Filter
Filter Wheel
   To Install a Filter
      1. Remove the filter access panel from the rear of the DS2 by
         unscrewing the two screws (one at the top and one at the bottom
         of the panel). Remove the filter access panel. The filters are
         mounted on the filter wheel.
      2. Locate the filter position in which the filter will be installed.
      3. Firmly grasp the exterior filter housing with a pair of needle nose
         pliers.
      4. Push the filter into the spring-loaded slot.
Wash Buffer
   Bottle A
Wash Buffer
   Bottle B
Waste Container
         Tip Waste
                                                          Door
                                                             Pump/Valve
                                                             Assembly
                                       Wash Head
                                       Aspirate Tube
                                     Wash Head
                                     Manifold
   Wash Head
Dispense Tube
Thumb Nut
                                                                Vacuum
                                                                Switch
                                                      Waste
                                                      Fitting
                                    Wash Head
                                    Manifold
Aspirate Tube
                                                                          Vacuum
                                                                          Line
Dispense Tube
                 Pinch Valve
                    Pump Outlet
      Bottle A Tube
                            90° Fitting
                                 Grommet
                              Bottle C Tubing                                  Pinch Valves
          11. Insert the replacement tubing for the Pump Valve Assembly in
              the reverse order of removing the tubing (steps 10 – 4).
          12. Lift the Wash Head off its cradle and replace the twin bore tube.
          13. Connect the free end of the tube to the Wash Head Manifold.
              Ensure the Wash Head is oriented as shown below.
Wash Head
Cradle
                  16. Remove and replace the tubing on the two Wash Buffer
                      containers and the Waste container.
   Wash Buffer
Container Tubes
                                  Waste Container
                                          Tubes
                                      Figure 112: Container Tubing
                  17. Connect the two Wash Buffer Containers and the Waste
                      Container to the DS2. Place the Tip Waste Container in it’s
                      proper location.
Requesting Service
      If the DS2 needs service, contact the Vendor from whom the system
      was purchased or contracted for. The Vendor will troubleshoot,
      assess the situation, and work with Dynex to find a solution to the
      service issue.
Limited Warranty
Warranty and Special Provisions
The Arm Calibration Tool is only available if the Matrix plug-in software has
been installed. This Tool should only be used by a trained technician.
   1. Double Click the Arm Calibration Tool icon or select the program
      using Start>All Programs>Dynex Technologies>DS-Matrix>Arm
      Calibration Tool.
   2. A window with a picture of the DS2 (cover open) and several buttons
      appears. Click the Arm Calibration button. The Arm Calibration
      Wizard screen appears with a window instructing the user that a
      Calibration Tool, a Dynex Microtiter plate, and a black delrin stepped
      collar will be needed to complete this procedure.
   3. Insert all five Sample Racks and both Reagent Racks into the rack
      carriers.
   4. The brass Calibration Tool should not be on the spigot. Detach and
       remove the brass calibration tool if it is attached to the spigot. Click
       Next.
Consumables Rack Position
   5. Insert the spigot into the rear hole located on the top edge of the
      consumables rack between the sample (blue) tip and deep well strip
      areas. On the dialog, click “Get Values” for the frame marked Rear
      Consumables Hole. Values will appear in the X,Y,Z boxes (See
      Figure A-1).
     Sample
   Tip Boxes
                                                      Rear
                                                      Hole
                                                      Front
                         Reader                       Hole
                         Cover
                         Handle
   6. Move the spigot to the front hole of the tips rack top edge. Click Get
      Values in the Front Consumables Hole frame. Values will appear
      in the X, Y, Z boxes. Click Next.
   8. Close the reader cover with the spigot pin still inserted in the handle.
      Click Get Values. Click Next.
   9. Take the tip out of the reader cover hole and gently move the arm up
      and back to home position. Load a box of blue sample tips into the
      sample tip rack. Load clear reagent tips into the reagent tip racks.
   10. Place the spigot fully into the tip marked Datum in Rack 1 (rear
       sample tip rack) as shown in the Figure A-3. Ensure the spigot is
       securely inserted into the tip and click Get Values.
                                                           Y Right
                   Datum
Rack 1
X Bottom
Rack 2
   11. Repeat step 10 for the tip locations marked “Y Right” and “X Bottom”
       after reinserting the spigot into the appropriate tips as indicated by
       the black circles in Figure A-4.
   12. Repeat the entire procedure for Rack 2 (front sample tip rack). Click
       Next.
Reagent Tip Rack Positions
   13. Ensure clear reagent tips are loaded into both racks in the positions
       shown in the following illustration.
Y Right
Datum
X Bottom
                                                                            Rack 2
                                    Rack 1
   14. Insert the spigot fully into the tips located in the reagent tip locations
       indicated in Figure A-5 below. Click Get Values for the frames
       marked “Datum”, “Y Right” and “X Bottom” for both sample racks.
       Click Next.
   Bar Code Reader Position
   16. Move the Barcode Reader to the far left position until it clicks into it’s
       home position. Insert the Spigot fully into the Barcode Reader hole.
       Do not force the spigot or compress the spring. Click Get Values in
       the frame labeled Barcode Position. Click Next.
Barcode
 Hole
Wash Head
                                                        Waste Hole
          17. Mount a clear reagent tip on the end of the Spigot. Move the Spigot
              and tip to the Waste Hole. Center the tip over the hole. Push the
              spigot down until the pins are as far down as possible without
              engaging the spring/eject mechanism in the spigot.
          18. Click Get Values for the frame marked Waste Position. Remove the
              reagent tip. Click Next.
          19. Insert the spigot fully into the wash head assembly. Engage and
              latch the pins on the Spigot into the catches on the wash head
              assembly. Click Get Values in the Washer Pickup frame. Remove
              the Wash Head assembly from the Spigot. Click Next.
          20. A message dialog box will come up which invites the user to save or
              to cancel the calibrations done so far. Ensure that the calibration tool
              is not present on the spigot and click Next to save the values and
              initiate a self-test.
          22. Manually open the Reader Cover. Click Present Upper Plate
              Carrier. Insert a standard microplate into the plate carrier.
          23. Engage and latch the Wash Head Assembly with the spigot. Be
              careful not to compress the wash head.
          24. Center the long aspirate pins in the wells in Column 1. Click Get
              Values for the frame marked Upper Plate Datum.
          25. Center the aspirate pins in the wells in Column 12. Click Get Values
              for the frame marked Upper Plate Y Right.
   26. Move the Spigot with the Wash Head Assembly attached back to the
       home position. Click Present Lower Plate Carrier. Insert a standard
       microplate into the plate carrier. Repeat the procedure in steps 24
       and 25 to Get Values for Columns 1 and 12 for the lower plate
       carrier.
   27. Park the Wash Head Assembly in its carrier and remove it from the
       Spigot. Click Next.
       Calibration Tool
             Holding Screw
                                Flat Side
   28. Mount the brass Calibration Tool onto the Spigot. The flat sides of
       the calibration tool must be to the left and right on the Spigot. Use a
       hex wrench to tighten the Holding Screw to hold the Calibration Tool
       in place. Click Next.
   29. Click Present Upper Plate Carrier, and follow the prompts to clear
       the arm and insert a plate.
30. Insert the Calibration Tool into well A1, click Get Values.
   31. Insert the Calibration Tool into well A12, Get Values. Repeat using
       well H1.
   32. Move the Spigot and the calibration tool free of the plate carriers.
       Click Present Lower Plate Carrier, and follow the prompts to clear
       the arm and insert a plate.
   33. Repeat Get Values for wells A1, A12 and H1 for the lower plate
       carrier. Click Next.
 X Bottom
 rd
3 Position
             34. Insert the Calibration Tool fully into Sample Tube position 1 in Rack
                 1. There should be no sample tube in the rack and the Tool should
                 go all the way down to the bottom of the sample rack. Click Get
                 Values for the Datum position. Repeat for the Y and X positions. (Y-
                 position is the first sample tube location on Rack 5. X-position is the
                 last sample tube location on Rack 1.) Click Next.
                                                                         Y Right
                                                                       2nd Position
   Datum
 st
1 Position
  X Bottom                                                                                     Delrin
  rd
 3 Position                                                                                    Collar
             35. Put the black Delrin Collar firmly into position 1 on the large tube
                 Reagent Rack. There should not be a reagent tube in this location.
             36. Insert the Calibration Tool into the collar as far as it will go. Click Get
                 Values.
   37. Repeat the procedure in steps 35 and 36 with the first tube position
       on Rack 2 (Y-Right position). Click Get Values.
   38. Repeat the procedure in steps 35 and 36 with the last small reagent
       tube position on Rack 1 (X-Bottom position). Click Get Values. Click
       Next.
   39. There should be no tubes in the control racks. The Delrin Collar is no
       longer needed for calibration and should be stored.
                                                            Y Right
                                                         nd
                       Datum                            2    Position
                      st
                     1 Position
                X Bottom
                rd
               3 Position
Rack 1 Rack 2
   40. Insert the Calibration Tool into the bottom of Datum position on
       Control Rack 1, Click Get Values, and repeat the procedure for Y-
       right and X- bottom positions as shown in Figure A-9.
   41. Repeat the procedure for all three positions on Control Rack 2. Click
       Next.
Calibration Finished
   42. Remove the Calibration Tool from the Spigot by loosening the small
       holding screw using a Hex wrench.
   43. The calibration is complete. To save all settings, click OK, otherwise
       click cancel, which will revert the settings back to the original values.
   44. The DS2 will return the Spigot to home position, and move the plate
       carriers into position.
Arm Speed
To review or change the settings for the speed of the Spigot Arm movement,
a Dynex-issued password is required. Only factory trained service personnel
are authorized to perform this procedure.
Click Exit.
I. Specifications
   Type         Max       Height       Internal     External            Bottom
              Volume       mm          Diameter     Diameter            Shape
                 mL                      mm           mm
Controls &      2000       44.45         8.69         10.3                V
Standards
Vial
Deep Well      2000        39.8          7.4              8               V
Strip
Large          25000       89.2         22.45            25.4             V
Reagent
Tube
Small          15000       75.9          17.6            20.5             V
Reagent
Tube
Sarstedt       10000        92          14.41            16              Flat
Monovette
12x75          4000        74.4          9.7             11.9             U
Sample Test
Tube
DS-Matrix allows the user to select from nine mathematical relationships to determine the
concentration of the compound of interest. In this appendix, each of the curve fits is defined.
Linear Regression
A Linear Regression is used to put a "best fit" straight line through a set of data points.
Equation C-1 describes a linear relationship between the measured variable and the
concentration of the compound of interest:
                Y = mX + b
        C-1     where: m = the slope
                b = the y-intercept defining the line
                X = the value for the concentration
                Y = the value for measured variable (O.D.)
The linear regression formulae for determining the best straight line through a series of data
points are provided in Equations C-2 and C-3:
The linear correlation coefficient (R) is used to determine how well the line fits through the
points of the graph. Since the linear correlation coefficient can be either positive or negative,
the value of R2 is normally reported. When comparing two sets of data (fits), the fit with the
larger value for R2 is a better fit for the data set; if R2 =1, all data points lie along the line of
best fit and if R2=0, the data points are highly scattered and there is a poor correlation.
Once the linear relationship best fit has been established for a set of known standards, the
concentration of an unknown sample can be determined via equation C-1.
           A
           b
           s
           o
           r
           b
           a
           n
           c
           e
Concentration
Concentration
Polygon Fit
A Polygon Fit of the data involves joining successive data points using straight lines in a
connect-the-dots fashion as shown in Figure C-2.
Equation C-5 is a quadratic equation, Equation C-6 is a cubic equation, and Equation C-7
is a quartic equation.
The procedures used to fit these functions to experimental data are similar to that for linear
regression and have a similar mathematical derivation. A polynomial regression tends to
minimize the deviations of the data points from the polynomial equation. The squaring
(cubing, etc.) of the deviations tends to minimize them since positive and negative
deviations tend to cancel out.
The regression coefficient is handled in a similar manner as with a linear relationship. When
R2 = 1, all the data points have been fitted to the curve, when R2 = 0, the data points are
scattered.
Examples of Quadratic and Quartic fits are shown in Figure C-3 and C-4 respectively.
Absorbance
Concentration
C-8
C-9
C-10
The Cubic Spline Curve Fit is not a best fit algorithm as it assumes that each data point is
correct. If the data is erratic, if it is known that the data has significant errors or if the
concentration/absorbance has many inflection points, this method can lead to significant
errors in the reported concentration.
At high concentrations, the Cubic Spline Curve Fit may break up (Figure C-5). In some
cases, a semi-log or log-log axis (instead of a linear fit) will improve the quality of the data.
Absorbance
                                    Concentration
                           Figure C-5: Cubic Spline Curve Fit
Akima Fit
The Akima fit constructs a smooth curve through the data points. The fitted curve appears
smooth and approximates a manually drawn curve (Figure C-6).
Absorbance
Concentration
The Akima Curve Fit involves applying a set of polynomials to the data points, which
determines the slope of the curve at each point .The Akima Fit is based on the assumption
that each data point is determined by five points, the point of interest and two points on
either side of it.
For example, the readings for a set of Standards increase in a non-linear fashion with
increasing concentration for a given test (right side of curve in Figure C-7) and the user
determines that a quadratic curve fit might give the best fit for such data. However, at
concentrations lower than the concentration of the lowest Standard, the curve may turn
sharply upwards as shown in the left side of the curve.
Absorbance
Concentration
Sigmoid Fit
Many immunoassays are characterized by an S-shaped or sigmoid curve fit (Figure C-8).
Absorbance
                         Concentration
             Figure C-8: Sigmoid Curve
C-11
The above definition assumes that b has a positive value. If b is negative, the definition of
parameters a and d are reversed.
Matrix uses an algorithm for estimating the four parameters and determining the best fit for
the sigmoid curve. The curve requires at least four standards (data points), one for each of
the four parameters. It is recommended that at least eight Standards are defined on a plate
to ensure satisfactory statistical significance.
The Sigmoid Curve Fit algorithm is iterative, requiring many complex calculations. Matrix
may take some time to process the results. If the curve cannot be drawn or data cannot be
processed, a Windows® “Illegal Operation – Shut Down” error message will occur. If this
happens, examine your data carefully to determine if the number of Standards should be
altered or a different type of curve fit should be used.
Semi-Log Fit
When a semi-log fit is used, the X axis presents log (power of 10) concentration. To obtain
the concentration of a sample, equation C-12 is used.
Log-Log Fit
When a log-log fit is used, the X axis represents the log of the concentration (concentrations
in powers of 10) and the Y axis represents log of the OD (measured variable). To obtain the
concentration of a sample, equation C-13 is used.
                                  EX: Verify(T,10%,2>=3
                                  Means the number of verified
                                  Test samples must be greater
                                  than or equal to 3. A verified Test
                                  sample falls within 10% of the
                                  mean of all other Test samples.
                       Following
                       Threshold
  POS / NEG       Well Label            The user can require that a
                                        specific (default or user defined)
                                        threshold Well Type be labeled
                                        All samples within the Well Type
                                        must fall into the required result
                                        range or the Well Type will fail
                                        QC and a warning will be
                                        displayed in the results.