Proline Promass E 200 Flow System
Proline Promass E 200 Flow System
TI01009D/06/EN/15.12
71190059
Table of contents                                                                                                                                   Proline Promass E 200
Table of contents
2                                                                                                                                                            Endress+Hauser
Proline Promass E 200
                        Document information
Symbols used            Electrical symbols
Symbol Meaning
                                         Direct current
                              A0011197
                                         A terminal to which DC voltage is applied or through which direct current flows.
                                         Alternating current
                              A0011198
                                         A terminal to which alternating voltage is applied or through which alternating current flows.
                                         Ground connection
                           )  A0011200
                                         A grounded terminal which, as far as the operator is concerned, is grounded via a grounding system.
                                         Equipotential connection
                                         A connection that has to be connected to the plant grounding system: This may be a potential equalization
                              A0011201
                                         line or a star grounding system depending on national or company codes of practice.
Symbol Meaning
                                         Allowed
                                         Indicates procedures, processes or actions that are allowed.
                              A0011182
                                         Preferred
                                         Indicates procedures, processes or actions that are preferred.
                              A0011183
                                         Forbidden
                                         Indicates procedures, processes or actions that are forbidden.
                              A0011184
                                         Tip
                                         Indicates additional information.
                              A0011193
                                         Reference to documentation
                                         Refers to the corresponding device documentation.
                              A0011194
                                         Reference to page
                                         Refers to the corresponding page number.
                              A0011195
                                         Reference to graphic
                                         Refers to the corresponding graphic number and page number.
                              A0011196
Symbols in graphics
Symbol Meaning
A, B, C, ... Views
                                                    Flow direction
                                         A0013441
Endress+Hauser                                                                                                                                   3
                                                                                                               Proline Promass E 200
Symbol Meaning
                                                Hazardous area
                              -                 Indicates a hazardous area.
                                     A0011187
Fc = 2 × Dm (n × w)
Fc = Coriolis force
Dm = moving mass
w = Rotational velocity
                      The amplitude of the Coriolis force depends on the moving mass Dm, its velocity n in the system and thus on
                      the mass flow. Instead of a constant rotational velocity, the Promass sensor uses oscillation w.
                      In the sensor, two parallel measuring tubes containing flowing fluid oscillate in antiphase, acting like a tuning
                      fork. The Coriolis forces produced at the measuring tubes cause a phase shift in the tube oscillations (see
                      illustration):
                      • At zero flow (when the fluid is at a standstill) the two tubes oscillate in phase (1).
                      • Mass flow causes deceleration of the oscillation at the inlet of the tubes (2) and acceleration at the outlet (3).
                                                     1                                2                                  3
                                                                                                                                    A0016771
                      The phase difference (A-B) increases with increasing mass flow. Electrodynamic sensors register the tube
                      oscillations at the inlet and outlet. System balance is ensured by the antiphase oscillation of the two measuring
                      tubes. The measuring principle operates independently of temperature, pressure, viscosity, conductivity and
                      flow profile.
                      Density measurement
                      The measuring tube is continuously excited at its resonance frequency. A change in the mass and thus the
                      density of the oscillating system (comprising measuring tube and fluid) results in a corresponding, automatic
                      adjustment in the oscillation frequency. Resonance frequency is thus a function of fluid density. The
                      microprocessor utilizes this relationship to obtain a density signal.
                      Volume measurement
                      Together with the measured mass flow, this is used to calculate the volume flow.
                      Temperature measurement
                      The temperature of the measuring tube is determined in order to calculate the compensation factor due to
                      temperature effects. This signal corresponds to the process temperature and is also available as an output signal.
4                                                                                                                     Endress+Hauser
Proline Promass E 200
Transmitter
Sensor
                        Input
Measured variable       Direct measured variables
                        • Mass flow
                        • Density
                        • Temperature
25 1 0 to 18 000 0 to 660
40 1½ 0 to 45 000 0 to 1 650
50 2 0 to 70 000 0 to 2 570
Endress+Hauser                                                                                                                                        5
                                                                                                                 Proline Promass E 200
DN x
8 3/8 85
15 ½ 110
25 1 125
40 1½ 125
50 2 125
To calculate the measuring range, use the Applicator sizing tool (® ä 49)
Operable flow range   Over 1000 : 1. Flow rates above the preset full scale value are not overridden by the electronics unit, with the
                      result that the totalizer values are registered correctly.
                      Output
Output signal         Current output
Resolution < 1 µA
Pulse/frequency/switch output
6                                                                                                                     Endress+Hauser
Proline Promass E 200
Pulse output
Frequency output
Switching output
Current output
                        4-20 mA
                        Failure mode                  Selectable (as per NAMUR recommendation NE 43):
                                                      • Minimum alarm:3.6 mA
                                                      • Maximum alarm:22 mA
                                                      • Adjustable value:3.59 to 22.5 mA
HART
Device diagnostics Device condition can be read out via HART Command 48
Endress+Hauser                                                                                               7
                                                                                                  Proline Promass E 200
Pulse/frequency/switch output
Pulse output
Frequency output
Switching output
Local display
       Backlight                    Additionally for device version with SD03 local display: red lighting indicates a device
                                    error.
       Operating tool
       • Via digital communication: HART protocol
       • Via service interface
Load Load for current output: 0 to 500 W, depending on the external supply voltage of the power supply unit
8                                                                                                         Endress+Hauser
Proline Promass E 200
500
400
300
200
100
                                     0
                                         14    16      18     20         22   24   26       28    30   32    34     36   Us [V]
                                                                                                                                      A0013563
                        1   Operating range
                        1.1 For order code for "Output", option A "4-20mA HART"/option B "4-20mA HART, pulse/frequency/switch output"
                            with Ex i and option C "4-20mA HART, 4-20mA"
                        1.2 For order code for "Output", option A "4-20mA HART"/option B "4-20mA HART, pulse/frequency/switch output"
                            with non-Ex and Ex d
                        Sample calculation
                        Supply voltage of the power supply unit: US =19 V
                        Maximum load: RB £ (19 V - 13.5 V) : 0.022 A = 250 W
Ex d type of protection
Ex nA type of protection
Endress+Hauser                                                                                                                             9
                                                                          Proline Promass E 200
Type of protection Ex ia
Type of protection Ex ic
IS type of protection
10                                                                                Endress+Hauser
Proline Promass E 200
Low flow cut off The switch points for low flow cut off are user-selectable.
Galvanic isolation All outputs are galvanically isolated from one another.
Manufacturer ID 0x11
                         Dynamic variables                The measured variables can be freely assigned to the dynamic variables.
                                                          Measured variables for PV (primary dynamic variable)
                                                          • Mass flow
                                                          • Volume flow
                                                          • Corrected volume flow
                                                          • Density
                                                          • Reference density
                                                          • Temperature
Endress+Hauser                                                                                                                      11
                                                                                                                     Proline Promass E 200
                      Power supply
Terminal assignment   Transmitter
2 1 3 2 1 3
                                         + –     + –
                                         3 4     1 2                                                    + –        + –
                                                                                                        3 4        1 2
                                                                                 A0013570                                                   A0018161
                      Maximum number of terminals, without integrated overvoltage Maximum number of terminals, with integrated
                      protection                                                  overvoltage protection
Supply voltage        An external power supply is required for each output. The following supply voltage values apply for the 4-20
                      mA and 4-20 mA HART current output:
Various power supply units can be ordered from Endress+Hauser: see "Accessories" section (® ä 50)
12                                                                                                                           Endress+Hauser
Proline Promass E 200
Current consumption     For every 4-20 mA or 4-20 mA HART current output: 3.6 to 22.5 mA
                             If the option Defined value is selected in the Failure mode parameter (® ä 7): 3.59 to 22.5 mA
A0015510
Connection examples
1 2 3 4 5 6
                                                                                                4...20 mA
                                                                    +                                               +
                                                                                                                           7
                                                                    -                                                -
A0015511
Endress+Hauser                                                                                                                    13
                                                                                                               Proline Promass E 200
1 2 3
                                                   +                                            4...20 mA
                                                                       +                                               +
                                                                                                                             4
                                                                       -                                               -
                                                   –
A0015512
                                          1                                             2
                                                                           –   +
                                                   _                                                                    +
                                                                                                                        –     3
                                                   +
                                   12345
                                                                                                                                    A0016801
                                         1                                              2
                                                                        _ +
                                                  _
                                                                                                                        +
                                                                                                                        _    3
                                                  +
A0016802
Terminals                • For device version without integrated overvoltage protection: plug-in spring terminals for wire cross-sections
                           0.5 to 2.5 mm2 (20 to 14 AWG)
                         • For device version with integrated overvoltage protection: screw terminals for wire cross-sections
                           0.2 to 2.5 mm2 (24 to 14 AWG)
14                                                                                                                    Endress+Hauser
Proline Promass E 200
Cable entries            • Cable gland (not for Ex d): M20 × 1.5 with cable Æ 6 to 12 mm (0.24 to 0.47 in)
                         • Thread for cable entry:
                           – For non-Ex and Ex: NPT ½"
                           – For non-Ex and Ex (not for CSA Ex d/XP): G ½"
                           – For Ex d: M20 × 1.5
Signal cable
                         Current output
                         • For 4-20 mA: standard installation cable is sufficient.
                         • For 4-20 mA HART: Shielded cable recommended. Observe grounding concept of the plant.
                         Pulse/frequency/switch output
                         Standard installation cable is sufficient.
Overvoltage protection   The device can be ordered with integrated overvoltage protection for ATEX/IECEx approvals:
                         Order code for "Accessory mounted", option NA "overvoltage protection"
                                 Depending on the temperature class, restrictions apply to the ambient temperature for device versions
                                 with overvoltage protection (® ä 21).
                         Performance characteristics
Reference operating      •    Error limits based on ISO 11631
conditions               •    Water with +15 to +45 °C (+59 to +113 °F) at 2 to 6 bar (29 to 87 psi)
                         •    Specifications as per calibration protocol ±5 °C (±9 °F) and ±2 bar (±29 psi)
                         •    Accuracy based on accredited calibration rigs that are traced to ISO 17025.
                                 To obtain measured errors, use the Applicator sizing tool (® ä 49)
                         Base accuracy
                         Mass flow and volume flow (liquids)
                         ±0.25 % o.r.
                         Mass flow (gases)
                         ±0.75 % o.r.
                                 Design fundamentals (® ä 18)
Endress+Hauser                                                                                                                      15
                                                                                                                             Proline Promass E 200
     Density (liquids)
     • Reference conditions:±0.0005 g/cm3
     • Standard density calibration:±0.02 g/cm3
       (valid over the entire temperature range and density range )
     Temperature
     ±0.5 °C ± 0.005 × T °C (±0.9 °F ± 0.005 × (T – 32) °F)
15 ½ 0.78 0.0287
25 1 2.16 0.0794
40 1½ 5.40 0.1985
50 2 8.40 0.3087
          E [%]
           2.5
2.0
1.5
1.0
0.5
                0
                    0       5   10     15    20   25   30    35    40     45    50   55   60    65     70     75   80    85     90   95   100 Q [%]
A0018212
     Flow values
     Flow values as turndown parameter depending on nominal diameter.
SI units
16                                                                                                                                   Endress+Hauser
Proline Promass E 200
US units
                        Accuracy of outputs
                        o.r. = of reading; o.f.s. = of full scale value
Current output
Accuracy ±10 µA
Pulse/frequency output
                        Density (liquids)
                        ±0.00025 g/cm3
                        Temperature
                        ±0.25 °C ± 0.0025 × T °C (±0.45 °F ± 0.0025 × (T–32) °F)
                        Temperature coefficient at       0.02 %/10 K, max. 0.35 % over the entire temperature range
                        zero point (4 mA)                –40 to +60 °C (–40 to +140 °F)
                        Temperature coefficient          0.05 %/10 K, max. 0.5 % over the entire temperature range
                        with span (20 mA)                –40 to +60 °C (–40 to +140 °F)
Pulse/frequency output
Endress+Hauser                                                                                                                          17
                                                                                                                       Proline Promass E 200
Influence of medium pressure   The table below shows the effect on accuracy of mass flow due to a difference between calibration pressure
                               and process pressure.
                               o.r. = of reading
DN [% o.r./bar]
[mm] [in]
8 3/8 no influence
15 ½ no influence
25 1 no influence
40 1½ no influence
50 2 –0.009
                               Installation
                               No special measures such as supports are necessary. External forces are absorbed by the construction of the
                               device.
Mounting location
A0015595
18                                                                                                                          Endress+Hauser
Proline Promass E 200
                        To prevent measuring errors arising from accumulation of gas bubbles in the measuring tube, avoid the following
                        mounting locations in the pipe:
                        • Highest point of a pipeline.
                        • Directly upstream of a free pipe outlet in a down pipe.
A0015596
8 3/8 6 0.24
15 ½ 10 0.40
25 1 14 0.55
40 1½ 22 0.87
50 2 28 1.10
Orientation             The direction of the arrow on the sensor nameplate helps you to install the sensor according to the flow direction
                        (direction of medium flow through the piping).
Orientation Recommendation
A Vertical orientation
A0015591
Endress+Hauser                                                                                                                            19
                                                                                                                                    Proline Promass E 200
                                                                           Orientation                                                  Recommendation
                                                                                                                                                  1)
                                     C       Horizontal orientation, transmitter head
                                             down                                                                                          Exception:
                                                                                                                                         (® å 7, ä 20)
                                                                                                                             A0015590
                                1)       Applications with high process temperatures may increase the ambient temperature. To maintain the maximum
                                         ambient temperature for the transmitter, this orientation is recommended.
                                If a sensor is installed horizontally with a curved measuring tube, match the position of the sensor to the fluid
                                properties.
                                                     1                                                                                  2
                                                                                                                                                       A0014057
Inlet and outlet runs           No special precautions need to be taken for fittings which create turbulence, such as valves, elbows or T-pieces,
                                as long as no cavitation occurs (® ä 29).
PT TT
3…5 × DN
4…8 × DN
A0016893
                                Rupture disk
                                Make sure that the function and operation of the rupture disk is not impeded through the installation of the
                                device. The position of the rupture disk is indicated on a sticker applied over it. If the rupture disk is triggered,
                                the sticker is destroyed. The disk can therefore be visually monitored.
20                                                                                                                                          Endress+Hauser
Proline Promass E 200
                                                                                 RUPTURE DISK
                                                                               12,5 BAR +/-20%@20°C
A0007823
                            Environment
Ambient temperature range   Measuring device                 –40 to +60 °C (–40 to +140 °F)
                            Local display                    –20 to +60 °C (–4 to +140 °F), the readability of the display may be impaired at
                                                             temperatures outside the temperature range.
                            ►      If operating outdoors:
                                   Avoid direct sunlight, particularly in warm climatic regions.
                            Temperature tables
                            The following interdependencies between the permitted ambient and fluid temperatures apply when operating
                            the device in hazardous areas:
                            SI units
                                 Nominal diameter           Ta           T6            T5            T4           T3            T2            T1
                                      [mm]                 [°C]        [85 °C]      [100 °C]      [135 °C]     [200 °C]      [300 °C]      [450 °C]
                            1)     The following applies for installations with overvoltage protection in conjunction with temperature class T5, T6 and
                                   approval codes IB, ID, IH, IJ, I4, BB, BD, BH, BJ, B2, C2, C5: Ta = Ta - 2 °C
                            US units
                                 Nominal diameter           Ta            T6           T5            T4            T3           T2            T1
                                      [in]                 [°F]        [185 °F]     [212 °F]      [275 °F]      [392 °F]     [572 °F]      [842 °F]
                            1)     The following applies for installations with overvoltage protection in conjunction with temperature class T5, T6 and
                                   approval codes IB, ID, IH, IJ, I4, BB, BD, BH, BJ, B2, C2, C5: Ta = Ta - 3.6 °F
Endress+Hauser                                                                                                                                      21
                                                                                                    Proline Promass E 200
     SI units
          Nominal diameter           Ta            T6           T5            T4            T3            T2            T1
               [mm]                 [°C]         [85 °C]     [100 °C]      [135 °C]      [200 °C]      [300 °C]      [450 °C]
     US units
          Nominal diameter            Ta            T6           T5           T4            T3            T2            T1
               [in]                  [°F]        [185 °F]     [212 °F]     [275 °F]      [392 °F]      [572 °F]      [842 °F]
     SI units
          Nominal diameter           Ta            T6           T5            T4            T3            T2            T1
               [mm]                 [°C]         [85 °C]     [100 °C]      [135 °C]      [200 °C]      [300 °C]      [450 °C]
     US units
          Nominal diameter            Ta            T6           T5           T4            T3            T2            T1
               [in]                  [°F]        [185 °F]     [212 °F]     [275 °F]      [392 °F]      [572 °F]      [842 °F]
22                                                                                                          Endress+Hauser
Proline Promass E 200
Ex ia, CCSAUS IS
                        SI units
                             Nominal diameter           Ta           T6            T5            T4            T3            T2            T1
                                  [mm]                 [°C]        [85 °C]      [100 °C]      [135 °C]      [200 °C]      [300 °C]      [450 °C]
                        1)     The following applies for installations with overvoltage protection in conjunction with temperature class T5, T6: Ta
                               = Ta - 2 °C
                        US units
                             Nominal diameter            Ta           T6            T5           T4            T3            T2            T1
                                  [in]                  [°F]       [185 °F]      [212 °F]     [275 °F]      [392 °F]      [572 °F]      [842 °F]
                        1)     The following applies for installations with overvoltage protection in conjunction with temperature class T5, T6: Ta
                               = Ta - 3.6 °F
                        SI units
                             Nominal diameter           Ta           T6            T5            T4            T3            T2            T1
                                  [mm]                 [°C]        [85 °C]      [100 °C]      [135 °C]      [200 °C]      [300 °C]      [450 °C]
                        1)     The following applies for installations with overvoltage protection in conjunction with temperature class T5, T6 and
                               approval codes ID, IG, IH, BD, BH, C4, C7: Ta = Ta - 2 °C
                        US units
                             Nominal diameter            Ta           T6            T5           T4            T3            T2            T1
                                  [in]                  [°F]       [185 °F]      [212 °F]     [275 °F]      [392 °F]      [572 °F]      [842 °F]
                        1)     The following applies for installations with overvoltage protection in conjunction with temperature class T5, T6 and
                               approval codes ID, IG, IH, BD, BH, C4, C7: Ta = Ta - 3.6 °F
Storage temperature –40 to +80 °C (–40 to +176 °F), preferably at +20 °C (+68 °F)
Endress+Hauser                                                                                                                                  23
                                                                                                                    Proline Promass E 200
                                Sensor
                                IP66/67, type 4X enclosure
Electromagnetic compatibility   As per IEC/EN 61326 and NAMUR Recommendation 21 (NE 21)
(EMC)                                Details are provided in the Declaration of Conformity.
                                Process
Medium temperature range        Sensor
                                –40 to +140 °C (–40 to +284 °F)
                                Seals
                                No internal seals
Pressure-temperature ratings The following material load diagrams refer to the entire device and not just the process connection.
                                                 [psi] [bar]
                                                        100
                                               1400                   PN 100
                                                         90
                                              1200
                                                         80
                                              1000       70
                                                         60           PN 63
                                                 800
                                                         50
                                                 600     40
                                                                      PN 40
                                                 400     30
                                                         20
                                                 200
                                                         10
                                                    0     0
                                                               -50             0          50     100      150         200 [°C]
                                                                -50     0          50   100 150 200 250 300 350 400 [°F]
                                                                                                                                       A0006904
24                                                                                                                         Endress+Hauser
Proline Promass E 200
                                      [psi] [bar]
                                             100
                                     1400
                                               90
                                     1200
                                               80          Class 600
                                     1000      70
                                               60
                                      800
                                               50
                                      600      40
                                                           Class 300
                                      400      30
                                               20
                                      200
                                               10          Class 150
                                         0      0
                                                    -50            0          50     100     150     200 [°C]
                                                     -50       0       50   100 150 200 250 300 350 400 [°F]
                                                                                                                A0006905
                                      [psi] [bar]
                                     1000     70
                                               60          63K
                                      800
                                               50
                                      600      40
                                                           40K
                                      400      30
                                               20
                                      200                  20K
                                               10
                                                           10K
                                         0      0
                                                    -50            0          50     100     150     200 [°C]
                                                     -50       0       50   100 150 200 250 300 350 400 [°F]
                                                                                                                A0006906
                                      [psi] [bar]
                                             100
                                     1400                  PN 100
                                              90
                                     1200     80
                                     1000     70
                                                    -50            0          50     100     150     200 [°C]
                                                     -50       0       50   100 150 200 250 300 350 400 [°F]
                                                                                                                A0006908
Endress+Hauser                                                                                                     25
                                                                                             Proline Promass E 200
     Tri-Clamp
     The clamp connections are suitable up to a maximum pressure of 16 bar (232 psi). Please observe the operating
     limits of the clamp and seal used, as they could be under 16 bar (232 psi) . The clamp and seal are not included
     in the scope of supply.
                    [psi] [bar]
                            50
                    600     40
                                          DN 8…40
                    400      30
                             20           DN 50
                    200
                             10
                       0      0
                                   -50            0          50       100         150         200 [°C]
                                    -50     0         50   100 150 200 250 300 350 400 [°F]
                                                                                                                  A0013664
     DIN 11851 allows for applications up to +140 °C (+284 °F) if suitable sealing materials are used. Please take
     this into account when selecting seals and counterparts, as these components can limit the pressure and
     temperature range.
                     [psi] [bar]
                    400      30
                             20
                    200
                             10           PN 16
                       0      0
                                   -50            0          50       100         150         200 [°C]
                                    -50     0         50   100 150 200 250 300 350 400 [°F]
                                                                                                                  A0012947
     SMS 1145 allows for applications up to 6 bar (87 psi) if suitable sealing materials are used. Please take this into
     account when selecting seals and counterparts, as these components can limit the pressure and temperature
     range.
26                                                                                                  Endress+Hauser
Proline Promass E 200
                                       [psi] [bar]
                                               50
                                       600     40
                                                            DN 8…40
                                       400     30
                                               20           DN 50
                                       200
                                               10
                                          0      0
                                                     -50            0          50       100         150         200 [°C]
                                                      -50      0        50   100 150 200 250 300 350 400 [°F]
                                                                                                                              A0013665
                                       [psi] [bar]
                                       600     40
                                       400     30           DN 8...40
                                               20
                                       200
                                               10           DN 50
                                          0      0
                                                     -50            0          50       100         150         200 [°C]
                                                      -50     0         50   100 150 200 250 300 350 400 [°F]
                                                                                                                              A0013666
                                       [psi] [bar]
                                       400     30
                                               20
                                       200
                                               10
                                          0      0
                                                     -50            0          50       100         150         200 [°C]
                                                      -50     0         50   100 150 200 250 300 350 400 [°F]
                                                                                                                              A0006912
Secondary containment   The sensor housing is filled with dry nitrogen and protects the electronics and mechanics inside.
pressure range
                        It does not fulfill any secondary containment function.
                        Reference value for the pressure loading capacity of the sensor housing: 15 bar (217.5 psi)
Rupture disk            To increase the level of safety, a device version with a rupture disk with a triggering pressure of
                        10 to 15 bar (145 to 217.5 psi) can be used.
                        Rupture disks cannot be combined with the separately available heating jacket (® ä 48).
Endress+Hauser                                                                                                                   27
                                                                                                                          Proline Promass E 200
Flow limit      Select the nominal diameter by optimizing between the required flow range and permissible pressure loss.
                     For an overview of the measuring range full scale values, see the "Measuring range" section (® ä 5)
                • The minimum recommended full scale value is approx. 1/20 of the maximum full scale value
                • In most applications, 20 to 50 % of the maximum full scale value can be considered ideal
                • Select a lower full scale value for abrasive substances (such as liquids with entrained solids): flow velocity
                  <1 m/s (<3 ft/s).
                • For gas measurement the following rules apply:
                  – The flow velocity in the measuring tubes should not exceed half the sonic velocity (0.5 Mach).
                  – The maximum mass flow depends on the density of the gas: formula (® ä 5)
Pressure loss To calculate the pressure loss, use the Applicator sizing tool (® ä 49)
                Re ³ 2300 *                                                 Dp = K · n
                                                                                         0.25
                                                                                                ·g
                                                                                                  1.85
                                                                                                         ·r
                                                                                                           –0.86
A0004626
                Re < 2300                                                                                K2 · n · g
                                                                                                                   0.25    2
                                                                            Dp = K1 · n · g +
                                                                                                               r
                                                                                                                                               A0004628
* To calculate the pressure loss for gases, always use the formula for Re ³ 2300.
28                                                                                                                             Endress+Hauser
Proline Promass E 200
                                 [mbar]
                                 10000
                                                                                                            DN 40
                                                                                                    DN 25
                                                                                           DN 15               DN 50
                                                                                    DN 8
                                  1000
100
10
                                    0.1
                                      0.001      0.01              0.1          1              10                   100   1000 [t/h]
A0013674
System pressure         It is important that cavitation does not occur, or that gases entrained in the liquids do not outgas. This is
                        prevented by means of a sufficiently high system pressure.
                        For this reason, the following mounting locations are recommended:
                        • At the lowest point in a vertical pipe
                        • Downstream from pumps (no danger of vacuum)
A0015594
Thermal insulation      In the case of some fluids, it is important that the heat radiated from the sensor to the transmitter is kept to a
                        minimum. A wide range of materials can be used for the required insulation.
                        Ensure that only up to 20 mm (0.79 in) of the transmitter neck is insulated so that the transmitter head is
                        completely free.
Heating                 Some fluids require suitable measures to avoid loss of heat at the sensor.
                        Heating options
                        • Electrical heating, e.g. with electric band heaters
                        • Via pipes carrying hot water or steam
                        • Via heating jackets
                               Heating jackets for the sensor can be ordered as accessories from Endress+Hauser (® ä 48).
Vibrations              The high oscillation frequency of the measuring tubes ensures that the correct operation of the measuring system
                        is not influenced by plant vibrations.
Endress+Hauser                                                                                                                             29
                                                                                                                         Proline Promass E 200
                     Mechanical construction
Design, dimensions   Compact version
Order code for "Housing", options B "GT18 two-chamber, 316L", C "GT20 two-chamber aluminum coating"
                                                 A                                                     D
                                             B          C                                        E           F
                                                                                                                                 H
                                                                                                                                     J
                                                                                                                                 G
                                                            M                                         L
A0013552
30                                                                                                                               Endress+Hauser
Proline Promass E 200
                                                                    C
                                                    B
                                              A
                                                    E
                                                                    D
+1,5 (+0.06)
A0015621
                                                                                 1)
                        Flange based on EN 1092-1 (DIN 2501 / DIN 2512N            ) / PN 40: 1.4404/316L (order code for "Process
                        connection", option D2S)
                        Surface roughness (flange): EN 1092-1 Form B1 (DIN 2526 Form C), Ra 3.2 to 12.5 mm
                              DN               A               B                C                 D                        E        F
                             [mm]            [mm]            [mm]             [mm]              [mm]                     [mm]     [mm]
                        1)    Flange with groove based on EN 1092-1 Form D (DIN 2512N) available (order code for "Process connection", option
                              D6S)
                        2)    Installation in accordance with NAMUR recommendation NE 132 optionally available (order code for "Process
                              connection", option D2N)
                        Flange based on EN 1092-1 (DIN 2501) / PN 40 (with DN 25 flanges): 1.4404/316L (order code for "Process
                        connection", option R2S)
Surface roughness (flange): EN 1092-1 Form B1 (DIN 2526 Form C), Ra 3.2 to 12.5 mm
                              DN               A               B                 C                D                        E        F
                             [mm]            [mm]            [mm]              [mm]             [mm]                     [mm]     [mm]
                        Flange based on EN 1092-1 (DIN 2501) / PN 40 with groove: 1.4404/316L (order code for "Process
                        connection", option D6N)
Surface roughness (flange): EN 1092-1 Form B1 (DIN 2526 Form C), Ra 3.2 to 12.5 mm
                              DN               A               B                 C                D                        E        F
                             [mm]            [mm]            [mm]              [mm]             [mm]                     [mm]     [mm]
Endress+Hauser                                                                                                                              31
                                                                                                        Proline Promass E 200
                                                              1)
     Flange based on EN 1092-1 (DIN 2501 / DIN 2512N            ) / PN 63: 1.4404/316L (order code for "Process
     connection", option D3S)
Surface roughness (flange): EN 1092-1 Form B1 (DIN 2526 Form C), Ra 0.8 to 3.2 mm
           DN               A                B                C                D                          E            F
          [mm]            [mm]             [mm]             [mm]             [mm]                       [mm]         [mm]
     1)    Flange with groove based on EN 1092-1 Form D (DIN 2512N) available (order code for "Process connection", option
           D7S)
                                                              1)
     Flange based on EN 1092-1 (DIN 2501 / DIN 2512N            ) / PN 100: 1.4404/316L (order code for "Process
     connection", option D4S)
Surface roughness (flange): EN 1092-1 Form B2 (DIN 2526 Form E), Ra 0.8 to 3.2 mm
           DN               A               B                 C                D                          E            F
          [mm]            [mm]            [mm]              [mm]             [mm]                       [mm]         [mm]
     1)    Flange with groove based on EN 1092-1 Form D (DIN 2512N) available (order code for "Process connection", option
           D8S)
+1,5 (+0.06)
A0015621
Flange based on ASME B16.5 / Cl 150: 1.4404/316L (order code for "Process connection", option AAS)
           DN              A               B                  C                 D                         E             F
          [mm]           [mm]            [mm]               [mm]              [mm]                      [mm]          [mm]
32                                                                                                              Endress+Hauser
Proline Promass E 200
Flange based on ASME B16.5 / Cl 300: 1.4404/316L (order code for "Process connection", option ABS)
                             DN             A             B                C               D                         E      F
                            [mm]          [mm]          [mm]             [mm]            [mm]                      [mm]   [mm]
Flange based on ASME B16.5 / Cl 600: 1.4404/316L (order code for "Process connection", option ACS)
                             DN             A             B                C               D                         E      F
                            [mm]          [mm]          [mm]             [mm]            [mm]                      [mm]   [mm]
+1,5 (+0.06)
A0015621
                        Flange JIS B2220 / 10K: SUS 316L (order code for "Process connection", option NDS)
                             DN              A             B               C               D                         E      F
                            [mm]           [mm]          [mm]            [mm]            [mm]                      [mm]   [mm]
Flange JIS B2220 / 20K: SUS 316L (order code for "Process connection", option NES)
                             DN              A             B               C               D                         E      F
                            [mm]           [mm]          [mm]            [mm]            [mm]                      [mm]   [mm]
8 95 70 4 ´ Æ15 14 15 232
15 95 70 4 ´ Æ15 14 15 279
Endress+Hauser                                                                                                                      33
                                                                                                      Proline Promass E 200
Flange JIS B2220 / 40K: SUS 316L (order code for "Process connection", option NGS)
           DN                  A           B                 C                  D                       E              F
          [mm]               [mm]        [mm]              [mm]               [mm]                    [mm]           [mm]
Flange JIS B2220 / 63K: SUS 316L (order code for "Process connection", option NHS)
           DN                  A           B                 C                  D                       E              F
          [mm]               [mm]        [mm]              [mm]               [mm]                    [mm]           [mm]
     VCO connections
                            A
                                    B
+1,5 (+0.06)
A0015624
34                                                                                                           Endress+Hauser
Proline Promass E 200
Tri-Clamp
                                                    B
                                                                                                          +1,5 (+0.06)
                                                                                                        C - 2,0 (-0.08)
                                            mm (in)
                                                                                                                                          A0015625
1", 1½", 2" Tri-Clamp: 1.4404/316L (order code for "Process connection", option FTS)
                                DN                      Clamp                      A                      B                       C
                               [mm]                      [in]                    [mm]                   [mm]                    [mm]
                        3A version available (order code for "Additional approval", option LP) in combination with Ra £0.8 µm Ra £0.4 µm (order
                        code for "Measuring tube material", option SB, SC)
½" Tri-Clamp: 1.4404/316L (order code for "Process connection", option FUW)
                                DN                      Clamp                      A                      B                       C
                               [mm]                      [in]                    [mm]                   [mm]                    [mm]
                        3A version available (order code for "Additional approval", option LP) in combination with Ra £0.8 µm Ra £0.4 µm (order
                        code for "Measuring tube material", option SB, SC)
Endress+Hauser                                                                                                                               35
                                                                                             Proline Promass E 200
                            A
                                B
                                                                             +1,5 (+0.06)
                                                                           C - 2,0 (-0.08)
                         mm (in)
                                                                                                             A0015628
Sanitary connection DIN 11851: 1.4404/316L (order code for "Process connection", option FMW)
                DN                                 A                  B                                C
               [mm]                               [in]              [mm]                             [mm]
8 Rd 34 ´ 1/8 16 229
15 Rd 34 ´ 1/8 16 273
25 Rd 52 ´ 1/6 26 324
40 Rd 65 ´ 1/6 38 456
50 Rd 78 ´ 1/6 50 562
                                                                             +1,5 (+0.06)
                                                                           C - 2,0 (-0.08)
                         mm (in)
                                                                                                             A0015628
     Threaded hygienic connection DIN 11864-1 Form A: 1.4404/316L (order code for "Process connection",
     option FLW)
                DN                                 A                 B                                C
               [mm]                               [in]             [mm]                             [mm]
8 Rd 28 ´ 1/8 10 229
15 Rd 34 ´ 1/8 16 273
25 Rd 52 ´ 1/6 26 324
40 Rd 65 ´ 1/6 38 456
36                                                                                                Endress+Hauser
Proline Promass E 200
                        Threaded hygienic connection DIN 11864-1 Form A: 1.4404/316L (order code for "Process connection",
                        option FLW)
                                      DN                                A                            B                            C
                                     [mm]                              [in]                        [mm]                         [mm]
50 Rd 78 ´ 1/6 50 562
                        3A version available (order code for "Additional approval", option LP) in combination with Ra £0.8 µm (order code for
                        "Measuring tube material", option SB)
X X
                                                                  C
                                             A
                                                 B
                                                       E
+1,5 (+0.06)
A0015627
å 18 Detail X: Asymmetrical process connection; the part shown in gray is provided by the supplier.
                        DIN 11864-2 Form A (flange with notch): 1.4404/316L (order code for "Process connection", option FKW)
                             DN                    A               B                C                 D                  E                  F
                            [mm]                 [mm]            [mm]             [mm]              [mm]               [mm]               [mm]
8 54 37 4 ´ Æ9 10 10 249
15 59 42 4 ´ Æ9 10 16 293
25 70 53 4 ´ Æ9 10 26 344
40 82 65 4 ´ Æ9 10 38 456
50 94 77 4 ´ Æ9 10 50 562
Endress+Hauser                                                                                                                                      37
                                                                                                      Proline Promass E 200
                            A
                                B
                                                                                      +1,5 (+0.06)
                                                                                    C - 2,0 (-0.08)
                         mm (in)
                                                                                                                      A0015623
Threaded hygienic connection ISO 2853: 1.4404/316L (order code for "Process connection", option FJW)
                 DN                               A 1)                      B                                  C
                [mm]                             [mm]                     [mm]                               [mm]
1) Max. thread diameter to ISO 2853 Annex A; 3A version also available (Ra £ 0.8 mm)
                                                                                       +1,5 (+0.06)
                                                                                    C - 2,0 (-0.08)
                         mm (in)
                                                                                                                      A0015628
Threaded hygienic connection SMS 1145: 1.4404/316L (order code for "Process connection", option FSW)
                DN                                 A                          B                                C
               [mm]                               [in]                      [mm]                             [mm]
38                                                                                                         Endress+Hauser
Proline Promass E 200
                                                               C
                                                B
                                           A
                                                E
                                                               D
+1,5 (+0.06)
A0015621
Flange based on ASME B16.5 / Cl 150: 1.4404/316L (order code for "Process connection", option AAS)
                            DN            A               B              C               D                        E       F
                            [in]         [in]           [in]            [in]            [in]                     [in]   [in]
Flange based on ASME B16.5 / Cl 300: 1.4404/316L (order code for "Process connection", option ABS)
                            DN            A               B              C               D                        E       F
                            [in]         [in]           [in]            [in]            [in]                     [in]   [in]
Flange based on ASME B16.5 / Cl 600: 1.4404/316L (order code for "Process connection", option ACS)
                            DN            A               B              C               D                        E       F
                            [in]         [in]           [in]            [in]            [in]                     [in]   [in]
Endress+Hauser                                                                                                                     39
                                                                                                        Proline Promass E 200
VCO connections
                             A
                                   B
                                                                                       +1,5 (+0.06)
A0015624
                  DN                               A                            B                                C
                 [mm]                             [in]                        [in]                              [in]
     Tri-Clamp
                               A
                                                                                         +1,5 (+0.06)
                                                                                      C - 2,0 (-0.08)
                             mm (in)
                                                                                                                          A0015625
     1", 1½", 2" Tri-Clamp: 1.4404/316L (order code for "Process connection", option FTS)
              DN                          Clamp              A                            B                         C
              [in]                         [in]             [in]                        [in]                       [in]
40                                                                                                           Endress+Hauser
Proline Promass E 200
½" Tri-Clamp: 1.4404/316L (order code for "Process connection", option FUW)
                                 DN                    Clamp                      A                         B                          C
                                 [in]                   [in]                     [in]                     [in]                        [in]
                        3A version available (order code for "Additional approval", option LP) in combination with Ra £0.8 µm Ra £0.4 µm (order
                        code for "Measuring tube material", option SB, SC)
                                               A
                                                   B
                                                                                                            +1,5 (+0.06)
                                                                                                          C - 2,0 (-0.08)
                                             mm (in)
                                                                                                                                             A0015628
                        Threaded hygienic connection SMS 1145: 1.4404/316L (order code for "Process connection", option FSW)
                                      DN                           A                             B                                  C
                                      [in]                        [in]                          [in]                               [in]
                        3A version available (order code for "Additional approval", option LP) in combination with Ra£0.8 µm (order code for
                        "Measuring tube material", option SB)
                        Weight in SI units
                        All values (weight) refer to devices with EN/DIN PN 40 flanges. Weight information in [kg]
                                    DN                                                      Weight [kg]
                                   [mm]
                                                        Order characteristic for "Housing",            Order characteristic for "Housing",
                                                                    option C                                       option B
                                                                Aluminum coating                                 1.4404/316L
8 6 8.5
15 6.5 9
25 8 10.5
40 13 15.5
50 22 24.5
Endress+Hauser                                                                                                                                  41
                                                                                                 Proline Promass E 200
            Weight in US units
            All values (weight) refer to devices with EN/DIN PN 40 flanges. Weight information in [lbs]
                        DN                                                   Weight [lbs]
                        [in]
                                            Order characteristic for "Housing",      Order characteristic for "Housing",
                                                        option C                                 option B
                                                    Aluminum coating                           1.4404/316L
½ 14.3 19.8
1 17.6 23.2
1½ 28.7 34.2
2 48.5 54.0
                                                  • Ex nA                                       Nickel-plated brass
                                                  • Ex tb
            Sensor housing
            • Acid and alkali-resistant outer surface
            • Stainless steel 1.4301/304
42                                                                                                          Endress+Hauser
Proline Promass E 200
                        Measuring tubes
                        • Stainless steel EN 1.4539 / ASTM 904L
                        • Surface quality:
                          – Ramax = 0.8 mm (32 min)
                          – Ramax = 0.4 mm (16 min)
                        Process connections
                        • For all process connections (except flanges as per JIS B2220):
                          Stainless steel 1.4404/316L
                        • For flanges as per JIS B2220:
                          Stainless steel SUS 316L
                                List of all available process connections (® ä 43)
                        Seals
                        Welded process connections without internal seals
Accessories
                        Operability
Operating concept       Operator-oriented menu structure for user-specific tasks
                        • Commissioning
                        • Operation
                        • Diagnostics
                        • Expert level
                        Quick and safe commissioning
                        • Guided menus ("Make-it-run" wizards) for applications
                        • Menu guidance with brief explanations of the individual parameter functions
                        Reliable operation
                        • Operation in the following languages:
                          – Via local display:
                             English, German, French, Spanish, Italian, Dutch, Portuguese, Polish, Russian, Turkish, Chinese, Japanese,
                             Bahasa (Indonesian), Vietnamese, Czech
                          – Via "FieldCare" operating tool:
                             English, German, French, Spanish, Italian, Dutch, Japanese
                        • Uniform operating philosophy applied to device and operating tools
                        • If replacing the electronic module, transfer the device configuration via the integrated memory (integrated
                          HistoROM) which contains the process and measuring device data and the event logbook. No need to
                          reconfigure.
Endress+Hauser                                                                                                                      43
                                                                                                                  Proline Promass E 200
Local operation Order code for "Display; Operation", option C "SD02" Order code for "Display; Operation", option E "SD03"
1 1
                                              –                                                      –
                                                  +                                                      +
                                                      E                                                      E
A0015544 A0015546
                  Display elements
                  • 4-line display
                  • In the case of order code for "Display; Operation", option E: white background lighting; switches to red in
                    event of device errors
                  • Format for displaying measured variables and status variables can be individually configured
                  • Permitted ambient temperature for the display: –20 to +60 °C (–4 to +140 °F)
                    The readability of the display may be impaired at temperatures outside the temperature range.
                  Operating elements
                  • In the case of order code "Display; Operation", Option C: local operation with 3 push buttons ( , , )                    E
                  • In the case of order code for "Display; Operation", option E: external operation via touch control; 3 optical
                    keys: , ,       E
44                                                                                                                        Endress+Hauser
Proline Promass E 200
                                                                      6                 8
                                   3
                                   2
                                        1                  4                5                 7
                                                                                                                         A0013764
                                    3                                                                  –
                                                                                                           +
                                                                                                               E
                                                  2                                 1
                                                                                                                         A0014019
                        1   Service interface (CDI = Endress+Hauser Common Data Interface) of the measuring device
                        2   Commubox FXA291
                        3   Computer with "FieldCare" operating tool with COM DTM "CDI Communication FXA291"
Endress+Hauser                                                                                                              45
                                                                                                                  Proline Promass E 200
C-Tick symbol            The measuring system meets the EMC requirements of the "Australian Communications and Media Authority
                         (ACMA)".
Ex approval              The measuring device is certified for use in hazardous areas and the relevant safety instructions are provided
                         in the separate "Safety Instructions" (XA) document. Reference is made to this document on the nameplate.
                                 The separate Ex documentation (XA) containing all the relevant explosion protection data is available
                                 from your Endress+Hauser sales center.
                         ATEX/IECEx
                         Currently, the following versions for use in hazardous areas are available:
                         Ex d
                                              Category (ATEX)                                       Type of protection
                         Ex ia
                                              Category (ATEX)                                       Type of protection
                         Ex nA
                                              Category (ATEX)                                       Type of protection
                         Ex ic
                                              Category (ATEX)                                       Type of protection
CCSAUS
                         Currently, the following versions for use in hazardous areas are available:
                         NI (Ex nA, Ex nL)
                         • Class I Division 2 Groups ABCD
                         • Class II, III Division 1 Groups EFG
                         IS (Ex i) and XP (Ex d)
                         Class I, II, III Division 1 Groups ABCDEFG
46                                                                                                                          Endress+Hauser
Proline Promass E 200
Functional safety              The measuring device can be used for flow monitoring systems (min., max., range) up to SIL 2 (single-channel
                               architecture) and SIL 3 (multichannel architecture with homogeneous redundancy) and is independently
                               evaluated and certified by the TÜV in accordance with IEC 61508.
                               The following types of monitoring in safety equipment are possible:
                               • Mass flow
                               • Volume flow
                               • Density
                                    The Manual for Functional Safety with all information about the SIL device is available from your Endress
                                    +Hauser sales office.
Pressure Equipment Directive   The devices can be ordered with or without a PED approval. If a device with a PED approval is required, this
                               must be explicitly stated in the order. For devices with nominal diameters less than or equal to DN 25 (1"),
                               this is neither possible nor necessary.
                               • With the identification PED/G1/x (x = category) on the sensor nameplate, Endress+Hauser confirms
                                 conformity with the "Basic Safety Requirements" specified of Appendix I of the Pressure Equipment Directive
                                 97/23/EC.
                               • Devices bearing this marking (PED) are suitable for the following types of medium:
                                 – Media in Group 1 and 2 with a vapor pressure greater than, or smaller and equal to 0.5 bar (7.3 psi)
                                 – Unstable gases
                               • Devices not bearing this marking (PED) are designed and manufactured according to good engineering
                                 practice. They meet the requirements of Art.3 Section 3 of the Pressure Equipment Directive 97/23/EC.
                                 The range of application is indicated in tables 6 to 9 in Annex II of the Pressure Equipment Directive.
Endress+Hauser                                                                                                                            47
                                                                                                                            Proline Promass E 200
                              Ordering information
                              Detailed ordering information is available from the following sources:
                              • In the Product Configurator on the Endress+Hauser website: www.endress.com ® Select country ®
                                Instruments ® Select device ® Product page function: Configure this product
                              • From your Endress+Hauser Sales Center: www.endress.com/worldwide
                                   Product Configurator - the tool for individual product configuration
                                   • Up-to-the-minute configuration data
                                   • Depending on the device: Direct input of measuring point-specific information such as measuring range
                                     or operating language
                                   • Automatic verification of exclusion criteria
                                   • Automatic creation of the order code and its breakdown in PDF or Excel output format
                                   • Ability to order directly in the Endress+Hauser Online Shop
                              Application packages
                              Many different application packages are available to enhance the functionality of the device. Such packages
                              might be needed to address safety aspects or specific application requirements.
                              The application packages can be ordered from Endress+Hauser either directly with the device or subsequently.
                              Detailed information on the order code in question is available from your local Endress+Hauser sales center or
                              on the product page of the Endress+Hauser website: www.endress.com.
                              HistoROM extended           Comprises extended functions concerning the event log and the activation of the measured
                              function                    value memory (data logger).
                                                          Event log:
                                                          Memory volume is extended from 20 message entries (basic version) to up to 100 entries.
                                                          Data logging (line recorder):
                                                          • Memory capacity for up to 1000 measured values is activated.
                                                          • 250 measured values can be output via each of the 4 memory channels. The recording
                                                            interval can be defined and configured by the user.
                                                          • Data logging is visualized via the local display or FieldCare.
                              Accessories
                              Various accessories, which can be ordered with the device or subsequently from Endress+Hauser, are available
                              for the device. Detailed information on the order code in question is available from your local Endress+Hauser
                              sales center or on the product page of the Endress+Hauser website: www.endress.com.
Accessories Description
                              Promass 200 transmitter     Transmitter for replacement or storage. Use the order code to define the following
                                                          specifications:
                                                          • Approvals
                                                          • Output
                                                          • Display / operation
                                                          • Housing
                                                          • Software
                                                                For details, see Installation Instructions EA00104D
                              Weather protection cover    Is used to protect the measuring device from the effects of the weather: e.g. rainwater, excess
                                                          heating from direct sunlight or extreme cold in winter.
                                                                For details, see Installation Instructions SD00333F
48                                                                                                                                  Endress+Hauser
Proline Promass E 200
Accessories Description
                               Heating jacket         Is used to stabilize the temperature of the fluids in the sensor.
                                                      Water, water vapor and other non-corrosive liquids are permitted for use as fluids. If using
                                                      oil as a heating medium, please consult with Endress+Hauser.
                                                      Heating jackets cannot be used with sensors fitted with a rupture disk.
                                                            For details, see Operating Instructions BA00099D
                               Commubox FXA195 HART   For intrinsically safe HART communication with FieldCare via the USB interface.
                                                            For details, see "Technical Information" TI00404F
                               Commubox FXA291        Connects Endress+Hauser field devices with a CDI interface (= Endress+Hauser Common
                                                      Data Interface) and the USB port of a computer or laptop.
                                                            For details, see "Technical Information" TI00405C
                               HART Loop Converter    Is used to evaluate and convert dynamic HART process variables to analog current signals
                               HMX50                  or limit values.
                                                            For details, see "Technical Information" TI00429F and Operating Instructions
                                                            BA00371F
                               Fieldgate FXA320       Gateway for the remote monitoring of connected 4-20 mA measuring devices via a Web
                                                      browser.
                                                            For details, see "Technical Information" TI00025S and Operating Instructions
                                                            BA00053S
                               Fieldgate FXA520       Gateway for the remote diagnostics and remote configuration of connected HART measuring
                                                      devices via a Web browser.
                                                            For details, see "Technical Information" TI00025S and Operating Instructions
                                                            BA00051S
                               Field Xpert SFX100     Compact, flexible and robust industry handheld terminal for remote configuration and for
                                                      obtaining measured values via the HART current output (4-20 mA).
                                                            For details, see Operating Instructions BA00060S
Endress+Hauser                                                                                                                                    49
                                                                                                                   Proline Promass E 200
                         Memograph M graphic      The Memograph M graphic data manager provides information on all the relevant measured
                         display recorder         variables. Measured values are recorded correctly, limit values are monitored and measuring
                                                  points analyzed. The data are stored in the 256 MB internal memory and also on a SD card
                                                  or USB stick.
                                                        For details, see "Technical Information" TI00133R and Operating Instructions
                                                        BA00247R
                         RN221N                   Active barrier with power supply for safe separation of 4-20 mA standard signal circuits.
                                                  Offers bidirectional HART transmission.
                                                        For details, see "Technical Information" TI00073R and Operating Instructions
                                                        BA00202R
                         RNS221                   Supply unit for powering two 2-wire measuring devices solely in the non-Ex area.
                                                  Bidirectional communication is possible via the HART communication jacks.
                                                        For details, see "Technical Information" TI00081R and Brief Operating Instructions
                                                        KA00110R
                         Documentation
                              The following document types are available:
                              • On the CD-ROM supplied with the device
                              • In the Download Area of the Endress+Hauser Internet site: www.endress.com ® Download
50                                                                                                                         Endress+Hauser
Proline Promass E 200
ATEX/IECEx Ex i XA00144D
cCSAus IS XA00151D
cCSAus XP XA00152D
                          Registered trademarks
                          HARTÒ
                          Registered trademark of the HART Communication Foundation, Austin, USA
                          TRI-CLAMP Ò
                          Registered trademark of Ladish & Co., Inc., Kenosha, USA
                          ApplicatorÒ, FieldCareÒ, Field XpertTM, HistoROMÒ
                          Registered or registration-pending trademarks of the Endress+Hauser Group
Endress+Hauser                                                                                                                        51