Technical Data
DefiMonitor EVO
Defibrillation
Impulse shape                                             Biphasic
Energy in the automatic mode                              According to patient impedance
Energy Levels in Manual Mode                              1J, 2J, 3J, 4J, 5J, 6J, 7J, 8J, 9J, 10J, 15J, 20J, 30J, 40J, 50J, 75J, 100J, 125J, 150J, 175J, 200J, 300J, 360J
Charge Time                                               < 6 s for 200 J (For a new, full charged battery at 25 °C), 8 s for 360 J (For a new, full charged battery at 25 °C)
Pacemaker detection
                                                          Detect pacer pulses of ±2 mV to ±700 mV with pulse widths of 0,1 msec to 2 msec and rise times 10 % of width but not more than 100 μs
Cardioversion                                             Manual activation of the cardioversion via the SYNC button, defibrillation via paddles as well as self-adhesive electrodes
Defibrillation electrodes
Adhesive electrodes
Shelf life                                                36 months
Cable length                                              2,0 m
Activ electroode surface                                  174 cm² overall
Paddels                                                   Adult Electrodes with built-in child‘s electrodes
Functions Paddles                                         Energy selection and charging / discharging as well as printing of data by operating the paddles
Resuscitation guidelines
Factory set                                               Guideline 2010 (ERC / AHA)
ECG
12-Channel-Leads                                          Extremities acc. to Einthoven (bipolar) I, II, III, extremities acc. to Goldberger (unipolar) aVR, aVL, aVF, thoracic wall leads acc. to Wilson
                                                          (unipolar) V1, V2, V3, V4, V5, V6
Heart rate                                                0,20 to 300 bpm
Analysis time                                             5–9s
Interpretative algorithm                                  12-channel interpretation analysis program of Glasgow University
Respiration
Technology                                                Impedance measurement via electrodes (Impedance Pneumography) resp. capnography / Nondispersive infrared spectroscopy
Transthoracic pacemaker (NIP)
Modes                                                     Demand, Async (FIX)
Stimulation frequency                                     30 ppm to 180 ppm
Impulse current                                           0 mA to 140 mA
Impulse shape                                             Rectilinear, Constant current
Pulse oximeter (SpO2)*
Model                                                     Nellcor ®
Range                                                     1 % to 100 %
Pulse rate                                                20 – 300 bpm
Accuracy                                                  70% – 100%“
NIBP*
Operation Mode                                            Manual and automatic measurement of interval
Technology                                                Oscillometric Measurement
Measurement Range Adult / Pediatric                       SYS 60 to 250 mmHg / MAP 45 to 235 mmHg / DIA 40 to 200 mmHg
Measurement Range Neonatal                                SYS 40 to 120 mmHg / MAP 30 to 100 mmHg / DIA 20 to 90 mmHg
IBP*
Measurement Range                                         -50 – 300mmHG
Channels                                                  2 parallel connections
Capnography (EtCO2)*
Type/EtCO2 -Module                                        Respironics® / Sidestream / Mainstream
Ambient-CO2-level                                         0 – 150 mmHg
Parameter                                                 EtCO2, InCO2
Temperature*
Measurement Method                                        Thermistor, YSI compatible
Range                                                     0 to 50 °C (32.0 to 122 °F)
Monitor
Model                                                     TFT-LCD color, dual mode
Dimensions                                                170 x 128 mm (diagonal 8.4“/ 212 mm)
Resolution                                                800 x 600 Pixel
Number of channels                                        4 or 12 ECG leads, cascaded can be shown
Integrated Printer
Model                                                     Thermal print
Channels                                                  12 channels
Paper                                                     Thermal paper folded
Paper Width                                               80 mm
Printer Speed                                             25/50 mm/s
Data management
Memory                                                    12-lead-ECG (100 records) / Event (250 records)
Storage medium                                            USB memory, SD card
Externel Interfaces
USB                                                       For software updates (with external data carrier)
SD Memory Card                                            For extended memory (for software updates)
Telemetry                                                 3G/GSM, WLAN
Power supply
AC Power                                                  100V – 240V, 50Hz / 60Hz
Rechargeable battery                                      Lithium-Ion, 14,4V/ 5200 mAh
Operation time                                            Up to 10 hours
Number of shocks                                          200 shocks @ 200 J (1 accu), 400 shocks @ 200 J (2 accus)
Recharging Time                                           Approx. 5 hours
Dimensions
Hight x Width x Depth                                     Approx. 350 mm x 430 mm x 220 mm
Weight                                                    Approx. 7 kg excluding paddles, accessories and energy supply
Accessory
Standard                                                  Accessory pouches included
Options
Fixation                                                  Vehicle wall charging station according to Norm EN1789
Ambient conditions
Protection class                                          IP55
Operating temperature                                     0 – 50 °C
Storage temperature                                        -20 – 70 °C
Rel. humidity                                             5 – 95 %, non-condensing
Art.Nr./No. 22922 | Release A01 | Release date: 12/2014
Right to technical and design changes reserved - Copyright by METRAX GmbH Rottweil / Technische und konstruktive Änderungen vorbehalten
Copyright by METRAX GmbH Rottweil