Gain Selection: X0.25, X0.
5, X1, X2 TEMP sensor: Standard configuration- skin
C30 Patient Monitor ECG TEMP sensor,
TM
Safety Standards Lead Type Cardio Tec 12-leads ECG Resolution: 0.1 C
ISO 13485:2003 approved, CE marking according to Analysis, 5 Lead and 3 lead Accuracy: 0.1 C (exclusive of error of
MDD93/42/EEC selectable sensor)
Physical Characteristics 12 lead cable I; II; III; aVR; aVL;aVF; V1-V6. Channel: T1, T2, TD (Temperature
Product Size: 190 mmX8 1.6m mX1 07m m 5 lead cable: RA; LA; RL; LL; V or R; L; N; F; C Difference)
Net Weight: 1.2kg 3 lead cable: RA; LA; LL or R; L; F Comen SPO2
Lead selection: 12-Lead I; II; III; aVR; aVL;aVF; Measurement & 1~100%
Operation Environment V1-V6. alarm range
Temperature: 0-40℃ 5-lead: I; II; III; aVR; aVL Resolution: 1%
Humidity: 15-85% 3-lead: I; II; III Accuracy: ±2% (70~100%,Ped, non-motion+
Power Supply DC 15V Gain Selection X0.25, X0.5, X1, X2, auto ±3% (70-100%, Neo,
External Power Supply: AC100-250V, 50/60Hz Sweep Speed 12.5, 25, 50mm/s non-motion);
Battery Type: Built-in Lithium-ion Heart Rate Range Adult: 15-300bpm 0-69% unspecified
Battery Capacity: 2600mA Pediatric/Neonate:15-350bpm Alarm range 0~100%
Battery Recharging Maximum 6.5 hours for charging; Resolution: 1 bpm Pulse rate: Range: 20~300bpm
Time: Accuracy: ±1% or ±1bpm (whichever is Resolution:1bpm
Battery working 4 hours for continuous working greater) Accuracy: ±3bpm(non-motion)
period: Filter ECG enlarge width: ±5bpm(motion)
Performance Specifications MON: 0.5-55Hz Alarm range:20~250bpm
Patient range: Adult, Pediatric, Neonatal DIA: 0.05-150Hz Nellcor SpO2
Display: 4.3” color TFT touch screen OPE:1-20Hz Measurement & 0~100%
Resolution: 480*272 ST SEGMENT -2.0mV - +2.0Mv (Automatic) alarm range
Trace 2 waveforms detection: Resolution: 1%
Sweep Speed: 12.5mm/s, 25mm/s, 50mm/s Arrhythmia Analysis: Yes Accuracy: ±2% (90-100%, MAX-A, MAX-AL,
Indicator: Dual alarm indicator lights pacemaker detection: detectable MAX-N, MAX-P, MAX-I and
Power indicator light
NIBP MAX-FAST sensors);
Battery indicator light
Method Automatic oscillometric ±3% (70-100%, D-YS, DS-100A,
QRS beep and alarm sound
Operating key sound Work mode: Manual / Automatic/Continuous OXI-A/N and OXI-P/I sensors);
Interface: Parameter module frame and Auto measurement: Adjustable(1-480min) 0-69% unspecified
interface
Measurement Unit: mmHg / Kpa selectable PR Measurement 20~300bpm
AC power input socket
Network interface Measurement types: Systolic, Diastolic, Mean Range:
MDR Range of systolic Adult Mode:40-270mmHg Resolution: 1bpm
Trend time: 48 hours trending
pressure: Pediatric Mode:40-200mmHg Accuracy:±3bpm
Alarm: User-adjustable High and Low
Neonate Mode 40-135mmHg Alarm range:20~250bpm
Limits ;
Range of diastolic Adult Mode:10-215mmHg Masimo SPO2
Prioritized audible and visual
alarm pressure: Pediatric Mode:10-150mmHg Measurement & 1~100%
Networking: Connected to Central Monitoring
Neonate Mode 10-100mmHg alarm range
System by hardwire/wireless
Range of mean Adult Mode:20-235mmHg Resolution: 1%
Connected to DVI
pressure: Pediatric Mode:20-165mmHg Accuracy: ±2% (70~100%,Ped, non-motion+
Respiration
Neonate Mode 20-110mmHg ±3% (70-100%, Neo,
Method: impedance method
Over-pressure both Hardware and software non-motion);
RR measurement Adult: 0-120bpm
protection: over pressure protection 0-69% unspecified
range: Neonate:0-150bpm
Accuracy: Less than 5mmHg Alarm range: 0~100%
Resolution: 1 rpm
Resolution 1mmHg Pulse rate: Range: 25~240bpm
RESP Apnea Alarm: 10s-40s
Temperature (Dual Resolution:1bpm
Alarm: Audible and visual alarm; alarm
Channel) Accuracy: ±3bpm(non-motion)
events recallable
Measurement range: 0-50 C ±5bpm(motion)
Sweep Speed: 12.5、25mm/s
Alarm range:25~240bpm
EtCO2(Sidestream)
CO2 Measurement 0 -150 mm Hg, 0 to 79%, 0 to
Range: 20kPa (at 760mmHg)
Accuracy: ± 2 mm Hg (0 – 40 mm Hg)
± 5% of reading (41 – 70 mm Hg)
± 8% of reading (71 –100 mm Hg)
± 10% of reading (101 –150 mm
Hg)
Respiration rate: 0-120bpm
Respiration rate ±2rpm (0-70rpm)
accuracy: ±5rpm (>70rpm)
Response time: <240msec (10% to 90%)
Delay time: <2s
EtCO2(mainstream)
CO2 Measurement 0 -150 mm Hg, 0 to 79%, 0 to 20kPa
Range: (at 760mmHg)
Accuracy: ± 2 mm Hg (0 – 40 mm Hg)
± 5% of reading (41 – 70 mm Hg)
± 8% of reading (71 –100 mm Hg)
± 10% of reading (101 –150 mm Hg)
Respiration rate: 0-120bpm
RR Accuracy: ±2rpm (0-70rpm)
±5rpm (>70rpm)
Delay time: <2s
Standard Configuration 1:
5-lead ECG,RESP, Dual-Temp, Comen SPO2,NIBP,Reserved
EtCO2 interface,
Standard Configuration 2:
12-lead ECG,RESP, Dual-Temp, Masimo SPO2,NIBP,Reserved
EtCO2 interface,
Optional:
EtCO2 module, Nellcor SPO2, packet, External battery