C50 Patient Monitor Specifications

This document provides the specifications for the Comen C50 patient monitor. The C50 offers 5-lead ECG, respiration, temperature, SpO2, NIBP, and heart rate monitoring. Optional configurations include dual IBP, EtCO2, AG, C.O., additional SpO2 and temperature channels, 3/12-lead ECG, and thermal recorder. The C50 has a 10.4" color touch screen, stores up to 6 hours of waveforms and trend data, and connects to central monitoring systems. Safety certifications include IEC 60601-1.

Uploaded by

Goutham Revuru
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
140 views6 pages

C50 Patient Monitor Specifications

This document provides the specifications for the Comen C50 patient monitor. The C50 offers 5-lead ECG, respiration, temperature, SpO2, NIBP, and heart rate monitoring. Optional configurations include dual IBP, EtCO2, AG, C.O., additional SpO2 and temperature channels, 3/12-lead ECG, and thermal recorder. The C50 has a 10.4" color touch screen, stores up to 6 hours of waveforms and trend data, and connects to central monitoring systems. Safety certifications include IEC 60601-1.

Uploaded by

Goutham Revuru
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 6

Specification: C 50

SHENZHEN COMEN MEDICAL INSTRUMENT CO.,LTD


Floor 10, Floor 11 and Section C of Floor 12 of Building 1A & Floor 1
to Floor 5 of Building 2, FIYTA Timepiece Building, Nanhuan Avenue,
Matian Sub-district, Guangming District, Shenzhen, Guangdong,
518106,P.R. China
Tel: +86-755-26408879
Fax: +86-755-26431232
Email: infoatszcomen.com
Web: en.comen.com

2021 V 1.3
Patient Monitor
C50

Standard Configuration: QRS beep and alarm sound


5-lead ECG, RESP, Temp(Single Channel), Operating key sound
Comen SpO2,NIBP, HR
Optional Configuration:
Dual-IBP, EtCO2, AG, C.O., Nellcor/Masimo SpO2, Dual-Temp, Interface:
3/12-lead ECG, Thermal Recorder, Suntech NIBP, Trolley, Parameter cable interface
Wall mount, Ground wire, AC power input
Safety Standards: USB port
IEC 60601-1 IEC 60601-1-8 IEC 60601-2-27 EN 1060-3 IEC 80601- : RJ45 port
2-30 IEC60601-2-34 IEC60601-2-49 ISO 80601-2-56 ISO 80601-2- Optional: VGA output
61 Multi-functional interface
Physical Characteristics: Data storage
Product Size: 291.7mm×146.5mm×250mm Alarm Event Recall: 200 groups
Weight: 3.3kg Wave Recall: 6 hours (8 waves)
IP grade IPX1 NIBP Recall: 2000 groups
Display: 10.4” color TFT touch screen Trend Graph: 160 hours
Resolution: 800*600 Trend Table: 160 hours
Trace: 8 waveforms Power-off Yes
Operation Environment: storage:
working Alarm: User-adjustable High and Low 3-level
Temperature 5-40°C Limits;
Humidity: ≤93% Prioritized audible and visual alarm
Power Supply 100-240V~, 50/60Hz±1Hz Network: Connected to Central Monitoring System
Battery Type: Rechargeable Lithium-ion battery by hardwire/wireless
Battery Capacity: 2200mAh (option: 4400mAh) Recorder:
Battery Recharging Type: Built-in; Thermal array
Time: Maximum 6 hours for charging; Channel: 3 channel waveforms
Battery backup: 2 hours for continuous working Speed: 25mm/s,50mm/s
Indicator: Record width: 50mm
One alarm indicator Real-time record
Power indicator time 8s, 16s, 32s or continual
Battery indicator Alarm record Yes
Respiration: ST SEGMENT
Method: RA-LL Impedance Method detection: -2.0mV~+2.0mV (Automatic)
RR measurement Arrhythmia
range: Adult: 0-120rpm Analysis: 26 types
Pediatric/Neonate:0-150rpm Pacemaker Detectable
Accuracy: 7~150rpm, ±2rpm or 2%, whichever is detection:
greater; Alarm: Yes, audible and visual alarm, alarm
0-6rpm unspecified events reviewable
Resolution: ±1 rpm 12 lead ECG Yes
RESP Apnea 10s-60s(Adu); 10s-40s (Ped/Neo) Analysis:
Accuracy: 5s NIBP:
Alarm: Audible and visual alarm; alarm events Method Automatic oscillation
reviewable Work mode: Manual / Automatic/Continual (5min, not
Sweep Speed: 6.25,12.5,25mm/s applicable to neonates)
Gain Selection: X0.25, X0.5, X1, X2, X4 Measurement
ECG: Time: Adjustable (1-480min)
Lead Type: CardioTecTM5-leads ECG Analysis, 12-Lead Measurement Unit:
and 3-leads selectable mmHg / kPa selectable
Lead selection 12-Lead I; II; III; aVR; aVL; aVF; V1-V6. Measurement Systolic, Diastolic, Mean
5-lead: I; II; III; aVR; aVL;aVF; V types:
3-lead: I; II; III Range of systolic
Waveform 5-lead: 2 -channel pressure: Adult Mode: 40-270mmHg
3-lead: 1 -channel Pediatric Mode: 40-200mmHg
Gain Selection: X0.125, X0.25, X0.5, X1, X2, X4, auto error Neonate Mode: 40-135mmHg
<±5% Range of diastolic
Sweep Speed: 6.25,12.5, 25, 50mm/s, error≤±10% pressure: Adult Mode: 10-215mmHg
Resp, lead Pediatric Mode: 10-150mmHg
disconnection Neonate Mode: 10-100mmHg
detection and AC waveform: Range of mean
active noise Current :<0.1µA; pressure: Adult Mode:20-235mmHg
control: Frequency 64kHz, ±10% Pediatric Mode:20-165mmHg
Heart Rate Neonate Mode 20-110mmHg
measurement Static pressure
Range: Adult: 15~300bpm range and accuracy:
Pediatric/Neonate:15~350bpm 0~300mmHg(0kPa~40.0kPa)
Accuracy: ±1% or ±1bpm (whichever is greater) ±3mmHg(±0.4kPa)
Protection: Withstand 4000VAC/50Hz voltage in Over-pressure
isolation, Again electrosurgical protection: Adult Mode: 297mmHg
interference and defibrillation Pediatric Mode: 240mmHg
Accuracy: ±1% or ±1bpm (whichever is greater) Neonate Mode: 147mmHg
Band width: Monitoring Mode: 0.5-40Hz Accuracy: ±3mmHg
Diagnosis mode: 0.05-150Hz Initial pressure
Surgery mode:1-20Hz range(mmHg): Adult: 80~240;Pediatric: 80~200;
ST mode: 0.05-40Hz Neonate:60~120
Alarm: Systolic, Diastolic, Mean Data averaging and
PR from NIBP: Measurement & alarm range: 40-240bpm other signal
Resolution: 1bpm processing time: 2s
Accuracy: ±3bpm or ±3%, whichever is Data refresh rate: 8s
greater PR Measurement
Nellcor SpO2: Range: 20--254bpm
Measurement Resolution: 1bpm
range 0-100% Accuracy: ±2bpm
Alarm range: 20-100% Alarm range: 20~254bpm
Resolution: 1% Perfusion index: 0.05%~20%
Accuracy: ±2% (70-100%, Adu/Ped, non-motion) Resolution: 0.01% (within 0.05%~9.99%
±3% (70-100%, Neo, non-motion) range) or 0.1% (within 10.0%~20.0%
1-69% unspecified range)
PR Measurement Temperature (Dual Channel)
Range: 20-300bpm Measurement &
Resolution: 1bpm alarm range: 0-50°C
Accuracy: ±3bpm (20-250bpm); Sensor: Skin/rectal TEMP sensor
Unspecified (251-300bpm) Resolution: 0.1°C
Alarm Range: 20~300bpm Accuracy: ±0.1°C (exclusive of error of sensor)
Masimo SpO2: Channel: T1, T2, TD (Temperature Difference)
Measurement & EtCO2 (Sidestream)
alarm range 1~100% Unit: mmHg, kPa
Resolution: 1% Measurement
Accuracy: ±2% (70-100%, Adu/Ped, non-motion) range: 0mmHg~150mmHg
±3% (70-100%, Neo, non-motion) Resolution 1mmHg or 0.1kPa or 0.1%
1-69% unspecified Accuracy 0mmHg ~40mmHg should be±2mmHg;
PR Measurement 41mmHg ~70mmHg should
Range 25~240bpm be±5%×reading;
Resolution: 1bpm 71mmHg ~100mmHg should
Accuracy: ±3bpm (non-motion) be±8%×reading;
±5bpm (motion); 101mmHg~150mmHg should
Alarm range: 25~240bpm be±10%×reading
Perfusion index: 0.02~20% Oxygen
Resolution: 0.01% (within 0.02%~9.99% compensation 0~100 mmHg
range) or 0.1% (within 10.0%~20.0% Equilibrium gas Helium, room air, nitrous oxide
range) IBP
Comen SpO2: Channel: 4 Channels
Measurement Measured Pressure:
& alarm range 0~100% ART, PA, CVP, RAP, LAP, ICP, LV, AO, UAP,
Resolution: 1% BAP, FAP, UVP, IAP, P1, P2, P3, P4
Accuracy: ±2% (70~100%, Adu/Ped, non-motion) Measurement
±3% (70-100%, Neo, non-motion) range: ART: 0~300mmHg
Unspecified (1-69%) PA: -6~120 mmHg
CVP: -10~40mmHg
RAP: -10~40mmHg AG resolution: CO2: 1mmHg
LAP: -10~40mmHg awRR: 1rpm
ICP: -10~40mmHg Accuracy: For all measured values complies with EN
LV: 0~300mmHg ISO 21647:2004 and EN 864:1996
AO: 0~300mmHg Alarm: EtCO2: 0mmHg~190mmHg
UAP: 0~300mmHg Fi CO2: 0mmHg~190mmHg
BAP: 0~300mmHg AwRR: 2mmHg~150mmHg
FAP: 0~300mmHg EtO2: 18% ~ 100%
UVP: -10~ 40mmHg FiO2: 18% ~ 100%
IAP: -10~40mmHg EtN2O: 0% ~ 100%
P1, P2, P3, P4: -50~300mmHg FiN2O: 0% ~ 82%
Accuracy: ±2% or ±1mmHg (whichever is greater) EtHal/EtEnf/EtIso/EtSev/EtDes: 0% ~ 25%
0.1kPa or 1mmHg FiHal/FiEnf/FiIso/FiSev/FiDes: 0% ~ 25%
Resolution: (-50mmHg~300mmHg) Others: Up to 4 waveforms displayed
Alarm Range: -50mmHg~300mmHg MAC value displayed
Pressure sensor: Sensitivity: 5 V/V/mmHg ISATM (AG) Sidestream Gas Analyzer
Impedance range: 300~3000Ω Method: Infrared gas measurement
PR from IBP: Measurement & alarm range: No Breaths
20bpm~350bpm Timeout range Adult: 10s, 15s, 20s, 25s, 30s, 35s,
Resolution: 1bpm 40s ,45s, 50s, 55s or 1min;
Accuracy: ±1bpm or ±1%, whichever is Pediatric and neonate: 20s, 25s, 30s, 35s
greater or 40s
Accuracy: ±5s
AG No Breath Alm
AG (complies with ISO 80601-2-55) Delay: 10s, 15s, 20s, 25s, 30s, 35s, 40s, 45s, 50s,
Method: Infrared Radiation Absorption 55s, 1min or Off
Characteristics Working
AG preheating time conditions: ISA AX+: 0~50°C (32~122°F); ISA OR+:
<20s 5~50°C (41~122°F)
Gas sorts: CO2, N2O, DES, ISO, ENF, SEV, HAL, O2 Storage conditions -40~70°C (-40~158°F)
(optional paramagnetic sensor) RH <4kPa H2O (non-condensing) 95% RH,
Measurement 30°C
range: CO2: 0~15%: ±(0.2kPa+reading×2%), Barometric
15~25%: unspecified pressure 52.5~120kPa (4572m)
N2O: 0~100 %: ±(2kPa+reading×2%) Water treatment Sampling tube: patented dehydration
HAL, ISO, ENF: 0~8%: ± tube
(0.15%+reading×5%); 8~25vol%: Data output: Fi and Et values
unspecified Waveform: Display up to 4 gas concentration
SEV: 0~10%: ± (0.15%+reading×5%); waveforms at a time
10~25vol %: unspecified Diagnostic
DES: 0~22%: ± (0.15%+reading×5%); parameter: Barometric pressure
22~25%: unspecified ISA sensor: 2~9-channel NDIR gas analyzer
O2: 0-100%: ± (1%+reading×2%) (measurement range: 4~10µm)
Data output: Fi and Et values Compensation: CO2 broadening effect
Calibration No calibration is required. The Monitor reported, even if it is lower than 0.15
will auto perform zeroing when powered vol%
on and perform auto zeroing every 24h Cardiac Output (C.O.)
(ISACO2) or 8h (ISA AX+/OR+) Method: Thermodilution
subsequently. Measurement
Preheating time ISA CO2: <10s; ISA OR+/AX+: < 20s Range: C.O.: 0.1~20L/min
Rise time CO2: ≤250ms; N2O: ≤ 350ms; BT: 25~43°C
anesthetic gases: ≤ 350ms; O2: ≤ IT: 0~25°C
450ms Resolution: C.O.: 0.1L/min
Overall system BT, IT: 0.1°C
response time <3s (2m sampling tube) Accuracy: C.O.: ±5% or ±0.1 L/min,whichever is
Respiration greater
detection Self-adaptive threshold (minimum CO2 BT, IT: ±0.1°C (exclusive of sensor)
concentration change: 1 vol%) Alarm Range: BT Hi limit: (LO limit +0.4)-43.0°C
RR 0~150 breaths/min BT Lo limit:25.0~(Hi limit-0.4) °C
Anesthetic gas Step: 0.1°C
threshold Threshold of main anesthetic gases (ISA
OR+/AX+): 0.15 vol%. The concentration *Notice: Specifications subject to changes without
of any identified anesthetic gas will be prior notice. All rights reserved by Comen