The F5 Multi-Parameter Monitor features a 12.1-inch color TFT LED screen capable of displaying multiple waveforms and supports various monitoring functions for adult, pediatric, and neonatal patients. It includes advanced features such as automatic patient ID generation, real-time arrhythmia analysis, and online software upgrades, along with a robust alarm system. The device is designed for versatility in clinical settings, offering multiple display modes and extensive connectivity options.
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
9 views4 pages

Brosur f5

The F5 Multi-Parameter Monitor features a 12.1-inch color TFT LED screen capable of displaying multiple waveforms and supports various monitoring functions for adult, pediatric, and neonatal patients. It includes advanced features such as automatic patient ID generation, real-time arrhythmia analysis, and online software upgrades, along with a robust alarm system. The device is designed for versatility in clinical settings, offering multiple display modes and extensive connectivity options.
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 4

F5 Multi-Parameter Monitor

(12.1 inches)

Features
♦ 12.1' color TFT LED screen supports display 9~13 waveforms, support HDMI output
♦ Support 7 channel ECG waveform display simultaneously
♦ Newly advanced ARM A8 microcontroller, calculating speed is 4 times faster than traditional products
♦ Generate ID automatically when register a new patient and complete patient information management. Support medical
history search by patient ID, name and mobile number
♦ Support data control / input by mouse / keypad
♦ Three-level acousto-optic alarm and support alarm review.
♦ Multi-display mode to meet different clinical requirement, standard interface, big font, OxyCRG, trend graph, NIBP review and
full leads ECG interface.
♦ 21 types of Arrhythmia analysis and real-time S-T segment analysis and pacemaker detection
♦ Suitable for monitoring adult, pediatric and neonatal patient.
♦ Support online software upgrading by net/USB
♦ Support wired/ wireless CMS based on TCP/IP (built-in WIFI module)

Standard Configuration:
♦ Anti-glare screen, Large capacity rechargeable battery, 3/5-Lead ECG, Respiration, NIBP, SpO2, Pulse Rate, Temperature
F5 Multi-parameter Monitor Specification

Upright Handle, Easy to carry Control Panel & Color LED Display Transducer Socket
(7 hotkeys & Japan SHARP LED) (Switzerland LEMO socket)

Interface

ECG Full Lead Standard Display

Big Font Display Oxy CRG

Trend Table NIBP Review


F5 Multi-parameter Monitor Specification
Dimension and Weight - Resolution: 1 bpm

- Dimension: 310mm*290mm*150mm - CMRR: Diagnostic Mode >90 dB


- Weight: 3.4 Kg, excluding accessories Monitor Mode >105 dB

Power Supply Surgery Mode >105 dB


- Voltage: AC100~240V, 50/60HZ, Power≤60W - Input impedance: ≥5.0 Mohm

Display - ECG signal range: ±6 mV


- 12.1" color TFT LED - Leakage Current: < 10 μA
- Resolution: 800*600 pixels - Baseline Recovery: < 5 s After defibrillation (MOR or OPS mode)

Battery - Bandwidth
- Type: Rechargeable lithium battery 14.8V, 2200mAH Surgery 1 ~ 25 Hz
- Charge Cycle: 500 times Monitor 0.5 ~ 40 Hz
- Working time: 2 hours (larger battery for 4-5 hrs) Diagnostic 0.5 ~ 130 Hz

Recorder (Option) HR
- Method: Thermal printer - Range Adult: 15 ~ 300 bpm,
- Paper width: 50 mm Neo/ Ped: 15~ 350 bpm
- Paper speed: 25/50 mm/s - Resolution: 1 bpm
- Trace: Max. 3 tracks - Accuracy: ± 1% or ± 1bpm, which greater
- Recording way: Real-time Recording, Review printing ST Segment
Alarm
- Range -2.0 ~ +2.0 mV
- Level: Low, medium and high
- indication: Auditory and visual - Accuracy -0.8mV~+0.8mV: ±0.02mV or ±10%, which is greater

Input device - Other range: unspecified


- Touch screen (option) Respiration
- Knob (Standard) - Method: Impedance between RA-LL, RA-LA
- Keypad input(Standard) - Gain: ×0.25,×0.50,×1,×2
- Mouse/ Keyboard input (option) - Respiration Rate: Adult: 0 ~ 120 BrPM

System Output & Extensible Interface Neonatal / Pediatric: 0 ~ 150 BrPM


- Ethernet Network: 1* RJ45 - Resolution: 1 BrPM
- 1* USB port - Accuracy: ±2 BrPM or ±2%, whichever is greater
- Video Output: 1*HDMI port (option) - Apnea Alarm: 10 ~ 40 s

Trend & Reviewing: NIBP


- Trend: 1440 hours - Method: Oscillometric
- Waveform review: 2 hours - Measure mode Manual, Auto, STAT
- NIBP measurement reviewing: 1000 groups - Measure Interval in AUTO Mode 1~ 480 min
- Alarm event: 2000 groups - Measure Range: Adult: SYS: 30 ~ 270 mmHg

Operating Environment DIA: 10 ~ 220 mmHg


- Temperature: 5 ℃~ 40℃ MEAN: 20 ~ 235 mmHg
- Humidity: 15% ~ 80% (non-condensing) Pediatric: SYS: 40 ~ 235 mmHg
- Atmosphere pressure: 86 KPa ~ 110 KPa DIA: 10 ~ 220 mmHg

Transportation and Storage Environment MEAN: 20 ~ 225 mmHg


- Temperature: -20℃~50℃ Neonate: SYS: 30 ~ 135 mmHg
- Humidity: 10%~90% (non-condensing) DIA: 10 ~ 110 mmHg
- Atmosphere pressure: 86 KPa ~ 110 KPa MEAN: 20 ~ 125 mmHg

Safety: - Static pressure accuracy: ±3mmHg


- IEC60601-1 Approved, CE marking according to MDD93/42/EEC - Resolution: 1mmHg
- With reference to RoSH Directive 2011/65/EU recasting - Accuracy: Maximum Mean deviation ±5mmHg

Performance: Maximum Standard deviation ±8mmHg


ECG - - Dual-overpressure Protection for hardware and software

- Lead mode: 3/5 Leads, I, II, III or I, II, III, AVR, AVL, AVF, V TEMP
12 Leads (option), I, II, III, AVR, AVL, AVF, V1~V6 - Channel: Dual-channel, provide T1; T2; △T1
- ECG waveform: 3/7 channels, 12 channel for option - Measuring and Alarm Range: 0 ~ 50 ℃
- Gain: × 2.5mm/mV (× 0.25), × 5.0mm/mV (× 0.5), × 10mm/mV (× 1), - Resolution: 0.1℃
×20mm/mV (× 2), Auto - Accuracy: ± 0.1℃

- Sweep speed: 12.5 mm/s, 25 mm/s, 50 mm/s SpO₂


- ECG waveform: 3 channels - Measurement and Alarm Range: 0 ~ 100 %
- Accuracy: ±1bpm or ±1%, whichever is greater - Resolution: 1%
- Accuracy: 70% ~ 100%, ±2 % IBP (Option)
0% ~ 69%, unspecified - Max Channel: 2
- Support Pitch tone - Label ART, PA, CVP, RAP, LAP, ICP, P1, P2
PR - Measurement way: Thermal resistance way
- Measuring and Alarm Range: 0~250 bpm - Press Sensor: Sensitivity 5 uV/V/mmHg, ±2%
- Resolution: 1 bpm Impedance 300-3000Ω
- Accuracy: ± 2bpm - Resolution: 1 mmHg

Nellcor-SpO₂ (option) - Unit: mmKg, kPa


- Measurement Range: 0 ~ 100 % - Measuring and alarm range:
- Resolution: 1 % ART 0 ~ 300 mmHg
- Accuracy: 70% ~ 100%, ±2 % (adult) PA -6 ~ 120 mmHg
70% ~ 100%, ±3 % (Neonate) CVP/RAP/LAP/ICP -10 ~ 40 mmHg
70% ~ 100%, ±2 % (Low Perfusion) P1/P2 -10 ~ 300 mmHg
0% ~ 69%, unspecified - Actualization interval: about 1 Sec.

PR Depth of Anesthesia (BIS) (Option)


- Measurement range: 25 ~ 300 - EEG sensitivity: +/- 475μV
- Resolution: 1bpm - Noise: < 2uVp-p (1-150 Hz)
- Accuracy: 20 ~ 250 bpm, ±3 bpm - CMRR: > 100 dB

Masimo SpO2 (option) - Input impedance: > 50 Mohm


- Measurement Range: 0 ~ 100 % - Sample rate: 1024 samples/s, 16 bits
- Resolution: 1 % - Index and display update: 0-100, filter 1-47 Hz, 1 sec. display update
- Accuracy: 70% ~ 100%, ±2 % (adult/ pediatric, non-motion conditions) - EEG waveform: ± 500 μV scaleable, filter 1–30 Hz, 3 sec. window length.
70% ~ 100%, ±3 % (neonate, non-motion conditions) - ESR: 0-100%
70% ~ 100%, ±3 % (motion conditions) - EMG 0-100 arbitrary unit: Curve and bar
0% ~ 69% unspecified - Index trend: 5 min and 60 min

PR - Alarms: IoC high and low user selectable limits


- Measurement range: 25 ~ 240 - Artifact rejection: Automatic: user can disable.
- Resolution: 1bpm - Sensor impedance range: 0 - 14 kohm
- Accuracy: ±3 bpm (non-motion condition) Multi-gas/O2 (AG) (Option)
±5 bpm (motion conditions) - Method: Infrared absorption

EtCO₂ ( Mainstream/ Sidestream) (Option) - Gas sorts: CO2, N2O, Des, Iso, Enf, Sevo,
- Measure method: Infrared spectrum Hal, O2 (Optional paramagnetic sensor)
- Measure Range: 0 - 19.7% (0 - 150 mmHg) - Measurement range: CO2: 0~30%, N2O : 0~105%
- Resolution: 0.1 mmHg O2, :0~105%, Enf, Iso, Hal: 0~30%,
- CO2 Accuracy: 0 ~ 40 mmHg, ±2 mmHg Sevo: 0~30%, Des:0~30%
41 ~ 70 mmHg, ±5% of reading - Data output: Fi and ET values
71 ~ 100 mmHg, ±8% of reading - Respiration rate: ±1BrPM (2~60BrPM)
101~ 150 mmHg, ±10% of reading 61~100BrPM (unspecified)
at 760 mmHg, ambient temperature of 35℃) - Other: Up to 4 waveforms displayed
- Breath Rate: 3 - 150 bpm Agent mixture detection

MAC value displayed


* Specifications subject to change without prior notice.

Configuration

3/5 12 EtCO2 Nellcor/ Central


Feature SpO2/ Dual Dual Depth of Multi-gas Printer Wall-mount Touch
Lead NIBP TEMP RESP Lead Main-stream/ MASIMO HDMI Monitoring
PR TEMP IBP Anesthesia (AG) 50mm /Trolley Screen
ECG ECG Side-stream/ SpO2 Station

F5 √ √ √ √ √ Opt Opt Opt Opt Opt Opt Opt Opt Opt Opt Opt Opt

Shenzhen Unicare Electronic Co., Ltd.


Add: 3D, Building4, Tian-an Industrial Zone, Nan You, Distributor:
Nanshan District, Shenzhen, 518054, P.R. China
Tel: +86-(0)755-3336 9088 Fax: +86-(0)755-3336 6818
Email: market@unicare-med.com www.unicare-med.com