Simple, Accurate, and Reliable
Single Parameter Analysis
ChemScan mini Fe Analyzer
The ChemScan mini Fe analyzer provides operators with reliable        ACCESSORIES
process chemistry measurements. This reduces the need for frequent
manual sampling or laboratory analysis while producing the best
                                                                                              Sample Extraction Accessory
water quality.
                                                                                              Provides a continuous flow of fresh
                                                                                              sample to the ChemScan mini
APPLICATIONS
                                                                                              analyzer. Designed to reject algae
• Analysis of Iron in potable water, wastewater and industrial
                                                                                              and other larger solids.
  processes
FEATURES
• Robust design for demanding operating environments
                                                                                              ChemScan Cartridge Filter Wand
• Blockage resistant internal sample tubing
                                                                                              For high-solids applications.
• No filtration required on samples with low solids
                                                                                              No pressurized air, water or
• Minimal replacement parts for low maintenance
                                                                                              chemicals required for cleaning.
• Sample Blank eliminates electrical/optical drift
• Simple field adjustable calibration
• Separate enclosures for electronic and sample handling
                                                                                              ChemScan mini Outdoor Enclosure
• LED Light source for 10+ years design life
                                                                                              A turnkey solution for mounting the
• Self-Cleaning to eliminate internal fouling
                                                                                              ChemScan analyzer and related
• Separate external sample line cleaning available
                                                                                              items.
• Full range of sampling accessories available for all applications
BENEFITS
                                                                                              Submersible Pump
• Assure process conformance
                                                                                              Provides a continuous flow of
• Control energy and chemical costs
                                                                                              fresh sample to sample extraction
• Confirm plant compliance in real-time
                                                                                              accessory.
• Improve process performance
• Low reagent and maintenance costs
                                                                                              Deck-Mounted Self-Priming Pump
                                                                                              Provides a continuous flow of
                                                                                              fresh sample to sample extraction
                                                                                              accessory. (when submersible pump
                                                                                              is not applicable)
                                                                         Discuss with your ChemScan representative the most suitable
www.In-Situ.com                                                          accessories for your application.
                                                                                                                                     Mini Fe Technical Specifications¹
                                                                                                                                                                                                Revised 6/15/2021
FUNCTIONS AND OUTPUTS                                                                                           PERFORMANCE SPECIFICATIONS²
ANALYZER OPERATION              Automated, Continuous Analysis of Water and Wastewater                          READING INTERVAL       8 to 5999 minutes
MEASUREMENT PRINCIPLE           Reagent-Assisted Optical Absorbance with sample zero correction
                                                                                                                DEFAULT READ           16 minutes
                                                                                                                INTERVAL
NUMBER OF PARAMETERS            One
                                                                                                                RESPONSE TIME          8 minutes minimum
PARAMETER OPTION                Iron
                                                                                                                ACCURACY               2% of value or 2x detection limit (whichever is greater) Per EPA SP 846
                                                                                                                                       (The detection limit is the low concentration stated in ranges below)
DATA COMMUNICATIONS             4-20 mA (2 outputs)
                                                                                                                PRECISION              Less than 0.5% of Range
DATA LOG                        Time Date, Date, Concentration, Diagnostic Info, 5,000 events
                                                                                                                ZERO DRIFT             Less than 0.5% of Range
NUMBER OF SAMPLE LINES          One                                                                             RANGES                 Method 1039 0.01 - 5.0 mg/L
                                                                                                                                       Method 1037 0.02-20.0 mg/L
REAGENT ADDITION                YES, Direct Reagent Injection
                                                                                                                INSTRUMENT SPECIFICATIONS
AUTO MAINTENANCE                Auto Clean
                                                                                                                SIZE                   26” tall x 9.5” wide x 7” deep (66 cm tall x 24 cm wide x 18 cm deep)
CALIBRATION                     Factory calibrated for reagent response, field adjustable
                                                                                                                WEIGHT                 27 lbs (12.25 kg)
SAMPLE PARAMETERS
                                                                                                                FINISH COATING         Fiberglass Reinforced Polyester (FRP)
SAMPLE PRESSURE                 Pressurized sample line required regulated to 2-10 psi (15-70 kPa), (sample     MATERIAL
                                conditioning and pressurizing accessories available)
                                                                                                                POWER                  120-240 VAC ±10%, 50-60 Hz, 70 VA
SAMPLE FLOW                     0.5 to 1.0 l/min. 1 L Flush Per Sample (0.13 to 0.26 GPM - 0.26 Gallon Flush)
                                                                                                                POWER CONNECTION       120 VAC US cord / plug set (Standard) (conduit connection optional)
FILTRATION REQUIREMENT          For samples with more than 150 mgl TSS ( filter required for WW influent and
                                primary effluent)                                                                POWER CONDITION        Dedicated branch circuit free from: surges/dips > 10%, RF and
                                                                                                                                       switching noise
STRAINER REQUIREMENT            #20 Mesh - Opening of 0.69 mm (0.027 inches) Provided
                                                                                                                OPERATOR INTERFACE     2 x 20 LCD and 4 x 4 Keypad
SAMPLE TEMPERATURE              50-140°F (10° - 60°C)
                                                                                                                SAMPLE CONNECTION      ¼” FNPT Fitting
SAMPLE TURBIDITY                60NTU or 150mg/l Suspended Solids
                                                                                                                WASTE CONNECTION       6 foot length of 5/8” ID clear vinyl tube provided (route to open drain)
OPERATING ENVIRONMENT
                                                                                                                MOUNTING               Wall (Standard)
ENCLOSURE RATINGS               Upper Enclosure: NEMA 4X Fiberglass Reinforced Polyester, Acrylic window
                                Lower Enclosure NEMA 4X Fiberglass Reinforced Polyester                         MAINTENANCE
AMBIENT TEMPERATURE             41 – 113°F (5° - 45°C)                                                          REAGENT                3 months at default read interval
                                                                                                                REPLACEMENT
RELATIVE HUMIDITY               0 - 100% (Non-Condensing)
                                                                                                                CLEANING SOLUTIONS     As required (3 months typical)
INSTALLATION                    Indoor or Sheltered (from rain and sun) Location
                                                                                                                REFILL
                                                                                                                PERISTALTIC MIXING     Replace after six months of operation
                                                                                                                PUMP HEAD
                                                                                                                PERISTALTIC MIXING     Replace after twelve months of operation
                                                                                                                PUMP FULL ASSEMBLY
                                                                                                                PERISTALTIC ZEROING/   Replace after two years of operation
                                                                                                                CLEANING PUMP
                                                                                                                HEAD
Notes:
1. Technical Specifications are subject to change without prior notice.
2. All performance specifications are based on analysis of drinking water standards
   under factory conditions
www.In-Situ.com