OTS Kalos Collimator RALCO 302
OTS Kalos Collimator RALCO 302
Confidential Information
10.05.2007 N
LEGAL WARNING
RALCO SRL WILL PURSUE ANY AND ALL LEGAL REMEDIES AVAILABLE UNDER
THE LAW TO RECOVER DAMAGES INCURRED BASED UPON THE ABOVE
VIOLATION(S).
Model Certifications
Revision Level:
The English version of this manual is a translation of the Italian text which
prevails in case of doubts.
Rev. N Date N
MRM/024/07
A 31.01.2007
MRM/025/07
B 06.09.2007 MRD/012/07
C 12.06.2008 MRE/015/08
D 06.11.2008 MRD/015/08
E 22.03.2010 MRD/003/10
F 23.09.2010 MRD/011/10
G 21.04.2011 MRD/005/11
14.07.2011 MRD/013/11
06.09.2011 MRE/042/11
H
26.09.2011 MRM/251/11
24.10.2011 MRD/018/11
I 09.03.2012 MRD/006/12
J 14.11.2012 MRD/012/12
K 12.03.2013 MRM/031/13
L 25.09.2013 MRD/017/13
16.01.2014 MRD/001/14
29.01.2014 MRE/006/14
M 10.09.2014 ACD/020/14
10.09.2014 ACD/026/14
13.11.2014 MRE/041/14
20.01.2017 MRD/001/17
N
14.09.2017 MRD/034/17
THIS INSTRUCTION MANUAL PROVIDES THE SPECIFICATIONS, DIMENSIONS, AND FUNCTIONS FOR A STANDARD
COLLIMATOR. PERSONALIZATIONS ARE AVAILABLE UPON REQUEST. WITH PERSONALIZED COLLIMATORS, THE CUSTOMER
MUST ENSURE THE FOLLOWING HAS BEEN PROVIDED:
• YOUR SPECIFIC CODE IS CLEARLY VISIBLE ON THE BACK LABEL AFTER “CUSTOMIZATION”, ABOVE THE SERIAL
NUMBER, TO ENSURE THE CORRECT PERSONALIZED COLLIMATOR HAS BEEN PROVIDED (SEE ABOVE).
• THE CHAPTER ENTITLED “PERSONALIZATIONS,” WHICH INCLUDES RELEVANT INFORMATION REGARDING YOUR
PERSONALIZATION.
• OTHER PROVIDED ATTACHMENTS WHICH REFER TO YOUR PERSONALIZATIONS.
CONTENTS
MTR302 F/A 1
MAINTENANCE
CLEANING RECOMMENDATIONS .......................73
DISINFECTION .................................................73
RECOMMENDED MAINTENANCE PROGRAM .......73
GENERAL
WARRANTY .....................................................75
REPAIRS .........................................................75
END OF LIFE DISPOSAL ...................................76
DISASSEMBLY .................................................76
TRANSPORT AND STORAGE .............................76
SAFETY/RESPONSIBILITY .................................77
RESIDUAL RISKS .............................................77
MTR302 F/A
2
Ralco srl Modello/Model R302 F/A
ADVISORY
THE FOLLOWING INSTRUCTION MANUAL MUST BE READ AND UNDERSTOOD IN ITS ENTIRETY BY THE INSTALLER
AND OPERATOR.
THE COLLIMATOR ELECTRONICS SUPPLIES A SIGNAL DEFINED AS READY. THE X-RAY EQUIPMENT MANUFACTURER
WILL MANAGE THIS SIGNAL AS THE X-RAY CONSENT. IN THE EVENT OF A COLLIMATOR FAILURE, THE LIABILITY
FOR THE CORRECT BYPASS OF THIS SIGNAL, IN ORDER TO ALWAYS ENSURE THE X-RAY CONSENT, DEVOLVES
UPON THE X-RAY EQUIPMENT MANUFACTURER.
REPORTING
INFORMATION REGARDING ACCIDENTS THAT HAVE OCCURRED WHILE USING THIS COLLIMATOR MUST BE REPORTED
IMMEDIATELY TO RALCO, SRL.
LIMITATION OF LIABILITY
RALCO IS NOT LIABLE IF THE PROVIDED INSTRUCTIONS ARE NOT COMPLIED WITH. FURTHERMORE, RALCO
IS NOT LIABLE IF ONE OR SEVERAL OF THE FOLLOWING INSTANCES APPLY:
R302 F/A - Advisory
IF THE UNIT IS SPECIFICALLY DESIGNED PER CLIENT SPECIFICATIONS AND THE CERTIFICATION WAS THE DUTY
OF THE CLIENT;
THE COLLIMATOR DESCRIBED HERE, IS USED ON RADIOLOGICAL SYSTEMS AND IS CLASSIFIED AS A TYPE IIB
ACCORDING TO ATTACHMENT IX. THE COLLIMATOR HAS BEEN DESIGNED AND MANUFACTURED IN COMPLIANCE
TO ATTACHMENT II OF LEGISLATIVE DECREE, 2 FEBRUARY 1997, N. 46, IMPLEMENTATION OF DIRECTIVE
93/42/CEE OF 14 JUNE 1993 AND SUCCESSIVE MODIFICATIONS APPLYING DIRECTIVE 2007/47/CE.
LIST APPLICABLE STANDARDS AND TO WHICH RALCO ADHERES TO:
EN 60601-1:2006
EN 60601-1-2:2007
EN 60601-1-3:2008
EN 60601-2-45:2012
EN 60601-2-54:2009
EN 60825-1:2007
ISO 9001:2008
UNI ESO 13485:2012
21 CFR SUBCHAPTER J (FDA)
8750 01 & 8750 81 (CSA)
CE
*RALCO CAN PROVIDE DOCUMENTATION REGARDING ITS ADHERENCE TO ANY OF THE ABOVE STANDARDS.
EN 60601-1 par. 5
• PROTECTION AGAINST ELECTRIC HAZARDS: "CLASS I" EQUIPMENT.
• PROTECTION AGAINST DIRECT AND INDIRECT CONTACTS: TYPE B EQUIPMENT WITH APPLIED PARTS.
• PROTECTION AGAINST WATER SEE PAGE: "COMMON EQUIPMENT".
• SAFETY OF OPERATION IN THE PRESENCE OF INFLAMMABLE ANAESTHETICS WITH AIR OR OXYGEN OR NITROUS
OXIDE: EQUIPMENT NOT SUITED TO APPLICATION IN THE PRESENCE OF INFLAMMABLE ANAESTHETIC MIXTURES
CONTAINING AIR OR OXYGEN OR NITROUS OXIDE.
• OPERATION CONDITIONS: EQUIPMENT FOR CONTINUOUS OPERATION AT INTERMITTENT LOADS - SEE CHAPTER
OPERATION INSTRUCTIONS.
• SHOULD LABEL DATA ON THE COLLIMATOR NOT CORRESPOND TO THE SPECIFICATIONS HEREIN, INFORM RALCO OF
THE NON CONFORMITY.
• VERIFICATIONS OF THE SPECIFICATIONS ARE TO BE PERFORMED ACCORDING TO THE INDICATED
EQUIPMENT STANDARDS.
R302 F/A - Advisory
THE INHERENT RISK OF USING COLLIMATORS IN RADIOLOGICAL SYSTEMS IS DEEMED REASONABLE AS DETERMINED
BY APPLICABLE STANDARDS. THE USE OF COLLIMATORS FOR RADIOLOGICAL SYSTEMS IS STRICTLY REGULATED.
RALCO FOLLOWS ALL APPLICABLE STANDARDS. IT IS UP TO THE INSTALLER AND OPERATOR TO ENSURE THAT
ALL POSSIBLE STEPS ARE FOLLOWED TO ENSURE THE HEALTH AND SAFETY OF THE PATIENT AND OPERATOR.
THE CONTACT WITH THE MOVING PARTS CAN BE CONSIDERED IMPOSSIBLE AND MAY OCCUR ONLY DURING
INSTALLATION OR MAINTENANCE (LOW HAZARD). THE INTERNAL PARTS ARE PROTECTED BY A MECHANICAL SAFEGUARD
IN THE FORM OF A METAL OR PLASTIC SHEATH WHICH IMPEDES ACCESS. ALSO MOTOR UNITS ARE NOT ACCESSIBLE
AS KNOBS ARE FIXED USING THE ALLEN SCREWS AND CAN BE REMOVED ONLY BY USING A TOOL, IN ACCORDANCE
WITH PARAGRAPH 5.92.3 OF EN 60601-1 AND EN 60601-2-54 PAR. 201.9.2.2.5. IN ADDITION THE COLLIMATOR
HAS BEEN TESTED IN ACCORDANCE WITH EN 60601-1, PARAGRAPH 4.8.2
SYMBOLS
IN THIS MANUAL THE FOLLOWING SYMBOLS ARE USED:
THE TRIANGLE EMPHASIZES WARNINGS AND CAUTIONARY MESSAGES WHICH ARE IMPORTANT FOR YOUR SAFETY
AND/OR EFFICIENT OPERATION OF THE EQUIPMENT. THESE WARNINGS MUST BE STRICTLY COMPLIED WITH.
INFORMATION PROVIDED WITH THE LAMP IS ADDITIONAL ADVICE FOR THE PROPER CONTROL AND PRACTICAL
USE OF THE UNIT.
RALCO INSTRUCTION MANUALS ARE AVAILABLE IN TWO VERSIONS: ITALIAN AND ENGLISH.
THE ENGLISH VERSION OF THIS MANUAL IS A TRANSLATION OF THE ITALIAN TEXT WHICH PREVAILS IN CASE OF DOUBTS.
E-MAIL: RALCO@RALCO.IT
SYMBOLS
417-iec
Alternating Current 01-14
503
417-iec
Direct Current 01-18
5031
417-iec
Continuous and Alternating Current 01-19
5033
417-iec
Protective Ground 01-20
5019
417-iec
Plus; Positive Polarity 01-27
5019
417-iec
Minus; Negative Polarity 01-28
5006
417-iec
Input 01-36
5006
417-iec
Output 01-37
5034
417-iec
Iris Diaphragm: Closed 01-70
5324
417-iec
Radiation Filter or Filtration 04-51
5381
R302 F/A - Symbols
417-iec
Light Indicator of Radiation Field 04-51
5381
417-iec
Beam Limiting Device: Open 04-55
5385
417-iec
Beam Limiting Device: Closed 04-56
5386
Electrostatic-Sensitive Device
DESCRIPTION
Multilayer, square field, manual collimation system intended for installation on
stationary X-ray equipment. This device has been designed and manufactured for
skeletal and thoracic investigations.
The X-ray field is defined by six pairs of shutters, four of which are lead-lined. The six pairs of
shutters move perpendicularly within the X-ray field. Two pairs of brass shutters are located
near the focus, two are located near the entrance window and two are located near the exit
window of the X-ray beam from the collimator. The latter shutters serve to accurately define the
X-ray field edges.
Shutter movements are manual, controlled by two knobs on the collimator front panel.
CHARACTERISTICS
Mounting Plane at 80 mm (3.14") from the focus.
Continuous Film Coverage from Min: 00 x 00 cm to Max: 48 x 48 cm at 100 cm (40") SID.
Maximum Radiation Leakage: 150 kVp - 4 mA.
High Luminosity provided by a quartz iodide lamp simulating the X-ray field.
Minimum Inherent Filtration: 2 mm aluminium equivalent.
Extra Variable Filtration consists of three filters and an open space for no filtration on an
aluminium disk, 1 mm thickness, with setting:
- no filtration
- 1 mm Al equivalent + 0.1 Cu
- 1 mm Al equivalent + 0.2 mm Cu
- 2 mm Al equivalent
GC 338 timer limiting projection lamp exposure time to 30 seconds thus extending lamp life
by preventing overheating.
This collimator may have the following optional items; a detailed description is provided in the
chapter OPTIONAL ITEMS.
RO DESCRIPTION
Final control testing (light intensity, light field/radiation field alignment, radiation
RO 063
leakage, light field border contrast ratio, general funcitonality and electronic)
RO 240 Focus-skin spacers, preventing X-ray exposure too close to the patient
Single laser aligning the collimator to the Potter Bucky with power supply
RO 242/1
included
RO DESCRIPTION
RO 271 Halogen lamp 150 W for high luminosity with cooling system
Auto-centering top plate with plastic flange allowing +/- 45° rotation of the
RO 318
collimator, the plastic flange is included
Plastic flange with a notch rotating +/- 50°, 20 mm thickness without stop
RO 441
(available only for collimators assembled with RO 318)
Plastic flange with a notch rotating +/- 90°, 20 mm thickness without stop
RO 442
(available only for collimators assembled with RO 318)
SPECIFICATIONS
Specifications of the basic version of the collimator are described here below.
24 V DC - 50/60 Hz - 6.5 A
Power Supply
Fuse for power supply protection collimator. Not
T8A
supplied by Ralco.
Motor Supplied by Collimator Board n.a.
Fuse for Power Supply Protection Motor. Not supplied
n.a.
by Ralco.
EN 60601-2-54
Accuracy of X-ray field as shown on the front display
n.a.
versus actual X-ray field dimension.
X-ray Field Indication Precision
Settings on an Index Scale < 2 % SID
EN 60601-2-54
SID: (optional)
Precision of Measurement with Retractable Tape < 2% SID
EN 60601-2-54
Maximum Radiation Leakage
Measured at 100 cm with X-ray Beam = 150 kVp - 4 mA < 40 mRh
EN 60601-1-3
Potentiometers
n.a.
Square Field
n.a.
Round Field
Maximum Load for Accessory Guides and Dimensions
for Accessory
GUIDES FOR ACCESSORIES (IF ASSEMBLED): PRIOR TO
INSERTING ACCESSORIES IN THE GUIDES CHECK ON THE
PERFECT ATTACHMENT OF THE SUPPORT WITH THE
MOUNTING SLOT ON THE COLLIMATOR (TOLERANCE MAX.
± 0,5 MM). A FAULTY ATTACHMENT COULD BE DANGER-
OUS AND IT COULD CAUSE THE FALL OF ACCESSORIES.
ATTENTION: Static load: 70 N (about 7.1 Kg)
INTERAXLE SPACING VARIES DEPENDING ON THE COLLIMA-
TOR MODEL:
Dynamic load: 15 Nm (approx.
R 72 - N.A. 3.06 Kg)
R 72 S - 154 MM (+/- 0.5 MM)
R 103 - 133 MM (+/- 0.5 MM)
R 104 - 147 MM (+/- 0.5 MM)
R 105 - 147 MM (+/- 0.5 MM)
R 104/A, 108, R 108 F, R 221 - 152 MM (+/- 0.5 MM)
R 225, R 302 - 177 MM (+/- 0.5 MM)
SEE THE CHAPTER OPTIONAL ITEMS IF THE GUIDES
ARE PROVIDED ON REQUEST.
Operation Environment
10°- 40°C
• Ambient Temperature
0%- 75%
• Relative Humidity
700 - 1060 hPa
• Atmospheric Pressure
Storage
-40° - 70°C
• Temperature
10% - 95%
• Humidity
500 - 1060 hPa
• Atmospheric Pressure
Weight 9.4 Kg
L: 268
Dimensions W: 195.5
H: 206.5
R302 F/A - Specifications
INSTALLATION
X-RAY TUBE COMPATIBILITY
• Using the dimensions in Fig. Tube Compatibility below, ensure the near port shutters can
be inserted in to the X-ray tube port.
• Check that the tube housing literature indicates congruent minimum inherent filtration (1
mm) and, that maximum radiation leakage is 30 mR/hour measured at 1 m from the source
when operating at its leakage technique factors (150 kVp - 4 mA).
• Source values (tube housing-collimator) must not be less than 3 mm Al for filtration and
must never exceed 100 mR/hr for radiation leakage [21 CFR sub-chapter J, part 1020.30
(m) (1)].
• The distance between the X-ray tube focus and the flange mounting plane (collimator
upper plate) must be: 80 mm (3.14"), tolerance +/- (0.04”).
• Use the X-ray tube housing literature to determine the distance from the focal spot to the
tube port face.
• Subtract the resulting distance from 80 mm (3.14") and determine the number of 1.5 mm
(0.06") spacers that, combined with the thickness of the mounting flange, will make up the
difference: (20 mm for the fixed flange; 18 mm for the rotating flange). Allowable tolerance
R302 F/A - Installation
is 1 mm. (0.04").
• Different flange models with their dimensions are listed in the Chapter DESCRIPTION,
paragraph Optional Items.
• Mount the flange and spacers to the tube port with 4 countersunk bolts and 4 countersunk
washers, the bolts must be long enough to be driven into the tube port face for at least 5
threads.
IMPORTANT
TO SAFEGUARD THE OPERATOR AND PATIENT AGAINST THE HAZARD OF
A FALLING COLLIMATOR, THE FOLLOWING INDICATIONS MUST BE RESPECTED.
• Unscrew the four mounting and centering adjustment Allen screws on the collimator until
the four tabs are withdrawn from the collimator top opening, see Fig. Mounting Flange, in
this Chapter.
MAKE SURE TO TIGHTEN THE ALLEN SCREWS SECURING THE CONTROL TABS.
APPROPRIATE TIGHTENING OF THE 4 ALLEN SCREWS ENSURES SECURE MOUNTING
OF THE COLLIMATOR. TIGHTENING FORCE USED MUST BE GREATER THAN 0.50 NM.
• Check to see that the distance between the collimator housing and the mounting flange is
equal in all directions and that the collimator face is parallel to the axis of the table.
• Loosen the screws and adjust if required.
R302 F/A - Installation
ELECTRICAL CONNECTION
THE WIRING DIAGRAM INCLUDED IN THIS CHAPTER REFERS TO THE STANDARD PRODUCT.
IT IS THE RESPONSIBILITY OF THE CUSTOMER WHO HAS REQUESTED AN ELECTRIC
CUSTOMIZATION, TO ENSURE THAT AN ELECTRIC DIAGRAM RELATING TO THE CUSTOMIZATION
HAS BEEN PROVIDED WITH THE DOCUMENTATION.
THE DEVICE MUST BE EXCLUSIVELY CONNECTED TO POWER NETWORK, WITH EARTH PROTEC-
TION, IN ORDER TO AVOID A RISK OF ELECTRICAL SHOCK.
Supply and signals to the collimator must be to 2007/47/CE standards. Devices that supply the
collimator must therefore feature double or reinforced insulation as provided by the General
R302 F/A - Installation
The device is electrostatic sensitive, consequently all the relating safety standards must
be complied with.
CABLES AND TERMINALS USED FOR THE INTERNAL CONNECTION OF THE COLLIMATOR MUST
BE SUITABLE FOR OPERATION AT TEMPERATURES OF 70°C AND COLLIMATOR CURRENT
ABSORPTION.
INCORRECT POWER SUPPLY COULD DAMAGE THE ELECTRONIC BOARDS AND/OR THE
LIGHT SOURCE.
SUPPLY MAY BE EITHER IN ALTERNATE OR DIRECT CURRENT.
MAKE CERTAIN THAT POLARITY IS RESPECTED.
R302 F/A - Installation
WIRING DIAGRAM
R302 F/A - Installation
GC 338 Timer
Connectors
1. Lamp output (max 9 A)
2. Push button input
3. Power Supply +12/24 V DC o 12-24 V AC/V DC A
J1 POWER SUPPLY AND OUTPUTS
4. Shared Lamp
5. Power Supply –12/24 V DC o 12-24 V AC/V DC B
6. Shared push button
1. +Laser Power Supply (5 V)
J2 LASER POWER SUPPLY 5VDC 2. GND Laser
1. V pp
2. +5 V
3. GND
J3 PROGRAMING CONNECTOR
4. PGD
5. PGC
6. NC
1. +Fan Power Supply (max 200 mA)
J4 – FAN POWER SUPPLY 12V DC
2. – Fan Power Supply
JUMPERS OFF - no jumper
ON - jumper
JP1, JP2 – LAMP TIME SELECTION JP1 JP2
OFF - 30 s OFF - 30 s
ON - 45 s OFF - 45 s
OFF - 60 s ON - 60 s
ON - 120 s ON - 120 s
JP3, JP4 – OPERATION MODE JP3 JP4
SELECTION
OFF - Timer OFF - Timer
ON - Timer reset by OFF - Timer reset by pressing
pressing the button the button
OFF - Power Supply ON - Power Supply
ON - Fan ON for 90 ON - Fan ON for 90 seconds
seconds longer than the longer than the lamp
lamp
R302 F/A - Installation
OPERATION INSTRUCTIONS
LIGHT/X-RAY FIELD SETTING
• Set the collimator at 100 cm SID.
• By rotating the knobs and with the knob index on the front panel, follow the tags of the
scale.
• Do not force the knobs.
• Activate the light field by pressing the related push-button on the collimator front panel.
• The collimator is ready to operate.
WARNING:
PROLONGED LIGHTING WITHOUT ALLOWING LIGHT SOURCE TO COOL CAUSES THE
COLLIMATOR TO OVERHEAT IN THE AREA NEAR THE LIGHT SOURCE.
FOR EMERGENCIES: MAXIMUM LIMIT ADIVSED IS 5 SUCCESSIVE LIGHT SOURCE OPER-
ATIONS. ALLOW THE COLLIMATOR TO COOL (ABOUT 10 MIN).
The X-ray field is set through the knobs that control the shutter movement. The table provides
the readout of the values which must be set with the knobs. The exact value to be set is
obtained by crossing the SID in use (the value indicated by vertical axis), with the cassette size
value in cm or inches (horizontal axis).
E.g.
To obtain the X-ray field 24 cm (LONG) x 35 cm (CROSS) at SID 150 cm, with reference to
the chart, set the Long knob indicator to 16.5 cm and the Cross knob indicator to 24 cm. See
Fig. Long Aperture and Fig. Cross Aperture.
SID in use
CALIBRATION
IF THE DISTANCE OF 100 CM (40") SID CANNOT BE DETERMINED, USE THE SID
VALUE CLOSEST TO 100 CM (40") AND CALCULATE THE MEASUREMENT TOLERANCES
AS THE APPROPRIATE PERCENTAGES OF THE DISTANCE.
IF THE X-RAY BEAM CANNOT BE POSITIONED VERTICALLY USE CLAMPS, MASKING TAPE,
OR OTHER MATERIAL AS REQUIRED TO POSITION THE X-RAY TUBE AND IMAGE RECEPTOR.
THE X-RAY TUBE AND THE IMAGE RECEPTOR MUST BE SET OUT AT THE SPECIFIED
SID DISTANCE PERPENDICULARLY TO THE X-RAY TUBE AS DESCRIBED IN THE FOLLOWING
PROCEDURE.
NOTE
• For required values is referred to the System Reference Manual. However, the values L1 +
L2 and W1 + W2 shall NEVER exceed 2 cm (representing 2% of the SID).
• Note the measured values in Chart below.
• Repeat preceding steps with small focus selected.
THE HEEL EFFECT WILL CAUSE THE FIELD TOWARD THE CATHODE TO BE SLIGHTLY
LESS SHARP THAN ON THE OTHER THREE SIDES. THIS IS NORMAL AND CANNOT BE
CORRECTED BY ADJUSTMENT. IN ADDITION, AN X-RAY TUBE OF 12° OR LESS TARGET
ANGLE WILL PRODUCE AN ASYMMETRICALLY SHAPED FIELD WHEN A LARGE FIELD SIZE
IS USED AT SHORT SID, BECAUSE OF ANODE CUT-OFF EFFECT. THIS IS NORMAL AND
MAY NOT BE CORRECTED BY ADJUSTMENT.
• Remove the part of the cover necessary to access the screws, see Chapter - COVER
REMOVAL
• Remove the light source protection heatsink by unscrewing the fixing screws . This allows
you to access the light source.
• If the light-field needs to be moved laterally, loosen (not remove) the fixing screws A.
• Adjust through screw B.
• When calibration is terminated, lock the screws A.
See the following Fig. Light Field Adjustment
Vertical Alignment
• Remove the part of the cover necessary to access the screws, see Chapter - COVER
REMOVAL.
• If adjustment is required loosen the two screws C holding the light support.
• If the light-field is smaller than the X-ray field, move away the light source by adjusting
screw D.
• If the light-field is bigger than the X-ray field, move the light source closer by adjusting
screws D.
• Tighten the two screws C. See Fig. Light Field Adjustment
COMPLIANCE VERIFICATION
The above HVL requirements can be met if it is demonstrated that the aluminium equivalent in
the primary beam is not less than that shown in the following table:
1
Dental X-ray Systems designed for use with intraoral image receptors and manufactured after
December 1, 1980.
2
Dental X-ray Systems designed for use with intraoral image receptors and manufactured before
or on December 1, 1980, and all other X-ray systems subjected to this section and manufac-
tured before June 10, 2006.
3
All X-ray systems, except dental X-ray systems designed for use with intraoral image receptors
subjected to this section and manufactured on or after June 10, 2006.
The information contained in the above table was extracted from the Code of Federal
Regulations FDA 21 1020.30 (m).
Type 100 Aluminium Alloy (as given in “ALUMINUM STANDARDS AND DATA” verification of
compliance).
Below 50 0.5
From 51 to 70 1.5
Over 70 2.2
The Aluminium equivalence of each component in the primary beam (X-ray tube and housing,
beam limiting device and any additional filtration in the system) is specified on the component,
in the technical data attached to the component or can be measured. Determine the total
aluminium equivalence in the primary beam and make sure that it is equal or greater than those
specified in the above Table Total Filtration of Primary Beam in Aluminium Equivalence.
Using the type 1100 Aluminium Alloy, tape a total of 2.5 mm of Aluminium to the window of
the collimator. Make an exposure using the same technique factors; record the reading.
c) Verify that the radiation read with the 2.5 mm Al in the beam is greater or equal to 50% of
the radiation read with no filtration in the beam.
Standard Absorber Method
The HVL determination obtained from the following procedures are to be compared with those
illustrated in the Table - Minimum Filtration Requirement - Beam Quality (HVL). The HVL
in millimetres of aluminium obtained during the test must be greater or equal than the values
listed in the above mentioned table.
a) Direct the central X-ray beam perpendicular and in the center of a RAD-Check instrument.
Determine the exact distance from the X-ray tube focal spot to the window of the collimator
(273 mm - 10.75"). Place the input area of the RAD-CHECK at an equal distance from the
collimator window. Collimate the beam to an area slightly larger than the detector.
b) Select a tube potential of 100 kVp and appropriate mA and seconds, with no added filtration
in the beam make an exposure and record the reading. Using a set of several sheets of 1100
Aluminium Alloy, each having a thickness of 0.5 or 1.0 mm, tape the filtration to the window
of the collimator. Make an exposure for each increments of filtration and record the reading.
c) Plot the exposure readings (log scale) versus the total added filtration thickness on semi-log
paper; see the sample hereunder.
d) Verify that HLV values in the useful beam for the above specific tube potential is not less
than the values shown in Table - Minimum Filtration Requirement - Beam Quality (HVL).
generale\12_verif_conf_graph.gif
CROSSHAIR ALIGNMENT
CHAPTER- ADJUSTMENTS, PARAGRAPH CROSSHAIR ALIGNMENT.
b) Place the Focus of the X-Ray tube at 100 cm. from the table top were the light field as been
projected. Open the collimator's shutters to assure that each quadrant of the light field is
larger than the measuring area of the photometer.
c) Check that the voltage specified by the manufacturer is applied to the lamp, make certain
that all surfaces in the light beam are clean and unobstructed.
d) Place a photometer capable of reading up to 160 lux in the centre of each of the four
quadrants of the light field.
e) Turn on the light beam and read the light intensity, subtract to it the ambient lighting,
previously determined.
f) Verify that the average illumination is higher than 160 lux.
g) Verify that the contrast ratio is performed between two points:
• The first point at 3 mm outside the edge of the light field.
• The second point at 3 mm inside the edge of the light field.
These measurements are to be performed with the probe of the lux metre set at 1 mm
aperture.
h) Record the measured values including all data regarding the instrument and voltage
employed.
EMC COMPLIANCE
Electromagnetic Emissions
The R302 F/A colimator is suitable for use in the specified electromagnetic environment. The purchaser or user of the R302
F/A should assure that it is used in an electromagnetic environment as described below:
RF emissions Class B
CISPR 11
This R302 F/A is suitable for use in all establishments, including
Harmonic emissions Class A
domestic establishments and those directly connected to the
IEC 61000-3-2
public low-voltage power supply network that supplies buildings
Voltage fluctuations/flicker Complies used for domestic purposes.
emissions
IEC 61000-3-3
R302 F/A - Compliance Verification
Electromagnetic Immunity
The R302 F/A collimator is intended for use in the electromagnetic environment specified below. The customer or the user
of the collimator should assure that it is used in such an environment.
Collimator R302 F/A is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled.
The customer or the user of the collimator can help prevent electromagnetic interference by maintaining a minimum distance
between portable and mobile RF communications equipment (transmitters) and the collimator as recommended below,
according to the maximum output power of the communications equipment.
100 12 12 23
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in metres (m) can
be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of
the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection
from structures, objects and people.
R302 F/A - Compliance Verification
COVER REMOVAL
(1).
(2).
(3).
(4).
DISCONNECT THE
CONNECTOR.
(5).
IN COLLIMATORS WITH
MEASURING TAPES, REMOVE
THE TAPE STOP BY LIFTING IT
OFF WITH A SCREW DRIVER.
GENTLY EASE THE TAPE INTO
ITS CONTAINER WITHIN THE
COLLIMATOR.
(6).
(7).
(8).
(9).
ACCESS THE CLUTCHES BY REMOVING THE TWO KNOBS AND/OR THE FRONT PANEL DEPENDING
ON THE COLLIMATOR MODEL.
Timer Board
ACCESS THE TIMER BOARD BY REMOVING THE TWO KNOBS AND/OR THE FRONT PANEL
DEPENDING ON THE COLLIMATOR MODEL.
R302 F/A - Cover Removal
Power Supply
Longitudinal Movement
ADJUSTMENTS
WARNING:
THIS IS PROVIDED AS AN AID TO THE END USER. RALCO IS NOT
LIABLE FOR ANY DAMAGES RESULTING FROM THE ALTERING OF PRESET
FACTORY CONFIGURATIONS.
• If a shutter control is too tight, loosen one of the two screws A on the U bolt that frictions
the shaft, see Fig. Friction Brake.
• Before replacing the knobs, close both shutters and mount/fix the knobs so that the field
index is set at “0” (completely closed), see paragraph Shutter Dial Adjustment in this
Chapter.
MECHANICAL MOTION STOP ADJUSTMENT
The purpose of these stops is to prevent stressing the control shafts of the shutters at the two
limits of travel. This is done by limiting rotation of the control shafts.
• Adjustment of these stops will not be required unless:
• The shutters can not be completely closed or opened to the largest size.
• Torsion is felt at one or both ends of the travel.
• The shutters themselves appear to limit the motion rather than the stops.
TROUBLESHOOTING
A faulty Collimator must not be used until it is repaired and checked.
The use of a faulty collimator might impair the safety of the operator and patient.
Before returning the collimator to Ralco for repair, please make sure that the fault is not caused
by one of the problems listed below. If the indications provided fail to solve your fault, please
make sure that you obtain a Return number (RMA) from Ralco for the collimator, see
Chapter - GENERAL, paragraph Repairs.
Light Source
Problem Cause Solution
Collimator supplied Check: Supply/Tension/Current/
incorrectly Polarity/Fuses
Check item
Faulty light source Replace if necessary, see
Chapter- SUBSTITUTIONS
The light source fails to
switch ON Check timer supply
Faulty timer Substitute if necessary, see
Chapter- SUBSTITUTIONS
Centering
Problem Cause Solution
The light field of the The mirror or light See Chapter- CALIBRATION,
collimator is not source are not paragraph Light Field to X-ray Field
centered positioned correctly Alignment
Field Size
Problem Cause Solution
Faulty field size Knobs are not Regulate, see Chapter-
indication regulated ADJUSTMENTS
Shutters
R302 F/A - Troubleshooting
SUBSTITUTIONS
SUBSTITUTION OF THE HALOGEN LAMP
WARNING: DO NOT TOUCH THE LAMP, THE SOCKET, OR THE LAMP BRACKET
WITH YOUR FINGER. THEY CAN BE VERY HOT AND CAUSE SEVERE BURNS.
WARNING: DO NOT TOUCH THE LAMP WITH YOUR FINGERS, EVEN WHEN
IT IS COLD.OIL FROM YOUR SKIN WILL CAUSE THE LAMP TO CRACK AND
POSSIBLY EXPLODE. IF YOU HAVE TOUCHED THE LAMP, WIPE THE SURFACE
WITH ALCOHOL, THEN HANDLE THE LAMP WITH A PIECE OF PAPER.
• Install the new timer by proceeding in a reverse order; pay particular attention to the
connection of the cables on the 4 or 6-way electric terminal depending in the model.
Fig. 1 Fig. 2
• Unscrew the 2 screws A holding the filter system on the front plate, see Fig. 3 and Fig. 4.
• Slide the filter system out from the collimator body.
• Substitute with an identical filter group.
• Check that disc rotation is uniform.
• Insert the filter system into the collimator body.
• Tighten the 2 screws A.
• Remount in reverse order.
R302 F/A - Substitutions
Fig. 3 Fig. 4
R302 F/A - Substitutions
SPARE PARTS
THE SPARE PARTS LISTED BELOW REFER TO THE STANDARD MODEL ONLY.
LABEL
LABEL 1A
LABEL 7
LABEL 6
R302 F/A - Spare Parts
Parts Breakdown
R302 F/A - Spare Parts
OPTIONAL ITEMS
This collimator may have the following optional items; a detailed description is provided in this
chapter.
• Impossibility of retracting the tape into its lodging because the grip of the
spring has been forced and hook-up is consequently distorted.
NOTE: DO NOT EXTEND THE TAPE BEYOND THE INDICATION OF STOP.
THE HEEL EFFECT WILL CAUSE THE FIELD TOWARD THE CATHODE TO BE SLIGHTLY LESS SHARP
THAN ON THE OTHER THREE SIDES. THIS IS NORMAL AND CANNOT BE CORRECTED BY
ADJUSTMENT. IN ADDITION, AN X-RAY TUBE OF 12° OR LESS TARGET ANGLE WILL PRODUCE AN
ASYMMETRICALLY SHAPED FIELD WHEN A LARGE FIELD SIZE IS USED AT SHORT SID, BECAUSE OF
ANODE CUT-OFF EFFECT. THIS IS NORMAL AND MAY NOT BE CORRECTED BY ADJUSTMENT.
R302 F/A - Optional Items
RO 082 Mirror
Glass mirror, 0.8 mm thickness, inherent filtration 1 mm Al equivalent. With this mirror, the
collimator equivalent filtration value of 2 mm Al is reduced to 1 mm Al.
R302 F/A - Optional Items
NOTE
RALCO RECOMMENDS FOLLOWING THE BELOW PROCEDURE, DURING THE INSTALLATION OF THE COLLIMATOR
EQUIPPED WITH THE SIEMENS TOP COVER TO THE X-RAY TUBE. THE CORRECTLY PERFORMED PROCEDURE
ALLOWS THE COLLIMATOR TO ROTATE FREELY.
PROCEDURE:
1. REMOVE THE COLLIMATOR FROM ITS PACKAGING AND PLACE IT ON A FLAT SURFACE,
2. UNSCREW ALL 3 FIXING SCREWS OF THE COLLIMATOR, SEE FIG. 1,
3. INSTALL THE COLLIMATOR ONTO THE FLANGE PREVIOUSLY MOUNTED TO
THE X-RAY TUBE.
4. TIGHTEN ALL 3 SCREWS COMPLETELY, SO THAT THE TABS WILL TOUCH
THE FLANGE SLIGHTLY,
5. ONCE THE SCREWS ARE COMPLETELY TIGHTENED LOOSEN ONE BY ONE ABOUT
1/2 TURN (180°) TO ALLOW THE COLLIMATOR TO ROTATE FREELY. SEE FIG. 2.
Fig. 1 Fig. 2
R302 F/A - Optional Items
Laser Adjustment
• Remove part of the cover to access the
point of adjustment, see Chapter- COVER
REMOVAL.
• The line is to fall on a perpendicular
cross-line on the plastic anti-dust panel
near the collimator controls, see Fig.
Laser Line.
• Adjust the position of the line by rotating
or moving the base of the laser system.
• To rotate the laser system, loosen
the Allen screw A, see Fig. Laser
Adjustment.
Fig. Laser Adjustment
• Tighten the Allen screw when the
laser beam falls on or is parallel to
the bisector line drawn on the anti-
dust panel.
• Shift the laser system by loosening the
two B screws holding the laser system
base to the beam limiting device front
plate.
• Move the base until the laser beam
falls over the perpendicular
bisector line on the anti-dust panel,
see Fig. Laser Alignment.
• Tighten the two B screws.
DO NOT APPLY EXCESSIVE Fig. Laser Alignment
FORCE TO THE SCREW. THE
LASER SHELL IS IN PLASTIC AND
EXCESSIVE PRESSURE COULD
CRACK THE PLASTIC AND POSSI-
BLY SHORT-CIRCUIT THE LASER.
R302 F/A - Optional Items
Substitution
• Disconnect supply.
• Remove the cover, see Chapter- COVER
REMOVAL.
• Unscrew the fixing Allen screws A, see
Fig. Laser Adjustment.
• Disconnect the timer cables from the terminal board - white 0 V, red 5 V.
• Remove the laser and substitute with an identical item.
• Tighten the screws.
• Check the laser alignment, see Chapter- ADJUSTMENTS.
• Remount the cover, see Chapter- COVER REMOVAL.
Classification EN 60825-1 par. 1 -5: Class 2 laser product => Red laser line.
CALIBRATION CONTROL
• Draw a cross (two lines at 90°) on a
sheet of paper and use this as reference.
• Switch the light ON and set the sheet of
paper at 1 m from the focus.
• Make sure the cross on the paper
coincides exactly with the two lines silk-
screened on the plastic window.
DO NOT APPLY EXCESSIVE FORCE TO THE SCREW. THE LASER SHELL IS IN PLASTIC AND EXCESSIVE
PRESSURE COULD CRACK THE PLASTIC AND POSSIBLY SHORT-CIRCUIT THE LASER.
SUBSTITUTION
Horizontal
• Disconnect supply.
• Remove the laser cover located on the collimator cover by unscrewing the two screws A,
see Fig. Horizontal Laser.
• Disconnect the timer cables from the terminal board - white 0 V, red 5 V.
• Remove the laser and substitute with an identical item.
• Tighten the screws.
• Check the laser alignment, see paragraph ADJUSTMENT.
• Remount the laser cover.
Vertical
• Disconnect supply.
• Remove the knobs and front panel, see Chapter- COVER REMOVAL.
• Loosen the Allen screw A, see Fig. Vertical Laser on the laser support.
• Disconnect the timer cables from the terminal board - white 0 V, red 5 V.
R302 F/A - Optional Items
Classification EN 60825-1 par. 1 -5: Class 2 laser product => Red laser line.
For the collimator to rotate to position "Ø", two of the four tabs retract to allow to rotation. Adjust
and tighten the four Allen screws as follows:
screws 1 and 3 = 7 turns until the tongue touches slightly the flange;
screws 2 and 4 = 5½ turns exactly;
See Fig. Mounting Flange in this chapter. Please Note: If Allen screws "2" and "4" are
tightened more than 5 ½ turns, the collimator will not rotate.
The collimator control tabs conform to EN60601-1. Secure the tabs along the edge of the
mounting flange but do not lock them. Excessive tightening of the tabs prevents collimator
rotation.
Check that the distance between the collimator housing and the mounting flange is equal in all
directions and, that the collimator face is parallel to the table axis.
Loosen the screws and adjust if required.
Mounting Flange
R302 F/A - Optional Items
DO NOT TOUCH THE DISSIPATER WITH YOUR HANDS; IT COULD BE HOT AND
CAUSE SEVERE BURNS.
GC-LED-5A Board
CONNECTORS
J2 - SUPPLY AND OUTPUTS 1. 20-30 V AC or 12-45 V DC
2. 20-30 V AC or 0 V DC
3. +5 V Fan
4. GND Fan
5. +5 V Laser
6. GND Laser
7. + Power-up/Power down external push-button
8. GND Power-up external push-button
J1 - LED OUTPUT + LED 5 A positive output
- GND Led 5A
J3 - PROGRAMMING 1. Vpp
CONNECTORS 2. +5 V
3. GND
4. PGD
5. PGC
6. NC
J4 - HOME SENSOR INPUT + 1. + External LED
EXTERNAL LED 2. - External LED
3. - Photosensor LED Cathode
4. + Photosensor LED Anode
5. Input signal (Photosensor output collector)
6. GND (Photosensor output emitter)
R302 F/A - Optional Items
JP3 - FAN POWER SUPPLY OFF: fan is operated by micro controller according to
settings made with LED
ON: fan is always supplied
JP4 - LASER POWER SUPPLY OFF: Laser is micro controller operated
ON: Laser is always supplied
JP5 - CONFIGURATION OF PUSH- OFF: LED and laser are controlled by only push button (on
BUTTON NUMBER OUTPUT J2)
ON: LED is controlled by standard push button (on J2),
laser is controlled independently by push button on JP6.
JP6 - BUTTON INPUT FOR LASER ON 1. Laser ON button pin 1
TIME. 2. Laser OFF button pin 2
JP8 - CONFIGURATION OF LED ON Laser is controlled independently by the timer, this
TIME configuration is applicable in the same manner for both
push buttons.
OFF: output status is changed by pressing button (if the
light is OFF will be ON and vice versa).
ON: the timer operates in the time renew mode. When the
button is pressed the time is renewed without the LED OFF
(if the LED is ON it will stay ON but with the time reset). The
same is applicable for the laser if is independent.
FAN The fan switches ON together with the LED, and operates
15 seconds longer after the LED OFF.
Trimmers
TR1- CURRENT CONTROL Adjust the output current intensity to the LED.
Step-less current adjustment from 0 to 5.1 A. Adjusting
current to approximately 4.7 A (+/- 5%) will increase LED
life without a significant loss of luminosity.
TR2 - TIME ADJUSTMENT Adjusts the time from 30s to 120s up to 90% of run,
maximum light ON time is 15 min. for safety reasons.
When the timer is regulated at over 120s (i.e. 100% of its
adjustment), the timer performs as power supply: it starts
by powering-up the LED and other features such as laser
and fan. In this case, maximum light ON time is 15 min. for
safety reasons.
LED SUBSTITUTION
• Disconnect supply.
R302 F/A - Optional Items
THE FRONT PANEL SHOWS THE 2 INDEX SCALES THAT ALLOWS THE COLLIMATOR TO
OPERATE EITHER IN THE STANDARD OR STITCHING MODE. SEE FIG. INDEX SCALE -
STANDARD MODE AND FIG. INDEX SCALE - STITCHING MODE.
THE MAXIMUM COLLIMATOR APERTURE IN THE STITCHING MODE IS 43 X 43 CM AT 100 CM SID.
R302 F/A - Optional Items
Field 2
Cross left knob - aperture 35 cm
Long central knob - position 2
Long right knob - position 2
See Fig. Field 2.
Fig. Field 2
Field 3
Cross left knob - aperture 35 cm
Long central knob - position 3
Long right knob - position 3
See Fig. Field 3.
R302 F/A - Optional Items
Fig. Field 3
MAINTENANCE
To ensure constantly safe performance of the collimator and its compliance with applicable
regulations, a maintenance program is indispensable.
It is the Owner's responsibility to supply or arrange for this service.
CLEANING RECOMMENDATIONS
• The collimator housing must be cleaned as prescribed by the sanitary regulations followed
by the operator.
• Disconnect supply.
• Use non abrasive cleaning products.
Care must be taken to prevent liquid from entering the collimator. Please Note: The
collimator cover is not watertight.
• Do not reapply power if inflammable liquids have leaked into the collimator.
See the following Maintenance Instructions.
• Clean the varnished and aluminium surfaces with a damp cloth only, using a neutral
cleansing agent then dry the surfaces with a soft cloth.
• Clean chrome surfaces with a dry soft cloth.
Do not spray water or detergent directly over the collimator. The unit’s liquid
protection level is IPx0.
DISINFECTION
The disinfection method use must conform with the currently applicable norms and directives
covering disinfection and protection against explosion hazards.
.
Never use caustic substances, solvents or abrasive detergents.
If products that could form explosive gas mixtures are used, allow the gas to
evaporate before starting the system.
• Disconnect supply.
• Disinfect the unit including accessories and cables with a dampened cloth.
• Do not spray the unit with the disinfectant because it could leak into the collimator.
field.
• Calibration procedures must be performed as described in this manual.
• Check once a week that the screws and tabs which serve to secure the collimator to the
flange/tube adapter are correctly tightened.
• Remove the covers and panels from collimator. Inspect the moving parts for signs of wear
or damage.
• Check the electric system and substitute parts that show wear.
MAKE SURE TO TIGHTEN THEALLEN SCREWS SECURING THE COLLIMATOR OR THE CONTROL TABS.
APPROPRIATE TIGHTENING OF THE 4 ALLEN SCREWS ENSURES SECURE MOUNTING OF THE
COLLIMATOR. TIGHTENING FORCE USED MUST NOT EXCEED 0.50 NM.
GENERAL
WARRANTY
This product has been manufactured and tested to the highest quality standards by Ralco, srl.
Ralco undertakes to replace and repair any collimator during a period of 24 months for
mechanical and 12 months for electrical parts (motors, potentiometers, electrical boards,
lasers) from the date of invoice (shipment date from Ralco).
The warranty applies provided the product has been handled properly in accordance with its
operating instructions and its intended use.
Warranty covers cost of all components and labor involved, unless:
• Product documents have been altered in any way or made illegible;
• The model or production number on the product has been altered, deleted, removed or
made illegible;
• Repairs or product modifications and alterations have been performed by unauthorized and
unqualified persons;
• Unauthorized repairs and/or modifications have been performed;
• Damage caused by misuse or neglect, incorrect installation or accidental damage;
• Damage occurred during transit due to shipping company, or incorrect packing by
customer;
• Unoriginal spare parts and accessories have been used.
In-warranty spares will be available only upon return to Ralco, at the customer's expense, of
the parts considered to be faulty to allow Ralco to assess the cause of the fault.
Components Not Covered by this Warranty:
• Consumable items such as lamp bulbs, lexan panels and mirrors;
• Cosmetic damage such as scratches;
• Any missing components when product arrives for repair.
REPAIRS
In the event the customer finds any Non Conformity in the product, please contact Ralco via e-
mail at the address: repairs@ralco.it (Ralco Repair Assistance).
To successfully resolve any Non Conformity, the following information must be provided:
• The model and serial number of the collimator found on the label;
• A detailed description of the problem (in Italian or English);
• Whether you want a repair, refurbishment, or model upgrade of the product.
Ralco and the customer will work together to resolve the problem by either providing
instructions, and/or field service or by sending replacement parts.
In the event the collimator needs to be repaired at Ralco, you will be provided with a RMA
(Return Merchandise Authorization) number as your consent to the return. If a product is
returned to Ralco without a RMA or without a description of the problem, the customer will incur
a €100 processing fee.
If the product is returned under warranty and no defect is found, the customer incurs the cost
of evaluation, testing and shipping.
R302 F/A - General
The shipping of the unit is at the customer’s expense if the unit is out of warranty.
The warranty period resets only on replaced parts once repairs are completed (shipment date
from Ralco back to customer), see paragraph, WARRANTY.
Ralco reserves the right to decide if the product is to be repaired or substituted.
SAFETY/RESPONSIBILITY
Ralco adheres to the directives governing manufacturers of electro-medical equipment:
Directive 2007/47/CE para.10 -Legislative Decree n° 46 para.10
Ralco shall not be held responsible when instructions provided in the present manual are not
complied with. Ralco shall not be held responsible if the collimator relates to one or several of
the following instances:
• The unit is of Ralco construction, built to client specifications with no CE marking.
• The unit has been modified by the OEM or end user.
• The unit has been installed without respecting the instructions provided in this manual.
• The unit is used without respecting the instructions provided in this manual.
• The unit has not been subject to routine functional inspection.
• The unit has not been subject to routine maintenance.
• The unit has been repaired with unoriginal spare parts.
• Ralco shall decline all responsibility for any damage, direct or indirect, caused to persons
or things by inappropriate accessories.
INFORMATION REGARDING ACCIDENTS THAT HAVE OCCURRED WHILE USING THE RADIOLOGICAL
COLLIMATOR MUST BE REPORTED IMMEDIATELY TO RALCO SRL.
RESIDUAL RISKS
The collimator has been constructed to current standards to meet the safety requisites of
directive 2007/47/CE. However, due to the presence of x-rays, the type of application implies
a residual risk derived from possible faults that could occur during operation of the unit.
The Instructions contained in the this Manual will ensure the correct use of the device and
reduce the causes of possible hazards.
The residual risks of the device are reasonable; they have been assessed and approved in the
related Risk Management Plan contained the Technical Report.
R302 F/A - General
Date:
Customer:
Information regarding possible accidents that may have occurred while using the collimator.
Directive 2007/47/CE states that accidents (such as death or grievous injury to a patient) that involve the
collimator described herein, must be reported to the Ministry of Health and to the Manufacturer.
The present form is provided to report to Ralco srl post-free.
Date:
Customer:
info_cl.fm
info_cl.fm