elite V5

Patient Monitor
Version 1.4

Main Unit Specification


Physical Specifications Channels 3
Dimension 333 mm (W) × 289 mm (H) × 211 mm (D) Recording Types Continuous real-time recording
Max Weight < 6.2 kg 8-second real-time recording
Including XM module, without options 20-second real-time recording
Automatic interval recording
Physiological alarm recording
Power Supply
Trend graph review recording
Line Voltage 100 V to 240 V~ Trend table review recording
Current 1.8 A to 0.75 A NIBP review recording
Frequency 50 Hz/60 Hz Arrhythmia review recording
Fuse T3.15 AH 250 VP Alarm review recording
Titration table recording
Hemodynamic calculation recording
Battery
C.O. measurement recording
Capacity 5000 mAh 12-lead analysis review recording
Operating Time ≥10 h with 2 new, fully charged Frozen waveform recording
batteries, at 25°C, (continuous ST view recording
SpO2 measurement and NIBP QT view recording
automatic measurement mode
at interval of 15 minutes,
ECG/TEMP module connected, Data Management
recording at interval of 10
minutes, brightness set to “1”) Data Review
≥8 h with 2 new, fully charged Trend graph/Trend Table Review 3 hours, at 1 s resolution
batteries, at 25°C, (continuous 150 hours, at 1 min resolution
SpO2 measurement and NIBP Alarm/Monitoring Event data Up to 200 sets
automatic measurement mode
NIBP Measurement Review 1200 sets
at interval of 15 minutes,
ECG/TEMP module connected, Arrhythmia events Up to 200 sets
sidestream CO2 connected, 12-lead Diagnosis Review Up to 50 sets
recording at interval of 10
minutes, brightness set to “1”) Data Storage (internal or external storage device)

Charge Time ≤10 h The monitor is on or in standby A single piece of patient data maximally contains the following information:
mode, 100% charge. Trend graph and trend table 240 hours, at 1 min resolution
≤9 h The monitor is on or in standby NIBP measurement review 1200 sets
mode, 90% charge. Alarm review 200 sets
≤6 h The monitor is off, 100% Arrhythmia event 200 sets
charge.
12-lead diagnosis review 50 sets
≤5.4 h The monitor is off, 90%
charge. Full disclosure waveforms 48 hours

Display Wi-Fi
Display screen 12.1-inch color TFT screen, touch screen is IEEE 802.11a/b/g/n
configurable Frequency Band 2.4 GHz ISM band & 5 G ISM
Resolution 800 × 600 band
Waves A maximum of 13 waveforms can be displayed
on the same screen Interfaces and others
VGA Output 1
Recorder USB Interface 4
Record Width 48 mm Nurse Call/Analog
Record Paper Width 50 mm Output/Defibrillator Synchronization 1
Paper Speed 12.5 mm/s, 25 mm/s, 50 mm/s Wired Network Interface 1

Features and specifications are subject to change without prior notice. No reproduction, copy or transmission may be made without written
permission. Not all products or features are available in all countries, contact Edan for local availability.
DVI Interface 1 (Dräger Minimodule) 86 mm (D)
PAM Interface 1 Max Weight < 1.2 kg
RS232 Interface 1 V-SpO2 module Dimension 134 mm (W) × 104 mm (H) ×
(Nellcor) 43 mm (D)
SD Card Slot 1
Max Weight < 0.2 kg
Anti-theft Lock 1
V-NIBP module Dimension 134 mm (W) × 104 mm (H) ×
(Omron, SunTech) 86 mm (D)
Data Transmission Max Weight < 0.65 kg
Data Export Ethernet / USB / SD V-BIS module Dimension 134 mm (W) × 104 mm (H) ×
Data Management Patientcare Viewer 43 mm (D)
Central Monitoring System MFM-CMS Max Weight < 0.2 kg
HIS/EMR connection HL7 MFM-CMS / GW1 Gateway V-EEG module Dimension 134 mm (W) × 104 mm (H) ×
Software 43 mm (D)
Max Weight < 0.6 kg
Safety Specifications V-RM module Dimension 134 mm (W) × 104 mm (H) ×
86 mm (D)
Compliant with IEC 60601-1: 2005+A1 :2012; IEC 60601-1-2:
Standards 2014; EN 60601-1: 2006+A1 :2013; EN 60601- Max Weight < 0.65 kg
1-2: 2015; IEC 80601-2-49: 2018 V-ICG module Dimension 134 mm (W) × 104 mm (H) ×
Anti-electroshock Type Class I equipment and internal powered 43 mm (D)
equipment Max Weight < 0.3 kg
Anti-electroshock CF: ECG, RESP, TEMP, IBP, C.O., SpO2, V-NMT module Dimension 134 mm (W) × 104 mm (H) ×
Degree NIBP, CO2, AG, EEG, RM, NMT 43 mm (D)
BF: BIS, ICG Max Weight < 0.6 kg
Ingress Protection IPX1 (protected against vertically falling water V-Link module Dimension 134 mm (W)× 104 mm (H) ×
drops) 43 mm (D)
Max Weight < 0.2 kg
Environmental Specifications PAM Dimension 503 mm (W) × 148 mm (H) ×
Temperature Working : +0°C to +40°C (32 ºF ~104 ºF) 169 mm (D)
Storage: -20°C to +55°C (-4 ºF ~131 ºF) Max Weight < 2.5 kg
Humidity Working: 15%RH to 95%RH (non-condensing)
Storage : 15%RH to 95%RH (non-condensing) ECG
Altitude Working: 86 kPa to 106 kPa Lead Mode 3 Electrodes: I, II, III
Storage : 70 kPa to 106 kPa 5 Electrodes: I, II, III, aVR, aVL, aVF, V
6 Electrodes: I, II, III, aVR, aVL, aVF, Va Vb.
10 Electrodes: I, II, III, aVR, aVL, aVF, V1- V6
Modules (elite V Series) Specification
Electrode Standard AHA, IEC
Physical Specifications
Display Sensitivity ×0.125, ×0.25, ×0.5, ×1, ×2, ×4, AUTO gain
V-C.O. module Dimension 134 mm (W) × 104 mm (H) ×
Sweep 6.25 mm/s, 12.5 mm/s, 25 mm/s, 50 mm/s
43 mm (D)
Bandwidth (-3dB) Diagnosis: 0.05 Hz to 150 Hz
Max Weight < 0.2 kg
Diagnosis 1: 0.05 Hz to 40 Hz
V-CO2 module Dimension 134 mm (W) × 104 mm (H) ×
43 mm (D) Monitor: 0.5 Hz to 40 Hz
(Respironics
Surgery: 1 Hz to 20 Hz
Mainstream) Max Weight < 0.2 kg
Enhanced: 2 Hz ~18 Hz
V-CO2 module Dimension 134 mm (W) × 104 mm (H) × Customized: High-pass Filter and Low-pass
(Respironics 86 mm (D) Filter
Sidestream) Max Weight < 0.65 kg CMRR Diagnosis: > 95 dB
V-CO2 module Dimension 134 mm (W) × 104 mm (H) × Monitor: > 105 dB
(EDAN Sidestream) 86 mm (D) Surgery: > 105 dB
Max Weight < 0.65 kg Enhanced: > 105 dB
V-CO2 module Dimension 134 mm (W) × 104 mm (H) × Diagnosis 1: > 105 dB (when Notch is turned
(Masimo, Mainstream) 43 mm (D) on)
Max Weight < 0.2 kg Customized: > 105 dB (Low-pass Filter < 40
Hz)
V-CO2 module Dimension 134 mm (W) × 104 mm (H) ×
86 mm (D) > 95 dB (Low-pass Filter > 40 Hz)
(Masimo, Sidestream)
Hum Filter In diagnosis, Diagnosis 1, monitor, surgery,
Max Weight < 0.65 kg
enhanced and customized, modes: 50 Hz/60 Hz
V-AG module Dimension 134 mm (W) × 104 mm (H) × (Hum Filter can be turned on or off manually)
(EDAN, Sidestream) 86 mm (D) Recovery Time After < 5 s (measured without electrodes as
Max Weight < 0.65 kg Defibrillation IEC60601-2-27:2011, Sect. 201.8.5.5.1
V-AG module Dimension 134 mm (W) × 104 mm (H) × requires.)
(Masimo Mainstream) 43 mm (D) ESU Protection Cut mode: 300 W
Max Weight < 0.2 kg Coagulation mode: 100 W
V-AG module Dimension 134 mm (W) × 104 mm (H) × Restore time: ≤ 10 s
(Masimo Sidestream) 86 mm (D) Pace Pulse Detecting one among I, II, III, aVR, aVL, aVF, V1-V6
Max Weight < 0.65 kg Lead
V-AG module Dimension 134 mm (W) × 104 mm (H) ×

Features and specifications are subject to change without prior notice. No reproduction, copy or transmission may be made without written
permission. Not all products or features are available in all countries, contact Edan for local availability.
Heart Rate Sweep 6.25 mm/s, 12.5 mm/s, 25 mm/s, 50 mm/s
Range ADU: 15 bpm to 300 bpm Apnea Alarm Time 10 s, 15 s, 20 s (Default), 25 s, 30 s, 35 s, 40 s
PED/NEO: 15 bpm to 350 bpm Setup
Accuracy ±1% or 1 bpm, whichever is greater
Resolution 1 bpm EDAN Module NIBP
Technique Oscillometry
PVC Mode Manual, Auto, Continuous, Sequence
Range ADU: (0 to 300) PVCs/ min Measuring Interval in 1/2/3/4/5/10/15/30/60/90/120/180/240/360/480
PED/NEO: (0 to 350) PVCs/ min AUTO Mode min and User Define

Resolution 1 PVCs/min Continuous 5 min, interval is 5 s


Measuring Parameter SYS, DIA, MAP, PR

Pauses/min Measuring Range


Adult Mode SYS: 25 mmHg to 290 mmHg
Range ADU/PED/NEO: (0 to 30) pauses/min
DIA: 10 mmHg to 250 mmHg
Resolution 1 pause/min
MAP: 15 mmHg to 260 mmHg
Pediatric Mode SYS: 25 mmHg to 240 mmHg
ST value DIA: 10 mmHg to 200 mmHg
Range -2.0 mV to +2.0 mV MAP: 15 mmHg to 215 mmHg
Accuracy -0.8 mV to +0.8 mV: ±0.02 mV or 10%, Neonatal Mode SYS: 25 mmHg to 140 mmHg
whichever is greater. Beyond this range: not DIA: 10 mmHg to 115 mmHg
specified.
MAP: 15 mmHg to 125 mmHg
Resolution 0.01 mV
Cuff Pressure
Measuring Range 0 mmHg to 300 mmHg
QT measurement Pressure Resolution 1 mmHg
Range 200 ms ~ 800 ms Maximum Mean Error ±5 mmHg
Resolution 4 ms Maximum Standard
Accuracy ±30 ms Deviation 8 mmHg
Maximum Measuring Adult/Pediatric: 120 s
Period Neonate: 90 s
QTc measurement
Range 200ms ~ 800 ms Typical Measuring
Period 20 s to 35 s (depend on HR/motion disturbance)
Resolution 1 ms
Dual Independent Channel Overpressure Protection
Adult (297 ±3) mmHg
ΔQTc measurement
Pediatric (245 ±3) mmHg
Range -600 ms ~ 600 ms
Neonatal (147 ±3) mmHg
Resolution 1 ms

Venipuncture pressure
Arrhythmia analysis
Adult Default: 60 mmHg
Asystole, Sustain VT, V-Fib/V-Tach, ExtremeTachy, ExtremeBrady, V-Tach, Options: 20 mmHg, 30 mmHg, 40 mmHg, 50
Vent Brady, Tachy, Brady, Wide QRS Tachy, Non-Sustain VT, Afib, Vent mmHg, 60 mmHg, 70 mmHg, 80 mmHg, 90
Rhythm, Acc. Vent Rhythm, Pause, Pauses/min High, PVCs High, R on T, mmHg, 100 mmHg, 110 mmHg, 120 mmHg
PVC Bigeminy, PVC Trigeminy, Pacer not Pacing, Pacer not Capture,
Missed Beat, VEB, PVC, Couplet, Run PVCs, IPVC, Irr Rhythm, PAC Pediatric Default: 40 mmHg
Bigeminy, Multiform PVCs, PAC Trigeminy, Low Voltage (Limb) Options: 20 mmHg, 30 mmHg, 40 mmHg, 50
mmHg, 60 mmHg, 70 mmHg, 80 mmHg
Neonatal Default: 30 mmHg
12-Lead ECG Synchronization Analysis
Options: 20 mmHg, 30 mmHg, 40 mmHg, 50
Average parameters of heart beat PR interval (ms) mmHg
Heart rate (bpm) QRS interval (ms)
Time limit of P wave (ms) QT/QTC (ms) SunTech Module NIBP
P-QRS-T AXIS Method Oscillometric
Mode Manual, Auto, Continuous, Sequence
RESP Measuring Interval in 1/2/3/4/5/10/15/30/60/90/120/180/240/360/480
Method Impedance between RA-LL, RA-LA AUTO Mode min and User Define
Measurement lead Options are lead I and II. The default is lead II. Measuring Type SYS, DIA, MAP
Calculation Type Manual, Automatic Measuring Range
Measuring Range Adult: 0 rpm to120 rpm Adult Mode SYS: 40 mmHg ~ 260 mmHg
Neo/Ped: 0 rpm to150 rpm DIA: 20 mmHg ~ 200 mmHg
Resolution 1 rpm MAP: 26 mmHg ~ 220 mmHg
Accuracy Adult: 6 rpm to 120 rpm: ±2 rpm Pediatric Mode SYS: 40 mmHg ~ 230 mmHg
0 rpm to 5 rpm: not specified DIA: 20 mmHg ~ 160 mmHg
MAP: 26 mmHg ~ 183 mmHg
Neo/Ped: 6 rpm to 150 rpm: ±2 rpm
0 rpm to 5 rpm: not specified Neonatal Mode SYS: 40 mmHg ~ 130 mmHg
DIA: 20 mmHg ~ 100 mmHg
Gain Selection ×0.25, ×0.5, ×1, ×2, ×3, ×4, ×5
MAP: 26 mmHg ~ 110 mmHg

Features and specifications are subject to change without prior notice. No reproduction, copy or transmission may be made without written
permission. Not all products or features are available in all countries, contact Edan for local availability.
Pressure Resolution 1 mmHg Nellcor: 20 bpm to 300 bpm
Maximum mean error ±5 mmHg Accuracy EDAN: ±2 bpm
Maximum standard Nellcor: ±3 bpm (20 bpm to 250 bpm)
deviation 8 mmHg Resolution EDAN: 1 bpm
Maximum measuring Adult/Pediatric: 130 s Nellcor: 1 bpm
period Neonate: 75 s PR (NIBP)
Overpressure Adult/Pediatric: < 300 mmHg Measuring range EDAN: 40 bpm to 240 bpm
protection Neonate: < 150 mmHg SunTech: 30 bpm to 220 bpm
Accuracy EDAN: ±3 bpm or 3.5%, whichever is greater
EDAN Module SpO2 SunTech: ±3 bpm or ±2%, whichever is greater
Measuring Range 0% to 100% Resolution EDAN:1 bpm
Resolution 1% SunTech: 1 bpm
Data Update Period 1s PR (IBP)
Accuracy Adult /Pediatric: ±2% (70% to 100% SpO2) Measuring range EDAN: 20 bpm to 300 bpm
Undefined (0% to 69% SpO2)
Accuracy EDAN: 30 bpm to 300 bpm: ±2 bpm or ±2%,
Neonate: ±3% (70% to 100% SpO2) whichever is greater;
Undefined (0% to 69% SpO2) 20 bpm to 29 bpm: undefined
Resolution EDAN: 1 bpm
PI (Perfusion Index)
Measuring Range Level display (default): 0-10, invalid PI value is IBP
-?-.
Channel Maximum 8
Percentage display: 0%~20%.
Technique Direct invasive measurement
Resolution Level display: 1
Percentage display: 1% (10% to 20%) Measuring Range Art: 0 mmHg to + 300 mmHg
0.1% (1.0% to 9.9%) PA: -6 mmHg to + 120 mmHg
0.01% (0.00% to 0.99%) CVP/RAP/LAP/ICP: -10 mmHg to + 40 mmHg
P1/P2: -50 mmHg to +300 mmHg
Nellcor Module SpO2 Resolution 1 mmHg
Measuring Range 1% to 100% Accuracy ±2% or ±1 mmHg, whichever is greater
Resolution 1% (not including sensor) ICP: 0 mmHg to 40 mmHg: ±2% or ±1 mmHg,
whichever is greater;
Data Update Period 1s
-10 mmHg to -1 mmHg: undefined
Accuracy
Pressure Unit kPa, mmHg, cmH2O
DS-100A, OXI-A/N(Adult)
D-YS (Adult and Pediatric)
EDAN G2 Sidestream Module CO2
OXI-P/I (Pediatric) ±3% (70% to 100% SpO2)
MAX-A, MAX-AL, MAX-N, MAX-P, Intended Patient Adult, pediatric, neonatal
MAX-I, MAX-FAST (Adult and Measure Parameters EtCO2, FiCO2, AwRR
Pediatric) ±2% (70% to 100% SpO2) Unit mmHg, %, kPa
MAX-A, MAX-AL, MAX-N, MAX-P, Measuring Range CO2: 0 mmHg to 150 mmHg (0% to 20%)
MAX-I, MAX-FAST (Adult and
Pediatric) ±3% (60% to 80% SpO2) AwRR: 2 rpm to 150 rpm
If sensor is used for neonate as recommended, the accuracy will be larger Resolution EtCO2/FiCO2: 1 mmHg
than adult by ±1. AwRR: 1 rpm
EtCO2 Accuracy
Typical conditions: ±2 mmHg, 0 mmHg to 40 mmHg
TEMP Ambient temperature:
(25 3) C
Technique Thermal resistance
Barometric pressure:
Position Skin, oral cavity, rectum (760  10) mmHg
Measure Parameter T1, T2, TD (the absolute value of T2 minus T1) Balance gas: N2
Channel 2 Sample gas flowrate:
Sensor Type YSI-10K and YSI-2.252K 100 ml/min
Unit °C, ºF ±5% of reading, 41 mmHg to 70 mmHg
Measuring Range 0°C to 50°C (32 ºF to 122 ºF) ±8% of reading, 71 mmHg to 100 mmHg
Resolution 0.1°C (0.1ºF) ±10% of reading, 101 mmHg to 150 mmHg
Accuracy Accuracy (not including sensor): ±0.1°C All conditions ±12% of reading or ±4 mmHg, whichever is
greater
Sensor accuracy: ≤ ±0.2°C
AwRR Accuracy ±1 rpm
Transient Response
Time ≤ 30 s Sample Gas Flowrate 70 ml/min or 100 ml/min (default), accuracy:
15 ml/min
Warm-up Time Display reading within 20 s; reach to the
PR designed accuracy within 2 minutes.
PR (SpO2) Response Time < 4 s (with 2 m gas sampling tube, sample gas
Measuring range EDAN: 25 bpm to 300 bpm flowrate: 100 ml/min)

Features and specifications are subject to change without prior notice. No reproduction, copy or transmission may be made without written
permission. Not all products or features are available in all countries, contact Edan for local availability.
< 4 s (with 2 m gas sampling tube, sample gas Technique Thermodilution Technique
flowrate: 70 ml/min) Measure Parameters C.O., TB, TI
Barometric Pressure Automatic (The change of barometric pressure C.O.: 0.1 L/min to 20 L/min
Measuring Range
Compensation will not add additional errors to the
measurement values.) TB: 23°C to 43°C (73.4ºF to 109.4ºF)
Zero Calibration Support TI: -1°C to 27°C (30.2ºF to 80.6ºF)
Calibration Support Resolution C.O.: 0.1 L/min
Apnea Alarm Delay 10 s, 15 s, 20 s (Default), 25 s, 30 s, 35 s, 40 s, Resolution TB, TI: 0.1°C (+0.1 ºF)
60s Accuracy C.O.: ±5% or ±0.2 L/min, whichever is greater
TB: ±0.1°C (not including sensor)
Respironics Sidestream and Mainstream Modules CO2 TI: ±0.1°C (not including sensor)
Applicable Patient Adult, pediatric and neonatal patients EDAN Module AG
Type
Intended Patient Adult, pediatric, neonatal
Measure Parameters EtCO2, FiCO2, AwRR
Measure Parameters Halothane (HAL), Isoflurane (ISO), Enflurane
Unit mmHg, %, Kpa (ENF), Sevoflurane (SEV),
Measuring Range EtCO2: 0 mmHg to 150 mmHg Desflurane (DES), CO2, O2, N2O, AwRR, and
FiCO2: 3 mmHg to 50 mmHg MAC
AwRR: 0 rpm to 150 rpm (Mainstream) Unit HAL, ISO, ENF, SEV, DES, N2O: %;
2 rpm to 150 rpm (Sidestream) CO2, O2: mmHg, %, kPa, default is %;
AwRR: bpm;
Resolution EtCO2 / FiCO2 : 1 mmHg
Measuring Range CO2: 0~15 vol%
AwRR: 1 rpm
N2O: 0~100 vol%
EtCO2 Accuracy ±2 mmHg, 0 mmHg to 40 mmHg
Halothane/ Enflurane/ Isoflurane: 0~8 vol%
±5% of reading, 41 mmHg to 70 mmHg
Sevoflurane: 0~10 vol%
±8% of reading, 71 mmHg to 100 mmHg
Desflurane: 0~22 vol%
±10% of reading, 101 mmHg to 150 mmHg
O2: 0~100%
±12% of reading, RR is over 80 rpm
(sidestream) Resolution N2O, O2: 1%
There will be no degradation in performance CO2, AG: 0.1%
due to Respiration Rate. (mainstream) AwRR Measurement range: 2 ~ 150 rpm
AwRR Accuracy ±1 rpm Measuring accuracy: ±1 bpm (120 bpm and
Zero Calibration Support below),
Apnea Alarm Delay 10 s, 15 s, 20 s (Default), 25 s, 30 s, 35 s, 40 s Not specified (120 bpm above)
Sample Gas Flowrate Resolution: 1 rpm
(sidestream) 50±10 ml/min Sampling Flow Rate 150 ml/min, accuracy ±15 ml/min
CO2 Rise Warm-up Time Display reading within 20 s; reach to the
Time/Response Time designed accuracy within 2 minutes
(mainstream) Less than 60 ms
Response Time < 4 s (with 2 m gas sampling tube, sample gas
Sensor Response Time < 3 seconds, including transport time and rise flowrate: 150 ml/min)
(sidestream) time
Barometric Pressure
User setup Masimo ISA analyzer AG
Compensation
Module Type
ISA AX+ Displaying the concentration of CO2, N2O, and
Masimo Mainstream Module CO2
two anesthesia agent and identifying the
Applicable Patient Adult, pediatric and neonatal patients anesthesia agent automatically (built-in module)
Type
ISA OR+ Displaying the concentration of CO2, O2, N2O,
Measure Parameters EtCO2, FiCO2, AwRR and two anesthesia agent and identifying the
Measuring Range EtCO2: 0 to 25 vol% anesthesia agent automatically (built-in module)
FiCO2: 0 to 25 vol% Measurement CO2, N2O, O2, Halothane (HAL), Isoflurane
Parameters (ISO), Enflurane (ENF),
AwRR: 0 rpm to 150 rpm
Sevoflurane (SEV), Desflurane (DES), AwRR,
Resolution EtCO2 / FiCO2: 0.10 % MAC
AwRR: 1 rpm Sampling Flow Rate 50±10 ml/min
Accuracy EtCO2: 0 to15 vol%: ±0.2 vol% + 2% of reading Compensations Automatic compensation for pressure,
15 to 25 vol%: Unspecified temperature and broadening effects on CO2
AwRR: ±1 rpm Warm-up Time < 20 s
Sample Gas Flowrate Total System Response < 4 seconds
(sidestream) 50±10 ml/min Time (with 2 m Nomoline Airway Adapter Set
CO2 Rise sampling line)
Time/Response Time ≤ 200 ms (Measured according to EN ISO Data Update Period 1s
(mainstream) 80601-2-55.)
Measurement Range CO2: 0 to 25 vol%
Sensor Response Time < 3 seconds, including transport time and rise O2: 0 to 100 vol%
(sidestream) time
N2O: 0 to 82 vol%
Zero Calibration Auto, Manual HAL, ENF, ISO, SEV, DES: 0 to 25 vol%
AwRR: 0 rpm to 150 rpm
C.O. Resolution CO2: 0.1%

Features and specifications are subject to change without prior notice. No reproduction, copy or transmission may be made without written
permission. Not all products or features are available in all countries, contact Edan for local availability.
HAL, ENF, ISO, SEV, DES: 0.1% BIS
N2O: 1% Technique Bispectral index, power spectrum analysis
O2:1%
Measure Parameters
AwRR: 1 rpm
Primary Parameter BIS: 0 to 100
Respiration Rate 0 to 150 ±1 breaths/min
Secondary Parameters SQI: 0% to 100%
Apnea Alarm Delay 20 s (Default), 25 s, 30 s, 35 s, 40 s
SR: 0% to 100%
EMG: 30 dB to 80 dB
Masimo IRMA module AG
SEF: 0.5 Hz to 30.0 Hz
Module Type IRMA Displaying the concentration of CO2, N2O and
AX+ two anaesthesia agent and indentifying two TP: 40 dB to 100 dB
anaesthesia agent BC (only applicable to BIS™ Extend Sensor): 0
Measurement CO2, N2O, HAL, Isoflurane (ISO), Enflurane to 30
Parameters (ENF), Sevoflurane(SEV), Desflurane (DES), ASYM (Available only with BISx4 and
AwRR, MAC Bilateral Sensor): 0% to 100%
Barometric Pressure Automatic Sweep 6.25 mm/s, 12.5 mm/s, 25 mm/s, 50 mm/s
Compensation BIS Trend 6 min, 12 min, 30 min, 60 min
Warm-up Time < 20 seconds (agent identification enabled and EEG Bandwidth 0.25 Hz~50 Hz
full accuracy)
Noise (EEG
Total System Response < 1 second (Measured according to EN ISO Waveform) < 0.3μV (0.25 Hz~50 Hz)
Time 80601-2-55.)
Data Update Period 1s
EDAN Module EEG
Measurement Range CO2: 0 to 25 vol%
Measuring Parameters SEF/MF/PPF: 0.5 Hz to 30 Hz
N2O: 0 to 82 vol%
Alpha/Beta/Theta/Delta/SR: 0 to 100%
HAL, ENF, ISO, SEV, DES: 0 to 25 vol%
AwRR: 0 to 150 rpm TP: 40 dB to 100 dB
Resolution CO2: 0.1% EMG: 30 dB to 80 dB
HAL, ENF, ISO, SEV, DES: 0.1% Sampling Rate 500 Hz
N2O: 1% Input Dynamic Range ±2 mVac
AwRR: 1 rpm Bandwidth 0.5 Hz to 30 Hz
Respiration Rate 0 to 150 ±1 bpm. The respiration rate I Max. DC Input Offset ±500 mV DC
displayed after three breaths and the average Voltage
value is updated every breath. (Measured at I/E
ratio 1:1 using breath simulator according to Common Mode > 105 dB @50 Hz
EN ISO 80601-2-55 fig. 201.101.) Rejection Ratio
Apnea Alarm Delay 20 s (Default), 25 s, 30 s, 35 s, 40 s Noise ≤ 3 uVp-p (0.5 Hz to 30 Hz)
Differential Input ≥ 15 MΩ
Impedance
Dräger Module AG
Electrode Impedance Range: 1 KΩ to 100 KΩ
Method Sidestream gas measurement Resolution: 1 KΩ
Infrared measurement: CO2, N2O, Anesthetic
agents Connection Type Bipolar, Unipolar
Paramagnetic measurement: O2 SEF Threshold 90%, 95%, 98%
Barometric Pressure Sweep 6.25 mm/s, 12.5 mm/s, 25 mm/s, 50 mm/s
Compensation Automated compensation Scale 50 μv, 100 μv, 200 μv, 500 μv, 1 mv, 2 mv
Gas Sampling Rate 200 mL/min ±20 mL/min Low-pass Filter 15 Hz, 30 Hz, 50 Hz,
Maximum time until 41 hrs (sample gas under BTPS conditions, High-pass Filte 0.16 Hz, 0.5 Hz, 1.0 Hz, 2.0 Hz, 5.0 Hz, Off
water trap requires ambient air 23 °C) Hum Filter Off, 50 Hz, 60 Hz
draining
EEG Trend Scale 6 min, 12 min, 30 min, 1 h
Total System Response
Time < 3s CSA Scale 20 min, 30 min, 1 h
Drift Compensation Automated cyclical zeroing, once per day (in DSA Scale 20 min, 30 min, 1 h
(zeroing) error-free operation)
Zeroing Duration < 20s RM
Measurement Range O2: (0 to 100) Vol% Measure Parameters Flow, Tidal Volume, Airway Pressure,
CO2: (0 to 13.6) vol% Respiration Rate
N2O: (0 to 100) vol% Numeric – PIP (Peak Inspiratory Pressure)
Halothane: (0 to 8.5) vol% – Pplat (Plateau Pressure)
Isoflurane: (0 to 8.5) vol% – PEEP (Positive End-Expiratory Pressure)
Enflurane: (0 to 10) vol% – Pmean (Mean Airway Pressure)
Sevoflurane: (0 to 10) vol% – PIF (Peak Inspiratory Flow)
Desflurane: (0 to 20) vol% – PEF (Peak Expiratory Flow)
Respiratory Rate Range: 0/min to 100/min (Respiratory rate is – TVi (Inspiratory Tidal Volume)
derived from the CO2 value) – TVe (Expiratory Tidal Volume)
Accuracy: 0/min to 60/min: ±1 /min – MVi (Inspiratory Minute Volume)
> 60 /min: not specified – MVe (Rxpiratory Minute Volume)
Resolution: 1 /min – I:E (Ratio of the Inspiratory Time and
Expiratory Time)
– Cdyn (Dynamic Compliance)

Features and specifications are subject to change without prior notice. No reproduction, copy or transmission may be made without written
permission. Not all products or features are available in all countries, contact Edan for local availability.
– RAWi (Airway Resistance-Inspired) Temperature Working: +0°C to +40°C (32ºF ~ 104ºF)
– RAWe (Airway Resistance-Expired) Transport and Storage: -20°C to +55°C (-4ºF ~
– NIP (Negative Inspiratory Pressure) 131ºF)
– RSBI (Rapid Shallow Breathing Index) Humidity Working: 15%RH to 95%RH (non-condensing)
– P0.1 (Airway Pressure at 100 msec After the
Transport and Storage: 15%RH to 95%RH
Start of Inspiration)
(non-condensing)
– AwRR (airway respiration rate)
Altitude Working: 86 kPa to 106 kPa
– EtCO2 (End-Tidal Carbon Dioxide)
– FiCO2 (Fraction of Inspired Carbon Dioxide) Transport and Storage: 70 kPa to 106 kPa
Loop F-V (flow-volume) loops
P-V (pressure-volume) loops. V-CO2 module (Respironics Sidestream & Mainstream)
Temperature Working: +0°C to +40°C (32ºF ~ 104ºF)
ICG Transport and Storage: -20°C to +55°C (-4ºF ~
131ºF)
Technique Thoracic electrical bioimpedance
Humidity Working: 15%RH to 90%RH (non-condensing)
Measuring Range SV: 0 ml/beat~250 ml/beat
HR: 40 bpm~250 bpm Transport and Storage: 15%RH to 90%RH
(non-condensing)
C.O.: 0 L/min~30 L/min
Altitude Working: 86 kPa to 106.6 kPa
Accuracy SV: Undefined
HR: ±2 bpm Transport and Storage: 70 kPa to 106.6 kPa
C.O.: Undefined
V-AG module (Masimo Sidestream & EDAN Sidestream)
V-NMT Module (Xavant Module) Temperature Working: +5°C to +40°C (41ºF ~ 104ºF)
Stimulation Mode TOF, TWI, PTC, DBS Transport and Storage: -20°C to +55°C (-4ºF ~
Stimulation Output Pulse width: 0.2 ms ±5% 133ºF)
Stimulation current: 0~80 mA, step is 1 mA Humidity Working: 15%RH to 95%RH (non-condensing)
Current accuracy: 10% Transport and Storage: 15%RH to 95%RH
(non-condensing)
Max skin resistance: 5 KOhm
Altitude Working: 86 kPa to 106 kPa
Supply voltage range: 4.5 V~6.5V
Transport and Storage: 70 kPa to 106 kPa

V-NMT Module (EDAN Module)


V-AG module (Masimo Mainstream)
Stimulation Mode TOF, ST, PTC, DBS, TET
Temperature Working: +10°C to +40°C (50ºF ~ 104ºF)
Stimulation Output Pulse width: 0.2 ms ±10%
Transport and Storage: -20°C to +55°C (-4ºF ~
Stimulation current: 0~80 mA, step is 1 mA 133ºF)
Current accuracy: ±10% or ±2 mA, whichever Humidity Working: 15%RH to 95%RH (non-condensing)
is greater
Transport and Storage: 15%RH to 95%RH
Max skin resistance: 0~5 kΩ, 80 mA (non-condensing)
Supply voltage range: 4.5 V~5.5 V Altitude Working: 86 kPa to 106 kPa
Transport and Storage: 70 kPa to 106 kPa
LiDCO (V-Link module)
Supported model LiDCO plus, LiDCO rapid V-AG module (Dräger Minimodule)
Measuring parameter CO, CI, SV, SI, SVR, SVRI, DO2, DO2I, SaO2, Temperature Working: +10°C to +40°C (50ºF ~ 104ºF)
CVP, HR, MAP, DIA, SYS, SVV, PPV, SPV
Transport and Storage: -20°C to +55°C (-4ºF ~
and HRV.
131ºF)
Humidity Working: 15%RH to 95%RH (non-condensing)
Communication with Other Device (V-Link module)
Transport and Storage: 15%RH to 95%RH
Heinen Löwenstein LeoniPlus Ventilator (non-condensing)
LeonPlus Anesthetic device Altitude Working: 86 kPa to 106 kPa
Leon Anesthetic device
Transport and Storage: 70 kPa to 106 kPa
Maquet SERVO-U, SERVO-n, SERVO-air,
SERVO-i or SERVO-s
V-RM module
Dräger Savina 300 Ventilator
Temperature Working: +10°C to +40°C (50ºF ~ 104ºF)
Evita V300 Ventilator
Babylog VN500 Ventilator Transport and Storage: -20°C to +55°C (-4ºF ~
131ºF)
Evita Infinity V500 Ventilator
Oxylog 3000+ Ventilator Humidity Working: 15%RH to 95%RH (non-condensing)
Fabius Plus XL Anesthetic device Transport and Storage:15%RH to 95%RH (non-
Atlan (A350) Anesthetic device condensing)
Perseus A500 Anesthetic device Altitude Working: 86 kPa to 106 kPa
Transport and Storage: 70 kPa to 106 kPa
Environmental Specifications
PAM, XM module, V-SpO2 module (Nellcor), V-NIBP module V-ICG module
(SunTech), V-IBP module, V-C.O. module, V-BIS module, V-CO2 Temperature Working: +10°C to +40°C (50ºF ~ 104ºF)
module (EDAN Sidestream), V-Link module, Recorder
Transport and Storage: 0°C to +50°C (32ºF ~

Features and specifications are subject to change without prior notice. No reproduction, copy or transmission may be made without written
permission. Not all products or features are available in all countries, contact Edan for local availability.
122ºF)
Humidity Working: 30%RH to 75%RH (non-condensing)
Transport and Storage: 15%RH to 95%RH
(non-condensing)
Altitude Working: 86 kPa to 106 kPa
Transport and Storage: 70 kPa to 106 kPa

V-NMT module (Xavant)


Temperature Working: +10°C to +40°C (50ºF ~ 104ºF)
Transport and Storage: 0°C to +50°C (32ºF ~
122ºF)
Humidity Working: 15%RH to 90%RH (non-condensing)
Transport and Storage: 15%RH to 90%RH
(non-condensing)
Altitude Working: 86 kPa to 106 kPa
Transport and Storage: 70 kPa to 106 kPa

Features and specifications are subject to change without prior notice. No reproduction, copy or transmission may be made without written
permission. Not all products or features are available in all countries, contact Edan for local availability.