Op 900 Manual
Op 900 Manual
USER’S MANUAL
     (OP-900A, OP-900B Series)   9.9.19
         optimascale.com
               TABLE OF CONTENTS
Safety Precautions 1
Features 2
Specifications 3
Power Supply 4
Displays 5
Calibration 10
Helpful Definitions 17
Connectors 18-19
Relay Output 25
Troubleshooting 26
Q&A 27
Contact Us                          28
                 SAFETY PRECAUTIONS
For safe operation of the weighing indicator, please follow these
instructions:
●● Calibration inspection and maintenance of the indicator are prohibited by non-pro-
   fessional staff
●● Please ensure that the indicator rests on a stable surface
●● The indicator is a piece of static sensitive equipment; Please cut off power during
   electrical connections
●● Touching the internal components by hand is prohibited
●● DO NOT exceed the rated load limit of the unit
●● DO NOT step on the unit
●● DO NOT jump on the scale
●● DO NOT use this product if any of the components are cracked
●● DO NOT use for purposes other then weight taking
●● To avoid damaging the battery do not keep charger plugged in once battery is fully
   charged
●● Make sure the weight is not over the Max capacity as it could damage the load cell
   inside
●● Material that has a static electric charge could influence the weighing. Discharge
   the static electricity of the samples, if possible. Another
   solution to the problem is to wipe both sides of the pan and the top of the case with
   an anti-static agent
                                                                                          1
                                     FEATURES
    ●●   LED or LCD display option
    ●●   Multiple weighing units: kg/lb (g/oz/lb:oz offered on LCD versions)
    ●●   Gross/Tare/Pre-Set Tare/Zero
    ●●   Multiple Hold functions
    ●●   Count weighing
    ●●   Accumulation weighing
    ●●   Overload / Underload indication
    ●●   Connects to multiple printers
    ●●   Splash proof keyboard and display
    ●●   Connects to a Remote Display/Scoreboard
    ●●   Power saving mode
    ●●   Ability to modify gravity based on different geographical locations
    ●●   RS232 output
    ●●   NTEP approved for 5,000 divisions
    ●●   Relay output (optional) and 4-20mA analog output (optional)
    ●●   Can connect to a PC or printer for data logging (optional)
    ●●   Wireless capability (optional)
    ●●   Rechargeable battery (optional)
Technical Parameters
    ●●   Accuracy class: 5000 e
    ●●   Resolution - Display: 30,000 ; ADC: 2,000,000
    ●●   Zero stability error: TK0 < 0.1μV//K
    ●●   Span stability error: TKspn < ± 6 ppm//K
    ●●   Sensitivity (internal): 0.3 μV / d
    ●●   Input voltage: -30 to +30mV DC
    ●●   Excitation circuit: 5 VDC, 4 wire connection, 6 load cell of 350ohm max
    ●●   AC power: AC 100-250V (use only the included 9V adapter supplied)
    ●●   Operation temperature: -10 °C ~ +40 °C
    ●●   Operation humidity: ≤90%RH
    ●●   Storage temperature: -40 °C ~ +70 °C (32-104°F)
2
SPECIFICATIONS
FIGURE 1: INDICATOR MEASUREMENTS
                                   3
                             POWER SUPPLY
AC Adapter
If the indicator is powered by an adapter, plug the adapter directly into the “DC” pin
located at the bottom of the indicator. We recommend to plug into a wall outlet to avoid
interference with other wirings.
Battery (Optional)
If you have an indicator with the rechargeable battery option, please charge the
internal battery fully before first time use. A 110 to 220V AC adapter should be
provided with your indicator. Please use only the AC adapter provided to prevent dam-
age to your indicator. To keep the battery in best condition, fully discharge the battery
every month by leaving the indicator on until the indicator powers off, and then recharge
fully. If the battery is not going to be used for a long period of time it is recommended to
remove it to avoid leakage.
    On OP-900A
    ●● When the Battery is low the battery indicator light flashes red
    ●● During charging the red light will stay lit
    ●● The light will turn green once fully charged
    ON OP-900B
    ●●    symbol will indicate battery’s charge
    ●●    symbol indicates that the battery needs to be charged
4
OP-900A (LED)
OP 900B (LCD)
                5
       DISPLAY AND KEY DESCRIPTION
    ON/OFF Powers the Indicator On or Off if held for 2 seconds
    HOLD   Holds the weight (5 Hold options; can be changed in parameter settings)
    TOTAL     1. Accumulates weights
              2. Works with “Print” to perform the accumulation function and check the
              accumulation result
    UNITS     Shifts between weighing units
    COUNT     Use the scale to count product based on a sample weight
    TARE      1. Resets the scale to zero when there is something on the scale
              (ex. Tare out the weight of a pallet to weigh only the product on it)
              2. Clears the tare to see the gross weight (pallet + product)
    ZERO      Zero’s the scale
    PRINT     Print data
              The scale is at zero
6
             OPERATING INSTRUCTIONS
Power On
  ●● Turn on the power by pressing the power button for 2 seconds. Once on, the scale
     will flash the voltage and then begin to auto-check and count down from 0-9
     sequentially before entering the weighing mode
     Note: Anything on the scale before powering on will automatically be tared out.
Zeroing
  ●● The zero function is used only when the scale is empty and is not at gross zero due
     to material build up
  ●● Pressing the ZERO key will reset your scale to 0
  ●● Depending on what your manual zero range parameter is set to, you can zero out
     any number within your set selection, after that you will receive an error and will
     need to tare out the weight
Unit Selection
  ●● To switch between measuring units ((kg/g/lb/oz/lb:oz) press the UNITS key
Tare Function
  ●● The Tare function is used when you only wish to see the current change in weight,
     not the entire amount of weight that is on the scale
  ●● When the indicator is in gross mode (gross light is shown) pressing the TARE key
     will Tare the current weight on the scale and enter the net mode (net light shown)
  ●● For example if you are using a container add the container to the scale, press tare
     and the display will show the tare symbol        and reset back to 0
  ●● Add your product to the scale to weigh without the weight of the container
  ●● To exit Tare mode press the TARE key again to enter gross mode and you will see
     the total weight of the container and the product
Note: If you remove the container the scale will show the minus weight of the container
  To use a pre-set tare weight
  ●● Press and hold the TARE key for 2 seconds
  ●● Input the tare weight using the arrow keys
  ●● Press print key to confirm
                                                                                        7
Counting Function
    ●● The counting function is used to count a high volume of identical parts. You can do
       this by setting a sample and then either adding to the sample or taking away from
       the sample to count the number of objects on the scale
    ●● In weighing mode: Put a sample weight on the scale (Sample options are 5, 10, 20,
       50, 100, 200, 500)
    ●● Then press the COUNT key to go to the counting mode (“pcs” will light up)
    ●● It will then display the sample number (ex. PCS 0) which you can change using
       the up or down arrows
    ●● Press the PRINT key to confirm your sample number
    ●● The scale is now ready to start counting, load your product on the scale and the
       indicator will show the quantity
    ●● To exit counting mode press the COUNT key
    ●● If you want to count a different product hold the PRINT and COUNT key together
       and the sample pieces will reset back to zero
Accumulation
    ●● The accumulation function is used to add multiple weights and total them together
    ●● In weighing mode load the first weight, once stable press the ACCUM key to enter
       the accumulation mode. The “total” indicator/light will display
    ●● The screen will show “n001” to indicate the first weight has been saved
    ●● Remove the first weight and press the ZERO key to stabilize the scale
    ●● Add the second weight to the scale
    ●● Once stable press the ACCUM key to add the weight to the accumulated total
    ●● The screen will show “n002” to indicate the second weight has been saved
    ●● Repeat previous steps until all desired weights have been added to the total
       (you can accumulate up to 999 different weights)
    ●● When you are done and want to display the accumulated total, press the ACCUM
       and PRINT key together. The accumulated number “n002” (the number of weights
       you are adding together) will flash on the display followed by the total
    ●● The total will display by flashing between 2 sets of numbers
    ●● There are 8 digits in total, the display will flash 4 at a time, the first 4 on the left and
       the last 4 on the right. For example if the first 4 digits are “0012” and the last 4 digits
       are “3456” the actual weight is 001234.56 or 1234.56 lbs/kg
    ●● If you want to print the accumulated total, hold the PRINT key for one second while
       the last 4 digits of the total are shown
    ●● To exit accumulation mode, wait for the last 4 digits to the right of the screen to
       appear, and then press and hold the ACCUM key for one second
    ●● “CLr n” will be displayed, asking you if you want to keep the data?
        ●● If NO you do not want to clear the accumulated total, then keep “CLr n”.
        ●● If YES you do want to clear the total, then use the arrow key to change to “CLr y”
    ●● Finally, press the PRINT key to select exit accumulation mode
8
Hold
 There are 4 different hold functions you can choose from in the C11 parameter
 1. Peak Hold: Grabs the highest weight (for materials testing, ie. tension and pulling
 force)
 ●● Press the HOLD key then add weight to the scale
 ●● The indicator will show the highest weight it recorded and hold it on the screen until
    a higher weight is placed on the scale
 2. Manual Hold: Grabs the current weight and holds it so it will not change/fluctuate
 ●● While weighing, press HOLD and the indicator will hold the current weight on the
    screen until HOLD is pressed again
 3. Auto Hold: If the weight on the scale is above 20d (20 x division) and is stable,
 the indicator will hold that weight on the screen for 3 seconds then go back to general
 weighing
 ●● Pressing the hold key is unnecessary, holding is done automatically when the scale
    is stable
 4. Average Hold: Used for animal weighing, the indicator will display the average
 weight sampled from 3 seconds
 ●● Add livestock to scale and press HOLD
 ●● Indicator screen will show “LOC” for 3 seconds, then display the average weight
    from those 3 seconds
 ●● Press HOLD again to exit holding mode
Print
 ●● If the indicator is connected to a printer and the weight on the scale is stable press the
    PRINT key to print the current weight
 ●● In accumulation mode hold the PRINT key for one second when the last 4 digits of the
    total weight are shown to print the total weight
 Note: In tare mode the printer can not print if negative weight is shown
                                                                                             9
              CALIBRATION PROCEDURE
1.   Turn on the scale by holding ON/OFF        for 2 seconds.
2.   Press HOLD       and PRINT       together to access the setup menu.
3.   If done correctly, the display should now show C01.
4.   Press PRINT      to access the C1 channel. The display should show [C1 #].
5.   Press ZERO to choose which unit you want to calibrate in (1 = kg, 2 = lb).
6.   Press PRINT      to set the value. The display will now show C02.
7.   Press PRINT      to access the C2 channel. The display should show [C2 #].
8.   Press ZERO to change the setting to the decimal places desired (The C2 channel is
     used to adjust the decimal point on the scale. A value of 1 means there is one digit
     behind the decimal point.)
9. Press PRINT        to set the value. The display will now show C03.
10. Press PRINT       to access the C3 channel. The display should show [C3 #].
11. Press ZERO to cycle through the values until the desired graduation appears.
     (The C3 channel adjusts the divisions on the scale. A value of 1 selected and C2 set
     to 1, the scale will read in 0.1 lb. increments.)
12. Press PRINT       to set the value. The display will now show C04.
13. Press PRINT       to access the C4 channel. The display will show [######].
14. Enter in the maximum capacity you want to use for this scale by using UNIT and
     COUNT to move the cursor left and right, and TARE and ZERO to move the
     values down and up. (The C4 channel is used to enter in the max capacity of the
     scale; Make sure this doesn’t exceed the max capacity of the scale; Max capacity
     divided by the increment set in C02 and C03 above cannot exceed 5000.)
15. Press PRINT       to set the value. The display will now show C05.
16. Press PRINT       to access the C5 channel. The display should show [C5       0].
17. The C5 channel calibrates zero on the scale. Make sure the scale is empty.
18. Press ZERO to change the value to 1.
19. Press PRINT        . The display will count down from 10-1 while the scale is
calibrating zero. When the display shows 0 the zero calibration is complete.
OP-900 offers 2 calibration methods, Single Point which uses one weight to calibrate or
Linear Calibration, which uses multiple (2-7) weights for a more accurate calibration.
20. Press PRINT      to continue. The display will now show C06.
21. Press PRINT      to access the C06 channel. The display will show [C6    0].
22. The C6 channel is used to calibrate the scale with a known weight. Press ZERO to
    set the value of C6 to [C6     1]. Press PRINT . The display will flash SPAN, and
    then show [######].
10
23. Enter the calibration weight value you will use (at least 10% of max capacity you set
    in C04 by using UNIT and COUNT to move the cursor left and right, and TARE
    and ZERO move the values down and up.
24. Place the calibration weight you have on the empty scale and press PRINT        .
25. The scale will count down from 10 to 0. Once 0 has been reached, the display will
    show CALEnd.
26. Press PRINT      to continue. The display will now show C07.
27. Press ACCUM        to save and exit the setup menu.
28. The scale has now been calibrated. The display will show the value of the calibration
    weight on the scale.
29. If the scale does not show the value of the calibration weight, check that the feet on
    the platform are not screwed in too tightly, and verify that the platform is level.
30. Unload the scale; the display should read 000000.
31. If the scale does not display 000000, check that the feet on the platform are not
    screwed in too tightly, and verify that the platform is level.
                                                                                      11
      INDICATOR PARAMETER SETTINGS
The parameter settings menu has a calibration section (C01 to C07 explained above) and a
parameter settings section (C08 and up).
To access the calibration section the seal switch (located at one corner of the PCB) must be
OFF. This will allow access to all C01 and up settings. If the seal switch is ON, then only C08 and up
can be accessed by the user. If you break the official seal by opening the back of the indicator to
access the seal switch, you may need to have the indicator recertified. Be sure to adjust the seal
switch back to the original setting after calibration/configuration has been performed.
12
Function             Parameter Settings/Options
Restore Default                0 = do not restore
Settings               C07     1 = restore to default settings
                               0 = turn off warning tone
Warning Tone           C08     1 = turn on warning tone
                               0 = turn off auto power off
Automatic                      10 = power off automatically if no change within 10 minutes
Power Off              C09     30 = power off automatically if no change within 30 minutes
                               60 = power off automatically if no change within 60 minutes
                               LED Version OP900A:
                               0 = turn off power saving setting
                               3 = turn off display if no change within 3 minutes
                               5 = turn off display if no change within 5 minutes
Power Saving Mode      C10     LCD Version OP900B:
                               0 = turn off the backlight
                               1 = backlight only when the weight changes or keyboard is pressed
                               2 = constant backlight
                               0 = turn off hold function
                               1 = Peak hold - Grabs the highest weight
                               2 = Manual hold - Grabs the current weight
                               3 = Auto hold - Automatically holds data when stable
Hold Function          C11     4 = Average hold - for animal weighing, averages the weight from a
                                   sample of 3 seconds
                               5 = Auto Average hold - Average hold without the need to press the
                                   hold key
Unit Conversion        C12     * See table 2
Upper Limit Alarm      C13     Set upper limit within the max. capacity
Lower Limit Alarm      C14     Set lower limit within the max. capacity
Inner Code Display     C15     check the inner code (raw data)
Set Date               C16     Set date from left to right: year/month/day
Set Time               C17     Set the time from left to right: hour/minute/second
                               Set the serial interface data output method:
                               0 = Turn off serial interface data output
                               1 = Continuous sending mode, for remote display
                               2 = Print to paper thermal ticket printer
Communication                  3 = Command request mode, for computer.
Setting                C18     4 = PC continuous sending mode, for computer
                               5 = PC/remote display, continuous sending mode
                               6 = Print to adhesive label thermal printer
                               7 = Print to Zebra/large adhesive label thermal printer
                               8 = Reserved
                               0 = 1200 (for OP-910 remote display)
                               1 = 2400
Baud Rate              C19     2 = 4800
                               3 = 9600 (for all printers and OP-910X, OP-910XL)
                               4 = 14400
                                                                                             13
Function                Parameter Settings/Options
                                  0 = turn off manually zero setting
                                  1 = ±1% max capacity
                                  2 = ±2% max capacity
Manual
Zero Range                C20     4 = ±4% max capacity
                                  10 = ±10% max capacity
                                  20 = ±20% max capacity
                                  100 = ±100% max capacity
                                  0 = no initial zero setting
                                  1 = ±1% max capacity
                                  2 = ±2% max capacity
Initial Zero Range        C21     5 = ±5% max capacity
                                  10 = ±10% max capacity
                                  20 = ±20% max capacity
                                  100 = ±100% max capacity
                                  0= turn off zero tracking
                                  0.5 = ±0.5d                                 d = division
                                  1.0 = ±1.0d
                                  2.0 = ±2.0d
Zero Tracking             C22     3.0 = ±3.0d
                                  4.0 = ±4.0d
                                  5.0 = ±5.0d
                                  Note: the zero tracking range can not be bigger than manual zero
                                  range
                                  0 = turn off zero tracking time
                                  1 = 1 second
Zero Tracking Time        C23     2 = 2 seconds
                                  3 = 3 seconds
                                  00 = turn off overload range
Overload Range            C24     01-99d = overload range setting             d = division
                                  0 = -9d
                                  10 = -10% max. capacity
Negative Display          C25     20 = -20% max. capacity
                                  50 = -50% max. capacity
                                  100 = -100% max. capacity
                                  0 = quick
Standstill Time           C26     1 = medium
                                  2 = slow
                                  1 = 1d                                        d = division
                                  2 = 2d
Standstill Range          C27     5 = 5d
                                  10 = 10d
                                  0 = turn off dynamic filter
                                  1 = 1 digital filter strength
Digital Filter                    2 = 2 digital filter strength
(for filtering moving             3 = 3 digital filter strength
weight such as            C28     4 = 4 digital filter strength
animals)                          5 = 5 digital filter strength
                                  6 = 6 digital filter strength
                                  Note: The higher the number, the higher the filter strength
14
Function                 Parameter Settings/Options
                                     0    = turn off noise filter
                                     1    = 1 digital filter strength
Noise Filter
                           C29       2    = 2 digital filter strength
                                     3    = 3 digital filter strength
                                     0    = yy.mm.dd
                                     1    = mm.dd.yy
Print Time and Date        C30       2    = dd.mm.yy
                                     3    = yy.mm.dd
                                     0    = 0 - 5V ouput
Analog Output Setting      C31       1    = 4 - 20mA output
Calibrate Current          C32       4    - 20mA current
                                     0 = turn off relay output
                                     1 = turn on relay output function 1
Relay Output Setting       C33       2 = turn on relay output function 2
                                     3 = Reserved menu
Gravity of Calibration               9.7000 - 9.9999
Location                   C36
Gravity of Destination     C37       9.7000 - 9.9999
Version No.                C38
                                     0    = auto mode
Print Mode                 C41       1    = gross mode
                                     2    = tare mode
Print Carriage Return      C42       0    - 9 (How much space between print outs)
Space Print                C43       0 - 9 (Where the data prints on the paper: 0 = left ; 9 = right)
                                     0    =   do not print the date
Date Print                 C44       1    =   print the date
                                     0    =   do not print the time
Time Print                 C45       1    =   print the time
                                     0    =   1200 (for remote display)
Baud Rade for                        1    =   2400
2nd RS232                  C48       2    =   4800
                                     3    =   9600
16
                    HELPFUL DEFINITIONS
Division: The amount of increments a scale offers. How accurate the scale can be
Initial Zero Range: The percentage of weight allowed on the scale when indicator is powered
on that will automatically zero.
example: If initial zero range is set to 10% of the max. capacity and your max. capacity is
100lbs, you can place up to 10lbs of weight on the scale and when the indicator is powered on, it
will automatically zero out the weight.
Manual Zero Range: The percentage of weight allowed on the scale where the indicator will let
you manually zero (anything above this percent will be tared)
Zero Tracking Range: A subset to the manual zero range; if the weight on the scale
is not stable, the zero tracking range still allows you to zero within a set division of the
scale
Zero Tracking Time: A subset to the zero tracking range, it is the time allowed for the
scale to fall within the zero tracking range tolerance and still qualify to be zero’d
Overload Range: Weight allowance that is out of the set calibrated range. Adds a
tolerance to the calibrated max. capacity without having to recalibrate.
example: If your scale has a max. capacity of 1000lbs with a division of 1 and you set
the overload range to 60, you can add 1060lbs of weight to the scale without it
displaying an error code
Negative Display: How far you can go in the negative direction before displaying an
error code
Standstill Range: How much the scale can fluctuate before being determined stable
Digital Filter: For filtering moving weight, such as animals, It changes how sensitive
the scale is to variations in movement.
Noise Filter: A filter for how susceptible the scale is to general variations
                                                                                             17
                               CONNECTORS
18
                 FIGURE 3: INNER TERMINAL BLOCK CONNECTION DIAGRAM
                                                                                        19
        RS232 SERIAL OUTPUT FORMAT
Follow the pin out of Table 5 below to connect the indicator the RS-232 Serial device
The serial output format depends on the settings for parameter C18. The serial output
consists of a string of ASCII characters. Here is a list of the serial parameters
   ●● 8 data bits
   ●● 1 stop bits
   ●● No parity
   ●● No handshaking
Note: With the RS232 Output option we have data logging software available as seen in
Figure 5.
20
Remote Display Continuous Sending Mode (C18=1)
For use with a Scoreboard/Remote Display
Note: Baud Rate must be set to 1200 (C19 = 0)
                                                State A
                                               Bits0,1,2
           0                        1                                2                   Decimal point position
           1                        0                                0                        XXXXXX0
           0                        1                                0                        XXXXXXX
           1                        1                                0                        XXXXX.X
           0                        0                                1                        XXXX.XX
           1                        0                                1                        XXX.XXX
                                 Bits3,4                                                       Division
                 0                                         1                                      X1
                 1                                         0                                      X2
                                                State B
                     BitsS                                                     function
                     Bits0                                                 gross=0, net=1
                     Bits1                                         Symbol: positive=0, negative=1
                     Bits2                                           Overload (or under zero)=1
                     Bits3                                                   dynamic=1
                     Bits4                                                unit: lb=0, kg=1
                     Bits5                                                   Constant 1
                     Bits6                                                   Constant 0
                                                State C
          Bit2                    Bit1                             Bit0                           unit
           0                        0                               0                          Kg or lb
           0                        0                               1                              g
           0                        1                               0                              t
                                  Bit 3                                                       printing=1
                                  Bit 4                                                    Extend display=1
                                  Bit 5                                                       Constant 1
                                  Bit 6                                                       Constant 0
                                                                                                              21
Print Mode (C18 = 2)
For printing on a non-adhesive ticket printer. Parameters 16, 17, 30, & 42-45 all effect
your ticket print out.
Normal weighing ticket printout example:
      Date:05/01/2017
      Time:  11:30:52
      Net:25.6lb
      Tare:10.3lb
      Gross:35.9lb
      Date:05/01/2017
      Time:  11:30:52
n00115.4lb
n00217.2lb
n00335.6lb
Total68.2lb
The R command will trigger the indicator to output the followind data format:
 <STX>      <POL>    XXXXX.XX      <SP>    <LB/KG>     <SP>      <GR/NT>      <CR>        <LF>
22
Computer Continuous Sending Mode (C18=4)
         ,            ,                                                  CR LF
{
                                                              {
 S1           S2          S 3		          Data		                S4
S1:    weight status, ST=standstill, US=not standstill, OL=overload
S2:    weight mode, GS=gross mode, NT=net mode
S3:    weight of positive and negative, “+” or “-”
Data:  weight value, including decimal point
S4:    “kg” or “lb”
CR:    carriage return
LF: line feed
                                                                                                      23
4-20 mA ANALOG OUTPUT (OPTIONAL)
                                                 FIGURE 6. CONNECTION DIAGRAM
The 4-20 mA analog output of the OP-900
scale is a voltage sourcing sensor that will
output current which is proportional to the
calibrated scale’s weight range (i.e. 4 mA = 0
LBS and 20 mA = 10,000 LBS). It is
important to note, that this is a sourcing
output sensor (i.e. the sensor will source
9VDC with an output current range of 4-20
mA). Unlike many other “loop powered” type 4-20 mA
sensors, an external supply voltage should not be connected
to the unit’s 4-20 mA circuit, however, the ground connection
(pin 6) of the DB-9 connector will need to be connected to
the same ground as the data acquisition device which will be
responsible for interoperating the 4-20 mA signal. This ground
connection is imperative, as both the data acquisition device’s
power supply and the scale’s internal 9VDC power supply will
need to be on the same ground plane for the output current to
be synced and measured correctly. Please reference Figure 6 as a visual clarification on
how to connect your scales 4-20 mA output to a data acquisition device.
Below is a list of important notes when using the indicator with the 4-20 mA option
  ●● Resolution: 1/1000
  ●● Outside Load: 100-350ohms
  ●● Inside connection: load input port pin “1” of J2, ground port pin “GND” of J2
  ●● Outside connection: load input port pin 1 of DB9, ground port pin 6 of DB9
  ●● To test the connection, connect a 250 ohm load; Locate a volt meter, and probe
     across the 250 ohm load. As the weight input to the indicator varies, the voltage of
     the volt meter will change accordingly
  ●● Pressing the TARE key will reset to output current to 4mA
  ●● 0-20mA output can be set by setting parameter C31 to 0
  ●● Please note that this option will disable the RS232 weigh data output (serial ASCII
     data) that comes in the standard indicator
  Calibration:
  ●● Press PRINT and HOLD key to go into configuration mode
  ●● Go to C32 and press PRINT key
  ●● The display should show [out-4] and output should be at 4mA
  ●● Press the up/down arrow keys and the [out-#] will increase/decrease
Note: X corresponds to the output current. For example if out-12 is displayed, then
12mA should be the output. If not, press the left and right arrow key to adjust/calibrate
24
               RELAY OUTPUT (OPTIONAL)
  ●● The indicator can output 4 signals, which when connected to outside equipment,
     can perform an automatic control function and an upper/lower limit alarm function.
  ●● Change parameter setting C33 following Table 6 below:
Table 7 below shows the DB9 port pinout for the relay output option. Please note that
this is optional and only available for the OP-900 indicator ordered with this option.
The relay output option will disable the RS232 weigh data output (serial ASCII data) that
comes in the standard indicator.
                                                                                               25
                       TROUBLESHOOTING
Error Codes
ERR4 During calibration signal is unstable After the platform is stable, start calibration
26
                                       Q&A
Q:   The scale does not turn on
A:   Make sure the power cord is plugged in, and that there is power. One easy way to
     test this is by connecting another appliance to the same outlet and see if it’s
     operational
Q:   The reading goes negative when a load is applied
A:   Try interchanging the Sig+ and Sig- wiring connected to the load cell and/or
     junction box (if one is used)
Q:   How do I resolve ERR6 error?
A:   Please follow the procedure below:
     1) Turn on the indicator and make sure nothing is on the scale, and that the scale
     is level and not wobbling
     2) Press and hold the “PRINT and HOLD” key simultaneously for a few of seconds
     3) The screen will read “C01”
     4) Using the arrow keys, change C01 to C20. You have to change the 1st digit
     from 0 to 2 first before you can change the 2nd digit 1 to a 0.
     5) Press “PRINT” key to enter C20 parameter
     6) Change the value of C20 on the right to 100 if possible using the up arrow key.
     If 100 is not available change to 20
     7) Press “PRINT” key to enter your selection
     8) The screen will read “C21” now
     9) Press “PRINT” key to enter C21 parameter
     10) Change the value on the right of C21 to 100 if available, 20 if not
     11) Press “PRINT” key to enter your selection
     12) he screen will read “C22” now
     13) Press “TOTAL” key to save and exit
     14) Power the indicator off and then on, and see if this resolves the ERR 6 issue.
     If not, then following the Q&A answers below for resolving “nnnnnn”and “uuuuuu”
     errors
Q:    How do I resolve “nnnnnn” and “uuuuuu” error?
A:   1) Check to see if the cable that runs from the indicator to the junction box is
     damaged. If it is, replace the cable.
     2) Open up the junction box (if available) and check to see if there is any water
     damage. If so, replace the junction box
     3) Make sure all the wires on all 5 terminal blocks (5 wires on each terminal block)
     are not loose. Re-tighten the screws even if the wires seem to be connected
     4) Recalibrate
     5) If steps 1-4 do not work, there is a possibility one or more load cells are
     defective (consult with support@optimascale.com for further instructions)
                                                                                      27