Item
minimum technical requirements Quantity
No
1 Biosafety cabinet class III 2 set
Description
Type: Class 3, ISO 3 air cleanliness
Minimum Technical requirement
Nominal workable size: 5 feet
Usable chamber dimension (HxD), without bench: ≥ 60 x 50 cm
Number of glove port: 4 (2 pair)
Arm cuffs diameter: 300 ± 15% mm
Glove type: Neoprene or Butyl
Airlock: Automatic
Front window glass thickness: ≥ 5 mm, anti-uv protected
Exhausted air vented through a system of double (2) HEPA/ULPA filter
Filter efficiency: 99.999% at 0.1- 0.3 μm or better
Initial air flow volume: ≥ 750 m3/h
Minimum negative pressure: ≥ -120 Pa
Noise: ≤ 65 dBA
Lamp type: Housing LED/fluorescent
Interior illumination: ≥ 750 lux
Work surface material: Stainless steel
Main body construction: Corrosion and chemical resistant material
Display: Digital
Display parameter: Airflow velocity, temperature, filter life, UV-lamp life, usable time
Audio/visual alarm for incorrect airflow velocity
UV lamp: ≥ 20 W, 254 nm
Accessories
(i) Additional Tap (Fuel Gas / Non-Fuel Gas / Vacuum)
(ii) Gas cylinders (tri gas), regulators and fittings
(iii) Passbox/transfer hatch
(iv) Pressure gauge
(v) Vacuum airlock
(vi) Stainless steel suport cart (trolly) with caster wheel with lock feature
(vii) Hanging bar (SS)
(viii) Thimble Duct Exhaust Transition
Power supply: 220-400 V, 50-60 Hz, European plug, singel/three phase main supply
Qty
2 Product name: Bioasafety cabinet Class 2-type 2 2 set
Minimum Technical requirement
Type: Class II-B2
Nominal size: 4 feet
External dimension (with table); DxH: (850 x 2400 mm) ± 10%
The external body construction material should be chemical and corrosive resistant and with anti bacterial
powder coating
Interior should be made up of stainless steel (316 or 304) with side tempered glass to ensure corrosion resistant
Cabinet interior shall process continuous, rounded corners
Base stand: Manufacturer recommended support base stand shall be included
Work surface height: ≥ 700 mm
Maximum window opening: ≥ 425 mm
Safety working height: ≥ 200 mm
Front window moving mechanism: Motorized
Front window glass thickness: ≥ 5 mm, anti-uv protected
Airflow Volume (Inflow): 350 m³/h or better
Airflow Volume (Down flow): 65%: 675 m³/h or better
Airflow Volume (Exhaust): 5%: 360m³/h or better
Should have 2 HEPA/ULPA filter with 0.12 µm particles ≥99.999%
Single/dual fan blower
Air speed should be adjustable to automatically to maintain constant air speed
UV lamp: ≥ 20 W, 254 nm
Lamp type: LED/fluorescent
Interior illumination: ≥ 750 lux
Display: Digital
Display parameters: Panel should display all required parameters
(i) Inflow and down flow velocities
(ii) Filter life
(iii) UV lamp life
(iv) System work time
Audio and or visual alarm facility should be available for
(i) Abnormal airflow velocity
(ii) Filter change
(iii) Front window over height
Noise: ≤ 65 dBA
Should come with minimum
(i) Cabinet with base stand
(ii) LED/fluorescent and UV lamp
(iii) SS water tap x 1 Nos
(iv) Gas tap x 1 Nos
(v) Drain ball valve
(vi) 2 x 13 A multifunctional water proof socket
(vii) Include a leak tight duck: Minimu 2 m ducting should be included
(viii) Leak proof damper in the duct above the cabinet
Power supply: 220-400 V, 50-60 Hz, European plug, singel/three phase main supply
Qty
3 Biosafety Cabinets Class I 2
Minimum Technical requirement
Type: Class 1
Nominal size: 3 feet
External dimension (with table); DxH: (700 x 2400 mm)±10%
The external body construction material should be chemical and corrosive resistant and with anti bacterial
powder coating
Interior should be made up of stainless steel (316 or 304) with side tempered glass to ensure corrosion resistant
Cabinet interior shall process continuous, rounded corners
Base stand: Manufacturer recommended support base stand shall be included
Maximum window opening: ≥ 425 mm
Safety working height: ≥ 200 mm
Work surface height: ≥ 700 mm
Display: Digital
Should have 2 HEPA/ULPA filter with 0.3 µm particles ≥99.97%
Airflow Velocity (Inflow): 0.75 m/s or better
Air speed should be adjustable to automatically to maintain constant air speed
UV lamp: ≥ 20 W, 254 nm
Noise: ≤ 65 dBA
Lamp type: LED/fluorescent
Interior illumination: ≥ 1000 lux
Audio and or visual alarm facility should be available for
(i) Abnormal airflow velocity
(ii) Filter change
(iii) Front window over height
Display: Digital
Display parameters: Panel should display all required parameters
(i) Inflow and down flow velocities
(ii) Filter life
(iii) UV lamp life
(iv) System work time
Front sliding sash interlock with fan and UV lamp
Should come with minimum
(i) Cabinet with base stand
(ii) LED/fluorescent and UV lamp
(iii) SS water tap x 1 Nos
(iv) Gas tap x 1 Nos
(v) Drain ball valve
(vi) 2 x 13 A multifunctional water proof socket
(vii) Include a leak tight duck: Minimum 2 m ducting should be included
(viii) Leak proof damper in the duct above the cabinet
Power supply: 220-400 V, 50-60 Hz, European plug, singel/three phase main supply
All standard accessories shall be supplied