XR115 Operator Manual
XR115 Operator Manual
File List
1. 5475521TPH_s1_r6.pdf
2. 5475521TPH_s1_r6.docx
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document.
Approval Information
Person Action Date and Time
305018930_visveshvaren__m s Approved 09/12/2013 09:01:58 am GMT
305025941_krupa__maria Approved 09/14/2013 05:08:17 am GMT
thomas
This page is generated automatically by the GEHC MyWorkshop System.
Printed documents are for Reference Only and may be out-of-date.
Check the database to ensure you have the correct revision.
Brivo XR 115
Mobile X-Ray System
Operator Manual
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
5475521, Revision 6
See the GEHC Myworkshop System to determine the status of this document.
Brivo XR 115
Operator Manual, English
© 2013 General Electric Company
All rights reserved.
GE Healthcare reserves the right to make changes in specifications and features shown
herein, or discontinue the product described at any time without notice or obligation.
Contact Information
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
Manufactured by:
Web address:
www.gehealthcare.com
Revision History
5475521-1EN, Revision 6
Approved Document - 5475521TPH_r6.pdf Page 8 of 97
Brivo XR 115
Preface
Chapter i
i-1
Preface
Important!
The purpose of this Operator manual is to provide instructions for the proper use of the
system. It is the responsibility of the user to read this manual carefully as it contains
information that is crucial to the safety of the operator and the patient.
All of the information contained herein is based on the current version of the system.
GE Healthcare reserves the right to improve and implement changes to the information
herein to reflect any changes that maybe necessitated by technological enhancements to
the system.
DANGER Danger is used to identify conditions or actions for which a specific hazard is
See the GEHC Myworkshop System to determine the status of this document.
known to exist which will cause severe personal injury, death, or substantial
property damage if the instructions are ignored.
WARNING Warning is used to identify conditions or actions for which a specific hazard is
known to exist which may cause severe personal injury, death, or substantial
property damage if the instructions are ignored.
CAUTION Caution is used to identify conditions or actions for which a potential hazard
may exist which will or can cause minor personal injury or property damage if
the instructions are ignored.
WARNING This X-ray unit must be used in strict compliance with the safety instructions
contained in this manual and must not be used for purposes other than those
for which it was intended.
X-Ray equipment if not properly used may cause injury. Accordingly, the instructions
contained herein should be thoroughly read and understood before you attempt to place
this equipment in operation. GE Healthcare or its authorized representatives will be glad to
assist and co-operate in installing and setting up this equipment for use.
Although the equipment incorporates a high degree of protection against diagnostically
irrelevant X-radiation, no practical design of equipment can provide or ensure complete
protection.
It is important to note that, no mandatory safety procedures can be implemented by GE
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
Healthcare or its authorized representatives to ensure that the operator or other users take
adequate precautions to prevent the possibility of authorized or unauthorized persons
carelessly, unwisely, or unknowingly exposing themselves or others to direct or secondary
radiation.
See the GEHC Myworkshop System to determine the status of this document.
It is important that everyone having anything to do with X-radiation be fully acquainted with
the recommendations as per the local regulation recommendations. For India, refer to Atomic
Energy Research Board (AERB), India regulations and take adequate steps to insure
protection against injury.
WARNING It is assumed that all persons authorized to use the equipment are cognizant of
the danger of excessive exposure to X-radiation and the equipment is sold with
the understanding that the General Electric Company, its agents, and
representatives have no responsibility for injury or damage which may result
from exposure to X-radiation
Various materials and devices are available for protection against radiation. It is urged that
such materials and devices be used to ensure safety of the user and patient.
5475521-1EN, Revision 6
Approved Document - 5475521TPH_r6.pdf Page 11 of 97
Brivo XR 115
This page is intentionally left blank
i-4
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document. Introduction
5475521-1EN, Revision 6
Approved Document - 5475521TPH_r6.pdf Page 12 of 97
Brivo XR 115
Chapter 1
Introduction
1-1
Introduction
1.1. Introduction
It is recommended to keep the Operator Manual accessible and readily available to the user
at all times. It is the responsibility of the facility owning the equipment to periodically review
the procedures and safety precautions provided in the manual. It is important to read and
understand the contents of this manual before attempting to use this product.
It is urged that the operating instructions be studied and followed carefully before any
operation is attempted.
NOTE: The manual is not intended to teach radiology or to aid in making any type of clinical
diagnosis.
It is necessary for the operator to have sufficient knowledge to competently perform various
diagnostic imaging procedures.
The system should be used only by a qualified Radiology Technician, Radiology Technologist,
Radiology doctors and related Health Care Professionals.
Radiology Technologist/Technician shall be qualified as per AERB –India regulations and
should be trained in performing radiography on patients in the department/ward.
Radiology doctors who shall use the system shall complete medical graduation with a post-
graduation certificate or shall be undergoing post-graduation in Radiology Services and
should be trained in performing radiography on patients in the department/ward.
The equipment can be deployed in locations that may include Orthopedic Clinics, Radiology
Imaging Centers, hospital Radiology Departments, or hospital Orthopedic Departments, ICU
(Intensive Care Unit), NICU (Neonatal Intensive Care Unit), PICU (Pediatric Intensive Care
Unit), CCU (Critical care Unit), Emergency Room, Operating Room, Nursing Home, Radiology
department, Outpatient Surgery Clinic and Oncology clinics/hospitals.
For information on safety and safety notices please refer to Chapter 2: Safety and
Regulatory of this manual. This chapter describes in detail, the safety information that has to
be understood by the operator before using the system.
NOTE: Additional information pertaining to safety will also be detailed in applicable sections
See the GEHC Myworkshop System to determine the status of this document.
5475521-1EN, Revision 6
Approved Document - 5475521TPH_r6.pdf Page 15 of 97
Brivo XR 115
This page is intentionally left blank
1-4
Safety and Regulatory
Chapter 2
Safety and Regulatory
This chapter explains the indications and contraindication for use, safety considerations,
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
equipment related and patient related precautions and symbols used for the safe and
successful operation of your equipment.
See the GEHC Myworkshop System to determine the status of this document.
Intensive Care Unit, Cardiac and Operating departments, for patients who have limited
mobility or in cases where it is unsafe or impractical to move them to a traditional RAD room.
The Brivo XR 115 is indicated for taking radiographic exposures of the skull, spinal column,
chest, abdomen, extremities, and other body parts with the patient sitting, standing, or lying
See the GEHC Myworkshop System to determine the status of this document.
WARNING It is forbidden to use this device to take exposures repeatedly and frequently
for the same patient.
CAUTION Always provide necessary radiation protection for your patient, the operator
and bystanders when operating the unit.
Prior to taking an X-ray, ensure that necessary precautions are taken to
prevent injury or damage that may result from exposure to X-radiation.
It is the User’s responsibility to provide means for audio and visual
communication between the Operator and the patient
The equipment is sold with the understanding that GE Healthcare, its agents, and
representatives have no responsibility for injury or damage which may result from improper
use of the equipment.
Always use the smallest possible field of exposure by closing the collimator
diaphragms appropriate. The scatter dose produced depends principally on the
volume of the irradiated object.
Ask visitors to step outside the room during an exposure
When X-raying bed patients, move them away, as far as possible, from nearby
See the GEHC Myworkshop System to determine the status of this document.
patients.
Use gonad shields for patients whenever possible.
CAUTION During an X-ray exposure, if people other than the patient and the operator,
are present in the room, it is the operator/ facility’s responsibility to ensure that
adequate safety practices are followed to prevent unwarranted exposure to x-
radiation.
WARNING While operating or servicing X-ray equipment, always keep a distance not less
than 2 meters from the focal spot and X-ray beam, protect body and do not
expose hands, wrists, arms or other parts of the body to the primary beam.
NOTE: Under most conditions, cumulative radiation dose to the operator will not exceed
recommended maximum permissible levels. However, a qualified radiation expert should
evaluate situations involving frequent exposures using high kVp and mAs technic to
determine if extra protection devices are necessary.
worn on the body. Even though this device only measures the radiation which
reaches the area of the body on which they are worn, they do provide a reasonable
indication of the amount of radiation received.
See the GEHC Myworkshop System to determine the status of this document.
WARNING
This system shall not be used under flammable environment
Preventive maintenance or regular checks may not always help in
identifying potential failure or malfunction of components such as X-ray
tubes etc.
Always be alert to safety when you operate this equipment. You must be
familiar enough with the equipment to recognize any malfunctions that
can be a hazard. If a malfunction occurs or a safety problem is known to
exist, do not use this equipment until qualified personnel correct the
problem.
The system should not be operated when the system components or cable
insulations are damaged or when there is any system components that are loose and
free. Please contact service for safety.
Perform periodic maintenance to ensure continued safe use of the equipment. Follow
recommended preventive maintenance as outlined in Chapter 7
There are no operator serviceable parts or adjustments inside. Only trained service
personnel authorized by GE Healthcare may remove the unit covers and are
permitted access to the internal parts of the equipment and only in accordance with
the instructions contained in the Service Manual.
GE Healthcare will provide necessary diagnostic diagrams, component part lists,
descriptions and calibration instructions to assist the service personnel in parts
repair.
The X-ray system should be wheeled only into rooms or spaces that are sufficiently
bigger in dimension than the system itself. To prevent damage to the system or injury
to the operator, do not force the unit into areas that are too small to accommodate it.
It is recommended that the minimum door size of the room is 6’ x 4’ and the room
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
WARNING Electric Shock Hazard! Do not remove covers. The system contains high voltage
circuits for generating and controlling X-rays. Never touch the patient and any
exposed low safety circuits or connectors simultaneously. Serious injury may
occur.
CAUTION GE Healthcare is responsible for the safety of its products only when
maintenance have been performed by GE Healthcare or its own agreed
distributors.
GE Healthcare cannot be held liable for any malfunction, damage, or danger
resulting from improper use of the system or non-compliance with the rules of
proper maintenance.
CAUTION Electric shock hazard! This equipment must only be connected to supply mains
with protective earth.
Do not remove covers or panels. The acquisition console and cabinets contain
high voltage circuits for generating and controlling X-Ray. Prevent possible
electric shock by leaving covers or panels on the equipment.
For the safety of the patient and operator, only devices that are certified to the
appropriate safety standards for medical devices and in good working
condition is to be used in proximity to this X-ray Unit.
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
To prevent accidental exposures and/or injury ensure that the system safety
circuits and devices must not, for any reason, be moved, modified, or omitted.
The unit should not be used in areas where danger of explosion exists.
CAUTION After positioning the unit, engage the parking brakes. This prevents unintended
movement of the system.
For safe and easy operation, always use the handles provided for moving
and/or re-positioning the system and its components
Observe equipment limitations when moving the mobile system up or down on
an inclined plane. In the transport (parked) position the mobile system can be
driven up to a 10-degree incline.
CAUTION Keep the patient in full view at all times and never leave the patient
unattended.
To avoid patient injury, always assist the patient as needed, from the
beginning to the end of an exam.
Make sure that patient connected lines; tubes, etc. do not become pinched or
pulled.
Thoroughly check there is no interference or possibility of collision between
the patient and other equipment.
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
GE Healthcare has tested this equipment and is in compliance with the applicable
regulations regarding EMC. Refer to Appendix B for information on Electromagnetic
Compatibility, Requirements and tests, which defines the maximum allowed emission levels
from electronic devices and the required immunity from interference caused by externally
generated electromagnetic fields.
It is not, however, possible to exclude radio signals coming from transmitters such as mobile
phones or similar mobile radio devices. These and other transmitting devices, including
those in compliance with the EMC standards, may influence the functioning of medical
apparatus when used in proximity with it.
WARNING Do not use devices which intentionally transmit radio frequency (RF) signals
(cellular phones, transceivers, or radio controlled products) in the vicinity of this
equipment, as it may result in performance outside the published
specifications.
CAUTION Mobile phones must be switched off in zones close to the unit. These rules must
be applied when the unit is switched on i.e. connected to the mains and ready
for use.
If you suspect any electromagnetic interference affecting or caused by the unit,
call service. Portable and mobile RF communications equipment can affect
medical electrical equipment
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
NOTE: This equipment generates, uses, and can radiate radio frequency energy. The equipment may
cause radio frequency interference to other medical, non-medical devices and radio
communications. To provide reasonable protection against such interference, this product complies
with emission limits for Group 1 Class A Medical Devices as stated in EN 60601-1-2. However, there is
no guarantee that interference will not occur in a particular installation.
See the GEHC Myworkshop System to determine the status of this document.
If this equipment is found to cause interference with other devices (which may be
determined by switching the equipment on and off), you should attempt to correct the
problem using one or more of the following measures:
CAUTION The manufacturer is not responsible for any interference caused by use of
interconnect cables other than those recommended by the manufacturer or
unauthorized changes or modifications made to this equipment; which could void
the user’s authority to operate the equipment.
NOTE: The medical staffs in charge of this equipment are required to instruct technologists, patients,
and other people who may be around this equipment, to fully comply with the above requirement.
CAUTION This system source assembly is designed to be used with only the Brivo XR 115
tube and collimator. Replacement of either of these components with different
types may render the system non-compliant to applicable radiation safety
standards and regulations.
Equipment) directive which states that the unit or its components must not be disposed as
unsorted municipal waste and must be collected separately.
Symbol Description
Hand crushing hazard. This symbol indicates that serious injury to the hand may
occur.
Symbol Description
See the GEHC Myworkshop System to determine the status of this document.
The mains ON LED glows green when the Mains wall socket and system rocker
switch is switched ON.
The POWER ON button is used to turn ON the system.
The fault LED glows RED when there is an error in the system.
Pressing the Reset button will bring back the technique parameters to its
previously set values.
Symbol Description
This symbol indicates the equipment is suitable for alternating current only.
2.11.4 Ground
Table 2-4 Ground
Symbol Description
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
Protective earth (ground) identifies any terminal which is intended for connection
of an external protective conductor to protect against electrical shock in case of a
fault.
See the GEHC Myworkshop System to determine the status of this document.
2.11.5 Collimator
Table 2-5 Collimator
Symbol Description
2.11.6 X-radiation
Table 2-6 X-ray emission
Symbol Description
SYSTEM RATING
COLUMN
LABEL
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document.
Total Serial
Manufa Equipment Manufact. Model
unit No. of
cturer classification Date no.
weight system
COLUMN
INCLINE LIMIT
LABEL
WARNING LABEL
This X-ray Unit may be dangerous to patient and operator unless
safe exposure factors, operating instructions and maintenance
schedules are observed. To be used by authorized personnel only.
See the GEHC Myworkshop System to determine the status of this document.
TUBE HEAD
X-RAY TUBE HEAD
RATING LABEL ASSEMBLY
Caution: X-rays
TUBE
IDENTIFICATION X-RAY TUBE
LABEL
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
COLLIMATOR
KNOB CONTROL COLLIMATOR
LABEL (x2) UNIT
Note: Collimator knob on the tube head can either represent the
image as above (without graduation above the knob) or as below
image (with graduations above the knob).
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
COLLIMATOR COLLIMATOR
KNOB CONTROL UNIT
See the GEHC Myworkshop System to determine the status of this document.
LABEL (x2)
The collimator enclosure may get hot during extended use. To avoid
burn injuries, exercise caution.
See the GEHC Myworkshop System to determine the status of this document.
LH Bracket RH Bracket
A tape is wound around each end of the bracket that holds the tube
head. The tape markings provide an indication of the degree to
which the tube assembly has been moved in the backward and
forward direction as shown in Figure 4-6b)
ONE LOGO ON
THE TUBE HEAD
GE LOGO AND TWO LOGO S
ON EITHER SIDE
OF THE CABINET.
5475521-1EN, Revision 6
Approved Document - 5475521TPH_r6.pdf Page 34 of 97
Brivo XR 115
This page is intentionally left blank
2-19
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document.
Safety and Regulatory
5475521-1EN, Revision 6
Approved Document - 5475521TPH_r6.pdf Page 35 of 97
Brivo XR 115
This page is intentionally left blank
2-20
General Information
Chapter 3
General Information
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
This chapter explains some of the basic operations and features of the Mobile X-ray Unit.
See the GEHC Myworkshop System to determine the status of this document.
CAUTION The system does not belong to the category of equipment designed for
continuous operation.
6 Console
4
6
3
7
8 2
9 1
10
11
NOTE: Location and appearance of some of the system components is subject to change.
5
6 4
7
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
8 3
9
See the GEHC Myworkshop System to determine the status of this document.
10
CAUTION The system is not used in direct contact with the patient; however, accidental
contact of some unit parts with the patient and operator is possible
NOTE: Contact with the operator is strictly for reasons linked to the use of the equipment (normal
operation). Contact of the system with the patient is non-invasive
11 10 9 8 7 6 5 4 3 1 2
See the GEHC Myworkshop System to determine the status of this document.
NOTE: If the RESET key is pressed when the system is in stand-by condition (non-error condition); the
input voltage to the system “Vin” will be displayed on the console. This information is for use of
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
authorized Service Personnel only. Pressing the reset key again, will return the system to stand-by
mode.
EXPOSURE
When the system is ready for exposure, the ‘Ready’ LED
READY
glows Green
Indicator
NOTE:
a) The system needs a minimum room lighting of 60 to 80 lux for normal visibility to
move/or use the system inside the ward/ICU. Outside the room/ward, minimum lighting is
required to move the system within the hospital/ diagnostic center
b) If the system operating room lighting is more than 80 lux, the prep and exposure indications
may not be clearly visible
3.5 Hand-switch
X-ray Exposures are made with the hand switch. The hand switch is a 2 relay, mechanical
switch which operates in 2 levels:
1. PREP
2. EXPOSE
Figure 3-3 Hand switch operation
1at level
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
LEVEL 1: Depressing the hand switch half way downwards until the READY indicator
glows green and an audible beep is heard, prepares the system to READY for exposure.
LEVEL 2: Depressing the hand switch button fully until the READY indicator turns from Green
to Yellow and an audible beep is heard, indicates that X-ray emission is in progress.
NOTE: The switch has to be released only at the end of the beep; which marks the
completion of exposure.
CAUTION
In case, an exposure does not terminate cut-off the power supply to the
equipment immediately.
If the prep and exposure LED does not light up during the procedure, do not use
the system. Log in a service call.
While taking an exposure, ensure that the hand switch passes through each
level of press one after the other, as described above. Pressing the hand switch
in quick succession before Prep is completed; results in a system inhibit error.
The system enters into fault mode (E25 On), with an error code display on the
console, a continuous beep and fault LED glowing red.
NOTE: If the hand switch is released midway before the exposure could be completed, the
system displays the kV and mAs values up to which the exposure occurred. The system
identifies this as an error and goes into fault mode. Error “E57 Abt,” is displayed on the
console with a continuous beep and the fault LED glowing red.
The OFF position is when the Prep/Expose button on top of the hand switch is not in partially
or fully depressed state.
The beep ON duration is equal to the duration of the X-ray ON time if it is greater
than 100msecs.
For all exposures with X-ray ON time less than 100msecs the beep ON duration is
100msecs
The ‘READY’ LED glows Yellow during exposure
See the GEHC Myworkshop System to determine the status of this document.
Long Beep Generated when the system is in fault condition, ‘FAULT’ LED glows RED.
CAUTION If the prep and exposure beep sounds are not heard, log a service call.
If the ambient sound, in the room where exposures are carried out, is greater
than 60dB, there is a possibility that the Prep and/or Exposure beeps may not
be heard.
CAUTION The default technique factors for the system are 60kV and 10 mAs. These
values are placeholders only. No exposures should be made until the user
selects values appropriate for the patient size.
Radiographic
Step Range
parameter
Tube voltage (kV) 2 40kV to 70kV 70kV to 40 kV
Tube voltage (kV) 5 70kV to 100kV 100kV to 70kV
0.1 to 250mAs 250mAs to 0.1mAs
mAs Renard (0.1, 0.5, 1, 1.5, 2, 2.5, 3, 3.5, 4, 5, (250, 200, 160, 125,100, 80, 63, 50,
6.5, 8, 10, 12.5, 16, 20, 25, 32, 40, 40, 32, 25, 20, 16, 12.5, 10, 8, 6.5,
50, 63, 80, 100, 125, 160, 200, 250) 5, 4, 3.5, 3, 2.5, 2, 1.5, 1, 0.5, 0.1)
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document.
5475521-1EN, Revision 6
Approved Document - 5475521TPH_r6.pdf Page 45 of 97
Brivo XR 115
This page is intentionally left blank
3-10
Moving and Positioning
Chapter 4
Moving and Positioning
This section describes how to move the unit and position the X-ray tube for image
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
acquisitions
See the GEHC Myworkshop System to determine the status of this document.
WARNING Ensure that the parking brake is released and swivel arm is locked to the
park/transport position before attempting to move the unit. If not properly
latched, moving the Brivo XR 115 could cause the arm to slide and strike an
operator or patient.
Do not drive or position the Brivo XR 115 unless standing directly behind it.
Failure to do this may result in loss of control causing serious injury to the
operator and/or equipment damage.
WARNING While moving or maneuvering the unit through elevators, doorways or aisles do
not use the unit as an anchor to open doors or keep doorways blocked.
Constant and continuous impact of unwarranted forces on the tube head may
See the GEHC Myworkshop System to determine the status of this document.
Potential Crush hazard: Take care not to drive the system over cables or over a
person’s foot.
Exercise caution that system wires/cables do not come in contact with moving
parts.
To prevent injury to the user due to spillage of liquids on the floor/pathway
exercise caution while moving the system.
NOTE: If the arm appears to sag after positioning at the desired SID, tighten the friction lock as
shown in Figure 4-3c), to maintain the required SID.
WARNING Potential Pinch Point: The area where the tube head connects to the swivel arm
may create a pinch point when the tube is raised. To avoid injury, the operator
should use the handles provided to raise the arm in up/down direction.
CAUTION Make sure that the other end of the power cord cable is connected to the
system as shown in the Figure 4-4.
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
If the power cord is disconnected from the system, the input power supply will
be lost and the mains ON indicator on the console will not glow green.
c) Turn the system ON by pressing the power ON button, on the console. The system
will take few seconds to boot up. In this time, it performs a few diagnostic self-tests to
See the GEHC Myworkshop System to determine the status of this document.
desired angle. This section describes how to setup and position mechanical components of
the Brivo XR 115.
NOTE: Always use both hands to grasp the tube head handles.
See the GEHC Myworkshop System to determine the status of this document.
A goniometer is present on the front face of the monoblock. The goniometer (shown
in Figure 4-7) is an angle indicator that indicates the degree to which the tube head
assembly is rotated in the clockwise or in the anti-clockwise direction as shown in
Figure 4-6 a) above.
Collimator
1. Two Collimator lamp switches are available- one on the console and the other on
the tube head assembly. To activate the light field lamp, press the collimator light
switch:
a. The collimator light field switch is a toggle: The light will turn ON with the first
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
press of the switch and will turn OFF with the second press of the switch. The
field light is timed and will automatically turn off 30 seconds after the first
press of the switch. To activate the field light for another 30 seconds, press
the button again.
See the GEHC Myworkshop System to determine the status of this document.
b. As long as the collimator light is ON, PREP cannot be started. Collimator light
should be switched off before initiating PREP.
c. The collimator light cannot be turned ON while the system is in prep mode.
NOTE:
a) Once the collimator lamp is switched ON, it takes approximately 7 seconds for the
light to reach its maximum intensity
b) Frequent use of the collimator light might lead to heating of the collimator back-
cover
CAUTION Take precaution not to direct the collimator light to the patient’s eye while
radiographing the skull/head. This light intensity may irritate the patient’s
eye. Instruct the patient to keep the eyes closed when the collimator light
is switched ON for positioning.
2. Using the collimator handles, the collimator can be rotated about its vertical axis
along +90° to -90°, as shown in Figure 4-9
CAUTION Make sure that no accessories or materials are located in the path of the
primary X-ray beam during exposure. This could result in bad image quality.
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
Always use the handles to rotate the collimator and tube head, and ensure that
the collimator is grasped in a way such that, the hand is not pinched or crushed
between the handles and the collimator.
While rotating the collimator or the tube head assembly to obtain the right
See the GEHC Myworkshop System to determine the status of this document.
WARNING Take precaution and check if the collimator is intact with the tube head
assembly. If the collimator assembly appears to be disconnected or has an
unintended movement with respect to tube head, do not use the system. Log
a service call.
3. Position the collimator and tube using the field light and cross hair shadow as
guides. The cross hair shows the field center.
4. Use the provided tape to measure SID (Source-to-Image Distance) in the
collimator.
CAUTION Use the largest SID appropriate for the anatomy being imaged, to keep the
patient dose as small as possible
5. Collimate by adjusting the size of the field with the two control knobs located on
the collimator.
a. At a SID of 40” (102cms), field sizes starting from 0”x0” up to 17”x17” (43x43
cm) can be obtained.
b. The knob selectors indicate the field size for a selected SID.
c. Dial numbers on the calibrated scale are for 1000mm and 900mm SID.
NOTE: Dimensions of X-ray field (FOV) and SID are provided on the collimator. Make sure you
are using the right field of X-ray and SID for specific examination. Refer to Section 2.12 - Light
field/X-ray field label for details
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
6. To allow for optional use of spectral filters, the collimator is provided with an
accessory rail, where filters can be inserted.
See the GEHC Myworkshop System to determine the status of this document.
5475521-1EN, Revision 6
Approved Document - 5475521TPH_r6.pdf Page 55 of 97
Brivo XR 115
This page is intentionally left blank
4-10
Image Acquisition
Chapter 5
Image Acquisition
This section details the process of acquiring images using the cassette.
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document.
NOTE: Do not raise the swivel arm when the system is placed close to the bed. This
may cause the tube head to hit the patient bed. Raise the swivel arm before reaching
close to the patient bed.
3. Connect the power cable to the wall mains power socket and turn ON the power.
CAUTION Take precaution not to pull out the cable connector from the system,
while plugging the power cord to the mains AC outlet.
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
A trip hazard could be created when the system’s power supply cord is
connected to the AC mains. Exercise caution while moving or walking in
the vicinity of the powered system.
To prevent damage to the system power supply cord, ensure that heavy
or sharp objects are not dropped on the cable
See the GEHC Myworkshop System to determine the status of this document.
4. Turn ON the rocker switch and the power ON button on the console
5. Position the cassette under the patient.
6. Unlock the foot brake and position the tube head in line to the patient and the
cassette. Confirm or adjust the SID as required using the measuring tape.
7. Set the Field of View to the required anatomy using the collimator knob.
8. Lock the foot brake.
9. Set the KV and mAs parameters as appropriate for the exam.
10. Instruct the patient appropriately before exposure.
11. Perform exposure using the hand switch.
a. Depress the hand switch partially (up to level I as shown in Figure 5-1 below) to
initiate prep. Wait until the READY LED glows green and a single audible beep is
heard.
b. This indicates that the system is ready to produce X-rays Press the switch
completely (up to Level II as shown in Figure 5-1 below) for exposure. X-ray ON
LED will glow yellow and will be continuously lit throughout the exposure time
along with an audible beep
12. After exposure is completed, delivered KV and mAs values will be displayed for up to 5
seconds
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
13. After each exposure is complete, the system enters into a cooling mode, indicated by
See the GEHC Myworkshop System to determine the status of this document.
the message “CLG Prd” which will be displayed in the KV and mAs display fields
CAUTION Do not power off the system while generator is in cooling mode. There is a
cooling mechanism incorporated inside the generator, which requires a
specific time, depending on the exposure time, to cool the generator
electronics before performing the next exposure.
14. After the cooling period elapses, the system console will display previously set
technique parameters
NOTE: Cooling period for each exposure will vary depending on the set exposure time
with a duty cycle of 1:30.
15. If required to perform an exposure on the same patient, repeat the steps from 5.
16. To power off the system, press the power OFF button on the console and turn OFF the
rocker switch.
17. Power OFF the mains supply, unplug the cable from the mains socket and wind the
cable around cable winder on the system.
CAUTION Ensure that the power cord is completely wound around the cable winder.
If not, the cable can drop down, and can come under the wheels of the
system during transport.
5475521-1EN, Revision 6
Approved Document - 5475521TPH_r6.pdf Page 60 of 97
Brivo XR 115
This page is intentionally left blank
5-5
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document. Image Acquisition
5475521-1EN, Revision 6
Approved Document - 5475521TPH_r6.pdf Page 61 of 97
Brivo XR 115
This page is intentionally left blank
5-6
Troubleshooting
Chapter 6
Troubleshooting
This section describes about the errors code, error display description along with the recommended
action that can be performed by the user before calling for service.
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document.
WARNING If the recommended action provided does not help in restoring the normal
functionality of the system, do not attempt to open the covers or repair the
system.
29 E40 IuU Input under Voltage These faults are generated when
defined operating specifications/
conditions are not met.
30 E41 IoU Input Over Voltage These faults will be automatically
cleared if operating conditions
comes back to defined specifications
31 E42 CAP Calibration Problem
(Maximum Iteration
exceeded)
32 E43 oCC Over Current Cathode
43 The unit does not power on when mains is switched Check if Power ON switch back light
ON illuminates
If not, there may be a loose contact
at the wall socket end or the wall
outlet is not receiving power.
Check local electrical circuit for trips.
If Power ON switch is illuminating:
(a) check that the Mains Cable is
properly connected to the power
socket on the system
(b) Switch off mains power. Wait for 2
minutes.
Switch ON mains power.
If the problem persists, call service.
44 No X-ray image even though the unit indicates In case of manual processing, verify
normal exposure. film development and storage
method. The films could be damaged
or the chemicals could be
contaminated. If these factors are
satisfactory, call for service.
5475521-1EN, Revision 6
Approved Document - 5475521TPH_r6.pdf Page 67 of 97
Brivo XR 115
This page is intentionally left blank
6-6
Maintenance
Chapter 7
Maintenance
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
This chapter explains the Maintenance process for your system. To assure continued performance of
this X-ray equipment, a periodic inspection program must be established.
7.1 General
In order to assure continued safe performance of this x–ray equipment, following are some
points to remember:
Regular periodic maintenance can help identify any potential problem.
Based on frequency of usage, it is the owner’s responsibility to ensure that periodic
maintenance is arranged.
Apart from routine maintenance, any abnormal noise, vibration or unusual
performance should be reported immediately to a GE representative.
Before calling for service, be sure the equipment is being operated in accordance
with the foregoing instructions.
Safe equipment performance also requires the use of service personnel specially trained on
this medical x–ray equipment. GE and its associates, maintain a world–wide organization of
stations from which one may obtain skilled x–ray service. If desired, arrangements can
usually be made to furnish periodic and/ or emergency service on a contract basis. A GE
representative will be glad to discuss this plan.
CAUTION Never use cleaners or solvents of any kind if you are uncertain of the nature of
the cleaning agent
Ensure that no liquid seeps into the unit, so as to avoid short-circuiting or
corroding the electrical and electromechanical parts.
Personal caution should be used when removing any accumulating foreign
Material/debris with live system components
NOTE:
a) The equipment need not be kept sterile.
b) Never use solvents or flammable solutions to clean hand switch.
c) Never use dripping cloth (or) immerse hand switch in water or cleaning solution.
d) Never use chemicals for cleaning the cables, as this may degrade the cable insulation
Check the hand switch for accumulation of dirt, contrast material, damage
to the plastic housing or loosening in the plastic housing
Check for any debris accumulation in the imaging pathway (collimator). If
any debris found, do not use the system. Call service.
Check effectiveness of the brakes
Every 6
months Check handle functionality – System push handle, Tube head handle,
Collimator handle, Friction lock, swivel arm handle and cassette bin handle
Repeatedly press the hand switch buttons (Prep and exposure) to check
that the switch moves smoothly from inactive to active positions.
Yearly
Log a service call with GEHC to replace the hand switch, if the buttons
motion is not smooth or if the buttons are getting stuck.
In the event of any observation of any fault, switch off the unit. Perform the checks as given
above.
NOTE: Please note that the System should be used only after the fault condition is reset.
CAUTION Only trained and qualified service personnel should be permitted access
to the internal parts of the equipment.
When re-starting the system after long storage (more than 3 months), perform the below
mentioned procedure to ensure the tube head operates at its optimum.
Step 1- Set the kV and mAs to the minimum value (40kV, 0.1mAS).
Step 2- Close the Collimator shutters.
Step 3- Take 5 exposures.
Step 4- Increase kV in steps of 10 kV and repeat step 3 till 100kV.
CAUTION Do not hang external loads or weights on the tube head or swivel arm. The arm
and base units are designed for its own weight and may not hold an additional
weight.
Do not step or stand on the base unit. Damage to equipment or injury to
personnel may occur if the unit is stepped or stood upon.
5475521-1EN, Revision 6
Approved Document - 5475521TPH_r6.pdf Page 73 of 97
Brivo XR 115
This page is intentionally left blank
7-6
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document. Specifications
5475521-1EN, Revision 6
Approved Document - 5475521TPH_r6.pdf Page 74 of 97
Brivo XR 115
Appendix A
Specifications
A-1
Specifications
5475521-1EN, Revision 6
Approved Document - 5475521TPH_r6.pdf Page 77 of 97
Brivo XR 115
This page is intentionally left blank
A-4
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document.
5475521-1EN, Revision 6
EMC Compliance Statement
B-1
EMC Compliance Statement
EMC Compliance Statement
B-1 Introduction
This equipment complies with IEC 60601-1-2 Edition 3.0 (2007-03) EMC standards for
medical devices.
This equipment generates, uses, and can radiate radio frequency energy. The equipment
may cause radio frequency interference to other medical and non-medical devices and
radio communications.
To provide reasonable protection against such interference, this product complies with the
radiated emission standard limits as per CISPR11 Group1 Class A standard limits. However,
there is no guarantee that interference will not occur in a particular installation.
If this equipment is found to cause interference (which may be determined by turning the
equipment on and off), the user (or qualified service personnel) should attempt to correct the
problem by one or more of the following measure(s):
Reorient or relocate the affected device(s)
Increase the separation between the equipment and the affected device (see
recommended separation distances in Table B-4)
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
Power the equipment from a source different from that of the affected device
Consult the point of purchase or service representative for further suggestions
The manufacturer is not responsible for any interference caused by using other than
recommended interconnect cables or by unauthorized changes or modifications to this
See the GEHC Myworkshop System to determine the status of this document.
equipment.
Unauthorized changes or modifications could void the users' authority to operate the
equipment.
All interconnect cables to peripheral devices must be shielded and properly grounded,
except when technologically prohibited. Use of cables not properly shielded and grounded
may result in the equipment causing radio frequency interference.
These systems are intended for use in non-domestic environments, and not directly
connected to the Public Mains Network that supplies buildings used for domestic purposes.
The magnetic field environment from a MRI device located nearby is a risk of interference.
All of the above are required to achieve the Electromagnetic Compatibility for a typical
installation of the system. Further detailed data and requirements are described in the Use
recommendations.
NOTE: This system complies with above mentioned EMC standard when used with supplied cables. If
different cable lengths are required, contact a qualified service representative for advice.
Electromagnetic Emission
Table B-1 Brivo XR 115 Electromagnetic Emission
Guidance and manufacturer’s declaration – Electromagnetic Emissions
The BRIVO XR115 is intended for use in the electromagnetic environment specified below.
The customer or the user of the BRIVO XR115 should assure that it is used in such an environment.
Electromagnetic Immunity
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
Conducted RF 3 Vrms,
6
800 MHz to 2.5GHz
A Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless)
telephones and mobile radios, amateur radio, AM and FM radio broadcast, and TV
broadcast cannot be predicted theoretically with accuracy. To assess the
electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey
should be considered. If the measured field strength in the location in which the system is
used exceeds the applicable RF compliance level above, the system should be observed
to verify normal operation. If abnormal performance is observed, additional measures
may be necessary, such as re-orienting or relocating the system.
B Over the frequency range 150 kHz to 80 MHz, field strengths should be less than V1.
These guidelines may not apply in all situations. Electromagnetic propagation is affected
by absorption and reflection from structures, objects, and people.
Recommended separation distances between portable and mobile RF communications equipment and BRIVO
XR115
The BRIVO XR115 is intended for use in an electromagnetic environment in which radiated RF disturbances are
controlled. The customer or the user of the system can help prevent electromagnetic interference by
maintaining a minimum distance between portable and mobile RF communications equipment (transmitters)
and the BRIVO XR115 as recommended below, according to the maximum output power of the
communications equipment.
Rated Maximum Output Power (P) Separation distance according to frequency of transmitter (in meters)
of Transmitter Watts(W) 150 kHz to 80MHz 80 MHz to 800 MHz 800 MHz to 2.5 GHz
V1 is in V and E1 is in V/m.
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption
and reflection from structures, objects, and people.
5475521-1EN, Revision 6
EMC Compliance Statement
B-6
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document. APR values
5475521-1EN, Revision 6
Approved Document - 5475521TPH_r6.pdf Page 84 of 97
Brivo XR 115
Appendix C
C-1
APR values for Brivo XR 115
APR values
NOTE: The above mentioned exposure techniques are for reference only. It is the
responsibility of the radiographer to make sure the exposure factors are correct before
exposing
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
Exposure factors are determined depending on the patient size and exposure conditions
(with grid or without grid).
The techniques guided above are without the use of a grid, on a Computed Radiography (CR)
system. If a gridded cassette is used, the exposure technique has to be increased to one step
See the GEHC Myworkshop System to determine the status of this document.
5475521-1EN, Revision 6
Approved Document - 5475521TPH_r6.pdf Page 87 of 97
Brivo XR 115
This page is intentionally left blank
C-4
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document.
5475521-1EN, Revision 6
Installation Instructions
D-1
Appendix D
Installation Instructions
D-2 Safety
Please refer to Chapter 2 Safety and Regulatory provided in this Operator Manual.
CAUTION Take necessary safety precautions while lifting or moving heavy objects during
installation.
Figure D.1
See the GEHC Myworkshop System to determine the status of this document.
a. Remove the shrink wrap and cut and remove the straps
b. Remove the top cover of the package
g. Unscrew the fasteners and take out the metal clamp on the front side of the
system as shown in Figure D.3
See the GEHC Myworkshop System to determine the status of this document.
Figure D.3
h. Remove the bolts and take out the metal clamp in the front of the positioner as
shown in the Figure D.4
Figure D.4
i. Remove the 2 rear blocks by tilting the system on either sides as shown in Figure
D.5
Figure D.5
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEHC Myworkshop System to determine the status of this document.
j. Tilt the system as shown in Figure D.6 to remove the blocks below the base of the
positioner.
Figure D.6
k. Release the break and unload the system from the pallet by pulling it over the
ramp as shown in Figure D.7
Figure D.7
l. Visually inspect the system and components for any damage that might have
occurred during shipping.
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
Check & ensure that the Power supply at the Wall Outlet is as per
recommendation in Appendix A, Section A-4 of the Operator Manual.
Ensure that the Neutral to Earth (Ground) voltage is less than 3 Volts. Connect the
Power Cord to the Wall supply.
Switch on the mains power and the Rocker switch located at the bottom of the
equipment. The rocker switch will glow green
Turn the system ON by pressing the power ON button, on the console. The system
will take few seconds to boot up. In this time, it performs a few diagnostic self-
tests to ensure that the system electronics is functioning.
When the console display lights up with default technique values (60KV and
10mAs) the unit is ready for operation.
3. Initial testing
a. Movements and Positioning
Check for arm balancing in all positions
b. Collimator
Open collimator shutters using the two knobs provided on the collimator.
Press the collimator light switch to turn on the light field.
Adjust the light field on the area to be irradiated, using the collimator knobs
The set light field will be the X-ray field while taking an X-ray. There is an
automatic timer of 30 seconds for collimator light. This will switch off the light
after 30 seconds.
If the light field adjustment is still required, press the collimator light switch
again. To switch off the light field, press the collimator switch.
CAUTION Check for debris in image chain path (Switch on the light and
visually check for any debris in the Collimator)
c. Exposure
NOTE: When re-starting the system after long (more than 3 months) storage,
take the unit through a ‘seasoning’ procedure to ensure the tube head
operates at its optimum. Follow the seasoning procedure as given in Section
7.6 Shipping and Long storage of this Operator Manual.
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
If the date of installation is less than 3 months, skip the seasoning procedure.
Set the technique parameters to 40kV and 0.1mAs.
Close the collimator shutters and depress the hand switch partially (up to level
1 as shown in Figure D.8,) to initiate prep. Wait until the READY LED glows
See the GEHC Myworkshop System to determine the status of this document.
Figure D.8
4. Installation Checklist
5475521-1EN, Revision 6
Installation Instructions
D-9
Revision History
Release/
Rev. Description of Content Reason for change or Change
Page Author Effectivity
No changed Control Number
Date
As per My
1 All Initial Release N/A Lalitha V
Workshop
Error codes and regulatory
Updated figures, error codes As per My
2 All contents were not mentioned in Lalitha V
and regulatory contents Workshop
Rev.1
Alignment and formatting Alignment and formatting error As per My
3 All Lalitha V
maintained in the manual of the text in Rev.2 Workshop
Manufacturers address and
Updated System Rating Label As per My
4 2-14 WEEE symbol updated in Lalitha V
image Workshop
system Rating Label
Collimator knob label (Rear
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
and Front shutter) with As per SCR, the collimator knob As per My
2-17 Lalitha V
graduations included in label is updated Workshop
5 Sec.2.12
Updated System Rating Label ISI mark to be included in the As per My
2-14 Lalitha V
image with ISI Mark System Rating Label Workshop