Lab 1

This document outlines the safety lab procedures for operating Philips CT systems, emphasizing the importance of safety and proper power-up procedures. It includes instructions for powering up the gantry and host computer, as well as emergency protocols and questions to assess understanding. The document is proprietary and confidential, restricted to authorized Philips personnel and licensed representatives.
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
35 views6 pages

Lab 1

This document outlines the safety lab procedures for operating Philips CT systems, emphasizing the importance of safety and proper power-up procedures. It includes instructions for powering up the gantry and host computer, as well as emergency protocols and questions to assess understanding. The document is proprietary and confidential, restricted to authorized Philips personnel and licensed representatives.
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 6

CTI2IL016 Family One - Safety Lab

Proprietary Notice:
This document and the information contained in it is proprietary and confidential information of Philips and
may not be reproduced, copied in whole or in part, adapted, modified, disclosed to others, or disseminated
without the prior written permission of the Philips Legal Department. Use of this document and the
information contained in it is strictly reserved for current Philips personnel and Philips customers who have a
current and valid license from Philips for use by the customer’s designated in-house service employee on
equipment located at the customer’s designated site. Use of this document by unauthorized persons is strictly
prohibited. Report violation of these requirements to the Philips Legal Department. This document must be
returned to Philips when the user is no longer licensed and in any event upon Philips’ first written request.

CSIP Level 2:
This document or digital media and the information contained in it is restricted proprietary and confidential
information that CANNOT BE DISTRIBUTED OR LICENSED AND IS ONLY FOR THE USE OF PHILIPS PERSONNEL
AND LICENSED REPRESENTATIVES. IT IS NOT TO BE DISTRIBUTED, SHARED, COPIED, TRANSFERRED OR
OTHERWISE PROVIDED TO ANYONE ELSE, ELECTRONICALLY OR OTHERWISE.

Warranty Disclaimer Language:


Philips provides this DOCUMENT without warranty of any kind, implied or expressed, including, but not
limited to, the implied warranties of merchantability and fitness for a particular purpose.

Limitations of Liability Language:


Philips has taken care to ensure the accuracy of this document. However, Philips assumes no liability for
errors or omissions and reserves the right to make changes without further notice to any products herein to
improve reliability, function, or design. Philips may make improvements or changes in the product(s) or
program(s) described in this document at any time.

© 2019 Koninklijke Philips N.V. All Rights Reserved.


CSIP Level 2: ATTENTION: THIS PAGE CONTAINS COPYRIGHTED MATERIALS THAT ARE CONFIDENTIAL AND/OR PROPRIETARY. ANY RELEASE OR DISTRIBUTION OF THIS MATERIAL, WITHOUT PERMISSION, IS A VIOLATION OF LAW.

Page 1 of 6
Objectives
When you have successfully completed this lab, you will be able to:
Perform all operations safely as defined in the safety lecture.

Prerequisite Lessons
Safety lecture

Tools/Reference Materials
System Safety Manuals

Time
15 Minutes

Introduction
The purpose of this module is to make the student aware aware of the general and specific safety issues
related to the Access CT, Access Dual CT, MX16 and Incisive CT as well as the system power UP procedure.
Failure to perform procedures in a safe manner can cause personal or equipment damage.
This lab will provide instructions on how to power up the system to the point where the system is ready for
tube warm up.

Procedure Preview
In this lab exercise, the learner will perform system power up, locate the E-stop buttons

Special Instructions/Precautions

<CAUTION:
Damage to the equipment and personnel can occur if you are not careful. If you do not know the outcome of
a motion before initiating, DON’T DO IT! Please seek help from the instructor if unsure.>

© 2019 Koninklijke Philips N.V. All Rights Reserved.


CSIP Level 2: ATTENTION: THIS PAGE CONTAINS COPYRIGHTED MATERIALS THAT ARE CONFIDENTIAL AND/OR PROPRIETARY. ANY RELEASE OR DISTRIBUTION OF THIS MATERIAL, WITHOUT PERMISSION, IS A VIOLATION OF LAW.

Page 2 of 6
Instructions
Power up procedures

1. Use the gantry power switch on the right side of the gantry to
switch on the gantry. This will supply power to all the parts in
the gantry and the patient table. Once the power is supplied
both the GHOST and the RHOST will need approximately 30
seconds to boot up.

2. Press the power “on” button to power up the Host computer.

3. After the HOST computer has booted up, click “OK” to logon to
Windows with username “CT”. Note: always use the username
as CT with no password for Windows login.

4. The screen will become blue and in the lower left corner a start
button can be found. Wait about 2 minutes after the logon to
Windows and then press the start button and select “Access
Dual” or “Access CT”, “MX16” or “Incisive CT” depending on the
scanner type.

5. Next this login screen will pop up. Login as “CT” (no password
required).

© 2019 Koninklijke Philips N.V. All Rights Reserved.


CSIP Level 2: ATTENTION: THIS PAGE CONTAINS COPYRIGHTED MATERIALS THAT ARE CONFIDENTIAL AND/OR PROPRIETARY. ANY RELEASE OR DISTRIBUTION OF THIS MATERIAL, WITHOUT PERMISSION, IS A VIOLATION OF LAW.

Page 3 of 6
Question:

Q1 - Did the gantry and / or the patient table move?

Q2 - If they did move, why did they move?

Q3 - If the patient table is out of the range (at mechanical outer limit – move the table all the way out
from the gantry), what would happen?

Emergency Stop Buttons

6. The red emergency stop buttons will power down the gantry and the patient table. The
emergency stop buttons are located on the gantry control panels (left front and right front,
MX16 also has control panels at the back) and on the Scan Control Box (CT box) on the
operator’s console.

Question

Q4 - Which message is displayed on the CT box (Access CT/MX16) and on the screen for Incisive
CT?

Q5 - Is there any power to the DMS? Why?

8. Reset the system to clear the message.

© 2019 Koninklijke Philips N.V. All Rights Reserved.


CSIP Level 2: ATTENTION: THIS PAGE CONTAINS COPYRIGHTED MATERIALS THAT ARE CONFIDENTIAL AND/OR PROPRIETARY. ANY RELEASE OR DISTRIBUTION OF THIS MATERIAL, WITHOUT PERMISSION, IS A VIOLATION OF LAW.

Page 4 of 6
Question

Q6 - How do you reset an E-stop situation? (Describe the sequence)

Q7 - Can you move the patient table?

Patient Table Clutch Switches

In case of an emergency the patient table can be moved out quickly by punching one of the green
buttons on the side of the patient table.

Question

Q8 - What is the function of the green patient table clutch release buttons under E-stop condition?

---- End of Lab ----

© 2019 Koninklijke Philips N.V. All Rights Reserved.


CSIP Level 2: ATTENTION: THIS PAGE CONTAINS COPYRIGHTED MATERIALS THAT ARE CONFIDENTIAL AND/OR PROPRIETARY. ANY RELEASE OR DISTRIBUTION OF THIS MATERIAL, WITHOUT PERMISSION, IS A VIOLATION OF LAW.

Page 5 of 6
Additional Information
N/A

Content Change Summary

Revision Date Modified By Revision Summary


(dd-MMM-yyyy) (Name)
04-DEC-2019 Frank Broeders Initial Release

Template Change Summary

Revision Date Modified By Revision Summary Approver Name


(yyyy-MMM-dd) (Name) & Date (yyyy-MMM-DD)
Initial Release
2015-JAN-01 D. Preston Changed copyright year to 2015 G. Argento 2015-JAN-01
2015-SEP-18 G. Argento Changed this column’s title G. Argento 2015-SEP-18
2016-JAN-06 D. Preston Changed copyright year to G. Argento 2016-JAN-06
2016
2016-NOV-23 G. Argento Changed copyright year to 2017 G. Argento 2016-NOV-23
2018-MAY-09 D. Preston Changed copyright year to 2018. Sue Hove 2018-May-10
To comply with GCS-ED-PR-00003:
Added “Content Change Summary”
table; changed this table to
“Template Change Summary”;
changed date format to YYYY-MMM-
DD; updated table headers. Changed
<Insert module title> to <Insert
Course code and title>. Deleted
“Healthcare (“Philips”) in legal
statements and made footer
applicable to all pages.

© 2019 Koninklijke Philips N.V. All Rights Reserved.


CSIP Level 2: ATTENTION: THIS PAGE CONTAINS COPYRIGHTED MATERIALS THAT ARE CONFIDENTIAL AND/OR PROPRIETARY. ANY RELEASE OR DISTRIBUTION OF THIS MATERIAL, WITHOUT PERMISSION, IS A VIOLATION OF LAW.

Page 6 of 6

You might also like