Sop 2

This document contains Standard Operating Procedures (SOPs) for various laboratory equipment including actophotometers, ampoule filling machines, ball mills, disintegration test apparatus, double cone blenders, and hand-operated tablet filling machines. Each SOP details the purpose, scope, responsibilities, operation procedures, cleaning, maintenance, and documentation requirements for safe and effective use of the equipment. The SOPs emphasize compliance with institutional guidelines and proper handling techniques to ensure safety and quality in laboratory operations.

Uploaded by

Tribikram
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
31 views28 pages

Sop 2

This document contains Standard Operating Procedures (SOPs) for various laboratory equipment including actophotometers, ampoule filling machines, ball mills, disintegration test apparatus, double cone blenders, and hand-operated tablet filling machines. Each SOP details the purpose, scope, responsibilities, operation procedures, cleaning, maintenance, and documentation requirements for safe and effective use of the equipment. The SOPs emphasize compliance with institutional guidelines and proper handling techniques to ensure safety and quality in laboratory operations.

Uploaded by

Tribikram
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 28

Standard Operating Procedure (SOP) for Actophotometer

1. Purpose

This SOP outlines the procedures for the safe and consistent operation of an actophotometer to
measure locomotor activity in small animals (typically mice or rats).

2. Scope

This SOP applies to all personnel authorized to use the actophotometer within an animal
research facility.

3. Responsibilities

● Researcher/Technician:
○ Conducts actophotometer testing following this SOP.
○ Ensures animal health and well-being during testing.
○ Maintains the actophotometer in good working condition.
● Animal Caretaker: Ensures animal health and well-being before and after testing.

4. Animal Care and Use

● Ensure all procedures comply with your institution's Animal Care and Use Committee
(IACUC) protocols.
● Use only healthy rodents that have been acclimated to the testing environment.
● Minimize stress and ensure proper handling techniques throughout the experiment.

5. Apparatus Preparation

5.1 Cleaning and Disinfection * Clean the actophotometer chamber and activity lanes with a
disinfectant appropriate for your facility. * Allow the chamber to dry completely before use.

5.2 Functional Check * Turn on the actophotometer and ensure all lights and sensors are
functioning properly. * Run a short test cycle (without animals) to verify the activity counter is
working.

6. Animal Preparation

6.1 Acclimatization * Allow the animals to acclimatize to the testing room for at least 30
minutes before testing.

6.2 Marking (Optional) * Depending on the experimental design, some protocols may involve
marking animals with a harmless substance to distinguish individuals during activity tracking
within the chamber. Follow IACUC guidelines for any animal identification methods.

7. Testing Procedure

7.1 Animal Placement * Gently pick up a rodent and place it individually into the designated
activity lane of the actophotometer chamber.

7.2 Testing Session * Start the actophotometer according to the chosen testing duration. * The
instrument will automatically detect and record animal movements (beam breaks) within the
chamber.

7.3 Data Collection * After the testing session, record the total activity count displayed by the
actophotometer for each animal. * Include relevant details like animal ID, weight, and any
observations during testing.

8. Post-Testing

8.1 Animal Recovery * Return the animal to its home cage after testing. * Monitor the animal
for any signs of distress and provide veterinary care if necessary.

8.2 Apparatus Cleaning * Clean the actophotometer chamber and activity lanes with
disinfectant after each testing session.

9. Maintenance

● Perform regular maintenance checks on the actophotometer to ensure proper functioning of


lights, sensors, and activity counters.
● Report any malfunctions or need for repairs to the designated personnel.

10. Reference Documents

● IACUC protocol approval for your specific actophotometer experiment.


● Manufacturer's instructions for the actophotometer model being used.

11. Additional Notes

● This SOP is a general guideline and specific details may vary depending on the
actophotometer model, animal species, and your research objectives.
● Consider familiarizing animals with the testing chamber through a short habituation session
(following IACUC guidelines) before data collection to minimize initial stress.
Standard Operating Procedure (SOP) for Ampoule Filling and Sealing Machine

1. Purpose

This SOP defines the procedures for safe and aseptic operation of the ampoule filling and
sealing machine to ensure consistent, sterile filling and sealing of ampoules.

2. Scope

This SOP applies to all personnel authorized to operate the ampoule filling and sealing machine
within a sterile environment.

3. Responsibilities

● Operator: Follows the SOP for aseptic operation of the machine.


● Supervisor/QA: Ensures the SOP is implemented and adhered to.

4. Pre-Requisites

● Training: Operators must be trained on the specific model of ampoule filling and sealing
machine being used, including aseptic techniques.
● Personal Protective Equipment (PPE): Wear sterile gloves, gown, and head cover during
operation.
● Materials: Ensure availability of sterile ampoules, filling solution, filters, and other required
materials.

5. Machine Setup and Preparation

5.1 Machine Cleaning and Sanitization: * Clean and sanitize the machine exterior with a
designated disinfectant according to manufacturer's instructions. * Ensure all internal
components used for product contact are sterile (refer to manufacturer's guidelines).

5.2 Machine Setup: * Verify gas supplies (e.g., nitrogen, oxygen, LPG) are adequate and
pressure is within specified range. * Set filling volume and sealing parameters according to
product specifications. * Install sterile filters on the filling line. * Load sterile ampoules into the
feeding hopper.

6. Operation Procedure

6.1 Aseptic Practices: * Maintain aseptic technique throughout the process. * Work within a
laminar airflow unit to minimize contamination risk.

6.2 Filling and Sealing: * Start the machine according to manufacturer's instructions. * Monitor
the filling process for proper volume and any irregularities. * Adjust filling parameters as needed
to maintain consistency. * Visually inspect ampoules for cracks, improper filling, or
contamination. * Collect and dispose of rejected ampoules in a designated waste container.
7. Cleaning and Maintenance

7.1 Post-Operation Cleaning: * Turn off the machine and gas supplies. * Disconnect and clean
filling lines and tubing according to aseptic protocols. * Wipe down the machine exterior with a
disinfectant solution. * Dispose of used filters and waste materials properly.

7.2 Scheduled Maintenance: * Perform routine maintenance tasks like lubrication and filter
replacements as per the manufacturer's schedule.

8. Documentation

● Record all relevant parameters like filling volume, sealing settings, and batch number.
● Document any deviations from SOP and corrective actions taken.
● Maintain cleaning and maintenance logs.

9. Reference Documents

● Manufacturer's instructions for the specific ampoule filling and sealing machine model.
● Aseptic handling procedures for your facility.
● Batch production records.

10. Additional Notes

● This SOP serves as a general guideline. Specific details may vary depending on the machine
model, product characteristics, and regulatory requirements of your facility.
● Always refer to the manufacturer's instructions and internal quality control procedures for
additional details and safety protocols.
Standard Operating Procedure (SOP) for Ball Mill

1. Purpose

This SOP outlines the safe and effective operation of a ball mill for grinding and mixing
materials.

2. Scope

This SOP applies to all personnel authorized to operate a ball mill in your facility, including
research labs or production areas.

3. Responsibilities

● Operator: Responsible for following the SOP for safe and proper ball mill operation.
● Supervisor/Lab Manager: Ensures the SOP is implemented and adhered to.

4. Precautions

● Only trained personnel are allowed to operate the ball mill.


● Never exceed the maximum speed or filling capacity for the specific ball mill model.
● Ensure proper ventilation when using the ball mill with dusty materials.
● Wear appropriate PPE like safety glasses, ear protection, and dust masks when necessary.
● Turn off the equipment immediately if abnormal sounds or vibrations are detected.

5. Operation Procedure

5.1 Preparation

● Identify the appropriate grinding media (balls) based on the material being processed.
● Weigh and measure the desired quantity of material to be ground.
● Secure the grinding chamber on the mill base according to manufacturer's instructions.

5.2 Loading

● Turn off the power supply before opening the grinding chamber.
● Load the grinding media (balls) into the chamber.
● Add the material to be ground, ensuring it doesn't exceed the recommended fill level.
● Close the grinding chamber securely.

5.3 Grinding

● Turn on the power supply and ensure all safety interlocks are engaged.
● Set the grinding parameters like speed and grinding time according to your material and
desired outcome.
● Start the grinding process and monitor the operation for any abnormalities.
● Limit grinding time to prevent overheating or damage to the mill or grinding media.
5.4 Unloading

● Turn off the power supply and allow the mill to come to a complete stop before
opening the chamber.
● Carefully open the grinding chamber and empty the contents.
● A sieve or filter may be used to separate the ground material from the grinding media.

6. Cleaning Procedure

6.1 After Operation

● Turn off the power supply before cleaning.


● Wipe down the exterior surfaces of the ball mill with a damp cloth.
● Dispose of residual material according to designated waste procedures.

6.2 Detailed Cleaning (if required)

● For thorough cleaning, disassemble the grinding chamber according to manufacturer's


instructions.
● Wash the grinding chamber, balls, and other components with a suitable solvent or cleaning
solution.
● Rinse thoroughly with clean water and allow everything to dry completely before reassembly.

7. Records

● Maintain a log of ball mill operation, including:


○ Date and time of operation
○ Material being processed
○ Grinding parameters used (speed, time)
○ Any deviations from the SOP and justifications
○ Cleaning and maintenance performed

8. Reference Documents

● Manufacturer's instructions for the specific ball mill model being used
● Safety Data Sheets (SDS) for the materials being processed

9. Additional Notes

● This SOP serves as a general guideline. Specific details may vary depending on the ball mill
model, materials being processed, and your facility's requirements.
● Always refer to the manufacturer's instructions and internal safety protocols for additional
details.
Standard Operating Procedure (SOP) for Disintegration Test Apparatus

1. Purpose

This SOP outlines the procedures for the safe and consistent operation of a disintegration test
apparatus to evaluate the disintegration time of solid dosage forms (tablets or capsules).

2. Scope

This SOP applies to all personnel responsible for performing disintegration testing in a
pharmaceutical or quality control laboratory.

3. Responsibilities

● Lab Technician:
○ Conducts disintegration tests following this SOP.
○ Records all test parameters and observations.
○ Maintains the disintegration test apparatus in good working condition.
● Quality Assurance (QA): Ensures the SOP is followed and reviewed periodically.

4. Equipment and Materials

● Disintegration test apparatus


● Distilled or deionized water (or other media as specified in the monograph)
● Beakers or vessels
● Thermometers
● Analytical balance
● Tablets or capsules to be tested
● Discs (if specified in the monograph)

5. Pre-Test Preparations

5.1 Apparatus Setup * Ensure the disintegration test apparatus is clean and free of dust. * Fill
the water bath with the appropriate medium (usually distilled water) at the specified temperature
(typically 37°C ± 2°C). * Set the stirring mechanism of the apparatus according to the
manufacturer's instructions.

5.2 Sample Preparation * Weigh and record the weight of individual tablets or capsules to be
tested according to the monograph specifications.

6. Test Procedure

6.1 Sample Loading * Place the specified number of tablets or capsules into the individual
baskets or tubes of the apparatus. * Add discs to each tube if required by the specific
monograph for the tested product.

6.2 Test Start * Turn on the apparatus and initiate the test cycle. * The test duration will vary
depending on the monograph specifications for the product being tested.

6.3 Monitoring and Observations * Monitor the disintegration process visually throughout the
test. * Record the time at which all tablets or capsules have disintegrated completely, or have
disintegrated to a specified state mentioned in the monograph. * A tablet or capsule is
considered disintegrated when no whole or nearly whole unit remains, and any remaining
fragments are no more than a specified size (refer to specific monograph).

7. Post-Test Procedures

7.1 Sample Collection * After the test is complete, carefully remove the baskets or tubes from
the water bath. * Collect any remaining fragments of the disintegrated tablets or capsules using
a suitable sieve if required.

7.2 Cleaning * Turn off the apparatus and allow it to cool down. * Empty and rinse the water
bath and all utensils used in the test with clean water. * Clean the disintegration test apparatus
according to the manufacturer's instructions.

8. Data Recording and Reporting

● Record the following information in a designated logbook or electronic data system:


○ Date of testing
○ Sample identification (name of the product)
○ Batch number of the tablets or capsules
○ Individual disintegration time of each tablet or capsule tested (if applicable)
○ Average disintegration time for the sample
○ Any observations made during the test (e.g., deviations from the monograph, incomplete
disintegration)
● Compare the results with the acceptance criteria specified in the product monograph.

9. Calibration and Maintenance

● Regularly calibrate the temperature sensor of the water bath according to the manufacturer's
instructions and using a certified thermometer.
● Perform routine maintenance on the disintegration test apparatus as recommended by the
manufacturer.
● Report any malfunctions or need for repairs to the designated personnel.

10. Reference Documents

● Manufacturer's instructions for the disintegration test apparatus model being used.
● Pharmacopoeia or compendium monographs for the specific tablets or capsules being
tested.
● Laboratory Quality Manual (if applicable)

11. Additional Notes

● This SOP is a general guideline and specific details may vary depending on the disintegration
test apparatus model, the specific dosage form being tested, and regulatory requirements.
● Always refer to the manufacturer's instructions and the relevant pharmacopoeia or
compendium monograph for the most current information and specific test parameters.
Standard Operating Procedure (SOP) for Double Cone Blender

1. Purpose

This SOP outlines the safe and effective operation of a double cone blender to achieve
homogeneous mixing of dry powders or granules.

2. Scope

This SOP applies to all personnel responsible for operating a double cone blender within your
facility, such as production floors or research labs.

3. Responsibilities

● Operator: Follows this SOP for safe and proper double cone blender operation.
● Supervisor/Production Pharmacist: Ensures the SOP is implemented and adhered to.

4. Precautions

● Only trained operators are allowed to operate the double cone blender.
● Never exceed the maximum weight capacity for the specific blender model.
● Ensure proper ventilation when using the blender with materials that may generate dust.
● Wear appropriate PPE like safety glasses, gloves, and dust masks when necessary.
● Turn off the equipment immediately if abnormal sounds or vibrations are detected.

5. Pre-Operation Procedures

5.1 Preparation

● Verify that the blender interior is clean and free of foreign materials.
● Ensure the lower lid is securely fastened.

5.2 Material Loading

● Turn off the power supply before opening the blender.


● Weigh and measure the desired quantities of materials to be blended according to the Batch
Record (BMR).
● Carefully load the materials into the upper lid of the blender.
● Close the upper lid securely.

6. Operation Procedure

6.1 Blending Process

● Turn on the power supply according to the manufacturer's instructions.


● Set the blending time according to the BMR or established process parameters for the
materials being mixed.
● Start the blender and monitor the mixing process visually.
6.2 In-Process Checks (Optional)

● Depending on your material properties and mixing requirements, you may choose to perform
short pauses during the blending cycle to check for homogeneity.
● Turn off the blender before opening the upper lid.
● Carefully open the upper lid and use a spatula to collect samples from different locations
within the blender for visual inspection.
● Resume blending if further mixing is required.

7. Post-Operation Procedures

7.1 Unloading

● Turn off the blender and allow it to come to a complete stop before opening the lower
lid.
● Carefully open the lower lid and discharge the blended material into a designated container.

7.2 Cleaning

● Turn off the power supply before cleaning.


● Wipe down the exterior surfaces of the blender with a damp cloth.
● For detailed cleaning, follow the manufacturer's instructions, which may involve:
○ Disassembling the blender (if applicable)
○ Washing internal components with a suitable cleaning solution
○ Rinsing thoroughly with clean water
○ Allowing everything to dry completely before reassembly

8. Documentation

● Maintain records of double cone blender operation, including:


○ Date and time of operation
○ Batch identification or reference number
○ Materials blended and their quantities
○ Blending time used
○ Any deviations from the SOP and justifications
○ Cleaning and maintenance performed

9. Reference Documents

● Manufacturer's instructions for the specific double cone blender model being used.
● Batch Records (BMRs) for blending operations.

10. Additional Notes

● This SOP serves as a general guideline. Specific details may vary depending on the blender
model, materials being processed, and your facility's requirements.
● Always refer to the manufacturer's instructions and internal quality control procedures for
additional details and safety protocols.
Standard Operating Procedure (SOP) for Hand-Operated Tablet Filling Machine

1. Purpose

This SOP outlines the safe and efficient operation of a hand-operated tablet filling machine to
ensure accurate and consistent filling of capsules with tablets or other solid dosage forms.

2. Scope

This SOP applies to personnel authorized to use a hand-operated tablet filling machine within a
pharmaceutical, nutraceutical, or veterinary production facility.

3. Responsibilities

● Operator: Follows this SOP for proper operation, cleaning, and maintenance of the machine.
Records all relevant data during operation. Reports any malfunctions or deviations from the
SOP to the supervisor.
● Supervisor: Ensures operators are trained on this SOP and verifies adherence.
● Quality Assurance (QA): Approves and reviews the SOP periodically.

4. Materials and Equipment

● Hand-operated tablet filling machine


● Empty capsules (appropriate size for tablets)
● Tablets to be filled (counted and verified for correct dosage)
● Powder tray (optional, for excess powder)
● Tamper (filling plate)
● Locking lever
● Lifting plate lever
● Nylon brush (optional)
● Calibration weight (optional, for specific machine models)

5. Pre-Operation Procedures

5.1 Preparation

● Ensure the workspace is clean, free of debris, and well-lit.


● Verify the machine is in good working order, free of rust or damage.
● Prepare the required number of empty capsules according to the production batch size.
● Count and verify the tablets for filling, ensuring they match the prescribed dosage and are
free of defects.

5.2 Machine Set-Up

● Place the empty capsule tray onto the machine's designated platform.
● Depending on the machine model, you may need to adjust the filling plate (tamper) to the
appropriate height for the capsule size.
6. Operation Procedure

6.1 Tablet Filling

● Locking Lever: Ensure the locking lever is disengaged (usually in the upright position).
● Lifting Plate: Use the lifting plate lever to raise the loading tray with empty capsules. This
allows access to the capsule bodies (without caps).
● Filling the Bodies: Carefully fill the empty capsule bodies with tablets. You may use a
counting tool or a scoop for accurate filling.
● Excess Powder (Optional): If the machine has a powder tray, use it to collect any excess
powder overflowing from the capsules during filling.
● Tampering: Lower the tamping plate (use the lifting plate lever) to compress the tablets
gently into the capsules. Ensure all capsules are filled uniformly. You may use a nylon brush
to brush away any excess powder on the tamping plate before lowering it.
● Calibration (Optional): Some machines may require weight calibration to ensure consistent
filling. Follow the manufacturer's instructions for calibration procedures using a certified
reference weight.

6.2 Capsule Closure

● Locking Lever: Engage the locking lever (usually by pushing it down) to secure the filled
capsule bodies and the caps.
● Lifting Plate: Raise the lifting plate lever again to compress the capsules and securely seal
them.
● Empty Capsule Tray Removal: Carefully remove the tray containing the filled capsules. You
may invert the tray to release the filled capsules.

7. Post-Operation Procedures

7.1 Cleaning

● Disassemble the machine according to the manufacturer's instructions (if possible) for
thorough cleaning.
● Clean all machine surfaces that come into contact with tablets or capsules using a suitable
cleaning solution.
● Pay close attention to the filling nozzle and tamping plate to remove any residual powder.
● Allow all parts to dry completely before reassembling the machine.

7.2. Documentation

● Record the following information in a designated logbook or electronic data system:


○ Date and time of operation
○ Batch number of tablets used
○ Lot number of empty capsules used
○ Number of capsules filled
○ Any observations or deviations from the SOP
○ Cleaning and maintenance performed
8. Maintenance

● Perform routine preventive maintenance on the machine as recommended by the


manufacturer. This may involve lubricating moving parts and inspecting for wear and tear.
● Report any malfunctions or need for repairs to the designated maintenance personnel.

9. Quality Control

● Regularly collect samples of filled capsules for weight verification to ensure they meet the
specified filling requirements.
● Visually inspect finished capsules for proper sealing and absence of cracks or defects.
● Follow established quality control procedures for your product.

10. Reference Documents

● Manufacturer's instructions for the specific hand-operated tablet filling machine model being
used.
● Batch records for the tablets used
● Quality control procedures for your product

11. Additional Notes

● This SOP
Standard Operating Procedure (SOP) for Homogenizer

1. Purpose

This SOP outlines the safe and consistent operation of a homogenizer to achieve uniform
product mixtures.

2. Scope

This SOP applies to all personnel responsible for operating homogenizers within your facility,
including research labs or production floors.

3. Responsibilities

● Machine Operator: Responsible for following the SOP for operating the homogenizer.
● Supervisor/Production Pharmacist: Ensures the SOP is implemented and followed.

4. Precautions

● Only trained operators are allowed to use the homogenizer.


● Never exceed the maximum recommended speed for the homogenizer.
● Avoid contact with moving parts, especially the homogenizing head and shaft.
● Turn off the equipment immediately if abnormal sounds are detected.
● Do not use the homogenizer with volatile or toxic substances unless proper procedures are in
place.

5. Operation Procedure

5.1 Preparation

● Verify that all homogenizer parts are clean and dry.


● Attach the appropriate stage label for the product being processed.
● Refer to the Batch Record (BMR) for specific homogenization parameters like speed and
processing time.

5.2 Operation

● Turn on the homogenizer for a short test run (around one minute) without the product to
ensure proper functioning.
● Begin homogenization according to the BMR instructions.
● Monitor the process to ensure proper mixing and achieve the desired consistency.
● Record all homogenization parameters (time, speed, pressure) and any deviations from the
SOP.

6. Cleaning Procedure

6.1 After Operation


● Label the homogenizer as "TO BE CLEANED".
● Turn off the power supply before cleaning.
● Wipe down the exterior of the homogenizer with a clean, lint-free cloth.

6.2 Detailed Cleaning

● Label the homogenizer again as "TO BE CLEANED".


● Turn off the power supply before cleaning.
● Rinse the homogenizer thoroughly with tap water.
● Wash the homogenizer with a 0.1% liquid soap solution, using a soft brush if necessary.
● Rinse again with tap water followed by deionized water.
● Collect a final rinse sample to ensure there is no residual product remaining.
● Pay close attention to cleaning between the homogenizer jaws to remove any sticky material.
● After successful cleaning, label the homogenizer as "CLEANED".

7. Records

Maintain records of:

● Homogenization parameters for each batch.


● Any deviations from the SOP and justifications for them.
● Cleaning and maintenance logs.

8. Reference Documents

● Manufacturer's instructions for the homogenizer model being used.


● Batch Records (BMRs)

Note: This SOP is a general guideline and specific details may vary depending on your
homogenizer model, application, and regulatory requirements. Refer to your manufacturer's
instructions and internal quality control procedures for additional details.
Standard Operating Procedure (SOP) for Ointment Filling and Sealing Machine

1. Purpose

This SOP outlines the safe, efficient, and consistent operation of an ointment filling and sealing
machine to ensure the quality and sterility of filled ointment tubes.

2. Scope

This SOP applies to all personnel authorized to operate the ointment filling and sealing machine
within a pharmaceutical or cosmetic manufacturing facility.

3. Responsibilities

● Operator:
○ Follows this SOP for the proper operation, cleaning, and maintenance of the machine.
○ Records all relevant data during operation.
○ Reports any malfunctions or deviations from the SOP to the supervisor.
● Supervisor: Ensures operators are trained on this SOP and verifies adherence.
● Quality Assurance (QA): Approves and reviews the SOP periodically.

4. Pre-Operation Procedures

4.1 Preparation

● Ensure the work area is clean and free of debris.


● Verify the machine is properly connected to power and air supply (if applicable).
● Prepare the designated ointment for filling, ensuring it is at the appropriate consistency and
temperature according to the product specifications.
● Gather empty ointment tubes and ensure they are free of defects or damage.

4.2 Machine Setup

● Follow the manufacturer's instructions for setting up the machine for the specific ointment and
tube size being used.
● This may involve:
○ Adjusting the filling nozzle size and position.
○ Setting the filling volume on the control panel.
○ Setting the sealing parameters (temperature, pressure, and dwell time) for the specific
tube material.

4.3 Calibration (if applicable)

● Some machines may require periodic calibration to ensure accurate filling volume.
● Follow the manufacturer's instructions for calibration procedures using certified reference
weights or volume measurements.
5. Operation Procedure

5.1 Machine Start-Up

● Turn on the machine according to the manufacturer's instructions.


● Allow the machine to preheat (if applicable) for the ointment to reach the desired filling
temperature.

5.2 Filling and Sealing

● Load empty tubes into the designated feeder system of the machine.
● Start the filling and sealing cycle.
● The machine will automatically fill each tube with the set volume of ointment and seal the top
of the tube.

5.3. In-Process Monitoring

● The operator should visually monitor the filling process to ensure proper filling and sealing of
each tube.
● Check for any leaks, spills, or incomplete seals and reject defective tubes.
● Record the number of tubes filled at regular intervals.

6. Post-Operation Procedures

6.1 Machine Shutdown

● Turn off the machine following the manufacturer's instructions.


● Allow the machine to cool down (if applicable) before cleaning.

6.2 Cleaning and Sanitation

● Thoroughly clean all ointment contact surfaces of the machine according to the
manufacturer's recommendations and approved sanitizing procedures.
● Pay close attention to the filling nozzle and hopper areas to prevent product buildup.
● Allow all parts to dry completely before restarting the machine.

6.3. Documentation

● Record the following information in a designated logbook or electronic data system:


○ Date and time of operation
○ Batch number of the ointment being filled
○ Lot number of empty tubes used
○ Number of tubes filled
○ Any observations or deviations from the SOP
○ Cleaning and sanitation procedures performed

7. Maintenance
● Perform routine preventive maintenance on the machine according to the manufacturer's
schedule.
● This may involve lubricating moving parts, inspecting seals and gaskets for wear, and
cleaning filters.
● Report any malfunctions or need for repairs to the designated maintenance personnel.

8. Quality Control

● Regularly collect samples of filled tubes for weight or volume verification to ensure they meet
the specified filling requirements.
● Visually inspect finished tubes for proper sealing and cosmetic appearance.
● Follow established quality control procedures for your product.

9. Reference Documents

● Manufacturer's instructions for the ointment filling and sealing machine model being used.
● Batch records for the ointment being filled
● Quality control procedures for your product

10. Additional Notes

● This SOP serves as a general guideline. Specific details may vary depending on the machine
model, ointment properties, and your facility's requirements.
● Always wear appropriate personal protective equipment (PPE) as required for handling the
ointment and cleaning agents.
● Aseptic techniques may be required for sterile ointments. Follow established aseptic
practices as needed.
Standard Operating Procedure (SOP) for pH Meter

1. Purpose

This SOP outlines the proper procedures for the safe, accurate, and consistent use of a pH
meter to measure the acidity or alkalinity (pH) of aqueous solutions.

2. Scope

This SOP applies to all personnel authorized to use a pH meter within your facility, such as
research laboratories, environmental monitoring stations, or industrial process control settings.

3. Responsibilities

● Operator:
○ Follows this SOP for the proper use and care of the pH meter.
○ Records all measurements and observations.
● Supervisor/Lab Manager: Ensures the SOP is implemented and adhered to.
○ Maintains calibration buffers and ensures proper storage.

4. Equipment and Materials

● pH meter with appropriate electrode


● Calibration buffers (typically pH 4.00, 7.00, and 10.01)
● Deionized or distilled water
● Rinse bottle with deionized or distilled water
● Beakers or other suitable containers for samples and buffers
● Kimwipes or other lint-free wipes

5. Pre-Measurement Procedures

5.1 Meter Preparation

● Ensure the pH meter is turned on and in proper working order.


● Consult the manufacturer's instructions for specific model operation.

5.2 Electrode Preparation

● Remove the electrode from its storage solution and rinse thoroughly with deionized water.
● Gently blot the electrode tip with a Kimwipe to remove excess water.

5.3 Calibration

● Standardize the pH meter using standard calibration buffers with known pH values (typically
pH 4.00, 7.00, and 10.01) according to the manufacturer's instructions.
● Follow the specific steps for your meter, which may involve:
○ Immersing the electrode in each buffer solution.
○ Waiting for the stable reading on the meter display.
○ Adjusting the meter's calibration knobs or buttons to match the buffer value.
○ Repeating the process with each calibration buffer.

6. Sample Measurement

6.1 Sample Preparation

● Ensure your sample is well-mixed and free of any large particles that may clog the electrode.

6.2 Measurement

● Rinse the electrode with deionized water and blot dry with a Kimwipe.
● Immerse the electrode tip into the sample solution.
● Allow the reading to stabilize on the meter display.
● Record the pH value of the sample.

6.3 Rinsing and Storage

● Rinse the electrode thoroughly with deionized water after each measurement.
● Store the electrode in the designated storage solution (usually KCl solution) when not in use.
Refer to the manufacturer's instructions for specific storage recommendations.

7. Maintenance

● Regularly inspect the electrode for any signs of damage or clogging.


● Follow the manufacturer's instructions for electrode cleaning and maintenance procedures.
● Report any malfunctions or need for repairs to the designated personnel.
● Maintain calibration buffers with fresh solutions according to their expiry dates.

8. Data Recording and Reporting

● Record the following information for each measurement:


○ Date and time of measurement
○ Sample identification
○ Measured pH value
○ Calibration buffer values and expiry dates used
○ Any observations made during measurement (e.g., unusual behavior of the electrode)

9. Quality Control

● Regularly perform calibration checks with standard buffer solutions to ensure the accuracy of
the pH meter.
● Consider using a reference electrode for more critical measurements, following the
manufacturer's instructions.

10. Reference Documents

● Manufacturer's instructions for the specific pH meter model being used.


● Safety Data Sheets (SDS) for calibration buffers and any chemicals used.
11. Additional Notes

● This SOP serves as a general guideline. Specific details may vary depending on the pH
meter model, samples being measured, and your facility's requirements.
● Always refer to the manufacturer's instructions and internal quality control procedures for
additional details.
Standard Operating Procedure (SOP) for Pole Climbing Apparatus

1. Purpose

This SOP outlines the safe and ethical use of a pole climbing apparatus to assess motor
function, coordination, and grip strength in rodents (typically mice or rats).

2. Scope

This SOP applies to researchers and technicians conducting pole climbing experiments on
rodents within an animal research facility.

3. Responsibilities

● Researcher/Technician:
○ Conducts pole climbing experiments following this SOP.
○ Ensures animal health and well-being during testing.
○ Maintains the pole climbing apparatus in good working condition.
● Animal Caretaker: Ensures animal health and well-being before and after testing.

4. Animal Care and Use

● Ensure all procedures comply with your institution's Animal Care and Use Committee
(IACUC) protocols.
● Use only healthy rodents that have been acclimated to the testing environment.
● Minimize stress and ensure proper handling techniques throughout the experiment.

5. Apparatus Preparation

5.1 Cleaning and Disinfection * Clean the pole and base of the apparatus with a disinfectant
appropriate for your facility. * Allow the apparatus to dry completely before use.

5.2 Functional Check * Ensure the pole is securely attached to the base and stable. * Visually
inspect the pole surface for any damage or slippery spots.

6. Animal Preparation

6.1 Acclimatization * Allow the animals to acclimatize to the testing room for at least 30
minutes before testing.

6.2 Habituation (Optional) * Consider including a habituation session where animals are
allowed to explore the pole climbing apparatus for a short period (following IACUC guidelines) to
familiarize them with the environment.

7. Testing Procedure

7.1 Animal Placement * Gently pick up a rodent and place it at the bottom of the pole, facing
upwards.

7.2 Climbing Observation * Observe the animal's attempts to climb the pole for a
predetermined time period or until it reaches a designated point on the pole. * Note the animal's
climbing ability, grip strength, and any coordination issues.

7.3. Scoring (Optional) * Depending on your experimental design, a scoring system may be
established to quantify climbing success (e.g., reaching a specific height, time spent climbing)
based on IACUC guidelines.

8. Post-Testing

8.1 Animal Recovery * Return the animal to its home cage after testing. * Monitor the animal
for any signs of distress and provide veterinary care if necessary.

8.2 Apparatus Cleaning * Clean the pole climbing apparatus with disinfectant after each testing
session.

9. Maintenance

● Regularly inspect the pole climbing apparatus for damage or wear and tear.
● Tighten any loose screws or components as needed.
● Report any malfunctions or need for repairs to the designated personnel.

10. Reference Documents

● IACUC protocol approval for your specific pole climbing experiment.

11. Additional Notes

● This SOP is a general guideline and specific details may vary depending on the pole climbing
apparatus design, animal species, and your research objectives.
● Consider using a textured or grooved pole surface to improve the animal's grip.
● Avoid using excessively tall poles or testing durations that could lead to exhaustion or injury.

12. Ethical Considerations

● Use the pole climbing apparatus only for approved research purposes.
● Minimize stress and discomfort to the animals during testing.
● Monitor animal behavior and stop the test if signs of distress are observed.
● Provide adequate rest periods between trials if necessary.
Standard Operating Procedure (SOP) for Rotarod Testing

1. Purpose

This SOP outlines the procedures for the safe, consistent, and ethical use of a rotarod
apparatus to assess motor coordination and balance in rodents (mice or rats).

2. Scope

This SOP applies to all personnel conducting rotarod experiments on rodents within your facility,
such as animal research labs.

3. Responsibilities

● Animal Caretaker: Ensures animal health and well-being before, during, and after testing.
● Researcher/Technician:
○ Conducts rotarod testing following this SOP.
○ Maintains the rotarod apparatus in good working condition.

4. Animal Care and Use

● Ensure all procedures comply with your institution's Animal Care and Use Committee
(IACUC) protocols.
● Use only healthy rodents that have been acclimated to the testing environment.
● Minimize stress and ensure proper handling techniques throughout the experiment.

5. Apparatus Preparation

5.1 Cleaning and Disinfection * Clean the rotarod arena and lanes with disinfectant
appropriate for your facility. * Allow the arena to dry completely before use.

5.2 Rotarod Settings * Ensure the rotarod apparatus is level and functioning properly. * Set the
desired initial speed, acceleration rate (if applicable), and maximum speed based on your
experiment. * Choose appropriate lane separators based on the size of the rodents being
tested.

6. Animal Preparation

6.1 Acclimatization * Allow the animals to acclimatize to the testing room for at least 30
minutes before testing.

6.2 Habituation (Optional) * Consider including a habituation session where animals are
placed on the stationary rotarod for a short period to get familiar with the environment (following
IACUC guidelines).

7. Testing Procedure

7.1 Animal Placement * Gently pick up a rodent and place it on the rotarod lane, facing
forward.

7.2 Testing * Start the rotarod according to the chosen settings. * Record the latency (time) to
fall for each animal. * A cut-off time limit may be set to avoid excessive stress on the animal.

7.3 Data Collection * Record the latency to fall for each animal in a designated log or data
collection sheet. * Include relevant details like animal ID, weight, and any observations during
testing.

8. Post-Testing

8.1 Animal Recovery * Return the animal to its home cage after testing. * Monitor the animal
for any signs of distress and provide veterinary care if necessary.

8.2 Apparatus Cleaning * Clean the rotarod arena and lanes with disinfectant after each
testing session.

9. Maintenance

● Perform regular maintenance checks on the rotarod apparatus to ensure proper functioning.
● Report any malfunctions or need for repairs to the designated personnel.

10. Reference Documents

● IACUC protocol approval for your specific rotarod experiment.


● Manufacturer's instructions for the rotarod model being used.

11. Additional Notes

● This SOP is a general guideline and specific details may vary depending on the rotarod
model, animal species, and your research objectives.
● Always refer to your IACUC protocols and institutional guidelines for animal care and use
during rotarod testing.
Standard Operating Procedure (SOP) for Student Simulator Use

1. Purpose

This SOP outlines the procedures for safe, effective, and ethical use of the student simulator
within the learning environment.

2. Scope

This SOP applies to all students, instructors, and staff who will be interacting with the student
simulator during simulations.

3. Responsibilities

● Students:
○ Follow this SOP for proper use of the student simulator.
○ Treat the simulator with respect and avoid causing any damage.
○ Report any malfunctions or issues with the simulator to the instructor immediately.
● Instructor:
○ Ensure students are familiar with this SOP before using the simulator.
○ Supervise student use of the simulator and provide guidance.
○ Maintain the simulator in good working condition and report any maintenance needs.
● Staff:
○ Assist instructors in setting up and maintaining the simulator environment.
○ Ensure a clean and safe learning environment for simulation activities.

4. Pre-Simulation

4.1 Instructor: * Review the learning objectives for the simulation scenario. * Prepare the
simulation environment based on the scenario requirements. * Brief students on the scenario,
including patient condition, roles, and expected behaviors. * Instruct students on the
functionalities and limitations of the simulator.

4.2 Students: * Review the scenario information and learning objectives provided by the
instructor. * Familiarize themselves with the simulator controls and functionalities. * Ask any
questions or clarify any doubts they might have about the scenario or simulator use.

5. Simulation Operation

5.1 Students: * Perform assigned roles within the simulation scenario, interacting with the
simulator as instructed. * Respond to simulated patient signs and symptoms following
established protocols. * Communicate effectively with other students and the instructor during
the simulation. * Maintain a professional and respectful demeanor throughout the simulation.

5.2 Instructor: * Monitor student performance and provide feedback during the simulation. *
Operate the simulator to introduce complications or changes in the patient's condition as
needed. * Pause or terminate the simulation if necessary to address student concerns or safety
issues.

6. Debriefing

6.1 Instructor: * Lead a debriefing session after the simulation to discuss student performance,
decision-making, and learning points. * Encourage open discussion and facilitate reflection on
actions taken and alternative approaches. * Provide constructive feedback to students focusing
on their strengths and areas for improvement.

6.2 Students: * Actively participate in the debriefing session by sharing their experiences and
insights. * Seek clarification and ask questions regarding any aspects of the simulation. * Reflect
on their performance and identify areas for personal growth.

7. Post-Simulation

7.1 Students: * Review their learning objectives and reflect on the key takeaways from the
simulation. * Document their learnings and experiences, if required.

7.2 Instructor: * Reset the simulator to its original state. * Document observations and student
performance during the simulation. * Store simulation data for future reference or debriefing
analysis (if applicable and following ethical guidelines).

8. Maintenance

● The instructor or designated staff will be responsible for routine maintenance of the simulator
according to manufacturer's instructions.
● Report any malfunctions or need for repairs to the appropriate personnel immediately.

9. Ethical Considerations

● All simulation activities must adhere to ethical principles of patient confidentiality, respect for
autonomy, and non-maleficence.
● Students should treat the simulator with respect, simulating care as they would with a real
patient.
● Data collected from simulations should be used for educational purposes only and
anonymized if necessary.

10. Reference Documents

● Manufacturer's instructions for the student simulator model being used.


● Program's curriculum and learning objectives for simulation activities.

Note: This SOP serves as a general guideline and may be adapted to specific student simulator
models, learning environments, and institutional regulations.