2 Software
RUO-IDV-05-0269-A
Release Date: May 2013
This document is provided to customers who have purchased AB SCIEX equipment to use in the
operation of such AB SCIEX equipment. This document is copyright protected and any
reproduction of this document or any part of this document is strictly prohibited, except as
AB SCIEX may authorize in writing.
Software that may be described in this document is furnished under a license agreement. It is
against the law to copy, modify, or distribute the software on any medium, except as specifically
allowed in the license agreement. Furthermore, the license agreement may prohibit the software
from being disassembled, reverse engineered, or decompiled for any purpose. Warranties are as
stated therein.
Portions of this document may make reference to other manufacturers and/or their products,
which may contain parts whose names are registered as trademarks and/or function as
trademarks of their respective owners. Any such use is intended only to designate those
manufacturers' products as supplied by AB SCIEX for incorporation into its equipment and does
not imply any right and/or license to use or permit others to use such manufacturers' and/or their
product names as trademarks.
AB SCIEX warranties are limited to those express warranties provided at the time of sale or
license of its products and are AB SCIEX’s sole and exclusive representations, warranties, and
obligations. AB SCIEX makes no other warranty of any kind whatsoever, expressed or implied,
including without limitation, warranties of merchantability or fitness for a particular purpose,
whether arising from a statute or otherwise in law or from a course of dealing or usage of trade,
all of which are expressly disclaimed, and assumes no responsibility or contingent liability,
including indirect or consequential damages, for any use by the purchaser or for any adverse
circumstances arising therefrom.
For research use only. Not for use in diagnostic procedures.
The trademarks mentioned herein are the property of AB Sciex Pte. Ltd. or their respective
owners.
AB SCIEX™ is being used under license.
© 2013 AB Sciex Pte. Ltd.
Printed in Canada.
2 of 78 RUO-IDV-05-0269-A
Contents
Chapter 1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Related Documentation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5
Technical Support . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5
System Components . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5
Supported Devices . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6
Peripheral Device Software Plug-in Vendors . . . . . . . . . . . . . . . . . . . . . . . . . . . .6
Adding AAO-Controlled Devices to the Hardware Profile . . . . . . . . . . . . . . . . . . .7
Chapter 2 Agilent Series Devices Configuration . . . . . . . . . . . . . . . . . . . . . . . . . 9
Configuration of Agilent Devices Through Serial Port Communication . . . . . . . . .9
Configuration of Agilent Devices Through GPIB Communication . . . . . . . . . . . .10
Configuration of Agilent Devices Through LAN (Ethernet) Communication . . . .10
Configuring the Network Interface Card . . . . . . . . . . . . . . . . . . . . . . . . . . . . .12
Configuration of Agilent Devices with CAN Cables . . . . . . . . . . . . . . . . . . . . . . .14
Chapter 3 Autosampler Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Agilent Autosampler . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .15
Connecting the Agilent Autosampler . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .16
CTC PAL Autosampler . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .19
Connecting the CTC PAL Autosampler . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .19
Gilson 215 Liquid Handler . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .21
Connecting the Gilson 215 Liquid Handler . . . . . . . . . . . . . . . . . . . . . . . . . . .21
Installing the Syringe on the Gilson 215 Liquid Handler . . . . . . . . . . . . . . . . .26
Gilson 233 XL Sampling Injector . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .26
Connecting the Gilson 233 XL Sampling Injector . . . . . . . . . . . . . . . . . . . . . .26
Installing the Syringe on the Gilson 402 Syringe Pump . . . . . . . . . . . . . . . . .29
PerkinElmer Series 200 Autosampler . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .30
Connecting the PerkinElmer Series 200 Autosampler . . . . . . . . . . . . . . . . . .30
Spark Holland Endurance Autosampler . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .33
Connecting the Spark Holland Endurance Autosampler . . . . . . . . . . . . . . . . .33
Other Autosamplers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .36
Chapter 4 Pump Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
Agilent Pumps . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .37
Connecting the Agilent Pump to the Computer . . . . . . . . . . . . . . . . . . . . . . . .38
Harvard 22 Syringe Pump . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .40
Connecting the Harvard 22 Syringe Pump . . . . . . . . . . . . . . . . . . . . . . . . . . .40
PerkinElmer Series 200 LC Pumps . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .41
Connecting the PerkinElmer Series 200 LC Pumps . . . . . . . . . . . . . . . . . . . .41
Chapter 5 Shimadzu Devices . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
Shimadzu LC-30 Series Devices . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .43
Setting Shimadzu Device Communications – SIL-HT (SCL-10Avp) . . . . . . . .44
Setting Shimadzu Device Communications for Use on a Network
– CBM-20A (CBM-20A lite) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .44
Setting Shimadzu Device Communications for Stand-alone use
RUO-IDV-05-0269-A 3 of 78
Contents
4 of 78 RUO-IDV-05-0269-A
Introduction
1
This guide is intended for customers and FSEs who are responsible for configuring devices to
work with the mass spectrometer. You can control devices automatically during LC/MS/MS data
acquisition through the Analyst® or Analyst TF software. The software supports LC pumps,
autosamplers, column ovens, switching valves, detectors, and analog-to-digital converters from
several manufacturers.
Some hardware setup and configuration is required so that the supported peripheral devices and
the mass spectrometer can communicate properly. Use the procedures in this guide to connect
and configure the peripheral devices and your system.
Related Documentation
The guides and tutorials for the instrument and the Analyst or Analyst TF software are installed
automatically with the software and are available from the Start menu: All Programs > AB SCIEX.
A complete list of the available documentation can be found in the Help. To view the Analyst
software Help, press F1.
Technical Support
AB SCIEX and its representatives maintain a staff of fully-trained service and technical
specialists throughout the world. They can answer questions about the instrument or any
technical issues that may arise. For more information, visit the Web site at www.absciex.com.
System Components
Figure 1-1 on page 6 is an example of how you can connect peripheral devices to the computer
and mass spectrometer. You do not have to configure your system exactly as shown. In this
example, the pumps, column ovens, autosamplers, and switching valves are controlled from the
computer by means of serial (RS-232) ports. The diode array detector is controlled by LAN
(Ethernet) cables.
For more information on how to configure peripheral devices to communicate with the computer,
see the section in this guide specific to each device.
RUO-IDV-05-0269-A 5 of 78
Introduction
Supported Devices
For an up-to-date list of the peripheral devices and firmware supported by the Analyst software,
see the most current Analyst software Release Notes and Installation Guide.
6 of 78 RUO-IDV-05-0269-A
Introduction
RUO-IDV-05-0269-A 7 of 78
Introduction
3. In the Available Devices dialog, in the Device Type list, click Software
Application.
The list of AAO software applications installed on the computer appears in the
Devices box.
4. Click the AAO software applications you want to add, and then click OK.
8 of 78 RUO-IDV-05-0269-A
Agilent Series Devices Configuration
2
This section provides information about configuring the Agilent series peripheral devices using a
standard serial (RS-232) port, GPIB (general purpose interface bus), or LAN (Ethernet)
communication, with or without CAN cables. An overview of each type of communication is
provided for the Agilent 1100, 1200, 1260, and 1290 series LC Systems.
Note: Use CAN cables with a RS-232, GPIB, or LAN (Ethernet) cable if configuring
multiple Agilent devices in a stack configuration. See Configuration of Agilent Devices
with CAN Cables on page 14.
Note: You must connect the diode array detector to the computer through GPIB or LAN
(Ethernet) communication.
If you connect an Agilent device (except a DAD) to the computer with a RS-232 cable, set the
DIP switches at the back of the device. The DIP switches configure parameters for the
communication protocol and instrument initialization procedures.
The following table shows the appropriate DIP switch settings for a baud rate of 19 200 bps for
the Agilent 1100, 1200, 1260, and 1290 series devices. If you create a new hardware profile that
includes an Agilent 1100, 1200, 1260, or 1290 series device, or if you add an Agilent device to an
existing hardware profile, set the DIP switches for a baud rate of 19 200, and then set the baud
rate to 19 200 in the Hardware Configuration Editor.
Note: You must restart the devices to apply the new baud rate.
RUO-IDV-05-0269-A 9 of 78
Agilent Series Devices Configuration
10 of 78 RUO-IDV-05-0269-A
Agilent Series Devices Configuration
Note: The 1290 module is shipped with all switches Down (Off). To perform any LAN
configuration, SW1 and SW2 must be Down. For all modules with on-board LAN
(G1315/65C/D, G1314D/E, G4212A, G4220A) the default is all switches Down. For
specific LAN modes, switches 3 to 8 must be set as required. For boot or test modes,
switches 1 and 2 must be Up (On).
RUO-IDV-05-0269-A 11 of 78
Agilent Series Devices Configuration
3. On the General tab, right-click Internet Protocol and then click Properties.
12 of 78 RUO-IDV-05-0269-A
Agilent Series Devices Configuration
Note: If you are using the BOOTP utility, you need the MAC address from
the label on the module. The MAC address is used to recognize which
device is used in the configuration.
6. Press Enter.
The Subnet mask field is automatically filled.
7. Click OK and then close the Local Area Connection 2 Properties dialog.
See the Agilent 1200 Series Diode Array and Multiple Wavelength Detector SL User
Manual.
Note: Make sure that the DAD is using the preset LAN settings (IP address
192.168.254.11).
RUO-IDV-05-0269-A 13 of 78
Agilent Series Devices Configuration
2. To confirm the connection between the computer and the DAD, click Start and then
click Run.
3. In the Open field, type ping 192.168.254.11.
The Ping utility makes four data transmissions to the IP address specified and
reports the reply if successful. If the data transmission fails, the utility reports
“Request timed out.”
4. Using the Analyst® software, add the DAD to a hardware profile. See the Analyst
software Help.
Note: If you connect a DAD to the computer using an Ethernet connection and you
connect the rest of the stack to the computer using a single RS-232 cable, you cannot
connect the DAD to the rest of the stack with a CAN cable.
To monitor and control the stack manually, you can connect a handheld Agilent series control
module to one of the CAN connections at the back of any Agilent device. The devices connected
by CAN cables in the stack must match the devices in the Analyst software hardware profile. In
the event of a fault in the CAN-linked stack, reboot all the devices in the stack.
For more information on configuring Agilent devices with CAN cables, see the Agilent
documentation.
Note: If you are switching a stack from CAN to another communication mode in the
Analyst software, the CAN cables must be disconnected from the device.
Note: All devices connected by CAN must be at the same level of firmware.
14 of 78 RUO-IDV-05-0269-A
Autosampler Configuration
3
This section provides information on the required autosampler hardware, how to connect the
autosampler to the computer and the mass spectrometer, and how to configure the most current
autosampler for external control.
The cables for the Agilent and Shimadzu autosamplers, and any other device that does not have
its own AUX I/O cable, are included with the mass spectrometer. The cable for the CTC
autosampler is included with the CTC autosampler.
Note: You can configure autosamplers that are not supported by the Analyst® software
to operate with the mass spectrometer through analog signals or through AAO–type
software. For information on configuring unsupported autosamplers to operate with a
mass spectrometer, see Peripheral Device Analog Synchronization on page 73.
Agilent Autosampler
The following Agilent autosamplers are supported by the Analyst software and are configured the
same way. For the most up-to-date list of supported devices, see the most current Analyst
software Release Notes.
Table 3-1 Supported Agilent 1100 Series Autosamplers
Autosampler Model number
Standard G1313A
Thermo-enabled standard G1327A
Well-plate G1367A
Thermo-enabled well-plate G1368A
Micro well-plate G1377A
Thermo-enabled micro well-plate sampler G1378A
Micro G1389A
Thermo-enabled micro autosampler G1387A
Thermostat module G1330B
RUO-IDV-05-0269-A 15 of 78
Autosampler Configuration
The following table lists the required hardware. For the latest version of firmware supported, see
the most current Analyst software Release Notes.
Table 3-5 Required Hardware for the Agilent Autosamplers
Cable Other parts needed
• RS-232 cable (PN WC024736) • Network interface card if using a LAN (Ethernet)
connection
• GPIB cable (PN WC021365)
• Agilent PN G5183-4649 (for a direct LAN connection)
• AUX I/O cable (PN 014474)
• Agilent PN G1530-61485 (for a LAN connection using a
hub)
16 of 78 RUO-IDV-05-0269-A
Autosampler Configuration
1 2 3 4 5
Item Description
1 Remote port
2 Serial port
3 CAN connectors
4 Agilent GPIB
5 DIP switches
RUO-IDV-05-0269-A 17 of 78
Autosampler Configuration
Table 3-6 Wiring for the Agilent Autosampler (TTL—Active Low) Injection
Input
Autosampler Mass spectrometer AUX I/O cable
Pin 9 (power 5V) Red with black stripes
Pin 10 (anode) Orange with black
stripes
Remote port (pin 3) White with black stripe Pin 22 (cathode) White with black stripes
0 0
Remote port (pin 1) Green with black stripe Pin 21 (ground) Green with black stripes
0 0
2. Connect the cathode wire (white with black stripes) and the ground wire (green with
black stripes) on the AUX I/O cable to the remote port at the back of the Agilent
autosampler.
3. Connect the cathode wire (white with black stripes) to Pin 3 of the remote port and
connect the ground wire (green with black stripes) to Pin 1 of the remote port.
Polarity is important.
Note: Make the connections to the remote port with a 9-pin DB push-lock
or solder-tail connector. If you use the Agilent remote cable to connect the
remote port to the AUX I/O cable, make the cable as short as possible.
4. Connect the other end of the AUX I/O cable to the mass spectrometer AUX I/O
connector.
18 of 78 RUO-IDV-05-0269-A
Autosampler Configuration
Wire the autosampler so that the autosampler injection triggers the mass spectrometer to begin
data acquisition. To do so, connect a pair of wires from the AUX I/O connector at the back of the
mass spectrometer to the remote port of the autosampler.
1 2
RUO-IDV-05-0269-A 19 of 78
Autosampler Configuration
Item Description
1 AUX I/O connector
2 Fast wash station connector
Inject marker (pin 3) White with black stripe Pin 22 (cathode) White with black stripes
0 0
Common (pin 4) Green with black stripe Pin 21 (ground) Green with black stripes
0 0
Caution: Potential Equipment Damage: Cover each connection and then the
entire cable assembly with insulating tape or heat shrink tubing to prevent
shorting to other wires or grounded metal parts.
2. Connect the white with black stripe AUX I/O wire to Pin 3 of the DB15 connector.
3. Connect the green with black stripe AUX I/O wire to Pin 4 of the DB15 connector.
4. Connect the DB15 male connector to the CTC PAL autosampler Interface 1
connector.
5. Connect the other end of the AUX I/O cable into the mass spectrometer AUX I/O
connector.
20 of 78 RUO-IDV-05-0269-A
Autosampler Configuration
Note: Make sure that the tray hardware configured in the system is listed
in the Tray Type and Tray Holder menus. See the manufacturer’s
documentation.
10. Press Esc to return to the previous window and then scroll down to select Inject.
11. In the next window that appears, highlight the Source line and then scroll between
the options. Select Immediate, and then press ENTER.
12. Press Esc twice to move back two windows.
13. Scroll down and then select Out Signals.
14. In the next window that appears, select Injected.
15. Highlight the Destination line, scroll between the options and then select SW-Out1.
16. Press F4 to return to the Home menu.
Wire the autosampler so that the autosampler injection triggers the mass spectrometer to begin
data acquisition. To do so, connect a pair of wires from the AUX I/O connector at the back of the
mass spectrometer to the remote port of the autosampler.
RUO-IDV-05-0269-A 21 of 78
Autosampler Configuration
Figure 3-3 I/O connection for the Gilson 215 liquid handler
Item Description
1 Output switch 1 (pins 3 and 4)
22 of 78 RUO-IDV-05-0269-A
Autosampler Configuration
Figure 3-4 Cable connections for the Gilson 215 liquid handler
4
Item Description
1 GSIOC cable to the Gilson 819 or 841
2 RS-232 port to computer
3 SW1
4 SW2
Injection Input
Autosampler Mass spectrometer AUX I/O cable
Interface 136 Pin 9 (power 5V) Red with black stripes
Pin 10 (anode) Orange with black
stripes
Output switch (pin 3) White with black stripe Pin 22 (cathode) White with black stripes
0 0
Output switch (pin 4) Green with black stripe Pin 21 (ground) Green with black stripes
0 0
Peripheral Devices Setup Guide Analyst® 1.6.2 Software
RUO-IDV-05-0269-A 23 of 78
Autosampler Configuration
2. Connect the cathode wire (white with black stripes) and the ground wire (green with
black stripes) on the AUX I/O cable to Output Switch 1 (Pins 3 and 4 of the Output
Port at the back of the Gilson 215 liquid handler) using the green connector supplied
with the liquid handler.
3. Connect the other end of the AUX I/O cable to the mass spectrometer AUX I/O
connector.
WARNING! Electrical Shock Hazard: Disconnect the power cord and wait at least
one minute before removing the cover. See the Gilson Injection Valve Actuator
operator safety manual before removing the cover.
1. Turn off the Gilson 215 liquid handler by pressing the On/Off button.
2. Turn off the Gilson injection valve actuator by pressing the On/Off button.
3. Disconnect the Gilson keypad and any other connections to the Gilson 215 liquid
handler.
4. At the back of the Gilson injection valve actuator, set the UNIT ID so that the white
dot is at position 9. For help locating the UNIT ID, see Figure 3-5 and Figure 3-6.
5. Remove the cover of the injection valve actuator.
6. Do one of the following:
• For the 819 injection valve actuator, set SW 1 so that the white dot is at
Position 0 (External).
• For the 841 injection valve actuator, set the jumper to External.
Note: When viewed from the front of the Gilson 819 injection valve
actuator, the SW1 switch is located on the left side. For the 841 injection
valve actuator, the jumper is located on the right side.
24 of 78 RUO-IDV-05-0269-A
Autosampler Configuration
1 2
Item Description
1 GSIOC port
2 UNIT ID
Item Description
1 GSIOC port
2 UNIT ID
RUO-IDV-05-0269-A 25 of 78
Autosampler Configuration
Wire the autosampler so that the autosampler injection triggers the mass spectrometer to begin
data acquisition. To do so, connect a pair of wires from the AUX I/O connector at the back of the
mass spectrometer to the remote port of the autosampler.
Figure 3-7 Back panel on the Gilson 233 XL sampling injector
r
1 2 3 4
Item Description
1 Relay output 1
2 DIP switches
3 GSIOC
4 RS-232 port
26 of 78 RUO-IDV-05-0269-A
Autosampler Configuration
Injection Input
Autosampler Mass spectrometer AUX I/O cable
Pin 9 (power 5V) Red with black stripes
Pin 10 (anode) Orange with black
stripes
Output switch (pin 3) White with black stripe Pin 22 (cathode) White with black stripes
0 0
Output switch (pin 1) Green with black stripe Pin 21 (ground) Green with black stripes
0 0
2. Connect the cathode wire (white with black stripes) and the ground wire (green with
black stripes) of the AUX I/O cable to the Output Switch 1 (Pins 1 and 3 of the Output
Port at the back of the Gilson 233) using the green connector supplied with the liquid
handler.
3. Connect the other end of the AUX I/O cable to the mass spectrometer AUX I/O
connector.
RUO-IDV-05-0269-A 27 of 78
Autosampler Configuration
Table 3-13 Gilson 402 Pump: Options and ID DIP Switch Settings
For this switch... 8 7 6 5 4 3 2 1
Set as... Left Left Right Left Left Left Left Left
(Off) (Off) (On) (Off) (Off) (Off) (Off) (Off)
5. Set the ID Number DIP switches at the back of the Gilson 233 XL sampling injector
as follows:
Table 3-14 Gilson 233 XL Sampling Injector: ID Number DIP Switch Settings
For this switch... 7 6 5 4 3 2 1 0
Set as... Down Up Up Down Up Down Up Down
(Off) (On) (On) (Off) (On) (Off) (On) (Off)
28 of 78 RUO-IDV-05-0269-A
Autosampler Configuration
b. Position the injection port at the far right end of the adapter, above the
switching valve.
Figure 3-11 Injection port at the far right end of the adapter
3. Turn on the Gilson 402 syringe pump by pressing the On/Off button at the back of
the device.
4. Turn on the Gilson 233 XL sampling injector by pressing the On/Off button at the
back of the device.
5. Restart the computer.
RUO-IDV-05-0269-A 29 of 78
Autosampler Configuration
Wire the autosampler so that the autosampler injection triggers the mass spectrometer to begin
data acquisition. To do so, connect a pair of wires from the AUX I/O connector at the back of the
mass spectrometer to the remote port (INJ 1 or INJ 2) signal of the autosampler.
30 of 78 RUO-IDV-05-0269-A
Autosampler Configuration
Table 3-16 Wiring for Perkin Elmer Series 200 Autosampler (Normally Open)
Injection Port
Autosampler Mass spectrometer AUX I/O cable
Pin 9 (power 5V) Red with black stripes
Pin 10 (anode) Orange with black
stripes
INJ 1 White with black stripe Pin 22 (cathode) White with black stripes
0 0
Green with black stripe Pin 21 (ground) Green with black stripes
0 0
Note: Polarity is not important; however, use the two middle slots of the
connector.
2. Connect the cathode wire (white with black stripes) and the ground wire (green with
black stripes) to the AUX I/O cable with one of the telephone-style connectors
supplied with the PerkinElmer Series 200 autosampler.
Figure 3-12 Wire connections to the telephone-style connector
1 2 4
3
Item Description
1 Inverted top piece—connector housing
2 AUX I/O cable
3 Bottom piece—connector housing
4 AUX I/O cable
5 Top piece—connector housing
6 Phillips screw
3. Insert the completed connector into the INJ 1 port on the side panel of the
autosampler.
RUO-IDV-05-0269-A 31 of 78
Autosampler Configuration
5. Connect the other end of the AUX I/O cable to the mass spectrometer AUX I/O
connector.
32 of 78 RUO-IDV-05-0269-A
Autosampler Configuration
Wire the autosampler so that the autosampler injection triggers the mass spectrometer to begin
data acquisition. To do so, connect a pair of wires from the AUX I/O connector at the back of the
mass spectrometer to the remote port of the autosampler.
Note: For more information about serial cables, see the autosampler
documentation.
RUO-IDV-05-0269-A 33 of 78
Autosampler Configuration
Item Description
1 Communication Interface connectors
3. Connect the serial cable from the serial port on the autosampler to the desired serial
port on the computer, noting the port number. Use any of the following serial cables:
• RS-232
• RS-422
• RS-485
34 of 78 RUO-IDV-05-0269-A
Autosampler Configuration
2. Connect the AUX I/O wire (white with black stripes) to Pin 1 of the DB15 male
connector.
3. Connect the AUX I/O wire (green with black stripes) to Pin 13 of the DB15 male
connector.
4. Insert the DB15 male connector into the P2 connector at the back of the
autosampler.
5. Insert the other end of the AUX I/O cable into the AUX I/O connector on the mass
spectrometer.
RUO-IDV-05-0269-A 35 of 78
Autosampler Configuration
Other Autosamplers
The instructions in this section are required only when AAO or the Analyst software support is not
available. You can synchronize any autosampler with the mass spectrometer for use with the
normally open autosampler contact closure inject signal. The autosampler is connected to the
mass spectrometer by means of an AUX I/O cable.
To synchronize other autosamplers, first create a hardware profile, and then choose the LC
synchronization trigger.
6. Click OK.
The Hardware Configuration Editor dialog appears.
7. Click Activate Profile.
A green check mark appears next to the hardware profile, indicating that the profile
is active.
36 of 78 RUO-IDV-05-0269-A
Pump Configuration
4
This section describes the required hardware for each pump, how to connect the pump to the
computer, and how to configure the pump for external control.
Agilent Pumps
The following Agilent pumps are supported by the Analyst® software and all are configured the
same way.
Table 4-1 Supported Agilent 1200 Series Pumps
Pumps Model number 1200
Binary G1312A, G1312B
Quaternary G1311A
Isocratic G1310A
Capillary G1376A
Nano G2226A
Note: The Agilent G4220A Binary Pump contains a software-controlled Purge valve.
The Analyst software allows users to control the purge option through the Acquisition
Method. For more information, see the most current Analyst software Release Notes.
The following table lists the required hardware. For the latest version of firmware supported, see
the most current Analyst software Release Notes. Depending on how the system is configured,
you may not require all of the following cables.
RUO-IDV-05-0269-A 37 of 78
Pump Configuration
Table 4-4 Required Hardware for Agilent 1100, 1200, 1260, and 1290 Series Pumps
Cable Other parts needed
• RS-232 cable (PN WC024736) • General purpose cable for Agilent devices (Agilent PN
G1103-61611)
• GPIB cable (PN WC021365)
The following parts are optional. The external relay
• CAN cable (ships with Agilent
contacts board (Agilent PN G1351-68701) is required to
system)
provide timed contact closure events during the LC
program. This option is not required for analog
synchronization of peripheral devices.
• Network interface card (PN 1016082) if using a LAN
(Ethernet) connection
• Agilent PN G5183-4649 (for a direct LAN connection)
• Agilent PN G1530-61485 (for a LAN connection using a
hub)
WARNING! Electrical Shock Hazard: See the Agilent Pump Safety Instructions
before configuring any AC mains-powered equipment.
WARNING! Electrical Shock Hazard: Disconnect the power cord and wait at least
one minute before removing the pump cover.
38 of 78 RUO-IDV-05-0269-A
Pump Configuration
Figure 4-1 Agilent pump (back) with external relay contacts board installed
1 2 3 4 5
Item Description
1 Remote connector
2 Serial port
3 CAN connectors
4 Agilent GPIB
5 DIP switches
Note: Depending on the firmware version, you may require an access code (300) in
the Agilent Nano Pump (G2226A).
RUO-IDV-05-0269-A 39 of 78
Pump Configuration
The following table lists the required hardware. For the latest version of supported firmware, see
the most current Analyst software Release Notes.
Table 4-5 Required Hardware for the Syringe Pump
Cable Other parts needed
RS-232 cable (PN WC024735) N/A
3. Press the STOP/START key until 96 is displayed and then press the ENTER key.
The baud rate is set to 9600.
40 of 78 RUO-IDV-05-0269-A
Pump Configuration
Note: To run a gradient with two PerkinElmer Series 200 micro pumps, you need two
cables and two serial ports.
The following table lists the required hardware. For the latest version of supported firmware, see
the most current Analyst software Release Notes.
Table 4-8 Required Hardware for the PerkinElmer Pumps
Cable Other parts needed
• RS-232 cable (PN WC024736) N/A
RUO-IDV-05-0269-A 41 of 78
Pump Configuration
42 of 78 RUO-IDV-05-0269-A
Shimadzu Devices
5
You can use the following three system controllers to connect to and control a Shimadzu HPLC
system using the Analyst® software:
• CBM-20A
• CBM-20A lite
• SIL-HT (SCL-10Avp)
Communications settings are similar for all three.
The CBM or SCL is required for the Analyst software to communicate with and control any
Shimadzu device. For the SCL, the only way to connect to the computer is using a RS-232
(serial) cable. The CBM uses serial or TCP/IP (Ethernet) connectivity, with TCP/IP being the
preferred mode of communication. For more information on controlling Prominence devices
using the Analyst software, contact your AB SCIEX FSE.
WARNING! Electrical Shock Hazard: See the Shimadzu CBM System Controller
Safety Instructions before configuring any AC mains-powered equipment.
RUO-IDV-05-0269-A 43 of 78
Shimadzu Devices
44 of 78 RUO-IDV-05-0269-A
Shimadzu Devices
• SUBNET MASK: Type the number supplied by the Network Administrator, and
then press ENTER.
• DEFAULT GATEWAY: Type the value supplied by the Network Administrator,
and then press ENTER.
8. TRS MODE sets the communications protocol parameters. Press 2 and then press
ENTER.
9. POWER CYCLE the unit to accept and save the changes.
RUO-IDV-05-0269-A 45 of 78
Shimadzu Devices
11. Right-click the network connection that you want to dedicate to the Shimadzu CBM
communications and then click Properties.
12. Click Internet Protocol (TCP/IP) and then click Properties.
13. Click Use the following IP address and then type the following:
• IP ADDRESS: 192.168.200.89
• SUBNET MASK: 255.255.255.0
• DEFAULT GATEWAY: Leave blank
14. Click OK to accept the changes.
15. Click CLOSE.
16. Shut down the computer.
17. Using a CAT 5 network cable, connect the Shimadzu CBM (lite) to the supplied
(Shimadzu) network switch.
Note: You can use a CAT 5 crossover cable instead of the network switch
to directly connect the computer to the Shimadzu CBM (lite).
18. Connect the computer to the network switch using the network card that was
configured for use with the Shimadzu system.
19. Turn on the computer and the CBM (lite) and wait for them to complete their
respective boot-up routines.
20. To determine whether proper communications have been established between the
computer and CBM (lite), start Microsoft Internet Explorer (other browsers may not
display properly), type the CBM (lite) IP address in the address bar
(192.168.200.99), and then click GO.
Note: Make sure that all pop-up blockers are turned off.
The Shimadzu Prominence LC CBM-20A screen appears for a few seconds followed
by the Status screen.
21. Make sure that the Serial number listed for the HPLC system under System Name
matches that of the unit to which you are connected and that its status is Ready.
22. Close Internet Explorer.
23. Start the Analyst software and then configure the HPLC system.
46 of 78 RUO-IDV-05-0269-A
Shimadzu Devices
3. Connect the RS-232 cable from the serial port at the back of the system controller to
any available serial port on the computer, noting the port number. See Figure 5-1
Back of the Shimadzu SCL system controller or Figure 5-2 Back of the Shimadzu
CBM system controller.
Figure 5-1 Back of the Shimadzu SCL system controller
1
2 3
4
5 6 7
5
10 10 8
9
Item Description
1 Remote Connector Channels 1 to 8 (fiber optic ports)
2 RS-232C connector
3 OPT LINK connector
4 AC REMOTE connector
5 EVENT OUT connectors
6 MAN. INJ. IN connector
7 ALARM IN connector
8 Power connector
9 Fuse holder
10 AC OUT connectors
RUO-IDV-05-0269-A 47 of 78
Shimadzu Devices
3
8 6 3
7
5 5 4
Item Description
1 Ethernet port
2 Remote Connector Channels 1 to 8 (fiber optic ports)
3 External I/O connectors
4 Power connector (AC IN)
5 AC output connectors (AC OUT)
6 RS-232 connector
7 Network indicators (100M/ACT/LINK)
8 Initialization button (INIT)
Configuring the System Controller for External Control (Analyst Software Version
1.3 or Later)
1. Turn on the Shimadzu SCL system controller.
The System Configuration window appears.
2. If the word FIXED appears in the upper right corner of the System Configuration
screen, press the F2 key (screen name FIX) to deselect FIX.
3. Press the F5 key (screen name MENU).
The Menu window appears.
4. Press the number 4 key.
The System screen appears.
5. Press the F3 key (screen name NEXT).
6. Set the Class VP to 5.x: Use the up and down arrow keys to select Class VP and
then use the left and right arrow keys to select 5.x.
48 of 78 RUO-IDV-05-0269-A
Shimadzu Devices
7. Set the Interface to RS-232C: Use the up and down arrow keys to select Interface
and then use the left and right arrow keys to select RS-232C.
8. Set the Baud Rate to 9600: Use the up and down arrow keys to select Baud Rate
and then use the left and right arrows to select 9600.
9. Set the Level to Enhanced: Use the up and down arrow keys to select Level and
then use the level and right arrows to select Enhanced.
10. Press the F5 key (screen name MENU).
11. Turn off the Shimadzu SCL System Controller.
12. Turn on the Shimadzu SCL System Controller.
The System Configuration screen loads.
13. Connect and configure individual devices to the system controller, following the
instructions in their respective guides.
14. Verify that each connected device is listed on the System Configuration screen on
the system controller.
Configuring the System Controller for External Control (Analyst Software Version
1.2 or Earlier)
1. Turn on the Shimadzu SCL System Controller.
The System Configuration window appears.
2. If the word FIXED appears in the upper right corner of the System Configuration
screen, press the F2 key (screen name FIX) to deselect FIX.
3. Press the F5 key (screen name MENU).
The Menu window appears.
4. Press the number 4 key - Response.
The System screen appears.
5. Press the F3 key (screen name NEXT).
6. Set the Class VP to 4.x: Use the up and down arrow keys to select Class VP and
then use the left and right arrow keys to select 4.x.
7. Set the Interface to RS-232C: Use the up and down arrow keys to select Interface:
8. Set the Baud Rate to 9600: use the up and down arrow keys to select Baud Rate.
9. Press the F5 key (screen name MENU).
10. Turn off the Shimadzu SCL System Controller.
11. Turn on the Shimadzu SCL System Controller.
The System Configuration screen loads.
12. Connect and configure individual devices to the system controller, following the
instructions in their respective guides.
13. Verify that each connected device is listed in the System Configuration window on
the system controller.
RUO-IDV-05-0269-A 49 of 78
Shimadzu Devices
Orange wire White with black stripe Pin 22 (cathode) White with black stripes
0 0
Brown wire Green with black stripe Pin 21 (ground) Green with black stripes
0 0
Note: Isolate these wires so they do not contact any other wires or metal.
3. On the free end of the AUX I/O cable, short together the following wires but do not
connect them to anything else:
• Red with black stripe (wire 9)
• Orange with black stripe (wire 10)
4. Connect the other end of the AUX I/O cable to the mass spectrometer AUX I/O
connector.
5. Verify that RELAY 1 in the SIL-HT (SCL-10Avp) is set to START while you are
configuring the Shimadzu system controller in the Analyst software.
50 of 78 RUO-IDV-05-0269-A
Shimadzu Devices
Connecting the Shimadzu CBM (lite) System Controller to the Mass Spectrometer
The AUX I/O cable (PN 014474) is used to connect the Shimadzu CBM (lite) system controller to
the mass spectrometer.
1. Connect the Shimadzu Event cable to Out 1 at the back of the CBM, shown in
Figure 5-2 Back of the Shimadzu CBM system controller on page 48, by pressing the
button above the terminal with a flat-head screwdriver and pushing the wire inside.
Make sure the wire is held securely inside the terminal.
2. Connect the wires from the free end of the AUX I/O cable as follows:
Table 5-4 Wire Connections
Use this AUX I/O wire... And connect to Event Cable...
White with black stripe (wire 22) Black wire
Green with black stripe (wire 21) White wire
Table 5-5 Wiring for the Shimadzu CBM (lite) System Controller
CBM Mass spectrometer AUX 1/O cable
Event Pin 9 (power 5V) Red with black stripes
cable Pin 10 Orange with black stripes
Black wire White with black stripe Pin 22 (cathode) White with black stripes
0 0
White wire Green with black stripe Pin 21 (ground) Green with black stripes
0 0
3. On the free end of the AUX I/O cable, short together the following wires but do not
connect them to anything else:
• Red with black stripe (wire 9)
• Orange with black stripe (wire 10)
4. Connect the other end of the AUX I/O cable to the mass spectrometer AUX I/O
connector.
5. Make sure that RELAY 1 in the CBM-20A lite is set to START while you are
configuring the Shimadzu system controller in the Analyst software.
Shimadzu Devices
Shimadzu recommends that the devices attached to the CBM system controller be identical to
those configured in the Analyst software hardware profile. Differences between the two
configurations can result in communication problems between the Analyst software, the
Shimadzu system controller, and the attached devices.
RUO-IDV-05-0269-A 51 of 78
Shimadzu Devices
If the vial detection sensor is ON, then missing autosampler vials or aborting a run during an
autosampler rinse creates Shimadzu device fault conditions. To correct these errors you will have
to intervene manually before the Analyst software can continue functioning normally. To recover
Analyst software control, perform the task indicated on the device display. Alternatively, follow the
Fault Recovery procedure to clear all conditions.
The Shimadzu preset run time is set at 90 minutes. If required, change the duration in the
method.
Note: The needle height in the method must match that of the current tray. The preset
value is not valid for all trays.
Fault Recovery
The Shimadzu HPLC equipment can generate three different error conditions that cause the
Analyst software to stop: warning, error, and fatal error.
Warnings
A warning is an informational notification of conditions such as a door open on a temperature
controlled module, solvent level, or temperature not ready. These conditions do not prevent the
Shimadzu system from operating properly; however, the Analyst software does not recognize
these warnings, generates an error, and then stops the sequence. Contact your local Shimadzu
Technical Support specialist for more information on how to minimize these conditions.
Errors
Any error condition on the Shimadzu system stops the Analyst sequence. The Shimadzu system
typically sounds an audible alarm in the event of an error until you acknowledge the error. Some
errors that may be encountered and the Shimadzu suggested action include the following:
• LEAK DETECT: Press CE to stop the alarm. Find and address the problem.
Thoroughly dry the area around the leak sensor of the affected module (and possibly
any module below it in the stack due to the internal drain system). Recover
according to the procedures below.
• PRESSURE OVER PMAX: Press CE to stop the alarm. Correct the problem.
Recover with one of the following procedures: To recover from a fault for systems
equipped with a CBM-20A lite system controller, ROM 1.11 or higher on page 53 or
To recover from a fault for systems equipped with a SCL-10Avp (SIL-HT) system
controller on page 54.
• MISSING VIAL: This error appears on the autosampler if it does not find a vial it is
asked to inject. The result of this condition can be dealt with in one of two ways
through the Analyst software in the Shimadzu Hardware configuration settings.
Select the autosampler model from the list and then click Configuration to display
the Autosampler Configuration dialog.
52 of 78 RUO-IDV-05-0269-A
Shimadzu Devices
Select the Skip sequence line if vial is missing check box and click OK. The
Analyst software skips that vial and continues running. If the check box is not
selected, then the Analyst software reports an error and stops the sequence.
The Skipped Vial notification appears on the SIL front panel display and the vial
number skipped is shown. Be sure to reconcile the data obtained in subsequent
runs.
Fatal Errors
The final level of error generated by Shimadzu equipment is a fatal error. Fatal errors are
normally generated by a mechanical failure and are generally associated with the autosampler
injection mechanism. The only way to recover from a fatal error is to power cycle the entire
system. If, after power cycling, the error occurs again, contact your local Shimadzu Service
Engineer for assistance.
To recover from a fault for systems equipped with a CBM-20A lite system
controller, ROM 1.11 or higher
For warnings and typical errors, the module experiencing the problem displays the condition on
its front panel display and the module and CBM display a RED status LED bar. The connect LED
on the CBM is no longer lit. The CBM-20A lite system controller works in the same way but has
no indication of the error because it is installed in a module.
1. Press CE to stop the alarm and clear the error.
2. Correct the cause of the error.
3. Press the black INIT button at the back of the CBM-20A lite for no longer than five
seconds. See Figure 5-2 Back of the Shimadzu CBM system controller on page 48.
The CBM status LED bar changes to green and the connect LED illuminates, thus
confirming that communication with the Analyst software has been restored.
4. If either the status LED does not change to green or the connect LED fails to
illuminate, continue with To recover from a fault for systems equipped with a SCL-
10Avp (SIL-HT) system controller on page 54.
RUO-IDV-05-0269-A 53 of 78
Shimadzu Devices
To recover from a fault for systems equipped with a SCL-10Avp (SIL-HT) system
controller
Note: In the event of a device fault, either within the Analyst software or at the device
itself, it may be difficult to reactivate or run the devices. If this occurs, perform the
following reboot sequence to regain control.
Note: You can control up to four pumps using the Shimadzu CBM system controller. A
provisional ROM is available for four-pump control through the SIL-HT (SCL-10Avp).
For more information, contact your local Shimadzu Representative.
54 of 78 RUO-IDV-05-0269-A
Shimadzu Devices
Note: For more information on setting up the system controller for other
pumps, see the appropriate Shimadzu manual.
RUO-IDV-05-0269-A 55 of 78
Shimadzu Devices
Note: The model number for each connected device appears on the
System Configuration screen. The message Remote appears on any
connected valve.
Note: The model number for each connected device appears on the
System Configuration screen. The message Remote appears on any
connected pump.
56 of 78 RUO-IDV-05-0269-A
Column Oven Configuration
6
This section provides information about the required hardware and how to connect a column
oven to the computer.
The Analyst® software supports the following column ovens:
• Agilent column oven.
• PerkinElmer Series 200 column oven.
• Shimadzu column ovens via the Shimadzu SCL-10Avp, CBM20A, or CBM20A lite
system controllers. See Shimadzu Devices on page 43.
The following table lists the required hardware. For the latest version of supported firmware, see
the most current Analyst software Release Notes.
Table 6-2 Required Hardware for the Agilent Column Ovens
Cable Other parts needed
• RS-232 cable (PN WC024736) • Network interface card (PN 1016082) if using a LAN
(Ethernet) connection
• GPIB cable (PN WC021365)
• Agilent PN G5183-4649 (for a direct LAN [Ethernet]
connection)
• Agilent PN G1530-61485 (for a LAN [Ethernet]
connection using a hub)
• CAN cable (ships with Agilent system)
RUO-IDV-05-0269-A 57 of 78
Column Oven Configuration
This procedure describes how to connect an Agilent column oven to the computer through
standard serial port communication.
1. Turn off the column oven.
2. Set the DIP switches at the back of the column oven. Make sure that the switches
are set for a baud rate of 19 200. For specific instructions on setting the DIP
switches, see Configuration of Agilent Devices Through Serial Port Communication
on page 9.
For the location of the DIP switches at the back of the column oven, see Figure 6-1.
Figure 6-1 Back panel of the Agilent column oven
1 2 3
Ba
Item Description
1 Serial connector
2 CAN connectors
3 DIP switches
3. Connect the RS-232 cable from the serial port at the back of the column oven to the
desired serial port on the computer, noting the port number.
58 of 78 RUO-IDV-05-0269-A
Column Oven Configuration
WARNING! See the PerkinElmer Series 200 Column Oven safety instructions
before configuring any AC mains-powered equipment.
RUO-IDV-05-0269-A 59 of 78
Column Oven Configuration
60 of 78 RUO-IDV-05-0269-A
Switching Valve Configuration
7
The Analyst® software supports the following switching valves:
• Valco two-position switching valve.
• Agilent switching valves. See Agilent Column Oven on page 57.
• Shimadzu internal valves using the Shimadzu CBM controller. See Shimadzu
Devices on page 43.
The following table lists the required hardware. For the latest version of firmware supported, see
the most current Analyst software Release Notes.
Table 7-1 Required Hardware for the Valco Valve
Cable Other parts needed
RS-232 cable (PN WC024740) WC027522 Valve kit and all accessories
RUO-IDV-05-0269-A 61 of 78
Switching Valve Configuration
Caution: Do not connect the round connector on this cable to the valve and
motor assembly at this time, as it will damage the valve setting.
2. Insert the five-wire connector of the Valco motor output cable into the receptacle at
the rear left of the Valco two-position actuator control module.
3. Connect the 10-wire Valco manual controller cable from the receptacle on the front
right of the Valco two-position actuator control module to the receptacle on the front
of the Valco manual controller.
The 10-wire cable should have a 10-wire connector on each end.
4. Connect the Valco power supply to the mains power.
5. On the Valco manual controller, cycle the actuator at least two times by pressing
Position A followed by Position B and so on.
Initialization is achieved when the position indicator lights on the actuator change
according to the position button pressed on the manual controller.
6. Insert the round connector of the motor driver output cable into the receptacle at the
rear underside of the valve and motor assembly.
7. Check the operation of the Valco kit by using the manual controller to change valve
positions several times.
62 of 78 RUO-IDV-05-0269-A
Switching Valve Configuration
8. Disconnect the Valco manual controller cable from the receptacle on the front of the
Valco two-position actuator control module. Store the manual controller and cable
until the next time it is needed.
2. Connect the 3-pin end of the RS-232 cable to the receptacle on the Valco two-
position actuator control module.
3. Connect the other end of the RS-232 cable to the desired 9-pin serial port on the
computer, noting the port number.
RUO-IDV-05-0269-A 63 of 78
Switching Valve Configuration
64 of 78 RUO-IDV-05-0269-A
Detector Configuration
8
The Analyst® software supports the following detectors:
• Agilent diode array detectors (DADs).
• Shimadzu UV-VIS detector (see Shimadzu Devices on page 43).
The following types of Agilent diode array detectors are supported by the Analyst software.
Table 8-1 Supported Detectors
Detector Model Number
Agilent DAD G1315 A, B, C, D
Agilent 1260 DAD G4212B
Agilent 1290 DAD G4212A
The Agilent G4212A and G4212B DADs have one lamp source instead of two, as in previous
DADs. As a result, the usable wavelength range has been changed to 190 nm to 640 nm.
The G4212A DAD supports slit widths up to 8 nm, and the G4212B DAD has a fixed slit width of
4 nm.
The following table lists the required hardware. For information about the latest version firmware
supported, see the most current Analyst software Release Notes.
Table 8-2 Required Hardware for the Agilent Detector
Cable Other parts needed
WC021365 (for GPIB connection) • Network interface card if using a LAN (Ethernet)
connection
Note: This cable is not required for
models that use a LAN (Ethernet) • Agilent PN G5183-4649 (for a direct LAN connection)
connection. • Agilent PN G1530-61485 (for a LAN connection using a
hub)
WARNING! Electrical Shock Hazard: See the Agilent Diode Array Detector safety
instructions before configuring any mains-powered equipment.
RUO-IDV-05-0269-A 65 of 78
Detector Configuration
1 2
Item Description
1 LAN port
2 GPIB port
4. Connect the other end of the cable (GPIB or LAN) to the computer.
66 of 78 RUO-IDV-05-0269-A
NIDAQ and Terminal Block Installation
9
Installing an ADC Card on a New Instrument
Current systems have the correct drivers installed. The list of supported devices may change.
See the most current Analyst® software Release Notes. The following are requirements for
earlier Analyst software versions:
• Analyst TF 1.5 software requires driver version 8.3 or higher
• Analyst 1.5 software requires driver version 8.3
• Analyst 1.4.1 software requires driver version 7.1 of the appropriate driver
• Analyst 1.4 software requires driver version 6.9.3
• All previous versions of the Analyst software require driver version 6.7
The current systems include the Measurement and Automation Explorer software. This software
is also installed on systems that had a GPIB board previously installed.
1. Insert the NIDAQ PCI board in the computer.
2. Open the Measurement and Automation Explorer software.
The left pane displays a list of available devices.
3. Expand the list to view the PCI 6032E ADC card.
If this card is in the list, then it is installed on the computer. There are some useful
tools within this software that you can use to monitor the input to the terminal block
without having to use the Analyst software. You can use an AA battery to supply a
test signal.
4. Attach the ADC terminal box.
The block is marked as having 16 analog channels mixed in with earthed channels
(marked as A1 1 to A1 16).
Note: Because the system uses Differential mode, the software has to
distinguish the voltage difference between the anode and cathode of the
variable wavelength detector, as opposed to grounding the cathode and
monitoring only the anode.
RUO-IDV-05-0269-A 67 of 78
NIDAQ and Terminal Block Installation
Figure 9-1 shows the polarity of the two cables: the positive polarity is channel 33
(A1 1) and the negative polarity is channel 66 (A1 9).
Figure 9-2 Cable connections
6. Make sure the DIP switches are set to Differential mode, as shown in Figure 9-2.
68 of 78 RUO-IDV-05-0269-A
NIDAQ and Terminal Block Installation
7. In the Analyst software, add the ADC card to the hardware profile as shown in
Figure 9-3. Make sure the settings are exactly as shown.
Figure 9-3 ADC configuration
Note: The ADC channel must be set to channel 2. This equates to channel
1 on the terminal block.
RUO-IDV-05-0269-A 69 of 78
NIDAQ and Terminal Block Installation
The Margin for Negative Absorbance field, shown in Figure 9-5, is variable. Adjust
it according to requirements.
Figure 9-6 Parameters
70 of 78 RUO-IDV-05-0269-A
NIDAQ and Terminal Block Installation
RUO-IDV-05-0269-A 71 of 78
NIDAQ and Terminal Block Installation
72 of 78 RUO-IDV-05-0269-A
Peripheral Device Analog Synchronization
A
The preferred method of synchronizing peripheral devices is through the Analyst® software
control. For devices that cannot be controlled through the Analyst software, synchronize through
the use of analog signals (contact closure).
RUO-IDV-05-0269-A 73 of 78
Peripheral Device Analog Synchronization
Figure A-2 Schematic of the AUX I/O interface and cable on the 5500 series of
instruments.
Ready Signal
The Ready signal is an autosampler Inject signal that is generated using a DPST (Double-Pole,
Single-Throw) relay. It provides either an NO (Normally Open) or NC (Normally Closed) contact
closure.
Note: The Ready signal is active only when the mass spectrometer is operated in LC
Sync mode. For more information on operating modes, see the Analyst software Help.
The Ready signal is activated when the LC/MS devices are ready to acquire data and are waiting
for an injection. As soon as the MS acquisition is started (by the START signal), READY is
deactivated. Do not confuse READY with the MS Ready status, which is not specific to the LC
Sync mode.
74 of 78 RUO-IDV-05-0269-A
Peripheral Device Analog Synchronization
Error Signal
The Error signal is used as an External Stop signal for any LC pumps connected to the ion
source to prevent accidental overflow of the source. An error is generated using a DPST relay
and provides either a NO or NC contact closure. The Error signal is active regardless of the MS
synchronization mode. The Error signal is activated for approximately five seconds when an MS
error occurs. The error type is non-specific and may include source, electronic, or vacuum
system failures.
Start Signal
The Start signal is given to the mass spectrometer to initiate data acquisition. This signal is
passed to the MS electronics through an optocoupler (a device that couples a light-emitting-diode
and a phototransistor to provide an isolated digital connection between the sender and receiver).
The Start signal may be any signal that creates a potential of between 2 to 8 volts across Pins 10
and 22. For example, a voltage pulse in the normal TTL range (2 to 5 volts) would be a START
signal.
By setting the MS synchronization trigger level, you can configure the Start signal as either Active
High or Active Low, as required. See Autosampler Configuration on page 15.
Use the biased +5V and ground signals provided on the AUX I/O port to:
• Generate the appropriate START using a contact closure.
• Generate TTL-level READY and ERROR signals.
RUO-IDV-05-0269-A 75 of 78
Peripheral Device Analog Synchronization
Note: Set the mass spectrometer Sync Mode to LC Sync in the acquisition method to
provide analog synchronization between the peripheral devices and the mass
spectrometer.
The following examples are used as guidelines for developing an analog synchronization
scheme for your peripheral devices. For more information about the types of signals generated
and required by the peripheral device, see the appropriate peripheral device documentation.
In both figures, in the center, wire colors are indicated as background/stripe.
Figure A-4 Analog synchronization scheme using contact closure signals
76 of 78 RUO-IDV-05-0269-A
CTC PAL Autosampler Setup Notes
B
This section provides an overview of the setup for the CTC PAL autosampler. With all versions of
the PAL autosampler, the only differences are in the frame size and the tray holders (or stacks)
bolted to the autosampler frame. In some cases, additional valves and accessories can be
attached.
The Analyst® software uses a software driver developed by CTC Analytics. The driver is
essentially the same as that used by the CTC software, Cycle Composer.
Note: The firmware required to operate the different models of autosampler is exactly
the same for all models when used with the Analyst software.
An FSE must configure the CTC autosampler firmware to indicate where the trays can be placed
and where everything is located in the X, Y, and Z dimensions. Use the handheld controller for
the autosampler to configure the PAL or use a separate utility from CTC to write the configuration
information into the autosampler's non-volatile memory.
The following terms are used to describe the Analyst software Batch Editor elements in relation
to the CTC.
Rack
CTC defines a rack as a drawer or tray that holds microtitre or vial plates. The Rack Position
designates where the rack is placed, and the Rack Code designates the type of rack.
Plate
CTC defines a plate as a microtitre plate or tray that holds vials. The Plate Code specifies the
type of plate and the plate position indicates where the plate sits on the rack.
Note: There is not a one-to-one mapping between a rack and the tray in CTC
terminology.
Tray
In the Analyst software, the term tray is used to define a physical location. A tray is a placeholder
for a location in which you can place different types of trays. The tray group indicates the tray
types you can use in each tray location.
The Analyst software imposes no restrictions on the number of tray types used in each location.
You can use all defined tray types in all tray locations, if required. With the Analyst software,
duplicate tray definitions are not required.
For every tray location on the autosampler, use the handheld controller for the autosampler to
verify and correct the position of each tray type. If any trays are incorrectly defined on the X, Y, or
Z dimension, the CTC driver cannot find the correct layout of the trays in the autosampler. This
RUO-IDV-05-0269-A 77 of 78
CTC PAL Autosampler Setup Notes
either causes the Analyst software to load the tray configuration incorrectly, which results in the
Batch Editor Locations tab displaying 6 tray locations, or it causes the Analyst software to not
indicate the trays that should be present.
Note: The AUX I/O triggers the mass spectrometer to start scanning through the
contact closure. If the mass spectrometer does not start scanning, it may be because
the CTC autosampler Sync Signal is not set to Immediate. This situation typically occurs
when the autosampler is being used as a standalone device without any controlling
software. The CTC autosampler has a handheld controller for the user to configure
settings in the autosampler. One of these settings is the Sync Signal. If you use the
autosampler by itself with no computer control, you may set this to wait for an external
ready signal. Under the Analyst software control, however, typically this is not needed.
Therefore, if the autosampler is configured incorrectly, it will sit and wait and not inject.
78 of 78 RUO-IDV-05-0269-A