RSA3000 Datasheet Espectrometro
RSA3000 Datasheet Espectrometro
Ultra-Real technology
Frequency: up to 4.5 GHz
Displayed average noise level (DANL): <-161 dBm (typical)
Phase noise: <-102 dBc/Hz (typical)
Level measurement uncertainty: <1.0 dB
4.5 GHz tracking generator
Min. RBW 1 Hz
Up to 40 MHz real-time analysis bandwidth
Multiple measurement modes
Various advanced measurement functions
EMI measurement application (option)
Vector network analyzer application
Multiple trigger modes and trigger masks
Density, spectrogram, and other display modes
PC software options
10.1'' capacitive multi-touch screen; supporting touch gestures
USB, LAN, HDMI and other communication and display interfaces
    RSA3000 Series Real-time Spectrum Analyzer
                                            Built-in Linux operating system reliable and stable interface
                                                                                                                     Built-in quad-core
                                                                                                                     processor
    10.1-inch capacitive                                                                                             high processing speed
     multi-touch screen
      supporting touch
                gestures
    Based on the Ultra-Real technology, the high-speed real-time measurement mode allows you to acquire the signals in
    the analysis bandwidth seamlessly and make data analysis. It also provides various display modes, such as Spectrogram,
    Density, and PVT. Besides, FMT function is also available.
2
FMT
Frequency mask trigger (FMT) to trigger the measurement by
sporadic or transient events in the spectrum
Composite displays
Spectrogram for gap-free display of the spectrum                      Density spectrum for you to visualize how frequently signals occur
 Integrates four measurement modes to address the challenges for multiple RF test requirements with
 one single instrument
 RSA3000 series provides EMI, RTSA, and VNA modes in addition to the traditional GPSA mode. Engineers may find it convenient to
 address multiple RF test challenges with just one instrument, effectively reducing their time and costs, greatly improving their working
 efficiency.
 Advanced measurement mode provides test items required                Quickly recall the limit line compliant with the CISPR standard
for the transmitter test such as multichannel power, ACP, and         (e.g. EN55011, EN55012, etc.) to carry out pre-test and monitor
occupied BW.                                                         the target point with three different detectors.
                                                                                                                                 3
     With the Density spectrum, you can find out the exceptional             In VNA mode, you can make S11, S21, and DTF measurements
    signals hidden behind the high-level signals, and capture them          for the components and circuit networks. The network
    accurately with the FMT.                                               characteristics of the components under test can be accurately
                                                                           demonstrated in Smith chart, Polar chart, and other formats.
    Various operation modes to improve your                                Multiple interfaces to improve the connectivity
     operation experience                                                     of the instruments
    The 10.1-inch capacitive multi-touch screen supports various           The instrument can be connected to a larger display/monitor via
    touch gestures, making it always keep up with the mainstream           the HDMI interface for better display effects. The Web Control
    development trend for screen operation. The gesture-enabled            function allows you to directly control the device by accessing the
    operation such as tapping, dragging, pinching & stretching makes       device IP address, improving the experience of remote control.
    the measurement action smooth and convenient, easy for you to
    operate the instrument. Meanwhile, the instrument still keeps the
    knob and key operation as what RIGOL traditional instruments
    have, optimizing the user-friendly interactive experience to a large
    extent. It also supports keyboard and mouse operation.
4
     Specifications
Specifications are valid under the following conditions: the instrument is within the calibration period, is stored for at least two hours
at 0℃ to 50℃ temperature, and is warmed up for 40 minutes. Unless otherwise noted, the specifications in this manual include the
measurement uncertainty.
Typical: characteristic performance, which 80 percent of the measurement results will meet at room temperature (approximately 25℃).
This data is not warranted and does not include the measurement uncertainty.
Nominal: the expected mean or average performance or a designed attribute (such as the 50 Ω connector). This data is not warranted
and is measured at room temperature (approximately 25℃).
Measured: an attribute measured during the design phase which can be compared to the expected performance, such as the amplitude
drift variation with time. This data is not warranted and is measured at room temperature (approximately 25℃).
NOTE: All charts in this manual are the measurement results of multiple instruments at room temperature unless otherwise noted. The specifications
(except the tracking generator specifications) listed in this manual are those when the tracking generator is off.
 Measurement Mode
 Measurement Mode
 General-Purpose Spectrum Analyzer (GPSA)
 Real-Time Spectrum Analyzer (RTSA)
 EMI Measurement Application (EMI)
 Vector Network Analyzer Application(VNA)
 Measurement Mode and Product Model Adaptation Table
                                     RSA3015N            RSA3030           RSA3030-TG           RSA3030N             RSA3045     RSA3045-TG       RSA3045N
             GPSA                         √                   √                   √                   √                  √           √                 √
             RTSA                         √                   √                   √                   √                  √           √                 √
              EMI                         O                   O                   O                  O                   O           O                O
              VNA                         √                   ×                   ×                   √                  ×           ×                 √
    Tracking Generator                    √                   ×                   √                   √                  ×           √                 √
Note: x indicates not supported; √ indicates standard configuration; O indicates optional configuration.
      The RSA3000N models include hardware capability not in the RSA3000-TG. The RSA3000-TG models cannot be used in VNA mode.
                                                                                                                                                       5
    GPSA Mode
    Frequency
     Frequency Readout Accuracy
     Marker Frequency Resolution                                      span/(number of sweep points - 1)
                                                                      ±(marker frequency readout x reference frequency accuracy + 1% x span + 10% x
     Marker Frequency Uncertainty
                                                                      resolution bandwidth + marker frequency resolution + LO sweep resolution[1])
     Frequency Counter
     Resolution                                                       1 Hz
     Uncertainty                                                      ±(marker frequency readout x reference frequency accuracy + counter resolution)
     Frequency Span
                                      Standard                        0 Hz, 100 Hz to maximum frequency
     Range
                                      Option RSA3000-BW1              0 Hz, 10 Hz to maximum frequency
     Resolution                                                       2 Hz
     Uncertainty                                                      ±[0.25% x span + span/(number of sweep points - 1) + 12 Hz]
     SSB Phase Noise
                                                                      20℃ to 30℃,fc = 500 MHz
                                  1 kHz                               <-90 dBc/Hz (typical)
                                  10 kHz                              <-100 dBc/Hz, <-102 dBc/Hz (typical)
     Carrier Offset
                                  100 kHz                             <-100 dBc/Hz, <-102 dBc/Hz (typical)
                                  1 MHz                               <-110 dBc/Hz, <-112 dBc/Hz (typical)
RSA3015N/RSA3030/RSA3030-TG/RSA3030N/RSA3045/RSA3045-TG/RSA3045N
    Residual FM
                                                                      20℃ to 30℃ , RBW = VBW = 1 kHz
    Residual FM                                                       <10 Hz (nominal)
    Bandwidth
                                                                      Set "Sweep Time Rule" to "Accy"
    Resolution Bandwidth Standard                                     1 Hz to 3 MHz, in 1-3-10 sequence
    (-3 dB)[2]           Option RSA3000-BW1                           1 Hz to 10 MHz, in 1-3-10 sequence
                                                                      10 Hz to 1 kHz, <15% (nominal)
    RBW Accuracy                                                      3 kHz to 1 MHz, <5% (nominal)
                                                                      3 MHz to 10 MHz, <15% (nominal)
    Resolution Filter Shape Factor (60 dB: 3 dB)                      <5 (nominal)
    Video Bandwidth (-3 dB)                                           1 Hz to 10 MHz, in 1-3-10 sequence
    Resolution Bandwidth (-6 dB) (Option RSA3000-
                                                  200 Hz, 9 kHz, 120 kHz, 1 MHz
    EMC)
6
Amplitude
 Measurement Range
                                                                    fc ≥ 10 MHz
 Range
                                                                    DANL to +30 dBm
 Maximum Safe Input Level[1]
 DC Voltage                                                         50 V
                                                                    +30 dBm, attenuation ≥ 40 dB, preamp off.
 CW RF Power
                                                                    -10 dBm, attenuation = 20 dB, preamp on.
 Maximum Damage Level
 CW RF Power                                                        +33 dBm (2 W)
RSA3015N/RSA3030/RSA3030-TG/RSA3030N/RSA3045/RSA3045-TG/RSA3045N RSA3015N/RSA3030/RSA3030-TG/RSA3030N/RSA3045/RSA3045-TG/RSA3045N
Note: [1] When fc < 10 MHz, the maximum safe input level is decreased.
                                                                                                                                                                       7
                                                            RSA3015N/RSA3030/RSA3030-TG/RSA3030N/RSA3045/RSA3045-TG/RSA3045N
    Level Display
    Logarithmic Scale                                   1 dB to 200 dB
    Linear Scale                                        0 to reference level
    Number of Display Points                            801
    Number of Traces                                    6
                                                        normal, pos-peak, neg-peak, sample, RMS average, voltage average, and
    Trace Detector
                                                        quasi-peak (Option RSA3000-EMC)
    Trace Function                                      clear write, max hold, min hold, average, view, blank
    Scale Unit                                          dBm, dBmV, dBμV, nV, μV, mV, V, nW, μW, mW, W
    Frequency Response
                                                        attenuation = 10 dB, relative to 50 MHz, 20℃ to 30℃
                     100 kHz to 3.0 GHz                 <0.7 dB, <0.5 dB (typical)
    Preamp off
                     3.0 GHz to 4.5 GHz                 <0.9 dB, <0.5 dB (typical)
                                                        attenuation = 0 dB, relative to 50 MHz, 20℃ to 30℃
                     100 kHz to 3.0 GHz                 <1.0 dB, <0.5 dB (typical)
    Preamp on
                     3.0 GHz to 4.5 GHz                 <1.2 dB, <0.5 dB (typical)
RSA3015N/RSA3030/RSA3030-TG/RSA3030N/RSA3045/RSA3045-TG/RSA3045N RSA3015N/RSA3030/RSA3030-TG/RSA3030N/RSA3045/RSA3045-TG/RSA3045N
8
Input Attenuation Switching Uncertainty
Setting Range                                      0 dB to 50 dB, in 1 dB step
                                                   fc = 50 MHz, relative to 10 dB, preamp off, 20℃ to 30℃
Switching Uncertainty
                                                   <0.3 dB
RSA3015N/RSA3030/RSA3030-TG/RSA3030N/RSA3045/RSA3045-TG/RSA3045N of RSA3015N/RSA3030/RSA3030-TG/RSA3030N/RSA3045/RSA3045-TG/RSA3045N
                                                                                                                                                                              9
     RF Input VSWR
                                                        attenuation ≥ 10 dB, preamp off
                             300 kHz to 3.0 GHz <1.6 (nominal)
     VSWR
                             3.0 GHz to 4.5 GHz         <1.8 (nominal)
     Distortion
                                                        fc ≥ 50 MHz, input signal level = -20 dBm, attenuation = 0 dB, preamp off.
     Second Harmonic Intercept (SHI)
                                                        +45 dBm
                                                        fc ≥ 50 MHz, two -20 dBm tones at input mixer spaced by 200 kHz, attenuation = 0 dB,
     Third-order Intercept (TOI)                        preamp off.
                                                        +10 dBm, +15 dBm (typical)
                                                        fc ≥ 50 MHz, attenuation = 0 dB, preamp off
     1 dB Gain Compression (P1dB)[1]
                                                        0 dBm (norminal)
of RSA3015N/RSA3030/RSA3030-TG/RSA3030N/RSA3045/RSA3045-TG/RSA3045N
     Spurious Response
                                                        input terminated with a 50 Ω load, attenuation = 0 dB, 20℃ to 30℃
     Residual Response
                                                        <-90 dBm, <-100 dBm (typical)
     Intermediate Frequency                             <-60 dBc
                                                        referenced to local oscillators, referenced to A/D conversion, referenced to subharmonic of
     System-related Sideband                            first LO, referenced to harmonic of first LO
                                                        <-60 dBc
                                                        mixer level = -30 dBm
     Input-related Spurious
                                                        <-60 dBc
 Note: [1] The frequency interval of the two-tone signals should be greater than 10 MHz.
10
Sweep
Sweep
                   span ≥ 10 Hz                                1 ms to 4,000 s
Sweep Time
                   zero span                                   1 μs to 6,000 s
                   span ≥ 10 Hz, RBW ≥ 1 kHz                   5% (nominal)
Sweep Time
Uncertainty        zero span (sweep time > 1 ms) 5% (nominal)
Trigger
Trigger
Trigger Source                                                 free run, external 1, external 2, video
                    span ≥ 10 Hz                               0 to 500 ms
Trigger Delay
                    zero span                                  0 to 500 ms
Tracking Generator
Tracking Generator Output
                                         RSA3015N                                 RSA3030-TG/N                                RSA3045-TG/N
Frequency Range                          100 kHz to 1.5 GHz                       100 kHz to 3.0 GHz                          100 kHz to 4.5 GHz
Output Level Range                       -40 dBm to 0 dBm
Output Level Resolution                  1 dB
                                         relative to 50 MHz
Output Flatness
                                         ±3 dB (nominal)
              @0 dBm of RSA3015N/RSA3030-TG/RSA3030N/RSA3045-TG/RSA3045N
                                                                                                    @0 dBm of RSA3015N/RSA3030-TG/RSA3030N/RSA3045-TG/RSA3045N
                                                                                                                                                                 11
 RTSA Mode
                                                       10 MHz
 Real-time Analysis Bandwidth                          25 MHz (Option RSA3000-B25)
                                                       40 MHz (Option RSA3000-B40)
                                                       maximum span, default Kaiser Window
 Min. Signal Duration for 100% POI at                  9.3 μs
 the Full-Scale Accuracy                               7.82 μs (Option RSA3000-B25)
                                                       7.45 μs (Option RSA3000-B40)
 Trace Detector                                        pos-peak, neg-peak, sample, average
 Number of Traces                                      6
 Window Type                                           Hanning, Blackman-Harris, Rectangular, Flattop, Kaiser, and Gaussian
                                                       provides 6 RBWs for each window, except the Rectangular;
                                                       for Kaiser window
                                                       Span                              Min. bandwidth             Max. bandwidth
                                                       40 MHz                            100 kHz                    3.21 MHz
 Resolution Bandwidth
                                                       25 MHz                            62.8 kHz                   2.01 MHz
                                                       10 MHz                            25.1 kHz                   804 kHz
                                                       1 MHz                             2.51 kHz                   80.4 kHz
                                                       100 kHz                           251 Hz                     8.04 kHz
 Max. Sample Rate                                      51.2 Msa/s
 FFT Rate                                              146,484/s (norminal)
 Number of Markers                                     8
 Amplitude Resolution                                  0.01 dB
 Frequency Point                                       801
                                                       Max. sample rate
 Acquisition Time
                                                       >156.5 μs
 Min. Signal Duration for 100% POI at Different RBWs
                                                       Duration Time (μs)
 Span                                                  RBW1            RBW2              RBW3             RBW4       RBW5            RBW6
 40 GHz                                                26.9            16.9              11.9             9.32       8.07            7.45
 25 MHz                                                38.9            22.9              14.9             10.9       8.82            7.82
 10 MHz                                                86.8            46.8              26.8             16.8       11.8            9.30
 1 MHz                                                 807             407               207              107        56.3            31.3
 Amplitude
 Amplitude Flatness                                    ±0.5 dB[1] (nominal)
 SFDR                                                  <-50 dBc/Hz (typical)
         Density
 Probability Range                                     0 to 100% (with a step of 0.1%)
 Min. Span                                             5 kHz
 Persistence Duration                                  32 ms to 10 s
         Spectrogram
 History Depth                                         8,192
 Dynamic Range Covered by Bitmap
                                 200 dB
 Color
           PVT
 Min. Acquisition Time                                 187.9 μs
 Max. Acquisition Time                                 40 s
 Trigger
 Trigger Source                                        free run, external 1, external 2, power(time), FMT
           FMT
 Trigger Diagram                                       density, spectrogram, normal, PVT
 Trigger Resolution                                    0.5 dB (nominal)
 Trigger Criteria                                      enter, leave, inside, outside, enter-leave, leave-enter
 Note:[1] Only applicable to the Normal measurement.
12
EMI Mode (Option RSA3000-EMI)
VNA Mode
Measurement Setup
                               RSA3015N                        RSA3030N                       RSA3045N
Frequency Range
                               10 MHz~1.5 GHz                  10 MHz~3.0 GHz                 10 MHz~4.5 GHz
Measurement Type               Reflection(S11), Transmission(S21), Distance-to-fault (DTF)
Measurement Bandwidth          1 kHz~10 MHz(in 1-3-10 sequence)
Data Points                    101~10001; default 201
Trace Type                     mem, math, clear write, average, max hold, min hold,
Number of Markers              8
Mechanical Calibration Kit     Open, Short, Load, Through; User Calibration Kit
Transmission Measurement S21
Port Output Power              -10 dBm (nom.)
Format                         Lin Mag, Log Mag, Phase, Group Delay
Magnitude Range                -500 G to 500 G
Magnitude Resolution           Log: 100f; Lin 1a
                               S21, RBW=10 kHz, Port1 level=0 dBm, Log Mag, Average=50
Dynamic Range
                               80 dB (nom.)
Reflection Measurement S11
Port Output Power              -10 dBm (nom.)
                               Lin Mag, Log Mag, Phase, Group Delay, SWR,
Format                         Smith Chart (Lin/Phase, Log/Phase, Real/Imag, R+j*X, G+j*B),
                               Polar Chart (Lin/Phase, Log/Phase, Real/Imag)
Magnitude Range                -500 G to 500 G
Magnitude Resolution           Log: 100f; Lin 1a
VSWR Range                     -500 G to 500 G
Corrected Directivity          S11, Log Mag, Average=50
(With CK106A)                  > 40 dB (nom.)
                                                                                                               13
                     Measured return loss of RSA3000N port 1                                                Measured return loss of RSA3000N port 1
 Distance to Fault(DTF)
 Port Output Power                                    0 dBm (nom.)
 Format                                               Lin Mag, Log Mag, SWR
 Maximum Distance (meter)                             8.0 x 1010 x Velocity Factor/Span
 Fault Resolution (meter)                             1.5 x 108 x Velocity Factor/Span
 Windows                                              Gaussian, Flattop, Rectangular, Hanning, Hamming
 Velocity Factor                                      0.1~1
General Specifications
 Display
 Type                                                          capacitive multi-touch screen
 Resolution                                                    1024 x 600 pixels
 Size                                                          10.1''
 Color                                                         24-bit color
 Printer Supported
 Protocol                                                      network printer
 Mass Memory
                          Internal Storage                     512 MB (nominal)
 Mass Memory
                          External Storage                     USB storage device (not supplied)
 Power
 Input Voltage Range, AC                                       100 V to 240 V (nominal)
 AC Frequency                                                  45 Hz to 440 Hz
 Power Consumption                                             55 W (typical), max. 90 W with all options
 Environment
                          Operating Temperature
                                                               0℃ to 50℃
                          Range
 Temperature
                          Storage Temperature
                                                               -20℃ to 70℃
                          Range
                          0℃ to 30℃                            ≤ 95% RH
 Humidity
                          30℃ to 40℃                           ≤ 75% RH
 Altitude                 Operating Height                     below 3,048 m (10,000 feet)
14
Electromagnetic Compatibility and Safety
                        complies with EMC Directive 2014/30/EU,
                        complies with or above the standard specified in IEC61326-1:2013/EN61326-1:2013 Group 1 Class A
                        CISPR 11/EN 55011
                        IEC 61000-4-2:2008/EN
                                              ±4.0 kV (contact discharge), ±8.0 kV (air discharge)
                        61000-4-2
                        IEC 61000-4-3:2002/EN
                                              3V/m (80 MHz to 1 GHz); 3V/m (1.4 GHz to 2 GHz); 1V/m (2.0 GHz to 2.7 GHz)
                        61000-4-3
 EMC                    IEC 61000-4-4:2004/EN
                                              1 kV power
                        61000-4-4
                        IEC 61000-4-5:2001/EN 0.5 kV (phase-to-neutral voltage); 1 kV (phase-to-earth voltage);
                        61000-4-5             1 kV (neutral-to-earth voltage)
                        IEC 61000-4-6:2003/EN
                                              3 V, 0.15 to 80 MHz
                        61000-4-6
                        IEC 61000-4-11:2004/EN voltage dip: 0% UT during half cycle; 0% UT during 1 cycle; 70% UT during 25 cycles
                        61000-4-11             short interruption: 0% UT during 250 cycles
                                                 complies with IEC 61010-1:2010 (Third Edition)/EN 61010-1:2010, UL 61010-1:2012
Safety
                                                 R4.16 and CAN/CSA-C22.2 No. 61010-1-12+ GI1+ GI2
                                                 Samples of this product have been type tested in accordance with RIGOL's reliability
                                                 test regulations and verified to be robust against the environmental stresses of
Environmental Stress                             storage, transportation, and end-use; those stresses include, but are not limited to,
                                                 temperature, humidity, shock, and vibration. The test methods are compliant with
                                                 standards specified in GB/T6587 Class 2 and MILPRF-28800F Class 3.
Size
                                                 410 mm x 224 mm x 135 mm
(W x H x D)
                                                 (16.14'' x 8.82'' x 5.32'')
Weight
Without Tracking Generator                       4.65 kg (10.25 lb)
With Tracking Generator                          4.95 kg (10.91 lb)
Calibration Interval
Recommended Calibration Interval                 18 months
Input/Output
                                                                                                                                15
 External Trigger Input/Output
                                           Impedance                       ≥ 1 kΩ (nominal)
 External Trigger Input 1                  Connector                       BNC female
                                           Level                           5 V TTL level
                                                       on trigger input    ≥ 1 kΩ (nominal)
                                           Impedance
                                                       on trigger output   50 Ω (nominal)
 External Trigger Input 2/Trigger Output
                                           Connector                       BNC female
                                           Level                           5 V TTL level
 IF Output
                                           Frequency                       430 MHz ± 20 MHz (nominal)
                                                                           RF input power (PRFin) ≤ -10 dBm, attenuation = 0,
                                                                           preamp off.
                                           Amplitude                       50MHz, PRFin ± 4 dB (nominal)
 IF Output
                                                                           other frequency, P RFin ± 4 dB + RF frequency
                                                                           response (nominal)
                                           Impedance                       50 Ω (nominal)
                                           Connector                       SMB male
 Communication Interface
                                           Connector                       A plug
 USB Host (4 ports)
                                           Protocol                        version 2.0
                                           Connector                       B plug
 USB Device
                                           Protocol                        version 2.0
                                           Connector                       100/1000Base, RJ-45
 LAN
                                           Protocol                        LXI Core 2011 Device
                                           Connector                       A plug
 HDMI
                                           Protocol                        HDMI 1.4b
16
  Order Information
               Description                                                                                Order No.
               Real-time Spectrum Analyzer, 9 kHz to 1.5 GHz (include TG and VNA)                         RSA3015N
               Real-time Spectrum Analyzer, 9 kHz to 3.0 GHz                                              RSA3030
               Real-time Spectrum Analyzer, 9 kHz to 4.5 GHz                                              RSA3045
 Model         Real-time Spectrum Analyzer, 9 kHz to 3.0 GHz (include TG)                                 RSA3030-TG
               Real-time Spectrum Analyzer, 9 kHz to 4.5 GHz (include TG)                                 RSA3045-TG
               Real-time Spectrum Analyzer, 9 kHz to 3.0 GHz (include TG and VNA)                         RSA3030N
               Real-time Spectrum Analyzer, 9 kHz to 4.5 GHz (include TG and VNA)                         RSA3045N
 Standard      Quick Guide (hard copy)                                                                    -
 Accessories   Power Cord                                                                                 -
               EMI Measurement Application (includes RSA3000-EMC)                                         RSA3000-EMI
               Preamplifier (PA)                                                                          RSA3000-PA
               High Stability Clock                                                                       OCXO-C08
               RBW 1 Hz to 10 MHz                                                                         RSA3000-BW1
               Real-time Analysis Bandwidth 25 MHz                                                        RSA3000-B25
 Option        Real-time Analysis Bandwidth 40 MHz                                                        RSA3000-B40
               Advanced Measurement Kit                                                                   RSA3000-AMK
               EMC Filter and Quasi-Peak Detector Kit                                                     RSA3000-EMC
               Spectrum Analyzer PC Software                                                              Ultra Spectrum
                                                                                                          S1210 EMI Pre-compliance
               EMI Pre-compliance Test Software
                                                                                                          Software
               High-performance Network Analysis Calibration Kit(frequency range: DC to 6.5 GHz)          CK106A
               Economical Network Analysis Calibration Kit(frequency range: DC to 1.5 GHz)                CK106E
               Include: N-SMA cable, BNC-BNC cable, N-BNC adaptor, N-SMA adaptor, 75 Ω-50 Ω
                                                                                                          DSA Utility Kit
               adaptor, 900 MHz/1.8 GHz antenna (2pcs), 2.4 GHz antenna (2pcs)
               Include: N(F)-N(F) adaptor (1pcs), N(M)-N(M) adaptor (1pcs), N(M)-SMA(F) adaptor (2pcs),
               N(M)-BNC(F) adaptor (2pcs), SMA(F)-SMA(F) adaptor (1pcs), SMA(M)-SMA(M) adaptor
                                                                                                          RF Adaptor Kit
               (1pcs), BNC T type adaptor (1pcs), 50 Ω SMA load (1pcs), 50 Ω BNC impedance adaptor
               (1pcs)
               Include: 50 Ω to 75 Ω adaptor (2pcs)                                                       RF CATV Kit
 Optional
               Include: 6 dB attenuator (1pcs), 10 dB attenuator (2pcs)                                   RF Attenuator Kit
 Accessories
               30 dB high-power attenuator, with the max power of 100 W                                   ATT03301H
               N(M)-N(M) RF Cable                                                                         CB-NM-NM-75-L-12G
               N(M)-SMA(M) RF Cable                                                                       CB-NM-SMAM-75-L-12G
               VSWR Bridge, 1 MHz to 3.2 GHz                                                              VB1032
               VSWR Bridge, 2 GHz to 8 GHz                                                                VB1080
               Near-field Probe                                                                           NFP-3
               Rack Mount Kit                                                                             RM6041
               USB Cable                                                                                  CB-USBA-USBB-FF-150
Warranty
Three years for the mainframe
                                                                                                                              17