Manual SL-9000S SL-9000L
Manual SL-9000S SL-9000L
HI 8033 - HI 8633
HI 8733 - HI 8734
HI 933000
Portable Multi-Range
Conductivity/TDS Meters
mS
ins rum
HI 8633
Conduct ns
ivity met
er
COND.
TEMP. ON
OF F
199.9
µS 1999
µS
19.99
mS 199.9
mS
TEMPE
RATUR
E
TABLE OF CONTENTS
2
PRELIMINARY EXAMINATION
GENERAL DESCRIPTION
3
FUNCTIONAL DESCRIPTION &
SPECIFICATIONS OF HI 8033
mS
1
Conductivity meter
HI 8033
µS
ppm salinity
0-199,9
CaCO3 0-19,990 µS
0-1999
OFF µS
0-19,990
°C k%
0 10 10 15
20 + 5
3 30
0
5
4
-
50 40 10 15
ins rum n s
4
FUNCTIONAL DESCRIPTION &
SPECIFICATIONS OF HI 8633
mS
1
6
ins rum n s
HI 8633 Conductivity meter
COND. ON
3 TEMP. OFF
2
199.9 1999
µS µS
4
19.99 199.9
mS mS
5
TEMPERATURE
5
FUNCTIONAL DESCRIPTION &
SPECIFICATIONS OF HI 8733
mS
1
5
ins rum n s
HI 8733 ON
Conductivity
meter OFF
2
199.9 1999
µS µS
3
19.99 199.9
mS mS
0 2.5
4 0.5 2
1 1.5
TEMPERATURE COEFFICIENT
6
FUNCTIONAL DESCRIPTION &
SPECIFICATIONS OF HI 8734
7
FUNCTIONAL DESCRIPTION &
SPECIFICATIONS OF HI 933000
8
OPERATIONAL GUIDE
• Each meter is supplied complete with a 9V battery. Slide off the
battery compartment cover on the back of the meter (see page 22).
Install the battery while paying attention to its polarity.
• Connect the probe to the meter securely by aligning the pins with
the socket and pushing the plug in.
With HI 933000, tighten the threaded ring.
• Make sure that the meter has been calibrated before taking any
measurements (see page 11 for calibration procedure).
• Immerse the conductivity probe into the
sample, with the holes on the shaft com-
pletely submerged.
If possible, use plastic beakers or con-
tainers to minimize any EMC interference.
• Tap the probe lightly on the bottom of the beaker to remove any
air bubbles which may be trapped inside the PVC sleeve.
• Turn the instrument on by pressing the ON
ppm salinity
µS
OFF 0-199,9
OFF µS
0-19,990
surement range.
• For HI 8033, HI 8633 and HI 8734 only: °C
9
With HI 8633 and HI 8734 press the COND/TEMP key (for
HI 8633) or TDS/TEMP (for HI 8734) to display the temperature
and adjust the temperature knob to that of the solution e.g. 20°C.
COND.
TEMP.
TDS ºC
TEMP.
10
CALIBRATION
Accessories needed:
• Use any calibration solution within the meter's range. The
solution should ideally be close to the samples being mea-
sured. Use for example HI 7030 or HI 8030, 12880 µS/cm
(=12.88 mS/cm) conductivity solution for HI 8033, HI 8633,
HI 8733 and HI 933000 and HI 7032, 1382 mg/L
(=2764 µS/cm) TDS solution for HI 8734
• ChecktempC or an accurate thermometer with 0.1°C resolution
(not necessary for HI 8733 and HI 933000)
• a small screwdriver.
PROCEDURE FOR HI 8033
0
7 03
• Pour sufficient quantity of a conductivity
HI
calibration solution (e.g. HI 7030/
HI 8030) into a beaker to cover the holes
on the probe. If possible, use plastic bea-
kers to minimize any EMC interference.
• Immerse the conductivity probe, making °C
µS
ppm salinity
0-199,9
CaCO3 0-19,990 µS
µS
range.
0-19990
• Turn the k% calibration knob until the display shows the conduc-
tivity reading at 25°C (see the conductivity vs. temperature chart
on page 18). E.g. @ 25°C, 12880 µS/cm = 12.88 mS/cm.
k%
10 15 mS
+ 5
0
5
-
10 15
11
• All subsequent measurements will be compensated to 25°C (77°F).
If you prefer to standardize the temperature compensation to
20°C (68°F) rather than 25°C (77°F), leave the °C knob at 18°C
(if the temperature of your solution is 18°C), adjust the trimmer
to read "11.67 mS" (see the conductivity vs. temperature chart at
page 18). All subsequent measurements will be compensated to
20°C.
• The calibration is now complete and the instrument is ready for
use.
The instrument should be re-calibrated at least once a month, or
when the probe or battery is changed.
Note: For more accurate results, it is advisable to use a calibration
solution close to the range to be measured. See the accessories
section on page 23 for a wide selection of conductivity solu-
tions.
PROCEDURE FOR HI 8633
0
703
HI
• Pour sufficient quantity of a conductivity
calibration solution (e.g. HI 7030/
HI 8030) into a beaker to cover the holes
on the probe. If possible, use plastic bea-
kers to minimize any EMC interference.
• Immerse the conductivity probe, making
°C
COND.
• Press COND/TEMP to display the temperature. TEMP.
TEMPERATURE
12
COND.
• Press COND/TEMP again to display conductivity
TEMP.
measurement.
ins rum
HI 8633
Conduct ns
ivity mete
r
COND.
TEMP. ON
199.9
OFF
mS
µS 1999
µS
19.99
mS 199.9
mS
TEMPER
ATURE
13
PROCEDURE FOR HI 8733
30 70
HI
• Pour sufficient quantity of a conductivity
calibration solution (e.g. HI 7030/
HI 8030) into a beaker to cover the holes
on the probe. If possible, use plastic bea-
kers to minimize any EMC interference.
• Immerse the conductivity probe in the
solution, making sure that holes are com-
pletely submerged.
• Wait for a couple of minutes for thermal
equilibrium to be reached.
• Tap the probe on the bottom, then shake it while rotating to
make sure no air bubbles remain trapped in the sleeve.
• Switch the instrument on by pressing ON
ON/OFF. OFF
1 1.5
page 20).
19.99
mS
• Select 19.99 mS/cm range by pressing
the appropriate range key.
mS
14
The instrument should be re-calibrated at least once a month, or
when the probe or battery is changed.
Note: For more accurate results, it is advisable to use a calibration
solution close to the range to be measured. See the accessories
section on page 23 for a wide selection of conductivity solu-
tions.
PROCEDURE FOR HI 8734
0
7 03
• Pour sufficient quantity of a TDS calibra-
HI
tion solution (e.g. HI 7032) into a beaker
to cover the holes on the probe. If pos-
sible, use plastic beakers to minimize any
EMC interference.
• Immerse the TDS probe, making sure
°C
TEMPERATURE
TDS
• Press TDS/TEMP again to display the TDS TEMP.
measurement.
15
• Adjust the calibration trimmer (see page 7,#6) on the side of the
instrument with the calibration screwdriver until the display shows
the TDS reading at 25°C (see the TDS vs. temperature chart on
page 19). E.g. @25°C "1382 mg/L".
TEMPER
ATURE
0
703
• Pour sufficient quantity of a conductivity
HI
calibration solution (e.g. HI 7030/
HI 8030) into a beaker to cover the holes
on the probe. If possible, use plastic bea-
kers to minimize any EMC interference.
• Immerse the conductivity probe in the
solution, making sure that holes are com-
pletely submerged.
• Wait for a couple of minutes for thermal
equilibrium to be reached.
• Tap the probe on the bottom, then shake it while rotating to make
sure no air bubbles remain trapped in the sleeve.
16
ON
• Switch the instrument on by pressing ON/OFF. OFF
mS
9 VOLT BATTERY
17
CONDUCTIVITY VERSUS
TEMPERATURE CHART
The conductivity of an aqueous solution is the measure of its ability to
carry an electrical current by means of ionic motion.
The conductivity invariably increases with increasing temperature.
It is affected by the type and number of ions in the solution and by
the viscosity of the solution itself. Both parameters are temperature
dependent. The dependency of conductivity on temperature is ex-
pressed as a relative change per degree Celsius at a particular
temperature, commonly as percent per °C.
For manual temperature compensation, refer to the following chart:
18
TDS VERSUS
TEMPERATURE CHART
The TDS value in aqueous solutions is directly proportional to conduc-
tivity. The ratio between the two parameters depends on the solution
and usually it is set to a factor of 0.5 (corresponding to a solution of
CaCO3). This means that 1 µS/cm is equal to 0.5 mg/L (ppm) of TDS.
For manual temperature compensation, refer to the following chart:
°C °F HI 7032 HI 7036
mg/L g/L
(ppm) (ppt)
0 32 758 6.82
5 41 876 7.88
10 50 999 8.99
15 59 1122 10.10
16 60.8 1148 10.33
17 62.6 1173 10.56
18 64.4 1200 10.78
19 66.2 1224 11.01
20 68 1251 11.24
21 69.8 1277 11.47
22 71.6 1303 11.71
23 73.4 1329 11.94
24 75.2 1358 12.18
25 77 1382 12.41
26 78.8 1408 12.65
27 80.6 1438 12.89
28 82.4 1461 13.13
29 84.2 1476 13.37
30 86 1515 13.61
31 87.8 1541 13.85
For instance, the TDS values of the calibration solutions at 25°C are
1382 mg/L or 12.41 g/L when using HI 7032 or HI 7036, respec-
tively.
At 20°C, the values are 1251 mg/L or 11.24 g/L, respectively.
With the solutions at 30°C, the values are 1515 mg/L or 13.61 g/L,
respectively.
19
DETERMINING THE TEMPERATURE
COEFFICIENT OF A SOLUTION (HI 8733)
Highly acidic, alkaline samples or solutions with high salt content
might have a different coefficient than the customary 2% per degree °C.
In order to calculate this coefficient follow the procedure below:
• Immerse the probe of HI 8733 in the sample and adjust the
TEMPERATURE COEFFICIENT knob to 0% (i.e. no compensation).
0 2.5
0.5 2
1 1.5
TEMPERATURE COEFFICIENT
• Condition the sample and probe to 25°C and note the conductiv-
ity reading, C25.
• Condition the sample and probe to a different temperature t°C
(approximately 10°C different from 25°C) and note the conduc-
tivity reading Ct.
• The temperature coefficient ß of the solution is calculated as given
by the following formula:
(C t - C25)
ß = 100 x
(t - 25) x C 25
The above procedure is suitable for determining the temperature
coefficient in a laboratory or where the temperature of the solution
can be controlled.
If this is not possible (e.g. on-site measurements), the following
procedure can be used providing the sample temperature varies by at
least 5°C or preferably 10°C:
• Immerse the probe of HI 8733 in the test solution and turn the
TEMPERATURE COEFFICIENT knob to 0% (no compensation).
• Check the conductivity reading and record the value. Make sure
the reading is stable, i.e. no greater variations than ±0.2 mS/
cm within a minute.
• Repeat the procedure when the temperature of the test solution
has changed by at least 5°C. Wait for the conductivity reading to
stabilize.
• Adjust the TEMPERATURE COEFFICIENT knob until the display
shows the same value as recorded earlier.
• The value indicated by the knob is the temperature coefficient of
the solution.
20
PROBE MAINTENANCE
21
BATTERY REPLACEMENT
When the battery becomes weak the meters will display "V" or an
additional blinking decimal point (HI 8733).
When the low battery indicator appears, the battery has only a few
hours left. A low battery will result in unreliable measurements.
It is recommended that the battery be replaced immediately.
Battery replacement must only take place in a non-hazardous area
using an alkaline 9V battery.
Slide off the battery compartment cover at the rear of the meter and
replace the 9V battery with a new one. Make sure the battery contacts
are tight and secure before replacing the cover.
HI 933000
9 VOLT BATTERY
HI 8033
HI 8633
SLIDE
HI 8733 OFF
HI 8734
BATTERY
SLIDE
OFF
22
SHOCKPROOF RUBBER BOOTS
( for H I 933000 only
only))
These rubber boots are specially
made to prolong the life of your
instrument and to prevent dam-
age due to accidental bangs and mS
falls.
They measure 155 x 90 x 45 mm
CONDUCTIVITY
METER
ON
OFF
HI 710008 = orange.
mS mS
HI933000
ACCESSORIES
CONDUCTIVITY & TDS BUFFER SOLUTIONS
HI 7030L 12880 µS/cm (µmho/cm), 460mL
HI 7030M 12880 µS/cm (µmho/cm), 230mL
HI 7031L 1413 µS/cm (µmho/cm), 460mL
HI 7031M 1413 µS/cm (µmho/cm), 230mL
HI 7033L 84 µS/cm (µmho/cm), 460 mL
HI 7033M 84 µS/cm (µmho/cm), 230 mL
HI 7034L 80000 µS/cm (µmho/cm), 460mL
HI 7034M 80000 µS/cm (µmho/cm), 230mL
HI 7035L 111800 µS/cm (µmho/cm), 460mL
HI 7035M 111800 µS/cm (µmho/cm), 230mL
HI 7039L 5000 µS/cm (µmho/cm), 460mL
HI 7039M 5000 µS/cm (µmho/cm), 230mL
HI 7032L 1382 ppm (mg/L), 460 mL
HI 7032M 1382 ppm (mg/L), , 230 mL
HI 7036L 12.41 ppt (g/L), 460 mL
HI 7036M 12.41 ppt (g/L), 230 mL
23
CONDUCTIVITY BUFFER SOLUTIONS IN FDA APPROVED
BOTTLES
HI 8030L 12880 µS/cm (µmho/cm), 460 mL
HI 8031L 1413 µS/cm (µmho/cm), 460 mL
HI 8033L 84 µS/cm (µmho/cm), 460 mL
HI 8034L 80000 µS/cm (µmho/cm), 460 mL
HI 8035L 111800 µS/cm (µmho/cm), 460 mL
HI 8039L 5000 µS/cm (µmho/cm), 460 mL
CONDUCTIVITY PROBES
HI 76301W 4-ring conductivity/TDS probe with 1 m (3.3') screened
cable (for HI 8033, HI 8633 and HI 8734)
19,5mm
DIA
0.77"
85mm
3.34"
11.5mm
DIA
0.45"
118.5mm
4.66"
19,5mm
DIA
0.77"
85mm
3.34"
11.5mm
DIA
0.45"
118.5mm
4.66"
OTHER ACCESSORIES
HI 710001 Soft carrying case for HI 8033, HI 8633, HI 8733,
HI 8734 and HI 933000 measuring 230x100x50 mm
CHECKTEMPC Electronic thermometer (range: -50.0 to 150.0°C)
HI 721313 Rugged carrying case for HI 8733 and HI 8633 measur-
ing 340 x 230 x 90 mm complete with conductivity
calibration solution (HI 7030S, 120 mL) and screwdriver
24
HI 721314 Rugged carrying case for HI 8734 measuring
340x230x90 mm complete with TDS calibration solu-
tion (HI 7032S, 120 mL) and screwdriver
HI 731326 20 small screwdrivers, length 90 mm, for calibration
purposes (except for HI 8033)
HI 710007 Blue shockproof rubber boot for HI 933000
HI 710008 Orange shockproof rubber boot for HI 933000
HI 710009 Blue rubber boot for HI 8033, HI 8633, HI 8733 and
HI 8734
HI 710010 Orange rubber boot for HI 8033, HI 8633, HI 8733
and HI 8734
MANCONDR2 Instruction manual
WARRANTY
All Hanna Instruments meters are warranted for two years against
defects in workmanship and materials when used for their intended
purpose and maintained according to instructions. The electrodes
and the probes are warranted for a period of six months. This
warranty is limited to repair or replacement free of charge.
Damages due to accident, misuse, tampering or lack of prescribed
maintenance are not covered.
If service is required, contact the dealer from whom you purchased
the instrument. If under warranty, report the model number, date of
purchase, serial number and the nature of the failure. If the repair is
not covered by the warranty, you will be notified of the charges
incurred. If the instrument is to be returned to Hanna Instruments,
first obtain a Returned Goods Authorization number from the Cus-
tomer Service department and then send it with shipping costs prepaid.
When shipping any instrument, make sure it is properly packaged for
complete protection.
To validate your warranty, fill out and return the enclosed warranty
card within 14 days from the date of purchase.
All rights are reserved. Reproduction in whole or in part is prohibited
without the written consent of the copyright owner,
Hanna Instruments Inc., 584 Park East Drive, Woonsocket, Rhode
Island, 02895 , USA.
Hanna Instruments reserves the right to modify the design, construc-
tion and appearance of its products without advance notice.
25
CE DECLARATION OF CONFORMITY
26
OTHER PRODUCTS FROM HANNA
27
HANNA LITERATURE
Hanna publishes a wide range of catalogs and hand-
books for an equally wide range of applications.
The reference literature currently covers areas such
as:
• Water Treatment
• Process
• Swimming Pools
• Agriculture
• Food
• Laboratory
• Thermometry
and many others. New reference material is con-
stantly being added to the library.
MANCONDR2
06/97
h t t p : / / w w w . h a n n a i n s t . c o m