01 Manometer

This document is a Certificate of Calibration issued to Bio Mark Pharmaceuticals for a Dwyer Manometer, calibrated on February 1, 2025, with the next calibration due on January 29, 2026. The calibration was performed under specific environmental conditions, and all observed values matched the standard with no deviations. The equipment is confirmed to be in good working condition.

Uploaded by

Noor Ulain
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as DOCX, PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
62 views1 page

01 Manometer

This document is a Certificate of Calibration issued to Bio Mark Pharmaceuticals for a Dwyer Manometer, calibrated on February 1, 2025, with the next calibration due on January 29, 2026. The calibration was performed under specific environmental conditions, and all observed values matched the standard with no deviations. The equipment is confirmed to be in good working condition.

Uploaded by

Noor Ulain
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as DOCX, PDF, TXT or read online on Scribd
You are on page 1/ 1

CERTIFICATE OF CALIBRATION

Issued By: Scientific Solution Certificate Number: BM/MM/SS/01


Calibration on: Feb 01, 2025 Calibration due on: 29 Jan, 2026

Customer Details:
Customer Name: Bio Mark Pharmaceuticals
Customer Address: Plot # 527, Sundar Industrial Estate, Lahore, Punjab.

Instrument Details:
Equipment Name: Manometer
Model No.: Dwyer
Serial No.: N/A
Manufacturer: USA
Location: Coating Tab II
Environmental Conditions:
Temperature: 24 °C ±2°C
Relative Humidity: 30 % ±10%
Calibration Standard: Pre-calibrated Manometer
Traceability Statement:
The calibration standard is Calibrated as per procedure traceable to NPSL, Pakistan maintained and controlled at
APC & I Centre, PCSIR Laboratories Complex Lahore. The NPSL maintains international traceability through
membership of the International Bureau of Weights and Measures (BIPM), France, and the Asia Pacific Metrology
Program (APMP).

CALIBRATION DATA:

Sr. NO. Standard OBSERVED DEVIATION


(inches of water column) (inches of water column) (inches of water column)
1 -1 -1 0.0
2 -2 -2 0.0
3 -3 -3 0.0
4 -4 -4 0.0
5 -5 -5 0.0
6 0.1 0.1 0.0
7 0.3 0.3 0.0
8 1.0 1.0 0.0
9 2.0 2.0 0.0
10 3.0 3.0 0.0
REMARKS: Based on the above tests, the Equipment has been calibrated satisfactorily and is in good working
condition.

Calibrated By: Checked By: _____________