Chapter Calibrations and Quality
3 Assurance
Run Calibration and QA Tests
Options for running scanner tests are available on the Utilities menu.
From SmartScan STUDIO scanning workflow, press to access the Utilities menu.
NOTE: Refer to the i-CAT FLX User Manual for information about using SmartScan STUDIO for
the scanning workflow.
Option Description
PanelCal Panel calibration should be performed once a week.
ShutterCal Shutter calibration should be performed once a week to ensure optimal
image quality.
ChairCal Chair calibration is to be performed by Service Technicians only.
GeoCal Geometric calibration should be performed once a year to ensure
optimal image quality. Run as needed if the image quality is degraded.
QA Line Pair QA Line Pair test checks the spatial resolution using the QA phantom.
Perform annually or when image quality is degraded.
QA Material test checks consistency of measurements in various
QA Material materials using the QA phantom. Perform annually or when image
quality is degraded.
QA Air Water test is a noise level and uniformity test. HU
QA Air Water measurements are taken at five different regions within the water
volume of the water phantom. Perform annually or when image quality
is degraded.
QA Pan QA Pan test is used to validate the PAN scan data capture using the
PAN phantom. Perform annually or when image quality is degraded.
Reprocess Exam Refer to i-CAT FLX User Manual for information.
Favorites Manager Refer to i-CAT FLX User Manual for information.
Roll-off Roll-off is to be used by Service Technicians and other qualified
personnel only.
032-0328-EN Rev K 3-1
Downloaded from www.Manualslib.com manuals search engine
i-CAT FLX Technical Guide
CAUTION
If the optional deadman handswitch is installed on the scanner, press and hold
handswitch before pressing the Scan button, and continue to hold handswitch down for
the duration of the exposure (X-ray light on). Early release of the handswitch stops the
exposure and the Fault light turns on. The patient will have to be re-scanned.
Panel Calibration
It is recommended that Panel Calibration be performed once a week.
PanelCal is performed in both landscape and portrait (V17 systems only) positions for 4 x 4 and
2 x 2 resolutions. Several tests run as part of the panel calibration. A pie chart displays status.
Run Panel Calibration
1. From Utilities menu, select PanelCal.
2. Ensure the field of view on the scanner is clear.
3. Press . The scanner initializes.
4. When prompted, press the Scan button on the operator control box. An audible alarm is
sounded and the X-ray ON light is illuminated during radiation exposure.
5. You will be prompted to press the Scan button for each test.
NOTE: The panel will rotate to the portrait position at the start of the portrait tests (V17
systems only).
6. When Panel Cal is complete, Calibration Complete is displayed.
7. Press to display Complete screen and select another option.
Panel Calibration Failure
If the calibration fails, the following message is displayed:
Panel Calibration Processing Failure
Press to exit.
Check that the field of view is clear of all objects and that there are no obstacles to the rotation
of the gantry. Re-run PanelCal. If failure persists, contact Technical Support.