Thermo Orion

The document provides instructions for using a conductivity calibration kit containing six resistors with known conductance values to verify the accuracy of a conductivity meter. The resistors can be connected to the meter one at a time and the reading compared to the actual conductance value listed on the resistor. Details are given on adjusting the meter settings and specifications for different resistor kits compatible with various meter models.

Uploaded by

hdabc123
Copyright
© Attribution Non-Commercial (BY-NC)
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
515 views2 pages

Thermo Orion

The document provides instructions for using a conductivity calibration kit containing six resistors with known conductance values to verify the accuracy of a conductivity meter. The resistors can be connected to the meter one at a time and the reading compared to the actual conductance value listed on the resistor. Details are given on adjusting the meter settings and specifications for different resistor kits compatible with various meter models.

Uploaded by

hdabc123
Copyright
© Attribution Non-Commercial (BY-NC)
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 2

Instruction Sheet

Thermo Scientific Orion Conductivity Calibration Kit


The conductivity calibration kit contains six resistors that simulate standard conductance values on a meter. The resistors may be used to verify proper meter function by comparing the meter reading to the conductance of the resistor. Note: The conductivity calibration kit verifies proper meter function only. After the meter function is verified, the conductivity cell must be calibrated. Calibrate the cell constant of the conductivity cell according to the meter and cell user guides.

Resistor Specifications

Each resistor is labeled alphabetically. The actual resistance and conductance values are printed on each resistor. The nominal resistance and conductance values are listed on the label of the calibration kit box for reference. Nominal Resistance Value 1000 K 0.1% 100 K 0.1% 10 K 0.1% 1 K 0.1% 100 0.1% 10 0.1% Nominal Conductance Value 1 S 10 S 100 S 1000 S 10 mS 100 mS

Resistor A B C D E F

Temperature Range 0 to 50 C 0 to 50 C 0 to 50 C 0 to 50 C 0 to 50 C 0 to 50 C

Verification of Meter Accuracy

Remove the conductivity cell and temperature probe from the meter. Perform a manual calibration on the meter and set the cell constant to 1.000 cm-1. Set the temperature value to 25.0 C and turn the temperature compensation off. Refer to the appropriate meter user guide for directions on cell constant and temperature adjustment. Attach a resistor to the meter by aligning the tabs on the resistor and meter. Push the resistor in until firmly in place. If the alignment is incorrect, the resistor and/or meter input can be damaged. When the meter indicates a stable conductivity value, compare the displayed value to the actual conductance of the resistor. The displayed conductivity value for each resistor should fall within the accuracy specification of the meter plus 0.5% of the actual resistor conductance. Since the resistor kits are made to test the accuracy of the meter, it is important that each displayed resistor value falls within the meter accuracy specification. If the displayed reading is not in the correct range, contact Technical Support. Within the United States call 1.800.225.1480 and outside the United States call 978.232.6000 or fax 978.232.6031. In Europe, the Middle East and Africa, contact your local authorized dealer.

2007 Thermo Fisher Scientific Inc. All rights reserved.

Ordering Information

Cat. No. 1010001 contains resistors with 8 pin MiniDIN connectors that are compatible with the 3-Star conductivity meter, 4-Star pH/conductivity meter, 5-Star pH/DO/ conductivity meter and 5-Star pH/ISE/DO/conductivity meter. Cat. No. 011220 contains resistors with 8 pin DIN connectors that are compatible with the 105A+, 115A+, 125A+, 145A+ and 150A+ meters. Cat. No. 011221 contains resistors with 8 pin waterproof DIN connectors that are compatible with the 130A, 131S, 135A, 136S, 162A, 550A and 555A meters.

Cat. No. 1010001 011220 011221 011008 01100910 011007 01100710 011006 01100610 011005 01100510 990106

Description Conductivity calibration resistor kit for Star Series conductivity meters, 8 pin MiniDIN connection Conductivity calibration resistor kit for 105A+, 115A+, 125A+, 145A+ and 150A+ meters, 8 pin DIN connection Conductivity calibration resistor kit for 130A, 131S, 135A, 136S, 162A, 550A and 555A meters, 8 pin waterproof DIN connection 100 S/cm conductivity/TDS standard, 5 x 60 mL bottles 147 S/cm conductivity standard, 10 pouches 1413 S/cm conductivity/TDS standard, 5 x 60 mL bottles 1413 S/cm conductivity/TDS standard, 10 pouches 12.9 mS/cm conductivity/TDS standard, 5 x 60 mL bottles 12.9 mS/cm conductivity/TDS standard, 10 pouches 111 mS/cm conductivity standard, 5 x 60 mL bottles 111 mS/cm conductivity standard, 10 pouches 0.1 M KCl conductivity standard, 475 mL bottle

Environmental Instruments
Water Analysis Instruments
166 Cummings Center Beverly, MA 01915 USA
254783-001 Rev.A 0807

Toll Free: 1-800-225-1480 Tel: 1-978-232-6000 Dom. Fax: 1-978-232-6015 Intl Fax: 978-232-6031 email: info.water@thermofisher.com www.thermo.com/water
2007 Thermo Fisher Scientific Inc. All rights reserved.