0 ratings0% found this document useful (0 votes) 272 views26 pagesHP 435B Operation PDF
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content,
claim it here.
Available Formats
Download as PDF or read online on Scribd
————— OPERATING AND SERVICE MANUAL
435B
POWER METER
(Including Options 001, 002, 003, and 004)
SERIAL NUMBERS
‘This manual applies directly to instruments with
serial numbers prefixed 2342A and 2342U.
With changes described in Section VI, this man:
| ual also applies to instruments with serial num-
bers prefixed 205A, 2041 and 238A.
For additional important information about serial
numbers, see INSTRUMENTS COVERED BY
MANUAL in Section 1.
iz HEWLETT
PACKARD
© Copyright HEWLETT-PACKARD COMPANY 1980, 1963
4501 PAGE MILL ROAD, PALO ALTO, CALIFORNIA, U.S.A.
OPERATING AND SERVICE MANUAL PART NO. 00435-0040,
Operating Manual Part No. 00435-00061
Operating and Service Microtiche Part No. 00435-90042 Printed: FEBRUARY 1984®
Model 4358
Safety Considerations
SAFETY CONSIDERATIONS
GENERAL
‘This product and related documentation must be
reviewed for familiarization with safety markings
and instructions before operation.
This product is a Safety Class I instrument (pro-
vided with a protective earth terminal).
BEFORE APPLYING POWER
Verify that the product is set to match the avail-
able line voltage and the correct fuse is installed.
SAFETY EARTH GROUND
An uninterruptible safety earth ground must be
provided from the main power source to the pro-
duct input wiring terminals, power cord, or sup-
plied power cord set.
Any interruption of the protective (grounding)
conducior (inside or outside the instrument) or
disconnecting the protective earth terminal will
cause a potential shock hazard that could resultin
personal injury. (Grounding one conductor of a
two conductor outlet is not sufficient protection).
In addition, verify that a common ground exists
between the unit under test and this instrument
prior to energizing either unit.
Whenever it is likely that the protection has been
impaired, the instrument must be made inoperative
and be secured against any unintended operation.
If this instrument is to be energized via an auto-
transformer (for voltage reduction) make sure the
common terminal is connected to neutral (that is,
the grounded side of the mains supply).
Servicing instructions are for use by service:
trained personnel only. To avoid dangerous elec:
tric shock, do not perform any servicing uniess
qualified to do so.
Adjustments described in the manual are per-
formed with power supplied to the instrument
while protective covers are removed. Energy avail-
able at many points may, if contacted, result in
personal injury.
Capacitors inside the instrument may still be
charged even if the instrument has been discon-
nected from its source of supply.
For continued protection against fire hazard, re-
place the line fuseis) only with 250V fuse(s) of the
same current rating and type (for example, normal
blow, time delay, ete.). Do not uge repaired fuses or
short circuited fuseholders.
SAFETY SYMBOLS
AX, Instruction manual symbol: the product
will be marked with this symbol when itis,
necessary for the user to refer to the in-
struction manual (see'Table of Contents for
page references),
; Indicates hazardous voltages.
Indicates earth (ground) terminal.
‘The WARNING sign denotes a
(WARMING) sitar Itc attention
procedure, practice, or the like,
which, ifnot correctly performed
oradhered to, could result in per-
sonal injury. Do not proceed be-
yonda WARNING sign until the
indicated conditions are fully
understood and met.
‘The CAUTION sign denotes a
hazard. It calls attention to an
operating procedure, practice, or
the like, which, if not correctly
performed or adhered to, could
result in damage to or destrue-
tion of part or all of the product.
Do not proceed heyond a CAU-
TION sign until the indicated
conditions are fully understood
and met.
CAUTIONGeneral Information
Model 4358
POWER SENSOR CABLE
MODEL 4358
re
ALIGNMENT TOOL
POWER CABLE
1-0
Figure 1-1, HP Model 4358 and Accessories SuppliedModel 4358
General Information
SECTION!
GENERAL INFORMATION
14. INTRODUCTION
‘This manual provides information pertaining to
the installation, operation, testing, adjustment
and maintenance of the HP Model 435B Power
Meter.
Figure 1-1 shows the Power Meter with accessories
supplied.
An operating manual is shipped with the instru-
ment. This is simply a copy of the first three sec-
tions of thismanual. Theoperating manual should
be kept with the instrument for use by the opera-
ior. Additional copies of the operating manual
may be ordered separately through your nearest
Hewlett-Packard office. The part number is listed
on the title page of this manual.
On thetitle page of this manual, below the manual
part number, is a “Microfiche” part number. This
number may be used to order 100 x 150 mm (4x6-
inch) microfilm transparencies of themanaul. The
microfiche package also includes the latest Man-
ual Changes supplement as well as all pertinent
Service Notes.
Instrument specifications are listed in Table 1-1.
‘These specifications are the performance stand-
ards or limits against which the instrument may
be tested,
1-2, INSTRUMENTS COVERED BY MANUAL
Options 001, 002, 003 and 004 of the Power Meter
are documented in this manual. The differences
are noted in the appropriate location such as
OPTIONS in Section I, the Replaceable Parts List,
and the schematic diagrams.
‘This instrument has a two-partserial number. The
first four digits and the letter comprise the serial
number prefix. The last five digits form the se
quential suffix that is unique to each instrument,
The contents of this manual apply directly to in-
struments having the same serial number pre-
fix(es) as listed under SERIAL NUMBERS on the
title page.
An instrument manufactured after the printing of
this manual may havea serial number prefix that
is not listed on the title page. This unlisted serial
prefix indicates that the instrument is different
from those documented in this manual. The man-
ual for this newer instrument is supplied with a
yellow Manual Changes supplement that con-
tains “change information” explaining how to
adapt the manual to the newer instrument.
In addition to change information, the supple-
ment may contain information for correcting er-
rors in the manual. To keep this manual as current
and accurate as possible, Hewlett-Packard recom-
mends that you periodically request the latest
Manual Changes supplement. The supplement for
this manual is keyed to the manual’s print date
and part number, both of which appear on the title
page. Complimentary copies of the supplement
are available from Hewlett-Packard.
For information concerning aerial number prefix
not listed on the title page or in the Manual
Changes supplement, contact your nearest
Hewlett-Packard office,
1-3, DESCRIPTION
‘The Power Meter and a compatible power sensor
are interconnected with the power sensor cable to
form a power measurement system. The system
power level range, frequency response, and load
impedance are dependent on the power sensor.
Accuracy of the power measurement system is,
ensured by the following Power Meter character-
istics:
a, Aninternal automatiezeroing circuit which
removes error dueto the ambient temperature out-
put of the power sensor’s power sensing device.
b, A calibration factor adjustment which ac:
counts for error due to the frequency response of
the power sensing device.
¢. Aninternal calibration reference which has
an output of 1 mW + 0.7% (502).
LaGeneral Information
Model 435B
Table 1-1, Specifications
Frequency Range:
100 kHz to 26.5 GHz (depending on power sensor
used),
reve fo
treba renee een
cain ana ame cab vt fl scale
Me sams 3,35 and dm
Cav ow
gta sr 4 8 with 9 cal
ae rg i002, 00 and Bm
een
1B 164 eer 068 with
in aemonee oat 10-0,
eta toaee cried att
Win ba decocen th ei ofa ene ranges of
gerbes yi niar ear ep
apm ccbew a 0s.
Aecuraey:
instrumentation 1% of full sale on all ranges
Zero: Automatic, operated by front-panel switch.
Zero Set 20.5% of full scale on most sensitive
ange, typical
Zero Carryover: 20.5% of fall scale when zeroed on
the most sensitive range.
Noise typical, at constant temperature, peak
| change over any oneminute interval); 20 pW
{B484A), 40 nW (481A, B482A, S48BA, 8485A),
{1 (B481HY, 84824, 40 uW 64818, 84828),
trv hour, typical) at constant temperatare
after 24-hour warcn-up); 40 pW (84844); 15 0
(BASLA, 84824, B489A, 84854); 1.5 AW (S481,
54824); 15 AW (6A81B, 84828)
Power ferent: Internal 50 MEIz oscillator with
‘Type N Female connector on front panel {or
rear panel, Option 068 only)
Power output: 1.00 mW.
Factory set to 0.7% traceable to tho National
Burenu of Standards
Accurney: #1.2% worst ease (20.9% rss) for
one year (0 t0 55°C)
SPECIFICATIONS
Aesponse Time:
(0 to 99% of reading, five time constants.)
Range 1 (most sensitive) <10.0 seconds.
Range 2 <3. seconds,
Range 3 1.9 seconds.
Ranges 4—10 -=500 milliseconds,
‘Typical, measured at recorder output)
al Factor:
16position switch normalizes meter reading to
account for calibration factor or effective eff
cieney.
Range 85% to 100% in 1% steps.
Cal Adjustment:
Front panel adjustment provides capability to
adjust gain of meter to match power sensor in
Recorder Output:
Proportional to indicated power with 1 volt cor |
responding to full scale; 1 kf output impedance:
BNC connector.
AF Blanking Output:
Provides a contact closure to ground when auto-
zero mode is engaged.
Power Consumption:
100, 120, 220, or 240V +5%, 10%.
100 and 120 volts, 48 to 66 Hz and 360—~440 Hz.
220 and 240 volts, 48 to 66 He.
20 V-A maximum.
Weight
Net, 2.7 kg (6.9 Ibs).
Dimensions: |
155 mm high (68/32 inches),
180 mm wide (5-1/8 inches).
279 mmn deep {11 inches).
Includes aneornontineadty. Adld-+2, 1% ontoptworanges when uring the MSA, 482A, 5480A and BASTA power senses:
rae eee ing ie ait BandWs2B power senses ad =.0%onthetop oranges when using theb4e1H ahd BZ power acnaor.
honeModel 435B
1-4, OPTIONS
4-5, Battery
‘The Model 435B, Option 001 Power Meter is sup-
plied with a rechargeable battery that provides up
to 16 hours continuous operation from a full
charge.
If the Power Meter was purchased without the
battery option, it may be ordered in kit form under
HP part number 00435-60012. The kit includes the
battery, the battery clamp, a 6-32 x 1/2-inch pan
head machine screw and installation instructions.
1-6. Input-Output Options
Option 002, A rear panel input connector is
connected in parallel with the front panel input
connector.
Option 003. A rear panel input connector is con-
nected in paralle! with the front panel input con-
nector. A rear panel POWER REF OUTPUT con-
nector replaces the standard front panel con-
nector.
Option 004, The 1.5 metre (5 ft.) power sensor cable
is not shipped with the Power Meter.
4-7, ACCESSORIES SUPPLIED
The accessories supplied with the Power Meter are
shown in Figure 1-1.
a, The 1.5 metre (5-foot) power sensor cable,
HP part number 11730A, is used to couple the
power sensor to the Power Meter. The 1.5 metre
cable is omitted when Option 004 is ordered.
b, ‘The line power cable may be supplied in sev-
eral configurations. Refer to the paragraph en-
titled Power Cables in Section II.
1-8. EQUIPMENT REQUIRED BUT NOT
SUPPLIED.
‘To form a complete RF power measurement sys-
tem, a power sensor, such as the HP Model 881A,
must be connected to the Power Meter via the
power sensor cable.
General Information
1-9, EQUIPMENT AVAILABLE
The HP Model 11683A Range Calibrator is recom-
mended for performance testing, adjusting and
troubleshooting the Power Meter. The Power
Meter’s range-to-range accuracy and auto-zero
operation can easily be verified with the calibra:
tor. It also has the capability of supplying a full
scale test signal for each range.
‘An extender board (HP part number 5060-0630)
may be used to place the Ad assembly printed
circuit board in a position that allows easy acces
to test points and components.
‘The following table lists the power sensor cable
aceessories and their lengths that are available
for use with the Power Meter. Order option 004 if
the standard 1.5 metre cable is not desired with a
cable accessory.
Power Sensor Cable Accessory | Cable Length
117308 3.1m (10
u7a0¢ 0)
117300 15.2m (50 f)
17306 30.5m (100 0)
11730F 1.0m (200 #0)
4-10, RECOMMENDED TEST EQUIPMENT
The test equipment shown in Table 1-2 is recom:
mended for use during performance testing, adjust
ments and troubleshooting, To ensure optimum
performance of the Power Meter, the specifica-
tions of a substitute instrument must equal or
exceed the critical specifications shown in the
table.
1-11, SAFETY CONSIDERATIONS
‘The Power Meter is a Safety Class I instrument
(provided with a protective earth terminal). This
instrument has been designed according to inter
national safety standards and has been supplied
in safe condition.
1aGeneral Information
Table 1-2. Recommended Test Equipment
Model 4353
Instrument Type
cr
1 Specifications
‘Suggested
Model a
Digital Voltmeter |
i
Function: DC, Resistance
Ranges:
Resistance: 200 2
‘Vale: 100 mV, 1000 mV, LOV, 100V
10M input impedance
51/2 digit resolution
‘Accuracy: #9.05% of reading
40.0280 of range
HP 24564 | PLAT
Vertical sensitivity: 0.2 V/division
Horizontal sensitivity: 1 ms/division
Frequency | Frequency Range: 200 Ha — 50 MElz HP 53144, A
Counter Sensitivity: 100 mVrms
| Accuracy: 0.01%
Oscilloscope Bandwidth: de to 50 MHz HP 1740,
| Power Meter
‘Range: capability to measure 1 mW
‘Transfer Aceuracy {input to output):
HP 43
Power Sensor |
Range: capability to measure 1 mW
HP 451A/Hor) P,A
HP 8482/1
Range Calibrator
HP 116838,
‘Thermistor
Mount
SWR: 1.05 at 50 MHz.
‘Accuracy:** 20.0% at 50 MHz
HP 4TeAHS | PLA
+P =Performance Tests, A= Adjustment; T= Troubleshooting
+ traceuboto NBS:Model 4358
Installation
SECTION II
INSTALLATION
2-1. INTRODUCTION
‘This section includes information on the initial
inspection, preparation for use, and storage and
shipment instructions for the Power Meter.
2-2. INITIAL INSPECTION
WARNING
To avoid hazardous electrical shock, do
not perform electrical tests when there
are signs of shipping damage to any por-
tion of the outer enclosure (covers and
panels).
Inspect the shipping container for damage. If the
shipping container or cushioning material is dam-
aged, it should be kept until the contents of the
shipment have been checked for completeness and
the instrument has been checked mechanically
and electrically. The contents of the shipment
should be as shown in Figure 1-1. Procedures for
checking electrical performance are given in See-
tion IV. If the contents are incomplete, if there is
mechanical damage or defect, or if the instrument
does not pass the electrical performance tests, no-
tify the nearest Hewlett-Packard office. If the
shipping container is damaged, or the cushioning
material shows signs of stress, notify the carrier
as well as the Hewlett-Packard office. Keep the
shipping materials for the earrier’s inspection.
2:3, PREPARATION FOR USE
2-4, Meter Zeroing
With the LINE switch set to OFF, the meter point-
or should be positioned directly over zero. Ifneces.
sary, insert a screwdriver into the mechanical
Meter Zero control (beneath the meter) and align
the pointer with zero. Back the adjustment off
slightly. The backlash in the control ensures
against a meter indication error caused by jarring
the instrument.
25. Range Switch Scale Selection
‘The RANGE switch has three seales on 2 remov-
able rings which correspond to the measurement
capabilities of compatible power sensors, The
range scales are 3W to 0.3 mW (+35 to -5 dBm),
100 mW to uW (+20t0-25 dBm) and 10 :Wto 0.3
(-20 to -65 dBm), Each seale listed indicates the
maximum and minimum full sealemeterreadings.
‘To select the correct RANGE switch knob assem-
bly scale (see Figure 2:
a. Unserew the outer (black) knob by turning it,
counterclockwise, Then, remove the outer knob.
b, Remove the two scale rings,
c. Determine which of the’ sealesis to be used
d. Place the other scale ring on the knob
assembly.
e. Place the selected ring on the knob assembly
with the selected scale out.
£ Line up the tabs of the seale rings with the
slot in the knob assembly.
g. Hold the scale rings in place with your
fingers. Thread the outer knob onto the knob
assembly. Lightly tighten the knob.
Scate BASE
Figure 2-1. Changing Range Switch ScaleInstallation
2-6. Power Requirements
‘The Power Meter requires a power source with an
output of 100, 120, 220, or 240 Vac +5%, ~10% sin-
gle phase, 100 and 120 volts, 48 to 66 Hz and 260
to 440 Hz, 220 and 240 volts, 48 to 66 Hz. Power
consumption is 20 V-A maximum.
If this instrument is to be energized via
an external autotransformer, makesure
the autotransformer common terminal
is connected to the earth terminal of the
power source.
2-7. Line Voltage Selection
BEFORE SWITCHING ON THIS IN-
STRUMENT, make sure the instrument
ig set to the voltage of the power source.
Figure 2-2 provides instructions for line voltage
and fuse selection. The line voltage selection card
and proper fuse are factory installed for 120 Vac
operation.
Fuses may be ordered under HP part numbers
2110-0234, 0.1 (20V slow blow) for 100/120 Vac
‘operation and 2110-0040 0.062A (250V slow bow)
for 220/240 Vae operation.
2-8. Power Cable
In accordance with international safety stan-
dards, this instrument is equipped with a three
wire power cable. When connected to an appropri-
ate ac power receptacle, this cable grounds the
instrument cabinet. The type of power cable plug
shipped with each instrument depends on the
country of destination. Refer to Figure 2-3 for the
part numbers of the power cable plugs available.
BEFORE SWITCHING ON THIS IN-
STRUMENT, the protective earth ter
minals of this instrument must be con
nected to the protective conductor of the
(Mains) power cord. The Mains plug
shall only be inserted in a socket outlet
provided with a protective earth con-
tact. The protective action must not be
negated by the use of an extension cord
(power cable) without a protective con-
ductor (grounding).
Model 435B
Operating voltage is shown
in module window,
SELECTION OF OPERATING VOLTAGE
4, Open caver door, pul the FUSE PULL lever and rotate to
Jat. Remove the tse
2, Remove te Line Voltage Selection Car, Position the card
50th ine voltage appears at tet cover, Push the card
firm into the sit
43, Rotate the Fuse Pull ever to ts normal position, knsrt a
use ofthe correct value Inthe holder. Close te cover
door.
WARNING
To avoid the possibility of hazardous electri
cal shock, do not operate this instrument at
line voltages greater than 126.5 Vae with line
frequencies greater than 66 Hz (leakage cur,
renis at these line settings may exceed
3.5 mA)
Figure 2-2. Line Voltage Selectien
2-9. Interconnections
‘The Power Meter and a power sensor are integral
parts of this measurement system. Before mea-
surements can be performed, the Power Meter and
sensor must be connected together with the power
sensor eable. (The cable is supplied with the Power
Moter.)
‘The power sensor cable couples the de supply and
sampling gate drive from the Power Meter to the
power sensor and the 220 Hz acoutputsignal from
the power sensor to the Power Meter.
The maximum voltage which may be
safely coupled to the Power Meter input
from the power sensor is 18 mVrms.
2-10. Operating Environment
‘The operating environment should be within the
following limitations:Model 4358
Installation,
aon mua |]
ofthat obtain
rave sev or ebatsor
ire riuoe ness wenscra
Taete*swrstaozie | cate-Toreton is
zareaoy
OPERATION
soor120v
OPERATION
PLUG: NEMA 5.15P
CABLE: 8120-1978
PLUG*: NEMA 6:15
CABLE: HP #120-0698
220/240
OPERATION
PLuG*: cEET-VII
CABLE*: HP 8120-1689
PLUGS: CEEZ2.V1
CABLE*: HP 8120-18
220240 A mola
orenaTion 22) orenaion
SJ)
SS) otuse: as 13634
60 GABLE: HP 8120-1351
"The number shown forthe cable isn HP part numer for complete eae inclu
Figure 2-3. Power Cable HP Part Numbers Versus Mains Plugs Available
Operating Environment (con's)
Temperature y Oto 55°C
Humidity 95% relative at 40°C
Altitude. . <4570 metres (15 000 feet)
2-11. Bench Operation
The instrument cabinet has plastic feet and a fold-
away tilt stand for convenience in bench opera-
tion. (The plastic feet are shaped to ensure self-
aligning of the instruments when stacked.) The
tilt stand raises the front of the instrument for
easier viewing of the control panel.
2-12, Rack Mounting
Instruments that are narrower than full rack
width may be rack-mounted using Hewlett
Packard adapter frames or combining cases.
Adaptor Frames. Hewlett-Packard accessory
adaptor frames are an economical means of rack
mounting instruments that are narrower than full
rack-width. A set of spacer clamps, supplied with
each adaptor frame, permits instruments of dif-
ferent dimensions to be combined and rack mount-
ed asa unit. Accessory blank panels are available
for filling unused spaces.
Combining Cases. Model 1051A and 1052A Com
bining Cases are metal enclosures that allow
combinations of one-third and one-half rackewidth
ingtruments to be assembled for use on a work-
bench or for mounting in a rack of standard 19-
inch spacing. Bach case includes a set of parti-
tions for positioning and retaining instruments
and a rack mounting kit. No tools are required for
installing the partitions. For bench use the cases
have the same convenient features as full rack-
width instruments, (i.e, fold-away tilt stands and
specially designed feet for easier instrument stack-
ing). Accessories available for the combining
cases include blank filler panels and snap.on full
width control panel covers.
2-43. Battery Operation LN
To operate the Power Meter on battery power, the
battery must be installed and charged, the line
power cable must he disconnected, and the LINE,
switch must be ON.
This task should be performed only by
service trained persons who are aware
of the potential shock hazard of working
on an instrument with protective covers
removed.
Battery Inst
To avoid hazardous electrical shock, the
line (Mains) power cable should be dis-
connected before attempting to install
the battery.Installation
Battery Operation (Cont'd)
WARNINGS] (Cont'd)
Do not short the battery terminals. This
may result in overheating which can
cause burns or increase risk of fire
Do not incinerate or mutilate the bat
tery, Tt might burst or release toxie
‘materials causing personal injury.
The battery is installed in the Power Meter as
follows (see Figure 2-4);
a, Remove the top cover.
b, Hold the battery above the Power Meter,
parallel to printed circuit board Ad, The battery
terminal lugs must face the circuit board.
c. Loosen the lugs. Move the battery down into
place and guide the lugs into the slots on the cir-
cuit board. The battery should now rest on the
aluminum deck.
Model 4853
d. Place the battery clamp over the battery and
secure it. The two prongs fit into slots on the rear
panel and the 6.32 x 1/2inch pan head machine
screw holds the forward end of the clamp in place.
e, Tighten the battery terminal lugs by hand.
Figure 2.
installed.
shows the Power Meter with battery
Battery Charging. ‘The battery is being charged if
the battery has been installed, theline power cable
is connected to the available line power, and the
LINE switch is ON. In thefully charged condition,
(@4hour charge time), the battery will supply
power for a minimum of 16 hours.
2-14, STORAGE AND SHIPMENT
2.15. Environment
‘The instrument should be stored in a clean, dry
environment. The following environmental lim-
itations apply to both storage and shipment:
‘Temperature . 55 to +75°C
Humidity : <95%% relative at 40°C
Altitude... +15 300 metres (50 000 feet)
Figure 2-5. Power Meter with Battery InstalledModel 4358
2-16. Packaging
Tagging for Service. If the instrument is being
returned to Hewlett-Packard for service, please
complete one of the blue repair tags located at the
end of this manual and attach it to the instrument,
Original Packaging. Containers and materials
identical to those used in factory packaging are
available through Hewlett-Packard offices. If the
instrument is being returned to Hewlett-Packard
for servicing, attach a tag indicating the type of
service required, return address, model number
and full serial number. Also mark the container
FRAGILE to ensure careful handling. In any cor-
respondence refer to the instrument by model
number and full serial number.
Other Packaging. The following general instruc-
tions chould be used for re-packaging with com-
mercially available materials:
Installation
a, Wrap the instrument in heavy paper or plas-
tic. (If shipping to a Hewlett-Packard office or ser-
vice center, attach a tag indicating the type of
service required, return address, model number
and full serial number.
>. Use a strong shipping container. A double-
wall carton made of 2.4 MPa (250 pound) test
material is adequate.
¢._Usea layer of shock-absorbing material 70
to 100 mm (3 to 4 inches) thick around all sides of
the instrument to provide firm cushioning and
prevent movement inside the container. Protect
the control panel with cardboard.
d. Seal the shipping container securely
, Mark the shipping container FRAGILE to
ensure careful handling.
f. In any correspondence, refer to the instru:
‘ment by model number and full serial number.®
Model 435B
Operation
SECTION Il
OPERATION
4, INTRODUCTION
‘This section provides complete operating instruc-
tions for the Power Meter. The instructions consist
of: panel features, operator's checks, operating
instructions, power measurement accuracy and
operator’s maintenance.
3-2. PANEL FEATURES
Front and rear panel features of the Power Meter
are described in Figures 3-1 and 3-2. These figures
contain a detailed description of the controls, indi-
cators and connectors.
3-3. OPERATOR'S CHECKS
NOTE
If the instrument does not operate prop-
erly and ig being returned to Hewlett
Packard for service, please complete one
of the blue repair tags located at the end
of this manual and attach it to the
instrument.
Upon receipt of the instrument, or to check the
Power Meter for an indication of normal operation,
follow the operational procedure shown in Figure
3-9, These procedures are designed to familiarize
the operator with the Power Meter and to provide
an understanding of the operating capabilities.
3-4. OPERATING INSTRUCTIONS
General operating instructions are contained in
Figure 3-4, The instructions will familiarize the
operator with the basic practices used when oper-
ating the Power Meter.
Any interruption of the protective
(grounding) conductor (inside or outside
the instrument) or disconnecting the
protective earth terminal is likely to
‘make this instrument dangerous. Inten
tional interruption is prohibited.
3-5. POWER MEASUREMENT ACCURACY
‘A power measurement is never free from error or
uncertainty, Any RF system has RF losses, mis-
match losses, mismatch uncertainty, instrumen-
tation uncertainty and calibration uncertainty,
‘Measurement errors as high as 50% are not only
possible, they are highly likely unless the error
sources are understood and, as much as possible,
eliminated.
3-6, Sources of Error and Measurement
Uncertainty
RF Losses. Some of the RF power that enters the
power sensor is not dissipated in the power sens-
ingelements. This RF lossis caused by dissipation
in the walls of waveguide power sensors, in the
center conductor of coaxial power sensors, in the
dielectric of capacitors, connections within the
sensor and radiation losses.
Mismatch. The result of mismatched impedances
between the device under test and the power sen-
sor is that some of the power fed to the sensor is,
reflected before it is dissipated in the load. Mis-
matches affect the measurement in two ways.
First, the initial reflection is a simple loss and is
called mismatch loss. Second, the power reflected
from the sensor mismatch travels back up the
transmission line until it reaches the source.
‘There, most of it is dissipated in the source impe-
dance, but some of it is re-reflected by the source
mismatch. The re-reflected power returns to the
power sensor and adds to, or subtracts from, the
incident power. For all practical purposes, the
effect the re-reflected power has upon the power
measurements unpredictable. This effectis called
mismatch uncertainty.
Instrumentation Uncertainty. Instrumentation un-
certainty describes the ability of the metering cir-
cuits to accurately measure the de output from the
power sensor's power sensing device. In the Power
Meter, this error is less than +1%.' It is important
to realize, however, that a 1% meter does not auto-
matically give 1% overall measurement accuracy.
Power Reference Uncertainty. The uncertainty of
the output level of the power reference oscillators
40.7%. This references normally used to calibrate
thesystem andis, therefore, apart ofthesystem's
total measurement uncertainty.
evo ftrarun 008
+ specifieston in Section when ange
B1Operation
Cal Factor Switch Resolution Error. The resolution
of the CAL FACTOR switch contributes a signifi:
cant ertor to the total measurement because the
switch has 1% steps. The maximum error possible
in each position is 40.5%.
3-7. Corrections for Error
Calibration Factor and Etfective Efficiency. The two
correction factors basic to power meters are cali-
bration factor and effective efficiency. Effective
efficiency is the correction factor for RF losses
within the power sensor. Calibration factor takes
into account the effective efficiency and mismatch
losses.
Calibration factor is expressed as a percentage
with 100% meaning the power sensor has no
losses. Normally the calibration factor will be
100% at 50 MHz, the operating frequency of the
internal reference oscillator,
‘The power sensors used with the Power Meter
have individually calibrated calibration factor
curves placed on their covers. To correct for RF
and mismatch losses, simply find the power sen-
sor’s calibration factor at the measurement fre-
quency from the curve or the table that is supplied.
with the power sensor, and set the CAL FACTOR,
switch to this value.
The CAL FACTOR switch resolution error of
+0.5% may be reduced by one of the following
methods:
1) Set the CAL FACTOR switch to the nearest
positions above and below the correction factor
given on the table. Interpolating between the
power levels measured provides the corrected
power level.
2) Leave the CAL FACTOR switch on 100%
after calibration. Then, make the measurement
and record the reading. Use the reflection coeffi-
cient, magnitude and phase angle, if such a table
is supplied with the power sensor, to calculate the
corrected power level.
3-8, Calculating Worst Case Uncertainty
Worst case uncertainty is the sum of the specified
uncertainties and mismatch uncertainty. Uncer-
tainty calculatigh is outlined in the following two
subsections and examples are worked out in Fig-
ures 3-5 and 36. For a more complete explanation
of measurement uncertainty refer to HP applica:
tion note AN-64-1 “Fundamentals of RF and Mi-
crowave Power Measurement”
a2
Model 435B
Specified Uncertainties. The specified uncertain-
ties which account for part of the total power mea:
surement uncertainty are:
a. Instrumentation 1%! or +0.05 dB.
b. Power reference +0.7% or £0.03 dB.
c. CAL FACTOR switch resolution, 0 to
(depending on Cal Factor).
a, Zoro set, 0.5% of full scale of lowest range
which is 15 nW.
e. Zero Carryover, #0.5%.
£ Noise and Drift, depends on the range and
type of sensor.
g. Calibration factor uncertainty, which
depends on sensor type, is listed in the sensor
manual.
Figure 3.5 gives an example of specified uncer-
tainty calculation.
Calculating Mismatch Uncertainty. Mismatch un-
certainty is the result of the source mismatch
interacting with the power sensor mismatch. The
magnitude of uncertainty is related to the magni-
tudes of the source and power sensor reflection
coefficients, which can be calculated from SWR.
Figure 3-6 shows how the calculations are made
and Figure 37 illustrates mismatch uncertainty
and total calculated uncertainty for two cases. In
the first case, the power sensor's SWR= 1.5, andin
the second case, the power sensor's SWR = 1.25.In
both cases the source has an SWR of 2.0. The
example shows the effect on power measurement
accuracy a poorly matched power sensor will have
as compared to one with low mismatch.
‘A faster, easier way to find mismatch uncertainty
is to use the HP Mismatch Error (uncertainty)
Limits/Reflectometer Calculator. The calculator
may be obtained, on request, from your nearest
Hewlett-Packard office by using HP part number
59520948,
‘The method of calculating measurement uncer-
tainty from the uncertainty in dB is shown by
‘gure 3-8, This method would be used when the
initial uncertainty calculations were made with
the Mismatch Error/Reflectometer Calculator.
‘Refer io Ingtrament accuracy specification in Section] when using theModel 4358
3-9. OPERATOR'S MAINTENANCE
‘The only maintenance responsibilities the opera-
tor should normally perform are primary power
fuse replacement, LINE switch lamp replacement
and rechargeable battery replacement.
Battery replacement is the only operation that
requires tools. A Pozidriv screwdriver is needed to
remove the battery clamp.
3-10. Fuses
‘The primary power fuse is found within the A6
Power Module Assembly on the Power Meter's
rear panel. For instructions on how to change the
fuse, refer to the paragraph entitled Line Voltage
Selection in Seetion II.
Make sure that only fuses with the
required rated current and of the speci
fied type (slow blow, timedelay,ete.)are
used for replacement. The use of re
paired fuses and the short-cireuiting of
fuse-holders must be avoided.
3-11. Lamp Replacement
‘The lamp is contained in a plastic lens which dou-
bles for a pushbutton on the LINE switch. When
Operation
the Power Meter LINE switch is ON and is being
operated by the available line power, the lamp
should be illuminated. If the lamp is defective,
remove the lens by pulling it straight out, Order
lamp (3131-0434) CD6 and replace the old
pushbutton-lamp assembly with the new one. To
replace the assembly, align the pins with the
notch in the receptacle and push straight in.
3-12. Battery Replacement
If the meter indicates that the battery is dis-
charged by a full downscale reading, and after
charging the battery still will only power the
Power Meter for a short period of time, the battery
is probably defective, The replacement battery,
BT1 (HP part number 1420-0096), may be ordered
through the nearest Hewlett-Packard office. Refer
to Battery Installation in Section II.
This task should be performed only by
service trained persons who are aware
ofthe potential shock hazard of working
on an instrument with protective covers
removed.Operation
Model 4358
@ Meter. Normaily indicates average RF power in
aBmor Watts. During battery operation the meter
continuously indicates battery eondition. A nor
malreading indicates the battery is charged: full
down-scale reading indicates the battery is dis
charged or is defective
© Meter Zer0. Mechanical adjustment used tozero the
meter when the LINE switch is OFF.
| @ LINE switen. Connecis line or battery power to the
Powor Meter circuits when the LINE switch is ON.
During battery operation, the lamp contained
| within the LINE awitch will not be illuminated
‘when the INSTRUMENT is ON.
© AANGE switch, Selects desired power range; keyed
to meter full-scale deflection; has three removable
sealee which are changed to match the range of
the power sensor.
FRONT PANEL FEATURES
(@ POWER REF OUTPUT. RF output of 1.00 mW 40.70%
into 600 at 50 MElz from on internal reference
oeeillator, Available for system calibration.
QCAL ADI. Screwdriver adjustment for calibrating
any power sensor and Power Meter as asystem, to
a known standard, |
© iu canner. ropa fom the power sensor via
the power sensor eable
CALFACTOR Switch. Changes the gain ofthe Power
Meier amplifier cireuits to compensate for mis
rnatch losses and effective effcieney of the power
sensor.
ZERO Switch. Te ZERO switeh activates a feed
back circuit, which automatically zeros the meter
pointer, and a rear panel RF blanking signal
Figure 3-1. Front Panel Controls, Connectors and Indicators
a4¢
Model 436B
Operation
tf
REAR PANEL FEATURES
@ POWER SENSOR INPUT. Option 002.and 003 havea
ear pane] input connector wired in parallel with
the front panel input connector.
© Power Module Assembly.
@ Window. saiety interiock; fuse cannot be removed
while power cable is connected to Power Meter.
FUSE PULL Handle, Mechanical interlock to guar.
antee fuse has been removed before Line Voltage
| Selection Card ean be removed,
@ Fuse. Rover to Section I for values.
| © Line Voltage Selection Card. Matches transformer
| primary 1 available ine voltage.
@ Receptacle. For power cable connection to avail-
able line voltage.
© POWER AEF OUTPUT. Takes the place of the front
panel POWER REF OUTPUT connector (Option
003 only).
© FECORDER OUTPUT. Provides a linear output with,
respect to the input power. +1.00 Vée corresponds
to meter full-scale, The minimum load which may
be coupled to the output is 1 Ma.
@ AF BLANKING OUTPUT. Contact closure to ground
when ZERO switch is pressed, May be used to
remove RF input signal during automatic zeroing
operation.
@ POWER AEF Switch. Opens or closes the circuit
from the power supply to the power reference oscil-
lator, Reduces current drain during battery opera
tion when OFF.
Figure 3-2. Rear Panel Contras, Connectors. and IndicatorsOperation Model 4353
OPERATOR'S CHECKS
1. BEFORE SWITCHING ON THIS INSTRUMENT, check that the power transformer primary is
matched to the available line voltage, the correct fuse is installed and the safety precautions are
taken, See Power Requirements, Line Voltage Selection, Power Cables and associated warnings
and cautions in section Il.
BEFORE CONNECTING LINE POWER TO THIS
INSTRUMENT, ensure that all devices connected to
this instrument are connected to the protective (earth)
ground,
BEFORE SWITCHING ON THIS INSTRUMENT, en-
ure that the line power (Mains) plug is connected to @
three-conductor line power outlet that has a protective
(earth) ground. (Grounding one conductor of a two
conduetor outlet is not sufficient.)
Do not twist the body of the power sensor when connect
ingordisconnecting it. This cancause majordamageto
the power sensor.
2, Set the meter indication to zero with the mechanical meter zero control. Back the control off
slightly,
3, Connect the power sensor to the Power Meter with the power sensor cable,
4. Connectthe power cable to the power outlet and power module receptacles. Setthe LINE switeh to
ON; the lamp within the switch lens should be illuminated.
5. Change the Power Meter’s RANGE switch scale so it corresponds to the range of the power
sensor. Refer to the paragraph entitled Range Switch Scale Selection in Section {1
6 Set the Power Meter Controls ax follows
RANGE switeh position fully cow
CAL FACTOR switch ......2-.+++-100%
POWER REF switch ORF
Press the ZERO switch and verify that the meter pointer moves to zero (0) and the RF BLANK:
ING OUTPUT is shorted to ground.
5, Set the RANGE switch to the position indicated in the following table. Then, connect the power
sensor (and adapter or attenuator as required) to the POWER REF OUTPUT and set the rear
pane] POWER REF ewitch to (ON). Verify that the meter reads approximately the same as
indicated in the table.
Figure 3-3. Operator's Checks (1 of 2)
36Model 435B
Operation
OPERATOR'S CHECKS
RANGE Switeh| Meter
Pee ae Piten | indcation
4481B and 84928 (remove attenuator) aw aw
| S161, stb2a, st, 2H 3mW | imi
8485A (HP 1250-1250 Adapter required) amw | imaW
8483A (HP 1250-0597 Mechanical Adapter required) amw | osemw
84844 (HP 11708A Reference Attenuator rrequired) 3 uw 1 aW
9. Step the CAL FACTOR switch through its range noting a small inerease in meter reading with
each successive step. Resot the CAL FACTOR switch to 1008,
10. Set the RANGE switch to the position indicated in the table below. Then, adjust the CAL ADJ
control for a full-scale moter reading for 500 power sensors and a 96% of full scale meter reading
for 750 power sensors.
Power Sensor ANGE Switch Patan
‘4818 and 64828 remove attenuator) Ww
| etsia, saa, eet, seaatt Timi
84854 (HD 1250-1250 Adapter required Taw
ESA (HP 12500507 Mechanical Adapter eguired) Tm
BAA (HP 11708A Reference Attenuator required) Law
11. Cheek at the rear panel RECORDER OUTPUT jack for an output of ~ 1 Vde.
12, To check operation using battery power, disconnect the power cable from the rear panel power
module receptacle and set the LINE switch to ON (the lamp within the switch lens will not be
illuminated). When a power measurement is made, a normal upscale reading indicates normal
operation; a full down-scale reading indicates the battery is discharged.
Figure 3-3. Operator's Checks (2 of 2)Operation
‘Model 4358
OPERATING INSTRUCTIONS
BEFORE SWITCHING ON THIS INSTRUMENT, check that the power transformer primary is.
matched to the available line voltage, the correct fuse is installed and safety precautions are
taken. See Power Requirement, Line Voltage Selection, Power Cables and associated warnings
and cautions in Section TL
BEFORE CONNECTING LINE POWER 10 THE
INSTRUMENT, ensure that all devices connected to
this instrument are connected to the provective (earth)
ground.
BEFORE SWITCHING ON THIS INSTRUMENT, en-
sure that the line power (Mains) plug is connected to a
three.conductor line power outlet that has a protective
(earth) ground. (Grounding one conductor of « two-
conductor outlet is noc sufficient.)
Donot twist the body of the power sensor when connect-
ing or disconnecting it. This can cause major damageto
the sensor.
Set the meter indication to zero with the mechanical meter zero control. Back the control off
slightly,
Connect the power sensor to the Power Meter with the power sensor cable.
Connect the power eable to the power outlet and power module receptacles, Set theLINE switch to
ON; the lamp within the switeh lens should be lit,
Change the Power Meter’s RANGE switch scale so it corresponds to the range of the power
sensor. Refer to the paragraph entitled Range Switch Scale Selection in Section IL.
Set the Power Meter switches as follows:
RANGE position. .....6s...10.2+. fully cow
CAL FACTOR. ieee 100%
POWER REF... OFF
Press the ZERO switch, allow 5 seconds for the zeroing operation to take place, and release the
switch,
Set the RANGE switch to the position indicated in the following table. Then, connect the power
sensor (and adapter or attenuator as required) to the POWER REF OUTPUT and set the rear
panel POWER REF switch to (ON). For 502 power sensors, adjust the CAL ADJ control for a
full-scale reading; the meter pointer should be aligmed with the CAL mark (full-scalereading) on
the meter face. For 750 power sensors, adjust the CAL ADJ control for 2 96% of full scale reading;
the meter pointer should be aligned with the 0.96 mark on the meter face.
38
Figure 3-4. Operating Instructions (1 of 2)Model 435B Operation
or OPERATING INSTRUCTIONS
Power Soar fANGE Swish Palon
4S1B and 84828 romove attenuator) we |
SABLA, BASRA, SAH, S480 TW
185A (HP 1260-1250 Adaptor required) mW |
483A (HP 1250-0597 Mechanical Adapter required)
S44A (HP 11708A Reference Attenuator required) | 1 AW
| im |
9 Disconnect the power sensor from the POWER REF OUTPUT and set the POWER REF switeh to
OFF.
10. Locate the calibration curve on the power sensor cover, Find the CAL FACTOR for the measure.
ment frequency; set the CAL FACTOR switch accordingly.
11, Set the RANGE switch such that full seale is greater than the power level to be measured.
See Operating Precautions in the power sensor Operat
ing and Service Manuals for maximum power levels
which may be safely coupled to this systera. Levels
which exceed the limits may damage the power sensor,
Power Meter, or both.
12, Connect the power sensor to the RF source. Read the power lavel in dBm or Watts on the panel
NOTE
When the battery is being used as the power supply for
the Power Meter, an automatic test circuit continually
monitors battery condition. When the battery voltage is
‘above a predetermined level, the meter indicates the
correct power level. When the voltage drops below the
threshold level, the meter reading is full downscale.
)
Figure 3-4, Operating Instructions (2 of 2)
39Operation Model 435E
SPECIFIED UNCERTAINTY CALCULATION
Conditions: Range —1 mW
Meter Reading ~ 0.7 mW
Sensor — 841A,
Frequeney — 1 GHz
CAL FACTOR — 99.5%
Instrumentation Uncertainty
Power Reference Uncertainty
CAL FACTOR Switch Resolution Uncertainty
Zero Set Uncertainty
Zero Carryover Uncertainty
Noise
Drift
Cal Factor Uncertainty
£0.06 4B
+003 4B
£0.02 dB
++0.00009 4B
£0.03 4B
0.00025 4B.
70.0008 dB
20.12.48
40,000015 mW
40,019 mW
0825 mW
Total Specified Uncertainties ~ 10.015 mW = 2985 9 sa.
O55 say
=10 225 = 205 8
NOTE: FS =% of full seale
R = %ofreading
ire 3-5. Specified Uncertainties
310Model 4358 Operation,
CALCULATING MEASUREMENT UNCERTAINTY,
1. Calculate the reflection coefficient from the given SWR.
PU, = {1+ [(0.2(0.833)] }?
= {us 0.067}?
= {uoer}* and {0.933}
= 1.188 and 0.871
= 0111
Pu fh
(ones)]*
sWR-1
6° SWRI
Power Sensor #1 Power Sensor #2, Power Source
SWR=15 SWR= 1.25 SWR = 2.0
is 125-1
&* T5ei 1.2541
= 0.388
2. Calculate the relative power and percentage power mismatch uncertainties from the reflection
coefficients. An initial reference level of 1 is assumed.
Relative Power Uncertainty
{2 = 0.2111)(0.299)) }*
{1+ 0.087}?
{1.037}? and {0.963 }*
1.075 and 0.927
Percentage Power Uncertainty
%PU = (PU-1) 100%
RPU, = (1.1381) 100% and (0.8711) 100%
= (0.188) 100% and (0.129) 100%
= 13.8% and 12.9%
%PU; = (1.0751) 100% and (0.9271) 100%
= (0.075) 100% and (0.073) 100%
= 15% and -1.8%
Figure 3-6. Caleulating Measurement Uncertainties (1 of 2)Operation
Model 4358
3.
Calculate the Measurement Uncertainty in dB
=) | #
MU,
Mu,
CALCULATING MEASUREMENT UNCERTAINTY
ws «8 [ona
10 [tog )]
10 (0.056)
+0.56 dB
10 [ioe (=2)]
10 [0.031]
40.31 4B
and
and
and
and
and
and
ost
10 [10s (*2)]
10 [0.060]
0.60 dB
1
1 [oe (22)
10 [-0.083]
0.83 dB
342
Figure 3-6. Calculating Measurement Uncertainties (2 of 2)Model 4358 Operation
INDICATED POWER VERSUS RANGE OF ACTUAL POWER
(Values from examples on Figures §-5 and 3-6.)
INDICATED POWER “TOTAL SPECIFIED WORST CASE UNCERTAINTY
UNCERTAINTIES (SPECIFIED & MISMATCH)
' t 0.84 mW (0.73 48h
t EXAMPLE 1
i SENSOR SWR = 1.5
' SOURCE SHR = 2.0 ACTUAL POWER
i |eaa% [+199% ANYWHERE IN
{ SHADED AREA
0.74 mt (-1,28 dBm)
10.65 ad (1.81 dBm)
O67 mW (2.4 dBm)
EXAMPLE?
AUPE ag 6.8 iv (0.98 8m
Savage SHR=20 ACTUAL POMER
suse erase ANvuniene Iv
0,74 mW |-1.29 dBm) cotati
7
/ Vein! J /
047 i (1.58 Bm) LLL LL
0.68 ni (1.81 dBm)
' 0.6 mW (=2.1 dBm)
Figure 3-7. Worst Case Effects of Specified and Mismatch Uncertainties5B
Operation
Model 43
CALCULATING MEASUREMENT UNCERTAINTY,
1. For this example the known values are: source SWR, 2.2 and power sensor SWR, 1.1
From
the Mismatch Error Calculator the mismatch uncertainty is found to be +0.24,—0.25 dB,
certainty is +0.50,~0.51 dB.
Add the specified uncertainties from Figure 3-5, («
26 dB). Our total measurement un-
3. Caleulate the relative measurement uncertainty from the following formula:
dB
aB
10
ey
10 tox( $4)
Po,
Py
(2)
4. Calculate the percontage Measurement Uncertainty.
aU (Py —Po) 100
= (1.1221) 100
= 412.2%
051
ven (6 )
0.389
(0.889 ~ 1) 100
11%
Figure 3-8. Calculating Measurement Uncertainty (Uncertainty in dB Known)
S14