CONTENTS
1. SAFETY INFORMATION…………………………….1
                                                  2. DESCRIPTION……………………..…………………3
                                                  3. OPERATING INSTRUCTION………………………..6
                                                      3.1 MEASURING VOLTAGE……………………….6
                                                      3.2 MEASURING CURRENT………………………6
                                                      3.3 MEASURING RESISTANCE…………………..7
                                                      3.4 MEASURING FREQUENCY……………………7
                                                      3.5 MEASURING CAPACITANCE…………………8
                                                      3.6 CONTINUITY TEST…..………………………...8
                                                      3.7 DOIDE TEST……………………………………..8
                                                      3.9 TRANSISTOR TEST…………………………….9
                                                  4. SPECIFICATIONS…………………………………….9
                                                  5. ACCESSORIES………………………………………13
                                                  6. BATTERY & FUSE REPLACEMENT………………14
Downloaded from www.Manualslib.com manuals search engine
                                                  1. SAFETY INFORMATION
                                                  This multimeter has been designed according to
                                                  IEC-1010 concerning electronic measuring instruments
                                                  with an overvoltage category ( CATⅡ) and pollution 2.
                                                  Follow all safety and operating instructions to ensure
                                                  that the meter is used safely and is kept in good
                                                  operating condition.
                                                  1.1 PRELIMINARY
                                                  ※ When using this meter, the user must observe all
                                                  normal safety rules concerning:
                                                  — Protection against the dangers of electrical current.
                                                  — Protection of the meter against misuse.
                                                  ※ Full compliance with safety standards can be
                                                  guaranteed only if used with test leads supplied. If
                                                  necessary, they must be replaced with the same model or
                                                  same electronic ratings. Measuring      leads must be in
                                                  good condition.
                                                  1.2 DURING USE
                                                  ※ Never exceed the protection limit values indicated in
                                                  specifications for each range of measurement.
                                                  ※ When the meter is linked to measurement circuit, do
                                                  not touch unused terminals.
                                                  ※ When the value scale to be measured is unknown
                                                  beforehand, set the range selector at the highest
                                                  position.
                                                  ※ Before rotating the range selector to change
                                                  functions, disconnect test leads from the circuit under
                                                    -1-
Downloaded from www.Manualslib.com manuals search engine
                                                  test.
                                                  ※ When carrying out measurements on TV or
                                                  switching power circuits, always remember that there
                                                  may be high amplitude voltages pulses at test points
                                                  which can damage the meter.
                                                  ※ Never perform resistance measurements on live
                                                  circuits.
                                                  ※ Always be careful when working with voltages
                                                  above 60V dc or 30V ac rms. Keep fingers behind the
                                                  probe barriers while measuring.
                                                  1.3 SYMBOLS
                                                           Important safety information, refer to the operating
                                                           manual.
                                                           Dangerous voltage may be present.
                                                           Earth ground
                                                           Double insulation (Protection class Ⅱ)
                                                  1.4 MAINTENANCE
                                                  ※ Before opening the meter, always disconnect test
                                                  leads from all sources of electric current.
                                                  ※ For continue protection against fire, replace fuse
                                                  only with the specified voltage and current ratings: F1:
                                                  2A/250V F2: 10A/250V
                                                  ※ If any faults or abnormalities are observed, the
                                                  meter can not be used any more and it has to be
                                                                                                       -2-
Downloaded from www.Manualslib.com manuals search engine
                                                  checked out.
                                                  ※ Never use the meter unless the back cover is in
                                                  place and fastened fully.
                                                  ※ To clean the meter, use a damp cloth and mild
                                                  detergent only, do not use abrasives or solvents on it.
                                                  2. DESCRIPTION
                                                  This meter is an autoranging professional instrument
                                                  with 3 3/4 digit LCD display, capable of performing
                                                  functions:
                                                  - DC voltage measurement
                                                  - AC voltage measurement
                                                  - DC current measurement
                                                  - AC current measurement
                                                  - Resistance measurement
                                                  - Capacitance measurement
                                                  - Frequency measurement
                                                  - Diode test
                                                  - Transistor test
                                                  - Audible continuity test
                                                  Names of components
                                                  1-4. buttons
                                                  5. Rotary switch
                                                  6. A input jack
                                                  7. mA input jack
                                                  8. V/Ω/Hz input jack
                                                  9. COM input jack
                                                  10. Transistor input jack
                                                    -3-
Downloaded from www.Manualslib.com manuals search engine
                                                  2.1 SELECT (Function selection button)
                                                  This button is used to select ACA or DCA measuring
                                                  mode.
                                                  2.2 RANGE (Auto/manual range selection button)
                                                                                             -4-
Downloaded from www.Manualslib.com manuals search engine
                                                  This key is the auto/manual key. Power on default is
                                                  auto measurement, and press one time to switch to the
                                                  manual measurement. In manual measurement mode,
                                                  the system will jump up one mode for each press until
                                                  the highest mode, then return to the lowest mode as a
                                                  loop. If press the key over 2 seconds, the system will
                                                  switch back to auto measurement status. Frequency
                                                  and capacitance measurement cannot use manual
                                                  measurement. Fox example, at voltage range the
                                                  changes are as the figure.
                                                  2.3 REL (Relative value measurement)
                                                  The button is for relative value measurement except Hz,
                                                  diode and continuity.
                                                  Push the button to get relative measurement mode,
                                                  “REL” annunciate display on LCD. But store the
                                                  displayed reading as a reference value. In the Relative
                                                  mode, the value shown on the LCD is always the
                                                  difference between the stored reference value and the
                                                  present reading. If the new reading is the same as the
                                                  reference value, the display will be zero.
                                                  2.4 DATA HOLD
                                                  When this button is pushed, the display will keep the
                                                  last reading and “ ” symbol will appear on the LCD.
                                                  Pushing it again returns the meter to normal mode.
                                                    -5-
Downloaded from www.Manualslib.com manuals search engine
                                                  2.5 INPUT JACKS
                                                  The unit has four input jacks that are protect against
                                                  overload to the limits. During use, connect the black test
                                                  lead to the COM jack and the red test lead as shown
                                                  below:
                                                                 Red lead
                                                   Function                             Input limits
                                                                connection
                                                      V           V/Ω/Hz         1000V dc or 750V rms ac
                                                      Ω           V/Ω/Hz             250V dc or rms ac
                                                                  V/Ω/Hz             250V dc or rms ac
                                                                  V/Ω/Hz             250V dc or rms ac
                                                    μA/mA           mA              500mA dc or rms ac
                                                      A              A                10A dc or rms ac
                                                  μA/mA and A ranges are protected by fuses.
                                                  3. OPERATING INSTRUCTION
                                                  3.1 MEASURING VOLTAGE
                                                  1. Connect the black test lead to the COM jack and the
                                                  red test lead to the V/Ω jack.
                                                  2. Set the rotary switch at the desired V       or V ~
                                                  range position and connect test leads across the
                                                  source or load under measurement.
                                                  3. Read LCD display. The polarity of the red connection
                                                  will be indicated when making a dc voltage
                                                  measurement.
                                                  3.2 MEASURING CURRENT
                                                  1. Connect the black test lead to the COM jack and the
                                                  read test lead to the mA jack for a maximum of 500mA.
                                                                                                    -6-
Downloaded from www.Manualslib.com manuals search engine
                                                  For a maximum of 10A, move the red lead to the A jack.
                                                  2. Set the function switch at μA, mA or A range to be
                                                  used and push “SELECT” button to select DCA or ACA
                                                  measuring mode.
                                                  3. Connect test leads in series with the load under
                                                  measurement.
                                                  4. Read LCD display. The polarity of the red lead
                                                  connection will be indicated when making a DC
                                                  measurement.
                                                  3.3 MEASURING RESISTANCE
                                                  1. Connect the black test lead to the COM jack and the
                                                  red test lead to the V/Ω/Hz jack. (NOTE: The polarity of
                                                  red lead connection is positive “+”)
                                                  2. Set the rotary switch at Ω position and connect test
                                                  leads across the resistance under measurement. Read
                                                  LCD display.
                                                  NOTE:
                                                  1. For resistance above 1MΩ, the meter may take a few
                                                  seconds to stabilize reading.
                                                  2. When the input is not connected, i.e. at open circuit,
                                                  the figure “OL” will be displayed for the overrange
                                                  condition.
                                                  3. When checking in - circuit resistance, be sure the
                                                  circuit under test has all power removed and all
                                                  capacitors are full discharged.
                                                  3.4 MEASURING FREQUENCY
                                                  1. Connect the black test leads to the COM jack and the
                                                    -7-
Downloaded from www.Manualslib.com manuals search engine
                                                  red test lead to the V /Ω/Hz jack.
                                                  2. Set the rotary switch at Hz position and connect test
                                                  leads across the source or load under measurement.
                                                  3.5 MEASURING CAPACITANCE
                                                  1. Set the rotary switch at desired    position.
                                                  2. Before inserting capacitor under measurement into
                                                  capacitance testing socket, be sure that the capacitor
                                                  has been discharged fully.
                                                  3. When measuring capacitors with shorter leads, a
                                                  testing adapter is provided with the meter. Insert the
                                                  adapter into the capacitance testing socket on the front
                                                  panel to continue measurements.
                                                  WARNING!
                                                  To avoid shock, be sure the capacitor measuring
                                                  adapter has been removed before changing to another
                                                  function measurement.
                                                  3.6 CONTINUITY TEST
                                                  1. Connect the black test lead to the COM jack and the
                                                  red test lead to the V/Ω/Hz jack (Note: the polarity of
                                                  red lead connection is positive “+”).
                                                  2. Set the rotary switch at       position and connect test
                                                  leads across two points of the circuit under testing. If
                                                  continuity exists ( i.e. resistance less than about 50Ω ),
                                                  built — in buzzer will sound.
                                                  3.7 DIODE TEST
                                                  1. Connect the black test lead to the COM jack and the
                                                                                                     -8-
Downloaded from www.Manualslib.com manuals search engine
                                                  red test lead to the V/Ω/Hz jack. ( NOTE: The polarity of
                                                  red lead connection is positive“ + ”)
                                                  2. Set the rotary switch at     position and connect the
                                                  red lead to the anode, the black lead to the cathode of
                                                  the diode under testing. The meter will show the approx.
                                                  Forward voltage drop of the diode.
                                                  3.8 TRANSISTOR TEST
                                                  1. Set the rotary switch at hFE position.
                                                  2. Identify whether the transistor is NPN or PNP type
                                                  and locate Emitter, Base and Collector lead. Insert
                                                  leads of the transistor to be tested into proper holes of
                                                  the testing socket on the front panel.
                                                  3. LCD display will show the approximate hFE value at
                                                  the test condition of base current 10µA and Vce 3.2V.
                                                  4. SPECIFICATIONS
                                                  Accuracy is specified for a period of one year after
                                                  calibration and at 18°C to 28°C (64°F to 82°F) with
                                                  relative humidity to 80%.
                                                  Accuracy specifications are given as:
                                                  ± % of reading ± number of least significant digits.
                                                  4.1 GENERAL
                                                  MAXIMUM VOLTAGE BETWEEN TERMINALS AND
                                                  EARTH GROUND: 1000V DC or 700 rms AC (sine)
                                                  FUSE PROTECTION: µA, mA: F 500mA/250V
                                                                   A: F 10A/250V
                                                    -9-
Downloaded from www.Manualslib.com manuals search engine
                                                  POWER SUPPLY: 9V battery, NEDA 1604 or 6F22
                                                  DISPLAY: LCD, 3999 counts MAX, updates 2-3/sec
                                                  MEASURING METHOD:
                                                  Dual - slope integration A/D converter
                                                  OVERRANGE INDICATION:
                                                  “OL” figure only on the display
                                                  POLARITY INDICATION:
                                                  “—” displayed for negative polarity
                                                  LOW BATTERY INDICATION:“               ”displayed
                                                  OPERATING TEMPERATURE:
                                                  0°C to 40°C (32°F to 104°F)
                                                  STORAGE TEMPERATURE:
                                                  -10°C to 50°C (14°F to 122°F)
                                                  DIMENSION: 91×189×31.5mm
                                                  WEIGHT: 310g (including battery)
                                                  4.2 AC VOLTAGE
                                                    Range   Resolution              Accuracy
                                                   400mV        0.1mV                   --
                                                      4V        1mV
                                                                           ± 0.5 % of rdg ± 0.2% of full
                                                     40V        10mV
                                                                           scale ± 2 digit
                                                    400V       100mV
                                                    750V         1V          ± 0.8 % of rdg ± 2 digits
                                                  Input Impedance: 10MΩ.
                                                  Frequency Range: 40Hz to 400Hz
                                                  Response: Average, calibrated in rms of sine wave
                                                  4.3 DC VOLTAGE
                                                                                                - 10 -
Downloaded from www.Manualslib.com manuals search engine
                                                    Range     Resolution            Accuracy
                                                   400mV       0.1mV          ± 0.8 % of rdg ± 2 digit
                                                      4V        1mV           ± 0.5 % of rdg ± 2 digit
                                                     40V        10mV          ± 0.5 % of rdg ± 2 digit
                                                    400V       100mV          ± 0.5 % of rdg ± 2 digit
                                                    1000V        1V          ± 0.8 % of rdg ± 2 digits
                                                  Input impedance: 10MΩ, more than 100MΩ at 400mV
                                                  range.
                                                  4.4 DC CURRENT
                                                   Range     Resolution              Accuracy
                                                   400µA        0.1µA         ± 0.8 % of rdg ± 2 digits
                                                    4mA          1µA          ± 0.8 % of rdg ± 2 digits
                                                    40mA        10µA          ± 0.8 % of rdg ± 2 digits
                                                   400mA       0.1mA          ± 1.2% of rdg ± 2 digits
                                                     10A        10mA          ± 2.0 % of rdg ± 5 digits
                                                  Overload protection: F 500mA fuse for µA and mA
                                                  ranges, F 10 A fuse for A range.
                                                  4.5 AC CURRENT
                                                   Range    Resolution             Accuracy
                                                                              ± (0.8~8.0)% of rdg
                                                   400µA        0.1µA
                                                                                    ± 3 digits
                                                    4mA          1µA        ± 0.8% of rdg ± 3 digits
                                                    40mA        10µA        ± 0.8% of rdg ± 3 digits
                                                   400mA       0.1mA        ± 1.2% of rdg ± 3 digits
                                                     10A       10mA         ± 2.0 % of rdg ± 5 digits
                                                    - 11 -
Downloaded from www.Manualslib.com manuals search engine
                                                  Overload protection: F 500mA fuse for µA and mA
                                                  ranges, F 10 A fuse for A range.
                                                  Frequency Range: 40Hz to 400Hz
                                                  Response: Average, calibrated in rms of sine wave
                                                  4.6 RESISTANCE
                                                     Range        Resolution              Accuracy
                                                      400Ω           0. 1Ω          ± 0.8% of rdg ± 3 digits
                                                      4KΩ              1Ω           ± 0.8% of rdg ± 1 digits
                                                     40KΩ             10Ω           ± 0.8% of rdg ± 1 digits
                                                     400KΩ           100Ω           ± 0.8% of rdg ± 1 digits
                                                      2MΩ             1KΩ           ± 0.8% of rdg ± 1 digits
                                                     40MΩ            10KΩ           ± 1.2% of rdg ± 2 digits
                                                  Max. open circuit voltage: 3.2V
                                                  4.7 CAPACITANCE
                                                      Range       Resolution               Accuracy
                                                        4nF          1pF
                                                       40nF         10pF
                                                      400nF        100pF
                                                                                          ± 3.0% ± 20
                                                        4µF          1nF
                                                       40µF         10nF
                                                      200µF        100nF
                                                  4.8 FREQUENCY
                                                     Range     Resolution                  Accuracy
                                                    100kHz       1Hz                       ±2%±5
                                                                                                  - 12 -
Downloaded from www.Manualslib.com manuals search engine
                                                  4.9 DIODE
                                                  Function  Resolution         Test         Open circuit
                                                                               current      voltage
                                                              1mV              25µA         3.0V
                                                  4.10 AUDIBLE CONTINUITY
                                                  Function                   Description
                                                              If resistance is lower than 50Ω, built-in
                                                                         buzzer will sound.
                                                  4.11 TRANSISTOR
                                                  Function  Range           Base current       Vce
                                                     hFE    1 to 1000           25µA             3.0V
                                                  5. ACCESSORIES
                                                  5.1 SUPPLIED WITH THE MULTIMETER
                                                  Test leads   Electric Rating 1500V, 10A       1 couple
                                                  Battery      9V NEDA 1604 or 6F22              1 pcs
                                                  Operating Manual                              1 pcs
                                                  Holster                                        1 pcs
                                                  Capacitance Testing Adapter                   1 pcs
                                                  5.2 HOW TO USE THE HOLSTER
                                                  The holster is used to protest the meter and to make the
                                                  measurement more comfortable, it comes with tow
                                                  stands installed together. The figure shows how to use
                                                  the holster to:
                                                    - 13 -
Downloaded from www.Manualslib.com manuals search engine
                                                  A: Support the meter with a standard angle.
                                                  B: Support the meter with a small angle using the little
                                                  stand
                                                  C: Hang the meter on the wall using the little stand.
                                                  Take the little stand off from the back side of the large
                                                  and insert it into holes located upper on the holster.
                                                  D. hold test leads
                                                  6. BATTERY & FUSE REPLACEMENT
                                                  If the sign“         ”appears on the LCD display, it
                                                  indicates that battery should be replaced.
                                                  Remove screws on the back cover and open the case.
                                                  Replace the exhausted battery with a new one.
                                                                                                  - 14 -
Downloaded from www.Manualslib.com manuals search engine
                                                  Fuse rarely need replacement and blow almost always
                                                  as a result of the operator’s error. Open the case as
                                                  mentioned above, and then take the PCB out from the
                                                  front cover. Replace the blown fuse with same ratings.
                                                          WARNING
                                                  Before attempting to open the case, be sure that test
                                                  leads have been disconnected from measurement
                                                  circuits to avoid electric shock hazard.
                                                  For protection against fire, replace fuse only with
                                                  specified ratings:
                                                  F1: 200mA/250V (quick acting)
                                                  F2: F20A/250V
                                                    - 15 -