CX506a Multitester Manual
CX506a Multitester Manual
CX506a
       MULTITESTER
       INSTRUCTION MANUAL
D-1034-157 CX506a(和・英)取扱説明書
CONTENTS
  [1] SAFETY PRECAUTIONS:Before use, read the following safety precautions   001
    1-1 Warning Instruction for safe use                                     001
    1-2 Explanation of Warning Symbols                                       002
    1-3 Overload Protections                                                 002
    1-4 Influence of the electromagnetic field                               002
  [2] APPLICATION AND FEATURES                                               003
    2-1 Application                                                          003
    2-2 Features                                                             003
  [3] NAME OF FUNCTIONS                                                      003
  [4] SCALE READING                                                          004
  [5] DESCRIPTION OF FUNCTIONS                                               005
    5-1 Selections, adjusters and switches                                   005
    5-2 How to Use the Stand                                                 005
  [6] MEASUREMENT PROCEDURE                                                  006
    6-1 Start -up Inspection                                                 006
    6-2 How to select an appropriate range                                   006
           (Selection of an appropriate range)
    6-3 Preparation for Measurement                                          006
    6-4 Voltage Measurement                                                  008
      6-4-1 DCV Measurement ( )                                              008
      6-4-2 ACV Measurement ( )                                              009
    6-5 DCA Measurement ( )                                                  010
    6-6 Resistance Measurement                                               011
      6-6-1 Resistance Measurement ( Ω )                                     011
      6-6-2 Terminal to Terminal Current ( LI )                              012
    6-7 Capacitance Measurement ( )                                          013
      6-7-1 C1,C2 Range                                                      013
      6-7-2 C3 Range                                                         015
    6-8 Transistor Measurement                                               016
      6-8-1 ICEO Measurement                                                 016
D-1034-157 CX506a(和・英)取扱説明書
   11. Never use the meter on the electric circuits that exceed 6kVA.
   12. Pay special attention when measuring the voltage of AC 33
       Vrms (46.7V peak) or DC 70V or more to avoid injury.
   13. Never apply an input signals exceeding the maximum rating
       input value.
   14. Never use the meter for measuring the line connected with
       equipment (i.e. motors) that generates induced or surge
       voltage since it may exceed the maximum allowable voltage.
   15. Never use the meter if the meter or test leads are damaged or
       broken.
   16. Never use uncased meter.
   17. Be sure to use a fuse of the specified rating or type. Never use a
       substitute of the fuse or never make a short circuit of the fuse.
   18. Always keep your fingers behind the finger guards on the
       probe when making measurements.
   19. Be sure to disconnect the test pins from the circuit when
       changing the function or range.
   10. Before starting measurement, make sure that the function and
       range are properly set in accordance with the measurement.
   11. Never use the meter with wet hands or in a damp environment.
   12. Never open rear case except when replacing batteries or fuse.
       Do not attempt any alteration of original specifications.
   13. To ensure safety and maintain accuracy, calibrate and check
       the tester at least once a year.
   14. Indoor use.
                                  — 1 —
D-1034-157 CX506a(和・英)取扱説明書
                                 — 2 —
D-1034-157 CX506a(和・英)取扱説明書
  2-1 Applications
    This instrument is portable multimeter designated for
    measurement of weak current circuit.
  2-2 Features
      High-Sensitivity(DC50kΩ /V)meter
      Capacitance measurement by built-in transistor oscillator
      IEC61010-1 MEASUREMENT CAT- Ⅲ MAX. 600V
      Wide measurement functions 26-ch switch
      Transistor check function
      Polarity reversal switch for DCV and DCA
                  Finger guard
    Test pin                        Test probe (Red)
                                                                Plug
                                                                       Clip
    Plug
                          Test lead (TL-21)                   Clip lead (CL-506)
                                         — 3 —
D-1034-157 CX506a(和・英)取扱説明書
     
                                                                             
     
                                                                             	
     
                                                                            
                                    — 4 —
D-1034-157 CX506a(和・英)取扱説明書
Zero position
                                  — 6 —
D-1034-157 CX506a(和・英)取扱説明書
START
    Connect the black plug        Short the red and       Stop using it and
     of the test lead to the        black test pins.        have it repaired.
     − terminal.
                                                
                                                        
                                       — 7 —
D-1034-157 CX506a(和・英)取扱説明書
Battery
                                   
                                   
                           
    3) Measurement procedure
       Shift the polarity reversal switch to + position.
        Connect the black plug of the test lead to the − input
        terminal and the red plug to the + input terminal.
        Set the function/range selector to an appropriate DCV
        range.
       Apply the black test pin to the negative potential side of the
        circuit to measure and the red test pin to the positive
        potential side.
       Read the pointer on V・A scale.
       After measurement, remove the red and black test pins from
        the circuit measured.
                                — 8 —
D-1034-157 CX506a(和・英)取扱説明書
                                             
                                                    
                                    
Outlet
                                    
                                    
                           
    3) Measurement procedure
       Shift the polarity reversal switch to + side.
        Connect the black plug of the test lead to the − input
        terminal and the red plug to the + input terminal.
        Set the function/range selector to an appropriate ACV
        range.
       Apply the black and red test pin to measuring circuit.
       Read the pointer on V・A scale.
        The AC3V range only uses the“AC3V”scale.
       After measurement, remove the red and black test pins
        from the circuit measured.
                                —   9   —
D-1034-157 CX506a(和・英)取扱説明書
                                     Resistor       Battery
                                                       
                                                
                                    
                                    
                            
                                — 11 —
D-1034-157 CX506a(和・英)取扱説明書
                             — 12 —
D-1034-157 CX506a(和・英)取扱説明書
                                              Capacitor
                                 
 	                 	
                                 
                                           
                                  
                          
                               — 13 —
D-1034-157 CX506a(和・英)取扱説明書
   Note
    Measuring frequency
     C1range : approx. 900Hz     C2range : approx. 800Hz
    Measuring voltage
     C1range : approx.8.0V (peak)/When 200pF is measured
     C1range : approx.0.5V (peak)/When 0.05µF is measured
     C2range : approx.4.0V (peak)/When 0.1µF is measured
     C2range : approx.0.7V (peak)/When 5.0µF is measured
    Application
     ・Test of the cord (Use the C1 range)
Cx 1
Cx 2
Cx 1>Cx 2
                               — 14 —
D-1034-157 CX506a(和・英)取扱説明書
  6-7-2 C3 range
    1) Application
       Measured large capacitor
    2) Measuring range:       C3 range: 1∼2000µF
    3) Measurement procedure
       Shift the polarity reversal switch to + side.
        Connect the black plug of the test lead to the − input
         terminal and the red plug to the + input terminal.
       Set the function/range selector to C3 range.
       Discharge the capacitance before measuring it.
       Short the test pins and turn the 0 Ω・C ∞ adjuster to have
         the pointer align exactly with ∞ of C3 scale.
       Apply the test pins to the measuring capacitor. Then, read
         the maximum reading on the C3 scale.
C3
                          
                                                             Capacitor
                                                                   
                                          C3      
                                                  
                                                         Wire
                              — 15 —
D-1034-157 CX506a(和・英)取扱説明書
                                                    transistor
                                                      
                                       
                                                
                                                            E: emitter
                                                          B: base
                                                            C: collector
                                — 16 —
D-1034-157 CX506a(和・英)取扱説明書
        Case of NPN-TR            	
                                              
                                          transistor
                                           
                                                           
                                         
                                                   
                                                           E: emitter
                                                        B: base
                                                           C: collector
                                 — 17 —
D-1034-157 CX506a(和・英)取扱説明書
    1) Application
       The probe is suitable for measuring voltage of high impedance
       circuits, such as CRT anode voltage of TV sets.
    2) Measuring range:
                                                            CRT
       HV PROBE (DC120mV)
                                                                      Anode
    3) Measurement procedure
        Shift the polarity                                             
          reversal switch to +
          position.
       Connect the black plug                                
          of the HV Probe to the                       Earth line
          −input terminal and                            
          the red plug to the +
                                                       
          input terminal.
        Set the function/range
          selector to HV PROBE position.
       First, connect the clip (black) of the probe to the earth line (−) in
          the circuit to be measured, and then apply the measuring pin on
          the probe body to your measuring point.
       Read the pointer on V・A (0∼30) scale as kV.
        After measurement, remove the measuring pin from the
          measured circuit, and then remove the clip.
                                  — 18 —
D-1034-157 CX506a(和・英)取扱説明書
  [7] MAINTENANCE
                                 WARNING
   1.This section is very important for safety. Read and understand
      the following instruction fully and maintain your instrument
      properly.
   2. The instrument must be calibrated and inspected at least once a
      year to maintain the safety and accuracy.
    If your instrument fails any of above check, do not use it, and have it
    repaired or replace it to new one.
  7-2 Calibration
    The manufacturer may conduct calibration and inspection. For
    more information, please contact the dealer or manufacturer.
                                 — 19 —
D-1034-157 CX506a(和・英)取扱説明書
                                  Spare fuse
                Battery (6F22)             Battery (R6)
                    9V X1                   1.5V X2
                          9V
                                          1.5V
Fuse
                                 — 20 —
D-1034-157 CX506a(和・英)取扱説明書
  8-1 Repair
    If the multimeter fails during use, check the following items before
    sending it for repair.
        Is the battery not exhausted ?
        Are the test leads not disconnected ?
        Is the fuse not blown ?
    We repair a defective product at cost. When mailing it to us for
    repair, do not use the same cardboard box in which it was
    delivered to you because it may receive damage in transit. Please
    send it in a box at least five times larger than the original box with
    enough cushioning material stuffed around it.
                                 — 21 —
D-1034-157 CX506a(和・英)取扱説明書
[9] SPECIFICATIONS
       *MEASUREMENT CATEGORY
        ・CATⅠ:Secondary electrical circuits connected to an AC
              electrical outlet through a transformer or similar device.
        ・CATⅡ:Primary electrical circuits in equipment connected to
              an AC electrical outlet by a power cord.
        ・CATⅢ:Primary electrical circuits of heavy equipment
              connected directly to the distribution panel, and
              feeders from the distribution panel to outlets.
                                — 22 —
D-1034-157 CX506a(和・英)取扱説明書
      Band width
      3V and 12V range : 40Hz∼30kHz                     30V range : 40Hz∼10kHz