Operating instructions
Electronic preset counter/timer
KALtrol-SP (KAT-SP/KAT-SPS)
1.1 Foreword                                                     personnel.
           Please read this instruction manual                   The device must compulsorily be
           entirely and carefully before installa-               protected with approved external
           tion and start-up. Please observe all                 fuses. The value of these fuses
           warnings and advices, both for your                   can be found in the technical
           own safety and for operating safety.                  information.
           The intended protection can be
           impaired If the device is not used                    This symbol is used on the device
           in accordance with this instruction                   to remind of the existence of dan-
           manual.                                               gers, which are referred to in this
                                                                 manual.
1.2 Safety instructions and warnings
            Please use the device only if its        1.3 Use according to the intended purpose
            technical condition is perfect. It       The preset counter counts pulses and times up
            should be used only for its intended     to max. 30 Hz and offers various operating
            purpose, in accordance with this         modes. The present counter simulta-neously
            instruction manual.                      processes a programmed preset value. Any
            Defective or damaged devices             other use is considered as non compliant with
            should be disconnected from the          the intended use.
            mains immediately and taken out of       The application area for this device lies in
            operation.                               industrial processes and controls, in the fields
            Repair work may be performed             of manufacturing lines for metal, wood, plastics,
            only by the manufacturer‘s repairs       paper, glass, textile and other like industries.
            service.                                 Overvoltages at the terminals of the device
            This device may be opened only           must be kept within the limits of Overvoltage
            for battery replacement. For this        Category II. The device must only be operated
            purpose, the device must be sepa-        when properly mounted in a panel and in accor-
            rated from all electric circuits and     dance with the section “Technical Data”.
            de-energized.                            The device is not suitable for use in areas with
            Only connect the device to the           explosion protection and for areas excluded in
            electricity networks provided to that    EN 61010-1. If the device is used to monitor
            purpose.                                 machines or processes in which, in the event of
            The safety of the system in which        a failure of the device or an error made by the
            the device is integrated is the          operator, there might be the risk of damaging
            responsibility of the installer.         the machine or causing an accident to the
            Disconnect all electricity networks      operators, it is your responsibility to take the
            prior to any installation or mainte-     appropriate safety measures.
            nance work.                              The device has been designed for indoor ope-
            Use exclusively cables approved in       ration. It may nevertheless be used outdoors,
            your country and designed for your       provided the technical data is complied with.
            temperature and power ranges.            In this case, take care to provide suitable UV
            Installation and service work shall be   protection.
            carried out exclusively by qualified
                                                                                              Page 1
1.4 Mounting in a control panel                                   The device must be protected exter-
           Mount the device away from heat                        nally for proper operation. Informa-
           sources and avoid direct contact                       tion about the prescribed fuses can
           with corrosive liquids, hot steam or                   be found in the technical data.
                                                       DANGER
           the like.                                              The relay output is not protected /
 CAUTION
           Provide a free space of 10 mm all                      overcurrent-limited internally in the
           around the device for its ventilation.                 device. Without suitable protection
           The device should be mounted so                        of the relay output, undesired heat
           that the terminals are out of the                      development or even fire may
           reach of the operator and cannot be                    occur. The relay output must be
           touched by him. When mounting the                      protected externally by the manuf-
           device, consider the fact that only                    acturer of the plant. It must also be
           the front side is classified as acces-                 made sure that, even in case of a
           sible for the operator.                                malfunction, the values stated in the
                                                                  technical data are under no circum-
1.5 Mounting instructions                                         stances exceeded.
1. Remove the mounting clip from the device.
2. Insert the device from the front into the panel   During installation, make sure that the voltage
   cut-out, ensuring the front-panel gasket is        of the counting and reset inputs and the wiring
   correctly seated.                                  of the output contacts are both powered by the
3. Slide the fixing clip from the rear onto the      same mains phase, in order not to exceed the
    housing, until the spring clamps are under        maximum permitted voltage of 250V.
    tension and the upper and lower latching lugs
    have snapped into place.                          The cables and their insulation must be desi-
Note: In case of proper installation, IP65 can be     gned for the planned temperature and voltage
reached on the front side.                            ranges. Regarding the type of the cables,
                                                      adhere to the applicable country and final device
1.6 Electrical installation                           standards. The cross sections allowed for the
This device is powered by an internal battery.        screw terminals can be found in the technical
                                                      data.
            The device must be disconnected
            from any power supply prior to any        Before starting the device, check the cables
            installation or maintenance work.         for proper wiring and tightening. The screws of
            Make sure that no more voltages           unused screw terminals must be screwed to the
 DANGER
            LIABLE TO CAUSE AN ELECTRO-               stop, so that they cannot loosen and get lost.
            CUTION are present.
            Installation or maintenance work
            may only be carried out by qualified
            personnel and in compliance with
            the applicable national and internati-
            onal standards.
            Take care to separate all extra-low
            voltages entering or exiting the
            device from hazardous electrical
            conductors by means of a double or
            reinforced insulation (SELV circuits).
Page 2
The device has been designed for overvoltage            1.8 Start-up
category II. If higher transient voltages cannot       • Is the device properly set and programmed
be excluded, additional protection measures               (function; max. counting frequency for coun-
must be taken in order to limit the overvoltage to        ters)?
the values of CAT II.
                                                       1.9 Failure possibilities and causes
Advice on noise immunity                               Impossible to use the keys:
All connections are protected against external         • Key lock input activated
sources of interference. The installation location     Counter does not count:
should be chosen so that inductive or capacitive       • Wrong or reversed wiring of the counting input
interference does not affect the device or its         • Setting of an input signal not matching the
connecting lines! Interference (e.g. from switch-         pulse generator.
mode power supplies, motors, clocked control-          • No ground connection between the pulse
lers or contactors) can be reduced by means of            generator and the counter.
appropriate cable routing and wiring.                  • Signal levels do not reach the switching
                                                          threshold of the counter.
Measures                                               Relay output signal missing:
• Use only shielded cable for signal and control      • Output contact short-circuited
   lines. Connect cable shield at both ends. Mini-     • Output signal (fugitive signal duration) pro-
   mum cable conductor cross-section 0.14 mm2.            grammed too short
• The shield connection to the equipotential          • Relay contact defective due to excessive load
   bonding should be as short as possible and          • Relay current below min. value
   with a contact area as large as possible (low-      Displaying noFunc:
   impedance).                                         • Battery is unloaded and must be replaced.
• Only connect the shields to the control panel, if
   the latter is also earthed.                         If, despite all, your device still does not operate,
• Install the device as far away as possible from     contact your local representative or call us
   noise-conducting cables.                            directly for technical support.
• Avoid routing signal or control cables parallel
  to power lines.                                      When sending your device back, please attach
                                                       a short description of the failure, of the program-
1.7 Cleaning and maintenance                           ming and of the connection diagram, in order to
The front side of the unit should only be cleaned      allow us to reproduce a possibly existing defect
using a soft damp (water!) cloth. Cleaning of          and to repair your device as quickly as possible.
the embedded rear side is not planned and is
the responsibility of the service personnel or of      2. Inputs
the installer.                                         2.1 Reset input terminal 6
                                                       This input is connected functionally in parallel
In normal operation, this device is maintenance-       with the red Set key and sets the counter to
free. Should the device nevertheless not ope-          zero in case of adding counting and to the pre-
rate properly, it must be sent back to the manuf-      set value in case of subtracting counting. The
acturer or to the supplier. Opening and repairing      reset function is dynamic.
the device by the user is not allowed and can
adversely affect the original protection level.
                                                                                                  Page 3
2.2 Counting input (Start input) terminal 5           during this countdown, the device switches back
Input damped to 25 Hz for pulse counting with         to the previous operating mode. If the keys are
the pulse preset counter or static start input with   released after 0 is reached, parameter setting is
the time preset counter (time counting active         activated and the first parameter is displayed.
when input is active, time indicator located on
top left flashes when time counting is active).       4.2 Setting the parameters
                                                      The key of the 1st decade allows switching
Note: The reset input (terminal 6) and the coun-      among the available parameters (e.g. add-sub)
ting input (terminal 5) are inputs with a common      and the key of the 6th decade allows switching
connection (terminal 7) and must be controlled        over to the following function. The duration of the
wit the same polarity.                                fugitive pulse is set using the keys of decades
                                                      1-3. After the last input (dP), the key of the 6th
2.3 Key lock input                                    decade allows jumping back to the start of the
Tre front keys (including the set key) are locked     programming routine; pressing simultaneously
when this input (terminal 2) is set to +3 VDC         the reset key and the keys of the 5th and 6th
(terminal 1).                                         decades allows saving the parameters.
3. Output                                             Note: Every time the programming routine is
Relay with potential-free contact, contact pro-       called up, the counter switches back to the
grammable as normally open or normally closed         basic parameter setting, i. e. the previous pro-
(terminal 3 and terminal 4). For the adding count,    gramming is lost and the counter and preset
the relay is activated when > preset value, for       values are reset to zero.
the subtracting count, it is activated when < 0.
The output signal is programmable as a fugitive       4.3 Parameters description
signal between 0.1 and 99.9 seconds or as a           (see also the function sequence)
bistable signal with the setting 0.0 (Loop OFF) no    Funct.Count:	Pulse preset counter operating
automatic repetition. When the relay is active, a     mode
colon symbol is displayed on the left end of the      Funct.Time: 	 Time preset counter operating
lower row. The output relay shall not exceed a        mode
switching frequency of 4 Hz.                          tMode:	Programmable time range (only
                                                                    for Funct.Time), SEC=seconds,
    etting the operating mode and the
4. S                                                                Min=minutes, hour=hours
   operating parameters                               Count Add: 	 Adding counting mode
4.1 New installation                                  Count Sub: 	 Subtracting counting mode
In case of a new installation of the counter, after   LooP off: 	 Automatic repetition off
a battery replacement that lasted more than 20                      When parameter LooP is „off“ and                                                                    	
seconds, or if the counter switched over to the                     dELAY time is 0.0, the output is
„NoFunc“ safety mode because of a too low                           activated when the preset value
battery voltage, the preset counter switches                        is reached and remains activated
                                                                    until reset is performed.
automatically to the programming mode. All
                                                      LooP on:	Automatic repetition on
parameters must then be input again (see 4.2).
                                                                    When parameter „LooP is „on“, the
Parameter modification is also possible during
                                                                    counter is reset automatically and
operation. For safety reasons, to access the                        the output emits a fugitive pulse
programming mode, the reset key and the keys                        for the time of the duration set in
of the 5th and 6th decades must be pressed                          dELAY.
simultaneously. The lower display line then dis-      rELAY no:	Normally open contact
plays the message INIT. If these 3 keys remain        rELAY nc:	 Normally closed contact
pressed, a countdown from 5 to 0 takes place          dELAY:	Output signal duration: fugitive
in one-second intervals. If the keys are released                   signal programmable from 0.01
Page 4
     – 99.9 seconds, 0.0 = bistable                   allows preventing malfunctions. After battery
     output (active until reset) program-             replacement in „noFunc“ mode, the device
     mable only for Loop off, in case of              requires re-parameterizing. Please read chapter
     Loop on, 0.0 is set automaticaly to              Battery replacement to that purpose.
     1.0 second.
                                                      
dP:	Programmable decimal point:                                 Note: In order to avoid loosing the
     pulse preset counter: 0 - 0.00000                           settings and values because of
     only for display purposes,                                  switching to Safe Mode, the service
     time preset counter 0 – 0.0                                 plan of the device should include
     (seconds) 0 – 0.00 (minutes,                                timely battery replacement in func-
     hours) determines the resolution.                           tion of the ambient temperature and
     If the relay output is switched from                        of the switching cycles of the relay.
     „no“ to „nc“, the relay condition
     must be taken over by pressing                   5.5 Battery replacement
     the reset key.                                   Before replacing the batteries, disconnect the
                                                      voltages or switch the device off. Disconnect
5. Operating the counter                              the connector from the device. This also makes
5.1 Setting or resetting                              sure that no untimely pulses set the device in
The red set key or a pulse on the reset input         an undefined state during battery replacement.
sets the counter to zero in adding mode and to        Then push the battery lid backwards and remo-
the preset value in subtracting mode.                 ve the 2 batteries. Position the new batteries
                                                      so that the „-“ pole is connected to the battery
5.2 Preset setting                                    holder marked with „-“ on the printed circuit.
The preset value is displayed in the lower dis-       After battery replacement, the display may go
play line. Setting is carried out with the 6 preset   on displaying the LowBat message for 2 more
keys, each assigned to a decade. The set or           minutes. Pressing the reset key can shorten this
modified preset value is taken over at the sub-       delay, as the device checks the current battery
sequent setting or resetting of the counter.          voltage every time this key is pressed. However,
                                                      this also resets the current counter value.
5.3 Overflow or underflow
                                                      
In adding mode, overflow passes from 999999                      Important: If battery replacement
to zero, in subtracting mode from zero to                        lasts longer than 20 seconds, the
999999. The output signal remains unchanged.                     parameters, the preset value and
The display flashes in case of overflow or                       the current counter value can be
underflow. The counter stops after the overflow                  erased. In order to prevent this loss
or underflow of an additional decade.                            of information, we strongly recom-
                                                                 mend replacing the batteries one
5.4 Lo-Bat indicator                                             after the other.
In case of a too low battery load, the message
Lo-bat is diaplayed and the display flashes.          Only use the batteries described in the Techni-
The batteries must be replaced when „Lo-bat“          cal Data. Otherwise, proper operation cannot be
is displayed. If the battery voltage decreases in     guaranteed.
such a way that safe relay switching cannot be
ensured any more, the counter switches to Safe
Mode and displays the message „noFunc“. In
this mode, the device does not count any more
pulses or times, and the relay remains in its
last position. Thys way, the user knows exactly
whether the plant is operating optimally, which
                                                                                              Page 5
5.6 Segments test                                    7. Connection example
Segments test is activated by pressing simul-
taneously the reset key and the keys of the          Device type:
5th and 6th decades. Init 5 is then displayed;       KAT-SP
release immediately the keys, the segment test       12 ... 250 VAC/VDC
is started and all display segments are displa-
yed. Pressing the 1st decade key calls up the        V AC/DC Key lock
programming mode; all parameters must then
be re-programmed.
                                                                                 Count/
                                                                                 Start
6. Terminal assignment
                                                                        Reset
                                                     Device type:
                                                     KAT-SPS
                                                     Switch Closure/NPN
                                                            Key lock
Terminal
no.
1          +3 VDC for key lock on terminal 2
2          Key lock input
                                                       Count/           Reset
3          Relay contact                               Start
4          Relay contact
           AC/DC optocoupler counting input /        8. Factory programming:
5          start input (.820) / NPN counting input   Function(Funct):          counter(Count)
           / start input (.850)                      Counting mode(Count):	 adding(add)
           AC/DC optocoupler reset input (.820)      Auto. repetition(LooP): 	 none(oFF)
6                                                    Output contact(rELAY):	 normally open
           / NPN reset input (.850)
                                                     Output signal duration(dELAY):
           Common AC/DC input (.820) / 0V                                      bistable output(0.0)
7
           input (.850) for terminals 6 and 5        Decimal point(dP):		      none(0)
Page 6
9. Interference suppression measures                 Application area:	Pollution level 2
The following interference suppression               Insulation
measures may be required: if the output swit-        Front side:	Double /
ches inductive loads (relays, solenid valves),                          reinforced insulation
interferences must be suppressed by means of         Relay:	            Basic insulation
a freewheeling diode for direct current and of       UL approval: 	     File no.: E128604
an RC element for alternating current. For safe
and EMC-compliant operation, use shielded            10.3 Mechanical data
cables for the control inputs (counting, reset and   Protection level:	   IP65 (font side)
key lock inputs). The shield must be connected       Weight:	             appr. 80 g
at both ends. If this is not possible, a capacitor
(e.g. 10 nf) may be integrated on one side in the    10.4 Inputs
shield connection.                                   Inputs:	Reset, counting and key
                                                                             lock inputs
10. Technical data                                   Input polarity
10.1 General technical data                          (for reset and counting/start inputs):
Display: 	2-line LCD display, each                  12 ... 250 V AC/DC:	bidirectional optocoupler
                       with 6 decades 999999;                                input
                       height 7 resp. 4,5 mm         NPN:	NPN input activated by
Operating temperature:	-20°C ... +65°C                                      transistor or contact
                       (not condensing)              Key lock input:	is switched to +3 V DC
Storage temperature: 	-25°C ... +70°C                                      (terminal 1)
Altitude:	             up to 2000 m                  Min. input pulse duration:
                                                     Reset input:	           50 ms
10.2 Electrical data                                 Key lock input:	        15 ms
Voltage supply:	2 replaceable lithium               Input switching level:
                        batteries type 1/2 AA        12 ... 250 V AC/DC:	 LOW: < 1 V AC/DC
                        Lithium 3.6 V                                        HIGH: 12...250 V AC/DC
Data backup:	min. 8 years for 5 x 106               NPN:	                   LOW: 0 ... 0,8 V DC
                        output relay switching                               HIGH: 2.5 ... 30 V DC
                        cycles and 25°C opera-       Input frequency
                        ting temperature             Pulse counter:	         max. 30 Hz
EMC                                                  Time counter:	Measuring pulses
Interference emission:	EN 55011 Class B                                     < 200 ms are not
Interference immunity:	EN 61000-6-2,                                        measured accurately
                        EN 61326-1                   Accuracy:	System error
Device safety                                                                +/- 100 ms per
Designed to:	           EN 61010                                             measuring pulse
Protection class:	      2 (front side)               Time accuracy:	         200 ppm
                                                     Input resistance: 	     110 kΩ
           Only the front side is classified as
           accessible for the operator.
                                                                                             Page 7
10.5 Outputs
Outputs: 	bistable relay with poten-
                         tial-free contact – pro-
                         grammable as normally
                         open or normally closed
Max. switching voltage:250 V AC / 30 V DC
Max. switching current: 2 A
Max. switching capacity:
                         60 VA / 30 W
Mechanical service life:1x107
Electrical service life: 5x105 (1A 30Vdc)
Output response time: < 20 ms, max. 4 Hz
Insulation coordination: Basic insulation
UL ratings:              2A, 30Vdc, 50°C
10.6 Fuses / overcurrent limitations
Input signals: In case of a failure, the current
must be limited to a safe value via the inputs.
We recommand a 0.1A/T fuse.
Relay: The maximum switching current of 2A
shall in no case be exceeded.
10.7 Connections
7 plug-in screw terminals, pitch 5.08, cable
cross section max. 2.5 mm.
10.8 Insulation coordination
Connections among each other
KAT-SP: only the counting and reset inputs
have double / reinforced insulation
KATS-SPS: generally only basic insulation
10.9 Consumables
Battery: order code: N060045 Tadiran Lithium
SL350/S ½ AA 3.6V 1.2Ah
Page 8
11. Programming menu
 Funktionsablauf
 Press Reset
 key together
 with
 5 + 6 button                                                                                                                                             change battery
                                   count down                                                                                                      <20s
                                   with 1sec clock                                                                   timefor
                             <5s                               <1s                                                   batterychange
                                                                                                                               >20s
                   Init 5             Init 4,3,2, 1                            Init 0
                =5s
             8.8.8.8.8.8                                                                                                                                              Operating
 Segmenttest : 8.8.8.8.8.8     1                                                                                                                                      Mode
                                         Funct.                                                            Funct.
                                                                                 1
                                         Count                                                              tImE
                                            6                                                             tModE                             select one
                                                                                                                                      1
                                                                                                       hour, Min, SEC                         value
                                         Count                               select one                   Count                             select one
                                                                  1                                                               1
                                        Add, Sub                               value                     Add, Sub                             value
                                           6                                                                 6
                                          LooP                               select one                     LooP                            select one
                                                                     1                                                            1
                                         oFF;on                                value                       oFF;on                             value
                                            6                                                                6
                                         rELAY                               select one                    rELAY                            select one
                                                                     1                                                                1
                                         no, nc                                value                       no, nc                             value
                                           6                                                                 6
                                                                 1                                                                1
                                         dELAY                               to change                     dELAY                          2 to change
                                                                         2
                                          01.0                                  value                       01.0                               value
                                                                 3                                                                3
                                           6                                                                 6
                                            dP                                                               dP
                                    0, 0.0, 0.00, 0.000,             1       select one                  SEC: 0, 0.0                  1     select one
                                     0.0000, 0.00000                           value                hour+Min: 0, 0.0, 0.00                    value
                                           6                                                                  6
                                                      PressResetkey
                                                      togetherwith
                                                      5+6button
                                                                               SAVE
                                                                                          1-2 Sec
                                                           Operating
                                                           Mode
                                                                                                                                                                      Page 9
12. Dimensions                         13. Scope of the delivery
Dimensions in mm [inch]                Counter KALtrol-SP
                                       2 lithium batteries
                                       1 plug-in screw terminal
                                       1 front bezel for screw mounting, installation
                                         cross-section 50 x 50 mm
                                       1 front bezel for clip mounting, installation cross-
                                         section 50 x 50 mm
                                       1 mounting clip
                                       1 template for front panel cut-out
                                       1 operating instructions manual
                                       14. Order code
                                       KAT-SP = 12 … 250 V AC/DC inputs
                                       KAT-SPS = Switch Closure/NPN inputs
                                       15. Battery ordinance
                                                  This device includes lithium batte-
                                                  ries. In compliance with the Battery
                                                  Ordinance, we want to point out the
                                                  following:
                                       Batteries shall not be disposed of in the residual
                                       waste; you are obliged by law to return them. You
                                       can send the complete devices back to us after
                                       use. If you can remove the batteries properly, you
                                       can also dispose of them at a municipal collection
                                       point or at a retailer collecting batteries.
                                       Specific provisions on the return of lithium
                                       batteries:
                                       Avoid short-circuits! Protect the poles of the
                                       battery with insulating adhesive tape. The poles
                                       of a lithium battery shall not come in contact
                                       with metal objects, neither by accident nor inten-
                                       tionally !
                                       Pollutant-containing batteries bear a symbol
                                       made of a crossed-out bin and the chemical
                                       symbol of the heavy metal that determines their
                                       classification as containing pollutants. Thank you
                                       for your help!
  Contact: Kessler-Ellis Products   TEL: 732-935-1320 Web:www.kep.com
Page 10