110 WATTS DC2-110
SINGLE/MULTI OUTPUT DC-DC OUTPUT SPECIFICATIONS
Total Output Power at 50°C(1) 80W Convection Cooled(13, 15)
FEATURES: (See Derating Chart) 110W 300LFM Forced-Air Cooled(12, 14, 16)
Output Voltage Centering Output 1: 0.5% (All outputs
• Compact 3” x 5” x 1.3” Size • IEC 60601-1 3rd ed. Medical Cert. Output 2: 5.0% at 50% load)
• 2 Year Warranty • IEC 60950-1 2nd ed. ITE Certification Output 3: 5.0%
• 18-36VDC Input • 0-70°C Operating Temperature Output 4: 5.0%
• One to Four Outputs • RoHS Compliant Output Voltage Adjust Range Output 1: 95 - 105%
• 4242VDC Reinforced Insulation • Optional Chassis/Cover Load Regulation Output 1: 0.5% (10-100%
• Under/Overvoltage Lockout • Power Good Signal Output 2: 5.0% load change)
• Size/Pin Compatible with REL-110 Series (4001-5 Models) 8.0%
(2001 Model) 6.0%
Output 3: 5.0%
Output 4: 5.0%
Source Regulation Outputs 1 – 4: 0.5%
Cross Regulation Outputs 2 – 4: 5.0%
Output Noise Outputs 1 – 4: 1.0%
Turn on Overshoot None
Transient Response Outputs 1 – 4
Voltage Deviation 5.0%
Recovery Time 500S
CHASSIS/COVER OPEN FRAME Load Change 50% to 100%
Output Overvoltage Protection Output 1: 110% to 150%
SAFETY SPECIFICATIONS Output Overpower Protection 110-160% rated Pout, cycle on/off, auto recovery
Underwriters Laboratories UL 60950-1:2007, 2nd Edition Start Up Time 5 Seconds
File E137708/E140259 AAMI/ANSI ES60601-1:2005/(R) 2012 INPUT SPECIFICATIONS
CB Reports/Certificates (including all Input Voltage Range 18-36 VDC
National and Group Deviations)
Input Under-Voltage Lockout
IEC 62368-1:2014, 2nd Edition
Turn-On Voltage 14.5-17.5 VDC
IEC 60601-1:2005/A1:2012
Turn-Off Voltage 14.0-17.0 VDC
UL Recognition
CAN/CSA-C22.2 No. 60950-1-07, 2nd Edition Input Overvoltage Shutdown 37.0-43.0 VDC
Mark for Canada
CAN/CSA-C22.2 No. 60601-1:2014 Maximum Input Current 8.5 A
File E137708/E140259
Reflected Ripple Current 5%
EN 62368-1:2014, 2nd Edition Efficiency 82% Typ., Full Power, 24VDC, varies by model
TUV
EN 60601-1:2006/A1:2013 ENVIRONMENTAL SPECIFICATIONS
Ambient Operating 0C to + 70C
RoHS Directive (Recast) (2011/65/EU of June 2011) Temperature Range Derating: See Power Rating Chart
Ambient Storage Temp. Range - 40C to + 85C
MODEL LISTING Temperature Coefficient Outputs 1 – 4: 0.02%/C
GENERAL SPECIFICATIONS
MODEL OUTPUT 1(20) OUTPUT 2(20) OUTPUT 3(19) OUTPUT 4(19) Means of Protection
Primary to Secondary 2MOOP (Means of Operator Protection)
DC2-110-4001 +3.3V/10A(17) +5V/6A +12V/2A -12V/2A Primary to Ground 1MOOP (Means of Operator Protection)
DC2-110-4002 +5V/10A(17) +3.3V/6A +12V/2A -12V/2A Secondary to Ground Operational Insulation(Consult factory for 1MOOP or 1MOPP)
DC2-110-4003 +5V/10A(17) +3.3V/6A +15V/2A -15V/2A Dielectric Strength(7, 8)
DC2-110-4004 +5V/10A(17) -5V/6A +12V/2A -12V/2A Reinforced Insulation 4242 VDC, Primary to Secondary
DC2-110-4005 +5V/10A(17) -5V/6A +15V/2A -15V/2A Basic Insulation 2121 VDC, Primary to Ground
DC2-110-4006 +5V/10A(17) +24V/2A +12V/2A -12V/2A Operational Insulation 707 VDC, Secondary to Ground
DC2-110-4007 +5V/10A(17) +24V/2A +15V/2A -15V/2A Power Good Signal(11) Logic high with input voltage above Vin min.
DC2-110-3001 +5V/10A(17) +12V/3A -12V/3A Remote Sense (singles only)(9) 250mV compensation of output cable losses
DC2-110-3002 +5V/10A(17) +15V/2A -15V/2A Mean-Time Between Failures 100,000 Hours min., MIL-HDBK-217F, 25° C, GB
DC2-110-2001 +3.3V/10A(17) +5V/6A Weight 0.65 Lbs. Open Frame
DC2-110-2002 +5V/10A(17) +12V/5A 1.15 Lbs. Chassis and Cover
DC2-110-2003 +5V/10A(17) +24V/3A EMC SPECIFICATIONS
DC2-110-2004 +12V/5A -12V/4A Electrostatic Discharge EN61000-4-2 8KV contact/ 15KV air discharge A
DC2-110-2005 +15V/4A -15V/3A Electrical Fast Transients/Bursts EN61000-4-4 2KV, 5KHz/100KHz A
DC2-110-1001 2.5V/22A(18) Surge Immunity EN61000-4-5 2KV line to earth/ 1KV line to line A
DC2-110-1002 3.3V/22A(18) MAXIMUM OUTPUT POWER vs. AMBIENT TEMPERATURE
DC2-110-1003 5V/22A(18) 110
DC2-110-1004 12V/9.2A
100
DC2-110-1005 15V/7.3A F O R C E D A I R C O O L I N G
DC2-110-1006 24V/4.6A 90
Output Power (Watts)
DC2-110-1007 28V/3.9A 80
DC2-110-1008 48V/2.3A 70
C O N V E C T I O N C O O L I N G
60
ORDERING INFORMATION
50
Consult factory for alternate output configurations.
Consult factory for positive, negative or floating outputs. 40
C O N V E C T I O N C O O L I N G
Please specify the following optional features when ordering: 30
W I T H C H A S S I S / C O V E R
20
CH – Chassis I/O – Isolated Outputs
CO – Cover TS – Terminal Strip 10
BD – Reverse Input Protection 0
0 10 20 30 40 50 60 70
Ambient Temperature (C)
All specifications are maximum at 25°C/110W unless otherwise stated, may vary by model and
are subject to change without notice.
INTEGRATED
POWER DESIGNS 300 Stewart Road ■ Wilkes-Barre, PA 18706 ■ Phone: (570) 824-4666 ■ Fax: (570) 824-4843 ■ Email: sales@ipdpower.com ■ Web: www.ipdpower.com
DC2-110 SERIES MECHANICAL SPECIFICATIONS APPLICATIONS INFORMATION
1. Each output can deliver its rated current but Total Output Power must not exceed 110W
as determined by the cooling method.
2. Generally, adequate cooling is provided when semiconductor case temperatures do not
exceed 70°C rise and transformer temperature does not exceed 60°C rise at any
specified ambient temperature.
3. Sufficient area must be provided around power supply to allow natural movement of air to
develop in convection-cooled applications.
4. This product is intended for use as a professionally-installed component within information
technology, industrial, and medical equipment and is not intended for stand-alone
operation.
5. A minimum load of 10% is required on Output 1 to ensure proper regulation of remaining
outputs.
6. Peak-to-Peak Output Ripple and Noise is measured directly at the output terminals of the
power supply, without the use of the probe ground lead or retractable tip (tip-and-barrel
method), 20 MHz bandwidth.
7. This product was type-tested and safety-certified using the dielectric strength test voltages
listed in Table 6 of IEC 60601-1:2005. In consideration of Clause 8.8.3, care must be
taken to insure that the voltage applied to a reinforced insulation does not overstress
different types and levels of insulation. Primary and secondary-to-ground capacitors may
need to be disconnected prior to performing a dielectric strength test on the power supply
or the end product. It is highly recommended that the DC test voltages listed in DVB.1,
Annex DVB of UL 60601-1 1st Edition are not exceeded during a production-line dielectric
strength test of the assembled end product. Please consult factory for further information.
8. This power supply has been safety-approved and final-tested using a DC dielectric
strength test. Please consult factory before performing an AC dielectric strength test.
9. Remote-Sense terminals may be used to compensate for cable losses up to 250mV
(single output models only). The use of a twisted pair, decoupling capacitors and an
appropriately-rated low-impedance capacitor connected across the load will increase
noise immunity.
10. Maximum screw penetration into bottom chassis mounting holes is 0.100 inches.
Maximum screw penetration into side chassis mounting holes is 0.250 inches.
11. Power Good feature provides a logic-high signal from an open collector transistor when
DC input reaches minimum operating voltage.
12. 300LFM minimum of airflow must be maintained one inch above all points of top-side
components or cover when forced-air cooling is required.
13. Total Power must not exceed 80W with convection cooling on open-frame models except
where noted.
14. Total Power must not exceed 110W with 300LFM forced-air cooling on open-frame
models.
15. Total Power must not exceed 65W with convection cooling and Chassis/Cover option.
16. Total Power must not exceed 110W with 300LFM forced-air cooling and Chassis/Cover
option.
17. Rated 8A maximum with convection cooling.
18. Rated 16A maximum with convection cooling.
19. Total current from Outputs 3 & 4 must not exceed 3A with convection cooling.
20. Total current from Outputs 1 & 2 must not exceed 12A with convection cooling.
CONNECTOR SPECIFICATIONS
P1 DC Input 0.156 friction lock header mates with Tyco 640250-4 or
equivalent crimp terminal housing with Tyco 3-640706-1 or
equivalent crimp terminal.
P2 DC Output 0.156 friction lock header mates with Tyco 770849-8 or
(Single) equivalent crimp terminal housing with Tyco 3-640707-1 or
equivalent crimp terminal.
P2 DC Output 0.156 friction lock header mates with Tyco 1-770849-0 or
(Multiple) equivalent crimp terminal housing with Tyco 3-640707-1 or
equivalent crimp terminal.
G Ground 0.187 quick disconnect terminal.
P3 P.G./Sense 0.100 breakaway header mates with Molex 50-57-9006 or
(Single) equivalent crimp terminal housing with Molex type 71851 or
equivalent crimp terminal.
P3 P.G. 0.100 breakaway header mates with Molex 50-57-9002 or
(Multiple) equivalent crimp terminal housing with Molex type 71851 or
equivalent crimp terminal.
REV. P 03/20/2019