70 WATTS                                                                                           REL-70
SINGLE/MULTI OUTPUT AC-DC                                                                        OUTPUT SPECIFICATIONS
FEATURES:                                                                                     Total Output Power at 50°C(1)       50W               Convection Cooled(16)(18)
                                                                                              (See Derating Chart)                70W               300LFM Forced-Air Cooled(15)(17)(19)
•   Compact 2.5 x 4.5" x 1.2" Size             •   IEC 60601-1 3rd ed. Medical Cert.          Output Voltage Centering            Output 1:          0.5% (All outputs at 50% load)
•   2 Year Warranty                            •   IEC 62368-1 2nd Certification                                                  Output 2,3,4:      5.0%
•   Universal 85-264V Input                    •   IEC 60601-1-2 4th ed. EMC                  Output Voltage Adjust Range         Output 1:         95 - 105%
•   One to Four Outputs                        •   Class B Emissions per EN55011/32           Load Regulation                     Output 1:         0.5%      (10-100% load change)
•   High Efficiency                            •   RoHS Compliant                                                                 Output 2:         5.0%
•   0-70°C Operating Temperature               •   Optional Chassis/Cover                                                         (4001-5)          8.0%
                                                                                                                                  (2001)            8.0%
                                                                                                                                  Output 3:         5.0%
                                                                                                                                  Output 4:         5.0%
                                                                                              Source Regulation                   Outputs 1 – 4:    0.5%
                                                                                              Cross Regulation                    Outputs 2 – 4:    5.0%
                                                                                              Output Noise                        Outputs 1 – 4:    1.0%
                                                                                              Turn on Overshoot                   None
                                                                                              Transient Response                  Outputs 1 – 4
                                                                                                    Voltage Deviation             5.0%
                                                                                                    Recovery Time                 500S
                                                                                                    Load Change                   50% to 100%
                                                                                              Output Overvoltage Protection       Output 1:        110% to 150%
                                                                                              Output Overpower Protection         110-160% rated Pout, cycle on/off, auto recovery
          CHASSIS/COVER                                       OPEN FRAME                      Hold Up Time                        16mS min., Full Power, 85V Input
                            SAFETY SPECIFICATIONS                                             Start Up Time                       4 Seconds, 120V Input
                                                    UL 62368-1:2014, 2nd Edition                                             INPUT SPECIFICATIONS
             Underwriters Laboratories              CAN/CSA-C22.2 No. 62368-1-14              Protection Class                    I
             File E137708/E140259                   AAMI/ANSI ES60601-1:2005/(R) 2012         Source Voltage                      85 – 264 Volts AC
                                                    CAN/CSA-C22.2 No. 60601-1:2014            Frequency Range                     47 – 63 Hz
             CB Reports/Certificates (including all IEC 62368-1:2014, 2nd Edition             Peak Inrush Current                 40A
             National and Group Deviations)         IEC 60601-1:2005/A1:2012                  Efficiency                          78% Typ., Full Power, 230V, varies by model
                                                                                              Power Factor                        0.95 (Full Power, 230V)
             TUV SUD America
                                                    EN 62368-1:2014, 2nd Edition                                 ENVIRONMENTAL SPECIFICATIONS
                                                    EN 60601-1:2006/A1:2013                   Ambient Operating                   0C to + 70C
                                                                                              Temperature Range                   Derating: See Power Rating Chart
             Low Voltage Directive                  (2014/35/EU of February 2014)             Ambient Storage Temp. Range         - 40C to + 85C
             RoHS Directive (Recast)                (2011/65/EU of June 2011)                 Temperature Coefficient             Outputs 1 – 4:       0.02% /C
             Electrical Equipment (Safety) Regulations 2016 SI No. 1101                                                           3,000m ASL – Operating – Medical 60601-1
                                                                                              Altitude                            5,000m ASL – Operating – ITE/AV – 62368-1
             Restriction of the Use of Certain Hazardous Substances in EEE Regulations                                            12,192m ASL – Non-Operating
             2012 SI No. 3032 + 2019 SI No.492
                                                                                                                        GENERAL SPECIFICATIONS
                                                                                              Means of Protection
                                    MODEL LISTING                                                  Primary to Secondary           2MOPP (Means of Patient Protection)
MODEL NO. OUTPUT 1                    OUTPUT 2           OUTPUT 3           OUTPUT 4               Primary to Ground              1MOPP (Means of Patient Protection)
                                                                                                   Secondary to Ground            Operational Insulation(Consult factory for 1MOPP)
REL-70-4001        +3.3V/6A           +5V/5A             +12V/2A(21)        -12V/2A(21)       Dielectric Strength(8, 9)
REL-70-4002        +5V/6A             +3.3V/5A           +12V/2A(21)        -12V/2A(21)            Reinforced Insulation          5656 VDC, Primary to Secondary
REL-70-4003        +5V/6A             +3.3V/5A           +15V/2A(21)        -15V/2A(21)            Basic Insulation               2121 VDC, Primary to Ground
REL-70-4004        +5V/6A             -5V/5A             +12V/2A(21)        -12V/2A(21)            Operational Insulation          707 VDC, Secondary to Ground
REL-70-4005        +5V/6A             -5V/5A             +15V/2A(21)        -15V/2A(21)       Leakage Current
REL-70-4006        +5V/6A             +24V/2A            +12V/2A(21)        -12V/2A(21)            Earth Leakage                  <300µA NC, <1000µA SFC
REL-70-4007        +5V/6A             +24V/2A            +15V/2A(21)        -15V/2A(21)            Touch Current                  <100µA NC, <500µA SFC
REL-70-4009        6.7V/5A            5V/4A              +15V/2A(21)        -15V/2A(21)       Power Fail Signal(14)               Logic low with input power failure 10 ms
                                                                                                                                  minimum prior to Output 1 dropping 1%
REL-70-3001        +5V/6A             +12V/2A                               -12V/2A(21)
                                                                                              Remote Sense (singles only)(10)     250mV compensation of output cable losses
REL-70-3002        +5V/6A             +15V/2A                               -15V/2A(21)
                                                                                              Mean-Time Between Failures          100,000 Hours min., MIL-HDBK-217F, 25° C, GB
REL-70-3003        +5.1V/6A           +7.5V/2A                              -7.5V/2A(21)
                                                                                              Weight                              0.60 Lbs. Open Frame
REL-70-3004        +3.3V/6A           +7V/5A             +12V/2A(21)
                                                                                                                                  1.00 Lbs. Chassis and Cover
REL-70-2001        +3.3V/6A           +5V/5A
                                                                                               EMC SPECIFICATIONS (IEC 60601-1-2:2014, 4 TH ED./IEC 61000-6-2:2005)
REL-70-2002        +5V/6A             +12V/4A
                                                                                              Electrostatic Discharge             EN 61000-4-2      8KV contact / 15KV air discharge A
REL-70-2003        +5V/6A             +24V/2A
                                                                                              Radiated Electromagnetic Field      EN 61000-4-3      80MHz-2.7GHz, 10V/m, 80% AM              A
REL-70-2004        +12V/3A            -12V/3A
                                                                                              Electrical Fast Transients/Bursts   EN 61000-4-4      2 KV, 5KHz/100KHz                       A
REL-70-2005        +15V/3A            -15V/2A
REL-70-2006        +5.5V/6A           -5.5V/5A                                                Surge Immunity                      EN 61000-4-5      2 KV line to earth / 1 KV line to line A
                                                                                              Conducted Immunity                  EN 61000-4-6      0.15 to 80MHz, 10V, 80% AM               A
REL-70-1001        2.5V/14A(20)
                                                                                              Magnetic Field Immunity             EN 61000-4-8      30A/m, 60 Hz.                            A
REL-70-1002        3.3V/14A(20)
                                                                                              Voltage Dips                        EN 61000-4-11     0% UT, 0.5 cycles, 0-315° 100/240V A/A
REL-70-1003        5V/14A(20)
                                                                                                                                                    0% UT, 1 cycles, 0°         100/240V A/A
REL-70-1004        12V/5.8A                                                                                                                         40% UT, 10/12 cycles, 0° 100/240V B/A
REL-70-1005        15V/4.7A                                                                                                                         70% UT, 25/30 cycles, 0° 100/240V B/A
REL-70-1006        24V/2.9A                                                                   Voltage Interruptions               EN 61000-4-11     0% UT, 300 cycles, 0°       100/240V B/B
REL-70-1007        28V/2.5A                                                                   Radiated Emissions                  EN 55011/32       Class B
REL-70-1008        48V/1.5A                                                                   Conducted Emissions                 EN 55011/32       Class B
                            ORDERING INFORMATION                                              Harmonic Current Emissions          EN 61000-3-2      Class A
Consult factory for alternate output configurations.                                          Voltage Fluctuations/Flicker        EN 61000-3-3      Compliant
Consult factory for positive, negative or floating outputs.
Please specify the following optional features when ordering:                                 All specifications are maximum at 25°C/70W unless otherwise stated, may vary by model and are
CH - Chassis                                           I/O - Isolated Outputs                 subject to change without notice.
CO - Cover                                             TS - Terminal Strip
               INTEGRATED
               POWER DESIGNS 300 Stewart Road ■ Wilkes-Barre, PA 18706 ■ Phone: (570) 824-4666 ■ Fax: (570) 824-4843 ■ Email: sales@ipdpower.com ■ Web: www.ipdpower.com
     REL-70 MECHANICAL SPECIFICATIONS                                                                                            APPLICATIONS INFORMATION
                                                                                   1.                           Each output can deliver its rated current but Total Output Power must not exceed 70W, as
                                                                                                                determined by the cooling method.
                                                                                   2.                           Generally, adequate cooling is provided when semiconductor case temperatures do not
                                                                                                                exceed 70°C rise and transformer temperature does not exceed 60°C rise at any
                                                                                                                specified ambient temperature.
                                                                                   3.                           Sufficient area must be provided around power supply to allow natural movement of air to
                                                                                                                develop in convection-cooled applications.
                                                                                   4.                           This product is intended for use as a professionally-installed component within information
                                                                                                                technology, industrial, and medical equipment and is not intended for stand-alone
                                                                                                                operation.
                                                                                   5.                           A minimum load of 10% is required on Output 1 to ensure proper regulation of remaining
                                                                                                                outputs.
                                                                                   6.                           This product includes only one fuse in the input circuit. In consideration of Clause 8.11.5
                                                                                                                of IEC 60601-1:2005, a second fuse may be required in the end product.
                                                                                   7.                           Peak-to-Peak Output Ripple and Noise is measured directly at the output terminals of the
                                                                                                                power supply, without the use of the probe ground lead or retractable tip (tip-and-barrel
                                                                                                                method), 20 MHz bandwidth.
                                                                                   8.                           This product was type-tested and safety-certified using the dielectric strength test voltages
                                                                                                                listed in Table 6 of IEC 60601-1:2005. In consideration of Clause 8.8.3, care must be
                                                                                                                taken to insure that the voltage applied to a reinforced insulation does not overstress
                                                                                                                different types and levels of insulation. Primary and secondary-to-ground capacitors may
                                                                                                                need to be disconnected prior to performing a dielectric strength test on the power supply
                                                                                                                or the end product. It is highly recommended that the DC test voltages listed in DVB.1,
                                                                                                                Annex DVB of UL 60601-1 1st Edition are not exceeded during a production-line dielectric
                                                                                                                strength test of the assembled end product. Please consult factory for further information.
                                                                                   9.                           This power supply has been safety-approved and final-tested using a DC dielectric
                                                                                                                strength test. Please consult factory before performing an AC dielectric strength test.
                                                                                   10.                          Remote-Sense terminals may be used to compensate for cable losses up to 250mV
                                                          
                                                                                                                (single output models only). The use of a twisted pair, decoupling capacitors and an
                                                                                                                appropriately-rated low-impedance capacitor connected across the load will increase
                                                                                                                noise immunity.
                                                                                   11.                          Maximum screw penetration into bottom chassis mounting holes is 0.100 inches.
                                                                                                                Maximum screw penetration into side chassis mounting holes is 0.250 inches.
                                                                                   12.                          To comply with emissions specifications, all four mounting hole pads must be electrically
                                                                                                                connected to a common metal chassis. Chassis/Cover option is recommended. Refer to
                                                                                                                Operating Instructions for additional information.
                                                                                  13.                          Common RF shielding precautions may need to be taken to assure emissions
                                                                                                                compliance. Refer to Operating Instructions for additional information.
                                                                                   14.                          Power-Fail (AC-Good) feature provides a logic-low warning signal from an open collector
                                                                                                                transistor output 10ms prior to loss of output from AC failure, 5V/10mA.
                                                                                   15.                          300LFM minimum of airflow must be maintained one inch above all points of top-side
                                                                                                                components or cover when forced-air cooling is required.
                                                                                   16.                          Total power must not exceed 50W with convection cooling on open-frame models.
                                                                                   17.                          Total power must not exceed 70W with 300LFM forced-air cooling on open-frame models.
                                                                                   18.                          Total power must not exceed 40W with convection cooling and Chassis/Cover option.
                                                                                   19.                          Total power must not exceed 70W with 300LFM forced-air cooling and Chassis/Cover
                                                                                                                option.
                                                                                   20.                          Rated 10A with convection cooling.
                                                                                   21.                          Rated 1.5A with convection cooling.
                                                                                        MAXIMUM OUTPUT POWER vs. AMBIENT TEMPERATURE
                                                                                                                     70
                                                                                                                     60                FORCED        AI R   COOLI   NG
                                                                                         Output Power (Watts)
                                                                                                                     50
                                                                                                                                    CONVECTI          ON    COOLI   NG
                                                                                                                     40
                                                                                                                     30
                                                                                                                                       CONVECTI ON          COOLI NG
                                                                                                                     20                WI TH CHASSI         S/ COVER
                                                                                                                     10
                                                                                                                      0
                                                                                                                          0       10            20           30          40   50         60         70
                                                                                                                                          Ambient Temperature (C)
                                                                                                                                 CONNECTOR SPECIFICATIONS
                                                                                   P1                            AC Input              0.156 friction lock header mates with Tyco 640250-3 or
                                                                                                                                       equivalent crimp terminal housing with Tyco 3-640706-1 or
                                                                                                                                       equivalent crimp terminal.
                                                                                   P2                            DC Output             0.156 friction lock header mates with Tyco 770849-8 or
                                                                                                                 (Single)              equivalent crimp terminal housing with Tyco 3-640707-1 or
                                                                                                                                       equivalent crimp terminal.
                                                                                   P2                            DC Output             0.156 friction lock header mates with Tyco 770849-8 or
                                                                                                                 (Multiple)            equivalent crimp terminal housing with Tyco 3-640707-1 or
                                                                                                                                       equivalent crimp terminal.
                                                                                   G                             Ground                0.187 quick disconnect terminal.
                                                                                   P3                            P.F./Sense            0.100 breakaway header mates with Molex 22-55-2061 or
                                                                                                                                                                                                           REV. V 10/06/2022
                                                                                                                 (Single)              equivalent crimp terminal housing with Molex type 71851 or
                                                                                                                                       equivalent crimp terminal.
                                                                                   P3                            Power Fail            0.100 breakaway header mates with Molex 50-57-9002 or
                                                                                                                 (Multiple)            equivalent crimp terminal housing with Molex type 71851 or
                                                                                                                                       equivalent crimp terminal.
INTEGRATED
POWER DESIGNS 300 Stewart Road ■ Wilkes-Barre, PA 18706 ■ Phone: (570) 824-4666 ■ Fax: (570) 824-4843 ■ Email: sales@ipdpower.com ■ Web: www.ipdpower.com