Continuous Vacuum Regulators: Service Manual
Continuous Vacuum Regulators: Service Manual
Regulators
Service Manual
MED
400
300
100
80 120 500
-mmHg 50
60 140 60 40
LOW
70 30 200
HIGH
40 High Flow
160 600 High Vacuum
80 20
20
180 100
90 10
200 700 100
a
F U L L VA C - kP
Hg
-m m
User Responsibility
This Product will perform in conformity with the description thereof contained in this manual and accompa-
nying labels and/or inserts, when assembled, operated, maintained and repaired in accordance with the
instructions provided. This Product must be checked periodically. A malfunctioning Product should not be
used. Parts that are broken, missing, plainly worn, destroyed or contaminated, should be replaced immedi-
ately. Should such repair or replacement become necessary, Ohmeda Medical recommends that a tele-
phonic or written request for service advice be made to the nearest Ohmeda Medical Service Office. This
Product or any of its parts should not be repaired other than in accordance with written instructions pro-
vided by Ohmeda Medical, or altered without the prior written approval of Ohmeda Medical’s Safety
Department. The user of this Product shall have the sole responsibility for any malfunction which results
from improper use, faulty maintenance, improper repair, damage, or alterations by anyone other than
Ohmeda Medcial.
Important: Federal law in the U.S.A. and Canada restricts this device to sale by or on the order of a li-
censed medical practitioner.
Important: This document is not to be reproduced in any manner, nor are the contents herein to be dis-
closed to anyone without the express authorization of Ohmeda Medical.
Technical Competence
The procedures described in this service manual should be performed by competent individuals who have
a general knowledge of and experience with devices of this nature. No repairs should ever be undertaken
or attempted by anyone not having such qualifications.
Genuine replacement parts manufactured or sold by Ohmeda Medical must be used for all repairs.
Read completely through each step in every procedure before starting the procedure; any exceptions may
result in a failure to properly and safely complete the attempted procedure.
2/Scope
2.1 North American Vacuum Regulators. ................................................................................................... 2-1
2.2 International Vacuum Regulators .......................................................................................................... 2-2
4/Operation
4.1 Equipment Set-up ................................................................................................................................. 4-1
4.2 Attaching the Safety Trap ...................................................................................................................... 4-2
4.3 Mode Selection .....................................................................................................................................4-3
4.4 Setting the Suction Level ...................................................................................................................... 4-4
4.5 Pre-Use Checkout Procedure ............................................................................................................... 4-4
4.6 Patient Set-up .......................................................................................................................................4-7
6/Troubleshooting
i
Table of Contents
8/Service Checkout Procedure
8.1 Set-up ...................................................................................................................................................8-1
8.2 Flow Test ..............................................................................................................................................8-1
8.3 Gauge Test ...........................................................................................................................................8-2
8.3.1 High Vacuum Gauges ONLY .............................................................................................................8-2
8.3.2 Standard and Low Vacuum
Gauges ONLY ....................................................................................................................................8-2
8.4 Regulation Test .....................................................................................................................................8-2
8.5 Low Vacuum Regulators ONLY ............................................................................................................8-2
8.6 Bleed Test .............................................................................................................................................8-2
8.7 Leak Test - Supply Side ........................................................................................................................8-3
8.8 Leak Test - Patient Side ........................................................................................................................8-4
9/Maintenance
9.1 General Maintenance of Suction Equipment ........................................................................................ 9-1
9.2 Recommended Maintenance Schedule ................................................................................................9-1
9.2.1 Maintenance Schedule .......................................................................................................................9-1
9.3 Repair Policy ......................................................................................................................................... 9-2
9.4 Technical Assistance ............................................................................................................................ 9-2
9.5 Return Instructions ................................................................................................................................9-2
10/Ordering Information
10.1 Illustrated Parts .................................................................................................................................10-1
10.2 Service Kits .......................................................................................................................................10-2
10.3 Fittings and Adapters ........................................................................................................................10-2
10.4 North American .................................................................................................................................10-3
10.4.1 Regulator Options ..........................................................................................................................10-3
10.4.2 Other Options .................................................................................................................................10-4
10.5 International ......................................................................................................................................10-5
10.5.1 Regulator Options ..........................................................................................................................10-5
10.5.2 Other Options .................................................................................................................................10-6
Appendix
Installation procedure for Adapters/Probes and Fittings. ........................................................................... A-1
ii
1/Precautions
1.1 Definitions
Note: A Note provides additional information to clarify a point in the text.
Important: An Important statement is similar to a note but of greater emphasis.
CAUTION: A CAUTION statement is used when the possibility of damage to the equipment exists.
WARNING: A WARNING statement is used when the possibility of injury to the patient or the operator exists.
w = Attention. Alerts you to a warning or caution in the text.
MAX = maximum
High Flow
High Vacuum = high flow, high vacuum
High Flow
Low Vacuum = high flow, low vacuum
I (On) = on
O (Off) = off
1.2 Warnings
After patient use, regulators may be contaminated. Handle in accordance with your hospital’s infection
control policy.
Clean and sterilize all suction equipment before shipment or service to ensure transportation personnel
and/or service personnel are not exposed to any hazardous contamination.
Connect the Vacuum Regulator to the vacuum source only and check its operation before attaching the
patient connection. Connection to pressure sources even momentarily could injure the patient or operator
and damage the equipment.
Do not use this device in the presence of flammable anesthetics. Static charges may not dissipate and a
possible explosion hazard exists in the presence of these agents.
If the Vacuum Regulator is repaired or disassembled in any manner, the Service Checkout Procedure
must be performed before using the equipment on a patient.
The patient port of the regulator must be occluded when setting the prescribed suction level so that the
patient does not receive higher than required suction levels.
The Low Vacuum Limit Setting Procedure must be followed when repairing Low Vacuum Regulators.
Failure to do this may result in suction exceeding the prescribed level.
The Pre-Use Checkout Procedure must be performed before using this equipment on each patient. If the
regulator fails any part of the Pre-Use Checkout Procedure, it must be removed from service and repaired
by qualified service personnel.
Do not use any Loctite® products to seal the fitting and adapter port threads (or products which contain
Methacrylate Ester as an active ingredient).
Only competent individuals trained in the repair of this equipment should attempt to service it.
To help prevent aspirate from entering the regulator, an Overflow Safety Trap should be attached prior to
its use. Aspirate in the regulator will impair the operation. The use of the Overflow Safety Trap will help
prevent this and extend the life of the suction equipment.
Use of lubricants other than recommended may degrade plastic or rubber components.
Prior to placing the unit back into service after repair or dissassembly, perform the Service Check-out
Procedure.
MAX
ME D
MED MED
400
100 300
100 80 120
HI
100 80 120 500
GH
80 120 60 Low Vacuum
-mmHg 60
-mmHg
140 140 200 -mm Hg
60 140
LOW
LOW
LOW
40 mm Hg 600
HIGH
40
HIGH
40 160
160 160
20 100
20 20 700
180 180
200 200
F U L L VA C F U L L VA C
Standard Three Mode Standard Two Mode Low Continuous High Three Mode
Continuous Continuous 6700-1227-900 Continuous
6700-1224-900 6700-1225-900 0-135 mm Hg 6700-1226-900
0-200 mm Hg-FULL VAC 0-200 mm Hg-FULL VAC 0-760 mm Hg
400 100
300 400 80
300 120
500
50 500 60
60 40 50 14 12
60 40 140 16 10
70 30 200 8
High Flow 70 30 200 18
High Vacuum High Flow 6 40
600 80 20 High Vacuum 20
600 80 20 4
100
90 10 100 2 20
90 10
700 100 100
700
Pa
- kP
a a -k
Hg
- kP
High Flow m
Hg Hg Low Vacuum -m
-m m -m m
Three Mode High Continuous Two Mode High Continuous Two Mode Low Continuous
6700-1229-900 6700-1230-900 6700-1233-900
CCW CCW CCW
0-760 mm Hg/0-100 kPa 0-760 mm Hg/0-100 kPa 0-135 mm Hg/0-18 kPa
Mode
Selector
Switch
400
300
500
50
60 40
70 30 200
High Flow
Vacuum 600 80
High Vacuum
20
Gauge 90
100
10
700 100
a
- kP
Hg
-m m
Adapter
Port
Suction
Control Knob
Positive Pressure Safety Located in the vacuum supply line to prevent pressurization
Relief Valve: of the patient connection by failed injector vacuum (venturi) units, or
inadvertent cross connection to pressurized gases
Weight
Low Vacuum models: 12 oz/340 grams
(Less fittings)
WARNING w Connect the vacuum regulator to the vacuum source only and check its operation before
attaching the patient connection. Connection to pressure sources, even momentarily, could
injure the patient or operator and damage the equipment.
Vacuum Port
Patient Port
Vacuum 400
300
500
Regulator 60
50
40
200
70 30
High Flow
600 High Vacuum
80 20
100
90 10
700 100
a
- kP
Hg
-mm
Disposable
Suction Filter
Connective
Tubing
Overflow
Safety Trap
Collection
Bottle
Underwater Seal Line 2 cm H2O
Made in USA
Equipment Set-up
Trap fitting
1. Raise the sleeve and insert the trap into the regulator fitting.
2. Turn the trap clockwise about one and a half turns to engage the threads. The trap does not need to be screwed
tight; an O-ring in the regulator fitting provides a vacuum seal. The trap should rotate freely to allow the desired
tubing positioning.
3. Lower sleeve to lock trap in position.
Regulator
Sleeve
DISS fitting
1. Insert trap into the regulator fitting. Situate the tubing in the desired position.
2. Turn the DISS wing nut clockwise to engage threads and tighten (there is no O-ring, so the vacuum seal
depends on a tight
connection).
Regulator
Wing nut
2 - Mode Switch
3 - Mode Switch
I/(On)
2 - Mode Switch
3 - Mode Switch
O/(Off)
3 - Mode Switch
MAX
400
300
1. Turn the mode selector switch to I (On). 500
60
50
40
70 30 200
100
10
Hg
-m m
Clamp
60
of the Pre-Use Checkout Procedure, it 70
High Flow
40
30 200
100
10
a
- kP
Hg
w Connect the vacuum regulator to the vacuum -m m
700 100
a
- kP
Hg
-m m
Clamp
3. Clamp tubing.
Regulator Setting 400
300
Standard & Low: increase the suction to 100 mm Hg (13 kPa) 500
60
50
40
200
High: increase the suction to 300 mm Hg (40 kPa) 600 80
70
High Flow
High Vacuum
30
20
100
90 10
Slowly open and close the clamped tubing to create various 700 100
- kP
a
flow rates through the regulator. Check that the suction level is -m m
Hg
Clamp
Release
Clamp
suction. 600 80
70
High Flow
High Vacuum
30
20
200
100
90
Turn the mode selector to I (On). 700 100
10
a
- kP
Hg
-m m
Clamp
12
60
14
140 16 10
18 8
6 40
20
4
2 20
Pa
-k
High Flow m
Low Vacuum -m
Clamp
400
300
500
50
60 40
70 30 200
High Flow
600 High Vacuum
80 20
100
90 10
700 100
a
- kP
Hg
-m m
700 100
a
- kP
Clamp
Made in USA
700 100
a
- kP
H g
-mm
Made in USA
20
200
100
90 10
700 100
a
- kP
Hg
-m m
Made in USA
5.1 Cleaning
wWARNING After patient use, regulators may be contaminated. Handle in accordance with
your hospital’s infection control policy.
5.1.1 Routine Exterior Cleaning
Routine cleaning of the regulator is recommended as a standard procedure after each use.
Wipe all exterior surfaces with a solution of water and mild detergent.
5.1.2 Internal Component Cleaning
CAUTION w Cleaning the gauge may result in damage.
The regulator requires cleaning if it becomes flooded with patient fluid as a result of misuse.
1. Refer to the Service - Disassembly and Assembly instructions.
2. All internal components, with the exception of the gauge, may be cleaned with a solution of warm water
and mild detergent, preferably an enzymatic cleaner.
3. Dry all components with a lint free cloth before assembly.
5.2 Sterilization
WARNING w Following sterilization with ethylene oxide, parts should be quarantined in a well
ventilated area to allow dissipation of residual ethylene oxide gas absorbed by the
material. Follow sterilizer manufacturer’s recommendations for specific aeration
periods required.
CAUTION w Do not steam autoclave or liquid sterilize the Continuous Vacuum Regulator. Severe
impairment to the operation of the regulator will result. The only acceptable method of
sterilization is with gas (ethylene oxide).
CAUTION w Sterilization with ethylene oxide mixtures may cause crazing (minute superficial crack-
ing) of some plastic parts. Crazing will be more pronounced when mixtures containing
Freon® are used.
Note: The Vacuum Regulator should only be sterilized if it is contaminated or maintenance is to be per-
formed.
1. The regulator should be sterilized with the mode selector switch in the | (On) position.
2. The only acceptable method of sterilization is with ethylene oxide. Ethylene oxide mixtures can be used
at temperatures of 125 to 135 °F (52 - 57 °C). If this temperature cannot be obtained, room temperature
sterilization with 100% ethylene oxide can also be used. Sterilization is not recommended as a standard
procedure after each use.
3. After each sterilization, check the condition of the internal filter. If the filter appears to have shrunk,
replace it before placing the regulator back in service.
4. Adequately aerate the regulator prior to disassembly, shipment or use. Aerate parts prior to reassembly.
B. No gauge indication but suction 1. Blocked gauge pressure 1. Clean or replace gauge
is being delivered sensing orifice
2. Gauge mechanism locked by 2. Clean or replace gauge
debris
3. Gauge mechanism damaged 3. Replace gauge
E. Insufficient flow through 1. Partial blockage in wall supply 1a. Confirm wall supply
regulator open air flow meets
minimum hospital
requirements
1b. Unblock if necessary
2. Partial blockage in regulator 2a. Perform the Flow Test
in the Service Check-
out Procedure Section
of this manual
2b. Clear blockage if
necessary
G. With the fitting port occluded, 1. Blocked filter and/or orifice 1. Clear orifice and replace
unable to decrease the filter
suction level and gauge
needle does not return to 2. Damaged gauge 2. Replace gauge
zero when switched to
O (Off)
H. No suction in any setting and 1. Positive pressure safety relief 1. Replace positive pressure
whistling noise from inside valve failure safety relief valve
the regulator
2. Vacuum relief valve failure 2. Replace o-ring and ensure
(low models ONLY) that the steel ball is present
and clean
2
1. Cover Screw (4)
2. Cover Assembly
3. Low Vacuum Limiting Screws
(2) (Not Shown) (On old versions
of the Low Models Only)
12
11
1
1. Mode Selector Switch
2. Switch Plate 2
3. Screw
4. O-ring, Vacuum Relief Valve 3 5
(Low Models Only) 10
5. Vacuum Relief Valve & O-ring 4
(Low Models Only) 13
14
6. Gauge Assembly (includes 8
gauge, o-rings, and lens) 9
7. Lens 7
8. O-ring (2), Gauge
9. Regulator Module
10. Positive Pressure Safety Relief Valve
11. Filter (Internal) 6
12. Base
13. O-ring, Stem
14. O-ring, Regulator Module
1. Remove the four cover screws from the back of the regulator.
2. Carefully pull the cover assembly off the back body.
3. To remove the gauge assembly, grasp the assembly and
pull straight out. The snap-fit lens can also be removed
(if applicable) for replacement.
4. To remove the regulator module from the cover assembly,
rotate the suction control knob counterclockwise until
the regulator module is free.
5. Using tweezers, remove the filter. Replace with a new filter.
6. Grasp the positive pressure safety relief valve, and pull
Pull down plastic tab
it from the back body. with right index finger
7. Remove the mounting screw from the switch plate while holding gauge
assembly and remove the switch plate and mode switch.
8. Low models ONLY: Grasp the vacuum relief valve
and pull it from the back body. Snap Fit Lens Removal
7.4.2 Assembly
1. Return the diaphragm to its original position (convoluted), with the molded number facing the retainer.
2. Assemble the diaphragm onto the retainer by lining up their center holes as shown in the diagram.
3. Insert the piston into the diaphragm and ensure that the retainer mates with the recess in the piston.
4. Insert the stem through the piston.
5. Holding the retainer/diaphragm/piston/stem as an assembly with your fingers, insert the stem screw
and tighten.
6. Place the spring in the housing.
7. Insert the retainer/diaphragm/piston/stem assembly through the spring and into the cavity in the
housing. Fit the diaphragm bead into the groove in the housing.
8. Install the cap and the two cap screws.
9. Apply a small amount of Dow® 111 Grease to the o-rings only.
10. Install o-rings onto the stem and housing.
Retainer
Stem
Stem Screw
Piston
Diaphragm
Diaphragm Assembly
8.1 Set-up
1. Verify that there is 500 mmHg ± 10 mmHg (67 kPa ± 1.3 kPa) vacuum on the supply gauge.
2. The supply open flow must be 50 LPM minimum.
3. Connect the supply vacuum to the adapter port.
MED
100
80 120
60
mmHg
140 Supply Vacuum
LOW
40
160
20
180 Supply Vacuum Reg. set to Minimum and
200 500 mm Hg ± 10 mm Hg Open Flow of
F U L L VA C (64 kPa ± 1.3 kPa) 50 LPM Minimum
Occluded
Port A
Port B
Sealed
Fitting Port
1/4" I.D.
Clear Tubing
1.0 In.
Water
Bubble Leak
Tester
Mode
Selector
Switch
MED Port B
100
80 120
mmHg
60 140
LOW
40
160
20 Clear Tubing
Vacuum 180
1.0 In.
Suction Water
Control Knob
Bubble Leak
Tester
For all Repairs Contact your nearest Ohmeda Medical office or authorized Ohmeda Medical Distributor.
* If current unit does not have limit screws inside the cover behind the
control knob and the relief valve has 4 vent holes, the Std/High part can
be purchased.
12
11
1
3 5
10
4
13
14
8
9
7
Regulator Module
Replacement Kit .............................. 6700-0030-700
Includes the following parts:
O-ring, Stem These parts are available
O-ring only in kit form. If other
Diaphragm parts are needed for re-
Stem Screw placement, the entire
Cap Screw (3) Module must be ordered
(6700-1225-800).
6
10 11 12 13 14 15
7
LOW
Female
HIGH
40
160
20
200
180
901
F U L L VA C
Tubing Nipple
6701
Ohmeda
Diamond 902
Tubing Nipple and
1224 1/2 Gallon Bottle,
6702 Standard Three with Wall Slide
Ohmeda Mode Continu- (Straight)
Hand-I-Twist ous
903
6703 Locking Gland and
DISS Nut & 80
MED
100
120
1/2 Gallon Bottle with
Gland 60
-mmHg
140
Floor Bracket
LOW
HIGH
40
160
20
180
200
F U L L VA C
6704 904
Schrader Locking Gland
1225
905
6705 Standard Two Locking Gland and
NCG Mode Overflow Safety Trap
Rectangle Continuous
MAX 906
Locking Gland and
6706 300
400
-mm Hg
500
Overflow Safety Trap
200
700
and 1/2 Gallon Bottle,
with Wall Slide (Straight)
907
6707 DISS Male
Puritan-
Bennett 1226
High Continuous 908
DISS Male & Overflow
6708 Safety Trap, DISS Wing
OES Nut & Gland
ME D
100
80 120
HI
GH
60 Low Vacuum
140
LOW
40 mm Hg
160
20
6709 909
MedStar DISS Male, Overflow
Safety Trap and 1/2
Gallon Bottle, with Wall
6711 Slide (Straight)
90° Tubing
1227
Nipple
Low Continuous
MED
100
80 120
-mmHg
60 140
LOW
HIGH
40
20
160
180
MPTS, Regulator, Regulator, Bracket,
F U L L VA C
200
MPTS Slide and 1/4 Gallon Bottle,
Brackets (2) & 45° Locking Gland
Retaining Clips (2), 6711-XXXX-916
Overflow Safety
Trap
6711-XXXX-914
1224
Standard Three Mode
Continuous
Regulator and
Regulator with 30°
MED
80
100
120
Dovetail Bracket,
Fitting, Overflow Overflow Safety
-mmHg
60 140
LOW
HIGH
40
160
20
180 Safety Trap Trap
200
F U L L VA C 6711-XXXX-930 6711-XXXX-935
1225
Standard Two Mode
Continuous Regulator & MPTS
(Slide) Bracket, Regulator and
Overflow Safety Locking Gland
Trap (Threaded)
MAX
6711-XXXX-913 6711-XXXX-919
400
300
500
200 -mm Hg
600
100
700
100
80 120
HI
GH
60 Low Vacuum
140
LOW
40 mm Hg
160
20
6700-12XX-900 0221-2685-802
500
400
300 w/o Inlet Fittings Diamond
50
60 40
70 30 200
High Flow
600 High Vacuum
80 20
100
90 10
700 100
- kP
a
-m m
Hg 6700-12XX-901 0204-6717-800
w/Tubing Nipple Hand-I-Twist
6700-12XX-902 0221-0205-800
6700-1229-9XX w/Tubing Nipple, NCG (Rectangle)
Three Mode 12" Tubing,
High Continuous 1/2 gal. Bottle,
(Anti-clockwise) Cap & Float 0221-0204-802
Assembly and Wall NCG (Round)
Slide (Straight)
400
0221-0106-300
300
500
50
60 40
70 30 200
High Flow
600 80
90
High Vacuum
20
10
100
6700-12XX-904 Puritan
700 100
- kP
a w/Locking Gland
Hg
-m m
0221-0163-300
6700-12XX-905 MedStar
w/Locking Gland and
Overflow Safety Trap
6700-1230-9XX 0204-9020-535
Two Mode Tubing Nipple
High Continuous 6700-12XX-906
(Anti-clockwise) w/Locking Gland,
Overflow Safety Trap, 0221-0152-300
12" Tubing, 1/2 gal. OES
Bottle, Cap & Float
500
400
60
50
40
300
Assembly and Wall Slide
600 80
70
High Flow
High Vacuum
30
20
200
(Straight) 0221-0690-731
700
90
100
a
10
100
Schrader
- kP
Hg
-m m
6700-12XX-911
w/Locking Gland, 6700-0094-700
Overflow Safety Trap, DISS Nut &
12"Tubing, 1/2 gal. Gland
6700-1231-9XX Bottle, Cap & Float
Three Mode Assembly and Wall
High Continuous "V" Slide 6700-0095-700
(Anti-clockwise) (International) 90° Tubing
Nipple
100
6700-12XX-912
140
120
16
14 12
10
80
60 w/Tubing Nipple,
12" Tubing, 1/2 gal.
18 8
6 40
20
4
-k
Hg
High Flow
Low Vacuum -m
m
Assembly and Wall
"V" Slide
(International)
6700-1233-9XX
Two Mode
Low Continuous
(Anti-clockwise)
400
300
500
50
600 80
70
High Flow
High Vacuum
30
20
200
Regulator and
700
90
100
10
100
Fitting, Overflow Dovetail Bracket,
- kP
-m m
a
Hg
Safety Trap Overflow Safety
6711-XXXX-930 Trap
6711-XXXX-935
1229
Three Mode
High Continuous
(Anti-Clockwise)
Regulator & MPTS
(Slide) Bracket,
400
Overflow Safety
500
60
50
40
300
Trap
200
600 80
70
High Flow
High Vacuum
30
20 6711-XXXX-913
100
90 10
700 100
a
- kP
Hg
-m m
1230
Two Mode
High Continuous
(Anti-Clockwise)
100
80
120
12
60
14
140 16 10
18 8
6 40
20
4
2 20
Pa
-k
Hg
High Flow m
Low Vacuum -m
1233
Two Mode
Low Continuous
(Anti-Clockwise)
10-6
10-6 6700-0072-000 11/30/00 10-6
Appendix
Installation procedure for Adapters/Probes and Fittings.
All adapters/probes and fittings should be sealed and installed properly to prevent leaks and to support the
equipment when mounted. Both vacuum regulator ports are 1/8-27 NPTF tapered pipe threads. It is
important to note that adapters/probes and fittings seal on the thread and may have threads exposed after
they have been tightened properly.
Prior to installing the adapter/probe or fitting, seal the thread with Teflon® (PTFE) tape or one of the follow-
ing lubricants:
Dow® 111 (Ohmeda P/N 6700-0074-200)
Ball Vac Kote (37951M) (Ohmeda P/N 0220-0091-300)
CAUTION w Do not use any Loctite products to seal the threads (or products which contain Meth-
acrylate Ester as an active ingredient).
The torque range for installing adapters/probes and fittings is 4.0 ft-lb (5.4 N-m) minimum to 10.0 ft-lb (13.6
N-m) maximum.
Adapters/probes and fittings which are not keyed for specific orientation, should be torqued to approxi-
mately 6.0 ft-lb (8.1 N-m).
Adapters/probes and fittings that are keyed to specific orientation, must be torqued initially to 4.0 ft-lbs.
Additional torque is applied only until orientation is correct.
®
Teflon is a registered trademark of the DuPont Company.
The addresses listed on this cover are current as of 6/01 please visit our
website www.ohmedamedical.com for any location changes.